PCE Instruments PCE-RDM 10 Users guide

Page 1
© PCE Instruments
PCE-RDM 10 | Geiger Counter
User Manual
Last change: 3 May 2024
v1.0
English
Page 2
© PCE Instruments
Contents
1 Safety notes ........................................................................................... 1
2 Specifications ........................................................................................ 2
3 Delivery scope ....................................................................................... 2
4 Description of the device ..................................................................... 3
5 Menu ....................................................................................................... 4
5.1 On/Off .............................................................................................................................. 4
5.2 Main screen ..................................................................................................................... 4
5.3 Memory ............................................................................................................................ 6
5.4 Settings ............................................................................................................................ 7
6 Contact ................................................................................................. 11
7 Disposal ............................................................................................... 11
Page 3
© PCE Instruments
1
1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.
The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
Do not expose the device to shocks or strong vibrations.
The case should only be opened by qualified PCE Instruments personnel.
Never use the instrument when your hands are wet.
You must not make any technical changes to the device.
The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
The device must only be used with accessories from PCE Instruments or equivalent.
Before each use, inspect the case for visible damage. If any damage is visible, do not
use the device.
Do not use the instrument in explosive atmospheres.
The measurement range as stated in the specifications must not be exceeded under
any circumstances.
Non-observance of the safety notes can cause damage to the device and injuries to the user.
Always keep the sensor clean.
Do not touch the sensor with your hands.
We do not assume liability for printing errors or any other mistakes in this manual. We expressly point to our general guarantee terms which can be found in our general terms of
business.
Page 4
© PCE Instruments
2
2 Specifications
Specifications
Explanations
Measurement range
0 ... 99.99 µSv/h
Resolution
0.01 µSv/h
Accuracy
0.25 cps / µSv/h
Sensor
Geiger-Müller tube
Types of radiation
beta / gamma / X-ray radiation
Unit
µSv/h
Power supply
3.7 V / 2000 mAh rechargeable LiPo battery
Automatic power off
OFF / 10 min / 20 min / 30 min / 60 min
Interface
USB-C
Display
2.8" 16-bit TFT colour display
Alarm
OFF / 0.30 µsv/h / 0.50 µsv/h / 1.00 µsv/h / 2.00 µsv/h
Memory
167 857 values
Environmental conditions
-20 ... +50 ºC / 20 ... 85 % RH
Dimensions
165 x 90 x 32 mm
Weight
220 g
3 Delivery scope
1 x PCE-RDM 10 geiger counter 1 x micro-USB cable 1 x device bag 1 x user manual
Page 5
© PCE Instruments
3
4 Description of the device
No.
Description
1
Display
2
ON/OFF / Menu / Enter
3
Arrow up
4
Memory / Back
5
Arrow down
6
Next 7 USB interface
1
3 4
5
2
6
7
Page 6
© PCE Instruments
4
5 Menu
5.1 On/Off
Switch the device on using the ON/OFF key. To switch it off, press and hold the ON/OFF key for several seconds.
5.2 Main screen
After switching on the device, the following screen is displayed.
1. Date and time You can display the current date and time here. To set the date and time, see chapter
5.4.1.
2. Alarm As soon as the alarm threshold is exceeded, the device emits an acoustic alarm. Pressing the arrow up key deactivates the audible alarm and the loudspeaker icon is displayed with an "X". To activate it, press the arrow up key again.
3. Battery You can display the current battery status here. To charge the battery, use the included USB cable. Connect the device to a PC using this cable.
4. REL | Real-time dose
Displays the current value in the unit µsv/h. It is updated every second to display the current radiation value.
2
3
10
1
4 5 6
7
8
9
Page 7
© PCE Instruments
5
5. AVG | average dose
Indicates the average value in the unit µsv/h. It records the total amount of energy over a certain period of time and converts it into a measured value for analysis. To set the measuring rate, see chapter 5.4.3.
The average value is displayed after 5 minutes of measurement. If you switch the device off and on again, the average value is reset to zero. You must wait another 5 minutes before you can see the value on the display.
6. ACC | Accumulated dose
Displays the accumulated dose. If the device is switched off, the accumulated dose is not lost. The timer is just stopped. When the device is switched on again, the count continues. To reset the accumulated dose to zero, see chapter 5.4.6.
7. ACC_Time
Indicates how long the cumulative dose has already been calculated.
8. Graph
You can have the development of the last 30 values displayed graphically. This graph is constantly updated.
9. Alarm
Displays the set alarm threshold. To set the alarm threshold, see chapter 5.4.2.
10. REC | Storage
Displays the total number of values saved in the memory. Storage of the values begins 5 minutes after the device is switched on (at this time, the average value can be displayed on the screen).
Note If you are on the main screen, you can switch off the screen (not the device) by pressing the arrow down key. To switch it back on, press the arrow down key again. This function is only active on the main screen.
Page 8
© PCE Instruments
6
5.3 Memory
Press the left key to access the memory. The "History List" screen is the index of the measurement history. The values are sorted by date. The number of saved values is displayed on the right-hand side.
Hint A new file is created each time the device is switched on (after 5 minutes of measurement).
Select the desired file using the up and down arrow keys. The selected file appears in red. To enter the selected file, press the ON/OFF key. The following screen is displayed:
Use the arrow up and down keys to move within the file. To clear the memory, press and hold the next key (right) in the memory overview until "Clearing"
is displayed. When "Clear Finish" appears in the display, the memory has been deleted. The device automatically switches back to measuring mode.
History List
ItemDateQuant 12024-01-08128
22024-01-1087
32024-01-1234 42024-01-1577 52024-01-1812
Page 9
© PCE Instruments
7
5.4 Settings
To call up the settings menu (MAIN SET), briefly press the ON/OFF key when you are on the main screen. The following screen appears:
Select the setting using the arrow up and down keys. The selected setting appears in red. To accept the setting, press the ON/OFF key.
5.4.1 Time Set | Date and time
Use the arrow up and down keys to change the value. Use the right key to change the position (year - month - day - hour - minute - second). Once you have set the date and time, press the left key (history) to return to the main menu.
MAIN SET
Time Set
Alarm set Record Set Language Power Off Set ACC Clean Version Info
Time Set
2024-01-18 16:00:03
Page 10
© PCE Instruments
8
5.4.2 Alarm Set | Setting alarm limit values
Use the arrow up and down keys to select the threshold value. Confirm with the ON/OFF key. Once you have set the alarm threshold, press the left key (history) to return to the main menu.
5.4.3 Record Set | Storage rate
Use the up and down arrow keys to select the storage rate. Confirm with the ON/OFF key. Once you have set the storage rate, press the left key (history) to return to the main menu.
Alarm set
OFF
0.30 usv/h
0.50 usv/h
1.00 usv/h
2.00 usv/h
Record Set
OFF 2 sec 5 sec
10 sec
20 sec
Page 11
© PCE Instruments
9
5.4.4 Language
Use the arrow up and down keys to select the menu language. You can select the following languages: Chinese, Japanese, Russian and English. Confirm with the ON/OFF key. Once you have set the language, press the left key (history) to return to the main menu.
5.4.5 Power Off Set | Automatic power off
Use the arrow up and down keys to select the desired time at which the meter should switch off automatically. Confirm with the ON/OFF key. Once you have set the automatic power off, press the left key (history) to return to the main menu.
Language
中人 日本語 pусскийязык
English
Power Off Set
OFF
10 min 20 min 30 min 60 min
Page 12
© PCE Instruments
10
5.4.6 ACC Clean? | Zero setting of the accumulated dose
Use the arrow up and down keys to select whether you want to reset the accumulated dose to zero. Confirm with the ON/OFF key. Then press the left key (history) to return to the main menu.
5.4.7 Version Info | Firmware information
You can view the firmware data in this screen. When you have finished, press the left key (history) to return to the main menu.
ACC Clean?
YES NO
Version Info
Build Date: 2022-10-17 Version v1.1
Page 13
© PCE Instruments
11
6 Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact us. You will find the relevant contact information at the end of this user manual.
7 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations.
If you have any questions, please contact PCE Instruments.
Page 14
© PCE Instruments
12
PCE Instruments contact information
Germany France Spain
PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L. Im Langel 26 23, rue de Strasbourg Calle Mula, 8 D-59872 Meschede 67250 Soultz-Sous-Forets 02500 Tobarra (Albacete) Deutschland France España Tel.: +49 (0) 2903 976 99 0 Téléphone: +33 (0) 972 3537 17 Tel. : +34 967 543 548 Fax: +49 (0) 2903 976 99 29 Numéro de fax: +33 (0) 972 3537 18 Fax: +34 967 543 542 info@pce-instruments.com info@pce-france.fr info@pce-iberica.es www.pce-instruments.com/deutsch www.pce-instruments.com/french www.pce-instruments.com/espanol
United Kingdom Italy Turkey
PCE Instruments UK Ltd PCE Italia s.r.l. PCE Teknik Cihazları Ltd.Şti. Trafford House Via Pesciatina 878 / B-Interno 6 Halkalı Merkez Mah. Chester Rd, Old Trafford 55010 Loc. Gragnano Pehlivan Sok. No.6/C Manchester M32 0RS Capannori (Lucca) 34303 Küçükçekmece - İstanbul United Kingdom Italia Türkiye Tel: +44 (0) 161 464902 0 Telefono: +39 0583 975 114 Tel: 0212 471 11 47 Fax: +44 (0) 161 464902 9 Fax: +39 0583 974 824 Faks: 0212 705 53 93 info@pce-instruments.co.uk info@pce-italia.it info@pce-cihazlari.com.tr www.pce-instruments.com/english www.pce-instruments.com/italiano www.pce-instruments.com/turkish
The Netherlands United States of America Denmark
PCE Brookhuis B.V. PCE Americas Inc. PCE Instruments Denmark ApS Institutenweg 15 1201 Jupiter Park Drive, Suite 8 Birk Centerpark 40 7521 PH Enschede Jupiter / Palm Beach 7400 Herning Nederland 33458 FL Denmark Telefoon: +31 (0)53 737 01 92 USA Tel.: +45 70 30 53 08 info@pcebenelux.nl Tel: +1 (561) 320-9162 kontakt@pce-instruments.com www.pce-instruments.com/dutch Fax: +1 (561) 320-9176 www.pce-instruments.com/dansk info@pce-americas.com www.pce-instruments.com/us
Loading...