PCE Instruments PCE-HT 224E Users guide

Page 1
© PCE Instruments
PCE-HT 224E Concrete Test Hammer
User Manual
Last change: 19 December 2023
v1.0
English
Page 2
© PCE Instruments
Contents
1 Safety notes ........................................................................................... 1
2 Specifications ........................................................................................ 2
2.1 Technical specifications ................................................................................................... 2
2.2 Delivery scope.................................................................................................................. 2
3 System description ............................................................................... 3
3.1 Device description ............................................................................................................ 3
3.2 Description of the operating keys ..................................................................................... 4
3.3 Technical data .................................................................................................................. 4
4 Measurement parameters .................................................................... 5
4.1 Concrete strength calculation ........................................................................................... 5
4.2 Concrete strength value in the standard setting ................................................................ 5
5 Operation ............................................................................................... 6
5.1 Start a simple test ............................................................................................................ 6
5.2 Test series in the measurement menu with default settings .............................................. 6
5.3 Calibration ........................................................................................................................ 8
5.4 Start new measurement ................................................................................................... 8
5.5 Enter part name ............................................................................................................... 8
5.6 Enter the number of measurements in the test zone ......................................................... 9
5.7 Test zone entry ................................................................................................................ 9
5.8 Repeat test of the current test zone .................................................................................11
5.9 „Basic" measurement settings .........................................................................................11
5.10 Review measurement data ..............................................................................................12
5.11 Delete measurement data ...............................................................................................13
6 Data transfer ........................................................................................ 13
6.1 Set up data transfer .........................................................................................................13
6.2 Establishing a connection to the PC ................................................................................14
6.3 Driver installation .............................................................................................................14
6.4 Starting and exiting the PC software ................................................................................15
7 Main menu ........................................................................................... 16
7.1 Setting measuring units ...................................................................................................16
Page 3
© PCE Instruments
7.2 Setting the upper / lower limit values ...............................................................................16
7.3 Rights management, password settings ..........................................................................17
7.4 Energy saving settings ....................................................................................................17
7.5 Date setting .....................................................................................................................18
7.6 Setting the counter ..........................................................................................................18
7.7 Battery status display ................................................................ ......................................18
7.8 Instrument data (Info of Instrument).................................................................................18
8 Calibration ........................................................................................... 19
8.1 Completion of a test series ..............................................................................................19
9 Maintenance and servicing ................................................................ 20
10 PC software ......................................................................................... 20
11 Appendix 1 - Troubleshooting ........................................................... 21
12 Appendix 2 - Conversion table for concrete strength ..................... 24
13 Contact ................................................................................................. 25
14 Disposal ............................................................................................... 25
Page 4
© PCE Instruments
1
1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.
The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
Do not expose the device to shocks or strong vibrations.
The case should only be opened by qualified PCE Instruments personnel.
Never use the instrument when your hands are wet.
You must not make any technical changes to the device.
The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
The device must only be used with accessories from PCE Instruments or equivalent.
Before each use, inspect the case for visible damage. If any damage is visible, do not
use the device.
Do not use the instrument in explosive atmospheres.
The measurement range as stated in the specifications must not be exceeded under
any circumstances.
Non-observance of the safety notes can cause damage to the device and injuries to the user.
We do not assume liability for printing errors or any other mistakes in this manual. We expressly point to our general guarantee terms which can be found in our general terms of
business.
Page 5
© PCE Instruments
2
2 Specifications
2.1 Technical specifications
Measurement range
10 ...100 N/mm²
Accuracy of the rebound value R
±0.1 R Display
OLED
Nominal kinetic energy
2.207 J
Compliance
Technical regulations for rebound measurement (JGJ/T23-2001)
Interface
Mini USB
Memory
1000 data units with 256 measurements each
Operating conditions
-4 ... 40 °C
Power supply
3.7 V Li-Ion battery
Power supply (mains adaptor)
Primary: 100 ... 240 V AC, 50/60 Hz Secondary: 5 V DC, 1 A
Dimensions
280 x 75 x 60 mm / 11 x 3 x 2.4 in
Weight
approx. 1.2 kg / 2.6 lbs
2.2 Delivery scope
1 x PCE-HT 224E material tester 1 x mini-USB cable 1 x emery stone in storage box 1 x charger 1 x transport case 1 x user manual
The software can be downloaded free of charge from our download area:
https://www.pce-instruments.com/english/download-win_4.htm
Page 6
© PCE Instruments
3
3 System description
3.1 Device description
1. Retaining nut
2. Release screw
3. Retaining claw
4. Retaining pin
5. Guide plate
6. Locking button
7. Housing body
8. Housing
9. Guide rod
10. Impact weight
11. Guide bush
12. Two-piece retaining ring
13. Felt ring
14. Impact body
15. Retaining nut
16. Return spring
17. Impact spring
18. Housing retaining ring
19. Housing, top
20. Signal generator
21. Sensor element
22. Signal transmitter
23. LCD
24. Measuring strips
25. Rechargeable battery
26. Guide rail
27. Control keys
28. Locking spring
29. Pressure spring
30. Rear cover
Page 7
© PCE Instruments
4
3.2 Description of the operating keys
3.3 Technical data
Impact energy
2.207 J ±0.100 J
Bending stiffness of the return spring
785 ±30.0 N/m
Internal length of the sensor
20.0 ±0.2 mm
Frictional force of the sensor
0.65 ±0.15 N
Radius of the impact body
R 25 ±1.0 mm
Active length of the impact spring
61.5 ±0.3 mm
Length of the impact body
75.0 ±0.3 mm
Display resolution
256 x 64 pixels
Max. number of test zones per component
16
Max. number of test points per test zone
16
Sensor type
contactless optocoupler sensor
Units of measurement, adjustable
MPa, N/mm²; kgf/mm²; psi
Communication options to the PC
serial interface (optional Bluetooth)
On/off key
Up key
Down/delete key
Enter key
Page 8
© PCE Instruments
5
4 Measurement parameters
4.1 Concrete strength calculation
The test angle and the surface of the test zone must be set accordingly if the concrete hardness tester is not in a horizontal position. This allows the device to automatically calculate the corresponding values during the measurement.
Concrete carbonization: The carbonization depth, if known, can be entered and taken into account.
The PCE-HT 224E is able to use 5 different concrete strength curves:
- no standard
- standard, ready-mixed concrete (national standard-pumping)
- standard, in-situ concrete (national standard-non pumping)
- concrete strength curve North America
- user-specific concrete strength curve
4.2 Concrete strength value in the standard setting
Each test zone should be tested 16 times. After 16 tests, the 3 highest and 3 lowest values are cancelled and the average value of the 10 remaining tests is calculated. This average value is then the measured value determined for the test zone.
Before starting the test, the test angle and the test direction must be set so that the concrete test hammer automatically takes the corresponding correction values into account.
Concrete strength and standard deviation After testing the 16 measuring points, the PCE-HT 224E automatically calculates the concrete strength value and the standard deviation. The determined values are saved.
Calculation of concrete strength in customised test mode If the parameters preset by the user do not correspond to the standard set of rules, these are to be regarded as customised user settings.
Calculation of the average concrete strength value in the test zone After the test procedure, the maximum and minimum values are not displayed, only the average rebound value.
Page 9
© PCE Instruments
6
5 Operation
5.1 Start a simple test
Individual measurements can be carried out immediately without further adjustment. First press the On/Off key, the following is displayed:
0.00R 00.0 MPa Then unlock the device as follows:
Place the device with the impact body (14) on a firm surface. Now press the device slowly against the surface at a 90° angle. At the same time, push the locking button (6) to release the lock. The impact body extends out of the housing when you move the device backwards.
To carry out a measurement, place the extended impact body at a right angle to the concrete surface.
Then press the device against the surface to be tested slowly but firmly. The impact body is tensioned. When the trigger position is reached, the impact body is released and strikes the surface.
- Caution! There is a serious risk of injury if handled incorrectly!
The measurement result is shown on the display as an R value and in MPa.
5.2 Test series in the measurement menu with default settings
After switching on the meter, press the Enter key twice to access the main menu. The following line appears: "Welcome and please take the next step"
Then use the arrow down key to select "Test mode setup" and confirm with "Enter". Once you have made your entry, test mode is switched on and you cannot return to a higher level of the selection menu without further ado.
Page 10
© PCE Instruments
7
You can set the following values here:
- Component name: e.g. GJXX
- Number of test procedures: (16)
- Number of test zones: (16)
- Concrete type: e.g: GB standard-pumping (ready-mixed concrete)
- Test angle: e.g.: 0° (horizontal)
- Test direction: top (from above), bottom (from below), side (sideways)
- Carbonization depth: e.g. 0.0 mm (standard setting) Once the setting has been completed, confirm it with "Enter" and return to test mode. The device
has automatically saved the measurement parameters. An unfinished test can be continued after switching the device off and on again. The test parameters can be changed during the test. To do this, press the Enter key to access the main menu, then select "Test mode" and continue to the "Testing zone set" selection window. When the set number of rebound tests has been made, the display shows "Testing zone is completed" and the next test series can be started after a 3-second pause.
The test series is completed after the last test of the last test zone. The tester then automatically switches to the evaluation area. You can then view the test ranges and results in this area.
Concrete curve
Limit value alarm
Battery status
Measuring angle
Measuring scale
Test direction
Carbonization depth MPa value
Rebound value
Measurement counter
Component no.
Test zone
00.0MPa
Name
Page 11
© PCE Instruments
8
5.3 Calibration
Calibration is carried out to check the accuracy of the device and to correct it if necessary. For this purpose, 16 test procedures must be carried out on a calibration unit intended for this purpose.
After the 16th test procedure, press "Enter" to confirm. The determined value is displayed. If the measured value of the device deviates too much from that of the calibration unit, see item 11 Appendix 1 - Troubleshooting.
5.4 Start new measurement
Use the arrow down key to select "New project" and confirm your selection with "Enter". A new test series is created. The default settings for the new test series are as described below: Component name like G J X X X, e.g. G J 0 0 1 Number of tests like xx, e.g. 16 Number of test zones like xx, e.g. 16
5.5 Enter part name
In the "New part" menu item, use the arrow keys to select the "Part name" sub-menu and confirm with "Enter" to enter the name. Set each individual digit using the arrow up key. There are 36 different options for each digit, from 0 - 9 and from A - Z. Use the arrow down key to select the next digit.
Page 12
© PCE Instruments
9
5.6 Enter the number of measurements in the test zone
Enter the number of tests (test time) in the "New part" menu item using the arrow keys. Use the arrow key to select a number between 8 and 16 and confirm your selection with "Enter".
After completing the entry, confirm it with "Enter" and continue to the test area selection.
5.7 Test zone entry
There are 4 important parameters that need to be set here: the test angle, the direction of impact to the surface, the depth of carbonization and the concrete strength curve.
5.7.1 Setting the test angle
There are 4 important parameters for the "Test zone". Use the arrow keys to select the test angle and confirm with "Enter". A total of 9 different angles are possible here. Confirm your selection with "Enter" and return to the parameter overview window.
Page 13
© PCE Instruments
10
5.7.2 Enter test surface
In the entry menu for the measurement parameters of a new test zone, "Test zone setup", the measurement directions can be selected using the arrow keys. Three measuring directions are possible here: top surface, bottom surface, side surface. Press "Enter" to confirm and return to the measurement parameter menu.
5.7.3 Setting the carbonization depth
In the selection menu for the "Carburized depth", the carbonization depth can be set within the range from 0.0 to 6.0 mm, using the arrow keys. 13 different settings are possible here: 0.0 mm,
0.5 mm, 1.0 mm, 1.5 mm, 2.0 mm, 2.5 mm, 3.0 mm, 3.5 mm, 4.0 mm, 4.5 mm, 5.0 mm, 5.5 mm,
6.0 mm. Confirm your entry with "Enter" and return to the "Test zone setup" menu.
5.7.4 Setting the strength curve
In the "Strength curve" menu, use the up and down arrow keys to select the required concrete strength curve. Then confirm the selection with "Enter". This will also take you back to the "Test zone setup" menu.
Once the setting is complete, confirm it with "Enter" and return to measuring mode.
Page 14
© PCE Instruments
11
5.8 Repeat test of the current test zone
Press "Enter" to return to the main menu. Then select "Test mode setup" and confirm with "Enter" to access the selection window for the test mode. Select "Re-testing the current zone" and press "Enter" to return to the test zone. The display is then reset to zero.
5.9 Basic" measurement settings
Press "Enter" to access the main menu. Select "Test mode setup" and confirm with "Enter". Then select "Basic measuring" and confirm with "Enter".
In this setting, only the measuring scale, the rebound value, the MPa value, the limit value alarm and the battery level indicator are displayed on the LCD. In this setting, 100 rebound measurements can be recorded at a time. This data can then be called up and displayed. When the memory is full, a new file is automatically created.
Measuring scale
Battery status display MPa value
Limit value alarm
Rebound value
Page 15
© PCE Instruments
12
5.10 Review measurement data
Press "Enter" to access the main menu. Then select "Review data" and press "Enter" to access the review menu.
1) Exit the current menu and call up the measurement menu
2) View part data
3) View the measured values of fully tested parts
4) View the measurement data of complete test zones
The data of the test zones of the last few tests are displayed on the LCD screen. The selection is made using the arrow keys; the menu can be exited by pressing the "Enter" key.
Page 16
© PCE Instruments
13
5.11 Delete measurement data
Press "Enter" to open the main menu. Then select "Memory" and confirm with "Enter" to access the memory management.
You have the following options here:
- Exit: Exit the current menu item and switch to measurement view
- Delete current part No. xxx: Remove the data record for the current part
- Delete all data: Delete all saved data records
- Memory capacity: Display of the current memory capacity in per cent
0 ... 100 %; the total storage capacity is up to 1000 data records for components.
6 Data transfer
6.1 Set up data transfer
Press "Enter" to access the main menu. Then select "Data transfer" and confirm with "Enter". You are now in the data transfer menu.
First connect the PCE-HT 224E to the PC using the mini-USB cable, then select the "Data transfer mode" menu and select "Cable". The "Bluetooth" option for wireless connection is not available. Selecting "Off" would switch off a Bluetooth connection.
Page 17
© PCE Instruments
14
Then select "Send to PC". The message "Transferring" appears on the display. The device is now ready for data transfer.
The software is intended for reading out and editing the test data of the PCE-HT 224 E concrete test hammer. The test data can be read out and saved and reports can be created and printed out. System requirements: - a PC with USB interface
- Windows XP, Win 7, Win 8, NetFramework 2.0, Win 10 and higher
6.2 Establishing a connection to the PC
Carry out the following steps to connect the concrete test hammer to the PC.
1. Connect the two devices using the mini USB cable supplied.
2. Switch on the concrete test hammer and press the Enter key to access the main menu. Select the menu item "Data Transfer" > "Data Transfer Mode" > "USB Cable".
3. Select "Data transfer" > "Send to PC". "Transferring" appears on the display and the data transfer begins.
6.3 Driver installation
Found new Hardware" appears on the computer screen when the concrete test hammer is connected to the PC for the first time. Copy the drivers to the directory X:/ Driver folder (software installation). Install the "Digital Rebound Hammer" software by double-clicking on "Setup\setup.exe".
Page 18
© PCE Instruments
15
6.4 Starting and exiting the PC software
Once the software has been successfully installed, a shortcut will automatically appear on the desktop. Double-click on it to start the programme.
The main view of the software is in standard Windows format and contains the title bar, the menu bar and the toolbar. The software can be used to analyse and document the measured assemblies and components, e.g. as an attachment to acceptance reports.
Important: Once the software has been installed, activate the connection between the PC and the rebound hammer via "CONNECT"!
Page 19
© PCE Instruments
16
7 Main menu
7.1 Setting measuring units
Select the various units via the "Unit setup" menu.
7.2 Setting the upper / lower limit values
Set the upper and lower limit values via the "Upper-Lower limit setup" menu.
The upper and lower limit values of the rebound value (00 ... 60) are set here. Use the up and down arrow keys to select the settings and confirm and save them with "Enter".
Page 20
© PCE Instruments
17
7.3 Rights management, password settings
Under the "Rights management" menu item, you can activate or remove password protection or change the password. The password is then valid for data retrieval, limit value entry and rights management. Use the arrow down key to move the pointer to the next letter, use the arrow up key to set the numbers from 0 to 9 and then confirm with the Enter key.
7.4 Energy saving settings
In the "Power-saving mode setup" menu, select the desired settings using the arrow keys and confirm them with "Enter".
In this menu item, you will find 3 selection options: Normal energy consumption, i.e. approx. 8 - 10 hours operating time when the device is fully
charged
Reduced operation, energy consumption approx. 60 %; in this setting approx. 10 - 13 hours operating time is common when the device is fully charged
Page 21
© PCE Instruments
18
Economy mode, energy consumption approx. 30 %; approx. 13 - 16 hours operating time is common in this setting when the device is fully charged
In the energy saving settings, you can switch back to the normal energy setting by pressing "Enter". If you then press "Enter" again and then start the measurement, you will return to the preset energy-saving modes.
7.5 Date setting
Set the date in the "Date of test setup" menu. To do this, press the arrow down key to move the cursor to the right and use the arrow up key to set the current digit.
7.6 Setting the counter
Select the "Counter" setting. Here you can see the number of saved test procedures. You can also reset the number of test procedures in this view. If you select "Clear" and confirm with "Enter", the display is reset to zero.
7.7 Battery status display
The battery capacity is displayed in per cent (0 ... 100 %).
7.8 Instrument data (Info of Instrument)
Display of device data (serial number and firmware version)
Page 22
© PCE Instruments
19
8 Calibration
After repair or service work, the device calibration can be checked in the "Calibration" setting. To do this, use the up and down arrow keys to select the "Calibration" setting and confirm with "Enter". Then carry out a total of 16 test procedures on a test anvil intended for this purpose. After completing the 16 test procedures, confirm them with the Enter key.
The following is displayed:
The device is now calibrated and the display shows the zero position.
8.1 Completion of a test series
Once a test zone has been completed, the device automatically switches to the next test zone. When all test zones have been measured, the device switches to the "Part test finished and take next step" display area (completion of a test series). This completes the current series of measurements. You will see the menu items in the display: Review test part result, Review test zone result, New part test, Re-testing the current zone.
In the "Review test zone result" view, the measurement results are displayed on the right-hand side and the associated measurement parameters at the bottom left.
View results in the test area In the "Review test part result" view, the test results of each test area can be seen on the right-
hand side of the display. The name and numbering can be found at the bottom left.
Page 23
© PCE Instruments
20
9 Maintenance and servicing
Maintenance measures are important for the PCE-HT 224E concrete tester under the following conditions.
More than 2000 rebound tests
Inconclusive test results
Number of rebound tests performed unknown
Simple maintenance measures can be carried out as described below:
Dismantle the untensioned hammer, open the upper housing by turning the knurled screw at the upper (thick) end anti-clockwise and remove the rebound spring, the firing pin and the measuring unit. To do this, you must also loosen the retaining nut at the front end and remove the felt seal and the bearing shells. Then carefully remove the guide rod with impact weight and impact body. To do this, gently push in the locking hook and pull out the unit. Note the positions of the guide rails and the locks for subsequent reassembly.
Carefully clean all internal components with compressed air, then slightly oil the central guide tube. All other components must not be oiled.
Clean the inside wall of the housing carefully. The holding torque of the locking unit should be approx. 0.5 ... 0.8 Nm.
The retaining nut (large knurled screw) must no longer be able to be turned; it must be slightly tightened.
Self-manufactured spare parts must not be used.
After repairs have been carried out, a self-calibration must be performed (see also 5.3
Calibration). The rebound value should be 80 ±2. Select the calibration function, then carry out the test procedure 16 times on the corresponding calibration anvil in this setting. Then confirm with "Enter" to save the calibration and exit this setting.
10 PC software
See chapter 6 Data transfer.
Page 24
© PCE Instruments
21
11 Appendix 1 - Troubleshooting
If the display value suddenly stops changing during operation of the device or the rebound value shows a large deviation, a mechanical fault has usually occurred. The troubleshooting options are listed in the table below:
The replacement of parts or the mechanical repair of components may only be carried out by authorised specialist personnel!
Error
Cause
Repair method
I
During the function check, the signal transmitter (22) blocks at the starting point.
1. The distance between the signal generator (20), signal transmitter (22) and guide rail (26) is too small.
Dismantle the signal transmitter (22) and carefully change the angle of the signal generator (20).
2. The signal generator (20) is broken.
Dismantle the signal transmitter (22) and replace the signal generator (20).
II
The signal transmitter (22) is already pushed up before the rebound test and displays the rebound value.
1. The contact angle between the signal generator (20) and the signal transmitter (22) is too large.
Dismantle the signal transmitter (22) and carefully bend the contact angle of the signal generator (20) together.
2. The clearance between the signal transmitter (22) and the guide rail (26) is too large.
Remove the signal transmitter (22) and set the frictional force between the signal transmitter (22) and the guide rail (26) (0.65 ±0.15 N).
III
During the rebound process, the signal transmitter (22) shakes, moves to a certain point and blocks there.
The signal transmitter (22) rubs or jumps against the housing body.
Dismantle the signal transmitter (22) and carefully smoothen the guides with a small file.
IV
The impact weight (10) triggers too early.
1. The retaining claw (3) is blunt.
Sharpen the edge of the retaining claw (3) with a file.
2. The detent on the impact weight (10) is broken/worn.
Change the angle of the impact weight (10) or replace it.
V
The rebound device cannot be pushed back and the impact weight (10) does not move back.
1. The locking spring (28) does not engage and does not catch the retaining claw.
Refit the locking spring (28) or restore the tension and position by bending it open.
2. The end of the retaining claw (3) is broken.
Replace the retaining claw (3).
3. The end of the retaining claw (3) is worn.
Carefully grind the retaining claw (3) back into shape.
Page 25
© PCE Instruments
22
VI
The impact weight (10) is stiff or does not release.
1. The thick end of the retaining claw (3) touches the surface of the impact weight (10).
Grind 1 mm off the thick end of the retaining claw (3).
2. The retaining claw (3) is too sharp-edged.
Grind the catch hook of the retaining claw (3) to the right angle.
VII
The impact body (14) does not move and does not extend, device does not work.
The locking button (6) is loose, the small spring in the inner part is not working properly.
Hold one end of the coil spring and twist the spring 30 times to restore the tension. Then refit the spring into the locking button (6).
VIII
The rebound value is relatively high.
1. The impact spring (17) is longer than 61.5 mm.
Adjust the position of the impact spring (17) in the guide bushing (11).
2. The starting position of the impact weight (10) is relatively high. The impact spring (17) is too long.
Turn the adjusting bolt on the back of the housing (30) outwards.
3. There is too much oil on the guide rod (9).
Dismantle the relevant components of the rebound body, clean them with a soft cotton cloth and then slightly oil them.
IX
The rebound value is relatively low.
1. The impact spring (17) is shorter than 61.5 mm.
Adjust the position of the impact spring (17) on the guide bushing (11).
2. The starting position of the impact weight (10) is relatively low.
Turn the adjusting bolt on the back of the housing (30) inwards.
3. The friction of the signal transmitter (22) is relatively high.
Set the distance angle between the signal transmitter (22) and the guide rail (26) accordingly and set a friction value of 0.65 ±0.15 N.
4. The impact weight (10) is dirty.
Clean the impact weight thoroughly.
5 The friction between the impact weight (10) and the guide rod (9) increases.
Slightly oil the guide rod (9).
Page 26
© PCE Instruments
23
X
The rebound value is unstable and fluctuates significantly.
1. The contact surface between the impact weight (10) and the impact body (14) is uneven.
Rotate the impact body (14) by 180° or replace it.
2. The centre guide rod (9) is dirty.
Dismantle and clean the device and then oil it slightly.
3. The centre guide rod (9) is bent.
Align the guide rod (9) accordingly.
4. The friction between the signal transmitter (22) and the guide rail (26) is not the same.
Adjust the friction between the signal transmitter (22) and the guide rail (26).
5. The contact between the signal generator (20) and the impact weight (10) is poor.
Adjust the angle of the signal generator (20) accordingly.
6 The guide rail (26) rubs or strikes against the housing body (7).
Carefully smoothen the surfaces of the signal transmitter (22) with a lapping stone; also smoothen the guide channel on the housing body if necessary.
7. The guide rail (26) is bent.
Realign the guide rail.
Page 27
© PCE Instruments
24
12 Appendix 2 - Conversion table for concrete strength
Rebound value
Concrete compressive strength in MPa = N/mm²
for the direction of impact
top
horizontal
downwards
90°
60°
45°
30°
-30°
-45°
-60°
-90°
20
10.3
10.3
10.3
10.3
10.3
13.1
13.7
14.3
14.9
21
10.3
10.3
10.3
10.3
11.4
14.3
14.9
15.5
16.2
22
10.3
10.3
10.3
10.3
12.5
15.4
16.0
16.7
17.4
23
10.3
10.3
10.3
10.4
13.7
16.7
17.4
18.0
18.8
24
10.3
10.3
10.5
11.6
14.9
17.9
18.6
19.3
20.0
25
10.3
10.8
11.6
12.7
16.2
19.2
20.0
20.8
21.5
26
11.0
12.0
12.8
14.0
17.5
20.6
21.4
22.1
22.8
27
11.9
13.3
14.0
15.3
18.9
22.1
22.8
23.6
24.5
28
13.4
14.6
15.4
16.7
20.3
23.5
24.3
25.0
25.9
29
14.8
16.0
16.7
18.0
21.8
25.0
25.9
26.7
27.6
30
16.2
17.5
18.2
19.6
23.3
26.5
27.4
28.2
29.1
31
17.6
18.9
19.6
21.0
24.9
28.2
29.1
30.0
30.9
32
19.1
20.8
21.2
22.7
26.5
29.8
30.7
31.6
32.5
33
20.8
22.0
22.7
24.5
28.2
31.6
32.5
33.5
34.4
34
22.4
23.6
24.5
26.0
30.0
33.3
34.2
35.2
36.1
35
24.1
25.2
26.0
27.8
31.8
35.2
36.1
37.1
38.2
36
25.9
27.1
27.9
29.6
33.6
36.9
37.9
38.9
39.9
37
27.8
28.8
29.6
31.6
35.5
38.9
39.9
41.0
42.0
38
29.6
30.7
31.6
33.5
37.5
40.7
41.8
42.8
43.9
39
31.6
32.5
33.5
35.4
39.5
43.7
43.9
45.0
46.1
40
33.6
34.6
33.5
37.5
41.6
44.8
45.9
47.0
48.1
41
35.5
36.5
37.5
39.5
43.7
47.0
48.1
49.2
50.4
42
37.7
38.7
39.7
41.8
45.9
49.0
50.2
51.3
52.5
43
39.7
40.7
41.8
43.9
48.1
51.3
52.5
53.6
54.8
44
42.0
43.0
44.1
46.3
50.4
53.4
54.6
55.8
57.0
45
44.1
45.2
46.3
48.5
52.7
55.8
57.0
58.2
59.5
46
46.5
47.6
48.7
51.0
55.0
58.0
59.2
60.0 47
48.7
49.9
51.0
53.4
57.5
48
51.3
52.5
53.6
56.0
60.0
49
53.6
54.8
56.0
58.5 50
56.8
57.5
58.8
60.0
Page 28
© PCE Instruments
25
13 Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact us. You will find the relevant contact information at the end of this user manual.
14 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations.
If you have any questions, please contact PCE Instruments.
Page 29
© PCE Instruments
26
PCE Instruments contact information
Germany France Spain
PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L. Im Langel 26 23, rue de Strasbourg Calle Mula, 8 D-59872 Meschede 67250 Soultz-Sous-Forets 02500 Tobarra (Albacete) Deutschland France España Tel.: +49 (0) 2903 976 99 0 Téléphone: +33 (0) 972 3537 17 Tel. : +34 967 543 548 Fax: +49 (0) 2903 976 99 29 Numéro de fax: +33 (0) 972 3537 18 Fax: +34 967 543 542 info@pce-instruments.com info@pce-france.fr info@pce-iberica.es www.pce-instruments.com/deutsch www.pce-instruments.com/french www.pce-instruments.com/espanol
United Kingdom Italy Turkey
PCE Instruments UK Ltd PCE Italia s.r.l. PCE Teknik Cihazları Ltd.Şti. Trafford House Via Pesciatina 878 / B-Interno 6 Halkalı Merkez Mah. Chester Rd, Old Trafford 55010 Loc. Gragnano Pehlivan Sok. No.6/C Manchester M32 0RS Capannori (Lucca) 34303 Küçükçekmece - İstanbul United Kingdom Italia Türkiye Tel: +44 (0) 161 464902 0 Telefono: +39 0583 975 114 Tel: 0212 471 11 47 Fax: +44 (0) 161 464902 9 Fax: +39 0583 974 824 Faks: 0212 705 53 93 info@pce-instruments.co.uk info@pce-italia.it info@pce-cihazlari.com.tr www.pce-instruments.com/english www.pce-instruments.com/italiano www.pce-instruments.com/turkish
The Netherlands United States of America Denmark
PCE Brookhuis B.V. PCE Americas Inc. PCE Instruments Denmark ApS Institutenweg 15 1201 Jupiter Park Drive, Suite 8 Birk Centerpark 40 7521 PH Enschede Jupiter / Palm Beach 7400 Herning Nederland 33458 FL Denmark Telefoon: +31 (0)53 737 01 92 USA Tel.: +45 70 30 53 08 info@pcebenelux.nl Tel: +1 (561) 320-9162 kontakt@pce-instruments.com www.pce-instruments.com/dutch Fax: +1 (561) 320-9176 www.pce-instruments.com/dansk info@pce-americas.com www.pce-instruments.com/us
Loading...