PCE Instruments VL7-100B-d15, VL7-100B-d20, VL7-100B-d21 Users guide

Page 1
Page 2
Viscolite VL7
Guarantee
Hydramotion Ltd will repair a Viscolite portable viscometer free of charge within 24 months of the date of purchase if satised that the fault is the result of defective materials and/or workmanship and if the instrument is returned to Hydramotion Ltd carriage prepaid and undamaged in transit.
This Guarantee shall not proper care by the owner or user, (ii) in this User Manual, or (iii ) normal wear and tear on the instrument .
Hydramotion Ltd shall in no case be liable for any loss of output, revenue or any other losses or costs, consequential or otherwise, howsoever incurred.
apply to any fault resulting from (i) negligence or lack of
a failure to follow the recommendations set out
© 2013 Hydramotion Ltd
The Vis colite and all it s associated t echnologie s are protected b y UK and Interna tional Patent Ap plications .
2
U01-04-013 R1 08/13
Page 3
Viscolite VL7
Viscolite VL7
portable viscometer
Contents
Introduction 5
1 Getting started
1.1 Packaging 7
1.2 Specification 7
1.3 Fitting the batteries 9
2 Using the Viscolite
2.1 Turning on 10
2.2 Operation check 10
2.3 Measuring viscosity 11
2.4 Measuring temperature 12
2.5 Temperature-corrected viscosity 12
2.6 General maintenance 13
2.7 Guidelines for optimum performance 13
3 Temperature correction
3.1 Introduction 14
3.2 E-format numbers 14
3.3 How to enter an E-format parameter 15
3.4 Setting the reference temperature 16
3.5 Saving changes 17
3.6 Determination of correction factors 18
4 Operational adjustment
4.1 Introduction 20
4.2 How to modify a numerical value 20
4.3 Accessing calibration and adjustment menus 21
U01-04-013 R1 08/13
3
Page 4
Viscolite VL7
4.4 Minor null adjustment 22
4.5 Stabilise reading (averaging filter) 22
4.6 Scale reading by any desired factor 23
4.7 Take account of f luid density 24
4.8 How to display viscosity in centistokes 24
4.9 Matching to reference instrument 25
5 Calibration and Maintenance
5.1 Introduction 26
5.2 Calibration check 26
5.3 Full recalibration 27
5.4 Instrument Verification Service 27
5.5 Service and repair 28
Appendix 1: Operating principles 29
Appendix 2: Menu overview 31
Troubleshooting/Alarms 33
Index 34
4
U01-04-013 R1 08/13
Page 5
Viscolite VL7
Introduction
The Viscolite® VL7 is an easy-to-use portable
instrument for the instant measurement of viscosity
anywhere at any time using a simple dip-in probe. No
prior setting-up or adjustment is required.
U01-04-013 R1 08/13
The sensor is a solid probe with no moving parts, seals
or bearings. It is connected by a coiled f lexible cable
to a display unit powered by four AA batteries. The
probe also incorporates an integral platinum resistance
thermometer (PRT). The whole system is supplied in a
compact carrying case for ease of transport.
The Viscolite can be used with any volume of fluid,
from a cupful to a tank. Readings are made instantly,
enabling rapid assessment of samples or batches in situ.
The solid sensor can be wiped clean in moments.
This combination of toughness, flexibility and
portability make the Viscolite the ideal instrument for
the spot measurement of viscosity at a remote site, in
the laboratory or in a busy factory.
5
Page 6
Viscolite VL7
How it works
The Viscolite is a “resonant” or “vibrational” viscometer. The sensor consists of a steel
shaft with an end mass (“bob”) which is made to vibrate at its natural frequency. The
actual vibrational movement is very small, but it is enough. As the vibrating sensor
shears through a fluid, it loses energy as a result of the viscous drag exerted by the f luid.
The viscosity of the fluid can therefore be determined by measuring the amount of
energy loss. This technique, combined with high-performance transducer architecture,
leads to extremely high repeatability.
Since a Hydramotion viscometer produces no compression or sound waves it can be used
in any vessel, regardless of size. The smooth, monolithic design of the sensor eliminates
any chance of solid or bubble entrapment. And as the shear rate is high, the effect of
external factors such as vibration or changes in fluid velocity will be negligible.
For more details of how the energy loss is converted into a measurement of viscosity, see
Appendix 1 (p. 29).
6
U01-04-013 R1 08/13
Page 7
Viscolite VL7
1.1 Packaging1.1 Packaging
First, inspect the contents of the carrying First, inspect the contents of the carrying
Please do not discard the Please do not discard the
Keep the Viscolite in its carrying case Keep the Viscolite in its carrying case
The metal probe can move slightly relative The metal probe can move slightly relative
1 Getting started
1.1 Packaging
First, inspect the contents of the carrying
case. Immediately report any apparent
shipping damage to Hydramotion Ltd or its
representatives as well as to the carrier.
Please do not discard the packaging. Use
it if you need to forward the instrument to
another recipient or return it to the factory.
Keep the Viscolite in its carrying case
whenever it is not in use.
The metal probe can move slightly relative
to the black acetal enclosure: this is quite
normal.
• For correct operation it is essential that the sensor shaft is perfectly straight. Do not
leave the transducer resting on the sensor shaft.
1.2 Specification
Viscosity range
model type VL7-100B-d15 0 – 5,000 centipoise (cP) model type VL7-100B-d20 0 – 1,000 cP model type VL7-100B-d21 0 – 10,000 cP
Temperature range –20 to +120 °C
Materials
sensor 316L stainless steel (unless otherwise specified) enclosure acetal polymer
Power source 4 × 1.5V alkaline batteries size AA
Options
POL high polished sensor surface 0.4 micron AA
U01-04-013 R1 08/13
7
Page 8
Viscolite VL7
Immersion depths: B viscosity or temperature A viscosity only
All dimensions to nea rest mm
flexible coil lead
display unit
Hydramotion
show units
CAL
inc
dec
viscosity
ON
OFF
iscolite 700
portable viscometer
VL7-100 B- d15
Figure 1
Viscolite VL7
Dimensions (mm):
display unit L 180 W 95 H 40 probe (overall) L 215 (d15 model), L 305 (d20/d21 models)
sensor probe
B
15
62
35
A
15
VL7-100B-d20 & -d21
30
115
sensor probe
B
22
A
20 (type d20)
21 (ty pe d21)
• The full designation of an instrument includes the model type and any applicable
Option. Thus “VL7-100B-d15-POL” indicates the d15 model with a high-polish finish.
• A Certificate of Calibration and Conformity is included with each Viscolite. It should
be carefully checked against the shipping documents and the serial number stamped on
the instrument, then stored in a safe place.
Record the details of your instrument
Model type: Serial number: Certificate of Calibration and Conformity number:
8
U01-04-013 R1 08/13
Page 9
Viscolite VL7
1.3 Fitting the batteries
• The Viscolite uses four 1.5V batteries, ty pe AA or
equivalent (LR6, AM 3, MN 1500).
TAG
SER.
MALTON YORK ENGLAND
TYPE
MODEL
Figure 2
Underside of display unit
• Locate the battery compartment on the underside
of the display unit (see Figure 3).
• Remove the lid by pressing the part marked
' and
sliding sharply backwards.
• Place the batteries in the battery cradle, taking care
that they are inserted the right way round.
• Clip the cradle firmly to the connector.
• Place the battery cradle in the compartment and
replace the lid.
Note:
• Alkaline batteries (“Duracell” or equivalent) should
give up to 50 hours’ use under normal operating
conditions.
• Rechargeable batteries are not suitable.
Figure 3
Fitting the batteries
U01-04-013 R1 08/13
9
Page 10
Viscolite VL7
2 Measuring viscosity
Figure 4
Display unit
2.2 Operation check
2.1 Turning on
• Press the * key on the display unit and
hold it down for a second or two.
• The display shows a succession of digits
while the microprocessor goes through a
start-up routine. It will then show “VL”
briefly before settling on the “ live viscosity”
reading. This is the dynamic viscosity
in centipoise (cP) of whatever fluid is
surrounding the sensor.
• The reading is updated every 1.5 seconds
approximately.
• Pressing the
other measurements, “
VC” (temperature-corrected viscosity). For
more details, see pages 12 and 13.
• A beeper sounds each time a key is pressed.
• To switch off, press the
) or ( keys will bring up two
t” (temperature) and
* key again.
• The display unit should read 0.0 (zero) so long as the sensor is
(i) perfectly clean, dry and
(ii) wholly in air and not touching any thing.
• The display should show
For details of the checking procedure, see Section 5.2.
• If the air reading is not exactly zero:
(1) check that the sensor is perfectly clean, dry and not in contact with anything.
(2) if necessary, perform a “minor null adjustment” as described in Section 4.4 (p. 22).
10
1.0 (exactly one) if the sensor is immersed in water at 20 °C.
U01-04-013 R1 08/13
Page 11
Viscolite VL7
Minimum immersion depths: B viscosity or temperature A viscosity only
Do not clamp below here
B A
VL7-100B-d20 and VL7-100B-d21 VL7-100 B- d15
Measuring viscosity
Do not clamp below here
B
A
Figure 5
2.3 Measuring viscosity
• Switch on the Viscolite.
• Immerse the sensor into the f luid to be measured to the depth shown in Figure 5.
• If the fluid is likely to be at a different temperature to that of the sensor, be sure to allow
suff icient time for complete temperature equalisation before taking a reading.
• The display unit shows the dynamic viscosity of the fluid in centipoise (cP), equivalent
to millipascal-second (mPa·s).
Instability in the reading may be caused by
(1) temperature differences between the fluid and the sensor,
(2) the presence of suspended solids in the f luid, or
(3) a notable yield stress in a fluid exhibiting plastic behaviour ( Bingham effect).
If the instability is caused by (2) or (3) (suspended solids or plastic behaviour), use the
probe to stir the fluid gently.
U01-04-013 R1 08/13
11
Page 12
Viscolite VL7
• The viscosity reading may be noticeably higher or lower than expected if:
(1) the temperature of the f luid is significantly different from that at which the
reference measurement was made, or
(2) the sensor is not immersed to the depths shown in Figures 1 and 5, or
(3) the fluid is non-New tonian and the reference measurement was made using a
significantly different shear rate*, or
(4) the fluid density is significantly higher or lower than the nominal factory setting of
1.0 g/cm
3
. In this case, enter the correct density value using the procedure described in
Section 4.8 (p. 24).
2.4 Measuring temperature
• Switch on the Viscolite.
• Ensure that the sensor is immersed to the depth shown in Figure 5.
• Press the
) key. The display will show “t” (temperature) brief ly, then the temperature
of the fluid surrounding the sensor.
• Allow time for the sensor to reach the fluid temperature before noting the reading.
• To return to viscosity measurement, press the
( key. The display will show “VL” briefly
before showing the current live viscosity again.
2.5 Temperature-corrected viscosity
• If the display unit is showing live viscosity, press the ) key twice or the ( key once.
The display will show “
• For details of how to set the reference temperature, see Section 3.4 (p. 16).
• The displayed reading is calculated using the measured viscosity and temperature and
applying two correction factors specific to the fluid under test. See section 3 (p. 14) for
details of how to enter these correction factors.
*The Viscolite operates at a very high shear rate, which helps to give it high repeatability. Many laboratory viscometers operate at low shear rates. If the f luid is shear-t hinning (pseudoplastic or t hixotropic), t he Viscolite readi ng will be lower than t hat obtained with a low shear-rate viscometer. If the f luid is shear-thickening (dilatant or rheopect ic), the Viscol ite wi ll give a higher reading. See also Section 4.9 (p. 25).
12
VC” briefly and then temperature-corrected viscosity.
U01-04-013 R1 08/13
Page 13
Viscolite VL7
• To return to viscosity measurement, press the ( key twice or the ) key once. The
display will show “
VL” brief ly before showing the current live viscosity again.
2.6 General maintenance
• Check periodically that the instrument gives a zero reading in air.
• Clean the sensor after each measurement. It is not usually necessary to switch the
Viscolite off while cleaning. The sensor can withstand vigorous cleaning so long as the
shaft is not bent or distorted in any way.
• It may be necessary to carry out a calibration check if:
(1) signs of appreciable wear or abrasion are observed, or
(2) the instrument is dropped or knocked sharply.
• Carry out a calibration check immediately if the shaft does not look perfectly straight.
• Further calibration checks are not necessary if the instrument is still giving a zero
reading in air and a reading of 1 in water at 20 °C.
• If required, certain minor adjustments are possible, as outlined in Section 4 below
(p. 20). For any other calibration problem, please contact Hydramotion Ltd or its
representative for details of the Instrument Verification Service.
2.7 Guidelines for optimum performance
• Do not operate the instrument outside its specification. In particular, do not subject the
sensor to extremes of heat or cold.
• Remove the batteries if the Viscolite is to remain out of use for a prolonged period.
• Whenever possible, keep the transducer in its carrying case.
Never leave the transducer resting on the sensor shaft.
• Never use the sensor as a lever or for any other non-measurement purpose.
U01-04-013 R1 08/13
13
Page 14
Viscolite VL7
3 Temperature-corrected viscosity
3.1 Introduction
• The Viscolite uses an internal algorithm to calculate temperature-corrected viscosity
from the measured live viscosity and the measured temperature. (For more details see
Appendix 1, p. 29.)
• To enable temperature-corrected viscosity readings to be calculated, it is necessary for
the user to enter
(1) the desired reference temperature, and
(2) correction factors specific to the f luid under test.
• To determine the correction factors, see Section 3.6 (p. 18).
• The correction factors (and some other parameters) are entered, stored and displayed
in “ E-format”. Details of how to enter these parameters are given in Section 3.3 (p. 15).
3.2 E-format numbers
• The structure of an E-format number is:
- E -
- E - is the exponent (always displayed as two digits).
Int is the integer component of the number (always four digits).
FrAC is the fraction component (decimal part) of the number (also four digits).
Int
FrAC
• To get the complete number, the sum of the integer and fraction components is
multiplied by 10 raised to the power given by the exponent. I.e.,
number = (
Int + 0.FrAC) × 10
For example:
0.61280473 = 6128.0473 × 10
E
- E -Int FrAC
–4
Thus the exponent - E - indicates how many places the decimal point should move to
the left (if negative) or right (if positive) of the number when expressed as “
14
Int.FrAC”.
U01-04-013 R1 08/13
Page 15
Viscolite VL7
• Other examples :
Number
0.6059132781
–58.8009340804
43629.5787797
- E - Int
– 04 6059 13 28
– 02 – 5880 0934
01 4362 9579
*Note rounding of last digit.
FrAC
*
3.3 How to enter an E-format parameter
• Check that the Viscolite is switched on and displaying live viscosity (VL), temperature
(
t), or temperature-corrected viscosity (VC).
• Press and hold down the
display will show “
Navigate through the menus to the desired parameter using the
indicated in Figure 7 (p. 16) or (for calibration and adjustment) Figure 11 (p. 21).
• With the desired parameter displayed (e.g.
• The display will show “
• To go to the integer component
component
• To change a displayed value:
(1) press
(2) Use the
(3) When the desired value has been reached, press
VISC”. The instrument is now in “ Setup Mode”.
FrAC from the exponent, press ). To return to the parameter, press *.
+. The value will flash.
) or ( keys to increase or decrease the value.
+ and ( keys simultaneously for a second or two. The
( or + keys as
P91), press + for 3 seconds (Figure 6 below).
- E -” briefly followed by the value of the exponent.
Int from the exponent, press (. To go to the fraction
+ again.
Figure 6 Entering numerical data in E-format: overview
U01-04-013 R1 08/13
15
Page 16
Viscolite VL7
3.4 Setting the reference temperature
• Put the instrument into Setup Mode as described in Section 3.3 (p. 15), so that “VISC” is
showing on the display.
Figure 7
Overview of the temperature-correction menu. Use the < or > keys to navigate through the sub-menus (e.g. to go from P90 to P91, etc).
Details of the Calibration and adjustment menus are given in Section 4 (p. 20).
16
U01-04-013 R1 08/13
Page 17
Viscolite VL7
Figure 8 General functions of keys
accesses a sub-menu,
CAL
enables a numerical value to be changed, or
fixes or burns a numerical value.
move up or down through a menu or a sub-menu, or
increase or decrease a numerica l value.
exits a sub-menu.
ON
• Press
( twice. The display will show “VC” (see temperature-correction menu, p. 16).
• Press
+. The display will show “P90” briefly, then its value.
• Press
( twice. The display will show “ t rEF” (reference temperature) briefly, then its
value in °C. The factory default setting is 15.0.
• Press
+. The display will begin to f lash.
• Use the
• Press
• Press
• Return to normal operating mode (“
• If required, save the change as described in Section 3.5 below before switching off the
( or ) keys to change the value as required.
+. The value will cease flashing.
*. The display will revert to “VC”.
VL”) using the ) and * keys as indicated on the
menu on p. 16.
Viscolite.
3.5 Saving changes
• Any newly-entered value will be lost on power-down unless it is “burned” into
permanent (non-volatile) memory. To do this:
(1) Navigate through the relevant menu(s) to
(2) Press
• The “burn” can be done at any time before powering-down the Viscolite. It is not
necessary to save the data after each entry.
U01-04-013 R1 08/13
+ for 3 seconds. The display will show “b uSY” briefly as the data is saved.
burn .
17
Page 18
Viscolite VL7
VISCOSITY
3.6 Determination of temperature correction factors
• The correction factors are called P90 and P91 (see p. 30). The f irst of these (P90) is
simply an offset and can be left at zero. The factory default setting for
P91 is also zero.
• There are two ways of determining
(i) Calculation of
P91
P91: (i) calculation and (ii) trial-and-error.
• To calculate P91, it is necessary to know the viscosity of the f luid at two temperatures.
Figure 9
Calculation of P91
V
1
V
2
The calculation is:
where
• If possible, use values of t
reference temperature.
typical viscosity/temperature characteristic
t
1
P91 =
V
V
t
2
TEMPERATURE (°C)
logeV1 – logeV
2
[1/(t1 + 273)] – [1/(t2 + 273)]
= viscosity at temperature t1 °C
1
= viscosity at temperature t2 °C
2
and t2 that are close to the f luid temperature and the
1
• Enter the calculated value of
P91 as described in Section 3.3 (p. 15).
• If required, save the value as described in Section 3.5 (p. 17) before switching off the
Viscolite.
18
U01-04-013 R1 08/13
Page 19
Viscolite VL7
(ii) Trial-and-error determination of P91
• This method can only be used if
(a) variation in f luid temperature is causing the live viscosity reading (
VL) to fluctuate,
and
(b) it is known that the temperature-corrected viscosity would be stable, i.e. truly
constant, despite the fluctuations in
• Using the
( or + keys as indicated in Figure 7 (p. 16), navigate through the
temperature-correction menu to
• Enter a value of 5000 (
• Navigate to
• Go back to
• Go back to
VC. Note the degree of f luctuation in the reading.
P91 and enter a higher value, say 5250 (- E - = 00, Int = 5250, FrAC = 0000).
VC and see if there is more or less fluctuation in the reading.
- E - = 00, Int = 5000, FrAC = 0000) as described in Section 3.3.
• If the reading is fluctuating less, go back to
amount of f luctuation in the
• If the reading is fluctuating more, go back to
amount of f luctuation in the
VL.
P91.
P91 and increase the value again. Check the
VC reading. Continue in this way until VC stabilises.
P91 and reduce the value. Check the
VC reading. Continue in this way until VC stabilises.
• If required, save the value as described in Section 3.5 (p. 17) before switching off the
Viscolite.
Figure 10
VISCOSITY / TEMP
VL / VC
TEMP
VISCOSITY / TEMP
VL
VC
TEMP
Trial-and-error
determination
of P91
U01-04-013 R1 08/13
TIME
P90 = P91 = 0 P91 CHANGED
VL
VC
TEMP
VISCOSITY / TEMP
TIME
TIME
P91 FINE TUNED
VL
VC
TEMP
19
Page 20
Viscolite VL7
4 Operational adjustment
4.1 Introduction
• Minor operational adjustments are carried out using the Calibration and adjustment
menus shown in Figure 11 (p. 21).
• The following adjustments can be made:
Reference Page
correct small non-zero reading of viscosity in air Section 4.4 22
stabilise reading (especially in high-noise conditions) Section 4.5 22
scale reading by any desired factor Section 4.6 23
take account of fluid density Section 4.7 24
express reading in centistokes (cSt) Section 4.8 24
match Viscolite to reference viscometer Section 4.9 25
4.2 How to modify a numerical value
• Parameters stored in “E-format” must be changed as described in Section 3.3 (p. 15).
• Other items are changed as follows:
(1) Navigate through the appropriate menu(s) until the desired item is reached.
(2) The name of the item will appear brief ly, followed by its value.
(3) Press
(4) Use the
(5) Press
(6) To go to the next item in the same submenu, press
(7) To go to the previous item in the same submenu, press
(8) To exit the submenu, press
20
+. The value will flash.
( or ) keys to decrease or increase the value as desired.
+again. The value will cease f lashing.
(.
).
*.
U01-04-013 R1 08/13
Page 21
Viscolite VL7
4.3 Accessing calibration and adjustment menus
• To access these menus from normal operating mode (“VL”, e tc):
(1) Bring the instrument into Setup Mode by pressing
display will show “
(2) Press
• Use the keypad to navigate through the menus as indicated below:
+. The display will show “ dAC”.
VISC”.
+ and ( simultaneously. The
Figure 11
Calibration and adjustment menus. Use the < or > keys to navigate through the sub­menus
U01-04-013 R1 08/13
21
Page 22
Viscolite VL7
4.4 Minor null adjustment (offset)
• If the VL reading is not zero when the sensor is perfectly clean and dry and completely
surrounded by air:
(1) Navigate to
(2) Press
(3) Press
(4) Modify the value of
Example : If the
FCAL. (To do so from dAC, simply press (.)
+. The display will show “SPA n” brief ly, followed by its value.
(. The display will show “OSEt” (offset) briefly, then its value.
OSEt using the procedure described in Section 4.2 (p. 20).
VL reading in air is 0.3 and the value of OSEt is 0.1, reduce the value of
OSEt to –0.2.
(5) Navigate down the submenu to
necessary to navigate back to the main operating mode to reach
(6) If desired, compare the adjusted value (
submenu.
(7) Save the change as described in Section 3.5 (p. 17) before switching off the Viscolite.
• Once back in normal operating mode, the main
reading.
VL and check that its adjusted value is zero. (It is not
VL — see Figure 11.)
VL) with the unadjusted value (V) in the same
VL display will show the adjusted
4.5 Stabilise reading (averaging filter)
• The Viscolite has a very high natural rejection of exterior noise and vibration. Under
normal circumstances the reading should be stable.
• If there is noticeable instability in the reading, or if very high sensivity readings are
required over a long period of time, an averaging filter can be used.
• The filter is set by changing
readings included in the averaging calculation.
• In fact, the displayed
recalculated each time the Viscolite display is updated (every 1.5 seconds, approx.).
• If
nV is set to 1, there is no averaging.
22
nV in the FCAL submenu. The value of nV is the number of
VL reading is the average of nU measurements. The average is
U01-04-013 R1 08/13
Page 23
Viscolite VL7
• To set the averaging filter:
(1) Navigate to
(2) Adjust the value as desired.
(3) Navigate down the submenu to “
(4) The filtered reading can be compared with the unfiltered reading (
submenu.
(5) If necessary, go back and increase
(6) If desired, the change can be saved as described in Section 3.5 (p. 17).
• Once back in normal operating mode, the main
reading.
nV in the FCAL submenu (see Figure 11, p. 21).
VLAV” (filtered live viscosity).
VLAV by a further amount.
VL display will show the averaged
no filtering light
filtering
heavier filtering
V) in the same
Figure 12
Application of
averaging filter
4.6 Scale reading by any desired factor
• The factory default setting is for the display to show dynamic (absolute) viscosity in
centipoise (cP), equivalent to millipascal-seconds (mPa ·s).
• To scale the reading by any desired factor:
(1) Navigate to
setting is 1.000.
U01-04-013 R1 08/13
SCAL in the LCAL submenu (see Figure 11, p. 21). The default factory
23
Page 24
Viscolite VL7
(2) Enter the appropriate conversion factor (see below).
(3) If desired, save the change as described in Section 3.5 (p. 17) before switching off.
Scaling from centipoise (cP)
unit symbol cP equivalent conversion factor
poise P 100 0.01
2
poiseuille Pl ( Pa·s, N·s/m
) 1000 0.001
• A Viscolite display can show only one decimal place. The
SCAL parameter may therefore
be used to bring more significant figures into the display. For example:
SCAL = 1 SCAL = 10 SCAL = 100
Reading on display
1.2 12 .3 12 3.4
Viscosity in cP 1.2 1.23 1.234
4.7 Take account of fluid density
• The factory default setting for f luid density is 1.000 g/cm3. To change this:
(1) Navigate to
(2) Enter the appropriate value for the fluid density in g/cm
d_n (nominal density) in the LCAL submenu (see Figure 11, p. 21).
3
.
(3) If desired, save the change as described in Section 3.5 (p. 17) before switching off.
4.8 How to display viscosity in centistokes (cSt) ( kinematic viscosity)
(1) If the f luid density is d, calculate the value of d2, i.e. the square of the density.
(1) Navigate to
(2) Enter the calculated value of
display will show viscosity in centistokes (i.e. kinematic viscosity).
d_n in the LCAL submenu (see Figure 11, p. 21).
2
d
. Once back in normal operating mode, the main VL
(3) If desired, save the change as described in Section 3.5 (p. 17) before switching off.
• Note that this procedure is fluid-specific.
24
U01-04-013 R1 08/13
Page 25
Viscolite VL7
4.9 Matching to reference instrument
• The Viscolite is rigorously calibrated at the factory to traceable standards using certified
test f luids.
• The instrument has very high repeatability. A Viscolite reading should therefore
compare very well with that obtained using any other viscometer that has been properly
calibrated and well maintained.
• If the Viscolite is being used correctly, its reading may not match that obtained with a
reference viscometer because:
(1) measurements are not being made at the same temperature;
(2) the fluid is non-Newtonian and the reference viscometer is operating at a
significantly different shear rate;
(3) the reference viscometer has not been calibrated or maintained properly;
(4) the operating parameters of the Viscolite do not match those of the reference
viscometer (for example, readings in centipoise are inadvertently being compared with
readings in other units); or
(5) the Viscolite transducer is worn, bent or damaged.
• If none of the above apply and the discrepancy remains unacceptable:
(1) note the reading obtained with the Viscolite. Call this reading
(2) Note the reading obtained with the reference viscometer under the same
measurement conditions. Call this reading
(3) Calculate the ratio
(4) Navigate to
(5) The adjusted reading (
same submenu.
U01-04-013 R1 08/13
VR/VL.
SPAn in the FCAL submenu and enter the calculated ratio.
VL) can be compared with the unadjusted reading (V) in the
VR.
VL.
25
Page 26
Viscolite VL7
5 Calibration and Maintenance
5.1 Introduction
• All instruments are factory-calibrated using a range of certified silicone oils.
• Alternative calibration fluids can be used by special request.
• Minor adjustments can be made as described above (Section 4).
• Factory calibration can easily be checked by taking readings in air and water (see
Sections 2.2 above and 5.2 below).
5.2 Calibration check
• To check calibration:
(i) Check that the sensor is clean
and dry.
(ii) Set up the instrument as shown
in Figure 13.
Figure 13
Calibration check
(iii) Ensure the transducer is stably
supported. Apply clamps to the
black plastic enclosure, not t he
metal shaft.
(iv) Switch the Viscolite on and
note the reading with the sensor
in air (i.e. not in contact with any
fluid or surface). Switch off.
(v) Immerse the sensor in clean
water at least to the minimum
immersion depth as shown in
Figures 1 and 5.
(vi) Switch on the Viscolite.
26
temperature probe
Do not clamp below here
clean water
temperature­controlled water bath
stirrer
U01-04-013 R1 08/13
Page 27
Viscolite VL7
(vii) Stir the water to ensure that it is all at a uniform temperature.
(viii) Allow plenty of time for the sensor and water to reach the same temperature.
This may take some minutes.
(ix) Note the viscosity and temperature readings (see Section 2.2).
Note: Do not hurry this test. Errors can result if sufficient time is not allowed for
temperature equalisation.
• Please contact Hydramotion Ltd or its representative if the readings are unacceptable
and the problem cannot be resolved using the procedures described in Section 4.
5.3 Instrument Verification Service
• Please contact Hydramotion Ltd or its representative if you wish to arrange a calibration
check or require a full recalibration.
• A Viscolite may be fully recalibrated:
(1) to comply with quality assurance requirements, or
(2) (by request) to adapt the instrument for use with a specific non-Newtonian fluid.
• Otherwise, full recalibration should only be necessary:
(1) after a damaged transducer has been repaired,
(2) if the sensor has become corroded or eroded, or
(3) if the Viscolite has been operated outside specification (e.g. subjected to extreme
heat).
5.4 Service and repair
• Please do not hesitate to contact Hydramotion at any time for advice or assistance.
There are no user-serviceable parts in the Viscolite. If a fault is suspected, please
contact Hydramotion or its representative. Do not attempt to repair the unit.
• If reporting an apparent fault, please provide the following information:
(a) Serial number of the unit (b) How long the Viscolite has been in service (c) The type/composition of the fluid(s) being measured
U01-04-013 R1 08/13
27
Page 28
Viscolite VL7
(d) Measurement conditions (especially temperature) (e) Observed and expected viscosity values ( f) Stability of the measurement(s)
( g) When the fault was  rst noticed
(h) Whether the fault is continuous or intermittent.
Prior to shipping:
(1) Clarif y any s uch m atter s a s cos ts, liab i lity, insuranc e, i mpor t/ e x por t
documentation, customs declarations etc.
(2) Fax or mail a Returned Goods COSHH Declaration, duly completed, to:
Hydramotion Ltd 1 York Road B Seven Street Malton York YO17 6YA England
When shipping:
(1) Use the original packaging.
(2) Ensure the Viscolite is securely packed.
(3) Include documentation clearly stating the serial number and the reason(s) for
return.
usiness Park
28
U01-04-013 R1 08/13
Page 29
Viscolite VL7
Appendix 1: Operating principles
How the Viscolite determines viscosity
The signal generated by the transducer is called the “ loss factor” L. It is converted to the
“raw” calculated viscosity
V = (1/d)(P30 + P31L + P32L
Here
d is the density of the fluid and P30, P31, P32, etc. are “ calibration constants”
which are unique to each instrument and must be empirically determined. The number
of terms used varies between transducer models. All the Calibration Constants are set
out on the Certificate of Calibration and Conformity and can be accessed from the
submenu.
In fact, the response of a resonant viscometer like the Viscolite is not purely with
viscosity but with the product of viscosity and de nsit y, i.e. viscosity × density. In
practice, however, repeatability is generally far more important than absolute accuracy.
Since viscosity changes on a far greater scale than density, changes in density can in most
cases safely be ignored.
V using the following polynomial equation:
2
+ P33L3 + P34L4 + ... )
LCAL
For maximum accuracy, the density
d of the fluid can be entered in the d_n parameter.
See Section 4.8 above.
Scaling
If the scaling factor SCAL (Section 4.6) is applied, the equation becomes:
SCAL
VL =
(
P30 + P31L + P32L
d
2
+ P33L3 + P34L4 + ... )
The “live” viscosity reading displayed in normal operational mode (
product of the “raw” viscosity
V and SCAL, i.e.
VL = V × SCAL
U01-04-013 R1 08/13
VL) is therefore the
29
Page 30
Viscolite VL7
Viscometer matching
When matching the Viscolite response to that of a reference instrument, the SPAn factor
is applied (Section 4.9). The viscosity reading displayed in normal operational mode (
is then the product of the calculated viscosity
VL and SPAn , i.e.
VL)
VL = VL × SPAn
Null adjustment
To correct a small non-zero reading in air, an offset OSEt is added (Section 4.4). The
viscosity reading displayed in normal operational mode (
product of the calculated viscosity
VL and SPAn , i.e.
VL) then becomes OSEt plus the
VL = (VL × SPAn) + OSEt
Temperature correction (Section 3)
To correct the measured viscosity to viscosity at a given reference temperature, the
Viscolite uses an algorithm based on the commonly-accepted Arrhenius relationship
viscosity = Ae
where T is the absolute temperature and A and B are constants depending on the
particular fluid. The following equation is used to calculate the temperature-corrected
viscosity
VC from a viscosity reading VL:
VC = {VL . e
P91[1/(trEF + 273) – 1/(tL + 273)]
B/T
} – P90
where
trEF = reference temperature (°C) (Section 3.4)
These parameters can all be accessed from
VC = viscosity corrected to reference temperature VL = calculated viscosity
tL = measured temperature (°C) P90 = correction factor (offset) P91 = correction factor
(Section 3.6)
VC in the temperature-correction menus.
30
U01-04-013 R1 08/13
Page 31
Viscolite VL7
Appendix 2: Menu overview
Item index
(sub)menu item description section(s) page(s)
burn save settings in non-volatile memory 3.5 17 LCAL d-n fl uid density 4.8 24 t - FCAL dEGC measured temperature P30 (etc) - E - exponent of E-format number 3.3 14 P30 (etc) FrAC fraction of E-format number 3.3 14 P30 (etc) Int integer component of E-format number 3.3 14 tESt L measured loss FCAL nV averaging sample number 4.5 22 t - FCAL OFSt temperature null offset FCAL OSEt viscosity null offset 4.4 22 LCAL P30 – 38 calibration constants Appx 1 29 t - VC P90, P91 temperature correction factors 3.6 18 t - FCAL rES PRT resistance LCAL SCAL scaling factor 4.6, 4.7 23, 24 FCAL SPAn gain factor 4.9 25 t live temperature 2.4 12 VISC t temperature menu 3.4 16 tESt tL live temperature t - VC tL live temperature Appx 1 29 t - VC trEF reference temperature 3.4 16 FCAL V unadjusted viscosity 4.4 22 tESt V unadjusted viscosity VC temperature-corrected viscosity 2.5 12 tESt VC temperature-corrected viscosity t - VC VC temperature-corrected viscosity FCAL VCAV fi ltered temperature-corrected live viscosity tESt VCAV fi ltered temperature-corrected live viscosity VISC viscosity main menu 3.3 15 VL live viscosity 2.1 10 tESt VL live viscosity t-VC VL live viscosity FCAL VLAV fi ltered live viscosity 4.5 22 tESt VLAV fi ltered live viscosity
U01-04-013 R1 08/13
31
Page 32
Viscolite VL7
Figure 14
Menu
overview
32
U01-04-013 R1 08/13
Page 33
Viscolite VL7
Troubleshooting
Problem Possible cause Remedy Section
No display Not switched on Switch on 2.1
Batteries f lat Replace batteries 1.3
Reading high Shear-thickening f luid 2.3, 4.9
Fluid temperature low Use temperature correction 2.3
Fluid density high Enter density value 4.8
Reading low Shear-thinning fluid 2.3, 4.9
Fluid temperature high Use temperature correction 2.3
Fluid density very low Enter density value 4.8
Insufficient immersion Increase immersion depth 2.3
Unstable/jittery reading
Reading not precise enough
Sensor warmer or colder than fluid
Suspended solids in fluid
Plastic fluid (Bingham effect)
High-frequency external noise
Fluid has very low viscosity
Allow equalisation 2.3
Stir gently; filter fluid if possible
Stir gently 2.3
Set averaging filter 4.5
Change scaling factor 4.6
Alarm displays
batteries low — save any changes without delay to avoid loss of entered data
bAtt
d EAd batteries exhausted — temporary memory no longer operational
S IG no signal from transducer*
*Please contact Hydramotion Ltd or its representative if you see this display.
U01-04-013 R1 08/13
2.3
33
Page 34
Viscolite VL7
General index
adjustments 20 air reading 10 alarm displays 33 averaging fi lter 22
bAtt 33
batteries 9, 13 battery cradle 9 beeper 10 Bingham 11 calibration and adjustment menus 21 calibration check 10, 13 calibration constants 29 centipoise 10, 23 centistokes 24 Certifi cate of Calibration and Conformity 8, 29 clamping 11, 26 cleaning 13 correction factors 12, 14
determination of 18
damage 7
dEAd 33
density 12, 24, 29 display unit 10 E-format 14, 15
fraction component 14
integer component 14 exponent 14 fi lter 22 instability (in reading) 11, 22 kinematic viscosity 24 live viscosity 10 loss factor 29 maintenance 13 matching 25, 30 materials 7 menu
calibration and adjustment 21
temperature correction 16 minor null adjustment 22, 30 model type 7, 8 non-Newtonian fl uids 12, 25 numerical value, modifi cation of 20 options 7
P30 etc 29
P90, P91 18, 19
packaging 7 plastic behaviour 11 platinum resistance thermometer 5 polished 7 power source 7 POL 7 PRT 5 range 7 recalibration 27 reference instrument, matching 25, 30 reference temperature 16 repeatability 6, 25 Returned Instrument Health & Safety Certifi cate 28 return to factory 28 scaling 23, 29 serial number 8 setting
fi lter 22
reference temperature 16 Setup Mode 15 shaft 7 shear rate 12, 25
SIG 33
specifi cation 7 stabilise reading 22 start-up 10 suspended solids 11 switching off 10 switching on 10 temperature
correction factors 13
equalisation 11
measurement 12 temperature-corrected viscosity 12, 14, 30 turning off 10 turning on 10 viscosity
dynamic 11
kinematic 24
measurement 10
measurement principles 29
34
U01-04-013 R1 08/13
Page 35
Loading...