PCE Instruments PCE-SLT, PCE-SLT-TRM, PCE-SLT-TRM-24V, PCE-SLT-TRM-24V-ICA, PCE-SLT-ICA Users guide

...
Page 1
© PCE Instruments
PCE-SLT Sound Level Meter (Including Transmitter PCE-SLT-TRM)
User Manual
Last change: 13 August 2024
v1.0
English
Page 2
© PCE Instruments
Contents
1 Safety notes ........................................................................................... 1
2 Specifications ........................................................................................ 2
3 Dimensions ............................................................................................ 3
4 Description transmitter ........................................................................ 4
5 Display description ............................................................................... 5
6 Initial commissioning ........................................................................... 7
6.1 Connecting the transmitter ............................................................................................... 7
6.2 Connecting the display ..................................................................................................... 7
7 Set measurement ranges ..................................................................... 7
8 Calibration ............................................................................................. 7
9 Alarm setting (control) ......................................................................... 7
10 Advanced menu .................................................................................... 8
10.1 Decimal point shift ............................................................................................................ 8
10.2 4 mA Parameter ............................................................................................................... 8
10.3 20 mA parameter ............................................................................................................. 8
10.4 Filter function ................................................................................................................... 8
10.5 Hysteresis for the control message................................................................................... 8
10.6 Hysteresis for the alarm function ...................................................................................... 9
10.7 Offset ............................................................................................................................... 9
10.8 Gain setting ...................................................................................................................... 9
10.9 Set RS232 unit ................................................................................................................10
11 RS232 ................................................................................................... 11
11.1 RS232 settings ................................................................................................................11
11.2 RS232 protocol ...............................................................................................................12
12 Reset system ....................................................................................... 12
13 Contact ................................................................................................. 12
14 Disposal ............................................................................................... 12
Page 3
© PCE Instruments
1
1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.
The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
Do not expose the device to shocks or strong vibrations.
The case should only be opened by qualified PCE Instruments personnel.
Never use the instrument when your hands are wet.
You must not make any technical changes to the device.
The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
The device must only be used with accessories from PCE Instruments or equivalent.
Before each use, inspect the case for visible damage. If any damage is visible, do not
use the device.
Do not use the instrument in explosive atmospheres.
The measurement range as stated in the specifications must not be exceeded under
any circumstances.
Non-observance of the safety notes can cause damage to the device and injuries to the user.
We do not assume liability for printing errors or any other mistakes in this manual. We expressly point to our general guarantee terms which can be found in our general terms of
business.
Page 4
© PCE Instruments
2
2 Specifications
Sound sensor
Measuring range
30 ... 130 dB in three ranges
Resolution
0.1 dB
Measuring accuracy
±1.5 dB
Frequency weighting
A
Power supply
via transmitter
Cable length
1.5 m
Wall bracket
screwable (included in the scope of delivery)
Envioronmental conditions
max. 80 % RH / 0 ... +50 °C / 32 ... 122 °F
Microphone
1/2" precision electret microphone
Standard
IEC 61672 Class II
Transmitter
Housing
ABS plastic
Power supply (depending on model)
90 ... 260 VAC 24 V DC (max. 1,5 A)
Calibration
adjustable via grub screw
Output
4 ... 20 mA
Protection class
IP 50
Environmental conditions
max. 80 % RH / 0 ... +50 °C / 32 ... 122 °F
Digital display
Input signal from transmitter
4 ... 20 mA
Display
-1999 ... 9999 (1 ... 3 decimal places, depending on measured value / user-selectable)
Programming data (decimal place, limit value...)
are stored in the device Control output
relay output: 0.5 A / 240 VAC
Alarm output
relay output: 0.5 A / 240 VAC
Power supply
90 ... 260 ACV / 50 ... 60 Hz
Dimensions / panel cut-out according to DIN
92 (+ 0.8) mm x 45 (+ 0.5) mm Weight
approx. 250 g / 0.56 lb
Envioronmental conditions
max. 80 % RH / 0 ... +50 °C / 32 ... 122 °F
Protection class
IP 60
Page 5
© PCE Instruments
3
3 Dimensions
Page 6
© PCE Instruments
4
4 Description transmitter
5 and 6 Supply voltage
3.1 Calibration potentiometer
7 and 8 Current output
3.2 Power light
1 to 4 Not used
3.3 Sensor connection (female)
3.4 Connection to mounting plate
3.5 Fixing holes for wall mounting
3.6 Sensor connection (male)
3.7 Sensor
3.8 Transmitter cover
3.9 Setting the measured value ranges
3.10 Light (measured value too low)
3.11 Light (measured value too high)
Page 7
© PCE Instruments
5
5 Display description
3.1 Display
3.2 Measured value display (indicator)
3.3 Value set display (indicator)
3.4 SET key
3.5 Decrease key
3.6 Increase key
3.7 Control display
3.8 Alarm display
3.9 Terminal strip
3.10 Mounting bracket
3.11 RS232 interface
Page 8
© PCE Instruments
6
T1 - T2 Power supply T3 - T4 Control relay T5 - T6 Alarm relay T7 - T8 Power supply 24 VDC T9 - T14 Not connected T15 - T16 4 ... 20 mA interface
Page 9
© PCE Instruments
7
6 Initial commissioning
6.1 Connecting the transmitter
First mount the connection terminal to a designated DIN rail or screw it to a designated surface. First connect the mains voltage. To do this, use connection 5 and 6 on the connection terminal. Ensure that the connection cable is initially voltage-free. Then connect the transmitter to the connection terminal. Finally, connect the sensor to the transmitter.
Note: For the 24 V version of the transmitter (PCE-SLT-TRM-24V), ensure that the supply ground is galvanically isolated from the signal ground.
6.2 Connecting the display
First mount the display using the mounting bracket. For the power supply, connect the mains cable to the T1 and T2 connections on the display connection terminal. Ensure that the mains supply cable is initially voltage-free. Now connect the transmitter to the display. To do this, connect pin 7 to T15 (positive) and pin 8 to T16 (negative).
7 Set measurement ranges
First open the transmitter cover. Then remove the internal rubber seal. The switches for setting the measurement range are now accessible. Use the table on the inside of the transmitter cover to set the measurement range. Then cover the switches again with the rubber seal and close the transmitter cover.
8 Calibration
Open the transmitter cover. Turn the potentiometer labelled "SPAN" to adjust the measured value. To make changes to the potentiometer, use a small slotted screwdriver.
9 Alarm setting (control)
The display has two separate alarm relays. A distinction is made between control and alarm. The difference is that when the alarm switches, the display also flashes, which is not the case with the control.
To programme the limit values of the two functions, proceed as follows: First press the "SET" key briefly. CtLo" appears on the display to set the lower control value. You can now set this value directly using the arrow keys. Press the "SET" key to confirm this value and return directly to the menu. To set other parameters, press the "SET" key continuously until you have reached your parameter. The menu is organised as follows.
CtLo lower control value CtHi upper control value ALLo lower alarm value ALHi upper alarm value
Once you have set all the parameters, press the "SET" key again to exit the menu.
Page 10
© PCE Instruments
8
10 Advanced menu
To access the extended menu, press the "SET" key for two seconds. This menu is organised as follows:
dPSt Decimal point shift 4-A4 mA parameter 20-A20 mA parameter FiLt Filter function CtHY Hysteresis for the control function ALHY Hysteresis for the alarm function oFSt Offset GAin Gain setting Unit Set RS232 unit
10.1 Decimal point shift
To move the decimal point, first press the "SET" key for two seconds. dPSt" appears on the display. Now press the arrow keys to access this configuration mode and to move the decimal point. Press the "SET" key to save the setting.
10.2 4 mA Parameter
To change the parameterisation for 4 mA, first press the "SET" key for two seconds. dPSt" appears on the display. Press the "SET" key again. "4-A" now appears on the display. Now press the arrow keys to access this configuration mode and to change the parameterisation for 4 mA. Press the "SET" key to save the setting.
10.3 20 mA parameter
To change the parameterisation for 20 mA, first press the "SET" key for two seconds. dPSt" appears on the display. Now press the "SET" key twice. The display now shows "20-A". Now press the arrow keys to access this configuration mode and to change the parameterisation for 20 mA. Press the "SET" key to save the setting.
10.4 Filter function
To change the parameterisation for the filter function, first press the "SET" key for two seconds. dPSt" appears on the display. Now press the "SET" key three times. FiLt" now appears on the display. Now press the arrow keys to access this configuration mode and to change the parameterisation for the filter function. The higher the value, the more filtering takes place. Press the "SET" key to save the setting.
10.5 Hysteresis for the control message
To change the parameterisation of the hysteresis for the control message, first press the "SET" key for two seconds. dPSt" appears on the display. Now press the "SET" key four times. CtHY" now appears on the display. Now press the arrow keys to access this configuration mode and to change the parameterisation for the hysteresis. Press the "SET" key to save the setting.
Page 11
© PCE Instruments
9
10.6 Hysteresis for the alarm function
To change the parameterisation of the hysteresis for the alarm function, first press the "SET" key for two seconds. dPSt" appears on the display. Now press the "SET" key five times. ALHY" now appears on the display. Now press the arrow keys to access this configuration mode and to change the parameterisation for the hysteresis. Press the "SET" key to save the setting.
10.7 Offset
To change the parameterisation for the offset, first press the "SET" key for two seconds. dPSt" appears on the display. Now press the "SET" button six times. The display now shows "oFSt". Now press the arrow keys to access this configuration mode and to change the parameterisation for the offset. Press the "SET" key to save the setting.
10.8 Gain setting
To change the parameterisation for the gain, first press the "SET" key for two seconds. dPSt" appears on the display. Now press the "SET" key seven times. “GAin" now appears on the display. Now press the arrow keys to access this configuration mode and to change the parameterisation for the gain. Press the "SET" key to save the setting.
Page 12
© PCE Instruments
10
10.9 Set RS232 unit
To change the unit for the RS232 interface, first press the "SET" key for two seconds. dPSt" appears on the display. Now press the "SET" key eight times. The display will now show "Unit". Now press the arrow keys to access this configuration mode and to change the parameterisation for the unit. The correct value can be found in the following table.
00 = No unit
33 = KHz
66 = mF
01 = C
34 = DCV
67 = MHz
02 = F
35 = DCµA
68 = uH
03 = %
36 = DCA
69 = dBm
04 = %RH
37 = DCmA
70 = Red
05 = pH
38 = ohm
71 = Gren
06 = %O2
39 = Kohm
72 = Blue
07 = mg/L
40 = Mohm
73 = Stau
08 = m/s
41 = mH
74 = mSEC
09 = knot
42 = H
75 = uSEC
10 = km/h
43 = nF
76 = SEC
11 = ft/m
44 = µF
77 = Kgc2
12 = ml/h
45 = hFe
78 = mmHg
13 = uS
46 = DIO
79 = mH2O
14 = mS
47 = WATT
80 = inHg
15 = Lux
48 = KWATT
81 = Kgcm
16 = Ftcd
49 = ACmV
82 = LBin
17 = dB
50 = ACV
83 = N-cm
18 = mV
51 = ACµA
84 = CMM
19 = PPM
52 = ACA
85 = CFM
20 = mg
53 = ACmA
86 = mbar
21 = Tesl
54 = COSƟ
87 = Pa
22 = bar
55 = Kg
88 = kPa
23 = PSI
56 = Lb
89 = uHg
24 = cmHg
57 = gram
90 = Torr
25 = iH20
58 = oz
91 = hPa
26 = ATP
59 = NewT
92 = m/s2
27 = RPM
60 = m/m
93 = mm/s
28 = in/m
61 = Hour
94 = mm
29 = cm/m
62 = Min
95 = cm/s
30 = COUT
63 = VA
96 = inch
31 = Hz
64 = KVA
97 = FtS2
32 = DEG
65 = KWHr
98 = in/s
Press the "SET" key to save the setting.
Page 13
© PCE Instruments
11
11 RS232
The PCE-SLT has an RS232 interface that can be connected via a 3.5 mm jack. The jack plug must be constructed as follows:
11.1 RS232 settings
To receive the data correctly, set the COM connection on your PC as follows:
Page 14
© PCE Instruments
12
11.2 RS232 protocol
The display transmits a 16-digit protocol. This is structured as follows:
D15
Start word
D14
4
D13
Value too high = 1 Value too low = 2
D12 and D11
Unit, see 10.9
D10
Positive polarity = 0 Negative polarity = 1
D9
The specified value indicates the position of the decimal point from left to right. Values from 0 ... 3 are possible.
D8 ... D1
The specified values show the measured value from D8 to D1. If the measured value is 1234, for example, the following is sent from D8 to D1: 00001234
D0
End word
12 Reset system
To reset the system, proceed as follows. Press and hold the "SET" and the "Decrease" key for five seconds. "rSt" flashes on the display. The system has now been reset. After this, the device returns to measuring mode. After the reset, the device may need to be re-parameterised.
13 Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact us. You will find the relevant contact information at the end of this user manual.
14 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations.
If you have any questions, please contact PCE Instruments.
Page 15
© PCE Instruments
13
PCE Instruments contact information
Germany France Spain
PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L. Im Langel 26 23, rue de Strasbourg Calle Mula, 8 D-59872 Meschede 67250 Soultz-Sous-Forets 02500 Tobarra (Albacete) Deutschland France España Tel.: +49 (0) 2903 976 99 0 Téléphone: +33 (0) 972 3537 17 Tel. : +34 967 543 548 Fax: +49 (0) 2903 976 99 29 Numéro de fax: +33 (0) 972 3537 18 Fax: +34 967 543 542 info@pce-instruments.com info@pce-france.fr info@pce-iberica.es www.pce-instruments.com/deutsch www.pce-instruments.com/french www.pce-instruments.com/espanol
United Kingdom Italy Turkey
PCE Instruments UK Ltd PCE Italia s.r.l. PCE Teknik Cihazları Ltd.Şti. Trafford House Via Pesciatina 878 / B-Interno 6 Halkalı Merkez Mah. Chester Rd, Old Trafford 55010 Loc. Gragnano Pehlivan Sok. No.6/C Manchester M32 0RS Capannori (Lucca) 34303 Küçükçekmece - İstanbul United Kingdom Italia Türkiye Tel: +44 (0) 161 464902 0 Telefono: +39 0583 975 114 Tel: 0212 471 11 47 Fax: +44 (0) 161 464902 9 Fax: +39 0583 974 824 Faks: 0212 705 53 93 info@pce-instruments.co.uk info@pce-italia.it info@pce-cihazlari.com.tr www.pce-instruments.com/english www.pce-instruments.com/italiano www.pce-instruments.com/turkish
The Netherlands United States of America Denmark
PCE Brookhuis B.V. PCE Americas Inc. PCE Instruments Denmark ApS Institutenweg 15 1201 Jupiter Park Drive, Suite 8 Birk Centerpark 40 7521 PH Enschede Jupiter / Palm Beach 7400 Herning Nederland 33458 FL Denmark Telefoon: +31 (0)53 737 01 92 USA Tel.: +45 70 30 53 08 info@pcebenelux.nl Tel: +1 (561) 320-9162 kontakt@pce-instruments.com www.pce-instruments.com/dutch Fax: +1 (561) 320-9176 www.pce-instruments.com/dansk info@pce-americas.com www.pce-instruments.com/us