Thanks for you’re selecting Personal Noise Dose Meter. To ensure that you can get
the most from it we recommend that you read the manual carefully before use.
The personal noise dose meter is test equipment to evaluate the noise condition of
working environment by measuring the accumulated noise exposure. With this test
data, company can ensure the related noise regulation is followed or decide whether
or not to apply certain noise reduction action to avoid the acoustic trauma and
noise-induced hearing loss occurs to the working personnel.
Protected by:
Taiwan: D174456
U.S.A.:
China:ZL 2015 3 0269895.4
D768,514
2.PRINCIPLE OF MEASURING ACCUMULATED NOISE EXPOSURE
DOSE is a parameter used to quantify noise exposure measured in a period, during
turns of % noise exposure relative to 90dBA for 8 hours.
i.e. 100% DOSE = 90dBA for 8 hours
This is known as the Criterion. Other criterion are available such as 100% DOSE =
85dBA for 8 hours to meet legislation in different countries.
If the person stayed in this environment for 4 hours, he would receive 50%DOSE.
Also if the noise level was say 93dBA, 3dB higher, a 100% DOSE would be
accumulated in only 4 hours, the above example is for a dose meter with 3dB
Exchange Rate, where a 3dB increase in sound level corresponds to a doubling
energy. A dose meter with 5dB Exchange Rate is also, available where the sound
must increase by 5dB to double the dose as required to meet noise legislation in
some countries.
3.INTRODUCTION TO THE DOSE METER
The Noise Dose Meter measure the frequency weighted noise exposures and peak
sound levels simultaneously.
• Keypad lock protects meter against accidental operation
• Timers support pre–selection of measurement time and duration
• USB PC interface for the charging unit
The meter is ideal for quick on–site surveys and for monitoring personal noise
exposure in accordance with European ISO or USA OSHA, MSHA, DOD AND
ACGIH standards.
The meter is one measurement ranges: 70 – 140dB, A C & Z RMS weightings, A, C
& Z Peak weightings, Fast, Slow & Impulse time weightings, and energy exchange
rates (Q) of 3, 4, 5 & 6. In dose measurement mode, it display % dose, % dose
projected for an 8 hour period, peak level and measurement duration.
For countries that are subject to European Union regulations or the equivalent ISO
Aeq,8hr
standards, the meter measure the daily sound exposure level (L
according to IEC 61252, which is identical with the L
required by ISO 1999),
EX,8h
, Pa2h
while for USA OSHA/MSHA regulations they measure the Time Weighted Average
level (TWA).
When used as a sound level meter it display sound level, time-averaged (L
level, peak and sound exposure level (SEL).
-2-
) sound
Aeq
Page 5
4.SPECIFICATION
Standard applied: IEC 61252: 2002, ANSI S1.25-1991 for dose meter and sound
Time Weightings: Fast, Slow and Impulse (RMS detector)
Exchange Rate: 3, 4, 5 or 6dB
Stabilization Time: 10 seconds from power on
Threshold and Criterion Levels:
From the setup, predefined in the built- in setup according to applicable standards.
User–defined setups in the following ranges:
Threshold Level: 70 – 90dB in 1dB steps
Criterion Level: 70 – 90dB in 1dB steps
Overload Indications: At 0.1dB above the top of the selected measurement
range.
Under-Range Indications: At 1dB below the bottom of the selected measurement
range.
High Level Detector: 115dB ( )
Clock: Real-time clock with calendar.
Memory: Results from 30 measurements and 6101 data sets can be stored for
later viewing and download.
Measurement Control:
Measurement Duration: May be set 10 to 55 min in 5min steps or 1 to 24 Hrs in
1hr steps.
Timers: Up to 99 timers (automatic start and stop) can be set (maximum up to
one month ahead of measurement time).
Timer Controlled Started / Stop: Up to 16 timers can be set up from the PC
software or meter keypad.
-3-
Page 6
Setup: Ten built–in (predefined) default setups are included.
NR
Nine additional user–defined setups can be stored.
Setup
Measurement Range (dB)
Time Weighting
Frequency Weighting
Peak Frequency Weighting
Exchange Rate
Threshold (dB)
Criterion Level (dB)
Allow User to Change Setup
Exceedance Time LAS
DOSE (Dose), PDOSE (Pdos), TWA (Twa), PTWA (PTwa), LAVG (Lavg), LEPd
(Lepd), PLEPd (PLepd), Exceedance Time (Las>105 or Las >115), LEQ (Laeq,
Lceq or Lzeq), SEL (La
Max/Min 18 parameters for A, C, Z and F, S, I weightings.
Statistical analysis 45 parameters L05, L10, L50, L90, L95 for A, C, Z and F, S, I
weightings for sampled at 20ms interval into 0.1dB wide classes.
Statistical analysis 5 parameters LAEQ05, LAEQ10, LAEQ50, LAEQ90, LAEQ95 for
LAEQ sampled at 1s intervals into 0.1dB wide classes.
Display parameters selecting by PC.
User1
–9
Output: USB interface from the charging unit (The charging unit must be power on).
Battery: 3.7V 470mAH Lithium Polymer (Flat cell).
Battery Life time: Typically > 15 hours at room temperature.
The meter will automatically stop the measurement and store the data before
the battery voltage gets too low.
Battery Indicator: Symbol indication battery capacity.
“” indicated flash when voltage is insufficient for operation.
Operating Temperature & Humidity: 0 – 50ºC (32 to 122ºF), 10 – 90%RH
Storage Temperature & Humidity: -10 – 60ºC (14 to 140ºF), 10 – 75%RH
Weight: Approx. 77g (2.7oz), less mounting device.
Accessories: Instruction manual, Charging unit, AC adaptor, Mounting clips,
Carrying case, Software CD, USB cable.
-4-
Page 7
CE Certification:
ty implies that sound level indications of 70dB or
CE-mark indicates compliance EMC Directive
EMC Emission
EMC Immunity
EN50081-1 (1992): Generic emission standard.
Part 1: Residential, commercial and light industry
EN50081-2 (1993): Generic emission standard.
Part 2: Industrial environment
CISPR22 (1993): Radio disturbance characteristics of
information technology equipment. Class B Limits
FCC Rules, Part 15: Complies with the Limits for a Class B
digital device
EN50082-1 (1992): Generic immunity standard.
Part 1: Residential, commercial and light industry
RF immuni
greater will be affected by no more than ±1.5dB
EN 50082-2 (1995): Generic immunity standard.
Part 2: Industrial environment RF immunity implies that sound
level indications of 70dB or greater will be affected by no more
than ±1.5dB
-5-
Page 8
5.NOMENCLATURE AND FUNCTIONS
5-1 The Display and keypad
1. Windscreen and microphone
2. LED indicator
Red: Charging indicator.
Blue: A flashing blue LED indicates the measured value has exceeded the alarm
setting value.
3. Display: View data and various indicators.
4.
key:
On/Off key: Press this key one time to turn on the meter.
Press and hold down this key, A “Power Off 3, 2, 1” countdown will
appear as the meter is powered off.
Enter key: Confirms action of other keys and settings.
-6-
Page 9
5. ►/ key:
Windscreen
► Run key: In STOP mode, Press and hold down this key, A “START 3, 2, 1”
countdown will appear as the meter is running.
Stop key: In Run mode, Press and hold down this key, A “STOP 3, 2, 1”
countdown will appear as the meter is stop, if the meter PAUSE mode
is setting to “OFF” otherwise A “PAUSE 3, 2, 1” countdown will appear
as the meter is pause.
6. Up key: Press this key to scroll through display views or setting up the parameters.
7.
Down key
parameters.
5-2 The Windscreen and Microphone
During use, it is essential that the meter is fitted with the Windscreen. The
Windscreen protects the microphone from potential erroneous results due to air
movement passing over it but also helps to protect from dust ingress, moisture or
minor impact damage.
To calibrate the meter it is necessary to remove the Windscreen to expose the
microphone. To unscrew the Windscreen by turning the black ring (not the foam!)
in a anticlockwise direction.
It will not normally be necessary to remove the Microphone but to do so unscrew
anticlockwise direction. When refitting the Microphone, screw the Microphone on
clockwise until it fits tightly.
Following calibration refit the Windscreen, screw the Windscreen in a clockwise
and take care do not over tighten; finger tight is sufficient.
: Press this key to scroll through display views or setting down the
Microphone
(Anticlockwise)
-7-
Page 10
5-3 The Mounting Clips
Upon delivery, the meter will be fitted with the Crocodile-style mounting clips on
flexible webbing to aid positioning as they can be fitted to virtually any clothing
type. The mounting clip is used for attaching the meter to clothing and should be
used where piercing the employees clothing is not desired.
The recommended location for mounting a personal noise exposure meter in a
position close to the ear (10 – 15 cm). The meter should therefore be mounted on
the apex of the shoulder to avoid reflections from the head which might otherwise
affect the measurements. Consult your local Legislation and Standards for more
information.
All mounting clips are attached the meter by two screws. By removing the screws
with the screwdriver the mounting clips can be changed.
-8-
Page 11
5-4 The Charger
The charger is a drop-in intelligent charger unit, required to download the meter
memory and/or to charger the meter internal Lithium Polymer battery. The charger
consists of the charger base, AC adaptor, link cable and the charger link plate.
Note that the meter should only be charged in non-hazardous areas in the
absence of any flammable atmospheres.
Link cable
Charger
Charger link plate
-9-
Page 12
6.CHARGING THE METER
The LCD display
when the meter is on
The LCD display
6-1 Charging
The meter use internal Lithium polymer battery. Ensure the meter is fully charged
prior to use by placing in the charger. Note the meter will fit into charger unit
regardless of which mounting clips are attached.
Ensure the AC adapter is connected and the supply is switched on. The meter will
automatically switch on and display how much charge is within the meter. After
approximately 20 seconds the display will blank. Pressing either the , or
key will wake the display for a further 20 seconds to check progress. When the
meter is charging the red LED will on. The meter will display “FULL” once
charging is complete and the LED on the front of the meter will turn purple.
and put on the
battery charger.
when the meter is
off and put on the
battery charger.
-10-
Page 13
6-2 Linking Chargers Together
The charger can be linked to the another charger via the link cable and the
charger link plate. Up to 4 chargers can be linked.
6-3 Battery
The meter battery care best practices.
The meter incorporates a Lithium Polymer battery as a power source that with
proper care should give years of operational service. (Three years is typical)
Proper care includes these best practices.
●After use, it is best to charge the meter using the recommended charger prior
to storage for a prolonged period. A prolonged period is considered to be 2 or
more months. If the meter is to be stored for an even longer period, it is best
to take the meter out of storage and re-charge the meter every 2 – 3 months.
●Charge the meter in the recommended charger before attempting to turn it on
if the meter has not been used for a prolonged period or was stored in a low
battery condition.
●It is best not to store the meter for any period of time in low battery condition.
●If not using the meter it is best to keep the meter in a powered up the
recommended charger.
-11-
Page 14
7.GENERAL OPERATION
7-1 Switching on the Meter
Once the meter is switched on by pressing the
sequence, shown in as follows. This displays the meter firmware version (e.g. Rev
V1.0) and the current time and date, followed by the meter series number and the
“Custom Text” screen. The series number and text can be configured using the
CD software by PC.
(
1sec)
(1sec)
, it will display a start-up
key
I
f the Auto scroll mode has been disabled pressing the or key will scroll
through the available results from the last run depending upon how the meter was
configured.
-12-
Page 15
(k
ey)
( key)
( key)
( key)
( key)
( key)
( key)
( key)
( key)
( key)
( key)
( key)
( key)
( key)
-13-
Page 16
7-1-1 Error Messages at Power Up
If there are any error messages they will appear after the start-up screens:
Battery < 1 Hr Memory
Warning messages appear when there is less than 1 hours charge or memory
full. Press any key to return to the default screen or alternatively these error
messages will time-out after 30 seconds. Please either re-charge, download
and /or erase data as required before continuing with a new measurement run.
7-2 Calibrating the Meter
It is best practice (and often mandatory) to field calibrate each meter both before
and after use. Which will also allow the target calibration level to be set as shown
on the calibrator’s own calibration certificate, nominally 94.0dB, 114.0dB or OFF
selecting by PC.
Note: The meter will not enter the calibration mode if a measurement run is taking
place. If a run is in progress, stop the run according to section 7-5.
The windscreen should be removed prior to calibration. Please refer to section 5-2.
Push the sound calibrator over the microphone as shown. The calibrator should be
pushed on without twisting until fully home and switched on.
z
H
K
1
F
B
F
B
d
d
O
4
4
9
1
1
L
E
V
R
E
O
L
-
T
D
A
N
R
U
B
I
O
L
S
A
C
-14-
Page 17
The meter will automatically recognize a 1kHz calibration tone and enter the
key to confirm you wish to calibrate,
key to cancel the calibration. It will
take a few seconds to automatically calibrate
114.0dB during which time a
Once the meter has successfully calibrated
The meter is ready to take a measurement
once the calibrator has been removed and
calibration mode.
Press
↵
or press
to 94.0dB or
progress bar is displayed as shown.
the “CAL OK” message will appear.
(key)
the Windscreen re-fitted.
You may need to consult local regulations or
standards if the “before” and “after”
calibration values differ by more then
(typically) +/- 0.5dB e.g. section 12.2 of ISO
9612: 2009 suggest that measurement is
re-taken.
-15-
Page 18
7-3 Starting a Measurement Run
Before starting a measurement run, ensure the Windscreen is re-fitted according
to section 5-2 and ensure there is sufficient battery life and memory by looking at
the default screen shown in section 7-1.
If necessary, charge the meter as described in section 6-1, if the memory is fully
either clear it as described in section 8-10 or download the data to PC.
Note: If the battery available is less than one hour or memory is fully, the run will
not start.
Timers are also available to automatically start and stop a measurement run or
halt a run after a fixed period. These options may be selected in the Configuration
menu, see section 8-9
To manually start a measurement run, press and hold down key which will
initiate a 3 second countdown as shown. The keys must remain depressed
during the countdown for the measurement run to begin; releasing during the
3-2-1 countdown will cease the operation.
Once the measurement has begun, the “” mark is flashing displayed in the top
left of the screen. The display will cycle between the section 7-1 screens if the
Autoscroll mode has been selected in the configuration menu (see section 8-8) or
may be scrolled manually by pressing either the or key.
7-4 Locking and Unlocking the Keys
It is recommended that the meter is locked during a run to prevent the wearer
tampering and this can be down automatically upon the start of a run having first
selected “Auto Lock” in the configuration menu (see section 8-7) or manually as
follows.
Press key 3 times to enter the “Lock” mode, the “” mark is displayed.
To unlock the meter, press key 3 times, the “” mark is disappear.
-16-
Page 19
7-5 Stopping a Measurement Run
If the meter is locked, unlock as described in section 7-4 above. To stop the
current measurement run, press and hold down the key for the period of the 3
second countdown (releasing during the countdown will cease the operation).
The display will then show the data from the measurement run that had just been
completed. See section 7-1.
A new measurement run can be started immediately by following section 7-3.
7-6 Pausing a Measurement Run
Some measurement strategies may require a measurement run to be PAUSED
e.g. during a worker’s lunch break. To enable this function it must first be selected
in the configuration menu (see section 8-5).
When a measurement run is in progress, the action of pressing and hold down the
key for the period of the 3 second countdown to enter the pause mode
(releasing during the countdown will cease the operation).
Note: During PAUSED mode, all time history profile data is recorded and “paused
event” markers are saved into the data set. Cumulative values shown on
the display represent the measured data and excluded the time and noise
level during the paused intervals.
From the PAUSED screen:
– to continue the run, press the key as prompted by the “” mark.
– to stop the run, press the key as prompted by the “” mark and “STOP
Sure?” confirmation message will appear for 30 seconds then default back to
the PAUSE screen if no action is taken.
– To stop the run, press thekey as prompted by the “” mark.
– to return to the paused mode press the key.
-17-
Page 20
7-7 Reviewing Measurement Run Data
As soon as a measurement run has been stopped, the meter will automatically
display result from last measurement run. The display will cycle between the
screens as section 7-1 if the Auto Scroll mode has been selected in the
configuration menu (see section 8-8) or may be scrolled manually by pressing
either the or key.
7-8 Switching Off the Meter
To switch off the meter, press and hold down the key for the duration of the 3
second 3-2-1 countdown (releasing during the countdown will cease the
operation).
8.CONFIGURING THE METER
The configuration menu is accessed by press the key to turn on the meter,
then press key to enter the configuration screen.
Press key to enter the setting mode and press key to exit.
The setting menu is used to:
LCD Contrast: Set the display’s contrast
RTC: Adjust the date and time
LCD APO: Set the display timeout
Dosimeter: Select the dosimeter setup
Pause: Select/deselect the Pause function
(key)
Any changes made will become the default configuration.
LED Alarm: Select/deselect visual alarms
Auto Lock: Select/deselect Auto Lock of the keys
Auto Scroll: Select/deselect Auto Scroll of the
display
Timers: Select/deselect Timers
Interval: Set the data memory interval time
Language: Select language
Clear Memory: Clear the memory
-18-
Page 21
8-1 LCD Contrast
Set the display contrast by press key 2 times then press , keys to choose
from 1 to 10 brightness levels, and press key to save, press key to exit.
(key)(key)
8-2 Date and Time
Adjust the date and time by press key to select “
to make adjustments. Press
left then press , keys to adjust. Press
key to move the cursor to right or press key to
( key)
(key)
RTC
” then press
key to save, press
(key)
key to exit.
key 2 times
8-3 Display Timeout ()
The display can be left permanently on (which will reduce battery life) or will
timeout after the selected interval of 5, 10, 20 seconds, 1 or 5 minutes.
Set the display time out by press key to select “LCD APO” then press key
2 times to make select. Press , keys to select, press key to save and
press key to exit.
( key)
(key)
key)
(
-19-
Page 22
8-4 Selecting Dosimeter Setups
The meter has up to 3 simultaneous dosimeters, Dosimeter-1 Dosimeter-2 and
Dosimeter-3 which only determine what is displayed on screen and not, but all is
calculated and saved to memory for later download and analysis. The
Dosimeter-1 cannot be disabled. Additional Dosimeter-2 and Dosimeter-3 may be
selected to display any of the other available configurations.
The meter can be configured to display any one of ten built-in (predefined) default
setups (OSHA-80, OSHA-90, MSHA-80, MSHA-90, DOD, ACGIH, ISO-85,
ISO-90, NR-15 and NHO-01) or any one of nine full custom configurable from
USER-1 to USER-9.
The Dose dates and Noise parameters available for display on screen and is
achieved using the meter CD software by PC.
8-4-1 Pre-difined Dosimeter Setups
( key)
(key)
(key)
Pre-difined the Dosimeter setup by press key to select “Dosimeter” then
press key 2 times to select the default setups. Press , keys to
select the desired setups from “OSHA-80” to “NHO-01” then press key
to save and press key 2 times to exit.
8-4-2 Customised Dosimeter Configurations
Any of the general noise parameters listed below can be configured.
Reference setup: Select from OSHA-80 to NHO-01
Time weighted: Select Fast, Slow, or Impulse
Threshold: (THnn) Adjustable 70 to 90dB in 1dB steps
Criterion: (Cnn) Adjustable 70 to 90dB in 1dB steps
Exchange Rate: (Qn) Adjustable 3 to 6dB in 1dB steps
Exceedance level (LAS): Adjustable 70 to 140dB in 5dB steps
Projected Work duration: Adjustable 10 to 55 min in 5 min steps or
1 to 24 Hrs in 1 hr steps
-20-
Page 23
( key)
(key)
(key)
In the pre-difined setup mode, press , keys to select the desired
customized configuration from “USER-1” to “USER-9” then press
to confirm. Press , keys to review the parameters, if you will change the
parameters press key to confirm then press , keys to change the
desired parameters value and press key again to save.
When all parameter configuration is finished, press key to exit.
key)
(
key
-21-
Page 24
8-5 Pause Mode ()
Enable the pause mode so that you may temporarily stop a measurement run, for
example, during the worker’s break. During a Pause, all data is recorded.
Cumulative values will represent the measured data and will exclude the paused
intervals.
( key)
(key)
( key)
Select the Pause function by press key to select “Pause” then press key
to review. Press key to accept or press key then press key to change
and press key to confirm.
8-6 LED Alarms ()
LED Alarm criteria can be set by using the meter CD software by PC.
( key)
(key)
( key)
Select the LED alarm by press key to select “LED Alarm” then press key
to review. Press
and press key to confirm.
key to accept or press key then press key to change
-22-
Page 25
8-7 Auto Lock ( )
Selecting Auto Lock “lock” the display in run mode.
( key)
(key)
( key)
S
elect the Auto Lock by press key to select “Auto Lock” then press
to review. Press key to accept or press key then press key to change
and press key to confirm.
8-8 Autoscroll ()
If selected the display will automatically scroll from one screen to the next
(otherwise the screen has to be advanced manually).
key
( key)
(↵ key)
( key)
Select the Autoscroll by press key to select “Auto Scroll” then press key
to review. Press
and press key to confirm.
key to accept or press key then press key to change
-23-
Page 26
8-9 Timers ( )
To select a timer run.
8-9-1 No-Cycle Timer Run
The no cycle timer means the Start-Time and Duration-Time is valid only one
time.
( key)
(key)
(key)
( key)
(key)
(key)
(key)
(key)
-24-
Page 27
To select no-cycle timer run by press key to select “Timers” then press
key to enter this mode. Press key to select “Delayed” then press key
to review the Start time, press the key to review the Duration time. If you will
change the Duration time and the Start time, using key to position the
cursor on the date or time element to adjust then using , keys to setting
the selected date or time element value including the “Enable” or “Disable”
selected. Press key to confirm then press key to exit.
8-9-2 Auto-Cycle Timer Run
The auto-cycle timer run means the start-time and the duration-time is valid
at the repeat setting number from 1 to 99 times.
(key)
( key)
(key)
( key)
In the Timers mode, press key to select “Repetitive” then press key
to review the Start-time and Duration-time, press key to review the Repeat
times. If you will change the data, using
time element to adjust then using , keys to setting the selected time
element value including the Repeat number from 01 to 99 and the “Enable”
or “Disable” selected. Press key to confirm then press key to exit.
(key)
(key)
key to position the cursor to the
-25-
Page 28
8-10 Data Memory Interval Time
( key)
(key)
(key)
(key)
(key)
The LAVG, LEQ, LCPK and LZPK time-history data is saved at an data
memory interval time for every run and are available following download.
Data memory interval time can be set by press key to select “Interval” then
press key to review. Press key to accept or press key then press
key to select 1, 5, 10 or 30 minute, 1 or 2 hours and press key to confirm.
-26-
Page 29
8-11 Language
( key)
(key)
(key)
(key)
(key)
Set your preferred language English, Portugues or Espanol.
Select language by press key to select “Language” then press key to
review, press key to accept or press key then press key to select
and press key to confirm.
8-12 Clear Memory
( key)
(key)
(key)
In the setting mode, press key to select “Clear Memory” then press key to
enter this mode. Press key to select “YES” then press
screen “Clear Memory Sure ?” is displayed to check that you wish to
progress since once cleared all stored runs will be lost.
Press key to confirm then press key to exit.
key to the additional
-27-
Page 30
9.RECALL THE MEMORIZED DATA
( key)
Press the key to turn on the meter, then press
configuration screen. Press key to select “
select the desired memorized data to be recall. Press key to confirm and display
the memorized data. Press , key or autoscroll display the memorized data.
Press key to the next memorized data to be recall.
Press key two times to exit.
Memory
key again to enter the
” then press , keys to
10. SOFTWARE INSTALLATION AND OPERATION
For the detailed instruction, please refer to the content of attached CD-ROM,
which has the complete instruction of software operation and relevant information.
-28-
Page 31
PCE Instruments contact information
Germany
PCE Deutschland GmbH
Im Langel 26
D-59872 Meschede
Deutschland
Tel.: +49 (0) 2903 976 99 0
Fax: +49 (0) 2903 976 99 29
info@pce-instruments.com
www.pce-instruments.com/deutsch
United Kingdom
PCE Instruments UK Ltd
Trafford House
Chester Rd, Old Trafford
Manchester M32 0RS