For safety operation pls. follow the safety instruction.
WARNING
Setting. Calibration Inspection and Maintain of the indicator is
prohibited by Non-professional staff.
WARNING
Pls. make sure the weighing display have good ground in using
WARNING
The indicator is the static and sensitive equipment,
cut off the power during electrical connections,
internal components touched by hand is prohibited,
and please take the measure of anti-static.
This weighing indicator is designed for bench scale. Floor scale, the basic
weighing function include: Hold. Print kg/lb conversion . optional: I/O , 4-20mA
output.
Pinter
RS232/RS485 serial interface or second display
I/O
4-20mA
1.2 technical parameter
Accuracy class 6000 e
Resolution display: 30, 000 ADC: 2,000,000
Zero stability error TK
Span stability error TK
Sensitivity (internal) 0. 3 μV /d
Input voltage -30~30mV DC
Excitation circuit 5 VDC, 4 wire connection,
AC powerAC100~250V
Operation temperature - 10 °C ~ + 40 °C
Operation humidity ≤90%RH
Storage temperature - 40 °C ~ + 70 °C
< 0.1μV//K
0
< ± 6 ppm//K
spn
Maximum connect 6 load cell of 350Ω
1
Page 5
1.3 Outline(mm)
2
Page 6
1.4 Battery instruction
1. . when you use the internal battery first time,you should charge
the battery fully, to prevent low voltage resulted from self leakage
of battery.
2. when the “battery” light is on, means low battery, pls. charge it
in time
3. the light turns to yellow during charging
4. when the light turns to green, means fully charged.
5. if battery is not used for long time, take it out to avoid the
leakage.
6. In order to keep the battery in best using condition, it is suggest
that you fully discharge the battery every month, the method is that
using the indicator till it is automatically power off.
3
Page 7
2.Installation and calibration
2.1 Power supply connection
The indicator is powered by adapter, you plug the adapter directly into
the “DC” pin at the back cover the indicator is ok.
2.2 Connection of load cell and indicator
The indicator can connect with 6 load cell of 350Ω at most, 4 wire or 6
wire load cell both ok.
There are two methods connection between load cell and indicator
A. quick disconnect, as below:
B. Terminal trip connection (inner connection)
1. the exciting voltage for the load cell is 5VDC, the largest output
current 120mA, maximum connect 6 pcs 350Ω load cell;
2. Load cell (or the signal cable for the junction box) is connected
with 5 bit
Terminal trip ( J5) on the circuit-board of weighing indicator.
3. Open Weighing indicator back cover, insert signal cable to the
terminal trip(J5), and make sure the screw is fixed tightly, the
connection as below:
4
Page 8
Weighing display Load cell
+EXC+ voltage
+IN+ output signal
HD ----------- Shielded wire
-IN- output signal
-EXC -IN HD +IN +EXC -EXC- voltage
2.3 Communication interface
RS232 : DB9 Pin or 3 Pin
DB9 definition
5 4 3 2 1
9 8 7 6
Pin function and definition as bellows:
DB9 joint Definition Function
2 TXD Sending data
3 RXD Receiving data
5 GND Ground interface
Note: if RS485, The connection pin is 2 and 5 pin.
3 Pin definition
5
Page 9
Inner connection
TXD RXD GND A B
Pin definitions:
Pins Definitions Function
TXD Sending data
RS232
RS485
RXD Receiving data
GND Ground
A RS485 output “A”port
B RS485 output”B”port
2.4 4-20mA output
Technical parameters:
Resolutions: 1/1000
Outside Load: 100-350Ω
Connection:
1.Inside connection: 4-20mA, load input port connect with “I” of J2, Ground port
connect with “ GND” of J2
2.Outside connection: 4-20mA, load input port connect with “1” pin of D89,
Ground port connect with “6” pin of DB9
Testing :
6
Page 10
Connect the 250Ω to the 4-20mA, adjust the multimeter to the current stalls, the
red pan connect the loading port, and the black pan connect” GND” to test the
output current.
Calibration:
1.Press” Print” and “ Total” go to C32, show[out-4], the output current should be
4mA.
2. If Press [↑]show[out-5] ,Pres[↓]show[out-20],the output current should be
20mA.
3. Adjust the current, for example, adjust to [out-20], the output current should be
12mA, if not 12mA. Press [←](“Zero” key)or[→](Net/gross key)to adjust the current
to 12mA
2.5 Relay output signal function
The indicator can output 4 signal , connect with the outside equipment
the indicator can perform automatic control function and upper limit
and lower limit alarm function. Perform the 4 kinds function through
setting C33, 4 signals
As below
C33=0
C33=1
Output
port
Out1 Close output function No output signal
Out2 Close output function No output signal
Out3 Close output function No output signal
Out4 Close output function No output signal
Out1 Open overload control function Output overload
Out2 Open compliance control function Output compliance
Out3 Open underload control function Output underload
Port definition Function
control signal
control signal
control signal
7
Page 11
Out4 Open stable control function Output stable control
signal
Out1 Weight>=C13 instant connection Feeding control
signal
Out2 Weight<=C14 instant connection Feeding control
signal
C33=2
C33=3 Preserved, no function.
Out3 OUT1、OUT2 close When OUT1、OUT2
connect will output
alarm signal,
connect with buzzer
Out4 Weight<=C14 constant connection
Weight>=C13 disconnect
Feeding start and
stop
For example:
Check weigher application. Connect indicator with yellow, green, red 3
lights. Yellow light on when overload, if ok the green light on. If
underload red light on. And can connect with buzzer. There would be
alarm remind when overload.
C33=2 Feeding control application
For hopper scales. The target is between is 50~500kg, the Output
card can perform open and close feeding valve.
Parameter setting C13=500kg C14=50kg
C13=500kg upper limit , close feeding valve
C14=50kg lower limit, open feeding valve.
Connection: 1. Connect directly control machine, pls. connect OUT 4,
suggest connect another manually switch as the stop switch in
emergency
2. indicator output signal will influence the ON/OFF switch to control
machine, OUT 2 ON, OUT 1 OFF, OUT 3 connect buzzer, alarm
remind when start and stop.
Instruction: when the weight over 500kg, OUT 1 OU3 instant
connection and break off, OUT 4 break off; when weight under
8
Page 12
50kg, OUT 2 OUT3 instant connection and break off, OUT 4
constant connection.
5 4 3 2 1
9 8 7 6
DB9 pin definition port
1 pin 1st output signal pin Out1
6 pin 1st output signal pin Out1
2 pin 2nd output signal pin Out2
7 pin 2nd output signal pin Out2
3 pin 3rd output signal pin Out3
8 pin 3rd output signal pin Out3
4 pin 4th output signal pin Out4
9 pin 4th output signal pin Out4
Inner connection pin definitions
01 01 02 02 03 03 04 04
9
Page 13
3. Basic operation
3.1 keypad
LED instruction
Weighing data
kg
lb
kg
lb
10
Page 14
Hold
Data hold
Gross
Net
Tare
Hi
OK
Lo
.
PCS
TOTAL
Gross weight
Net weight
tare
The weighing data is stable
Weight is zero
Overload
ok
Underload
Decimal
Show the counting status.
Go to accumulation mode
Keys function
keys Key name Key function
1.work with “ZERO” TARE”“ON/OFF” key
Print
Zero
Tare
Gross weight
to perform many functions. 2.Print
Zero the weight within tolerance
At G.W mode, get the tare weight. At N.W
mode, clear the tare, get the G.W
At N.W mode, check the G.W, after 3
seconds back to N.W automatically
11
Page 15
Counting operation
Counting
HoldFreezes displayed weight on screen
1. Accumulation
Accumulation
2. work together with “ Print” to perform
The accumulation function and check the
accumulation result
Power on/off
3.2
Power on & off
Press 2 seconds to power on or power off, after power on the
indicator show”000000-999999”. After self inspection. It go to the weighing
mode. Pls. check it whether 6 bits LED/LCD display and the status light is
good or not.
3.3 Zero operation
1. Initial zero setting
When power on the indicator, if the weight on the scale is within the
initial zero tolerance, indicator show zero automatically.
2. Manually Zero setting
When the scales is stable, and not the negative display, you can zero
Press 2 seconds to power on or power off
the weight within tolerance by press keys.
3.4 Tare operation
12
Page 16
Press “TARE” key, the gross weight is tared, indicator show the Net
weight, the “Net”“tared” status light is on. At tare mode, Press” TARE”
key, clear the tare weight, the indicator will show the gross weight.
3.5 Accumulation operation
At Zero mode, load weight till stable, Press go to accumulation
mode, ”Total” light on, display” n 001”, and then display loaded weight; unload
the weight , back to zero, load the second weight again till stable. Press
display”n002” then display the second loaded weight. Repeat it agin
and again, maximum 999 times.
Check the accumulation
Press “ON/OFF “key and hold it then press “TOTAL”” key, display ”n**”, (it is
the accumulating times) then show total weight. there are 8 digits totally. It
shows the first 4 digits then the last 4 digits. For example, the first 4 digits
is”0012”, the last 4 digits is”34,56” It means the actual weight is “1234.56”
EXIT the accumulation function
When the indicator show the last 4 digits, Press hold it, the
indicator show “ clr n”, it means don’t clear the total Weight, Press “PRINT”
key to exit it; if you want to clear total weight, Press “ZERO” or “TARE” key,
“clrn” change to “clry” it means clear total weight ,then Press “PRINT” to clear
the total weight and exit accumulating mode.
3.6 Print
If the weighing is stable, after connect with printer, press” PRINT”
can print the weight. Note: at tare mode, print with tare. if negative
weight,, can not print. Set C30 for time format.
13
Page 17
3.7 Hold
There are two different hold function. Peak hold function and data hold
function. And the setting is different accordingly.
C11=3 Automatic hold function C11=0 close hold function.
3.8 COUNT
1.At weighing mode, load the weights on the platform scales, Press”
Count” the indicator show” PCS 0” press” Zero” key input the quantity,
press” Print” to confirm it
2 . Load the goods on the platform scales, then the indicator will show
the quantity.
3. Press” Count” back to weighing mode.
4. If you want to weigh different goods, at weighing mode, put the
sample on the platform scales, press” Count” the indicator show”0”
Press “ Count” hold it and then press “ ON/OFF” the indicator
show” PCS 0”, press “Zero” input the sample quantity, press“ Print” to confirm it. Then repeat the step 2 and 3.
4. Calibration and Parameter setting
4.1 Enter setting
There have two methods to enter the setting menu:
1. when the switch “ CAL” is off, press the “PRINT” hold it and
then press” HOLD” enter C08-39 setting.
2. Take out the sealing screw on the back of indicator, then press
“CAL” , at the “SPAN” position as below. press “PRINT” hold it and
then press “TOTAL” key , enter C01-C39 setting.
14
Page 18
The key functions in setting:
Enter
Up
Down
Left
Power switch. exit setting
4.2. Step of calibration operation:
According to the second method which can enter setting menu,
C01-C39
step
Method of operation
display Remark
1 [C01 ] After you enter calibration
mode, it display [C01 ]
2 press[C01 1]Weight unit
option:1=kg
2=lb
3 press
press
pressor
[C02 ]
[C02 0]
[C02 2]
15
Set decimal digits
option:0/1/2/3/4
Select decimal digit
Page 19
4 press
press
pressor
5 press
press
pressor/
press
press
press
press
6
7 press
press
pressor
press
pressor
press
[C03 ]
[C03 1]
[C03 5]
[C04 ]
[0100.00]
[0100.00]
[C05 ]
[C05 0]
[C05 1]
[CAL 9]
。。。。。。
[0000.00]
[C06 ]
[C06 0]
[C06 1]
[SPAN ]
[0100.00]
[0080.00]
[CAL 9]
……
[0080.00]
[CAL End]
example: two decimal point:
[C02 2]
Set graduation
option:1/2/5/10/20/50
Select required graduation
example:graduation 5:[C03 5]
Max capacity
example:max weighing 100kg:
[0100.00]
Zero calibration
Option
0=no need zero calibration
1=need zero calibration
calibration zero please choose 1 and
ensure scale is empty and “stable”
light is on
Ensure zero calibration, countdown.
Till show[0.00](example for two
decimal point)。
calibration
option:
0=No need calibration
1= need calibration
Load weights on scales
according to max. capacity.
Suggest close to the max
capacity, at least 10% of max.
capacity.
For example: the weights is
80kg
As bellows:
Input the 0080.00, count down ,
16
Page 20
then indicator shows 0080.00 ,
calibration is over.
If you want to set application
function parameter. Press
“PRINT” if you want to exit
press “TOTAL”
8
press
press
pressor
[C07 ]
[07 0]
[07 1]
Default parameters setting
option:0=non-restore default
parameters
1=restore default parameters
Note: after the above
parameters setting finish, please
do not set default parameters to
avoid the original setting
parameters is lost.
4.3 Application function parameters setting chart
Function
warning
tone
Setting
Item
C08
warning
tone
parameters setting and instruction
Options: 0 = close warning tone
1 = open warning tone
option:0=close auto power off
10= power off automatically if no
Automatic
power off
C09
Automatic
power off
30= power off automatically if no
change within 10 minute.
change within 30 minute.
60= power off automatically if no
change within 60 minute.
LED Version:
Power saving
setting
C10
Power saving
setting
option:0= close power saving setting
3= close display if no change within 3min.
5= close display if no change within 5 min.
LCD Version:
17
Page 21
Hold
function
Kg/lb
conversion
Upper/lower
limit alarm
Inner Code
display
C11
Hold mode
C12
Kg/lb
conversion
C13
Upper limit
alarm value
C14
Lower limit
alarm value
C15
Check inner
code
0=Close he backlight
1= backlight when the weight change or
press the keyboard
2=constant backlight
option:0=close hold function
1=Peak hold /2=Data Hold
Instruction:
Peak-hold: it shows the max. data,
mainly application for materials testing,
such as tension and pulling force.
Date-hold: it shows current weight value.
Mainly application for animal weighing.
C12=0 stop kg/lb conversion
C12=1 kg/lb conversion is ok
You can set it within the max. capacity limit
enter C15 to check the inner code
Date and time
Communication
setting
C16
Date
C17
Time
C18
Serial interface
data output method
18
Enter C16, you can set the date,
from left to right: year/month/day
Enter C17, you can set the time from
left to right: hour/min./sec.
option:0= Close serial interface data
output
1=Continuous sending, connect
second display
Page 22
Zero range
C19
Baud rate
C20
Manually zero
range
C21
Initial zero range
C22
Automatically zero
tracking range
2=Print method, connect printer.
3=Command request method ,
connect computer.
4=PC continues sending format,
connect computer.
5=PC/ second display continuous
sending format.
option:
0=1200/1=2400/2=4800/3=9600
Option:
0= close manually zero setting
1=±1% max capacity
2=±2% max capacity
4=±4% max capacity
10=±10% max capacity
20=±20% max capacity
100=±100% max capacity
option:0= no initial zero setting
1=±1% max capacity
2=±1% max capacity
5=±1% max capacity
10=±1% max capacity
20=±1% max capacity
Options:
0= close zero tracking
0.5=±0.5d
1.0=±1.0d
2.0=±2.0d
3.0=±3.0d
4.0=±4.0d
5.0=±5.0d
Zero tracking
Note: 1. d = division
2. the zero tracking range can
19
Page 23
not bigger than manual zero range.
00= close overload range
Overload range
Negative display
Standstill time
C23
Automatically zero
tracking time
C24
Overload range
C25
Negative display
range
C26
Standstill time
C27
Standstill range
Options:
0= close zero tracking time
1= 1 second
2= 2 seconds
3= 3 seconds
Dynamic filter
Instruction:
Dynamic filter is
collecting the data
filter before loaded
weight stable.
When loaded
weight easily
shaking (for
example animal) ,
you can set this
filter to make
weight display
more stable
C29
Noise filter
option:0= close dynamic filter
1=1 digital filter strength
2=2 digital filter strength
3=3 digital filter strength
4=4 digital filter strength
5=5 digital filter strength
6=6 digital filter strength
Note : Pls setting dynamic filter
strength carefully, the No. is
bigger, more stable. if the
loaded weight shake not too
much. The setting is less than 3
option:0=close noise filter
1=1 digital filter strength
2=2 digital filter strength
3=3 digital filter strength
5.5 PC or Second display continuous sending format
25
Page 29
6. Maintenance
6.1 Regular error and solution
ERROR REASON SOLUTION
UUUUUU
1.Overload
2. wrong connection
with load cell
3. load cell has quality
problem.
1. reduce the weight
2. check load cell connection
3. inspection load cell. Check
the input and output
nnnnnnn
ERR1
ERR2
ERR3
1. calibration is no
good
2. wrong connection
3. load cell has quality
problem
During calibration, not
input the weights or
the weight is overload
During calibration , the
weights is below than
Min. required weights
During calibration, the
input signal is
negative
26
1. check scale is resisted or
not, foot is kept level or not.
2. check load cell connection.
3. checking load cell: check
input and output resistance to
judge it is good or not.
Input the correct weights
The calibration weights
Minimum is 10% of Max. cap.
Recommend 60%-80% of Max.
Cap.
1.check the connection is
correct
2.check load cell is no
problem
3.recalibration if still wrong
change the PCB
Page 30
ERR4
During calibration, the
signal is unstable
After the platform is stable, start
calibration
ERR5 Change PCB
6.2 Daily maintain
1. Protect the indicator from strong sunlight to prolong the using life
2. Good connection between load cell and indicator. Far from away
from strong electric field, magnetic field.
3. Power off the indicator when lightning
4. Power off the indicator firstly before plug and unplug
6.3 Restore default parameter
Enter to calibration, Set C07=1. Press” PRINT” then press” TOTAL” to
exit saving setting. All parameter will back to default
Note: Pls. do not restore default parameter easily if you are not
professional staff or not yet calibrate the scales.
Default parameter
Parameter
C01
C02
C03
C04
C05
C06
C07
C08
C09
C10
instruction Default
Calibration
Decimal digits
Resolution
Max. capacity
Empty calibration
Capacity calibration
Restore default
Warning tone
Power-off automatically
Power saving mode
1
0
1
10000
0
0
0
1
0
0
27
Page 31
C11
Hold function
0
C12 Prohibit kg/lb conversion 1
C13 Upper limit alarm 000000
C14 Under limit alarm 000000
C15 Inner code
C16 Date setting
C17 Time setting
C18 Serial interface data output 0
C19 Serial interface Baud rate 3(9600)
C20 Zero manually 10
C21 Initial zero 10
C22 Zero tracking range 0.5
C23 Zero tracking time 1
C24 Overload range 9
C25 Negative range 10
C26 Standstill time 1
C27 Standstill range 2
C28 Dynamic filter 0
C29 Noisy filter 2
C30 Print format 0
C31 Analog signal options 1
C32 4~20mA testing 4
C33 Relay output setting 1
C34 Muti PC communication add. 0
C35
Wireless communication channel
6
C36 Calibration location gravity 9.7936
C37 Destination gravity 9.7936
C38 Version No. check
C39 Reserved menu