PCE Instruments PCE-SW 3000N, PCE-SW 5000N, PCE-SW 1500N Users guide

Page 1
PCE-SW N Series
Scales and Weighing Beams
USER MANUAL
Edition:10052001A
www.pce-instruments.com
Page 2
Safety Instruction
For safety operation pls. follow the safety instruction.
WARNING
WARNING
Pls. make sure the weighing display have good ground in using
WARNING
The indicator is the static and sensitive equipment, cut off the power during electrical connections, internal components touched by hand is prohibited, and please take the measure of anti-static.
LIST
Page 3
1. Instruction ............................................................................................. 1
1.1 Main function ................................................................................. 1
1.2 Technical parameter ........................................................................ 1
1.3 Drawing .......................................................................................... 2
1.4 Battery instruction ........................................................................... 3
2. Installation and calibration .................................................................... 4
2.1 Power supply connection .............................................................. 4
2.2 Loadcell connection ....................................................................... 4
2.3 communcation interface ................................................................ 5
2.4 4~20mA ......................................................................................... 6
2.5 Relay output signal ........................................................................ 6
3. Basic operation ....................................................................................... 9
3.1 Keyboards...................................................................................... 9
3.2 Power on....................................................................................... 11
3.3 Zero function ................................................................................. 11
3.4 Tare function ................................................................................. 11
3.5 Total function ................................................................................. 11
3.6 Print function ................................................................................ 13
3.7 Hold function ................................................................................ 13
4. Calibartion and Technical parameter setting ........................................ 13
4.1 Enter calibartion ........................................................................... 13
4.2 Caliaration.................................................................................... 14
4.3 Technical parameter setting......................................................... 16
5. Output format ........................................................................................ 22
5.1 Second display continuous output .............................................. 22
5.2 Conputer continous output .......................................................... 22
5.3 Serial interface receive PC command ......................................... 23
5.4 Print output format ....................................................................... 24
5.5 PC/ Second display continuous output format ............................ 24
6. Maintain ................................................................................................. 25
6.1Toubleshooting for common problems ......................................... 25
6.2 Daily maintenance ........................................................................ 26
6.3 Default parameter recovery ......................................................... 26
6.4 Packing list................................................................................... 29
Page 4
1. Instruction
This weighing indicator is designed for bench scale. Floor scale, the basic weighing function include: Hold. Print kg/lb conversion . optional: I/O , 4-20mA output.
1.1 Main function
Weighing function:
Zero, tare, G.W, N.W, accumulation. printing, animal -weighing. kg/lb convert. Overload remind. Print format: S.N. G.W N.W Tare. Date, Time
Options:
Pinter RS232/RS485 serial interface or second display I/O 4-20mA
1.2 technical parameter
Accuracy class 6000 e Resolution display: 30, 000 ADC: 2,000,000 Zero stability error TK Span stability error TK Sensitivity (internal) 0. 3 μV /d Input voltage -30~30mV DC Excitation circuit 5 VDC, 4 wire connection,
AC power AC100~250V
Operation temperature - 10 °C ~ + 40 °C Operation humidity ≤90%RH Storage temperature - 40 °C ~ + 70 °C
< 0.1μV//K
0
< ± 6 ppm//K
spn
Maximum connect 6 load cell of 350Ω
1
Page 5
1.3 Outlinemm
2
Page 6
1.4 Battery instruction
1. . when you use the internal battery first time,you should charge the battery fully, to prevent low voltage resulted from self leakage of battery.
2. when the battery light is on, means low battery, pls. charge it
in time
3. the light turns to yellow during charging
4. when the light turns to green, means fully charged.
5. if battery is not used for long time, take it out to avoid the
leakage.
6. In order to keep the battery in best using condition, it is suggest
that you fully discharge the battery every month, the method is that using the indicator till it is automatically power off.
3
Page 7
2.Installation and calibration
2.1 Power supply connection
The indicator is powered by adapter, you plug the adapter directly into
the DC pin at the back cover the indicator is ok.
2.2 Connection of load cell and indicator
The indicator can connect with 6 load cell of 350Ω at most, 4 wire or 6 wire load cell both ok.
There are two methods connection between load cell and indicator
A. quick disconnect, as below:
B. Terminal trip connection (inner connection)
1. the exciting voltage for the load cell is 5VDC, the largest output
current 120mA, maximum connect 6 pcs 350Ω load cell;
2. Load cell (or the signal cable for the junction box) is connected with 5 bit Terminal trip ( J5) on the circuit-board of weighing indicator.
3. Open Weighing indicator back cover, insert signal cable to the terminal trip(J5), and make sure the screw is fixed tightly, the connection as below:
4
Page 8
Weighing display Load cell
+EXC + voltage +IN + output signal HD ----------- Shielded wire
-IN - output signal
-EXC -IN HD +IN +EXC -EXC - voltage
2.3 Communication interface
RS232 : DB9 Pin or 3 Pin
DB9 definition
5 4 3 2 1
9 8 7 6
Pin function and definition as bellows:
DB9 joint Definition Function
2 TXD Sending data 3 RXD Receiving data 5 GND Ground interface
Note: if RS485, The connection pin is 2 and 5 pin.
3 Pin definition
5
Page 9
Inner connection
TXD RXD GND A B
Pin definitions
Pins Definitions Function
TXD Sending data
RS232
RS485
RXD Receiving data
GND Ground
A RS485 output Aport
B RS485 outputBport
2.4 4-20mA output
Technical parameters:
Resolutions: 1/1000 Outside Load: 100-350Ω
Connection:
1. Inside connection: 4-20mA, load input port connect with I of J2, Ground port
connect with GND of J2
2. Outside connection: 4-20mA, load input port connect with 1 pin of D89,
Ground port connect with 6 pin of DB9
Testing :
6
Page 10
Connect the 250Ω to the 4-20mA, adjust the multimeter to the current stalls, the red pan connect the loading port, and the black pan connect GND to test the output current.
Calibration:
1. Press Print and Total go to C32, show[out-4], the output current should be
4mA.
2. If Press []show[out-5] Pres[↓]show[out-20]the output current should be 20mA.
3. Adjust the current, for example, adjust to [out-20], the output current should be 12mA, if not 12mA. Press [](Zero key)or[](Net/gross key)to adjust the current to 12mA
2.5 Relay output signal function
The indicator can output 4 signal , connect with the outside equipment the indicator can perform automatic control function and upper limit and lower limit alarm function. Perform the 4 kinds function through setting C33, 4 signals
As below
C330
C331
Output
port
Out1 Close output function No output signal Out2 Close output function No output signal Out3 Close output function No output signal Out4 Close output function No output signal Out1 Open overload control function Output overload
Out2 Open compliance control function Output compliance
Out3 Open underload control function Output underload
Port definition Function
control signal
control signal
control signal
7
Page 11
Out4 Open stable control function Output stable control
signal
Out1 Weight>=C13 instant connection Feeding control
signal
Out2 Weight<=C14 instant connection Feeding control
signal
C332
C333 Preserved, no function.
Out3 OUT1OUT2 close When OUT1OUT2
connect will output alarm signal, connect with buzzer
Out4 Weight<=C14 constant connection
Weight>=C13 disconnect
Feeding start and stop
For example: Check weigher application. Connect indicator with yellow, green, red 3
lights. Yellow light on when overload, if ok the green light on. If underload red light on. And can connect with buzzer. There would be alarm remind when overload.
C33=2 Feeding control application For hopper scales. The target is between is 50~500kg, the Output
card can perform open and close feeding valve. Parameter setting C13=500kg C14=50kg C13=500kg upper limit , close feeding valve C14=50kg lower limit, open feeding valve. Connection: 1. Connect directly control machine, pls. connect OUT 4,
suggest connect another manually switch as the stop switch in emergency
2. indicator output signal will influence the ON/OFF switch to control
machine, OUT 2 ON, OUT 1 OFF, OUT 3 connect buzzer, alarm remind when start and stop.
Instruction: when the weight over 500kg, OUT 1 OU3 instant connection and break off, OUT 4 break off; when weight under
8
Page 12
50kg, OUT 2 OUT3 instant connection and break off, OUT 4 constant connection.
5 4 3 2 1 9 8 7 6
DB9 pin definition port
1 pin 1st output signal pin Out1 6 pin 1st output signal pin Out1 2 pin 2nd output signal pin Out2 7 pin 2nd output signal pin Out2 3 pin 3rd output signal pin Out3 8 pin 3rd output signal pin Out3 4 pin 4th output signal pin Out4 9 pin 4th output signal pin Out4
Inner connection pin definitions
01 01 02 02 03 03 04 04
9
Page 13
3. Basic operation
3.1 keypad
LED instruction
Weighing data
kg
lb
kg lb
10
Page 14
Hold
Data hold
Gross
Net
Tare
Hi
OK
Lo
.
PCS
TOTAL
Gross weight Net weight tare The weighing data is stable Weight is zero Overload
ok Underload
Decimal Show the counting status.
Go to accumulation mode
Keys function
keys Key name Key function
1.work with ZERO TARE ON/OFF key
Print
Zero
Tare
Gross weight
to perform many functions. 2.Print Zero the weight within tolerance
At G.W mode, get the tare weight. At N.W mode, clear the tare, get the G.W
At N.W mode, check the G.W, after 3 seconds back to N.W automatically
11
Page 15
Counting operation
Counting
Hold Freezes displayed weight on screen
1. Accumulation
Accumulation
2. work together with Print to perform The accumulation function and check the accumulation result
Power on/off
3.2 Power on & off
Press 2 seconds to power on or power off, after power on the indicator show000000-999999. After self inspection. It go to the weighing
mode. Pls. check it whether 6 bits LED/LCD display and the status light is good or not.
3.3 Zero operation
1. Initial zero setting When power on the indicator, if the weight on the scale is within the
initial zero tolerance, indicator show zero automatically.
2. Manually Zero setting When the scales is stable, and not the negative display, you can zero
Press 2 seconds to power on or power off
the weight within tolerance by press keys.
3.4 Tare operation
12
Page 16
Press TARE key, the gross weight is tared, indicator show the Net weight, the Net tared status light is on. At tare mode, Press TARE key, clear the tare weight, the indicator will show the gross weight.
3.5 Accumulation operation
At Zero mode, load weight till stable, Press go to accumulation
mode, Total light on, display n 001, and then display loaded weight; unload the weight , back to zero, load the second weight again till stable. Press
displayn002 then display the second loaded weight. Repeat it agin
and again, maximum 999 times.
Check the accumulation
Press ON/OFF key and hold it then press TOTAL”” key, display n**, (it is the accumulating times) then show total weight. there are 8 digits totally. It shows the first 4 digits then the last 4 digits. For example, the first 4 digits is0012, the last 4 digits is34,56 It means the actual weight is 1234.56
EXIT the accumulation function
When the indicator show the last 4 digits, Press hold it, the indicator show clr n”, it means dont clear the total Weight, Press PRINT
key to exit it; if you want to clear total weight, Press ZERO or TARE key, clrn change to clry it means clear total weight ,then Press PRINT to clear the total weight and exit accumulating mode.
3.6 Print
If the weighing is stable, after connect with printer, press PRINT can print the weight. Note: at tare mode, print with tare. if negative weight,, can not print. Set C30 for time format.
13
Page 17
3.7 Hold
There are two different hold function. Peak hold function and data hold function. And the setting is different accordingly.
C11=3 Automatic hold function C11=0 close hold function.
3.8 COUNT
1.At weighing mode, load the weights on the platform scales, Press Count the indicator show PCS 0 press Zero key input the quantity, press Print to confirm it 2 . Load the goods on the platform scales, then the indicator will show the quantity.
3. Press Count back to weighing mode.
4. If you want to weigh different goods, at weighing mode, put the
sample on the platform scales, press Count the indicator show0 Press Count hold it and then press ON/OFF the indicator show PCS 0”, press Zero input the sample quantity, press Print to confirm it. Then repeat the step 2 and 3.
4. Calibration and Parameter setting
4.1 Enter setting
There have two methods to enter the setting menu:
1. when the switch CAL is off, press the PRINT hold it and
then press HOLD enter C08-39 setting.
2. Take out the sealing screw on the back of indicator, then press CAL , at the SPAN position as below. press PRINT” hold it and then press TOTAL key , enter C01-C39 setting.
14
Page 18
The key functions in setting
Enter
Up
Down
Left
Power switch. exit setting
4.2. Step of calibration operation:
According to the second method which can enter setting menu, C01-C39
step
Method of operation
display Remark
1 [C01 ] After you enter calibration
mode, it display [C01 ]
2 press [C01 1] Weight unit
option1=kg
2=lb
3 press
press press or
[C02 ] [C02 0] [C02 2]
15
Set decimal digits option0/1/2/3/4 Select decimal digit
Page 19
4 press
press press or
5 press
press press or /
press press press press
6
7 press
press
press or
press
press or press
[C03 ] [C03 1] [C03 5]
[C04 ] [0100.00] [0100.00]
[C05 ] [C05 0] [C05 1] [CAL 9] 。。。。。。 [0000.00]
[C06 ] [C06 0]
[C06 1]
[SPAN ] [0100.00]
[0080.00] [CAL 9] …… [0080.00] [CAL End]
exampletwo decimal point [C02 2] Set graduation option1/2/5/10/20/50 Select required graduation example:graduation 5:[C03 5] Max capacity
examplemax weighing 100kg: [0100.00] Zero calibration
Option 0=no need zero calibration 1=need zero calibration calibration zero please choose 1 and ensure scale is empty and stable light is on Ensure zero calibration, countdown. Till show[0.00](example for two decimal point)
calibration option: 0=No need calibration 1= need calibration Load weights on scales according to max. capacity. Suggest close to the max capacity, at least 10% of max. capacity.
For example: the weights is 80kg As bellows: Input the 0080.00, count down ,
16
Page 20
then indicator shows 0080.00 , calibration is over. If you want to set application function parameter. Press PRINT if you want to exit press TOTAL
8
press press press or
[C07 ]
[07 0] [07 1]
Default parameters setting option:0=non-restore default parameters 1=restore default parameters Note: after the above parameters setting finish, please do not set default parameters to avoid the original setting parameters is lost.
4.3 Application function parameters setting chart
Function
warning
tone
Setting Item C08
warning tone
parameters setting and instruction
Options: 0 = close warning tone
1 = open warning tone
option0=close auto power off 10= power off automatically if no
Automatic
power off
C09
Automatic power off
30= power off automatically if no
change within 10 minute.
change within 30 minute. 60= power off automatically if no change within 60 minute.
LED Version:
Power saving
setting
C10
Power saving setting
option0= close power saving setting 3= close display if no change within 3min.
5= close display if no change within 5 min. LCD Version:
17
Page 21
Hold
function
Kg/lb
conversion
Upper/lower
limit alarm
Inner Code
display
C11
Hold mode
C12
Kg/lb conversion
C13
Upper limit alarm value
C14
Lower limit
alarm value
C15
Check inner code
0=Close he backlight 1= backlight when the weight change or
press the keyboard
2=constant backlight
option0=close hold function
1=Peak hold /2=Data Hold Instruction: Peak-hold: it shows the max. data,
mainly application for materials testing, such as tension and pulling force.
Date-hold: it shows current weight value.
Mainly application for animal weighing.
C12=0 stop kg/lb conversion C12=1 kg/lb conversion is ok
You can set it within the max. capacity limit
enter C15 to check the inner code
Date and time
Communication
setting
C16
Date
C17
Time
C18
Serial interface data output method
18
Enter C16, you can set the date, from left to right: year/month/day Enter C17, you can set the time from left to right: hour/min./sec. option:0= Close serial interface data output 1=Continuous sending, connect second display
Page 22
Zero range
C19
Baud rate
C20
Manually zero range
C21
Initial zero range
C22
Automatically zero tracking range
2=Print method, connect printer. 3=Command request method connect computer. 4=PC continues sending format, connect computer. 5=PC/ second display continuous sending format. option 0=1200/1=2400/2=4800/3=9600 Option: 0= close manually zero setting 1=±1% max capacity 2=±2% max capacity 4=±4% max capacity 10=±10% max capacity 20=±20% max capacity 100=±100% max capacity option0= no initial zero setting
1=±1% max capacity 2=±1% max capacity
5=±1% max capacity 10=±1% max capacity 20=±1% max capacity
Options: 0= close zero tracking
0.5=±0.5d
1.0=±1.0d
2.0=±2.0d
3.0=±3.0d
4.0=±4.0d
5.0=±5.0d
Zero tracking
Note: 1. d = division
2. the zero tracking range can
19
Page 23
not bigger than manual zero range.
00= close overload range
Overload range
Negative display
Standstill time
C23
Automatically zero tracking time
C24
Overload range
C25
Negative display range
C26
Standstill time
C27
Standstill range
Options: 0= close zero tracking time 1= 1 second 2= 2 seconds 3= 3 seconds
option 01d99d remarkd =division
Option0=-9d 10=10% max. capacity 20=20% max. capacity 50=50% max. capacity 100=100% max. capacity Option: 0= quick 1= medium 2= slow
Option: 1= 1d 2=2d 5=5d 10=10d D= division
20
Page 24
Digital filter
C28
Dynamic filter Instruction Dynamic filter is collecting the data filter before loaded weight stable. When loaded weight easily shaking (for example animal) , you can set this filter to make weight display more stable
C29
Noise filter
option0= close dynamic filter 1=1 digital filter strength 2=2 digital filter strength 3=3 digital filter strength 4=4 digital filter strength 5=5 digital filter strength 6=6 digital filter strength Note Pls setting dynamic filter
strength carefully, the No. is bigger, more stable. if the loaded weight shake not too much. The setting is less than 3
option0=close noise filter 1=1 digital filter strength 2=2 digital filter strength 3=3 digital filter strength
Analog output setting
4~20mA current calibrate
Relay output setting
C30
Print time and date
C30=0 yy.mm.dd C30=1 mm.dd.yy C30=2 dd.mm.yy C30=3 yy.mm.dd
C31 output type C31=0 0~5Vouput
C31=1 4~20mA output
C32 calibrate
Refer to 2.5
current C33 Relay output C33=0 close relay output
C33=1 Open relay output function 1 C3=2 Open relay output function2 C33=3 Preserved menu
21
Page 25
Muti communication add. Wireless
C34 Communication add.
C35 C35=0~99 signal communication Gravity of
C36 C36=9.7000~9.9999 calibration location Gravity of
C37 C37=9.7000~9.9999 destination Version No. C38 Preserved menu C39
C34= 0~99 Add. Code
22
Page 26
5. Output format
T X
W C X X X X X X X X X X X X C R C
5.1 Second display continuous sending format
Output continuous format
S
S
S
S
W
W
A
B
1
2 3 4 5 6
State A
Bits0,1,2 0 1 2 Decimal point position 1 0 0 XXXXXX0 0 1 0 XXXXXXX 1 1 0 0 0 1 1 0 1
Bits3,4 Division 0 1 X1 1 0 X2
K S
XXXXXX XXXXXX XXXXXX
State B
BitsS function Bits0 gross=0, net=1 Bits1 Symbol: positive =0,negative =1 Bits2 Overload(or under zero)=1 Bits3 dynamic=1
Bits4 Bits5 Constant 1 Bits6 Constant 0
23
unitlb=0, kg=1
Page 27
State C
,
Bit2 Bit1 Bit0 unit
0 0 0 Kg or lb 0 0 1 g 0 1 0 t
Bit 3 printing=1
Bit 4
Bit 5 Constant 1 Bit 6 Constant 0
5.2 Computer continuous sending format
Extend
display=1
,
CR LF
S 1 S 2 S 3 Data S 4
S1: weight status, ST= standstill, US= not standstill, OL= overload S2: weight mode, GS=gross mode, NT=net mode S3: weight of positive and negative, +” or –“ S4: kg” or lb Data: weight value, including decimal point CR: carriage return LF: line feed
5.3 Serial interface reception command
RS232COM serial interface can receive simple ASCII command. Command word and role as follows:
Command
T TARE Save and clear tare Z ZERO Zero gross weight
NAME Function
24
Page 28
P PRINT Print the weight R G.W/N.W Read gross weight or net weight C Kg/lb Kg/lb conversion G G.W Check gross weight at net weight mode
R command receive data format
5.4 Print format
ID.NO. 004 Serial No. Date: XX.XX. XX (yy.mm.dd) Time: XX.XX.XX (hh.mm.ss) GROSS 8.88kg (gross weight) TARE 2.88kg tare NET 6.00kg net weight
5.5 PC or Second display continuous sending format
25
Page 29
6. Maintenance
6.1 Regular error and solution
ERROR REASON SOLUTION
UUUUUU
1. Overload
2. wrong connection with load cell
3. load cell has quality problem.
1. reduce the weight
2. check load cell connection
3. inspection load cell. Check the input and output
nnnnnnn
ERR1
ERR2
ERR3
1. calibration is no good
2. wrong connection
3. load cell has quality problem
During calibration, not input the weights or the weight is overload
During calibration , the weights is below than Min. required weights
During calibration, the input signal is negative
26
1. check scale is resisted or not, foot is kept level or not.
2. check load cell connection.
3. checking load cellcheck
input and output resistance to judge it is good or not. Input the correct weights
The calibration weights Minimum is 10% of Max. cap. Recommend 60%-80% of Max. Cap.
1. check the connection is
correct
2. check load cell is no
problem
3. recalibration if still wrong
change the PCB
Page 30
ERR4
During calibration, the signal is unstable
After the platform is stable, start calibration
ERR5 Change PCB
6.2 Daily maintain
1. Protect the indicator from strong sunlight to prolong the using life
2. Good connection between load cell and indicator. Far from away
from strong electric field, magnetic field.
3. Power off the indicator when lightning
4. Power off the indicator firstly before plug and unplug
6.3 Restore default parameter
Enter to calibration, Set C07=1. Press PRINT then press TOTAL to
exit saving setting. All parameter will back to default
Note: Pls. do not restore default parameter easily if you are not
professional staff or not yet calibrate the scales.
Default parameter
Parameter
C01 C02 C03 C04 C05 C06 C07 C08 C09 C10
instruction Default
Calibration Decimal digits Resolution Max. capacity Empty calibration Capacity calibration Restore default Warning tone Power-off automatically Power saving mode
1 0 1
10000
0 0 0 1 0 0
27
Page 31
C11
Hold function
0 C12 Prohibit kg/lb conversion C13 Upper limit alarm 000000 C14 Under limit alarm 000000 C15 Inner code C16 Date setting C17 Time setting C18 Serial interface data output 0 C19 Serial interface Baud rate 39600 C20 Zero manually 10 C21 Initial zero 10 C22 Zero tracking range 05 C23 Zero tracking time 1 C24 Overload range 9 C25 Negative range 10 C26 Standstill time 1 C27 Standstill range 2 C28 Dynamic filter 0 C29 Noisy filter 2 C30 Print format 0 C31 Analog signal options 1 C32 4~20mA testing 4 C33 Relay output setting 1 C34 Muti PC communication add. 0 C35
Wireless communication channel
6 C36 Calibration location gravity 9.7936 C37 Destination gravity 9.7936 C38 Version No. check C39 Reserved menu
28
Page 32
6.4 Packing list
Packing list
S/N ITEM NAME UNIT QTY PACKING
Weighing
1
indicator
2 Plastic bag PCS 1 3 bag PCS 1
China/DC9V PCS 1 US/DC9V PCS 1
4 Adapter
5 USER MANUAL 6 RS232 3 PIN OR DB9 PCS 1
LOADCELL
7
PLUG Signal cable Φ5/3 core shield
8
9 Power cable 3 coreΦ0.75mm PCS 1
10 Bracket Wall-mounted PCS 1
11 Certificate PCS 1
12 Packing list PCS 1
UK/DC/9V PCS 1 EU/DC9V PCS 1 AU/DC9V PCS 1 OTHERS PCS 1
5 PIN Quick disconnect
signal cable
PCS
PCS 1
PCS
PCS
1
1
1