User manuals in various languages (français,
italiano, español, português, nederlands, türk, polski,
русский, 中文) can be found by using our
product search on: www.pce-instruments.com
Last change: 1 July 2021
v1.0

Contents
1 Safety notes ........................................................................................... 1
2 Features ................................................................................................. 1
3 Specifications ........................................................................................ 2
3.1 Delivery contents .............................................................................................................. 3
4 Device ..................................................................................................... 3
5 Operation ............................................................................................... 4
5.1 pH calibration ................................................................................................................... 4
5.2 Calibrate conductivity ................................................................ ....................................... 4
5.3 pH measurement .............................................................................................................. 4
5.4 Conductivity measurement ............................................................................................... 5
5.5 Redox measurement ........................................................................................................ 5
6 Key functions ........................................................................................ 5
7 Battery .................................................................................................... 6
8 Contact ................................................................................................. 6
9 Disposal ................................................................................................. 6
© PCE Instruments

© PCE Instruments
1
1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The
device may only be used by qualified personnel and repaired by PCE Instruments personnel.
Damage or injuries caused by non-observance of the manual are excluded from our liability and
not covered by our warranty.
• The device must only be used as described in this instruction manual. If used otherwise,
this can cause dangerous situations for the user and damage to the meter.
• The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
• Do not expose the device to shocks or strong vibrations.
• The case should only be opened by qualified PCE Instruments personnel.
• Never use the instrument when your hands are wet.
• You must not make any technical changes to the device.
• The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
• The device must only be used with accessories from PCE Instruments or equivalent.
• Before each use, inspect the case for visible damage. If any damage is visible, do not
use the device.
• Do not use the instrument in explosive atmospheres.
• The measurement range as stated in the specifications must not be exceeded under
any circumstances.
• Non-observance of the safety notes can cause damage to the device and injuries to
the user.
We do not assume liability for printing errors or any other mistakes in this manual.
We expressly point to our general guarantee terms which can be found in our general terms of
business.
If you have any questions please contact PCE Instruments. The contact details can be found at
the end of this manual.
2 Features
• 6 in 1 pH tester
• IP57 housing
• Automatic temperature compensation
• Battery operation
• 3-point pH calibration
• Exchangeable electrodes
• Max-Min hold function
• Automatic electrode recognition

© PCE Instruments
2
3 Specifications
0 ... 90°C / 32 ... 194°F
0 ... 2000 μS, 2 ... 20 mS
± 2% of measurement range
0 ... 50°C / 32 ... 122°F
0 ... 1300 ppm, 1.3 ... 13 ppt
± 2% of measurement range
0 ... 50°C / 32 ... 122°F
0 ... 1000 ppm, 1 ... 12 ppt
± 2% of measurement range
0 ... 50°C / 32 ... 122°F
0 ... 90°C / 32 ... 194°F
195 x 40 x 36 mm / 7.7 x 1.6 x 1.4 in

© PCE Instruments
3
3.1 Delivery contents
1 x pH meter PCE-PH 30
1 x pH electrode
1 x conductivity electrode
1 x pH 4 calibration liquid
1 x pH 7 calibration liquid
1 x 1413 μS / cm calibration liquid
1 x storage solution
1 x wrist strap
4 x 1.5 V AAA battery
1 x user manual
4 Device
Battery compartment
Display

© PCE Instruments
4
5 Operation
Remove the black protective cap from the device and optionally unscrew the electrode cap if you
want to use a different electrode. To change the electrode, you can simply remove it and attach
another one.
Note: Do not touch the sensors of the electrodes!
5.1 pH calibration
Since the electrode is subject to a time drift, it should be calibrated regularly. The interval depends
on the intended use and the required accuracy.
• Make sure that the pH electrode is mounted and the pH icon is shown on the display.
• Clean the electrode with distilled water.
• Immerse the electrode in the calibration solution pH 7, stir carefully and wait until the
value has stabilized.
• Press and hold the CAL key until "Cal" and the calibration value are displayed.
• Calibration is performed automatically and is finished when "End" is displayed. The
instrument returns to measuring mode.
• Clean the electrode with distilled water.
• Now repeat the calibration with the pH 4 calibration solution.
Notes:
• Always carry out a two-point calibration. Start at pH 7, followed by pH 4 (or 10).
• When not using the pH electrode, always keep the electrode moist with the storage
liquid.
5.2 Calibrate conductivity
1. Make sure that the conductivity electrode is mounted to the instrument. The display
shows COND, TDS or Salt.
2. Immerse the conductivity electrode in the 1413µS/cm calibration solution and wait until
the measured value has stabilized.
3. Press and hold the CAL button until "Cal" is displayed and the calibration value flashes.
Calibration is performed automatically and is finished when "End" is displayed. The
instrument returns to measuring mode.
Notes:
• A calibration with 12.88mS/cm can also be performed optionally. It is suitable for
measurements with high conductivity.
• If the measured value is not 0µS/cm in air after calibration, calibrate the instrument in
air.
5.3 pH measurement
1. Clean the pH electrode with distilled water.
2. Dip the electrode tip into the liquid to be examined and move the tip slightly back and
forth.
3. Wait briefly until the display shows a stable pH value.

© PCE Instruments
5
5.4 Conductivity measurement
1. Clean the conductivity electrode with distilled water.
2. Dip the electrode tip into the liquid to be examined and move the tip slightly back and
forth.
3. Wait briefly until the display shows a stable value.
Notes:
• The display shows "----- " if the value exceeds the measurement range.
• The measuring unit changes automatically depending on the measurement range.
• Clean the electrode with distilled water after the measurement.
5.5 Redox measurement
1. Make sure that the Redox electrode is mounted on the instrument. The display
automatically shows "ORP". Calibration is not necessary but can be performed to check
the ORP electrode with the appropriate calibration solution.
2. Clean the electrode with distilled water and immerse it in the liquid to be measured.
3. Wait for a short time until a stable value is displayed.
Notes:
• The display shows "----- " if the value exceeds the measurement range.
• When not using the electrode, always keep the electrode moist with the storage liquid.
•
6 Key functions
On/off / calibration key
1. Press and release the key to switch the unit on or off.
2. Press and hold the key while the instrument is switched on to enter calibration mode.
Measuring key
1. Press the key to switch between the measuring functions conductivity, TDS and
salinity. This is only possible with the conductivity electrode!
2. Press and hold the key to toggle between degrees Celsius and degrees Fahrenheit.
Hold/Max&Min key
1. Press and release the key to freeze the displayed value.
2. Press and hold the key to enter the max/min selection. Press and release the key
again to toggle between Max and Min.
3. Press and hold the key again to exit the Max/Min mode.
Note:
• The display does not switch off automatically in Max/Min mode.

7 Battery
When the battery icon flashes, replace the batteries as follows:
1. Loosen the 2 screws at the top of the battery compartment.
2. Open the battery compartment and replace the 4 AAA batteries. Observe correct
polarity.
3. Close the battery compartment and tighten the screws.
8 Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact
us. You will find the relevant contact information at the end of this user manual.
9 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament
applies. Due to the contained pollutants, batteries must not be disposed of as household waste.
They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use
them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with
your local waste regulations.
If you have any questions, please contact PCE Instruments.
© PCE Instruments
6

© PCE Instruments
7
PCE Instruments contact information
Germany France Spain
PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L.
Im Langel 4 23, rue de Strasbourg Calle Mayor, 53
D-59872 Meschede 67250 Soultz-Sous-Forets 02500 Tobarra (Albacete)
Deutschland France España
Tel.: +49 (0) 2903 976 99 0 Téléphone: +33 (0) 972 3537 17 Tel. : +34 967 543 548
Fax: +49 (0) 2903 976 99 29 Numéro de fax: +33 (0) 972 3537 18 Fax: +34 967 543 542
info@pce-instruments.com info@pce-france.fr info@pce-iberica.es
www.pce-instruments.com/deutsch www.pce-instruments.com/french www.pce-instruments.com/espanol
United Kingdom Italy Turkey
PCE Instruments UK Ltd PCE Italia s.r.l. PCE Teknik Cihazları Ltd.Şti.
Unit 11 Southpoint Business Park Via Pesciatina 878 / B-Interno 6 Halkalı Merkez Mah.
Ensign Way, Southampton 55010 Loc. Gragnano Pehlivan Sok. No.6/C
Hampshire Capannori (Lucca) 34303 Küçükçekmece - İstanbul
United Kingdom, SO31 4RF Italia Türkiye
Tel: +44 (0) 2380 98703 0 Telefono: +39 0583 975 114 Tel: 0212 471 11 47
Fax: +44 (0) 2380 98703 9 Fax: +39 0583 974 824 Faks: 0212 705 53 93
info@pce-instruments.co.uk info@pce-italia.it info@pce-cihazlari.com.tr
www.pce-instruments.com/english www.pce-instruments.com/italiano www.pce-instruments.com/turkish
The Netherlands Hong Kong United States of America
PCE Brookhuis B.V. PCE Instruments HK Ltd. PCE Americas Inc.
Institutenweg 15 Unit J, 21/F., COS Centre 1201 Jupiter Park Drive, Suite 8
7521 PH Enschede 56 Tsun Yip Street Jupiter / Palm Beach
Nederland Kwun Tong 33458 FL
Telefoon: +31 (0)53 737 01 92 Kowloon, Hong Kong USA
info@pcebenelux.nl Tel: +852-301-84912 Tel: +1 (561) 320-9162
www.pce-instruments.com/dutch jyi@pce-instruments.com Fax: +1 (561) 320-9176
www.pce-instruments.cn info@pce-americas.com
www.pce-instruments.com/us