If √ chosen, press key to enter zero
calibration, or return directly.
When there is nothing on the scale and it’s
stable, press key to start zero
calibration.
The indicator is sampling the A/D data of
zero point, it ends when it became 0% from
100%.
Press key to return without change
Press or key to choose √ or X,
If √ chosen, press key to enter weight
calibration, or return directly.
Zero Calibration
Weight Calibration
Page 14
13
Put the weight on the scale
Input the weight value the same as the
calibration weight
Press key to move the cursor.
Press key to change the digits.
Press key to save the change and
start to calibrate.
Press key to cancel and return.
The indicator is sampling the A/D data of
calibration point, it ends when it became 0%
from 100%.
The weight value is too small.
Trouble shooting:
1、 The weight is really too light.
2、Wrong connection of the signal
cable or excitation cable
Calibration done.
Full Calibration
This calibration is jointed the zero calibration and weight (capacity) calibration, prease refer
to above steps.
Calibration parameters
Page 15
14
This parameter can be retrieved and can
be modified, after tested well, please record
those three parameters for use in future.
When the data in EEPROM get problem,
reload those three parameters, it doesn’t
need to do calibration again.
If the same model indicator to be changed
and those three parameters reloaded, it
doesn’t need to do calibration, either.
Input the weight value the same as the
calibration weight
Press key to move the cursor.
Press key to change the digits.
Press key to save the change and
enter next parameter.
Press key to cancel and return.
Expansion Display
If 10 times resolution display needed, open
the expand display:
Press key to cancel and return.
Press or key to chose √ or X
Press key to confirm and return.
Press key or key to choose the
menu to be set, press key to enter.
Press key to return the previous step.
9.4.2 Operation Set
Page 16
15
Tare Operation
Choices: Disable tare
Enable tare
Press or key to change
Press key to save and return.
Press key to cancel and return.
Choices: 0%, 2%, 20%
Press or key to change
Press key to save and return.
Press key to cancel and return.
Choices: 0%, 2%, 20%
Press or key to change
Press key to save and return.
Press key to cancel and return.
0:None; 1:0.5d;
2: 1d; 3: 2d;
4: 3d; 5: 4d;
6: 5d;
Press or key to change
Press key to save and return.
Press key to cancel and return.
Zero upon Power on
Zero by Keypad
Zero Tracking
Vibration checking
Page 17
16
0:None
1:Vibration checking range = 1d;
2:Vibration checking range = 2d;
3:Vibration checking range = 3d;
4:Vibration checking range = 4d;
5:Vibration checking range = 5d;
Press or key to change
Press key to save and return.
Press key to cancel and return.
Digital filtering
Set range of 1-9, the more, the weight is more
stable and the longer time to be stable.
Normally, if the digital filtering number set to be
higher, the advance volume should be bigger, so
on the batching mode, the advance volume needs
to be adjusted again after the filtering number
changed.
Press or key to change
Press key to save and return.
Press key to cancel and return.
On or Off
Press or key to change
Press key to save and return.
Press key to cancel and return.
Buzzer Set
Page 18
17
9.4.3 Batching Set
Press or key to chose the menu to
be set.
Press key to enter.
Press key to return previous step.
Choices:4 materials batching
Single material weight lose
Free setpoint.
Press or key to change
Press key to save and return.
Press key to cancel and return.
Batching mode:
Material quantity
Page 19
18
If there is no I/O board, it can be set to be 1 only.
If there is the I/O board, it can be set from 1-4.
Press or key to change
Press key to save and return.
Press key to cancel and return.
Feeding door
1:single door single speed
2:double doors double speed
If there is no I/O board, only M1 can be set
Press key to move cursor.
Press key to modify the chosen digits.
Press key to save and return to previous
step.
Press key to cancel and return to previous
step.
This parameter only valid for double doors and
double speed
All open: On fast feeding, open two doors at the
same time.
One open: On fast feeding, open fast feeding
door.
Press or key to change
Press key to save change and return.
Press key to cancel and return.
Choice:Auto or manual
Press or key to change
Press key to save change and return.
Press key to cancel and return.
Feeding mode
Feeding action
Page 20
19
Discharge mode
Choice:Auto or manual
Press or key to change
Press key to save change and return.
Press key to cancel and return.
Choice:0%~9.9%×F.S
Only the weight on the scale is less than this
range and the cursor and output is effective,
otherwise it’s taken as not empty.
Press key to move cursor.
Press key to modify the chosen digits.
Press key to save and return to previous
step.
Press key to cancel and return to previous
step.
Choice:0~9
When it set to be 0, it means no adjust
Press or key to change
Press key to save change and return.
Press key to cancel and return.
Press key or key to choose the
menu to be set, press key to enter.
Press key to return the previous step.
Empty scale range
Frequency Adjust
Time Parameter
Delay start time
Page 21
20
Choice:0.0~9.9 seconds
After batching start, after the delay time, it
can start to tare, open the door and start to feed.
Press key to move cursor.
Press key to modify the chosen digits.
Press key to save and return to previous
step.
Press key to cancel and return to previous
step.
Choice:0.0~9.9 seconds
The comparing time of fast and fine
feeding both to be set in this menu.
Press key to move cursor.
Press key to modify the chosen digits.
Press key to save and return to
previous step.
Press key to cancel and return to
previous step.
Press key or key to choose
Choice:0.0~9.9 seconds, the stabling time
after batching finish of material 1
Comparing time forbidden
Delay checking time
Page 22
21
Choice:0.0~9.9 seconds.
When it reach to the range of empty
scale, it will close the door after this time, to
make sure full discharge.
Press key to move cursor.
Press key to modify the chosen digits.
Press key to save and return to
previous step.
Press key to cancel and return to
previous step.
the menu to be set, press key
to enter.
Press key to return the previous
step.
Choice:0.0~9.9 seconds, the stabling time
after batching finish of material 2
Choice:0.0~9.9 seconds, the stabling time
after batching finish of material 3
Choice:0.0~9.9 seconds, the stabling time
after batching finish of material 4
Delay time of discharge
Tolerance checking frequency
Page 23
22
Choice:0~99
Checking tolerance after the times of
batching, if it was set of 0, it will not check.
Press or key to change
Press key to save change and return.
Press key to cancel and return.
Tolerance range set
Choice:0%~9.9%×Target weight
When it checking, if value of the actual
weight minus the target weight is less than
this value, it will be taken as ok, if not, it’s
out of tolerance range.
Press key to move cursor.
Press key to modify the chosen
digits.
Press key to save and return to
previous step.
Press key to cancel and return to
previous step.
×:The first material doesn't need tare
√:The first material need tare
Press or key to change
Press key to save change and return.
Press key to cancel and return.
It will stop after this cycle times.
Press or key to change
Press key to save change and return.
Press key to cancel and return.
First material tare
Cycle times of batching
Adjust range
Page 24
23
Choice:0%~9.9%×Target weight
Only the value of actual weight minus target
weight is less this range, it goes to adjust, when
it’s set to be 0, this function off.
Press key to move cursor.
Press key to modify the chosen digits.
Press key to save and return to previous
step.
Press key to cancel and return to previous
step.
9.4.4 Communication protocol
Press key or key to choose the
menu to be set, press key to enter.
Press key to return the previous step.
No output
Continuous output format A
Command output format A
Continuous output format B
Command output format B
MODBUS RTU output
7 digits even parity check
7 digits odd parity check
8 digits even parity check
8 digits odd parity check
Press or key to change
Press key to save change and return.
Press key to cancel and return.
Choice: No transmitting
Transmitting
Press or key to change
Press key to save change and return.
Press key to cancel and return.
Choice: 1—64
Keep the uniqueness of address in the same
network
Press key to move cursor.
Press key to modify the chosen digits.
Press key to save and return to previous
step.
Checksum
Indicator address
Page 26
25
Press key to cancel and return to
previous step.
9.4.5 Analog quantity output
Press key or key to choose the
menu to be set, press key to enter.
Press key to return the previous step.
Choice:0~+5V; 0~+10V;
-5V~+5V; -10V~+10V;
4mA~20mA; 0mA~20mA;
0mA~24mA;
Press or key to change
Press key to save change and return.
Press key to cancel and return.
Press key or key to
choose the menu to be set, press
key to enter.
Press key to return the
previous step.
Output type choose
Zero calibration (coarse-small-fine)
Page 27
26
Press key or key to
choose the menu to be set, press
key to enter.
Press key to return the
previous step.
Press key or key to choose the
menu to be set, press key to enter.
Press key to return the previous step.
Full scale calibration (coarse-small-fine)
9.4.6 Diagnosis and maintenance
Page 28
27
Reload
Choose √ to reload the default parameters
Reloading…
Choose √ to enter
Menus
Press key or key to choose
the menu to be set, press key to
enter.
Press key to return the previous step.
Input the full capacity of load cells
Press key to move cursor.
Press key to modify the chosen
digits.
Press key to save and return to
previous step.
Press key to cancel and return to
previous step.
Free calibration
Page 29
28
Input the sensitivity of load cells
Press or key to change
Press key to save change and return.
Press key to cancel and return.
Zero calibration
Adjust the present weight manually
Press key to move cursor.
Press key to modify the chosen
digits.
Press key to save and return to
previous step.
Press key to cancel and return to
previous step.
Input interface test
Choose √ to enter test menu
Page 30
29
Input test interface
Output interface test
Choose √ to enter test menu
Output test interface
Choose √ to backup parameters
Input the password
Backup parameters
Page 31
30
Password wrong
Backup
Retrieve the backup parameters
Choose √ to retrieve the backup parameters
Indicates that there is no backup parameters
Retrieving…
Choices:
Tolerance output, fast feeding output, fine
feeding output, discharging output.
If the original output interface broken or
with problem, you can mapping that
interface to this alternate one by changing
the cables.
Press or key to change
Alternate output mapping
Page 32
31
Press key to save change and return.
Press key to cancel and return.
9.5 Recipes Parameters set
Basic + 4 materials type
Press key or key to choose the
menu to be set, press key to enter.
Press key to return the previous step.
IO + 4 materials type
Press key or key to choose the
menu to be set, press key to enter.
Press key to return the previous step.
Basic + 1 material weight losing
Press key or key to choose the
menu to be set, press key to enter.
Press key to return the previous step.
Basic + setpoints
Press key or key to choose the
menu to be set, press key to enter.
Press key to return the previous step.
Page 33
32
M1 recipe set
Press key or key to choose
the menu to be set, press key to
enter.
Press key to return the previous step.
Target weight:
The final weight filled in the
tank/container, etc.
Press key to move cursor.
Press key to modify the chosen
digits.
Press key to save and return to
previous step.
Press key to cancel and return to
previous step.
Page 34
33
Fine weight:
The relative weight between TARGET and
FAST (when it finish the fast filling and
change to fine filling), it’s the relative value
but not actual value.
Press key to move cursor.
Press key to modify the chosen
digits.
Press key to save and return to
previous step.
Press key to cancel and return to
previous step.
PREACT: Decide the weight of end of the fine
feeding (TARGET - PREACT).
Press key to move cursor.
Press key to modify the chosen
digits.
Press key to save and return to
previous step.
Press key to cancel and return to
previous step.
M2、M3、M4 recipes the same as M1
One number can preset of 3 recipes
If a new recipe, you need to set the recipe
number, if it’s the recipe already in the
indicator, you can modify it directly.
Press or key to change
Press key to save change and return.
Press key to cancel and return.
Recipe numbers set
Upper and lower weight limitations
Page 35
34
It needed to be set when it using for losing
weight filling
Lower limitation: the minimum remaining
after the end of discharging.
Upper limitation: the approximated
maximum weight it can reach.
During feeding, if the weight <(target + lower
limitation), the indicator will open the
supplement feeding valve until the weight
reach to upper limitation and then start to
discharge.
Note: If the target weight < (Upper limitation
– Lower limitation), every start will end after
once time discharge.
If the target weight > (Upper limitation –
Lower limitation), it will discharge more
times until it reach to the target value.
Press key to move cursor.
Press key to modify the chosen
digits.
Press key to save and return to
previous step.
Press key to cancel and return to
previous step.
Press key to move cursor.
Press key to modify the chosen
digits.
Press key to save and return to
previous step.
Press key to cancel and return to
previous step.
Setpoint 1
Setpoint 2 and setpoint 3 the same as above set.
Page 36
35
Press key to move cursor.
Press key to modify the chosen
digits.
Press key to save and return to
previous step.
Press key to cancel and return to
previous step.
Press key or key to choose
the menu to be set, press key to
enter.
Press key to return the previous step.
Press key or key to choose
the menu to be set, press key to
enter.
Press key to return the previous step.
Language set
Press key or key to choose
the menu to be set, press key to
enter.
Press key to return the previous step.
Preaction 1 set
Preaction 2 and preaction 3 the same set as above
9.6 Printing table format
9.7 System set
Page 37
36
10. Appendix: Communication Protocol
Screensaver time set
If it’s set to be 0, the function off
Press key or key to choose
the menu to be set, press key to
enter.
Press key to return the previous step.
Screen contrast adjust, the bigger number,
the more bright the screen.
Press or key to change
Press key to save change and return.
Press key to cancel and return.
Data format(18 bytes)
STX A B C X X X X X X X X X X X X CR
CKS
I
Ⅱ
Ⅲ
Ⅳ
Ⅴ
Ⅵ
BIT
Status A
Status B
Status C
Bit0
Decimal point position:
001:XXXXX0
010:XXXXXX
011:XXXXX.X
100:XXXX.XX
101:XXX.XXX
Ⅰ、<STX>ASCII start sign (02H)。
Ⅱ、status A、B、C,refer to below table
Ⅲ、Weight value, gross weight or net weight, 6 ASCII digits without sign and decimal point.
Ⅳ、Tare value,6 ASCII digits without sign and decimal point.
Ⅴ、<CR>ASCII Enter sign (ODH)。
Ⅵ、<CKS> checksum
Page 38
37
Bit5
Always 1
Always 1
Bit6
Always 0
Always 0
Bit7
Always 0
Always 0
10.2 Command transmitting format A
No.
Command sign
Note
1
“C” or “c”
Clear tare
2
“P” or “p”
Print gross tare net
3
“T” or “t”
Tare
4
“Z” or “z”
Zero
No.
Command letter
Indicator response
1
READ+ENTER NEWLINE
Refer to <Indicator weight reading and
status>
2
TARE+ ENTER NEWLINE
Refer to <Tare weight reading and status>
3
TARE ON+ ENTER
NEWLINE
Tare operation
Works: return YES+ ENTER NEWLINE
No work:return NO?+ ENTER NEWLINE
4
TARE OFF+ ENTER
NEWLINE
Clear tare
Works:return YES+ ENTER NEWLINE
No work:return NO?+ ENTER NEWLINE
5
ZERO ON+ ENTER
NEWLINE
Zero operation
Works:return YES+ ENTER NEWLINE
No work:return NO?+ ENTER NEWLINE
6
CAL XXXXXX+ ENTER
NEWLINE
XXXXXX is loading weight value, W, without
decimal point
If W=0,zero calibration.
If W≠0,loading weight calibration.
Indicator response:
Zero calibration done:
ZERO YES+ ENTER NEWLINE
Zero calibration vibrate:
ZERO E1+ ENTER NEWLINE
Loading weight calibration done
SPAN YES+ ENTER NEWLINE
Loading weight calibration vibrate:
10.3 Continuous transmit format B
10.4 Command output format B
Page 39
38
SPAN E1+ ENTER NEWLINE
Input data is too small:
SPAN E2+ ENTER NEWLINE
Input data is too big:
SPAN E3+ ENTER NEWLINE
Loading weight is too big:
SPAN E4+ ENTER NEWLINE
Indicator weight reading and status, the indicator response format:
Address
Digits
Note (read only)
40001
Current gross weight value
40002
Current display weight value
40003
.0
1=material 1 in fast feeding
.1
1=material 1 in fine feeding
.2
1=scale is empty
.3
1=reach to setpoint 1 (setpoint type)
.4
1=material 2 in fast feeding
.5
1=material 2 in fine feeding
.6
1=reach to setpoint 2 (setpoint type)
.7
1=reach to setpoint 3 (setpoint type)
.8
1=material 3 in fast feeding
.9
1=material 3 in fine feeding
.10
1=supplement feeding (single material losing weight mode)
.11
1=Indicator above tolerance or lower than tolerance
.12
1=material 4 in fast feeding
Tare weight reading and status, the indicator response:
10.5 MODBUS output format
MODBUS is the master-slave type network communication protocol, this indicator acts as the
subordinate in the network for called by the host system, the format is RTU and support 03 and 06
functions. The parameter needs to be set as F4.1=5 and the address of Modbus to set in F4.5, the mapping
address definitions as the below table:
Page 40
39
.13
1=material 4 in fine feeding
40004
.0
01~11:current recipe numbers(1~3)
.1
.2
.3
.4
1=batching run
.5
1=batching suspense
.6
1=discharging
.7
1=feeding finish, waiting for discharging
.8
0000=0.001;0011=0.01;0110=0.1;1001=1;
0001=0.002;0100=0.02;0111=0.2;1010=2;
0010=0.005;0101=0.05;1000=0.5;1011=5;
1100=10;1101=20;1110=50;
current division.
.9
.10
.11
.12
1=batching tolerance
.13
1=scale vibrate
.14
1=manual feeding mode
.15
1=manual discharging mode
40005
1~4 material batching mode: actual batched weight of current material 1.
Single material losing weight mode: the current batched value of this filling.
(the batching value kept till start of next cycle).
40006
1~4 material batching mode: actual batched weight of current material 2.
Single material losing weight mode: the current accumulated value of this
filling.
(the batching value kept till start of next cycle).
40007
actual batched weight of current material 3. (the batching value kept till start
of next cycle).
40008
actual batched weight of current material 4. (the batching value kept till start
of next cycle).
Address
Mapping(read and write)
40009
Tare value
40010
Target value of material 1
40011
Target value of material 2
40012
Target value of material 3
40013
Target value of material 4
40014
Fast feeding value of material 1
40015
Fast feeding value of material 2
40016
Fast feeding value of material 3
40017
Fast feeding value of material 4
40018
Preaction quantity of material 1
40019
Preaction quantity of material 2
40020
Preaction quantity of material 3
40021
Preaction quantity of material 4
40022
Range of empty scale
40023
Preaction adjust frequency
40024
Start delay time
40025
Discharge delay time
40026
Forbid comparing time
40027
Delay check time of material 1
40028
Delay check time of material 2
40029
Delay check time of material 3
40030
Delay check time of material 4
40031
Lower tank weight
40032
Upper tank weight
40033
Weight value of setpoint 1
40034
Weight value of setpoint 2
40035
Weight value of setpoint 3
40036
Preaction quantity of setpoint 1
Page 41
40
40037
Preaction quantity of setpoint 2
40038
Preaction quantity of setpoint 3
40039
Quantity used of material 1 (higher digits)
40040
Quantity used of material 1 (lower digits)
40041
Quantity used of material 2 (higher digits)
40042
Quantity used of material 2 (lower digits)
40043
Quantity used of material 3 (higher digits)
40044
Quantity used of material 3 (lower digits)
40045
Quantity used of material 4 (higher digits)
40046
Quantity used of material 4 (lower digits)
40047
0
1=Zero calibration done
1
1=Loading weight calibration done
2
1=Writing value less than 1% during loading weight calibration
3
1=Writing value more than full scale during loading weight calibration
4
1=Loading weight is not enough during loading weight calibration
5
1=running, calibration not allowed
40048
Floating number of current weight
40049
Address 位Mapping(Write only)
40101
.0
0001~0100:appoint the material number to be operated
.1
.2
.3
.4
Start manual feeding, material bits 0-3 given:
(Eg: If 0001: material 1 batching,
If 0010: material 2 batching,
If 0011: material 3 batching,
If 0100: material 4 batching)
.5 / .6
/
.7 / .8
Start auto feeding
.9
Suspend batching or discharging
.10
Stop emergency
.11
Start discharge(after feeding end and discharging type is manual)
.12
Tare(not batching, not vibrate and tare allowed)
.13
Clear tare(not batching, not vibrate and tare allowed)
.14
Zero(not batching, not vibrate)
.15
Continue batching or discharging
40102
.0
0001~0011:Choose working recipe number
.1
.2
.3
.4
10:Chose manual batching during all cycles.
11:Chose auto batching during all cycles.
.5
.6
10:Chose manual discharging during all cycles.
11:Chose auto discharging during all cycles.
.7
40103
Scale calibration:
0:Zero calibration
XXXXX:loading weight calibration(XXXXX is the value of the loading weight)
Page 42
41
10.6 Multi materials jointed work with relay connected
The below chart is for connecting of 4 materials batching, the input and output can be connected
as common source or common grounded, for single material, connect the three relay to the
connecting terminals of fast feeding, fine feeding and discharging.