PCE Instruments PCE-MO 2010-ICA, PCE-MO 2010 Users guide

Page 1
10A MICRO-OHM METER
PCE-MO 2010
INSTRUCTION MANUAL
Page 2
INDEX
1. INTRODUCTION.............................
2. SAFETY NOTES.............................
3. FEATURES.....................................
4. SPECIFICATIONS..........................
5. GENERAL.......................................
6. INSTRUMENT LAYOUT..................
7. MEASUREMENT.............................
8. MAINTENANCE..............................
9.
INTERFACE
CONNECTION
AND
OPERATION.............................
10. Contact..........................................
11. Disposal.........................................
PAGE
1 2 3 4 5
6-7
8-12
13
14-22
23 23
Page 3
1. INTRODUCTION
NOTE
This meter has been designed and tested According to CE Safety Requirements for Electronic Measuring Apparatus, IEC / EN 61010-1 and other safety standards. Follow all warnings to ensure safe operation.
WARNING
READ "SAFETY NOTES" (NEXT PAGE) BEFORE USING THE METER.
CAT IV
-
Is for measurements performed at the source of the low-voltage installation.
Is for measurements performed in the building Installation.
Is for measurements performed on circuits directly connected to the low-voltage installation.
CAT III
-
CAT II
-
-1-
Page 4
2. SAFETY NOTES
Read the following safety information carefully Before attempting to operate or service the meter.
Use the meter only as specified in this manual. Otherwise, the protection provided by the meter may be impaired.
Rated environmental conditions : (1) Indoor Use. (2) Installation Category III. (2) Pollution Degree 2. (3) Altitude up to 2000 meters. (4) Relative humidity 80% max. (5) Ambient temperature 0~40°C.
Observe the International Electrical Symbols listed below :
Meter is protected throughout by double insulation or reinforced insulation.
Warning ! Risk of electric shock.
Caution ! Refer to this manual before using the meter.
-2-
Page 5
3. FEATURES
Maximum testing current of 10A
Ranges:
200.0μΩ / 2000μΩ
20.00mΩ / 200.0mΩ
2.000Ω / 20.00Ω
200.0Ω
10A
1A
100mA
10mA
Maximum resolution of 0.1μΩ at 10A
Four terminal measurement
4 Lines × 20 Characters LCD
Optical USB to RS-232 data transmission
Well isolated from contact
2 built-in optical LEDs for data transfer
Memory function (200 measurement results)
Auto power off
Low battery indication
Power source:
Rechargeable battery DC14.8V 5200mAh 76.96Wh
Safety Standard: IEC/EN 61010-1 CAT III 300V EN 61326-1
-3-
Page 6
4. SPECIFICATIONS
4-wire method measurement
-4-
Measuring ranges / resolution
200.0μΩ / 0.1μΩ 2000μΩ / 1μΩ
20.00mΩ / 0.01mΩ
200.0mΩ / 0.1mΩ
2.000Ω / 1mΩ
20.00Ω / 0.01Ω
200.0Ω / 0.1Ω
Accuracy
200.0μΩ ~ 2000μΩ: ±(2% + 8dgt)
20.00mΩ ~ 200.0Ω: ±(1% + 4dgt)
Testing current
10A: 200.0μΩ / 2000μΩ 1A: 20.00mΩ / 200.0mΩ 100mA: 2.000Ω / 20.00Ω 10mA: 200.0Ω
Page 7
5. GENERAL
Operating temperature: 0~40°C.
Operating humidity: 80% RH.
Dimensions:
322(L) × 260(W) × 135(D)mm
Weight: Approx. 3137g (battery included)
Power source: Rechargeable battery DC14.8V 5200mAh 79.69Wh
Accessories: Instruction manual Test leads Data transmission cable Software for PC interface Shoulder belt Charger
Safety Standard: IEC/EN 61010-1 CAT III 300V
EN 61326-1
-5-
Page 8
6. INSTRUMENT LAYOUT
⑵ ⑶
⑿ ⒀ ⒁
⒃ ⒄
C1
P1
200.0µ
20.00m
2.000
200.0
P2
2000µ
20.00
C2
ESC ENTER ON
TEST
> 1s SEL
SAVE
OFF STOP
CAT III
300V
PCE-MO
10A MICRO-OHM METER
Optical Adapter for Data Transmission
Charge
⑻ ⑼ ⑹ ⑺ ⑽
-6- (1) 4 Lines × 20 characters LCD
Page 9
(2) 10A Button (3) 1A Button (4) 100mA Button (5) 10mA Button (6) ▲ Button (7) ▼ Button (8) ESC / SEL Button (9) ENTER / SAVE Button
(10) ON / OFF Button
(11) TEST / STOP Button (12) C1 Terminal (13) P1 Terminal (14) P2 Terminal (15) C2 Terminal (16) Charge LED (17) Charge Socket
(18) Optical Adapter for
Data Transmission
Test Leads
-7-
Page 10
7. MEASUREMENT
1. Power On :
Press "ON/OFF" button to turn the power on or off. Hold the "ON/OFF" button that can display the voltage of battery.
The system will be auto off when it has no operation after 3 minutes.
2.Before testing, you must make sure that the resistance under test does not have any voltage. When the voltage exceeds approx. 5 volts or the probe is connected to a power supply of more than approx. 5 volts,a warning message will be displayed.
3.Select test current & scale range : There are 4 buttons "10mA", "100mA", "1A" and "10A" which can choose the operation current and the scale range. "10mA" button for 200.0 Ω only. "100mA" button can choose 20.00 Ω & 2.000 Ω.
"1A" button can choose 200.0 mΩ & 20.00 mΩ. "10A" button can choose 2000μΩ & 200.0μΩ.
The screen will display date, time, test current and maximum resistance.
-8-
Page 11
4.The measurement function Press "TEST/STOP" button to measure the resistance value. It needs a few seconds to complete the measurement, so press "TEST/STOP" button again that can interrupt the measurement.
4-wire method
R
X
Charge
C
1 P1 CAT III 300V P2 C2
When the measurement is finished, the screen will display the resistance value and testing current.
-9-
Page 12
5.Record saving function When the measurement is finished, press "SAVE/ENTER" button to save the result of measurement. At the same time, the last line will display the number of records. The memory can record up to 200 records.
6.Minor function Press "ESC/SEL" button and hold over 1 second to enter the minor function.
Use the "UP" and "DOWN" button to move the function selection cursor, and press "SAVE/ENTER" to enter the minor function, and press "ESC/SEL" button to return back to the main page.
Date/Time ADJ
-10-
Page 13
Press "ESC/SEL" button to leave the Date/Time ADJ. Press "UP" or "DOWN" button to set the correct number. Press "TEST/STOP" button to switch to the next unit of date or time and go on the required adjustment. After all adjustments are completed, press "SAVE/ ENTER" button to confirm and save the date and time.
Press "ESC/SEL" button to return back to the minor function menu.
RECORD DISPLAY
Press "UP" or "DOWN" button to reach the record that you want to look up. Press "ESC/SEL" button to return back to the minor function menu.
-11-
Page 14
RECORD CLEAR
Press "SAVE/ENTER" button to clear all records in the memory. Press "ESC/SEL" button to return back to the minor function menu.
REMEMBER TO CHARGE THE BATTERY!
ALWAYS recharge the battery when the Low Battery info appears on the display. Leaving the battery uncharged for a long period of time may affect the lifespan of the battery.
Charging interval:
Every 3 months
-12-
Page 15
8. MAINTENANCE
Charge:
When low battery warning information "LoBAT" appears, please charge the battery.
1. Plug one end of charger into the charge socket
and the other end into the ACV power socket.
2. If ACV plug socket is energized, the charge
LED is lit on indicating that charge is underway. If the ACV plug socket isn’t energized, remove to another power-energized one making power charge.
3. After the voltage reaches approx. 16.0V~16.5V,
charge process is complete. (It can be observed on LCD display.)
No measurement can be performed during the charging process.
-13-
Page 16
9. INTERFACE CONNECTION AND OPERATION
(A) Microohm Meter Setup Installation:
1. Install the "Micoohm Setup" on the personal computer.
2. Click the "Next" button.
3. To install into a different folder, click Browse and Select another folder. If it's not necessary, click the "Next" button.
-14-
Page 17
4. Click the "Next" button.
5. All the files are installing to your personal computer.
-15-
Page 18
6. The Microohm Meter program has been successfully installed and then click "Finish" button.
7. The software driver file: "CDM 2.08.24 WHQL Certified x64-64bit" and "CDM 2.08.24 WHQL Certified x86-32bit" files.
(Note: New Windows system will install the
software driver automatically. It's necessary to install the software driver only if it is the old windows system)
-16-
Page 19
(B) Windows Comm Port setting:
1. Plug CA-232 data transmission cable into the personal computer USB port.
2. On the windows system, right click the "This PC" icon to find manager.
①Right click "This PC"
3. Click the "Device Manager" to find ports (COM & LPT).
-17-
Ports(COM & LPT)
②Click "Device Manger"
Page 20
4. Click the Ports (COM & LPT) to find USB Serial Port (COM6).
(Note: USB Serial Port will indicate
automatically different "COM" port)
-18-
③Click Ports(COM &LPT) USB Serial Port(COM6)
Page 21
(C) Microohm Meter Software Comm Ports
Setting:
1. Connect CA-232 data transmission cable to the Insulation Tester.
2. Click the "Microohm Meter" icon to the personal computer.
3. On the Microohm Meter windows, select the correct "Comm Port" and click the "connect button".
-19-
①Comm Ports Selection
②Click connect button
Page 22
4. The Comm Port selection is workable when the red signal turns green signal.
5. Click the "Download Log from Device" button
can show the data.
①Click the Download
Log from Device
②Click Save Log
-20-
Green Signal
Page 23
6. Click the "Save Log" button to save the data records to the personal computer.
(D)
Microohm Meter Windows Interface Introduction:
RS232 Connection
Main
Recording
-21-
Page 24
Cleaning and Storage :
Periodically wipe the case with a damp cloth and detergent. Do not use abrasives or solvents.
Due to our policy of constant improvement and development, we reserve the right to change specifications without notice.
-22-
WARNING
To avoid electrical shock or damage to the
instrument, do not get water inside the case.
Page 25
10 Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact us. You will find the relevant contact information at the end of this user manual.
11 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations.
If you have any questions, please contact PCE Instruments.
RoH
-23-
Pb
Pb
Cd
Hg
www.pce-instruments.com
Page 26
PCE Instruments contact information
Germany
PCE Deutschland GmbH Im Langel 26 D-59872 Meschede Deutschland Tel.: +49 (0) 2903 976 99 0 Fax: +49 (0) 2903 976 99 29 info@pce-instruments.com www.pce-instruments.com/deutsch
France
PCE Instruments France EURL 23, rue de Strasbourg 67250 Soultz -Sous -Forets France Téléphone: +33 (0) 972 3537 17 Numéro de fax: +33 (0) 972 3537 18 info@pce -france.fr www.pce -instruments.com/french
Spain
PCE Ibérica S.L. Calle Mula, 8 02500 Tobarra (Albacete) España Tel. : +34 967 543 548 Fax: +34 967 543 542 info@pce-iberica.es www.pce-instruments.com/espanol
United Kingdom
PCE Instruments UK Ltd Trafford House Chester Rd, Old Trafford
Manchester M32 0RS United Kingdom Tel: +44 (0) 161 464902 0
Fax: +44 (0) 161 464902 9 info@pce-instruments.co.uk www.pce-instruments.com/english
Italy
PCE Italia s.r.l. Via Pesciatina 878 / B -Interno 6 55010 Loc. Gragnano Capannori (Lucca) Italia Telefono: +39 0583 975 114 Fax: +39 0583 974 824 info@pce -italia.it www.pce -instruments.com/italiano
Turkey
PCE Teknik Cihazları Ltd.Şti. Halkalı Merkez Mah.
Pehlivan Sok. No.6/C 34303 Küçükçekmece - İstanbul Türkiye Tel: 0212 471 11 47 Faks: 0212 705 53 93 info@pce-cihazlari.com.tr www.pce-instruments.com/turkish
The Netherlands
PCE Brookhuis B.V. Institutenweg 15 7521 PH Enschede Nederland Telefoon: +31 (0)53 737 01 92 info@pcebenelux.nl www.pce-instruments.com/dutch
United States of America Denmark
PCE Americas Inc. 1201 Jupiter Park Drive, Suite 8 Jupiter / Palm Beach 33458 FL USA Tel: +1 (561) 320-9162 Fax: +1 (561) 320-9176 info@pce-americas.com www.pce-instruments.com/us
PCE Instruments Denmark ApS Birk Centerpark 40 7400 Herning Denmark Tel.: +45 70 30 53 08 kontakt@pceinstruments.com www.pce-instruments.com/dansk
© PCE Instruments