SOIL PH METER
Model : PCE-PH20S
our purchase of this SOIL PH METER marks a step forward for you into the field o precision measurement.
lthough this METER is a complex and delicate instrument, its durable structure will allow many years of use i proper operating techniques a r e developed. Please read the following instructions carefully and always keep this manual within easy reach.
OPERATION MANUAL
Page 2
TABLE OF CONTENTS
.
.
.
1. FEATURES................................................................1
2. SPECIFICATIONS......................................................
3. FRONT PANEL DESCRIPTIONS...................................3
3-1 Display............................................................... 3
3-2 Power button ( button )................................. 3
3-3 Hold button ...................................................... 3
3-4 Display reverse button ( button ).................... 3
3-5 pH electrode BNC socket..................................... 3
3-6 BNC protection cover...........................................3
3-7 Battery compartment/Cover................................ 3
3-8 pH electrode handle........................................... 3
3-9 pH electrode sensing head................................... 3
3-10 pH electrode BNC plug...................................... 3
3-11 pH electrode protection bottle............................
4. pH CALIBRATING PROCEDURE..................................
4-1 Calibrating Consideration.....................................
4-2 Requiring Equipment for Calibration......................4
4-3 Calibration procedures....................................... 4
1
3 4
4
5. MEASURING PROCEDURE..........................................5
6. REPLACEMENT OF BATTERY......................................5
7. OPTIONAL ACCESSORIES..........................................6
Page 3
1. FEATURES
2.SPECIFICATIONS
* Complete set with pen pH meter, soil pH electrode,
pH 4.0 /pH 7.0 standard buffer solution.
* Spear Tip pH electrode ( included ), ideal for
penetration pH measurement into soil, meats, cheese and other type of measurements requiring sample penetration. The electrode features a very durable glass measuring spear packaged in a rugged virtually unbreakable epoxy body.
* Separate soil pH electrode, easy for general purpose
usage and remote measurement.
*
Auto calibration for pH 4, pH 7 or pH 10.
* B uild in reverse display button. * Data hold. * Auto power off. * Compact size, light weight. * Optional carrying case. * Water resistance on the front panel. * Application : Agriculture, Gardening, Food
industrial, Education, School, Colleges, Laboratory, Industrial and Quality control.
2-1 Meter
Display LCD, size : 20 mm x 28 mm.
* User can reverse the display direction
by push button.
Measurement 0 to 14 pH x 0.01 pH Range Resolution 0.01 pH
1
Page 4
Accuracy ± 0.07 pH(pH5 to pH 9)
± 0.1 pH(pH4 to pH4.9, pH9.1 to pH10) ± 0.2 pH(pH1 to pH3.9, pH10.1 to pH13)
* 23± 5 * After calibration.
Input 10^12 ohms. Impedance Temperature Not necessary to make adjustment. Compensation Sampling Approx. 0.8 second. Time Hold Function To freeze the display reading value. Calibration Auto calibration, method pH 7, pH 4 ( or pH 10 ) pH electrode BNC connector. connector Operating 0 to 50 ( 32 to 122 )℃℉ Temperature Operating Less than 80% RH. Humidity Power Supply DC 1.5V battery ( UM-4/AAA ) x 4 PCs. Power Approx. 4.8 mA. Consumption Dimension Meter :180 x 40 x 32 mm ( 7.1 x 1.6 x 1.3 inch ).
Electrode : Body size : 12 mm dia. x 160 mm. Electrode cable length : approx.. 1000 mm.
Weight 220 g/0.48 LB (included electrode).
2
Page 5
Standard Instruction Manual......................... 1 PC.
.
.
Accessories
Soil pH electrode............................ 1 PC.
pH 4.0 buffer solution, PH-04A.......
pH 7.0 buffer solution, PH-07A.......
1 PC. 1 PC.
2-1 Soil pH Electrode ( included )
pH Measuring 1 to 13 pH ( typical 0 to 14 pH ). Range pH Operation 5 to 60 ( 41 to 140 ).℃℉ Temperature pH Electrode Combination type. Structure Reference Ag/AgCl, Double Junction. of pH Electrode Zero Potential 7± 1 pH. for pH Value Repeatability 0.05 pH. Response time 2 minutes. Body Material Epoxy. Connector BNC. Mechanical With protection bottle ( soaker Protection bottle ) on the electrode head. Weight 68 g ( including cable & socket ). Dimensions Body length - 160 mm.
Body Dia. - 12 mm. Cable length - 1000 mm.
3
Page 6
3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display 3-2 Power button ( button ) 3-3 Hold button 3-4 Display reverse button ( 3-5 pH electrode BNC socket 3-6 BNC protection cover 3-7 Battery compartment/Cover 3-8 pH electrode handle 3-9 pH electrode sensing head 3-10 pH electrode BNC plug 3-11 pH electrode protection bottle
4
button )
Page 7
4. pH CALIBRATING PROCEDURE
4-1 Calibrating Consideration
The pH meter already calibrated by mV signal that simulated from the ideal pH ELECTRODE mV output ( base on 25 environment ). However due to (a) An ideal electrode will produce 0 mV at pH 7.00, but most electrodes are slightly off. (b) The measuring environment temperature may not near 25 ( 15 to 35 ). ℃℃ If intend to keep instrument combined electrode within high accuracy, it is necessary to make the following calibration procedures ( 4-3 ).
4-2 Requiring Equipment for Calibration
1)Combination soil pH electrode (included).
2)Buffer solutions :
pH 7.00 buffer solution (included). pH 4.00 buffer solution (included).
4-3 Calibration procedures
pH 7 calibration
1)Connect the pH electrode plug ( 3-10, Fig. 1 ) to the BNC socket ( 3-5, Fig. 1 ) and place electrode into pH 7.00 buffer solution.
2)Power on the instrument by pressing the " Power button " ( 3-2, Fig. 1 ).
3)Press the " Hold button " ( 3-3, Fig. 1 ) once. the display will show the " HOLD " symbol. Press the " show following screen in sequence.
button " ( 3-3, Fig. 1 ) once, the display will
5
Page 8
screen 1
screen 2
flash CAL
screen 3
CAL
CAL flashing
7.00
7.00 flashing
screen 4
End
screen 5 ( return to normal screen )
7.00
6
Page 9
pH 4 calibration
1)Rinse the electrode in distilled water. Place electrode into buffer pH 4.00 solution.
2)All the calibration procedures are same as the above pH calibration, except :
a. Change the buffer solution to pH 4.00. b. The calibration screen value will change to 7.00
instead of 4.00
Remark :
* Except calibrate pH 7, pH 4, meter also can calibrate
pH 10.
* The calibration procedures should start from the pH 7 at
first, then pH 4 ( or pH 10 ).
* During the calibration, the circuit will sensing solution
value automatically. If the buffer solution value already beyond 7± 1 pH. 4± 1 pH, 10± 1 pH, the LCD will show " Err ".
* When the LCD show the screen 2, it allow the user to
adjust another calibration value precisely ( for example
7.01, 6.99... ) by pressing the " button " ( 3-2, Fig. 1 ) , " button " ( 3-4, Fig. 1 )'
* Rinse the electrode in distilled water before change
the buffer solution.
* Repeat above pH, pH calibration procedures two
times at least.
* After calibration, the " Electrode sensing head "
( 3-9, Fig. 1 ) should insert into the " Protection bottle " ( 3-11, Fig. 1 ).
7
Page 10
5. MEASURING PROCEDURE
After the meter along the pH electrode is calibrated ( section 4-3 ) , the unit is ready for measuring.
1)Connect the pH electrode plug ( 3-10, Fig. 1 ) to the BNC socket ( 3-5, Fig. 1 ) and place electrode into the measured solution.
2)Power on the meter by pressing the " Power button " ( 3-2, Fig. 1 ).
3)Place the electrode into the measured soil or other type penetration material ( such as cheese, meats... ) or solution, then the instrument will display the pH value.
* Electrode ( attached ) is the
rough and durable penetration electrode. However user should operate the electrode very carefully.
* The electrode is the consumer
accessory, after the electrode is used, if the damage of the Electrode Sensing Head ( Spear Tip Head ), it is without warranty.
* After make the measurement, please rinse the
electrode in distilled water.
* After calibration, the " Electrode sensing head " ( 3-9,
Fig. 1 ) should insert into the " Protection bottle " ( 3-11, Fig. 1 ).
Data hold
During the measurement, press the " Hold button " ( 3-3, Fig. 1 ) once will hold the measured value & the LCD will display a " HOLD " symbol.
* Press the " Hold Button " once again will release the
data hold function.
8
Page 11
Display reverse
1)When the LCD display show " " ,
During the measurement, press the " Display reverse button " ( 3-4, Fig. 1 ) once will reverse the display direction for the user convenience.
* Press the " Display reverse button " once again will
return to the normal display.
Auto power off
The meter is built the " Auto power shut off " function to save battery life. If any function buttons are not pushed within 10 minutes, the power will be off automatically.
6. REPLACEMENT OF BATTERY
It is necessary to replace the battery. However, in-spec measurement may still be made for several hours after Low Battery Indicator appears before the instruments instrument become inaccurate.
2)
To replace the battery, rotate and remove the " Battery Cover " ( 3-7, Fig. 1 ), take out the old batteries, install new batteries :
DC 1.5V battery ( UM-4/AAA ) x 4 PCs.
3)When install the batteries, should make attention the battery polarity.
4)After install the batteries, reinstall the battery cover again.
9