
Temperature Controller PCE-RE21R, PCE-RE21S, PCE-RE21T
User manuals in various languages (français,
italiano, español, português, nederlands, türk, polski,
русский, 中文) can be found by using our
product search on: www.pce-instruments.com
Last change: 1 October 2019
v1.0

© PCE Instruments
Contents
1 Safety notes ........................................................................................... 1
2 Main features ......................................................................................... 2
3 Technical parameters ........................................................................... 2
4 Front Panel Descrition .......................................................................... 3
5 Size ......................................................................................................... 3
6 Display the status and basic operation .............................................. 4
6.1 System parameter setting................................................................................................. 4
6.2 Set value setting ............................................................................................................... 5
6.3 “At”PID parameter auto-tuning.......................................................................................... 5
7 Parameter list and function .................................................................. 5
7.1 Field parameters .............................................................................................................. 5
7.2 Field parameter ................................................................................................................ 6
8 Default settings ..................................................................................... 7
9 Wiring diagram ...................................................................................... 8
10 Note ........................................................................................................ 8
11 Warranty ................................................................................................. 9
12 Disposal ................................................................................................. 9

© PCE Instruments
1
1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The
device may only be used by qualified personnel and repaired by PCE Instruments personnel.
Damage or injuries caused by non-observance of the manual are excluded from our liability and
not covered by our warranty.
• The device must only be used as described in this instruction manual. If used otherwise,
this can cause dangerous situations for the user and damage to the meter.
• The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
• Do not expose the device to shocks or strong vibrations.
• The case should only be opened by qualified PCE Instruments personnel.
• Never use the instrument when your hands are wet.
• You must not make any technical changes to the device.
• The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
• The device must only be used with accessories from PCE Instruments or equivalent.
• Before each use, inspect the case for visible damage. If any damage is visible, do not
use the device.
• Do not use the instrument in explosive atmospheres.
• The measurement range as stated in the specifications must not be exceeded under
any circumstances.
• Non-observance of the safety notes can cause damage to the device and injuries to
the user.
We do not assume liability for printing errors or any other mistakes in this manual.
We expressly point to our general guarantee terms which can be found in our general terms of
business.
If you have any questions please contact PCE Instruments. The contact details can be found at
the end of this manual.

© PCE Instruments
2
2 Main features
- White large characters color LCD display and new higher contrast LCD with
greatly improved visibility.
- Easy to Use, from Model Selection to Setup and Operation.
- The External waterproof sealing ring and the Internal waterproof sealing ring make
the instrument reach the good waterproof performance.
- Plastic Handle waterproof type key, is a hard surface that won't scuff, Operation
feel clear and smooth.
- Input type: K, E, J, N thermocouples and Pt100 platinum resistance freely
programmable.
- Use digital calibration technology for input Measurement accuracy:0.3%FS,
maximum resolution is 0.1℃.
- Use advanced artificial intelligent control algorithm, no overshoot and with the
function of auto tuning (AT) and self-adaptation.
- With worldwide power supply of AC/DC100-240V or DC12-24V.
3 Technical parameters
AC/DC 100~240V or DC 12-24V (-15%, +10% / 50-60HZ)
85% to 110% of rated supply voltage
Approx. 5.2 VA at 100 to 240V AC,
Approx. 3 VA at 12 to 24 VDC
0.3% FS ±1 measurement unit
Input type and
Measuring range
K (-50~+1300°C),
E (0~800°C),
J (0~1000°C),
N (0~1300°C),
Pt100 (-200~+600°C)
≤ 0.5 sec. (when digital filter parameter InF=0)
one-stop regulating, Intelligent Fuzzy PID algorithm
Relay output 3A / 250V AC; 5A / 30V DC
SSR output 12V DC / 50mA (Used to drive SSR)
Triac output NO TRIAC 1A / 250V AC
Electromagnetic
compatibility (EMC)
IEC61000-4-4 (Electrical Fast Transient Burst) ±4KV / 5KHz;
IEC61000-4-5 (Surge), 4KV
Isolation
withstanding voltage
Between power relay contact or signal terminals ≥ 2300V DC;
between isolated electroweak terminals ≥ 600V
Temperature: 0~60 °C; Humidity ≤ 90% RH

© PCE Instruments
3
4 Front Panel Description
1. No. 1 display, PV or specified parameter
2. No. 2 display, SP or specified parameter value
3. Temperature unit: °C/°F
4. Operation indicators:
OUT indicators, AL 1 indicator, AL2 indicators
5. Setup key: For accessing parameter table and
conforming parameter modification
6. Enter key: For confirm and change to another
Parameter
7. Data shift key
8. Data decrease key
9. Data decrease key

© PCE Instruments
4
6 Display the status and basic operation
6.1 System parameter setting
In the basic display state, press and hold key for 2 seconds. Enter the field parameters, set
the LOC = 800. Then press the button to confirm and enter the system parameter setting
state. , , key can directly modify the parameter values. Press the key to
reducing the data, press the to increase the data. Waiting to modify the value of the decimal
point will flash (like a cursor). Press key and hold, you can quickly increase / decrease in value,
and the speed will be automatically accelerated. Also press the key to move to modify the
data location (cursor), the operation is more efficient. key can be stored to modify
parameter values and display the next parameter, press and hold key 2 seconds, and can
return to the previous parameter; can immediately exit the parameter setting state.

© PCE Instruments
5
6.2 Set value setting
In the basic display state, press , , key can directly modify the given value.
6.3 “At”PID parameter auto-tuning
Press and hold for 2 seconds, the At parameter appears,"At" parameter "OFF" is set to "ON"
and then press the key to confirm instrument can start the implementation of the autotuning
Given function, the instrument in the basic display state display will flash the word "At",
the instrument after 2 oscillation cycle ON-OFF control can automatically calculate the PID
parameters. If you want to advance to give up auto-tuning , "At" parameter “ON” is set to “OFF”
and then press key to confirm.
Given tuning parameter values obtained are not identical, to perform auto-tuning function,
should be first given value set in the most commonly used value or middle value, if the system
is good insulation properties of the furnace, the given value should be set in the system uses
the maximum, and then Execute the start of the operation of auto-tuning function. Reasons to
learn, auto-tuning after the initial use, the effect may not be the best, you need a period of time
(usually the same time auto-tuning control) before they can get the best results.
7 Parameter list and function
7.1 Field parameters
In the basic display state, press and hold key 2 seconds, enter the field parameters,
AL1 high
limit alarm
value
“HAL1” is the absolute value alarm or deviation value
alarm, by “ALtd” parameter definition. When the value
set to Max. will disable this function. (3200)
AL1 low
limit alarm
value
“LAL1” is the absolute value alarm or deviation value
alarm, by “ALtd” parameter definition. When the value
set to Min. will disable this function. (-999)
AL2 high
limit alarm
value
“HAL1” is the absolute value alarm or deviation value
alarm, by “ALtd” parameter definition. When the value
set to Max. will disable this function. (3200)
AL2 low
limit alarm
value
“LAL1” is the absolute value alarm or deviation value
alarm, by “ALtd” parameter definition. When the value
set to Min. will disable this function. (-999)
run:Run control state StoP:Stop state, No.2 display
flashing display "StoP".
LOC<9000, will automatically for 0, allowed to modify
the SV Set Value. LOC≥9000, are not allowed to
modify the SV Set Value. Set the LOC = 800, then
press key to confirm, can enter the following
system parameters.

© PCE Instruments
6
7.2 Field parameter
In the field parameters,set Loc=800, then press the key to enter the system parameters.
Avoid frequent alarm on-off action because of the
fluctuation of PV
ALtd=0,AL1is the is the. deviation value alarm,AL2
absolute value alarm
ALtd=1,AL1 and AL2 is the absolute value alarm.
ALtd=2,AL1 and AL2 is the deviation value alarm.
onoF: on-off control. For situation not requiring high
precision.
FPId: advanced artificial intelligence ”FUZZY PID“
control.
Selection of
heating
refrigeration
onr: Reverse acting. Increase in measured variable
causes a decrease in the output, such as heating
control.
ond: Direct acting. Increase in measured variable
causes an increase in the output, such as
refrigerating control.
Proportional band in PID and APID control. Instead
of percentage of the measurement range, the unit is
the same as PV. Generally, optimal P, I, D and CP
can obtained by auto tuning. They can also be
manually inputted if you already know the correct
Values.
No derivative effect when d=0
CP reflect the instrument operator to adjust the
speed, the size of the CP that affect the control
accuracy. With SSR, SCR output control cycle
preferable to shorter, usually 0.5-3.0 Seconds. The
relay switch output is generally in 15-40 sec. When
the output relay switches, the CP will be limited to
3 seconds, and self-tuning At will automatically set
the CP as the appropriate value, taking into account
the control accuracy and mechanical switch life.
When the control mode CntL = onoF, the action of
the CP as an output disconnect or power-on output
ON Delay time.
HYS is used for on-off control to avoid frequent onoff action of relay. For a reverse acting (heating)
system, when PV > SV, output turns off; when
PV<SV-HYS, output turns on. For a direct acting
(cooling) system, when PV<SV, output turns off;
when PV>SV+HYS, output turns on.
Input spec: K、E、J、N、Pt(Pt100)
0 (no decimal), 0.0 (one decimal place).

Sc is used to shift input to compensate the error
caused by transducer, input signal, or auto cold
junction compensation of thermocouple.
PV after compensation=PV before compensation +
Sc It is generally set to 0. The incorrect setting will
cause measurement inaccurate.
The value of InF will determine the ability of filtering
noise. When a large value is set, the measurement
input is stabilized but the response speed is slow.
Generally, it can be set to 1 to 3. If great interference
exists, then you can increase parameter “InF”
gradually to make momentary fluctuation of
measured value less than 2 to 5. When the
instrument is being metrological verified, “InF” s can
be set to 0 or 1 to shorten the response time.
°C:celsius equals
°F:fahenheit equals
Minimum value that SV is allowed to be.
Maximum value that SV is allowed to be.

© PCE Instruments
8
9 Wiring diagram
10 Note
(1) Can not normally display measured values in the display window alternating show:
"orAL" character, indicating that the input of the measurement signal abnormalities or Out of
range; check Int parameter settings, and then the input sensor signals are the same category,
if it is determined the same, check the input sensing Signal not pick the wrong line, if it is
determined to not pick the wrong line, check whether the sensor problem, replace another
sensor to try.
(2) The instrument used by the local environmental temperature and humidity can not
exceed the scope of the provisions, the instrument around should allow sufficient space for
heat dissipation, instrument internal temperature rise due to thermal radiation will lead to a
result of the measurement accuracy and service life of the affected. In this Case, to be taken
to a forced cooling fan or other measures to reduce the ambient temperature.
(3) the extension or connection then the thermocouple leads should be used with the
thermocouple type to match the compensation conductor; extend or connect the
thermoelectric Resistance of the lead, you should use the minimal resistance of the wire and
cable away from power lines and load connection, in order to avoid signal interference.

© PCE Instruments
9
11 Warranty
You can read our warranty terms in our General Business Terms which you can find here:
https://www.pce-instruments.com/english/terms.
12 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament
applies. Due to the contained pollutants, batteries must not be disposed of as household waste.
They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use
them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with
your local waste regulations.
If you have any questions, please contact PCE Instruments.

© PCE Instruments
10
PCE Instruments contact information
Germany France Spain
PCE Deutschland GmbH
PCE Instruments France EURL
PCE Ibérica S.L.
Im Langel 4
23, rue de Strasbourg
Calle Mayor, 53
D-59872 Meschede
67250 Soultz-Sous-Forets
02500 Tobarra (Albacete)
Deutschland
France
España
Tel.: +49 (0) 2903 976 99 0
Téléphone: +33 (0) 972 3537 17
Tel. : +34 967 543 548
Fax: +49 (0) 2903 976 99 29
Numéro de fax: +33 (0) 972 3537 18
Fax: +34 967 543 542
[email protected] [email protected] [email protected]
www.pce-instruments.com/deutsch
www.pce-instruments.com/french
www.pce-instruments.com/espanol
Germany United Kingdom Italy
PCE Produktions- und
PCE Instruments UK Ltd
PCE Italia s.r.l.
Entwicklungsgesellschaft mbH
Unit 11 Southpoint Business Park
Via Pesciatina 878 / B-Interno 6
Im Langel 26
Ensign Way, Southampton
55010 Loc. Gragnano
D-59872 Meschede
Hampshire
Capannori (Lucca)
Deutschland
United Kingdom, SO31 4RF
Italia
Tel.: +49 (0) 2903 976 99 471
Tel: +44 (0) 2380 98703 0
Telefono: +39 0583 975 114
Fax: +49 (0) 2903 976 99 9971
Fax: +44 (0) 2380 98703 9
Fax: +39 0583 974 824
[email protected] [email protected] [email protected]
www.pce-instruments.com/deutsch
www.pce-instruments.com/english
www.pce-instruments.com/italiano
The Netherlands China Hong Kong
PCE Brookhuis B.V.
PCE (Beijing) Technology Co., Limited
PCE Instruments HK Ltd.
Institutenweg 15
1519 Room, 6 Building
Unit J, 21/F., COS Centre
7521 PH Enschede
Zhong Ang Times Plaza
56 Tsun Yip Street
Nederland
No. 9 Mentougou Road, Tou Gou District
Kwun Tong
Telefoon: +31 (0)53 737 01 92
102300 Beijing, China
Kowloon, Hong Kong
[email protected] Tel: +86 (10) 8893 9660
Tel: +852-301-84912
www.pce-instruments.com/dutch
[email protected] [email protected]
www.pce-instruments.cn
www.pce-instruments.cn
United States of America Turkey
PCE Americas Inc.
PCE Teknik Cihazları Ltd.Şti.
711 Commerce Way suite 8
Halkalı Merkez Mah.
Jupiter / Palm Beach
Pehlivan Sok. No.6/C
33458 FL
34303 Küçükçekmece -
İstanbul
USA
Türkiye
Tel: +1 (561) 320-9162
Tel: 0212 471 11 47
Fax: +1 (561) 320-9176
Faks: 0212 705 53 93
[email protected] [email protected]
www.pce-instruments.com/us
www.pce-instruments.com/turkish