PCE Instruments PCE-VS11, PCE-VS12 Instruction Manual

PCE-
PCE-
PCE Americas Inc. 711 Commerce Way Suite 8 Jupiter FL-33458 USA From outside US: +1 Tel: (561) 320-9162 Fax: (561) 320-9176 info@pce-americas.com
www.pce-instruments.com/english
www.pce-instruments.com
PCE Instruments UK Ltd.
Units 12/13
Southpoint Business Park
Ensign way
Hampshire / Southampton
United Kingdom, SO31 4RF
From outside UK: +44
Tel: (0) 2380 98703 0
Fax: (0) 2380 98703 9
info@industrial-needs.com
Inhaltsverzeichnis
1.Application.............................................................................................................4
2.The Devices at a Glance..........................................................................................4
3.Connectors.............................................................................................................. 5
3.1.Power Supply..................................................................................................5
3.2.Relay Output...................................................................................................5
3.3.USB Interface..................................................................................................6
4.Parametrization.......................................................................................................6
4.1.Device Identification.......................................................................................6
4.2.Settings............................................................................................................7
4.2.1.General.....................................................................................................7
4.2.2.Monitoring Mode.....................................................................................7
4.2.3.Gain.........................................................................................................7
4.2.4.Filters and Integrators..............................................................................8
4.2.5.Warning and Alarm Limits.......................................................................8
4.2.6.Switching Output.....................................................................................8
4.2.7.Factory Settings / Calibration...................................................................9
4.3.LED Status Indicators......................................................................................9
5.Measuring in the Time Domain...............................................................................9
6.Measuring in the Frequency Range (FFT)............................................................11
7.Teach-in Function.................................................................................................12
8.Measuring Points on Rotating Machines..............................................................13
8.1.General..........................................................................................................13
8.2.Attachment.................................................................................................... 13
8.3.Attachment Recommendations to ISO 10816-1.............................................13
9.Vibration Monitoring with Standard Limits..........................................................14
10.Installing the PC Software..................................................................................17
11.Integration of VS11/12 with other Software........................................................18
12.Firmware Update................................................................................................18
13.Technical Data....................................................................................................20
1. Application
The PCE-VS11/12 vibration switches are designed to monitor vibration amplitudes on ro ­tating machines (see. Chapter 9). When a given amplitude is exceeded an alarm sig ­nal or automatic shutdown is triggered via the relay output. Similarly, the devices can be used as impact detectors, for example, to report collisions.
The PCE-VS11 and PCE-VS12 devices measure and monitor vibration in both the time and frequency domain, for this reason they can selectively monitor individual frequency band components.
The devices possess a piezoelectric precision accelerometer and micro-controller­based electronics. This ensures high reliability and reproducibility. The devices are configured via a USB interface and free software. Due to its wide range of settings PCE-VS11/12 can be adjusted to every application, from the measurement of low vibra­tions to the detection of high-frequency shock accelerations.
2. The Devices at a Glance
PCE-VS11:
PCE-VS12:
4
3. Connectors
3.1. Power Supply
The PCE-VS11 vibration switch operates with DC voltage in monitoring mode, the termi­nals "+ U" (positive) and "0V" (negative/ground) must be connected inside the cas­ing. The supply voltage range is 5 to 26 V. Power consumption is less than 100 mA.
During parameter setting the PCE-VS11 obtains its power via the USB cable. The PCE-VS12 is powered by connecting a USB cable to the 8-pin socket. Alternatively, a DC voltage of 5 to 30 V can be connected at the terminals 4 (positive pole) and 7 (minus/ground) of the 8-pin socket (Figure 2).
The supply voltage connection is protected against false polarity.
3.2. Relay Output
The devices contain a PhotoMOS relay. The relay switching behavior can be pro ­grammed with the PCE-VS1x software (see. Chapter 4.2.6). The relay terminals are gal­vanically isolated from the rest of the circuit.
The PCE-VS11 relay output is connected via screw terminals inside the housing (Figure
1).
PCE-VS12 has relay terminals located at contacts 1 and 2 of the 8-pin socket (Figure 2).
Metra offers connection cables for PCE-VS12 with an 8-pin connector for power supply and relay output.
Please note that the relay is only suitable for switching small loads (see. Chapter Technical Data). No overload protection is provided.
5
Figure 2: Outside view of the PCE-VS12 socket with terminal numbers
Figure 1: Open PCE-VS11 with terminals for power supply / relay output and USB socket
PCE-
PCE-
PCE-
PCE-
PCE-
PCE-
$
recognized. Otherwise, you can start the search manually by clicking on "Search PCE-VS1x". The computer then searches from the entered COM port number and ends with COM50. You can also manually change the COM port. This may be useful if several PCE-VS11/12 are connected to the computer at the same time. The program works with COM port numbers 1 to 50.
On the top right you will see a status bar. If the green framed “OK” signal is shown the connection has been established. If the connection is interrupted the red framed “ERROR” signal will be shown.
4.2. Settings
4.2.1. General
The current settings are read as soon as the device is detected. In the line next to the COM port number you can see the type, version (3 digits for hardware and 3 digits for software), serial number and date of last calibration. This information cannot be edited. The device name can be overwritten and transferred to the device by pressing “Enter”.
Press the “Save” button to save the settings as an XML file and “Load” to upload them into the program. The adjustable parameters are assigned to the function blocks “Gain”, “Filters/Integrators”, “Warning”/Alarm” and “Switch Output”.
All entries will immediately be transferred to PCE-VS11/12 and retained even after dis­connecting the supply voltage.
4.2.2. Monitoring Mode
PCE-VS11/12 has two monitoring modes to choose from:
Monitoring in the time domain with RMS and peak values (see Chapter 5)
Monitoring in the frequency domain with frequency-band-dependent limit values
(see Chapter 6)
Select the mode under “Monitoring”. The most recently selected mode and the cor ­responding limits will remain active after closing the program or interrupting the USB connection. The same applies to the teach-in function (see. Chapter 7).
4.2.3. Gain
The gain can be selected from values 1, 10 and 100 via the “Fix” menu. The “Auto” setting automatically selects the most appropriate gain range. In this case the gain menu is grayed-out.
Most monitoring tasks can be carried out using automatic gain (auto). It is advanta ­geous because it achieves a better resolution when measuring low vibration ampli­tudes at higher gain levels. On the other hand unexpected high amplitudes do not cause overload.
There are, however, applications for which automatic gain selection is inappropriate, for example, at amplitudes that constantly fluctuate around a switching point or fre ­quent single shocks.
7
4.2.4. Filters and Integrators
PCE-VS11/12 can monitor vibration acceleration or vibration velocity. A range of high and low pass filters are available for selection. The widest frequency range is 0.1 Hz to 10 kHz for acceleration, and 2 to 1000 Hz for velocity. The frequency range is ad ­justed via a drop-down menu. The three vibration velocity ranges can be found at the end of the menu. For information on customary frequency ranges in monitoring rotating machinery, see Chapter 9.
Setting the filters and integrators is only relevant when monitoring in the time do ­main (RMS and peak). In the FFT mode they are deactivated.
4.2.5. Warning and Alarm Limits
You can select the monitoring value from the “RMS/Peak” menu. RMS values are typically used for measuring vibration, and peak values for single impacts.
The alarm limit determines the switching threshold of the relay output. It is entered in m/s² for acceleration or mm/s for velocity. The permissible value range is 0.1 to
500.0.
The warning limit is entered as a percentage of the alarm value.
Values ranging from 10 to 99% are permissible. The warning limit can be used to in­dicate pre-alarm status via the LEDs before the alarm is triggered (see Chapter 4.3).
The “teach-in-factor” is an automatic measuring function for the alarm limit (see Chapter 7). It determines how far the alarm limit is set above the currently measured maximum value. The teach-in warning limit is always set at 50%.
It is only necessary to preset the monitoring variables and the alarm limit when mea­suring in the time domain (RMS and peak). In FFT mode the alarm limit is set in the FFT window (see Chapter 6).
4.2.6. Switching Output
The PCE-VS11/12 contains a PhotoMOS relay switch. The switching function can be specified in the options menu. The relay opens (n.c.) or closes (n.o.) in response to a warning or alarm signal.
The power-on delay is the delay between switching on the power and activation of the monitoring function. It helps prevent false alarm signals after switching on the device caused by the transient response of the signal processing.
The delay range is 0 to 99 seconds.
The power-on delay is the delay between the alarm threshold being exceeded and the relay switching. At zero the relay reacts immediately.
If a minimum time duration should apply to exceeding the alarm limit, a switching delay of up to 99 seconds can be entered.
The “Hold time” is the time when the amplitude falls below the alarm limit until the relay returns to normal status. This setting can be useful if a minimum alert duration is required. The range is from 0 to 9 seconds.
8
4.2.7. Factory Settings / Calibration
By clicking the “Set defaults” button all parameters are restored to factory settings (acceleration 2-1000 Hz, automatic gain, limit value 10 m/s², pre-alarm to 50%, teach-in factor 2, relay close when alarm is triggered, switching delay 10 s, alarm delay 0 s, hold time 2 s).
The calibration password (“Cal. Password”) only needs to be entered by calibration labs.
4.3. LED Status Indicators
The PCE-VS11 signalizes the current status via four green/red LEDs. Whenever the de­vice is ready for operation all LEDs light up. The LEDs have the following configu ­ration:
4 x green: no warning / no alarm
2 x green/ 2 x red: warning limit exceeded
4 x red: alarm limit exceeded
The LEDs show the current vibration level in relation to the limit values.
They may differ from the current switching status of the relay if the switching delay or the hold time has not yet elapsed.
5. Measuring in the Time Domain
Besides vibration monitoring with switch output, the PCE-VS12 can be used in conjunc ­tion with PC software to record and display RMS and peak values with the selected filer and integrator settings.
For this purpose switch to the tab “RMS/Peak”. The upper window contains the nu ­merical display for RMS and peak. The time chart plots the course of the vibration quantity selected under “Plot” (Figure 4).
9
PCE- PCE-
By right clicking on the chart area you can automatically scale the chart (auto scaling Y). Double-clicking on the scale label of the Y-axis allows you to manually rescale the axis by overwriting it.
Scaling the frequency axis (X) is unnecessary because it is fixed by the frequency range of the FFT (1/10 kHz). The Y-axis can be displayed by a linear or logarithmic scale. To export the chart data the same options as in the time­domain measurements are available (see Section 9).
Input fields for 10 amplitudes and 10 frequencies are lo­cated under the chart menu. Here you can specify a limit line which is placed in the frequency spectrum and signals
an alarm when the limit is exceeded. The limit line en­ables you to selectively monitor spectral components. This may be advantageous in order to monitor a particular component from among a mix of vibration frequencies.
For the switching mode, warning limit and delay time the settings described in sec­tions 4.2.5 and 4.2.6 apply.
In the row with 10 frequencies you can enter any desired value in the range of 1 Hz to 1000 or 10000 Hz (depending on the filter range selected). The only condition is that the frequencies ascend from left to right. The amplitude entered below the fre­quency in m/s² is the limit of the next lower frequency up to this frequency. If you need less than 10 basic parameters you can also enter the maximum frequency of 1000 or 10000 Hz with the corresponding amplitude limit further to the left.
In this case the values to the right of the maximum frequency will be ignored.
The limit curve can be shown or hidden on the chart. The VS11/12 limit monitoring nevertheless always remains active.
7. Teach-in Function
The PCE-VS11 has a teach-in function for calibrating the alarm limit. A PC is not re­quired for this function. To use the teach-in function the vibration switch has to be mounted onto the object to be measured, which should be in ready-to-monitor oper­ating status.
To activate the teach-in function remove the screw cover labeled “teach-in” and briefly press the button underneath with a long, non-conductive object. When doing this be careful to avoid impacts on the casing.
According to the selected monitoring mode, the vibration switch will now determine the alarm threshold based on the available values.
This can take between 4 and 40 s, during which the LEDs remain unlit. Meanwhile the following processes will run in the vibration switch:
With RMS and peak monitoring in the time domain the selected monitoring
quantity with the set filter range is measured for a few seconds. The resulting RMS and peak values are multiplied by the teach-in factor (programmed under set-up) and saved as the alarm limit. The warning limit is set at 50 %.
Before activating the teach-in function please select a suitable filter range.
12
Figure 8: Chart Menu
With FFT monitoring in the frequency domain the frequency spectrum up to
10 kHz is measured and averaged for a few seconds and the results are recorded. Subsequently, the largest spectral line is determined. If this line is below 1kHz, the analysis will be repeated with a 1 kHz band width. The frequency range will then be subdivided into ten equally wide intervals of 100 or 1000 Hz. For each of these ranges the amplitude with the largest spectral line, is multiplied by the teach-in factor, and set as the limit. If the maximum lies on the margin of an in­terval, the next interval will also be set at this limit. The warning limit is also set at 50 %.
In this way the alarm limit can be determined without knowledge of the actual accel ­eration and velocity. The teach-in factor determines the permissible tolerance.
Attention: Please do not touch the PCE-VS11 during the teach-in process.
8. Measuring Points on Rotating Machines
8.1. General
For monitoring the condition of a machine selecting suitable measuring points is de­cisive. Whenever possible trained personnel with experience in machine monitoring should be called upon.
It is generally advisable to measure machine vibrations as near as possible to their source. This helps keep measuring signal distortions, due to transferred components, to a minimum. Suitable measuring location points include rigid parts such as bearing housings and gearbox housings.
Measuring point locations unsuitable for measuring vibration are light or mechani ­cally flexible machine parts, such as metal sheets or cladding.
8.2. Attachment
The PCE-VS11/12 devices have robust aluminum casing with an M8 thread pin for at­tachment. The devices should be attached by hand only. Please do not use tools.
8.3. Attachment Recommendations to ISO 10816-1
The ISO 10816-1 standard recommends bearing housings or their immediate sur ­roundings as preferred measuring location points for measuring machine vibrations (Figures 9 to 12).
For the purpose of machine monitoring it is usually sufficient to take measurements in only one direction, either vertically or horizontally
On machines with horizontal shafts and rigid foundations the largest vibration am­plitudes occur horizontally. On flexible foundations strong vertical components oc­cur.
For the purpose of acceptance tests, measuring values should be recorded in all three directions (vertical, horizontal and axial) on all bearing locations at the center of the bearing.
The following illustrations are examples of suitable measuring location points.
ISO 13373-1 provides recommendations for measuring location points on various machine types.
13
Figure 9: Measuring Points on Vertical Bearings
Figure 10: Measuring Points on Flange Bearings
Figure 11: Measuring Points on Electric Motors
Figure 12: Measuring Points on Ma­chines with Vertical Rotors
9. Vibration Monitoring with Standard Limits
Deriving statements about the condition of a machine from monitoring vibration limit values requires some experience. If no specific values from previous measure­ment results are available, in many cases you can refer to the ISO 10816 recommen­dations. In these sections of the standard the vibration severity zone limits for vari­ous machine types are defined. The guidelines can be used for an initial evaluation of a machines condition. The four zone boundaries characterize the machine in cate ­gories according to vibration severity:
A: New Condition
B: Good condition for unrestricted continuous operation
C: Poor condition – permits restricted operation only
D: Critical condition – danger of machine damage
In the annex of part 1 of the ISO standard (revised in 2009) general zone boundaries are provided for machines which are not dealt with separately in other parts of the standard.
14
X
Y
Z
X
Y
Z
Z1
Z2
X2
Y1
Y2
X1
Z
X1
Y1
X2
Y2
X3
X4
Y3
Y4
v
eff
10 – 1000 Hz
45 mm/s
28 mm/s
18 mm/s
14.7 mm/s
Zone
C/D
4,5 – 14,7
mm/s
11.2 mm/s
9.3 mm/s
Zone
B/C
1,8 – 9,3
mm/s
7.1 mm/s
4.5 mm/s
Zone
A/B
0,71 – 4,5
mm/s
2.8 mm/s
1.8 mm/s
1.12 mm/s
0.71 mm/s
0.45 mm/s
0.28 mm/s
D Risk of machine damage
C Restricted operation
B Unrestricted long-term operation possible
A Newly commissioned
Table 1: Typical Limit Values for Vibration Severity to ISO 10816-1
The ISO standard indicates that small machines such as electric motors with a power rating of up to 15 kW tend to lie around the lower zone boundaries, whereas large machines such as motors with flexible foundations lie around the upper zone limits.
15
In part 3 of ISO 10816-2:2009 you will find the zone boundaries for vibration sever­ity on machines with a power rating of 15 kW bis 50 MW (2).
Machine type
Large machines with
300 kW to 50 MW
Medium sized ma-
chines with 15 to 300
kW
Electric motors with a
shaft height of above
315 mm
Electric motors with a
shaft height between
160 and 315 mm
Foundation flexible rigid flexible rigid
v
eff
10 – 1000 Hz
> 11 mm/s D D D D
> 7.1 mm/s C D D D
> 4.5 mm/s B C C D
> 3.5 mm/s B B B C
> 2.8 mm/s A B B C
> 2.3 mm/s A B B B
> 1.4 mm/s A A A B
< 1.4 mm/s A A A A
D Risk of machine damage
C Restricted operation
B Unrestricted long-term operation possible
A Newly commissioned
Table 2: Classification of Vibration Severity to ISO 10816-3
16
/
7
PCE-
PCE-
https://www.pce-instruments.com/english/download-win_4.htm
PCE-
PCE-
PCE-
PCE-
PCE-
PCE-
https://www.pce-instruments.com/english/download-win_4.htm
Click the “Load” button and enter the directory of the downloaded firmware file vs1x.hex. Then click “Send” to start the update process. The progress will be indi­cated by a bar graph. After a successful update the vibration switch will restart and “Firmware Updater” will be closed.
Please do not interrupt the update process. After update errors you can restart “Firmware Updater”.
19
13. Technical Data
Sensor
Piezoelectric accelerometer, inbuilt
Monitoring Modes
Real RMS and Peak Frequency analysis
Measuring Ranges Acceleration Velocity
0.01 –1000 m/s² Frequency dependent
Acceleration Filters
0.1-100; 0.1-200; 0.1-500; 0.1-1000; 0.1-2000; 0.1-5000;
0.1-10000; 2-100; 2-200; 2-500; 2-1000; 2-2000; 2­5000; 2-10000; 5-100; 5-200; 5-500: 5-1000; 5-2000; 5­5000; 5-10000; 10-100; 10-200; 10-500; 10-1000; 10­2000; 10-5000; 10-10000; 20-100; 20-200; 20-500; 20­1000; 20-2000; 20-5000; 20-10000; 50-200; 50-500; 50­1000; 50-2000; 50-5000; 50-10000; 100-500; 100-1000; 100-2000; 100-5000; 100-10000; 200-1000; 200-2000; 200-5000; 200-10000; 500-2000; 500-5000; 500-10000; 1000-5000; 1000-10000 Hz
Velocity Filters
2-1000; 5-1000; 10-1000 Hz
Frequency Analysis
360 line FFT; peak of acceleration Frequency ranges: 5-1000, 50-10000 Hz Update rate: 1/s; Window type: Hann
Teach-in Function (VS11)
For teaching in the alarm threshold, via button inside cas­ing
Relay Output
Via screw terminals inside the casing (PCE-VS11) or via 8 pin connecter Binder 711 (PCE-VS12) PhotoMOS relay; SPST; 60 V / 0.5 A (AC/DC); isolated switch mode (n.o./n.c.) and hold time programmable
Alarm Delay
0 – 99 s
Alarm Hold Time
0 – 9 s
Status Indicators
4 LEDs; green: OK; red/green: warning; red: Alarm
USB Interface
USB 2.0, full speed, CDC mode,
PCE-VS11: via micro USB socket inside casing PCE-VS12: via 8-in socket Binder 711 with Cable VM2x-USB
Power Supply
5 to 26 V DC (VS11) / < 100 mA or USB
Operating Temperature
-40 – 80 °C
Protection grade
IP67
Dimensions, Ø x h (without connectors)
50 mm x 52 mm (PCE-VS11); 50 mm x 36 mm (PCE-VS12)
Weight
160 g (PCE-VS11); 125 g (PCE-VS12)
20
Loading...