Contents
1 Safety notes ........................................................................................... 1
2 Technical specifications ...................................................................... 2
3 Delivery Scope ...................................................................................... 2
4 Installing the Batteries ......................................................................... 3
5 Display ................................................................................................... 3
6 Keypad ................................................................................................... 4
7 Filling the Electrolyte Solution ............................................................ 5
8 Connecting the Dissolved Oxygen Probe .......................................... 5
9 Switching the Tester On and Off ......................................................... 6
10 Prior to Use ............................................................................................ 6
11 Setup Menu ............................................................................................ 6
11.1 Setting the default option .................................................................................................. 7
12 Setting the salinity coefficient ............................................................. 8
13 Setting the ambient pressure .............................................................. 8
14 DO Calibration in % Saturation Mode ................................................. 9
14.1 Single point calibration - 100% saturation ......................................................................... 9
14.2 2 points calibration ........................................................................................................... 9
15 Temperature Calibration .................................................................... 10
16 Dissolved Oxygen Measurement ....................................................... 11
17 Auto-Hold ............................................................................................. 11
18 Electrode Care and Maintenance ...................................................... 12
19 Troubleshooting .................................................................................. 12
20 Addendum: Preparation of the Zero Oxygen Solution .................... 12
21 Contact ............................................................................................... 13
22 Disposal ............................................................................................... 13
© PCE Instruments

© PCE Instruments
1
1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The
device may only be used by qualified personnel and repaired by PCE Instruments personnel.
Damage or injuries caused by non-observance of the manual are excluded from our liability and
not covered by our warranty.
• The device must only be used as described in this instruction manual. If used otherwise,
this can cause dangerous situations for the user and damage to the meter.
• The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
• Do not expose the device to shocks or strong vibrations.
• The case should only be opened by qualified PCE Instruments personnel.
• Never use the instrument when your hands are wet.
• You must not make any technical changes to the device.
• The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
• The device must only be used with accessories from PCE Instruments or equivalent.
• Before each use, inspect the case for visible damage. If any damage is visible, do not
use the device.
• Do not use the instrument in explosive atmospheres.
• The measurement range as stated in the specifications must not be exceeded under
any circumstances.
• Non-observance of the safety notes can cause damage to the device and injuries to
the user.
We do not assume liability for printing errors or any other mistakes in this manual.
We expressly point to our general guarantee terms which can be found in our general terms of
business.
If you have any questions please contact PCE Instruments. The contact details can be found at
the end of this manual.

© PCE Instruments
2
2 Technical specifications
Measuring range oxygen content
Resolution oxygen content
Measuring accuracy oxygen content
Measuring range oxygen saturation
Resolution oxygen saturation
Measuring accuracy of oxygen saturation
0 ... +40 °C / 32 ... 104 °F, automatic
60.0 ... 112.5 kPa, 450 ...850 mmHg
0 ... +60 °C / 32 ... 140 °F, max. 80 % r.H.
Dimension (length x radius)
3 Delivery Scope
1 x Dissolved Oxygen Meter PCE-WO2 10
1 x Oxygen probe with measuring cable
2 x Membrane
1 x Electrolyte solution (30ml)
2 x 1.5 V AAA batteries
1 x User manual
1 x Carrying case

© PCE Instruments
3
4 Installing the Batteries
• Twist the electrode collar counter clockwise, pull the 6-pin connector away from the
tester.
• Insert two AAA batteries into the battery compartment, note polarity.
• Align the slot on connector, gently push the connector into the tester.
• Twist the electrode collar clockwise until it is tight.
5 Display
PCE-WO2 10 pocket dissolved oxygen tester is equipped with an easy-to-read LCD display that
used to show the measured values and mode icons. The
following table describes the function of each icon.
INDEX:
Indicates the tester is in the measurement mode.
Indicates the tester is in the calibration mode.
Indicates the tester is in the setting mode.
Indicates the measuring value has been locked.
Indicates the temperature compensation is enabled.
When the battery is depleted, the icon will disappear.

© PCE Instruments
4
6 Keypad
The tester has a succinct membrane keypad, names and symbols describe each function key
controls.
INDEX:
• Switches the tester ON/OFF.
• Locks the measured value, press the key again to resume measuring.
• Exits the calibration or setting and returns to the measurement mode.
• Press the key to start the calibration.
• Press and hold the key to enter the setup menu.
• In the setting mode, press the key to select the default option.
• Confirms the calibration or selected option.

© PCE Instruments
5
7 Filling the Electrolyte Solution
• Take out the dissolved oxygen probe and electrolyte solution from the packaging. Unscrew
the membrane cap.
• Fill the membrane cap halfway with electrolyte solution.
• Screw the membrane cap onto the probe, excess electrolyte solution will drain out.
• Be sure the cathode of probe makes contact with membrane cap, the electrolyte solution in
membrane cap should be without an air bubble.
8 Connecting the Dissolved Oxygen Probe
Insert the 6-pin connector of probe into the connector socket on the tester. After the connection
is completed, DO NOT pull on the cable. Always make sure that the connector is clean and dry.

© PCE Instruments
6
9 Switching the Tester On and Off
• Press the Meas key to switch on the tester, the display shows measured values.
• Press and hold the Meas key for 5 seconds, the tester will switch off.
• To enable the Auto-Power Off function, please refer to chapter SETUP MENU.
10 Prior to Use
Switch on the tester 10 to 15 minutes and wait for the probe to polarize.
11 Setup Menu
PCE-WO2 10 pocket dissolved oxygen tester contains an integrated setup menu that is used to
customize the displayed option to meet measurement
requirements.
Set the salinity coefficient of
sample.
Set the barometric pressure
coefficient.
Set the number of calibration
points.
Set the default measurement
unit.
Calibrate the temperature.
When the option is enabled,
the tester will automatically
sense a stable reading and
lock the measurements.
When the option is enabled,
the tester will automatically
switch off if no key is pressed
within 8 minutes.
Reset the tester to factory
default settings.

© PCE Instruments
7
11.1 Setting the default option
• Press and hold the Cal key for 3 seconds to enter the setup menu, the tester shows a
menu item and default option.
• If necessary, press the Cal key again to select the desired option.
• Press the Enter key to confirm, the tester moves to the next menu item.
• Repeat the steps above until the tester returns to the measurement mode. Setting is
completed.
• During the setting process, press the Meas key, the tester will exit the setup menu and
return to the measurement mode.
• If you do not want to calibrate the temperature, press the Enter key to skip the °C/CAL
or °F/CAL option.
• The Reset function will restore the tester back to factory default settings. If enabled, all
of the calibration data and selected parameters will be lost or reset, the tester must
be recalibrated.

© PCE Instruments
8
12 Setting the salinity coefficient
In order to get an accurate reading, the salinity of the sample should be specified which is done
with the help of a coefficient.
Salt dissolved in water will influences oxygen content of water. If your sample contains high levels
of salinity, ensure that setting an applicable salinity
coefficient prior to measurement.
• In the measurement mode, press and hold the Cal key for 3 seconds to enter the
setup menu, the display shows (salinity coefficient) option.
• Press the Cal key again to set the value.
• Press the Enter key to confirm, the display shows the next menu item.
• Press the Meas key to exit the setting and return to the measurement mode.
13 Setting the ambient pressure
To ensure an accurate reading, the ambient pressure must be set in the unit mmHg.
The following table will help to convert your ambient pressure into the required unit mmHg on
the basis of your current height.
• In the measurement mode, press and hold the Cal key for 3 seconds to enter the
setup menu, the display shows .
• Press the Enter key, the display shows (Barometric Pressure Coefficient)
option.
• Press the Cal key to set the value.
• Press the Enter key to confirm, the display shows the next menu item.
• Press the Meas key to exit the setting and return to the measurement mode.

© PCE Instruments
9
14 DO Calibration in % Saturation Mode
PCE-WO2 10 pocket dissolved oxygen tester can be calibrated quickly in air. In the percentage
saturation mode, the tester is able to perform either 1 or 2 points calibration. For single point
calibration, we recommend that you perform 100% saturation calibration in the air-saturated
water. If the 2 points calibration is selected, the zero oxygen solution needs to be used.
14.1 Single point calibration - 100% saturation
• Make sure that you have selected 1 point
calibration in the setup menu.
• Press the Cal key, the tester shows “100% /
CAL1”.
• Hold the dissolved oxygen probe in the air
at 100% relative humidity or place the probe
into the air-saturated water for 15
minutes.
• Press the Enter key. Wait for the reading to
stabilize, the tester automatically shows
END. Single point calibration is completed.
14.2 2 points calibration
• Make sure that you have selected 2 points
calibration in the setup menu.
• Repeat the steps 1.2 to 1.4 above. When
the first calibration point is completed, the
display will show “0% / CAL2”. The
tester prompts you to continue with second
point calibration.
• Immerse the dissolved oxygen probe into
the zero oxygen solution, stir the probe
gently.
• Press the Enter key to confirm. Wait for the
reading to stabilize (this can often take more
than 10 minutes), the display automatically
shows END. The tester returns to the
measurement mode. Calibration is
completed.
Performing a percentage saturation
calibration will simultaneously calibrate the
corresponding mg/L (or ppm) concentration value.

© PCE Instruments
10
15 Temperature Calibration
During the measurement process, if the temperature reading displayed differs from that of an
accurate thermometer, the tester needs to be calibrated.
• Press and hold the Cal key for 3 seconds to
enter the setup menu.
• Press the Enter key until the tester shows
°C/CAL or °F/CAL.
• Press the Cal key to enter the temperature
calibration mode.
• Press the Cal key again to modify the
temperature value (Resolution: 0.5°C).
• Press the Enter key to confirm, the tester
shows the next menu item.
• Press the Meas key to return to the
measurement mode. Calibration is
completed.

© PCE Instruments
11
16 Dissolved Oxygen Measurement
PCE-WO2 10 pocket dissolved oxygen tester is suitable for measuring the water, wastewater,
brine and other liquids. If the sample is belong to the seawater or other water containing large
amounts of salt, please setting the salinity coefficient before measurement. Some gas and steam
such as chloride, sulfur
dioxide, sulfureted hydrogen, ammonium, carbon dioxide and iodin can permeate the membrane
via diffusion. So their existence will influence the measurement of dissolved oxygen. If the sample
contains the solvent, grease, sulfide and alga, the membrane on the probe will be blocked,
damaged or eroded.
• Connect the dissolved oxygen probe to tester and wait for 15 minutes to polarize the
probe.
• If necessary, to set the barometric pressure and salinity coefficient in the setup menu
(Refer to page 5).
• Immerse the probe in the sample solution, make sure the temperature sensor on the
probe is fully immersed.
• Stir the probe gently. Record the measured value when the reading is stable.
17 Auto-Hold
PCE-WO2 10 pocket dissolved oxygen tester contains an Auto-Hold function. If enabled, the
tester will automatically sense a stable reading and lock the
measurements. The icon appears on the display. If disabled, press the Hold key, the tester
will immediately lock the displayed value. Press the Meas key to resume measuring.

© PCE Instruments
12
18 Electrode Care and Maintenance
• Always keep the membrane of the dissolved oxygen probe is wet or moist.
• If you do not use the probe for long periods, please screw off membrane cap and rinse
the cathode, anode and membrane with deionized water, then soak up residual water
on them with filter paper. Install the probe again.
19 Troubleshooting
DO probe does not connect
to tester
Check the connector of probe.
Measured value is out of
range
Check the DO membrane whether
clogged, dirty or broken.
Electrolyte solution is
depleted
Refilling electrolyte solution.
Zero oxygen solution is
contaminated
Replace the calibration solution.
20 Addendum: Preparation of the Zero Oxygen Solution
Dissolve 500mg of sodium sulfate (Na2SO3) reagent and a small amount of cobalt (II) chloride
hexahydrate (CoCl
2
• 6H
2
O) in the 250mL distilled water, mix the solution until the reagent is
completely dissolved.

21 Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact
us. You will find the relevant contact information at the end of this user manual.
22 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European
Parliament applies. Due to the contained pollutants, batteries must not be disposed of as
household waste. They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either reuse them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance
with your local waste regulations.
If you have any questions, please contact PCE Instruments.
© PCE Instruments
13

© PCE Instruments
14
PCE Instruments contact information
Germany France Spain
PCE Deutschland GmbH
PCE Instruments France EURL
PCE Ibérica S.L.
Im Langel 4
23, rue de Strasbourg
Calle Mayor, 53
D-59872 Meschede
67250 Soultz-Sous-Forets
02500 Tobarra (Albacete)
Deutschland
France
España
Tel.: +49 (0) 2903 976 99 0
Téléphone: +33 (0) 972 3537 17
Tel. : +34 967 543 548
Fax: +49 (0) 2903 976 99 29
Numéro de fax: +33 (0) 972 3537 18
Fax: +34 967 543 542
[email protected] [email protected] [email protected]
www.pce-instruments.com/deutsch
www.pce-instruments.com/french
www.pce-instruments.com/espanol
Germany United Kingdom Italy
PCE Produktions- und
PCE Instruments UK Ltd
PCE Italia s.r.l.
Entwicklungsgesellschaft mbH
Unit 11 Southpoint Business Park
Via Pesciatina 878 / B-Interno 6
Im Langel 26
Ensign Way, Southampton
55010 Loc. Gragnano
D-59872 Meschede
Hampshire
Capannori (Lucca)
Deutschland
United Kingdom, SO31 4RF
Italia
Tel.: +49 (0) 2903 976 99 471
Tel: +44 (0) 2380 98703 0
Telefono: +39 0583 975 114
Fax: +49 (0) 2903 976 99 9971
Fax: +44 (0) 2380 98703 9
Fax: +39 0583 974 824
[email protected] [email protected] [email protected]
www.pce-instruments.com/deutsch
www.pce-instruments.com/english
www.pce-instruments.com/italiano
The Netherlands China Hong Kong
PCE Brookhuis B.V.
PCE (Beijing) Technology Co., Limited
PCE Instruments HK Ltd.
Institutenweg 15
1519 Room, 6 Building
Unit J, 21/F., COS Centre
7521 PH Enschede
Zhong Ang Times Plaza
56 Tsun Yip Street
Nederland
No. 9 Mentougou Road, Tou Gou District
Kwun Tong
Telefoon: +31 (0)53 737 01 92
102300 Beijing, China
Kowloon, Hong Kong
[email protected] Tel: +86 (10) 8893 9660
Tel: +852-301-84912
www.pce-instruments.com/dutch
[email protected] [email protected]
www.pce-instruments.cn
www.pce-instruments.cn
United States of America Turkey
PCE Americas Inc.
PCE Teknik Cihazları Ltd.Şti.
711 Commerce Way suite 8
Halkalı Merkez Mah.
Jupiter / Palm Beach
Pehlivan Sok. No.6/C
33458 FL
34303 Küçükçekmece -
İstanbul
USA
Türkiye
Tel: +1 (561) 320-9162
Tel: 0212 471 11 47
Fax: +1 (561) 320-9176
Faks: 0212 705 53 93
[email protected] [email protected]
www.pce-instruments.com/us
www.pce-instruments.com/turkish