PCE Instruments PCE-PH 23, PCE-PH 23-SP Users guide

Page 1
© PCE Instruments
PCE-PH 23 pH Meter
User Manual
Last change: 14 May 2020
v1.0
English
Page 2
© PCE Instruments
Contents
1 Safety notes ........................................................................................... 1
2 Before first use ...................................................................................... 2
2.1 Inserting / replacing the batteries ...................................................................................... 2
2.2 Soaking the electrode ....................................................................................................... 2
3 Device description ................................................................................ 2
3.1 Meter ................................................................................................................................ 2
3.2 Display ............................................................................................................................. 2
3.3 Operating modes of the meter .......................................................................................... 3
3.4 Key description................................................................................................................. 3
4 Operation ............................................................................................... 4
4.1 On / off ............................................................................................................................. 4
4.2 Set date and time ............................................................................................................. 4
4.3 Select buffer solutions ...................................................................................................... 4
4.4 pH value calibration .......................................................................................................... 4
4.5 Measurement ................................................................................................................... 5
4.6 HOLD mode ..................................................................................................................... 5
4.7 Saving readings ............................................................................................................... 5
4.8 Viewing and deleting data ................................................................................................ 5
5 Electrode replacement ......................................................................... 5
6 Error messages ..................................................................................... 6
7 Specifications ........................................................................................ 6
8 Warranty ................................................................................................. 7
9 Disposal ................................................................................................. 7
Page 3
© PCE Instruments
1
1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.
The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
Do not expose the device to shocks or strong vibrations.
The case should only be opened by qualified PCE Instruments personnel.
Never use the instrument when your hands are wet.
You must not make any technical changes to the device.
The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
The device must only be used with accessories from PCE Instruments or equivalent.
Before each use, inspect the case for visible damage. If any damage is visible, do not
use the device.
Do not use the instrument in explosive atmospheres.
The measurement range as stated in the specifications must not be exceeded under
any circumstances.
Non-observance of the safety notes can cause damage to the device and injuries to the user.
We do not assume liability for printing errors or any other mistakes in this manual. We expressly point to our general guarantee terms which can be found in our general terms of
business. If you have any questions please contact PCE Instruments. The contact details can be found at
the end of this manual.
Page 4
© PCE Instruments
2
2 Before first use
2.1 Inserting / replacing the batteries
Open the battery compartment cover and insert new batteries. Observe correct polarity. The unit can continue to operate for approximately 2 to 3 hours after the low battery indicator "BAT" appears on the display.
2.2 Soaking the electrode
The electrode must be soaked before first use or after storage. Remove the cover of the pH meter and soak the electrode in a potassium chloride solution with 3mol/l for 2 hours.
3 Device description
3.1 Meter
1. Battery compartment cover
2. LC display
3. Keypad
4. Electrode holder
5. Electrode and unit for
automatic temperature compensation (ATC)
The electrode cap is not shown.
3.2 Display
1. DATE Month-day-year (MM-DD-YY)
2. HOLD Display for Hold mode
3. MEASURE Display for measuring mode
4. BUF Selected buffer solutions
5. TIME 24 hours-minutes (HH:MM)
6. pH pH reading display
7. °C Temperature reading display
8. Number of calibration point (1P/2P/3P) or type of calibration
Note: °C or °F is pre-set.
1
2
3
4
5
Page 5
© PCE Instruments
3
3.3 Operating modes of the meter
1. Measuring mode in the measuring mode
2. Calibration mode In calibration mode, the 1-point, 2-point or 3-point calibration is performed.
3. Hold mode In Hold mode, the measurement is interrupted and the currently measured value is held on the display.
4. Data retrieval mode In data retrieval mode, you can display measured values from the memory.
5. Date and time setting mode The date and time setting mode is used to correct the time and date.
6. Buffer selection mode for the calibration solutions In this mode, the buffer solution set for the calibration is selected: either
7.00 (pH values 7.00/4.01/10.01) or 6.86 (pH values 6.86/4.00/9.18).
3.4 Key description
Key
Operating mode
Duration of keystroke
Function
HOLD
Any but HOLD mode
2 s or more
Meter on/off
Measuring mode
0 s
Holds current reading
Hold mode
0 s
Back to measuring mode
Date and time setting mode
0 s
Moves cursor to next data entry
CAL
Measuring mode
2 s
Jumps to calibration mode
Measuring mode
10 s or more
Jumps to date and time setting mode
Calibration mode
0 s
Leaves calibration mode
Date and time setting mode
2 s
Leaves date and time setting mode
Data retrieval mode
2 s
Clears memory and returns to measuring mode
STORE
Measuring mode
2 s
Saves current pH reding along with date and time
Date and time setting mode
0 s
Increases current reading
Data retrieval mode
2 s or more
Leaves data retrieval mode
Buffer selection mode
2 s or more
Leaves buffer selection mode
SCROLL
Measuring mode
2 s
Starts data retrieval mode
Measuring mode
5 s or more
Starts buffer selection mode
Data retrieval mode
0 s
Scrolls through saved data
Date and time setting mode
0 s
Decreases current reading
Buffer selection mode
0 s
Selects a new buffer solution set
Page 6
© PCE Instruments
4
4 Operation
4.1 On / off
Press the HOLD key for 2 seconds or longer. The instrument will enter measurement mode. To switch off the meter, press the HOLD key again for 2 seconds. The instrument will turn off automatically when no key has been pressed for 10 minutes.
4.2 Set date and time
1. Press the "CAL" key for 10 seconds to start date and time setting mode
2. Press the "HOLD" key to select the date or time value
3. Press the "STORE" or "SCROLL" key to increase or decrease value
4. Press the "CAL" key for 2 seconds Exit and return to measuring mode
4.3 Select buffer solutions
1. Press "SCROLL" key for 5 seconds to start buffer selection mode
2. Press "SCROLL" key again to select the buffer solution set:
7.00 for (7.00/4.01/10.01) or
6.86 for (6.86/4.00/9.18)
3. Press "STORE" for 2 seconds Exit and return to measuring mode
4.4 pH value calibration
1. Press the "CAL" key for 2...3 seconds to start calibration mode
2. Immerse the electrode in the 7.00 or 6.86 buffer solution. "WAIT" is displayed.
3. When the "WAIT" indicator disappears, the calibration for this value is finished.
4. Press "CAL" to complete the 1-point calibration or
5. repeat steps 2 and 3 with pH 4.01/4.00 or
10.01/9.18 for 2-point calibration
6. Press "CAL" to complete the 2-point calibration or
7. repeat step 5 for 3-point calibration
8. Automatic return to measuring mode
Note: Clean the electrode with distilled water before each change of calibration solution. For accurate measurement results, it is recommended to perform pH calibration weekly and after each change of electrode.
Page 7
© PCE Instruments
5
4.5 Measurement
In measuring mode, immerse the instrument in the liquid to be tested. Ensure that the electrode is immersed to a maximum of 2 mm below the electrode holder and that there are no solids in the water. These could influence the measurement result.
4.6 HOLD mode
1. When the pH reading is stable, press the "HOLD" key to hold it in the display
2. Press "HOLD" key to continue measurement
Note: In HOLD mode, the meter cannot be turned off.
4.7 Saving readings
1. Press the "STORE" key for 2...3 seconds in HOLD mode or measuring mode
2. "DATA SAVE" and "Mxx" indication on the display when data are stored in memory space Mxx
3. The instrument automatically returns to measuring mode.
Note: The instrument can store up to 50 measurements. When all memory spaces are occupied, the first memory space is overwritten first, then the following ones.
4.8 Viewing and deleting data
1. Press the "SCROLL" key for 2...3 seconds in measuring mode to call up the last saved measurement
2. Press the "SCROLL" key again to display the previously saved measurement
3. Press "STORE" for 2 seconds Exit and return to measuring mode or
4. Press "CAL" key for 2...3 seconds Clear memory completely and return to measuring mode
5 Electrode replacement
1. Unscrew electrode holder from the instrument
2. Remove old electrode from holder
3. Place new electrode in the holder
4. Place new electrode on the meter so that the holder fits into the electrode connections
5. Screw electrode holder onto the instrument
6. Soak electrode in 3M KCL for 2 hours
7. Calibrate the meter
Page 8
© PCE Instruments
6
6 Error messages
pH
display
Temperature
display
Mode
Possible causes
Any
“OVER“
Measurement
a) a) Temperature not within the range 0 ...
99.9 °C b) b) No temperature sensor c) c) Temperature sensor defective > Change the temperature of the liquid or > insert new temperature sensor
“OVER“
0 … 99.9°C
Measurement
a) pH value not within the range 0 14 b) Meter must be re-calibrated
“OVER“
0 … 60°C
pH calibration
Offset value or slope value of the electrode outside the specified range > Use new buffer solutions > Change the electrode
Any
“OVER“
pH calibration
Temperature of buffer solution not 0...60 °C > Change temperature
7 Specifications
Measuring range pH measurement
0.00 ... 14.00 pH
Resolution pH measurement
0.01 pH
Measuring accuracy pH measurement
± 0.02 pH ± 1 digit
Buffer detection pH measurement
pH 7.00, 4.01, 10.01 or pH 6.86, 4.00,
9.18
Temperature compensation pH measurement
Automatically 0.0 ... 99.9°C / 32 ... 211.8°F
Buffer detection Temperature measuring range
0.0 ... 60°C / 32 ... 140°F
Offset detection pH electrode
± 90 mV at pH 7.00 ± 98.3 mV / -81.7 mV at pH 6.86
Slope pH electrode
± 30% at pH 4.00, 4.01, 9.18, and 10.01
input resistance
> 1012 Ω
Technical Specification for Digital pH Meter PCE-PH 23: Temperature measurement
Measuring range temperature
0.0 ... 99.9°C / 32 ... 211.8°F
Resolution temperature
0.1°C / 0.18°F
Measuring accuracy temperature
± 0.3°C / 0.5°F
Temperature sensor
Thermistor, 10 kΩ at 25°C
Temperature unit
°C°F, default
General technical specification for the Digital pH Meter PCE-PH 23
Power supply
4 x LR44
Lifespan batteries
200 hours
Operating temperature
0.0 ... 50.0°C / 32 ... 122°F
Display
98 x 64 pixels, graphic LCD
Casing
ABS, IP67
Weight
107 g / < 1 lb
Page 9
© PCE Instruments
7
8 Warranty
You can read our warranty terms in our General Business Terms which you can find here:
https://www.pce-instruments.com/english/terms.
9 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations.
If you have any questions, please contact PCE Instruments.
Page 10
© PCE Instruments
8
PCE Instruments contact information
Germany France Spain
PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L. Im Langel 4 23, rue de Strasbourg Calle Mayor, 53 D-59872 Meschede 67250 Soultz-Sous-Forets 02500 Tobarra (Albacete) Deutschland France España Tel.: +49 (0) 2903 976 99 0 Téléphone: +33 (0) 972 3537 17 Tel. : +34 967 543 548 Fax: +49 (0) 2903 976 99 29 Numéro de fax: +33 (0) 972 3537 18 Fax: +34 967 543 542 info@pce-instruments.com info@pce-france.fr info@pce-iberica.es www.pce-instruments.com/deutsch www.pce-instruments.com/french www.pce-instruments.com/espanol
Germany United Kingdom Italy
PCE Produktions- und PCE Instruments UK Ltd PCE Italia s.r.l. Entwicklungsgesellschaft mbH Unit 11 Southpoint Business Park Via Pesciatina 878 / B-Interno 6 Im Langel 26 Ensign Way, Southampton 55010 Loc. Gragnano D-59872 Meschede Hampshire Capannori (Lucca) Deutschland United Kingdom, SO31 4RF Italia Tel.: +49 (0) 2903 976 99 471 Tel: +44 (0) 2380 98703 0 Telefono: +39 0583 975 114 Fax: +49 (0) 2903 976 99 9971 Fax: +44 (0) 2380 98703 9 Fax: +39 0583 974 824 info@pce-instruments.com info@pce-instruments.co.uk info@pce-italia.it www.pce-instruments.com/deutsch www.pce-instruments.com/english www.pce-instruments.com/italiano
The Netherlands China Hong Kong
PCE Brookhuis B.V. PCE (Beijing) Technology Co., Limited PCE Instruments HK Ltd. Institutenweg 15 1519 Room, 6 Building Unit J, 21/F., COS Centre 7521 PH Enschede Zhong Ang Times Plaza 56 Tsun Yip Street Nederland No. 9 Mentougou Road, Tou Gou District Kwun Tong Telefoon: +31 (0)53 737 01 92 102300 Beijing, China Kowloon, Hong Kong info@pcebenelux.nl Tel: +86 (10) 8893 9660 Tel: +852-301-84912 www.pce-instruments.com/dutch info@pce-instruments.cn jyi@pce-instruments.com www.pce-instruments.cn www.pce-instruments.cn
United States of America Turkey
PCE Americas Inc. PCE Teknik Cihazları Ltd.Şti. 711 Commerce Way suite 8 Halkalı Merkez Mah. Jupiter / Palm Beach Pehlivan Sok. No.6/C 33458 FL 34303 Küçükçekmece - İstanbul USA Türkiye Tel: +1 (561) 320-9162 Tel: 0212 471 11 47 Fax: +1 (561) 320-9176 Faks: 0212 705 53 93 info@pce-americas.com info@pce-cihazlari.com.tr www.pce-instruments.com/us www.pce-instruments.com/turkish