Please read this manual carefully and completely before you use the device for the first time. The
device may only be used by qualified personnel and repaired by PCE Instruments personnel.
Damage or injuries caused by non-observance of the manual are excluded from our liability and
not covered by our warranty.
•The device must only be used as described in this instruction manual. If used otherwise,
this can cause dangerous situations for the user and damage to the meter.
•The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
• Do not expose the device to shocks or strong vibrations.
• The case should only be opened by qualified PCE Instruments personnel.
• Never use the instrument when your hands are wet.
• You must not make any technical changes to the device.
• The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
• The device must only be used with accessories from PCE Instruments or equivalent.
• Before each use, inspect the case for visible damage. If any damage is visible, do not
use the device.
• Do not use the instrument in explosive atmospheres.
• The measurement range as stated in the specifications must not be exceeded under
any circumstances.
•Non-observance of the safety notes can cause damage to the device and injuries to
the user.
We do not assume liability for printing errors or any other mistakes in this manual.
We expressly point to our general guarantee terms which can be found in our general terms of
business.
If you have any questions please contact PCE Instruments. The contact details can be found at
the end of this manual.
o The default setting is none
o Press the ◄ or ► button to select the value, then use the ▲ and ▼ buttons
to toggle between none / wep_open / wep / wpa_tkip / wpa_aes / wpa2_tkip
/ wpa2_aes / wpawpa2_tkip / wpawpa2_aes
o Press the SETUP/ENTER button to confirm your change.
o Press the ▼ button to move the selection to Password.
• Password:
o The default password is 12345678
o Press the ◄ or ► button to select the value, then use the ▲ and ▼ buttons
to input numbers or letters, after pressing the EXIT/SHIFT button.
o Press the SETUP/ENTER button to confirm your change.
o Press the ▼ button to move the selection to IP.
• IP:
o The default IP address is 192.168.0.254
o Press the ◄ or ► button to select the value and then use the ▲ and ▼
buttons to input numbers for a new IP address.
o Press the SETUP/ENTER button to confirm your change.
o Press the ▼ button to move the selection to Port.
• Port:
o The default port is 200
o Press the ◄ or ► button to select the value and then use the ▲ and ▼
buttons to input numbers for a new port (max. 65535 numbers).
o Press the SETUP/ENTER button to confirm your change.
o Press the ▼ button to move the selection to Gateway.
• Gateway:
o The default gateway is 192.168.0.1
o Press the ◄ or ► button to select the value and then use the ▲ and ▼
buttons to input numbers for a new gateway.
o Press the SETUP/ENTER button to confirm your change.
o Press the ▼ button to move the selection to Unit.
• Unit:
oPress the ◄ or ► button to select the value, and then use the ▲ and ▼
buttons to toggle between C and F.
o Press the SETUP/ENTER button to confirm your change.
o Press the ▼ button to move the selection to Default Settings.
• Default Settings:
o Restore to factory settings (refer to point 5.6).
•Press the EXIT/SHIFT ( ) button to save the changed settings. The device then needs
~25 sec. to reconnect (Screen 7 or 8) before returning to the measurement screen
(Screen 5 or 6).
5.5 Measurement and Wi-Fi connection
• Screen 3 is the temperature and humidity measurement screen.
• Pt1000 measurement mode
o Insert the probe into the PT 1K input (Fig.1 3.18).
o Press the ► button to toggle between the PT1000 measurement screen
(Screen 4) and the Temp/Humidity measurement more (Screen 3)
•A Wi-Fi symbol ( ) will flash in the upper, right hand corner of both measurement
screens to indicate an existing connection.
•After starting the device, enter the settings menu (Screen 9) and set the following
parameters:
NetMode: AP
B.SSID: RHW-925
C.IP: 192.168.0.254
D.Port: 200
• Connect to the device with your smart phone Wi-Fi.
• Start the MeterApp and select Meter 1 on the first screen (Image 2).
• Press the settings button (Image 3, step 2).
• In the settings menu (Image 4) you can set/change the following settings:
o Meter name (default is Meter 1)
o CH No: channel number in accordance with the device channel number
o Sampling time: must be set, in minutes. Check back after doing the logging
function.
oIP: must be set in accordance with the device. The default IP address is
192.168.0.254.
o Port: must be set in accordance with the device. The default port is 200.
o Save File: select the log file save path here.
o Alarm Beeper: device beeps when the conditions on Image 6 are met.
o Alarm SMS: device sends an SMS as an alarm to your smart phone when
the conditions on Image 6 are met.
•The Save button (Image 5) stores the set settings and returns the device to Image 3. The
Wi-Fi icon (Step 3) will now flash green, indicating that the device is connected.
•When in the project screen (Image 3), tap on CH1 (step 4) to enter the settings screen (image
6) and set the desired parameters. Then press the return button to go back to the project
screen (Image 3). When the measured values reach the conditions set in Image 6, the
numbers will become red and the device will either beep or send an SMS.
•When in the project screen (Image 3), use the buttons in Step 5 to switch the display directly
between Meter1 to Meter4.
• Re-connect Use the reconnect button if the device loses connection.
• Auto Log: the log will be saved based on the sampling time set in Image 4.
• Manual Log: press the Manual Log button to save the log manually. The log file is in xml
formal.
•View Log: view the recorded data directly on your smartphone (Image 7).
o Determine the SSID for the meter.
o Setting a password is not necessary.
• Set the meter to Client mode with the following parameters:
o Net mode: Client
o SSID: RHW-925 (or used defined) Internet
o IP: 192.168.0.254
o Port: 200
Refer to Screen 6.
5.9 Client mode (external network, fixed IP)
• Simple schematic network:
• Set the meter to Client mode with the following parameters:
o Net mode: Client
o SSID: RHW-925 (or used defined)
o IP: 192.168.0.254
o Port: 200
o Gateway: 192.168.0.1 (download IP Tools from the App Store to learn how
to set this parameter).
•In the port forwarding settings of your IP router, you can set the parameters for SSID,
IP address and port such as here:
Image 8
•Enter the needed parameters in Image 4 of the MeterApp. Set the IP address according
• Set the meter to Client mode with the following parameters:
o Net mode: Client
o SSID: RHW-925 (or used defined)
o IP: 192.168.0.254
o Port: 200
o Gateway: 192.168.0.1 (download IP Tools from the App Store to learn how
to set this parameter).
•In the port forwarding settings of your IP router, you can set the parameters for SSID,
IP address and port such as in Image 8.
•Apply for free transfer to www. noip. com. The name of the application will be
transferred to the address input of your IP Router.
Image 9
•Enter the needed parameters in Image 4 of the MeterApp. Set the IP address according
D9
Decimal Point (DP), position from right to left
0 = No DP
1 = 1 DP
2 = 2 DP
3 = 3 DP
D8 – D1
Display reading: D1 = LSD, D8 = MSD
If the display reads 1234, the D8 to D1 is: 00001234
D0
End word
Baud rate
9600
Parity
No parity
Data bit no
8 Data bits
Stop bit
1 Stop bit
The device has a 3.5 mm socket (3-17, Fig. 1) for a RS232 computer interface.
The connector output is a 16 digit data stream, which can be utilized to the user's specific
application.
A RS232 lead with the following connection will be required to link the instrument with the PC,
NB USB port or serial port.
The 16 digits data stream will be displayed in the following format:
You can read our warranty terms in our General Business Terms which you can find here:
https://www.pce-instruments.com/english/terms
.
7 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament
applies. Due to the contained pollutants, batteries must not be disposed of as household waste.
They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use
them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with
your local waste regulations.
If you have any questions, please contact PCE Instruments.
PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L.
Im Langel 4 23, rue de Strasbourg Calle Mayor, 53
D-59872 Meschede 67250 Soultz-Sous-Forets 02500 Tobarra (Albacete)
Deutschland France España
Tel.: +49 (0) 2903 976 99 0 Téléphone: +33 (0) 972 3537 17 Tel. : +34 967 543 548
Fax: +49 (0) 2903 976 99 29 Numéro de fax: +33 (0) 972 3537 18 Fax: +34 967 543 542
info@pce-instruments.com info@pce-france.fr info@pce-iberica.es
www.pce-instruments.com/deutsch www.pce-instruments.com/french www.pce-instruments.com/espanol
Germany United Kingdom Italy
PCE Produktions- und PCE Instruments UK Ltd PCE Italia s.r.l.
Entwicklungsgesellschaft mbH Unit 11 Southpoint Business Park Via Pesciatina 878 / B-Interno 6
Im Langel 26 Ensign Way, Southampton 55010 Loc. Gragnano
D-59872 Meschede Hampshire Capannori (Lucca)
Deutschland United Kingdom, SO31 4RF Italia
Tel.: +49 (0) 2903 976 99 471 Tel: +44 (0) 2380 98703 0 Telefono: +39 0583 975 114
Fax: +49 (0) 2903 976 99 9971 Fax: +44 (0) 2380 98703 9 Fax: +39 0583 974 824
info@pce-instruments.com info@pce-instruments.co.uk info@pce-italia.it
www.pce-instruments.com/deutsch www.pce-instruments.com/english www.pce-instruments.com/italiano
The Netherlands China Hong Kong
PCE Brookhuis B.V. PCE (Beijing) Technology Co., Limited PCE Instruments HK Ltd.
Institutenweg 15 1519 Room, 6 Building Unit J, 21/F., COS Centre
7521 PH Enschede Zhong Ang Times Plaza 56 Tsun Yip Street
Nederland No. 9 Mentougou Road, Tou Gou District Kwun Tong
Telefoon: +31 (0)53 737 01 92 102300 Beijing, China Kowloon, Hong Kong
info@pcebenelux.nl Tel: +86 (10) 8893 9660 Tel: +852-301-84912
www.pce-instruments.com/dutch info@pce-instruments.cn jyi@pce-instruments.com
www.pce-instruments.cn www.pce-instruments.cn
United States of America Turkey
PCE Americas Inc. PCE Teknik Cihazları Ltd.Şti.
711 Commerce Way suite 8 Halkalı Merkez Mah.
Jupiter / Palm Beach Pehlivan Sok. No.6/C
33458 FL 34303 Küçükçekmece - İstanbul
USA Türkiye
Tel: +1 (561) 320-9162 Tel: 0212 471 11 47
Fax: +1 (561) 320-9176 Faks: 0212 705 53 93
info@pce-americas.com info@pce-cihazlari.com.tr
www.pce-instruments.com/us www.pce-instruments.com/turkish