Monitor with real time data logger, save the measuring
*
data along the time information ( year, month, date,
minute, second ) into the SD memory card and can be
down load to the Excel, extra software is no need.
User can make the further data or graphic analysis by
themselves.
*
Show Humidity and Temperature value in the same LCD.
0.1 %RH resolution for the humidity reading,
*
0.1 degree resolution for the Temp. reading.
* Used the precision capacitance type humidity sensor,
professional and high accuracy.
* SD card capacity : 1 GB to 16 GB.
* Sampling adjustment : 5/10/30/60/120/300 seconds
and auto function.
* Large LCD display, easy readout.
* Microcomputer circuit, high reliability.
* Low power consumption and long battery life when
use battery power.
* DC 1.5V ( UM-4, AAA ) battery x 6 PCs or DC 9V
adapter in.
* RS232/USB computer interface available.
*Patented
2. SPECIFICATIONS
2-1 General Specifications
CircuitCustom one-chip of microprocessor LSI
circuit.
DisplayLCD size : 60 mm x 50 mm
MeasurementHumidity and Temperature
Memory CardSD memory card, 1 GB to 16 GB.
Page 4
Datalogger
A
A
.
Sampling Time
5
/10/30/60/120/300/600 seconds
or Auto.
Default sampling time is 60 seconds.
*
*
The " Auto " sampling . means when
the measuring value is changed ( > ±
1 %RH or > ±
1 ) will save the℃
data one time only.
Advanced
setting
* SD
memory card Format
* Set clock time ( Year/Month/Date, Hour/Minute/
cond )
Se
* Set sampling time
* Set beeper sound ON/OFF
* Set SD card Decimal character
* Select the Temp. unit to or ℃℉
* Set RS232 data output ON/OFF
Update TimeApprox. 1 second if measuring data
of Displayi
s changed.
Data OutputRS 232/USB PC computer interface.
* Connect the optional RS232 cable
UPCB-02 will get the RS232 plug.
*
Connect the optional USB cable
USB-01 will get the USB plug.
Operating 0 to 50 .℃
Temperature
Operating Less than 85% R.H.
Humidity
Power Supply
*
Alkaline or heavy duty DC 1.5 V battery
( UM4, AAA ) x 6 PCs, or equivalent.
*
DC 9V adapter input. ( AC/DC power
adapter is optional ).
Battery life
f use the new battery ( alkaline type )
I
and sampling time set to 60 seconds,
the battery life will be > one month
typically.
Weight282 g/0.62 LB.
Dimension132 x 80 x 32 mm
( 5.2 x 3.1 x 1.3 inch )
AccessoriesInstruction manual...........................1 PC
Included
Hanging unit ( with sticker ).............
1 PC
OptionalSD Card ( 2 GB ).
AccessoriesAC to DC 9V adapter.
USB cable, USB-01.
RS232 cable, UPCB-02.
Data Acquisition software, SW-U801-WIN.
Page 5
2-2 Electrical Specifications (23±5 )℃
Humidity
Range10 % to 95 % R.H.
Resolution
Accuracy70% RH :≧
Temperature
0.1 % R.H.
± (3% reading + 1% RH).
< 70% RH :
± 3% RH.
Range0
Resolution0.1 degree
Accuracy - 0.8 .℃℃
@ Above specification tests under the environment RF Field Strength
3. FRONT PANEL DESCRIPTION
less than 3 V/M & frequency less than 30 MHz only.
to 50 ,32 to 122 .℃℃℉ ℉
- 1.5
.℉℉
Fig. 1
Page 6
3-1 Display
3-2 Logger button, Enter button
3-3 button, Time button▲
3-4 button▼
3-5 SET button
3-6 Humidity/Temp. sensor
3-7 Hanging holes
3-8 Stand
3-9 Battery cover/Battery compartment
3-10 Screw of the battery cover
3-11 Reset button
3-12 RS-232 output terminal
3-13 SD card socket
3-14 DC 9V power adapter input socket
3-15 Hanging unit ( with sticker )
4. MEASURING PROCEDURE
1)Install the batteries into the battery compartment :
oose the " Screw of the battery cover " ( 3-10, Fig. 1 )
* L
and take away the " Battery Cover " ( 3-9, Fig. 1 ) from
the meter.
* Replace with DC 1.5 V battery ( UM4/AAA, Alkaline/Heavy
duty type ) x 6 PCs, and reinstate the cover.
* Make sure the battery cover is secured after changing
the battery.
2)The " Display " ( 3-1, Fig. 1 ) will show both the
humidity and the temperature value that sensing from
Humidity/Temp. sensor " ( 3-6, Fig. 1 ).
5. DATALOGGER
5-1 Preparation before execute datalogger function
a. Insert the SD card
Prepare a " SD memory card " ( 1 GB to 16 GB, optional ),
insert the SD card into the " SD card socket " ( 3-13, Fig. 1)
with the correct direction exactly.
b. SD card Format
If SD card just the first time use into the meter, it
recommend to make the " SD card Format " at first,
please refer chapter 7-1 ( page 11 ).
c. Time setting
If the meter is used at first time, it should to adjust the
clock time exactly, please refer chapter 7-2 ( page 11 ).
Page 7
d. Decimal format setting
The numerical data structure of SD card is
default used the " . " as the decimal, for
example "20.6" "1000.53" . But in certain
countries ( Europe ...) is used the " , " as the
decimal point, for example " 20, 6 "
"1000,53". Under such situation, it should
change the Decimal character at first, details
of setting the Decimal point, refer to Chapter
7-5, page 13.
e. 3 Information of LCD display
*If the Display show :
It means that the SD card exist the problem
CH-or the SD card memory is full, it should
CArdchange SD memory card.
* If the Display show :
It means that the battery is low voltage.
LobAtUnder such condition, the Datalogger
function is disable.
* If the Display show :
It means that the SD card is not plugged
nointo the meter.
CArd
Page 8
5-2 Datalogger
*
*
a. Start the datalogger
Press the " Logger button ( 3-2, Fig. 1 ) > 2 seconds
continuously, until the Display show the indicator "
DATALOGGER ", release the " Logger Button " ( 3-2, Fig.
1 ), then the measuring data along the time information
will be saved into the memory circuit.
Remark :
How to set the sampling time, refer to Chapter 7-3,
page 12.
How to set the beeper sound is enable, refer to
Chapter 7-4, page 12.
c.. Finish the Datalogger
During execute the Datalogger function ( Display show the
" Datalogger " indicator ), press the " Logger button " ( 3-2,
Fig. 1 ) > 2 seconds continuously, until the Display
indicator " DATALOGGER " is disappeared, release the
" Logger Button " will finish the Datalogger function.
Before take away the SD card from the
meter, it should execute the
procedures of " Finish the Datalogger ",
otherwise some existing recent save
data may loss.
5-3 Check time information
Press " Time button " ( 3-3, Fig. 1 ) > 2 seconds
continuously, the LCD display will present the time
information of Year/Month/Date, Hour/Minute/Second and
the sampling value.
Page 9
5-4 SD Card Data structure
1)When the first time, the SD card is used into the meter,
the SD card will generate a folder :
HTC01
2)If the first time to execute the Datalogger, under the
route HTC01\, will generate a new file name
HTC01001.XLS.
After exist the Datalogger, then execute again, the data
will save to the HTC01001.XLS until Data column reach
to 30,000 columns, then will generate a new file, for
example HTC01002.XLS
3)Under the folder HTC01\, if the total files more
than 99 files, will generate anew route, such as
HTC02\ ........
4)
The file's route structure :
HTC01\
HTC01001.XLS
HTC01002.XLS
.....................
HTC01099.XLS
HTC02\
HTC02001.XLS
HTC02002.XLS
.....................
HTC02099.XLS
HTCXX\
.....................
.....................
Remark :
XX : Max. value is 10.
Page 10
6. Saving data from the SD card
(
)
to the computer ( EXCEL software )
1)After execute the Data Logger function, take away the
SD card out from the " SD card socket " ( 3-13, Fig. 1 ).
2)Plug in the SD card into the Computer's SD card slot
( if your computer build in this installation ) or
insert the SD card into the " SD card adapter ". then
connect the " SD card adapter " into the computer.
Power ON the computer and run the " EXCEL software ".
3)
Down load the saving data file ( for example the file
name : HTA01001.XLS, HTA01002.XLS ) from the SD
card to the computer. The saving data will present into
the EXCEL software screen ( for example as following
EXCEL data screens ) , then user can use those EXCEL
data to make the further Data or Graphic analysis
usefully.
EXCEL data screen
for example
Page 11
EXCEL graphic screen ( for example )
.
.
.
.
7. ADVANCED SETTING
Under do not execute the Datalogger function, press the
" SET button " ( 3-5, Fig. 1 ) > 2 seconds continuously will
enter the " Setting " mode., then release the " SET button ".
Following press the " SET button " (3-5, Fig. 1 ) once a
while in sequence to select the seven main function, the
display will show :
Sd F..... SD memory card Format
dAtE.....
SP-t......Set sampling time
bEEP....
dEC......
t-CF......Select the Temp. unit to or ℃℉
rS232...
Set clock time ( Year/Month/Date, Hour/Minute/
Second )
Set beeper sound ON/OFF
Set SD card Decimal character
Set RS232 data output ON/OFF
Page 12
Remark :
During execute the " Setting " function, if within
5 seconds , do not press any buttons further, the
LCD Display will return to normal screen.
7-1 SD memory card Format
When the Display show " Sd F "
1)Use the " Button " ( 3-3, Fig. 1 ) or "
3-4, Fig. 1 ) to select the upper value to " yES " or
" no ".
yES - Intend to format the SD memory card
no - Not execute the SD memory card format
2)If select the upper to " yES ", press the " Enter Button
" ( 3-2, Fig. 1 ) once again, the Display will show text
" yES Enter " to confirm again, if make sure to do the
SD memory card format, then press " Enter Button "
once will format the SD memory clear all the existing
data that already saving into the SD card.
Button " (▲▼
7-2 Set clock time ( Year/Month/Date,
Hour/Minute/ Second )
When the Display show " dAtE "
1)Use the " Button " ( 3-3, Fig. 1 ) or "
( 3-4, Fig. 1 ) to adjust the value ( Setting start from
Year value ). After the desired value is set, press the
" Enter button " ( 3-2, Fig. 1 ) once will going to
next value adjustment ( for example, first setting
value is Year then next to adjust Month, Date, Hour,
Minute, Second value ).
Button " ▲▼
Page 13
Remark :
The adjusted unit will be flashed.
2)After set all the time value ( Year, Month, Date, Hour,
Minute, Second ), press the " SET button " ( 3-5, Fig.
1 ) once will save the time value, then the screen will
jump to Sampling time " setting screen ( Chapter 7-3 ).
Remark :
After the time value is setting, the internal clock will
run precisely even Power off if the battery is under
normal condition ( No low battery power ).
After the desired value is set, press the " Enter Button "
( 3-2, Fig. 1 ) to save the adjusting value with default.
Remark :
The " Auto " sampling time means when the measuring
value is changed ( > ±
1 %RH or > ± 1
) will save℃
the data to the memory circuit one time.
7-4 Set beeper sound ON/OFF
When the Display show " bEEP "
1)Use the " Button " ( 3-3, Fig. 1 ) or "
( 3-4, Fig. 1 ) to select the data to " yES " or " no ".
Button " ▲▼
Page 14
yES - Meter's beep sound will be ON with default
when save each data.
no - Meter's beep sound will be OFF with default.
when save each data.
After select the upper text to " yES " or " no ", press the
2)
" Enter Button " ( 3-2, Fig. 1 ) will save the setting
function with default.
7-5 Decimal point of SD card setting
When the Display show " dEC "
The numerical data structure of SD card is used the " . "
as the decimal with default, for example "20.6"
"1000.53" . But in certain countries ( Europe ...) is
used the " , " as the decimal point, for example " 20,6 "
"1000,53". Under such situation, it should change the
Decimal character at first.
Use the " Button " ( 3-3, Fig. 1 ) or "
1)
( 3-4, Fig. 1 ) to select the upper text to " USA " or
" Euro ".
USA - Use " . " as the Decimal point with default.
Euro - Use " , " as the Decimal point with default.
2)After select the text to " USA " or " Euro ",
press the " Enter Button " ( 3-2, Fig. 1 ) will save the
setting function with default.
Button " ▲▼
Page 15
7-6 Select the Temp. unit to or ℃℉
When the Display show " t-CF "
1)Use the " Button " ( 3-3, Fig. 1 ) or "
( 3-4, Fig. 1 ) to select the upper Display text to " C " or
" F ".
C - Temperature unit is ℃
F - Temperature unit is ℉
2)After Display unit is selected to " C " or " F ", press the
" Enter Button " ( 3-2, Fig. 1 ) will save the setting
function with default.
Button " ▲▼
7-7 Set RS232 data output ON/OFF
When the Display show " rS232 "
1)Use the " Button " ( 3-3, Fig. 1 ) or "
( 3-4, Fig. 1 ) to select the upper Display text to "yES " or
" no ".
yES - RS-232 output terminal ( 3-12, Fig. 1 )
will send the RS232 signal output.
no - RS-232 output terminal ( 3-12, Fig. 1 )
will not send the RS232 signal output.
Button " ▲▼
2)After Display text is selected to " yES " or " no ",
press the " Enter Button " ( 3-2, Fig. 1 ) will save the
setting function with default.
Page 16
8. POWER SUPPLY from DC
1)When the left corner of LCD display show " ", it
ADAPTER
The meter also can supply the power supply from the
DC 9V Power Adapter ( optional ). Insert the plug of
Power Adapter into " DC 9V Power Adapter Input Socket
" ( 3-14, Fig. 1 ).
9. BATTERY REPLACEMENT
is necessary to replace the battery. However, in-spec.
measurement may still be made for several hours after
low battery indicator appears before the instrument
become inaccurate.
Loose the " Screw of the battery cover " ( 3-10, Fig. 1 )
2)
and take away the " Battery Cover " ( 3-9, Fig. 1 ) from
the instrument and remove the battery.
3)Replace with DC 1.5 V battery ( UM4/AAA,
Alkaline/heavy duty ) x 6 PCs, and reinstate the cover.
Make sure the battery cover is secured after changing
4)
the battery.
10. SYSTEM RESET
If the meter happen the troubles such as :
CPU system is hold ( for example, the key button can
not be operated... ).
Then make the system RESET will fix the problem.
The system RESET procedures will be either following
method :
During the power on, use a pin to press the " Reset
Button " ( 3-11, Fig. 1 ) once a while will reset the
circuit system.
Page 17
11. RS232 PC SERIAL INTERFACE
The instrument has RS232 PC serial interface via a 3.5 mm
terminal ( 3-12, Fig. 1 ) if the RS232 function already select
to " ON ", refer to chapter 7-7, page 14.
The data output is a 16 digit stream which can be
utilized for user's specific application.
A RS232 lead with the following connection will be
required to link the instrument with the PC serial port.
MeterPC
(9W 'D" Connector)
Center Pin..........................Pin 4
(3.5 mm jack plug)
Ground/shield.......................Pin 2
2.2 K
resistor
Pin 5
The 16 digits data stream will be displayed in the
following format :