PCE Instruments PCE-PH 25-ICA, PCE-PH 25 Users guide

Page 1
Operation Manual
PCE-PH 25
Page 2
1
CONTENTS
INITIAL INSPECTION AND ASSEMBLY ........................2
BEFORE YOUR FIRST USE ........................…...................2
A. Replace the Batteries ....................................................2
B. Soak the Electrode.... ....................................................2
C. Setup and Calibrate the Electrode and Meter…..........2
OVERVIEW PCE-PH 25 …………....................................3
A. Meter Description...........................................................3
B. VisionPlus Graphical Display .......................................3
OPERATION MODES AND KEYPAD OPERATIONS 4
A. Operation Modes............................................................4
B. Keypad Operations .......................................................4
USING PCE-PH 25……………............................................6
A. Power ON/OFF .............................................................. 6
B. Temperature Unit Select Mode................................... 6
C. pH Calibration............................................................... 7
D. Conductivity Calibration.............................................. 8
E. Measure ..........................................................................10
F. Hold Data ........... ......................................................... 10
G. Save Data ...........................................................…...... 10
H. Recall Data ...................................................................11
I. Erase Data......................................................................11
ONGOING MAINTENANCE...............................................12
A. Replace Electrode........................................................12
ERROR DISPLAYS AND TROUBLESHOOTING .......12
SPECIFICATIONS ................................................................13
WARRANTY ...........................................................................16
Page 3
2
INITIAL INSPECTION AND ASSEMBLY
Carefully unpack the instrument and accessories. Inspect for damages made in shipment. If any damage is found, notify th producer immediately. All packing materials should be saved until satisfactory operation is confirmed.
BEFORE YOUR FIRST USE
A. Replace the Batteries
1. Take off the battery cover as shown in the right figure.
2. Remove all of the old batteries and insert a new set of batteries ensuring the polarities are correct.
B. Soak the Electrode
1. Remove the electrode cap
covering the PCE-PH 25 meter.
2. Soak the electrode in a pH 4 solution for 10 minutes before first use or after storage.
C. Setup and Calibrate the Electrode and Meter
PCE-PH 25 must be setup and calibrated before your first use. Please follow the instructions detailed in section
USING pH-Meter PCE-PH 25.
Page 4
3
PCE-PH25 OVERVIEW
A. Meter Description
1. Battery cap
2. LCD screen
3. Keypad
HoId
Store
Mode
ScroII
4. Electrode collar
5. Electrode & ATC assembly (Electrode cap is not shown.)
B. Graphical Display
7.00
1
HOLD
2
10.00
M1
3
125.2
pH
Fã¡
5
6
7
4
BUF:7/4/10
REC
8
1. pH BUFFER mode indicator
2. HOLD mode indicator
3. DATA NUMBER indicator
4. CALIBRATION buffer(s)
5. BATTERY indicator
6. pH / CONDUCTIVITY / TDS / SANILITY reading
7. TEMPERATURE reading
8. RECALL DATA mode indicator
Page 5
4
OPERATION MODES AND KEYPAD OPERATIONS
A. Operation Modes
The pH-meter has 6 operation modes:
1. Measure Mode. Measure Mode is used to make pH / conductivity / TDS / salinity and temperature measurements.
2. Calibration Mode. Calibration Mode is used to perform 1, 2 or 3 point pH calibration and conductivity calibration.
3. Hold Mode. Hold Mode is used to display held measured values for increased ease of use.
4. Temperature display set mode. This Mode is used to select °C or °F.
5. pH Buffer Select Mode. Buffer Select Mode is used to select the buffer set, which can either be 7.00(7.00/4.01/10.01) or
6.86(6.86/4.00/9.18).
6. Recall Data Mode. Recall Data mode is used to display measured data which have been stored in the memory.
B. Keypad Operations
Key
Operation
Mode
Duration
Function
Hold
All, except Hold
1 second
Turns meter on/off.
Conductivity Calibration
0 second
Saves the Calibration reading.
Recall Data
0 second
Returns to Measure Mode.
Measure
0 second
Holds current measurement reading. Press again to resume measuring.
Hold
0 second
Returns to Measure Mode.
Page 6
5
Key
Operation
Mode
Duration
Function
Mode
Measure
0 second
Selects display mode. Pressing this key changes the display sequentially to display pH/Temp, Conductivity/Temp, TDS/Temp and Salinity/Temp modes.
pH Measure
2 seconds
Enters pH Calibration mode.
Conductivity Measure
2 seconds
Enters Conductivity Calibration mode.
TDS Measure
10 seconds
Selects Temperature °C or °F mode.
Recall Data
0 second
Selects whether or not to erase all stored data.
Store
Measure
2 seconds
Stores current pH/Temp, Conductivity/Temp, TDS/Temp and Salinity/Temp reading into memory.
Temperature display mode Selects
0 second
Selects Temperature °C
pH Buffer Select
0 second
Selects 7.00 Buffer Sets and Leaves Buffer Select Mode.
Recall Data
0 second
Decreases current number.
Page 7
6
Key
Operation
Mode
Duration
Function
Scroll
Measure
2 seconds
Enters Recall Data Mode.
Temperature display mode Selects
0 second
Selects Temperature °F
pH Measure
5 seconds
Enters pH Buffer Select Mode.
pH Buffer Select
0 second
Selects 6.86 Buffer Sets and Leaves Buffer Select Mode.
Recall Data
0 second
Increases current number.
USING PCE-PH 25
A. Power ON/OFF
1. Press and hold “Hold” key for 1 second. The unit will turn on and enter the “Measure Mode”. Repeat the process to
turn off the unit.
2. The unit will also automatically turn off after 10 minutes of no key activity.
B. Temperature Unit Select Mode
1. The PCE-PH 25 meter is factory pre-set in “°C”.
2. To change to °F, press “Mode” key to enter the “TDS Measure Mode”.
3. Press and hold the Mode” key for 10 seconds, the unit will enter the Temperature Unit Select Mode.
4. Press “Store” key to select “°C” and press Scroll key to select °F”.
TEMPERATURE UNIT
SELECT
MODE: EXIT
SCROLL: ¡ãF
STORE: ¡ãC
Page 8
7
C. pH Calibration
a. pH Buffer sets Select
1. In the pH Measure Mode, press and hold the Scroll” key for 5 seconds to enter the “pH Buffer sets Select Mode.
2. Press Store key to select 7.00 (7.00/4.01/10.01) or press Scroll key to select 6.86 (6.86/4.00/9.18) .
3. Press Mode key to exit and return to pH Measure Mode.
SELECT pH BUFFER
MODE: EXIT
SCROLL: 6.86
STORE: 7.00
Note: There is no need to repeat this procedure every time unless one decides to change the buffer settings.
b. pH Calibration
1. Press “Mode” key to enter the “pH Measure Mode”.
2. Clean and dip meter into either pH 7.00 or 6.86 buffer solution. Press Mode key and hold for 2 seconds to enter the pH Calibration Mode. A ‘WAIT” icon will flash. The pH/EC80 meter will perform end point sensing to determine when the calibration reading is stable.
BUF:7
7.00
25.2
7.00 WAIT
¡ãC
pH
3. When the “WAIT” icon disappears, one point/first point calibration is now complete. Press Mode” key to exit or to wait 5 seconds for the meter to enter two points / second point calibration.
4. Repeat Steps 2 and 3 for two points/second point calibration by dipping the meter in pH 4.01/4.00 or
10.01/9.18 buffers.
Page 9
8
5. Repeat Steps 2 and 3 for three points / third point calibration.
6. When “WAIT” icon disappears. Three points / third point calibration is now complete. The unit will automatically return to Measure Mode.
Note: For accurate measurements, it is recommended that pH calibration is preformed once a week and after replacing the electrode.
D. Conductivity Calibrate
a. Preparing Standard Solutions Suitable conductivity standards are available commercially or
the user can prepare them using research grade reagents. Here are some standard solutions the user can prepare to
calibrate the probe of the model PCE-PH 25.
1. Standard solution of 1413uS at 25°C: Accurately weight out
0.746 grams of research grade dried Potassium Chloride (KCL). Dissolve in 1000ml of distilled water.
2. Standard solution of 12.90mS at 25°C: Accurately weight
out 7.4365 grams of research grade dried Potassium Chloride (KCL). Dissolve in 1000ml of distilled water.
[Note: You can store the remaining solution in a plastic container for one week but the air space between the cap and the solution must be kept to an absolute minimum. Storing the excess solution below 4°C can increase the storage life. If you have any doubt of the accuracy of the stored solution, a fresh batch should be prepared.]
b. Conductivity Calibration
1. Press “Mode” key to enter the “Conductivity Measure Mode”.
2. Clean and dip meter into Standard Solution. Allow temperature reading to stabilize. Press and hold Mode key for 2 seconds to enter the Conductivity Calibration Mode. The “CAL” icon appears on the LCD.
3. Press “Store”, “Scroll” and “Hold” keys to sequentially view previous calibration settings.
Page 10
9
TEMP.COE.: 1.91%
147
uS/cm
TDS FACTOR: 0.65 TEMP.REF.: 25¡ãc
CAL
25.2
¡ãc
TDS FACTOR: The default factor value is 0.65. To change the TDS factor,
use the “Store” and “Scroll” keys to adjust the value between 0.30 and 1.00. Press “Hold” key to save the new
value and the unit will automatically go into the next
calibration parameter. If “Mode” key is pressed instead of the “Hold” key, any changes made will be cancelled and
the previous calibration settings will be retained.
TEMP. COE.:
The unit uses the temperature coefficient to calculate temperature compensated conductivity. The default value is 1.91%. To change the temperature coefficient, use the
Store” and “Scroll” keys to adjust the value between 0 and 4.00%. Press “Hold” key to save the new value and
the unit will automatically go into the next calibration parameter. If “Mode” key is pressed instead of the “Hold key, any changes made will be cancelled and the previous calibration settings will be retained.
TEMP. REF.:
The unit uses the temperature reference value to calculate temperature compensated conductivity. The default value is 25°C. To change the temperature coefficient, use the Store” and “Scroll” keys to adjust the value between 15°C and 25°C. Press “Hold” key to save the new value and the unit will automatically go into the next calibration parameter. If “Mode” key is pressed instead of the “Hold” key, any changes made will be cancelled and the previous calibration settings will be retained.
4. Immerse the probe in a standard of known conductivity, preferably a standard in the middle range of the solutions to
be measured. Immerse the probe (at least 2” to 3” or
5~7cm from the tip) into standard solution without touching the sides of the calibration container. Shake the probe lightly to remove any air bubbles trapped in the conductivity cell.
Page 11
10
Allow temperature to stabilize. The message “RANG” (range) may appear briefly on the display indicating auto­ranging. This is normal. After temperature stabilization, use the “Store” and “Scroll” keys to adjust the conductivity value to that of the conductivity standard at 25°C. Press “Hold” key to calibrate. The unit displays “CAL OK” to indicate a successful calibration. Conductivity calibration is now complete and the unit will automatically switch to the “Conductivity Measurement” mode.
E. Measure
In the “Measure Mode”, dip the meter into the test solution. Press “Mode” key to select: pH/Temperature, Conductivity/ Temperature, TDS/ Temperature and Salinity/Temperature.
F. Hold Data
1. When the readings is stable, press “Holdkey once to lock the reading.
2. Press “Hold” key again to unlock reading and the unit will return to the “Measure Mode”. The unit is now ready for another measurement.
HOLD
147
uS/cm
25.2
¡ãc
G. Save Data
Press and hold “Store” key for 2-3 seconds in the “Hold Mode” or “Measure Mode”, the “SAVE” and “M-XX” icons will appear indicating the reading has been saved and stored in memory location XX. The unit will automatically return to the “Measure Mode” and ready to perform other tasks.
M2
SAVE
147
uS/cm
125.2
¡ãF
Page 12
11
Note: The Model PCE-PH 25 is equipped with a non-volatile memory that can store up to 50 different sets of readings (pH/Conductivity/TDS/Salinity and temperature). Non-volatile memory will be retained even if power is lost. When all 50 memory locations are used up, the next set of data saved will overwrite the data stored in the first location and so on.
H. Recall Data
1. In “Measure Mode”, press and hold “Scroll” key 2-3 seconds, the “REC” icon appears on the LCD and the last set of saved data will appear.
2. By pressing the “Storeor Scroll key, previous saved data will appear.
3. Press “Hold” to exit and return to “Measure Mode”.
REC
M2
SAVE
147
uS/cm
125.2
¡ãF
I. Erase Data
1. In “Recall Data Mode”, press Modeto enter Erase Data Mode.
2. Press Store” key to erase ALL stored data and return to “Measure Mode.
3. Press Scroll” key to not erase ALL stored data and return to “Measure Mode.
ERASE ALL DATA ?
SCROLL: NO
STORE: YES
Page 13
12
ONGOING MAINTENANCE
A. Replace Electrode
1. Unscrew the electrode collar to
remove the electrode & ATC assembly as shown in the right figure.
2. Remove the old electrode
assembly from the electrode collar.
3. Insert a new electrode assembly
and make sure the electrode fit back into the meter correctly.
4. Screw back the electrode collar.
ERROR DISPLAYS AND TROUBLESHOOTING
Display
Mode
Main
Display
Secondary
Display
Possible cause(s) [Action(s)]
pH
Measure
or
pH
Calibration
UNDR
UNDR
Temperature< -5.0°C . [Bring solution to a higher temperature.] [Replace electrode.]
OVER
OVER
Temperature> 60.0°C . [Bring solution to a lower temperature.] [Replace electrode.]
pH
Calibration
UNDR
0.0 to 60.0°C
Offset @ 7.00pH: mV>100mV; Offset@6.86pH: mV>91.7mV; New slope> ideal slope by +30% [Use a new buffer solution.] [Replace electrode.]
OVER
0.0 to 60.0°C
Offset @ 7.00pH: mV< -100mV; Offset@6.86pH: mV<-108.7mV; New slope<ideal slope by -30% [Use a new buffer solution.] [Replace electrode.]
Page 14
13
Display
Mode
Main
Display
Secondary
Display
Possible cause(s) [Action(s)]
pH
Measure
UNDR
-5.0 to
60.0°C
pH value<0.00 pH. [Replace the test solution or recalibrate the meter.]
OVER
-5.0 to
60.0°C
pH value>14.00 pH. [Replace the test solution or recalibrate the meter.]
Conduc-
tivity
Measure
or
Conduc-
tivity
Calibration
UNDR
UNDR
Temperature< -5.0°C . [Bring solution to a higher temperature.] [Replace electrode.]
OVER
OVER
Temperature> 60.0°C . [Bring solution to a lower temperature.] [Replace electrode.]
Conduc-
tivity
Calibration
CAL
UNDR
-5.0 to
60.0°C
Correction of slope beyond -30%. [Use a new standard solution.] [Replace electrode.]
CAL
OVER
-5.0 to
60.0°C
Correction of slope beyond +30%. [Use a new standard solution.] [Replace electrode.]
Conduc-
tivity
Measure
OVER
-5.0 to
60.0°C
[The conductivity value of the test solution is greater than 20mS/cm.] [Clean or Replace electrode.]
SPECIFICATIONS
pH
Range
Resolution
Accuracy
0.00 to 14.00 pH
0.01 pH
±0.01 pH
Conductivity
Range
Resolution
Accuracy
0uS/cm to 20.00mS/cm
Minimum 1uS/cm
±1%FS
Page 15
14
TDS
Range
Resolution
Accuracy
0mg/L to 20.00g/L
Minimum 1mg/L
±1%FS
Salinity
Range
Resolution
Accuracy
0.0 to 10.0ppt
0.1ppt
±1%FS
Temperature
Range
Resolution
Accuracy
-5.0 to 60.0 °C
0.1 °C
±0.5
23 to 140 °F
0.1 °F
±0.9
pH pH buffer recognition:
pH 7.00, 4.01, 10.01 or pH 6.86, 4.00, 9.18 pH Temperature compensation:
AUTO -5.0 to 60.0 °C
pH Buffer Temperature range: 0.0 to 60.0°C pH Electrode Offset recognition:
±100 mV at pH 7.00; +91.7 mV / -108.7 mV at pH 6.86
pH Electrode Slope recognition:
±30% at pH 4.00, 4.01, 9.18 and 10.01 Input impedance: >10
12
Ω
Conductivity Reference Temperature: 15.0 to 25.0 °C Temperature Coefficient: 0.00% to 4.00% TDS TDS Constant: 0.30 to 1.00
Page 16
15
Temperature
Temperature sensor: Thermistor, 10 kΩat 25°C Temperature unit: °C or °F General
Power: LR44 x 4
Automatic shut off function: 10 minutes of non-use
Data logging capabilities: 50 data sets
Ambient temperature range: 0.0 to 50.0 °C
Relative Humidity: At 95%RH
Case: IP67 water-tight case
Dimensions (W x D x H): 44mm x 25mm x200mm
Weight : 110 g (Batteries included)