PCE Instruments PCE-IR 1300, PCE-IR 1800, PCE-IR 1600 User Manual

PCE Americas Inc. 711 Commerce Way Suite 8 Jupiter FL-33458 USA From outside US: +1 Tel: (561) 320-9162 Fax: (561) 320-9176 info@pce-americas.com
www.pce-instruments.com/english
www.pce-instruments.com
PCE Instruments UK Ltd.
Units 12/13
Southpoint Business Park
Ensign way
Hampshire / Southampton
United Kingdom, SO31 4RF
From outside UK: +44
Tel: (0) 2380 98703 0
Fax: (0) 2380 98703 9
info@pce-instruments.com
Content
1. Introduction
1.1 User interface
1.2 Display
1.3 External function introduction
2. Operation
2.1 Starting the measurement
2.2 Hold-function
2.3 Switch off
2.4 Setup the laser
2.5 Setup the display backlight
2.6 Setup the parameters
2.7 Setup emissivity
2.8 Setup the Min/Max function
2.9 Setup high and low alarm
2.10 Setup the temperature unit
2.11 Setup offset
3. Data Logger
3.1 Storing data
3.2 Datalogger recall
3.3 Clear the memory
4. Software
4.1 System requirements
4.2 Installation and connection
4.3 Download the data
4.4 Software interface
5. Specifications
5.1 Optical specifications
5.2 General specifications
Appendix: Emissivity
1
1. Introduction
Welcome to select PCE-IR 1000 Series portable infrared thermometer. Each single product passed through a quality process. Please read the operators
manual carefully before you start to use the unit.
1.1 User interface
Mode-Button
The button in middle that marked with a circle is the mode button. In hold mode, each time you press it, you will enter to another function setup, e.g. press one time to enter to emissivity setup, press it again will save emissivity and enter to MIN/MAX setup and so on.
Up/Laser
With this button you can setup the laser or increase the chosen value or read MAX/MIN result in HOLD mode.
Down/Display backlight
With this button to you can setup the backlight or reduce the chosen value.
1.2 Display
A. LCD backlight B. MAX/MIN C. Laser D. Alarm activation E. Temperature Unit F. Infrared temperature G. HOLD H. Emissivity
MODE but t on
UP but t on DOWN but t on
2
1.3 External function introduction
1) Trigger
2) Battery compartment cover
3) USB interface
4) Telescope
4
1
2
3
3
2. Operation
2.1 Starting the measurement
Turn on: Press the trigger to turn on the unit. Measuring: Aim the thermometer at a target. Pull the trigger and keep it pressed.
The display will show the current temperature value.
2.2 Hold-Function
If you release the trigger, the unit will enter to HOLD mode. The display shows
HOLD.
2.3 Switch off
The unit automatically switches off after 7 seconds if no button is pressed..
2.4 Setup the laser
Pull the trigger(keep it pressed) and then press up button ( ) ∧ to activate or
deactivate the laser. The laser symbol in the display(only if the trigger is pulled)
indicates the active laser.
2.5 Setup the display backlight
Pull the Trigger(keep it pressed) and then press the down button to
activate/deactivate the display backlight. The symbol in the display will indicate that.
WARNING: Do not point the laser directly at the eyes of persons or animals! Do not stare into the laser beam. Avoid indirect exposure via reflective surfaces.
4
2.6 Setup the parameters
You can setup the emissivity, max/min display, high/low alarm, temperature unit,
temperature offset as follows:
Release the trigger to enter to the Hold mode. Each time you press the mode button will enter to another function interface. The currently selected mode will show on the display by indicator flashes.
Now, you can choose or modify the requested value by pressing the up and down buttons. Press mode button to save the value and enter to next function. If you have not activated any button for 7 seconds, the unit will not save the current modification value and switches down automatically.
After enter the selecting mode, the following flashing signals will show on the display in turn.
2.7 Setup emissivity
In HOLD mode, press the mode button () to set the emissivity.
1) Emissivity(E= ) flashing
2) Press up button ( ) to increase the emissivity
3) Press down button () to reduce the emissivity
5
2.8 Setup the Min/Max function
In HOLD mode, press the mode button repeatedly until enter into the Min/Max
mode. Press the up/down button to setup which mode you want.
MIN indicates the minimum value of current measurement; MAX indicates the
maximum temperature value of current measurement.
2.9 Setup high and low alarm
1) High alarm
In Hold mode, press the mode button ) until enter the high alarm mode. Press
the up/down button to set the alarm on or off.
When the signal of “)” shows on the display, it means the high alarm is
activated.
After set high alarm, you can press the mode button to set the temperature alarm value.
2) Low alarm
In Hold mode, press the mode button utill enter to the low alarm mode.
Press the up/down button to set the alarm on or off.
When the signal “)”shows on the display, it means the low alarm is activated.
After set low temperature alarm, you can press the mode button to set the temperature alarm value.
6
2.10 Setup the temperature unit
In Hold mode, press the mode buttonuntil enter to the temperature unit
selecting mode, and press the up/down button to select the requested temperature unit.
Centigrade scale Fahrenheit scale
2.11 Setup offset
In Hold-mode, press the mode button repeatedly until enter into the temperature offset mode. Then, press the up/down button to setup the temperature offset value.
7
3. Data logger
PCE-IR 1000 series thermometer has the internal data
logger with a maximum capacity of 2000
measurement protocols.
3.1 Storing data
In measurement mode, release the trigger to enter to HOLD mode, and press the down button, there will be a line of small numbers; an icon showing a disk and an icon of “MAX” appear above the current temperature result.
The small numbers indicate the memory position. You can choose the memory position from 1~2000 with the up/down buttons. Press the mode button can save the max temperature of this measur ement to the current position.
3.2 Datalogger recall
In measuring mode, keep the trigger pressed and then press the mode button to
recall the saved data.
Users can choose the memory position by press the up/down buttons.
8
3.3 Clear the memory
In the Storing Data mode, choose the memory position to “0” by pressing the down button. Press the mode button () will clear the memory. The three buzzer signals confirms the successful clearing.
9
4. Software
PCE-IR 1000 series portable infrared thermometer has
special software and connecting cable.
Users can set the unit, download the loggered
data and recording temperature curve
and through the software.
4.1 System Requirements
Vista, XP or Windows 2000 operation system USB interfacet Hard disk with at least 30MByte free space At least 128MByte RAM CD-ROM Driver
4.2 Installation and connection
Please insert the CD to CD-ROM driv er. Please start the SETUP.EXE on the
CD-ROM, and install the software follow the instructions of the wizard.
USB interface driver installation
1) Insert USB Cable into computer’s USB port. Computer will nd the new hardware, and the following dialogue box appears:
2) Choose the option of “NO, not for the time being (T)”, click “next” and the following dialogue box appear:
3) Choose “install from the list or the designate place (advanced) (S)” click “next”, the following dialogue box appear.
4) Click “browse” to nd the USB le in CD.
5) Click “conrm”, the following dialogue box appear:
6) Click “next”, sometimes the system will come up the followin
g the prompt,
then click “still continue”.
01
7) After completing the driver installation, it will new hardware again, repeat the above operations until system prompt “New hardware has been installed successfully, you can use it now”.
8) Right click “My computer” to see “attributes”, choose
hardware device
manager
port as follows:
We can see there is a new port named “Infrared Thermometer Adapter” with
the port number behind. By now, the USB Cable Driver installation is completed.
9) After the successful start of the software, the communication will be displayed in the status bar of the windows. Choose the correct COM-Port via: [MENU: Setup\ interface]. There will be a signal of [PCE-IR 1000 connected] show on the left bottom of the windows after connect successfully.
4.3 Download the data
To download the loggered data from the infrared thermometer, please press the
download menu: [measure
download data], and all the loggered data will be
indicated on the screen.
4.4 Software interface
1) Digital display
After PCE-IR 1000 portable infrared thermometer connected to your personal computer
and the software is started successfully, the target temperature will show on
the left in digital form.
11
2) Starting the measurement
Please press the measuring key: [MeasureÆ begin]
3) Scaling of the temperature axis Global scaling: temperature range of the diagram is automatically adapted to
the respective peak values.
Local scaling: temperature range of the diagram will be dynamically adapted
to the respective peak values.
Manual scaling: can be done at any time using the control elements of the
temperature axis.
4) Stop the measurement
To stop the current measurement, please press the stop key [Menu: measurement/ stop]. The save key [Menu: File\save as] opens an explorer window to select destination and file name.
5) Diagram setting
The menu item settings [Menu: diagram\setting] enable the settings for data protection.
Color: temperature graph and digital display. Initial time: time frame on x-axis at the beginning of measurement
21
5. Specifications
5.1 Optical Specifications
PCE-IR 1300 D:S=120:1
PCE-IR 1800 / PCE-IR 1600 D:S=300:1
31
5.2 General specifications
Model
PCE-IR 1300
Temperature range 0~1300
System accuracy
±2 (0~200 )
±1% (200~1300 )
Repeatability
±1 (0~200 )
±0.5%(200~1300 )
Optical resolution
120:1
Response time
300ms (95%)
Emissivity 0.100~1.000adjustable
Aim mode
Laser telescope aim
Signal processing
MAX/MIN Scan/Hold Function
Alarm functions Audible a nd visible HIGH/LOW alarm
Data storage
2000 point
Backlight
Three color
Outputs/digital
USB cable and software
Storage temperature
-20~60 no battery
Ambient temperature
0~50 no green laser
Relative humidity
10~95% non condensing
Power
Ni-MH rechargeable battery
Size
264x203.5x60mm
61
General specifications
Model
PCE-IR 1600
Temperature range 385~1600
System accuracy 0.3% 2 (at ambient temperature 23 5 )
Repeatability 0.1% 1 (at ambient temperature 23 5 )
Optical resolution 300:1
Response time 100ms (95%)
Emissivity
0.100~1.000 adjustable
Aim mode
Laser telescope aim
Signal processing
MAX/MIN Scan/Hold Function
Alarm functions
Audible a nd visible HIGH/LOW alarm
Data storage
2000 point
Backlight Three color
Outputs/digital
USB cable and software
Storage temperature
-20~60 no battery
Ambient temperature 0~50
Relative humidity
10~95% non condensing
Power
Ni-MH rechargeable battery
Size 264x203.5x60mm
51
General specifications
Model
PCE-IR 1800
Temperature range 650~1800
System accuracy 0.3% 2 (at ambient temperature 23 5 ) Repeatability
0.1% 1 (at ambient temperature 23 5°C )
Optical resolution
300:1
Response time
100ms (95%)
Emissivity
0.100~1.000 adjustable
Aim mode
Laser telescope aim
Signal processing MAX/MIN Scan/Hold Function
Alarm functions
Audible a nd visible HIGH/LOW alarm
Data storage
2000 point
Backlight Three color
Outputs/digital
USB cable and software
Storage temperature
-20~60 no battery
Ambient temperature
0~50
Relative humidity
10~95% non condensing
Power
Ni-MH rechargeable battery
Size
264x203.5x60mm
17
Appendix: Emissivity
Emissivity: the radiant intensity launched from the object depends on the temperature of the target and the object surface material’s radiation characteristic.
Emissivity table
Material Emissivity
8-14µm Aluminum oxidized 0.2-0.4 Asbestos 0.95 Asphalt 0.95 Basalt 0.7 Ceramic 0.95 Concrete 0.95 Copper oxidized 0.4-0.8 Fabric 0.95 Glass plate 0.85 Gold 0.01-0.1 Gravel 0.95 Ice 0.98 Iron oxidized 0.5-0.9 Karborundum 0.9 Lead Oxidized 0.2-0.6 Paper Each color 0.95 Plastics Transparent>0.5mm 0.95 Rubber 0.95 Sand 0.9 Snow 0.9 Soil 0.9-0.98 Steel oxidized 0.7-0.9 Water 0.93 wood Natural 0.9-0.95
Loading...