PCWI Compact P20, Compact P40, Compact P60 Operator's Handbook Manual

®
PRECISION INSTRUMENTATION
Operator’s Handbook
for
Compact
P20, P40 & P60
POROSITY (HOLIDAY) DETECTORS
Complies with the requirements of:
Australian Standard AS3894.1-2002,
ASTM G62-87(1998), NACE RP0274-98,
NACE RP0490-2001, NACE RPO188-99, ASTM D4787-93(1999),
JIS G-3491, JIS G-3492, ANSI/AWWA C214-89,
ANSI/AWWA C213-94 and ISO 2746:1998
CE Marked PCWI Compact Detectors comply with the requirements of EMC Directives 89/336/EEC EMC and its amending directives.
CERTIFICATE OF CALIBRATION
Model : Compact Detector Manufacturer: PCWI Technology Pty Ltd Serial No:. ……………….. Tolerance:. ±5%
Offi e:
Postal:
Email: Web: Phone: Intl: Fax:
c
13 Alhambra Avenue CARDIFF NSW 2285 Newcastle - Australia
PO Box 900 HUNTER REGION MC NSW 2310
@
sales pcwi.com.au www.pcwi.com.au
(02) 4954 3900
+61 2 4954 3900 (02) 4954 3999
Conformance: This detector has been tested at a series of voltages from
0 to 60,000 Volts against a Certified Meter and High Voltage Probe, Report Number 209852 / Pulse Crest Meter Report Numbers 01-01851-A, 01-01851-B, 01-01851-C, 01-04258-A, 54019, 54819 and 723 0135/03, using AS 2886-1986 as a guide. The Detector complies with PCWI Standards of Manufacture.
PCWI’s Laboratory complies with the requirements of AS ISO/IEC 17025-1999 ‘general requirements for the competence of testing and calibration laboratories’. All calibrations are traceable to National and International Standards.
Note: It is recommended that the unit be periodically checked
prior to use with a Certified PCWI Crestmeter.
SIGNED: ___________________
Laboratory Technician
PCWI Technology Pty Ltd ABN 96 0 783 373 54
Quality System
Quality Endorsed Company
ISO9001
Lic 13066
CONTENTS
1.0 Safety Precautions...................................................................2
2.0 Operation...............................................................................3
3.0 Specifications .........................................................................4
4.0 Control Panel Layout................................................................5
5.0 Voltage Recommendations....................................................6-7
6.0 T roubleshooting...................................................................8-9
7 .0 Optional Accessories.........................................................10-11
8.0 Warranty..............................................................................12
9.0 Service.............................................................inside back page
INTRODUCTION
Thank you for choosing the PCWI Compact for pre-installation and post installation corrosion detection.
PCWI have designed this instrument with care, to provide ongoing corrosion detection efficiency under a wide variety of coating application conditions and to ascertain Porosity for the many protective coatings currently in use. Under reasonable care in operation, the unit will provide many years of trouble free detection.
To support the unit, PCWI maintain a comprehensive range of Electrodes (probes) - extending the versatility of the Compact from large to small and from accessible to inaccessible surfaces.
PCWI, in a continuing desire to achieve the maximum in corrosion detection competence, welcomes user enquiries and recommendations.
Y ours sincerely
Paul V an Gaal
WARNING
Unlike continuous direct current units where the electrical field drops to zero when contact is made and where the probe is earthed out via the body, Pulsed Detectors will continue to shock at full voltage. Caution should be taken to avoid contact with the live electrode.
Please consult these operating instructions before use.
1
1.0 SAFETY PRECAUTIONS:
All hand-held high voltage test equipment should be operated by responsible, trained and authorised personnel.
The unit must be earthed to both the item under test and to ground.
CAUTION
The Detector output can be up to 65,000 volts. Should the operator accidentally make contact with the test electrode, they may experience a mild shock or zap, and in order to avoid this possibility , the wearing of rubber gloves is recommended.
Furthermore, the operator should enjoy good health and not suffer from a cardiac condition. If the operator has a
pacemaker , then they should not use this equipment.
This equipment should only be used for the purpose for which it was designed, ie: checking the porosity, or electrical breakdown, of dielectric or insulating materials.
It is also recommended that testing should be carried out well clear of personnel not involved in the testing procedure, or in such a position whereby the surprise of receiving an electric shock could cause a related accident, if for example, tests being carried out close to moving or rotating machinery, or in such an unstable position that the operator could fall and injure themselves.
DANGER
2
It is recommended that the operator should have an assistant, to ensure that unauthorised personnel are kept well clear of the testing area, and generally assist when necessary with the testing procedure. It is also recommended that the Detector not be operated within close proximity of sensitive electronic apparatus, such as computer equipment.
Do not use the test equipment in any combustible or flammable atmosphere, as a test voltage can cause an arc or spark to be generated and an explosion could result.
Always consult the plant or safety officer before carrying out a test procedure.
When testing tank internals, be certain the tank does not contain solvents remaining from the painting procedure.
Coating Thickness Range
Applied coats should be cured, thickness tested, visually inspected and accepted – before high voltage porosity testing is carried out. Coating thickness should be above 150µm; coatings below this thickness should be tested with a wet sponge unit.
2.0 OPERATION
Connect the probe and earth leads to the unit. Connect the earth clamp to the metallic substrate of the item to be tested – substrate should be earthed to ground. Select the probe best suited for the test and attach to the probe handle. Fit the Fuse (if not already fitted). Turn the unit on. Test the batteries to ensure that they are charged. Adjust the voltage control to the required test voltage. Place the probe near the metal substrate. A spark should occur (if not re-check all leads and connections). The unit should now be ready for use. Re-check the output – adjust if necessary. Place the probe on the coated surface and move at approximately one metre per four seconds.
A fault is indicated by:
A spark at the probe – this can usually be seen and heard. A light flashes on the front panel of the unit. An audible sound – buzzer is mounted inside the unit.
Note: A definite flaw should be made in the coating and located with the
designated test voltage, therefore proving that the unit is locating the type of fault you wish to find. Probes must be kept in full contact with the surface, gaps in or between the probe and the coating may result in flaws going undetected. Wire brushes, rubber and coil spring probes should all be kept in good condition. Probes other than fine wire brushes may require higher voltages.
3
Loading...
+ 11 hidden pages