PCWI DC15, DC30 Operator's Handbook Manual

Can be used in accordance with:
Australian Standard AS3894.1
ASTM G62, ASTM D5162, ASTM D4787, NACE SP0274,
NACE SP0490, NACE SP0188, ANSI/AWWA C213,
ANSI/AWWA C214, ISO 2746
Operator’s Handbook
For
Compact
High Voltage
DC15 & DC30
POROSITY (HOLIDAY) DETECTORS
®
POROSITY DETECTOR ACCURACY:
+5% of reading or + 50V below 1kV (Note: at the time PCWI tested the Porosity Detector at the reference temperature & humidity).
Refer to the PCWI Manufacturer’s Calibration Certificate provided with the Porosity Detector.
Range: 0kV to 30kV Recertification available.
1
CONTENTS
1.0 Safety Precautions 4
2.0 Operation 5-6
3.0 Specifications 6
4.0 Control Panel Layout 7
5.0 Voltage Recommendations 8-9
6.0 Troubleshooting 10-11
7.0 Optional Accessories 12-13
8.0 PCWI NiMH Battery & Charger Clip 14-15
9.0 Warranty 16
10.0 Service & Maintenance 17
WARNING
This Equipment generates a high voltage and should be used with care. Please consult these operating instructions before use.
2
INTRODUCTION
Thank you for choosing the PCWI Compact for pre-installation and post installation corrosion detection.
PCWI have designed this instrument with care, to provide ongoing corrosion detection efficiency under a wide variety of coating application conditions and to ascertain Porosity for the many protective coatings currently in use. Under reasonable care in operation, the unit will provide many years of trouble free detection.
To support the unit, PCWI maintain a comprehensive range of Electrodes (probes) - extending the versatility of the Compact from large to small and from accessible to inaccessible surfaces.
PCWI, in a continuing desire to achieve the maximum in corrosion detection competence, welcomes user enquiries and recommendations.
Yours sincerely
Paul Van Gaal
3
1.0 SAFETY PRECAUTIONS:
All hand-held high voltage test equipment should be operated by
responsible, trained and authorised personnel.
The unit must be earthed to both the item under test and to ground.
CAUTION
The Detector output can be up to 35,000 volts. Should the operator accidentally make contact with the test electrode, they may experience a mild shock or zap, and in order to avoid this
possibility, the wearing of rubber gloves is recommended. Furthermore, the operator should enjoy good health and not suffer from
a cardiac condition. If the operator has a pacemaker, then they should
not use this equipment.
This equipment should only be used for the purpose for which it was designed, ie: checking the porosity, or electrical breakdown, of dielectric or insulating materials.
It is also recommended that testing should be carried out well clear of personnel not involved in the testing procedure, or in such a position whereby the surprise of receiving an electric shock could cause a related accident, if for example, tests being carried out close to moving or rotating machinery, or in such an unstable position that the operator could fall and injure themselves.
It is recommended that the operator should have an assistant, to ensure that unauthorised personnel are kept well clear of the testing area, and generally assist when necessary with the testing procedure. It is also recommended that the Detector not be operated within close proximity of sensitive electronic apparatus, such as computer equipment.
DANGER
Do not use the test equipment in any combustible or flammable atmosphere, as a test voltage can cause an arc or spark to be generated and an explosion could result.
Always consult the plant or safety officer before carrying out a test procedure.
When testing tank internals, be certain the tank does not contain solvents remaining from the painting procedure.
4
Coating Thickness Range
Applied coats should be cured, thickness tested, visually inspected and accepted – before high voltage porosity testing is carried out. Coating thickness should be above 150µm; coatings below this thickness
should be tested with a wet sponge unit.
2.0 OPERATION
Connect the probe and earth leads to the unit. Connect the earth clamp to the metallic substrate of the item to be tested – substrate should be earthed to ground. Select the probe best suited for the test and attach to the probe handle. Fit the Fuse (if not already fitted). Turn the unit on. Check the batteries to ensure that they are charged. Turn the voltage control clockwise to the required test voltage.
Note: If the neon is in place allow an additional 200v.
Place the probe near the metal substrate. A spark should occur (if not re-check all leads and connections). Place the probe on the surface to be tested. Re-check the output – adjust if necessary. Turn alarm sensitivity until the alarm is just off. Place the probe on the coated surface and move at approximately one metre per four seconds.
A fault is indicated by: A spark at the probe – this can usually be seen and heard. A light flashes on the front panel of the unit. An audible sound – buzzer is mounted inside the unit. Digital test voltage drops drastically. The neon in the probe flashes.
Note: A definite flaw should be made in the coating and located with the designated test voltage, therefore proving that the unit is locating the type of fault you wish to find.
Probes must be kept in full contact with the surface, gaps in or between the probe and the coating may result in flaws going undetected.
Wire brushes, rubber and coil spring probes should all be kept in good condition. Probes other than fine wire brushes may require higher voltages.
Wet and contaminated coatings should not be tested until dry and clean (see also troubleshooting pages).
5
Loading...
+ 14 hidden pages