Instructions For Use
LED Videoscope PCE-VE 700
Visualization system for industrial endoscopy
PCE Instruments UK Ltd.
Units 12/13
Southpoint Business Park
Ensign way
Hampshire / Southampton
United Kingdom, SO31 4RF
From outside UK: +44
Tel: (0) 2380 98703 0
Fax: (0) 2380 98703 9
info@industrial-needs.com

TPA500-PCE-00
Version: 0
2016 - Feb - 17
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 2 of 28

Content
0 Important Information About this Document....................................................................... 5
0.1 Scope of Validity, Identification, Purpose...................................................................................... 5
0.2 Target Group.................................................................................................................................. 5
0.3 Using and Storing this Document.................................................................................................. 5
1 General Information About the Device................................................................................. 6
1.1 Scope of Delivery...........................................................................................................................6
1.2 Intended Use.................................................................................................................................. 6
1.3 Safety Information.......................................................................................................................... 6
1.4 Device Description..........................................................................................................................7
1.4.1 Performance Characteristics and Function........................................................................7
1.4.2 Visual Overview..................................................................................................................8
1.5 Marking......................................................................................................................................... 11
1.5.1 Pictograms and Information on the Device and Packaging.............................................11
1.5.2 Information on the Nameplate......................................................................................... 11
1.6 Service Department Contact Details............................................................................................11
2 Use...........................................................................................................................................12
2.1 Handling Endoscopic Probes....................................................................................................... 12
2.2 Installation and Connecting the Power Supply............................................................................ 12
2.3 Connection of a Monitor...............................................................................................................13
2.4 Connection of a PC......................................................................................................................13
2.5 Connection of the Camera Head................................................................................................. 14
2.6 Connection of an Endoscopic Probe........................................................................................... 14
2.7 Switching On and Off...................................................................................................................15
2.8 Setting of the Window Size..........................................................................................................15
2.9 Image Focus.................................................................................................................................15
2.10 Setting of the White Balance........................................................................................................16
2.11 Setting of the Light Amplification on the Screen..........................................................................16
2.12 Conducting of Inspections............................................................................................................ 17
3 Cleaning.................................................................................................................................. 18
3.1 Cleaning of the Controller............................................................................................................ 18
3.2 Cleaning of the Camera Head and Cable................................................................................... 18
3.3 Cleaning of Endoscopic Probes...................................................................................................19
4 Maintenance and Repairs..................................................................................................... 20
4.1 Changing the Fuse.......................................................................................................................20
4.2 Troubleshooting Table..................................................................................................................21
4.3 Repairs..........................................................................................................................................23
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 3 of 28

5 Product Data.......................................................................................................................... 24
5.1 Technical Data..............................................................................................................................24
5.1.1 Controller.......................................................................................................................... 24
5.1.2 Camera Head................................................................................................................... 24
5.2 Ambient Conditions...................................................................................................................... 25
5.2.1 Transport and Storage Conditions................................................................................... 25
5.2.2 Operating Conditions....................................................................................................... 25
5.3 Spare Parts and Accessories.......................................................................................................25
6 Disposal.................................................................................................................................. 26
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 4 of 28

0 Important Information About this Document
0.1 Scope of Validity, Identification, Purpose
0 Important Information About this Document
0.1 Scope of Validity, Identification, Purpose
These instructions for use apply to the following device:
Description: LED BORESCOPE
Item numbers: PCE-VE 700
These instructions for use are an integral component of the device and contain all
the information required by users and operators for safe and proper use.
0.2 Target Group
These instructions for use are intended for all persons who are entrusted with the
operation and use of the product.
0.3 Using and Storing this Document
These instructions for use must be stored in a defined location so that they may
be accessed at all times by the target group.
In the event of the sale of this device or its relocation, this document must be
handed over to the new owner.
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 5 of 28

1 General Information About the Device
1 General Information About the Device
1.1 Scope of Delivery
The scope of delivery for the device includes:
one camera head with connection cable
one controller (CCU)
one external power supply with mains cable
one USB connection cable
one composite video cable
one S-video connection cable
one documentation software CD (test version)
one transport and storage case
one set of instructions for use
1.1 Scope of Delivery
Check scope of
delivery!
Check the delivery against the delivery note for completeness and damage.
The delivery left our premises in perfect condition. If you have any cause for com-
plaint, however, please contact our Service Department.
1.2 Intended Use
The PCE-VE 700 camera system is used for endoscopic inspection of industrial
components, such as injectors or hydraulic parts.
1.3 Safety Information
Our products are developed and manufactured to the highest quality standards.
Risks – despite the
high quality
Although this device corresponds to the current state of the art, risks could still
arise during initial operation, use, cleaning or maintenance.
Therefore, it is important that you read through these instructions for use carefully
and observe the warnings indicated.
The device must be operated only in a fault-free condition in accordance with the
intended use and the instructions for use. Before each use, check that the device
and accessories to be used are free of damage and in full working order.
Keep the original packaging. Transport and store the product in its original pack-
aging and use it to return goods if service support is required.
Routinely inspect all electrical equipment. Ensure that the electrical parts and con-
nections are undamaged. Replace the defective components and cables if need-
ed.
WARNING! Unauthorized modifications to the device. Risk of serious injury to
persons. Do not make any unauthorized modifications.
WARNING! Operating the device in an open condition. Risk of electric shock. Do
not open the device. Operate only when closed.
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 6 of 28

WARNING! High-intensity light source. Risk of injury to eyes. Do not look directly
into the light exit of the endoscope or of the camera head.
WARNING! Electrical connections installed improperly. Risk of fire, short circuit,
or electric shock. Make sure that the electrical connections are installed in accor-
dance with the relevant national technical regulations.
1.4 Device Description
1.4.1 Performance Characteristics and Function
1 General Information About the Device
1.3 Safety Information
Inspections in industrial
areas
Wide selection of
probes
Different image trans-
mission systems
The PCE-VE 700 camera system is a modular endoscopic visualization system
for use in industrial areas. The system consists of a camera head with integrated
LED light source and a camera controller.
A wide selection of rigid, semi-flexible and flexible endoscope probes is available
in different lengths and diameters for the camera system.
The brightness of the durable and maintenance-free LED can be continuously ad-
justed. The light is directed through optical fibers to the probe tip and is emitted
there into inspection area.
The probes are equipped with rod lens systems or quartz optical fiber bundles.
Probes with rod lens systems are distinguished by a high image quality and are
well-suited for straight boreholes. Probes with quartz fiber bundles are more flexi-
ble and permit the inspection of curved and very thin channels.
The image transmission system directs the image to the image sensor in the cam-
era head that converts the optical image into electrical signals. The controller re-
ceives and renders them on the monitor.
A connected PC with appropriate software enables the capture of videos and stat-
ic images. An additional video output permits the connection of further peripherals
such as printers, recorders or frame grabbers.
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 7 of 28

1.4.2 Visual Overview
Camera Head
1 General Information About the Device
1.4 Device Description
Figure 1-1: Camera head.
Camera head
Focusing
Controller (CCU) connection cable
LED light source
FlexiLock Lock for interchangeable probes
Front glass panel
Protective cap
Connection plug
Connection cap
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 8 of 28

Controller - Front View
1 General Information About the Device
1.4 Device Description
Figure 1-2: Front view of the controller.
ON/OFF switch
Window function
Gain control
White balance
LED brightness control
Socket for the connection cable on the camera head
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 9 of 28

Controller - Rear View
1 General Information About the Device
1.4 Device Description
Figure 1-3: Rear View of the Controller.
Power supply connection (12 V DC)
Service interface
Video output (S-video)
Fuse holder with fuse
Video output (composite/FBAS)
PC output (USB)
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 10 of 28

1 General Information About the Device
1.5 Marking
1.5.1 Pictograms and Information on the Device and Packaging
This section describes the different pictograms that feature on the device or the
packaging.
CE mark
Permissible storage temperature
Permissible relative air humidity during storage
Permissible atmospheric pressure during storage
1.5.2 Information on the Nameplate
This section describes the different pictograms found on the nameplate located on
the underside of the device.
1.5 Marking
Adhere to the instructions for use
Item Number
Serial number
Manufacturer
Date of manufacture
Separate collection for WEEE (waste of electrical and electronic equipment)
CE mark
1.6 Service Department Contact Details
If you have any questions about our devices, their installation, or use, or you wish
to arrange servicing, please contact one of our subsidiaries or your dealer.
You will find contact details on the back of this document.
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 11 of 28

2 Use
2.1 Handling Endoscopic Probes
Please follow the following instructions for handling the product in order to avoid
damaging the endoscopic probes:
The probe shaft and tip are mounted in extremely thin metal or plastic
tubes, depending on the design and version you have. Always hold the
probe at its main body.
The probes' bending ability will vary depending on the design and the ma-
terials being used. Do not bend rigid probes. Do not bend flexible probes
any more than 3° within the first 10 mm away from the body of the probe.
Any more than that may damage it. Refer to the attached data sheet for
each probe to see whether it is rigid or flexible.
Never keep applying pressure to the tip of the probe if you encounter re-
sistance. The optical components could be scratched or damaged.
Move the probe gently during inspections. Damage often results from
careless movements, particularly if the probe undergoes some kind of im-
pact either when it is being inserted or being tilted when it is being with-
drawn.
2 Use
2.1 Handling Endoscopic Probes
Damaged probes can be repaired in many cases. In order to avoid making
the damage any worse, do not continue using damaged probes and have
them repaired.
2.2 Installation and Connecting the Power Supply
Set the controller up and connect it to the power supply.
NOTICE! Using a different power adapter can damage the product and possibly
invalidate the warranty. Only use the original power adapter that came with the
product.
Proceed as follows:
1. Check that the mains voltage at the site of operation corresponds to the power
supply specifications.
2. Install the device.
Install safely!
When installing the device, ensure that:
it is positioned horizontally on an even, non-slip surface that is able to
support its weight
it is protected from water drips and splashes
during operation it is not subjected to any vibrations and remains free of
any vibration
the power supply is sufficiently ventilated and not covered
it is outside the range of any areas where explosions could occur
it is not placed anywhere near flammable materials
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 12 of 28

2 Use
2.3 Connection of a Monitor
3. Connect the connection on the AC adapter to the 12V port on the rear of the
controller and secure it firmly.
4. Connect the power adapter to the power supply.
Route the cables safe-
5. Route the power cord so that it is not damaged and no one can trip over or get
ly!
caught in it.
Completed!
2.3 Connection of a Monitor
Connecting monitors
Proceed as follows:
1. Connect the S-VIDEO OUT connection to the monitor.
Use the supplied S-video cable.
If the monitor that is being used does not support S-Video, use the composite
output and the composite cable that came with the controller.
Note: The S-video connection provides the best image quality.
2. If necessary, adjust the monitor so that it will receive the PCE-VE 700 signal.
Follow the instructions supplied with the monitor.
Note: Both units must be switched on to adjust the monitor.
Completed!
2.4 Connection of a PC
Connecting PCs
If you want to record your inspections, connect the controller to a PC. An appro-
priate software must be installed on the PC to record videos and still images.
Proceed as follows:
1. Install the documentation software that you want to use on your PC.
2. Connect the USB 2.0 connection to the PC.
Use the supplied USB cable to do this.
Completed!
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 13 of 28

2.5 Connection of the Camera Head
Connecting camera heads
Proceed as follows:
1. Remove the connection cap from the connection plug.
2. Align the red mark on the connection plug to match the socket on the con-
troller.
Now slide the connection plug fully home into the socket on the device.
Completed!
2.6 Connection of an Endoscopic Probe
Connecting endoscopic probes
Select the endoscope probe in accordance with the object to be examined.
Proceed as follows:
1. Open the lock on the camera head and remove the protective cap.
2 Use
2.5 Connection of the Camera Head
Figure 2-1: Removing the protective cap on the camera head.
2. Check and if necessary, clean the glass surfaces on the head and on the
probe with an alcohol swab.
3. Insert the probe into the camera head and close the lock. Make sure that the
lock engages when closed.
Figure 2-2: Connecting the probe to the camera head.
4. Make sure that the probe does not move and is fixed firmly in place.
Re-align the probe if necessary.
Completed!
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 14 of 28

2.7 Switching On and Off
Switching on the device
Proceed as follows:
1. Press the ON switch on the front panel.
> The device is ready for immediate operation.
Completed!
Switching off the device
Proceed as follows:
1. Press the OFF switch on the front panel.
> The LED in the ON/OFF switch goes off.
Completed!
2.8 Setting of the Window Size
2 Use
2.7 Switching On and Off
Setting window sizes
The image size displayed on the monitor varies depending on the probe that is
used. Optimize the picture display by selecting the appropriate window size.
Proceed as follows:
1. Press the Windows button on the front of the controller.
2. Select a window by repeatedly pressing the button that is about the size of or
2.9 Image Focus
Adjusting the focus
The image must be brought back into focus each time the probe is changed, even
if the same probe is being used.
Proceed as follows:
1. Aim the endoscope on a sharply delineated object within the probe's working
> A window will appear on the screen.
smaller than the image shown on the monitor.
Completed!
distance.
2. Adjust the focus on the camera head until the image becomes sharp.
Completed!
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 15 of 28

2.10 Setting of the White Balance
Setting the white balance
Proceed as follows:
1. Attach the endoscope probe that you want to use for the inspection.
2. Switch the device on.
3. Aim the endoscope at a white surface that is within the respective probe's
working distance, but which is not more than max. 5 cm away, and bring the
image into focus.
4. Press the White balance button on the front of the controller.
> The message, White balance, will be displayed on the screen.
Wait until the message White balance OK appears on the screen.
Completed!
2.11 Setting of the Light Amplification on the Screen
2 Use
2.10 Setting of the White Balance
Setting the light amplification
Adapt the screen image to a dark environment by intensifying the monitor display.
Note: You will not notice any difference when adjusting the setting in a bright or
highly reflective environment.
Proceed as follows:
1. Press the GAIN button on the front of the controller.
> The display on the screen will intensify.
> The current setting will display on the screen ("ACG 08 dB").
> The LED light amplification that you have set will remain unchanged.
2. Keep pressing the GAIN to continue amplifying the amount of light shown on
the screen.
> The current setting will display on the screen ("ACG 12 dB").
Pressing the button one more time will set your light amplification on the
screen back to 0 dB.
Completed!
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 16 of 28

2.12 Conducting of Inspections
Conducting inspections
NOTICE! Sensitive optical, mechanical and electronic components. Damage
resulting from rough handling. Handle the device with appropriate care. Do not
bump or drop the product. Do not kink or pinch the probe; do not press the tip of
the probe against or past any immovable object.
Proceed as follows:
2 Use
2.12 Conducting of Inspections
Selecting the probe
Switching on the de-
vice
Setting window sizes
Focus the image
White balance
Conducting inspections
1. Choose a probe that is right for the purpose for which you intend to use it, and
which is suitable for the object you want to examine, and attach the probe to
the camera head.
2. Switch the controller and the monitor on.
If you want to record the examination, then turn the PC that the probe is con-
nected to on and launch the documentation software.
3. Set the window size when you see the endoscopic image on the monitor (see
section 2.8).
4. Bring the endoscopic image into focus (see section 2.9).
5. Set the white balance (see section 2.10)
6. Carefully insert the probe into the object you want to examine and begin your
inspection.
Regulate the LED light intensity by rotating the regulator on the controller.
If the image displayed on the screen appears too bright or too dark, then
optimize the image using the light intensifier on the computer screen (see
section 2.11).
Cleaning the probe
7. After conducting your inspection, carefully pull the probe out of the object with-
out tilting the probe.
8. Disconnect the probe from the camera head, place the protective cap back on
the head and close the lock.
9. Clean the probe with a soft cloth. If necessary, use a mild cleaning solution
(see section 3.3).
10. Switch off the devices.
Completed!
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 17 of 28

3 Cleaning
3.1 Cleaning of the Controller
Cleaning the controller
NOTICE! The controller should only be wiped down using a damp cloth. Excess
cleaning solution can come into contact with the electrical parts and damage the
controller. Do not use wet sponges or cloths. Do not immerse the device in any
fluid.
Proceed as follows:
1. Switch the controller off and disconnect it from the power supply.
2. Cleaning the device.
Wipe the external surfaces with a soft wipe. Use a slightly dampened cloth
that has been moistened with water, a mild soap solution or isopropyl alcohol.
Stubborn stains can be removed with a mild, ammonia-based detergent.
3 Cleaning
3.1 Cleaning of the Controller
Do not use abrasives or solvents, since such agents may damage the paint
and the labeling on the front panel.
3. Do not turn the controller back on until it is completely dry.
Completed!
3.2 Cleaning of the Camera Head and Cable
Cleaning the camera head
NOTICE! Do not immerse the camera plug in any fluid.
Proceed as follows:
1. Disconnect the camera cable from the controller, and use the protective cap to
cover the connection plug.
2. Clean the components.
Wipe the external surfaces with a soft wipe. Use a mild cleaning agent. Do not
use any abrasive materials or solvents.
Remove any debris or soiling from optical surfaces using a swab dipped in al-
cohol (70 % ethanol).
3. Dry the components with a soft cloth.
Connect the device to the controller only after it is completely dry.
Completed!
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 18 of 28

3.3 Cleaning of Endoscopic Probes
Cleaning the probe
NOTICE! Sensitive optical, mechanical and electronic components. Damage
resulting from rough handling. Handle the device with appropriate care. Do not
bump or drop the product. Do not kink or pinch the probe; do not press the tip of
the probe against or past any immovable object.
Proceed as follows:
1. Clean the probe with a soft cloth that has been slightly moistened with a
cleaning solution. Do not use any metal brushes, abrasives or solvents.
When using cleaning agents, be sure to follow manufacturer specifications
and also make sure that the cleaning substances are suitable for using on the
probe.
Remove any debris or soiling from optical surfaces using a swab dipped in al-
cohol (70 % ethanol).
2. Use clear water to rinse the rest of the cleaning agent off.
3 Cleaning
3.3 Cleaning of Endoscopic Probes
3. Dry the components with a soft cloth.
Completed!
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 19 of 28

4 Maintenance and Repairs
4.1 Changing the Fuse
Fuse replacement
Proceed as follows:
1. Isolate the device from its power supply and disconnect the power cord.
NOTICE! Pulling the power cable will damage it. Disconnect the power at the
plug.
2. Detach the fuse holder with a counterclockwise quarter turn. Use a screwdriv-
er or a coin.
3. Carefully remove the fuse holder.
WARNING! Risk of fire due to unsuitable electrical fuse. Use the specified
fuse only and never bypass the fuse.
4. Replace the faulty fuse.
5. Push the fuse holder fully into the shaft and press slightly while screwing it
one-quarter clockwise turn.
4 Maintenance and Repairs
4.1 Changing the Fuse
Completed!
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 20 of 28

4.2 Troubleshooting Table
Issue Possible Causes Fixes
4 Maintenance and Repairs
4.2 Troubleshooting Table
Image too dark, too little illumina-
tion
No image on the monitor
Glass surfaces on camera head
are soiled
Light output set too dark Increase the light output at the
White balance not set correctly Set the white balance as de-
No power supply Connect and switch on the de-
Fuse defective Replace the fuse (see Mainte-
Monitor connected incorrectly to
the controller
Connection cable not connected
or defective
Clean the glass surfaces (see
Cleaning chapter)
light source
scribed in the Use chapter.
vices (see Use chapter)
nance and Repairs chapter)
Connect monitor to the controller
correctly (see Use chapter)
Connect the camera head to the
controller. Check the plug for
moisture. If the cable is defec-
tive, send it in for repair
Monitor not tuned in to the cor-
rect input signal
No light on the probe tip
Unknown Contact manufacturer
Monitor not configured correctly Check monitor settingsPoor color rendering
White balance not set correctly Set the white balance as de-
Colored lines through the image Defective video cable Replace the video cable or send
Image blurred
Image out of focus Adjust the focus for a sharp im-
Probe tip soiled Clean the probe tip as described
Set monitor for the correct input
signal
Ensure that the probe is connect-
ed correctly to the camera head
Increase the light intensity on the
controller
scribed in the Use chapter.
the product for repair
age on the camera head
in the Cleaning chapter
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 21 of 28

Issue Possible Causes Fixes
4 Maintenance and Repairs
4.2 Troubleshooting Table
No operation No power supply
Table 4-1: Troubleshooting table.
Check the mains connection and
connect the power if necessary
Check the fuse and replace if
necessary
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 22 of 28

4.3 Repairs
4 Maintenance and Repairs
4.3 Repairs
Should you need to arrange a repair for the device, please contact one of our
branch offices. Contact details can be found on the back of these instructions.
When you send in equipment, please enclose as accurate a fault description as
you can, and note down on the delivery slip the part number and serial number of
the product.
Send only thoroughly cleaned goods for repair. The device should be returned in
its original packaging, if possible.
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 23 of 28

5 Product Data
5.1 Technical Data
5.1.1 Controller
Dimensions (W x H x D) 169 x 62 x 247 mm
Weight 1.2 kg
Power consumption 10 VA
Power supply GlobTek®, Inc., GTM21089-1512-T3
Fuse T 800, 250 V, 5 x 20 mm
Brightness control automatic
Signal noise ratio (SNR) 46 dB
Outputs
Protection class II
5 Product Data
5.1 Technical Data
Input: Power supply 100-240 V AC 50-60 Hz;
max. 0.5 A
Output: 12 V DC; max. 1.25 A
S-Video (Y/C)
Composite video (FBAS)
USB 2.0
5.1.2 Camera Head
Weight 0.2 kg
Length/connection cable 2.3 m
Image sensor 1/4" CCD
TV systems PAL
Active image components 752 (H) x 582 (V)
Resolution (H) 625 lines
Scanning system 2:1 line skip
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 24 of 28

5.2 Ambient Conditions
5.2.1 Transport and Storage Conditions
Store the device in a dry, well-ventilated, temperature-controlled room, where it is
protected from dust.
When placing into storage or storing temporarily, make sure that the device can-
not be damaged by its surroundings. Therefore, the product is best stored either
separately or in a carrying case.
To avoid damage, keep out of direct sunlight and away from high temperatures,
high humidity and radiation.
Temperature 0 °C to +50 °C
Relative humidity 30 % to 90 %
Atmospheric pressure 70 kPa to 106 kPa
5.2.2 Operating Conditions
Temperature +10 °C to +40 °C
Relative humidity 30 % to 90 %
Atmospheric pressure 70 kPa to 106 kPa
5 Product Data
5.2 Ambient Conditions
5.3 Spare Parts and Accessories
Use original spare parts and accessories only.
Designation Item Number
EU power cord 05.0700R.CORD.EU
Power cord, country-specific Item number on request
Fuse 0.8 A (T) 5 x 20 mm 94605.
Table 5-1: Spare parts and accessories.
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 25 of 28

6 Disposal
6 Disposal
When disposing of or recycling the device and its components, you must adhere
to the applicable national regulations governing waste disposal and recycling.
Any product carrying this symbol must be disposed of separately through dedicat-
ed electrical and electronic devices recycling. Within the EU, such disposal is tak-
en care of by the manufacturer free of charge.
Instructions For Use LED BORESCOPE PCE-VE 700 | 17 - Feb - 2016 | Version: 0 Page 26 of 28

PCE Deutschland GmbH
Im Langel 4
59872 Meschede
Germany
Telefon: +49 (0) 2903 976 99 0
Fax: +49 (0) 2903 976 99 29
E-Mail: info@pce-instruments.com