Contents
1 Safety notes ........................................................................................... 1
2 Scope of delivery .................................................................................. 2
3 Specifications ........................................................................................ 2
4 Device description ................................................................................ 3
5 Display description ............................................................................... 3
6 On / off .................................................................................................... 4
7 Measurement ......................................................................................... 4
8 Calibration ............................................................................................. 4
8.1 New calibration reference ................................................................................................. 4
8.2 Zero point ......................................................................................................................... 4
9 Automatic Power Off............................................................................. 4
10 Battery .................................................................................................... 4
11 Notes on the measurement .................................................................. 5
12 Contact ................................................................................................. 5
13 Disposal ................................................................................................. 5
© PCE Instruments

© PCE Instruments
1
1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The
device may only be used by qualified personnel and repaired by PCE Instruments personnel.
Damage or injuries caused by non-observance of the manual are excluded from our liability and
not covered by our warranty.
• The device must only be used as described in this instruction manual. If used otherwise,
this can cause dangerous situations for the user and damage to the meter.
• The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
• Do not expose the device to shocks or strong vibrations.
• The case should only be opened by qualified PCE Instruments personnel.
• Never use the instrument when your hands are wet.
• You must not make any technical changes to the device.
• The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
• The device must only be used with accessories from PCE Instruments or equivalent.
• Before each use, inspect the case for visible damage. If any damage is visible, do not
use the device.
• Do not use the instrument in explosive atmospheres.
• The measurement range as stated in the specifications must not be exceeded under
any circumstances.
• Non-observance of the safety notes can cause damage to the device and injuries to
the user.
We do not assume liability for printing errors or any other mistakes in this manual.
We expressly point to our general guarantee terms which can be found in our general terms of
business.
If you have any questions please contact PCE Instruments. The contact details can be found at
the end of this manual.

© PCE Instruments
2
2 Scope of delivery
1 x gloss meter PCE-RM 100
1 x white & black standard
1 x white & black calibration reference
1 x mains adaptor
1 x cleaning cloth
1 x carrying case
1 x user manual
3 Specifications
Optical measurement angle
141 x 45 x 79 mm / 5.6 x 1.8 x 3.1 in

© PCE Instruments
3
4 Device description
"Power" key for switching the meter on and off
Decrease value and reset zero point
Measurement on black surface
Measurement on white surface
Indication that a measurement is made
Battery indicator - When this icon lights up, the built-in battery is discharged and
needs to be recharged
Reading for measurement on black surface (RB)
Reading for measurement on white surface (RW)

© PCE Instruments
4
6 On / off
To turn on the meter, press the "POWER" key. To turn off the power, press and hold the "POWER"
key until "OFF" is displayed.
7 Measurement
To perform a measurement, first place your sample on the black side of the included white & black
standard. Then place the meter on your sample and press "RB" to perform the first measurement.
You will see the first reading in the "RB" area. Then place the sample and the meter on the white
side of the included white & black standard and press "RW" to make a measurement. After the
measurement, you will be shown the value of the last measurement in the "RW" area and the
ratio between the two values in the upper section of the display.
Note: During a measurement, make sure that no light radiation enters the measuring chamber.
Exposure to light from outside affects the measured value.
8 Calibration
To perform a calibration of the measuring instrument, place the instrument on the white side of
the calibration reference provided without any sample. Then press the "CAL" key. The display will
show "CAL" and the calibration will be performed. After calibration, the set calibration standard
value is displayed. Compare this value with the value on the calibration reference you have used.
If the values are the same, the calibration is completed. If they do not match, this calibration
reference must first be entered in the instrument (see chapter 8.1). Then the calibration must be
repeated. Then make a test measurement to check whether the calibration was successful. If this
is not the case, repeat this procedure.
8.1 New calibration reference
If you have a new calibration standard, it must first be entered in the instrument. To do this, keep
the "POWER" key pressed until the display shows "CAL". Now set the displayed value to the
value of your new calibration reference. Press the "POWER" key again to accept the set value.
8.2 Zero point
To reset the zero point of the meter, place the meter on the black side of the included calibration
reference without any sample. Then press the "ZERO" key. CAL" appears on the display and a
measurement is performed. After the measurement, the set calibration standard is displayed.
Then make a test measurement to check whether the zeroing was successful. If this is not the
case, repeat this procedure.
9 Automatic Power Off
To save battery power, the meter has an automatic power off function. If you do not press any
keys for 10 minutes, the meter turns off automatically.
10 Battery
As soon as the battery icon appears on the display, the measuring instrument must be
charged to ensure proper functioning. Use the supplied charger to recharge the battery. The
charging cycle takes approximately 4 hours. After charging, the charger must be disconnected
from the meter.

11 Notes on the measurement
To ensure that the measured values are within the parameters, regular calibrations should be
carried out.
The sample preparation must comply with national standards and regulations.
The sensor system should always be covered to extend its lifetime. It is therefore advisable to
place the meter on the black surface of the white & black standard when no measurement is
made.
12 Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact
us. You will find the relevant contact information at the end of this user manual.
13 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European
Parliament applies. Due to the contained pollutants, batteries must not be disposed of as
household waste. They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either reuse them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance
with your local waste regulations.
If you have any questions, please contact PCE Instruments.
© PCE Instruments
5

© PCE Instruments
6
PCE Instruments contact information
Germany France Spain
PCE Deutschland GmbH
PCE Instruments France EURL
PCE Ibérica S.L.
Im Langel 4
23, rue de Strasbourg
Calle Mayor, 53
D-59872 Meschede
67250 Soultz-Sous-Forets
02500 Tobarra (Albacete)
Deutschland
France
España
Tel.: +49 (0) 2903 976 99 0
Téléphone: +33 (0) 972 3537 17
Tel. : +34 967 543 548
Fax: +49 (0) 2903 976 99 29
Numéro de fax: +33 (0) 972 3537 18
Fax: +34 967 543 542
[email protected] [email protected] [email protected]
www.pce-instruments.com/deutsch
www.pce-instruments.com/french
www.pce-instruments.com/espanol
Germany United Kingdom Italy
PCE Produktions- und
PCE Instruments UK Ltd
PCE Italia s.r.l.
Entwicklungsgesellschaft mbH
Unit 11 Southpoint Business Park
Via Pesciatina 878 / B-Interno 6
Im Langel 26
Ensign Way, Southampton
55010 Loc. Gragnano
D-59872 Meschede
Hampshire
Capannori (Lucca)
Deutschland
United Kingdom, SO31 4RF
Italia
Tel.: +49 (0) 2903 976 99 471
Tel: +44 (0) 2380 98703 0
Telefono: +39 0583 975 114
Fax: +49 (0) 2903 976 99 9971
Fax: +44 (0) 2380 98703 9
Fax: +39 0583 974 824
[email protected] [email protected] [email protected]
www.pce-instruments.com/deutsch
www.pce-instruments.com/english
www.pce-instruments.com/italiano
The Netherlands China Hong Kong
PCE Brookhuis B.V.
PCE (Beijing) Technology Co., Limited
PCE Instruments HK Ltd.
Institutenweg 15
1519 Room, 6 Building
Unit J, 21/F., COS Centre
7521 PH Enschede
Zhong Ang Times Plaza
56 Tsun Yip Street
Nederland
No. 9 Mentougou Road, Tou Gou District
Kwun Tong
Telefoon: +31 (0)53 737 01 92
102300 Beijing, China
Kowloon, Hong Kong
[email protected] Tel: +86 (10) 8893 9660
Tel: +852-301-84912
www.pce-instruments.com/dutch
[email protected] [email protected]
www.pce-instruments.cn
www.pce-instruments.cn
United States of America Turkey
PCE Americas Inc.
PCE Teknik Cihazları Ltd.Şti.
711 Commerce Way suite 8
Halkalı Merkez Mah.
Jupiter / Palm Beach
Pehlivan Sok. No.6/C
33458 FL
34303 Küçükçekmece -
İstanbul
USA
Türkiye
Tel: +1 (561) 320-9162
Tel: 0212 471 11 47
Fax: +1 (561) 320-9176
Faks: 0212 705 53 93
[email protected] [email protected]
www.pce-instruments.com/us
www.pce-instruments.com/turkish