PCE Instruments PCE-PM62, PCE-PM150, PCE-PM300 User Manual

USER MANUAL
COUNTING - SYSTEM PCE -IS01
PCE Americas Inc. 711 Commerce Way Suite 8 Jupiter FL-33458 USA From outside US: +1 Tel: (561) 320-9162 Fax: (561) 320-9176 info@pce-americas.com
www.pce-instruments.com/english
www.pce-instruments.com
PCE Instruments UK Ltd.
Units 12/13
Southpoint Business Park
Ensign way
Hampshire / Southampton
United Kingdom, SO31 4RF
From outside UK: +44
Tel: (0) 2380 98703 0
Fax: (0) 2380 98703 9
info@pce-instruments.com
USER MANUAL
____________________________________________________________________________________________________________________
2
Contents:
CAPACITY BALANCE
1. General description.................................................................................................. 4
2. Completeness .......................................................................................................... 4
3. Technical Data ......................................................................................................... 4
4. Keys and indicators ................................................................................................. 5
5. Security rules ........................................................................................................... 6
6. Environment protection ............................................................................................ 6
7. Preparation ............................................................................................................. 7
8. General operating rules ........................................................................................... 7
9. Connection with external devices (option) ............................................................... 8
10. Special functions description ................................................................................... 9
11. Legend ................................................................................................................... 10
12. Normal weighing .................................................................................................... 10
13. Weighing with tare ................................................................................................. 10
14. Pieces counting (F..-PCS) ..................................................................................... 11
15. Autozeroing (F..-AUt) ............................................................................................. 12
16. Printer cooperation settings (F..-LPt) ..................................................................... 12
17. Serial port RS232C working mode setting ............................................................. 13
18. Function Menu customisation (F..-ACt) ................................................................. 14
19. Maintenance and repairs of small defects ............................................................. 15
USER MANUAL
____________________________________________________________________________________________________________________
3
REFERENCE BALANCE
1. General information ................................................................................................ 16
2. Completeness ........................................................................................................ 16
3. Technical Data ....................................................................................................... 16
4. Keys and indicators ................................................................................................ 17
5. Security rules .......................................................................................................... 18
6. Environment protection ........................................................................................... 18
7. Preparing the balance to work ................................................................................ 19
8. Operation principles ............................................................................................... 19
9. Testing and calibration ........................................................................................... 20
10. Connecting the balance to a computer or a printer ................................................. 20
11. Special functions description .................................................................................. 22
12. Legend ................................................................................................................... 23
13. Normal weighing ..................................................................................................... 23
14. Weighing with tare .................................................................................................. 23
15. Sensitivity calibration (F..-CAL) .............................................................................. 24
16. Piece counting (F..-Pcs) ......................................................................................... 25
17. Autozeroing (F..-AUt) .............................................................................................. 26
18. Serial port working mode (F..-LPt) .......................................................................... 26
19. Serial port transmission parameters ....................................................................... 27
20. Per cent weighing ( PerCEnt ) ............................................................................... 28
21. Change of the units gram/carat ( CarAt ) ............................................................... 29
22. Actualization of the special functions ( ActlV ) ........................................................ 29
23. Attitude of the date (dAtE)-function ........................................................................ 30
24. Error messages ...................................................................................................... 31
25. Maintenance and repairs of small defects .............................................................. 31
Technical Data ............................................................................................................... 32
Declaration of Conformity APPROVAL of EC NO. TYPE PL 06 004 ......................... 34
USER MANUAL
____________________________________________________________________________________________________________________
4
CAPACITY BALANCE
1. General description
PCE-PM series electronic platform scales are perfect solution for use in all applications.
All scales are metrologicaly tested - calibration or EC verification certificate on demand. Scales with EC verification are marked with the green metrology sticker ‘M’.
The scales are made in two main versions: with LED display (basic) and with an accumulator and LCD display (optional).
NACE classification: 29.24.23.
2. Completeness
A standard package consists of:
1. Scale
2. Feeder (ZN12V/500mA)
3. User manual
4. Guarantee card
3. Technical Data
Type
PCE-PM62
PCE-PM150
PCE-PM300
Maximum loading
60kg
150kg
300kg
Minimum loading
400g
1000g
2000g
Reading unit (d)
20g
50g
100g
Verification unit (e)
20g
50g
100g
Accuracy class
III
Working temperature
-10 ÷ +40°C
Tare range
-60kg
-150kg
-300kg
Weighing time
<4s
Pan dimensions
400x400mm
600x500mm
600x500mm
Scale base dimensions
400x400x100mm
600x650x740mm
600x650x740mm
Scale height
740mm
Power supply
230V, 50Hz, 6VA / 12V 500mA
Scale weight
23kg
35kg
35kg
USER MANUAL
____________________________________________________________________________________________________________________
5
4. Keys and indicators
key - switch-on / switch-off (standby),
key
HR
- increased indication resolution,
key
B/G
- gross weigh indication switch,
key
MENU
- special functions menu,
key
0
- zero,
key - print-out,
key
T
- tare (subtract package weight from weighed mass)
indicator
0
- zero indicator
indicator
- result stabilisation indicator
indicator
NET
- net weight indicator (indication with subtracted tare)
indicator
.
- gross weigh indicator (after pressing B/G key)
LCD version:
indicator
OFF
- standby (switched-off with OFF key),
indicator
MODE
- special function mode indicator,
bar indicator
- total load indicator (graduated 0-100%)
indicator
MIN
- weighing result below MIN threshold,
indicator
OK
- weighing result between MIN and MAX thresholds,
indicator
MAX
- weighing result beyond MAX threshold.
weighing
result
function
menu
printout
gross
weight
readability
increace
metrological
data
switch-on / switch-off
zero
tare
zero
indicator
result
stabilisation
net weight
indicator
pieces counting indicator
autozeroing
indicator
USER MANUAL
____________________________________________________________________________________________________________________
6
5. Security rules
To avoid electrical shock or damage of the scale
or connected peripheral devices, it is necessary
to follow the security rules below.
All repairs and necessary regulations can be made by authorised personnel only.
To avoid fire risk use a feeder of an appropriate type (supplied with the scale). Pay attention that supply voltage is compatible with specified technical data.
Do not use the scale when its cover is opened.
Do not use the scale in explosive conditions.
Do not use the scale in high humidity.
If the scale seems not to operate properly, unplug it from the mains
and do not use until checked by authorised service.
6. Environment protection
A ccording to legal regulations it if forbidden to
dispose wasted electronic equipment in waste
containers.
WEEE-Reg.-Nr. DE64249495
USER MANUAL
____________________________________________________________________________________________________________________
7
7. Preparation
1. Take the scale out of a package removing protective foils.
2. Take a pan off and remove protective elements from below a pan.
3. Place the scale on a stable ground not affected by mechanical vibrations and airflows.
4. Level the scale with rotating rear legs so that the air bubble in water-level at the back of the scale is in the middle. Lock legs with a nut.
correct wrong
5. Put on a pan.
6. Remove everything from a pan and plug the scale to the mains with a ground connector. After self-tests and result stabilisation, the scale displays zero indication.
8. General operating rules
1. Before each measurement make sure that zero indicator is displayed. If zero
indicator does not displayed or “----“ communicate appears, press
→0←
key and
wait until zero indication and zero indicator appears.
2. The scale is equipped with a tare equal to its range. To tare the scale press
→T←
key (left or right). Storing a tare value does not extend measuring range, but only subtracts it from a load placed on a pan. To make weight control easier and to avoid range overdrawing, the scale is equipped with a load indicator (graduated in percentages).
3. Weighing result should be read when the indicator " " lights, which signalises stabilisation of a result.
4. When the scale is not used but it is necessary to be ready to work immediately, it
can be switched off by pressing I/ key. The scale reading system is then switched off to "standby" mode (signalled by the indicator "OFF"). To switch the scale on press I/ key. The scale is immediately ready to operate maximum accuracy (after self
tests).
5. Weighed sample should be placed in the centre of the pan.
USER MANUAL
____________________________________________________________________________________________________________________
8
6. Protect the scale against dust, aggressive dusts and liquids. To clean it is advised to wash the scale with water and dry it afterwards.
9. Connection with external devices (option)
The scale may send data to external devices (a computer or a printer) through RS232C or RS485 port.
When cooperating with a computer, the scale sends weighing result after initialising
signal from a computer or after pressing key.
When cooperating with a printer data is send automatically after result stabilisation, but next transmission is possible after removing previously weighted sample.
The scale sends following information: successive number of weighing and weighing result (see: printer cooperation mode setting).
The scale may be equipped with the second interface (optional) used for example for continuous weighing result transmission to an additional, external display.
When cooperating with the scale, a computer should be equipped with a program which enables processing data from the scale – available by PCE Group with extra charge.
We also offer:
Computer wires
Thermal printers
Label printers
Place the scale on a platform to avoid dropping weighed objects on the pan.
Do not overload the scale more then 20% of maximum load (Max). Do not press a pan with a hand.
RxD (receive)
TxD (transmit)
ground
casing
RS232C
A
A
Output
B
Input
B
ground
casing
RS485
USER MANUAL
____________________________________________________________________________________________________________________
9
Computer cooperation protocol:
Transmission parameters: 7bits, 1stop, parity ODD, 9600bps.
Signal description: ComputerScale: initialising signal S I CR LF (53h 49h 0Dh 0Ah), ScaleComputer: weighing result according to the diagram below (16 Bytes):
Note: network number different than zero (in F..-rS function) changes scale working
mode.
Specific bytes description:
Byte
1
- sign “-“ or space
Byte
2
- space
Byte
3÷4
- digit or space
Byte
5÷9
- digit, decimal dot or space
Byte
10
- digit
Byte
11
- space
Byte
12
- k,l,c,p or space
Byte
13
- g,b,t,c or %
Byte
14
- space
Byte
15
- CR
Byte
16
- LF
Connecting cable WK-1 (scale – computer / 9-pin interface):
10. Special functions description
All scales, beside basic functions like weighing and tare, are equipped with the set of additional functions: calibration, pieces counting, autozeroing and serial port working mode setting. Other functions: recipe ingredients summing, density calculation and other special functions can be enabled as an option on customer request (described in additional brochure when ordered).
To display all available function, enter Function Menu with F key. Functions are displayed with successive numbers: F1-LIC, F2-AUt, etc. When special functions are displayed, “MODE” indicator is displayed. Function ACt allows to choose a list of
special functions to be displayed in Function Menu (e.g. most frequently used).
To make clear how to manage with each function, in further part of instruction descriptions are replaced with pictures.
USER MANUAL
____________________________________________________________________________________________________________________
10
11. Legend
- press a key
- scale indication
- forced change
- automatic change
12. Normal weighing
When the pan is empty and indication is different than zero press "
→0←
" key.
Weighing result should be read when
the indicator " " lights.
13. Weighing with tare
The scale is equipped with tare equal to its range. To display gross weight press
B/G key.
USER MANUAL
____________________________________________________________________________________________________________________
11
14. Pieces counting (F..-PCS)
This function enables to count identical pieces, e.g. turnbuckles or buttons. A measurement is performed in two phases:
- first phase - single piece weight calculation on the basis of defined pieces amount (5, 10, 20, 50, 100, 200 or 500 pieces),
- second phase – pieces counting.
It is advised that single piece weight is not less than one reading unit and sample weight used in first phase is bigger than 100 reading units.
To leave function press MENU key and then with “→Τ←” key chose “F..-PCS
and “F..-0”. Note:
Err-3 communicate signalises that a sample was not put on the pan. The same communicate appears if single piece weight is less than one reading unit (it is possible to count pieces but measuring error is bigger). To chose previously used pieces amount select “..” in first phase. During pieces counting T key function does not change.
In scales equipped with LCD display, weighing unit is visible and "" sign is
replaced with "pcs".
USER MANUAL
____________________________________________________________________________________________________________________
12
15. Autozeroing (F..-AUt)
When F..-Aut function is activated, the
scale automatically ensures stable zero indication if the pan is empty or if zero indication was acquired by pressing “
→T←
.” key.
To leave the function press MENU key, then with “
→T←
” key chose “F..-AUt
and “F..-0”.
Note: Autozeroing function is activated automatically for 10 min. after switching-on.
16. Printer cooperation settings (F..-LPt)
Activate the function for automatic serial port working mode (cooperation with a printer). After activation the scale prints a header. Weighing result with a successive measurement number is printed automatically after result
stabilisation (without using key).
To select computer cooperation mode ( key activated and weighing results
without successive numbers) press
MENU key, then with “→T
” key chose
F..- LPt” and “F..- 0”.
USER MANUAL
____________________________________________________________________________________________________________________
13
17. Serial port RS232C working mode setting
The function enables to set the following transmission parameters (standard parameters underlined):
- transmission speed (bAud: 1200, 4800,
9600),
- the number of bits in a byte (bit: 7, 8),
- parity control (PArItY: 0, 1; Odd: 0, 1),
- network number when working in
multistand computer system (when working as a single scale the value should be “0”)
- continuous transmission – without using key, approx. 10 results per second
(SEnd: 0, 1).
Default parameters underlined.
To set desired transmission parameters
activate “F.._rS” function, choose
appropriate parameter and press
→T←
key to accept needed parameter value. The example at the left presents how to set transmission speed value to 9600bps.
To leave the function choose “out
option.
0.00
F. . - r S
F. . - 1
F. . - 0
baud
-9600
-4800
b It
nr
out
Odd
PArItY
Se n d
T
T
T
T
T
T
T
T
T
T
USER MANUAL
____________________________________________________________________________________________________________________
14
18. Function Menu customisation (F..-ACt)
This function enables to select special functions that will be displayed after
pressing MENU key. Easy access to the
most useful functions will shorten operation time and make work more comfortable.
Operation sequence shown on the picture, presents how to add RS232C parameters
setting function (F..-rS) to the Function
Menu.
To remove a function from the Function
Menu choose “F..-0” in the last operation.
0.00
F. . - A C t
F. . - 1
F. . - 0
F1 - AUt
F. . - r S
T
T
T
USER MANUAL
____________________________________________________________________________________________________________________
15
19. Maintenance and repairs of small defects
1. The scale should be kept clean.
2. Take care that no dirt gets between the platform and the scale base. If found any, remove the pan (lift it up), remove dirt and then replace the pan.
3. In case of improper operation caused by short-lasting power supply decay, unplug the scale from the mains and then plug it again after few seconds.
4. If the scale is switched on with empty pan and “Err-b” communicate appears, the
load cell has been mechanically damaged.
5. It is forbidden to make any repairs by unauthorised persons.
6. To repair the scale, please contact our nearest service.
Communicate Possible cause Remedy
C-1 ... 6
(over 1 min.)
selftests failed
if displayed more than 1 minute, contact an authorised service
the scale was switched on with loaded pan
remove a load from the pan
mechanical damage of the load cell
contact an authorised service
L
pan missing
put the pan on
mechanical damage
contact an authorised service
H
overloading
remove the load from the pan
mechanical damage
contact an authorised service
indicator does
not appear
unstable ground vibrations air flows
place the scale on a
stable ground not affected by mechanical vibrations
and airflows
scale is damaged
contact an authorised service
- - - - - -
taring in progress
as above
- -
taring could not be finished (the load is too small or B\G key was used)
zero the scale or press B\G key again
- -
the load is too big to be zeroed
tare the scale (→T←)
USER MANUAL
____________________________________________________________________________________________________________________
16
REFERENCE BALANCE
1. General information
LSM series balances are destined for high accuracy laboratory weighing where EC verification is not required. Additionally, balances may be equipped with special functions for specific applications:
- PCE-LSM..A series – with recipe ingredients summing function for drugstores,
- PCE-LSM..F series – for precise paint weighing in automatic dosing process.
All balances are metrologically tested - calibration or EC verification certificate on demand.
To ensure highest accuracy and reliable results, periodical testing and calibration is advised. To calibrate the balance use proper calibration weight – II or III class weight standard (F1 or F2 OIML class) as stated in Technical Data table (available with extra charge).
2. Completeness
A standard set consists of:
1. Balance
2. Feeder (ZN12V/500mA)
3. Guarantee card
4. User Manual
3. Technical Data
Type
PCE-LSM200
PCE-LSM2000
PCE-LSM 6000
Capacity (Max)
200g
2000g
6000g
Min load (Min)
0,02g
0,5g
5g
Reading unit (d)
0,001g
0,01g
0,1g
Verification unit (e)
0,01g
0,1g
1g (0,5g)
Tare range
- 200g
- 2000g
- 6000g
Accuracy class
II
Working temperature
+18 ÷ +33 °C
Weighing time
<5s
Pan dimensions
φ115mm
225 x 165 mm
Balance base dim.
240x275x90mm
Power
~230V 50Hz 6VA / =12V 160mA
Balance weight
4kg
Calibration weight (OIML)
F2 200g
F2 2000g
F2 2000g
USER MANUAL
____________________________________________________________________________________________________________________
17
4. Keys and indicators
key
I/
- switch On / switch Off (standby),
Indicator
Max, Min, d, e, II
- Meteorological parameters,
key
- Calibration / Acceleration of the examination
key
- Print,
key
T
- Taring function/confirmation of the input
key
MENU
- Entrance into the MENU /special functions
Display
OFF
- signals switch- Off the balance with the key (standby),
Display
- Special funktions (%; ct; n; pcs )
Display
0
- active automatic zero *,
Display
- load indicator (0-100%),
- Enlargement of the indicated number
- Comma,
→T
- Shift around next place,
MENU - Conclusion of storage.
OFF AUTMODE
0
50
100
SUM
Max Min e d
USER MANUAL
____________________________________________________________________________________________________________________
18
5. Security rules
To avoid electrical shock or damage of the balance or connected peripheral devices, it is necessary to
follow the security rules below.
All repairs and necessary regulations can be made by authorised
personnel only.
To avoid fire risk use a feeder of an appropriate type (supplied
with the balance). Pay attention that supply voltage is compatible
with specified technical data.
Do not use the balance when its cover is opened.
Do not use the balance in explosive conditions.
Do not use the balance in high humidity.
If the balance seems not to operate properly, plug it out of the
mains and do not use it until checked by authorised service.
6. Environment protection
According to legal regulations it if forbidden to
dispose wasted electronic equipment in waste
containers.
Please return wasted balance to the point of purchase or other company specialised in recycling of wasted electronic components.
(WEEE-Reg.-Nr. DE64249495).
USER MANUAL
____________________________________________________________________________________________________________________
19
7. Preparing the balance to work
1. Take all content out of a package.
2. Put the pan on the mandrel sticking out of the balance cover (except LSM10 and LSM12 – simple place the pan on pan supports).
3. Level the balance with rotating rear legs so that the air bubble in water-level at the back of the balance is in the middle.
4. Place the balance on a stable ground not affected by mechanical vibrations and airflows.
5. Plug a feeder plug into 12V socket placed at the back of the balance.
Correct Wrong
Leave the pan empty and plug a feeder to the mains. After self-tests and result stabilisation, the balance displays zero indication.
8. Operation principles
1. Weighed load should be placed in the middle of the pan.
2. The balance is equipped with a tare equal to its range. To tare the balance press "
→T←
" key. Storing a tare value does not extend measuring range, but only subtracts it from a load placed on the pan. To make weight control easier and to avoid range overdrawing, the balance is equipped with a load indicator (graduated in percentages).
3. Weighing result should be read when the indicator " " lights, which signalises stabilisation of a result.
4. When the balance is not used but should be ready to work immediately, it can be switched off by pressing " " key. The balance reading system is then switched off and the balance is in "standby" mode (signalled by the indicator "OFF"). To switch the balance on press "
→T←
" key. The balance is immediately ready to
operate maximum accuracy (after self tests).
5. The balance mechanism is a precise device, sensitive to mechanical strokes and shocks.
6. The balance should not be used to weigh ferromagnetic materials due to accuracy decrease.
USER MANUAL
____________________________________________________________________________________________________________________
20
9. Testing and calibration
To ensure reliable results it is advised to check balance accuracy with a calibration weight before and after each measuring session. It is necessary to calibrate the balance when permissible error.
10. Connecting the balance to a computer or a printer
The balance may send data to a computer or a printer through RS232C port.
When cooperating with the balance, a computer should be equipped with a program which enables processing data from the balance. Demo versions and program descriptions are available on the website:
We also offer:
Computer wires
Thermal printers
Label printers
Do not overload a balance more then 20% of maximum load (Max). Do not press the pan with a hand.
During transportation take off the pan (lift it up) and preserve a mechanism mandrel from any damages (excluding LSM10 and LSM12).
Masse
USER MANUAL
____________________________________________________________________________________________________________________
21
Serial port working modes:
Standard mode
The balance sends weighing result after initialising signal from a computer or after pressing “P” key.
Automatic mode
(when cooperating with a printer)
When cooperating with a printer data is send automatically after result stabilisation; next transmission is possible after removing previously weighted sample.
To select serial port working mode use – LPt- function (see further part of the manual).
Standard transmission parameters: 8bits, 1stop, no parity, 4800bps. To change
transmission parameters use –rS- function (see further part of the manual).
Data transmission protocol description ( Long protocol )
1. Standard mode:
ComputerBalance: initialising signal S I CR LF (53h 49h 0Dh 0Ah) Balance Computer: balance indication in the following format (16Bytes)
Further Bytes:
Byte
1 - sign or space
Byte
2, 11&14
-
space
Byte
3÷4
-
digit or space
Byte
5÷9
-
digit, decimal point or space
Byte
10 - digit
Byte
12 - k, l, c, p or space
Byte
13 - g, b, t, c or %
Byte
15 - CR (0Dh)
Byte
16 - LF (0Ah)
USER MANUAL
____________________________________________________________________________________________________________________
22
Connection cable WK-1 (balance – computer / 9-pin):
Connection cable WD-1 (balance - PCE-BP1 printer):
PCE-BP1 printer internal switches set-up:
SW-1
SW-2
SW-3
SW-4
SW-5
SW-6
SW-7
SW-8
on
off
on
off
off
on
off
off
11. Special functions description
All balances of the type PCE LSM order except the meteorological basic functions according to standard: taring and the weighing also still over an additional pallet at special functions how: Item counting, Auto-zero and the attitude of the RS- interface including the prescription weighing and the weighing in carat. After pressing of the key
MENUE can you see the broad pallet at special functions of the balance. They know the
functions and at the most for itself to activate accordingly. For the attitudes the function
ACtIV” is intended. During the switching procedure the message appears “MODE” on
the announcement.
The indication „hand” clarifies the respective proceeding.
BALANS
PRINTER
BALANS
COMPUTER
USER MANUAL
____________________________________________________________________________________________________________________
23
23.04 g
0.00 g
43.04 g
23.13 g
0.00 g
63.24 g
T
T
12. Legend
- press a key
- balance indication
- forced change
- automatic change
13. Normal weighing
When the pan is empty and indication is different than zero press "
→T←
" key.
Weighing result should be read when
the indicator " " lights.
14. Weighing with tare
The balance is equipped with tare equal to its range.
USER MANUAL
____________________________________________________________________________________________________________________
24
15. Sensitivity calibration (F..-CAL)
To calibrate the balance use proper calibration weight as stated in Technical Data table. Calibration with a weight standard smaller than advised affects in accuracy decrease. Calibration process must be performed very carefully because wrong calibration parameter values stored in memory affects balance accuracy.
calibration:
Switch the change into On-position
( illustration nr.1 ) . With the help of the
key you call MENUE for the user
functions. Press the key
→T←
during
the function „ CALIbr” are indicated. Press the key - - around the value of
the test weight indicate sereval times to let, which is to be used for the calibration. Their selection you can confirming by
→T←
you press the key.
After that the comunicate "LOAD"
appeared, presents you the test wight, and confirms you choice with the
→T←
key.
Wait to the adjustment are final.
After the appearance of the comunicate
"unLOAD" your remove the test wight.
With calibrated balances, you switch the adjustment switch, which causes an
expiring of the "Pr ON" comunicate .
T
m
m
unLOAD
m
m
--------
LO A D
CC
- C A L-
1
2
...
M ENU
T
T
0.00g
0.00g
C ALIb r
C AL o n
C AL o FF
...
ON
OFF
Illustration nr.1
USER MANUAL
____________________________________________________________________________________________________________________
25
16. Piece counting (F..-Pcs)
This function serves the counting of e.g. Screws, buttons etc. The wighing (item counting) takes place in two phases:
- first phase: determination of the mass of detail with the help of a samble of a certain number of pieces like e.g.: 5, 10, 20, 50, 100, 200 or 500 Pcs,
- second phase: item counting in the quantity which can be wighed.
Nr.1: Balancing mode [g]- put up the
detail quantity e.g. 5-screws on the balancing plate. The balances indicates the wight.
Nr.2: Balancing mode [pcs]- pressures
the key MENUE, you select with the
key
→T←
the option PcS. On the
display is indicated, the number of items and pieces (5,...,500 pieces). With the help of the key
→T←
you on
select your presented number of items e.g. 5-pieces. Now, you can begin your item counting
Over to terminate the piece counting you press the key MENUE and select afterwards the key
→T←
, you: "PCS"
and "PCS oFF". Attention:
Err-3 communicate signalises that a sample was not put on the pan. The same communicate appears if single piece weight is less than one reading unit (it is possible to count pieces but measuring error is bigger). To chose previously used pieces amount select “..” in first phase. During pieces counting –T- key function does not change. When counting very small details it is advised to deactivate autozeroing function.
0 pcs
0.00 g
12.05 g
23.04 g
PC S
PC S o FF
PC S o n
5
5
12
12 pcs
M ENU
T
T
T
T
USER MANUAL
____________________________________________________________________________________________________________________
26
0.01g
AUtOtAr
AUT
AUt o n
AUt o FF
0.00g
0.00g
LPt
LPt o n
LPt o FF
0.00g
M ENU
M ENU
T
T
T
T
17. Autozeroing (F..-AUt)
The activation of the function causes an automatic maintrance of the zero position of the balance begins in the time in that the platform is not loaded or if the zero position took place by the pressing of the key
→T←
. Over to lock
you press the function the key MENUE
and select with the help of the key
→T←
, you "AUtOtAr" and afterwards
"AUtoFFr".
Note: After switching on of the balance,
is the function 10 min. long activated.
18. Serial port working mode (F..-LPt)
With the activation of the function one
knows the attitudes of the LPt- port
haven, makes in the automode (printer). After presenting and removing the sample of the platform of the balance an automatic dispatching and the expression of the data at the printer take place like e.g.: the result and the weighing number. In order into the normal balancing mode back to arrive ,
press the key MENUE and afterwards
the key
→T←
and in the following the
key "LPt" and "LPt oFF”. The result
expression takes place then after pressing the key and the
stabilization of the balance.
USER MANUAL
____________________________________________________________________________________________________________________
27
19. Serial port transmission parameters
The function enables to set the following transmission parameters (standard parameters underlined):
- transmission speed (bAud: 1200, 4800,
9600),
- the number of bits in a byte (bit: 7, 8),
- parity control (PArItY: 0, 1; Odd: 0, 1),
- continuous transmission – without using
, key, approx. 10 results per second
(SEnd: 0, 1).
Parameter the according to standard are unterlined.
To set desired transmission parameters
activate "rS232" function, choose
appropriate parameter and press
→T←
key to accept needed parameter value. The example at the left presents how to set transmission speed value to 9600bps.
To leave the function, choose "out "
option.
0.00g
r S2 3 2
baud
-9600
-4800
b It
out
Odd
PArItY
Se n d
M ENU
T
T
T
T
T
T
T
T
USER MANUAL
____________________________________________________________________________________________________________________
28
20. Per cent weighing ( PerCEnt )
This function permits a separate weighing in per cent. The process takes place in two phases:
1. Weighing of the 100% single mass
2. Weighing of any portion of a mass as % - portion of the measures weighed in the first phase.
In the dependence on the accepted mass as test weight [%] – the alignment result is indicated to the weighed mass in different formats. For the test load within the range 0÷3,5% of the balancing range the result in form "100,0", within the range becomes 3,5÷35% - "100,00", and above 35% - " 100,000 ".
This function has the following options:
-"PEr oFF" – switching the function off,
-"PEr on" – storage of the current
indicated value as 100% transition to
the annoucement to [%].
-"PEr Con"- function remains actively,
independently of it is in which mode one momentarily.
Note:
In the time in that the balance in [%] –
indicates, remains the key
→T←
unchanged in its function.
0.00 g
19.07 g
PErC En t
PEr o FF
PEr o n
100.00%
50.00%
100%
50%
M ENU
T
T
USER MANUAL
____________________________________________________________________________________________________________________
29
21. Change of the units gram/carat ( CarAt )
22. Actualization of the special functions ( ActlV )
1.60g
CArAt
ct
CAr o n
C a r o FF
8.00
M ENU
T
T
0.00
ACtIV
r S o n
r S o FF
AUtOtAr
r S2 3 2
T
T
T
This function possible makes the
selection for a set of special functions, which are to be indicated to the key
MENUE after pressing. This is to serve
to reduce the whole pallet at axisting
functions. The user selects only the functions, which he needs for his
applications.
In order to differentiate the function
"ACtIV" from the MENUE-list to, is on
the left side the indication set.
On the picture an example is shown, how one RS 232 –
function into
(choice) the function menue integrates.
Around the function from (choice) the
function menu to exclude select the
option "rS on " and "rS oFF " .
Switching on the function on causes the change of the balancing unit between carat/gram (1 carat=0,2g).
Suggested this is with ”ct
” by the
indication on the right side of the
announcement.
USER MANUAL
_____________________________________________________________________________________________________________________
30
23. Attitude of the date (dAtE)-function
0.00g
d AtE
d A t SEt
d At o FF
h13 - 18
h13 - 1
8
h13 - 1
9
h13 - 19
0.00g
d04 - 05
r - 06
- -
M ENU
T
T
T
T
This function permits the attitude of the
current date and the time to the balance. This function contains the following options:
- "dAt oFF
" deactivation of the date
and the time in the
expression.
- "dAt on
" activation of the data and
the time in the expression.
( key )
- "dAt SEt" change of the current of the
Date and the time.
The example shows the application of
the option "dAt Set” .
USER MANUAL
_____________________________________________________________________________________________________________________
31
24. Error messages
Communicate Possible cause Remedy
"C-6"
selftests in progress
/ electronic unit damage
wait 1 minute
" - - - - "
zeroing in progress
/ mechanical damage
wait 1 minute
check if the balance is stable,
not affected by any vibrations
"L"
to small zero load
/ mechanical damage
check if there are
all necessary pan elements
"H"
overloading
/ mechanical damage
remove a load from the pan
"Err-b"
wrong initial load at start-up
/ mechanical damage
remove a load from the pan
and restart the balance
If a remedy does not have any effect and the communicate is still displayed, contact your dealer or service centre.
25. Maintenance and repairs of small defects
1. The balance should be kept clean.
2. Take care that no dirt gets between a casing and the pan. If found any, remove the pan (lift it up), remove dirt and then replace the pan.
3. In case of improper operation caused by short-lasting power supply decay, unplug the balance from the mains and then plug it again after few seconds. If “CALIb” communicate appears, calibrate the balance or contact a service centre.
4. It is forbidden to make any repairs by unauthorised persons.
To repair the balance, please contact our nearest service.
USER MANUAL
_____________________________________________________________________________________________________________________
32
COUNTING SYSTEM
Technical Data
CAPACITY BALANCE
REFERENCE BALANCE
Max. e d
d/ Piece
Max.
e d d/ Piece
[kg]
[g]
[g]
[g] [kg]
[g]
[g]
[g]
PCE-IS 01
60 - 20
0,8 2000
-
0,01
0,01
PCE-IS 02
150 - 50 2
2000
-
0,01
0,01
PCE-IS 03
300
-
100
4 2000
-
0,01
0,01
PCE-IS 01
60
20
20
0,8 2000
0,1
0,01
0,01
PCE-IS 02
150
50
50 2
2000
0,1
0,01
0,01
PCE-IS 03
300
100
100 4
2000
0,1
0,01
0,01
USER MANUAL
_____________________________________________________________________________________________________________________
33
Put up the sample with a defined number of elements (Reference number of items) on the plate of the reference balance (see illustration nr.1) Press you on the quantity
balance the key MENU and
switching and select the function
„F. - PC” and ” F. - 1”.
Select the option ”rS”.
Register with the help of the keys the number of elements (reference number of items) in the sample.:
→0← - Number up,
T - Transition to the next
number,
MENU - Conclusion.
Placing the number of you elements (reference number of items) on the plate of the quantity balance and read you result.
In order to lock the function, you
press the key MENU and after-
wards with the help of the key
T select you „F-PC”and”F- 0”.
Fn -PC
USER MANUAL
_____________________________________________________________________________________________________________________
34
Declaration of Conformity APPROVAL of EC NO. TYPE PL 06 004
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con disitintivo CE
Dichiarazione di coformitá per apparecchi contrassegnati con la marcatura CE
English We hereby declare that the product to which this declaration refers conforms
with the following standards.
Deutsch Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit
den nachstehenden Normen übereinstimmt.
Français Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la
présente déclaration, est conforme aux normes citées ci-après.
Español Manifestamos en la presente que el producto al que se refiere esta declaración
est´´a de acuerdo con las normas siguientes
Italiano Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito citate.
PCE-PM (60)62, PCE-PM(150)152, PCE-PM(300)302
1.EN 55022 standard Limits and methods of measurement of radio disturbance characteristics of
information technology equipment and IEC 61000-4-3 Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test, harmonised with the Council Directive 89/336/EEC
2.EN 61010-1:2004 standard Safety requirements for electrical equipment for measurement, control and laboratory use. General requirements harmonized with the directive 73/23/EEC (Low Voltage Directive).
3.PN-EN 45501 – the main polish and european standard concern Non-automatic weighing instruments, its metrological parameters, measuring of ranges, accuracy, influence of electromagnetic fields, disturbance of short electric impulse, influence of ambient temperature and humidity, gravity etc. subject to conformity implementing Council Directive 90/384/EEC amended by Council Directive 93/68/EEC
Signature: Date: 02.08.2006
PCE-GROUP Europe OHG Management
Zur Umsetzung der ElektroG (Rücknahme und Entsorgung von Elektro- und Elektronikaltgeräten) nehmen wir unsere Geräte zurück. Sie werden entweder bei uns wiederverwertet oder über ein Recyclingunternehmen nach gesetzlicher Vorgabe entsorgt.
-Reg.-Nr. DE64249495
USER MANUAL
_____________________________________________________________________________________________________________________
35
Declaration of Conformity APPROVAL NO. of EC PL 06 007
PCE-LSM 200, PCE-LSM 2000, PCE-LSM 6000
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con disitintivo CE
Dichiarazione di coformitá per apparecchi contrassegnati con la marcatura CE
English We hereby declare that the product to which this declaration refers conforms
with the following standards.
Deutsch Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit
den nachstehenden Normen übereinstimmt.
Français Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la
présente déclaration, est conforme aux normes citées ci-après.
Español Manifestamos en la presente que el producto al que se refiere esta declaración
est´´a de acuerdo con las normas siguientes
Italiano Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce
è conforme alle norme di seguito citate.
marked with CE mark comply the following:
1. EN 55022 standard Limits and methods of measurement of radio disturbance
characteristics of information technology equipment and IEC 61000-4-3 Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test harmonized with the Council Directive 89/336/EEC (Electromagnetic compatibility directive).
Date: 06.10.2006
Signature: _____ __________
PCE-GROUP Europe OHG
Management
Zur Umsetzung der ElektroG (Rücknahme und Entsorgung von Elektro- und Elektronikaltgeräten) nehmen wir unsere Geräte zurück. Sie werden entweder bei uns wiederverwertet oder über ein Recyclingunternehmen nach gesetzlicher Vorgabe entsorgt.
WEEE-Reg.-Nr. DE64249495