User manuals in various languages (français,
italiano, español, português, nederlands, türk, polski,
русский, 中文) can be found by using our
product search on: www.pce-instruments.com
Last change: 21 October 2019
v1.0

Contents
1 Safety notes ........................................................................................... 1
2 Delivery contents .................................................................................. 2
3 Technical specifications ...................................................................... 2
4 Device description ................................................................................ 3
5 Display description ............................................................................... 4
6 On/off ...................................................................................................... 5
7 Automatic Power Off............................................................................. 5
8 Measurement ......................................................................................... 5
9 Calibration ............................................................................................. 5
10 New calibration standard ..................................................................... 5
11 Saving/viewing/deleting measured data ............................................. 6
12 Recharging the battery ......................................................................... 6
13 What to bear in mind before and during a measurement ................. 6
14 Contact ................................................................................................. 7
15 Disposal ................................................................................................. 7
© PCE Instruments

© PCE Instruments
1
1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The
device may only be used by qualified personnel and repaired by PCE Instruments personnel.
Damage or injuries caused by non-observance of the manual are excluded from our liability and
not covered by our warranty.
• The device must only be used as described in this instruction manual. If used otherwise,
this can cause dangerous situations for the user and damage to the meter.
• The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
• Do not expose the device to shocks or strong vibrations.
• The case should only be opened by qualified PCE Instruments personnel.
• Never use the instrument when your hands are wet.
• You must not make any technical changes to the device.
• The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
• The device must only be used with accessories from PCE Instruments or equivalent.
• Before each use, inspect the case for visible damage. If any damage is visible, do not
use the device.
• Do not use the instrument in explosive atmospheres.
• The measurement range as stated in the specifications must not be exceeded under
any circumstances.
• Non-observance of the safety notes can cause damage to the device and injuries to
the user.
We do not assume liability for printing errors or any other mistakes in this manual.
We expressly point to our general guarantee terms which can be found in our general terms of
business.
If you have any questions please contact PCE Instruments. The contact details can be found at
the end of this manual.

© PCE Instruments
2
2 Delivery contents
1 x gloss meter PCE-GM 55
1 x reference standard
1 x cleaning cloth
1 x carrying case
1 x user manual
3 Technical specifications
± 1 gloss units (on reference standard)
± 0.5 gloss units (0 ... 99.9 GE)
0 ... 40°C / 32 ... 104°F, max. 85% rh
140 x 45 x 75 mm / 5.5 x 1.8 x 3 in

ANGLE key to set measuring angle (ANGLE)
CAL key for calibration and to decrease parameters
MEAS key to make a measurement
DEL key to delete reading
RD key to view memory and increase parameter
USB interface via 2.5 mm jack

© PCE Instruments
4
5 Display description
Indicates that you are in calibration mode
Indicates that battery must be recharged
Indicates that a measurement is made

© PCE Instruments
5
6 On/off
To turn on the meter, press the on/off key. To turn off the meter, press and hold the on/off key
until “OFF“ is displayed.
7 Automatic Power Off
The gloss meter has an automatic power off function to save battery power. If no key is pressed
for 10 minutes when a single measurement is made, the meter will turn off automatically. When
a continuous measurement is made, the meter will power off automatically after an hour. This
function cannot be deactivated in measuring mode.
8 Measurement
The gloss meter hast wo measuring functions:
In single measuring mode, the user can press the MEAS key to make a measurement. will
be displayed during a measurement. The reading will be saved directly after a measurement.
In continuous measuring mode, the meter automatically makes continuous measurements
without requiring the user to make any entries. A measurement is made approx. every three
seconds. The readings are saved continuously.
To switch between these measuring functions, press and hold the on/off key until “SC” is
displayed after “OFF” was shown.
To change the measuring angle, press the ANGLE key.
9 Calibration
The gloss meter should be calibrated regularly. To calibrate the meter, place it onto the calibration
standard. Now make one measurement per angle and compare the readings to the values on the
reference.
If one reading is outside the parameters, you can automatically set the value for this angle by
pressing the CAL key while the meter is still on the reference.
10 New calibration standard
If you want to use a new calibration standard, this standard must first be added. To add the
standard, press and hold the on/off key (while the meter is on) until “CAL” appears on the display
(after ”OFF“ and ”SC“). Then press the CAL key. “CAL” will be displayed, followed by the angle
and the pre-set value. You can now change the reference value with the arrow keys. With the
ANGLE key, you can select the next measuring angle. After setting the reference values for all
measuring angles, press the ANGLE key to confirm the settings and return to measuring mode.

© PCE Instruments
6
11 Saving/viewing/deleting measured data
This gloss meter has two functions which can be seen on the display.
The meter is in measuring mode
The meter is in memory readout mode
In measuring mode, each reding will be saved. A measurement is made as soon as the measuring
icon is displayed. A maximum of 254 values can be saved. When the memory is full, the
oldest memory item will be replaced.
To enter memory readout mode, press the RD key. The display will switch from “M“ to “R“. Use
the arrow keys to choose the saved reading you want to view.
With the DEL key, you can delete a single reading.
To return to measuring mode, press the MEAS key. The display will switch from “R“ to “M”.
To delete the complete memory at a time, press and hold the DEL key for three seconds in
measuring mode.
Note: If you try to enter memory readout mode when no measuring data are available, the display
will show “Err1“. Memory readout mode can only be opened if data have been saved.
12 Recharging the battery
To make sure you can still use the meter, it should be recharged when the battery icon
appears in the display. To charge the meter, use the charger that comes with it.
If the rechargeable battery needs to be replaced, please contact PCE Instruments.
13 What to bear in mind before and during a measurement
Before a measurement, make sure that no external light enters the measuring chamber as this
can affect the readings. When there are considerable temperature fluctuations, make sure the
meter becomes acclimatised to the new environmental conditions. Then re-calibrate the meter
before making a measurement. If the meter is used for a long period of time such as more than
an hour, regular calibrations must be made. Make sure to store the calibration reference safely
everytime it is removed.

14 Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact
us. You will find the relevant contact information at the end of this user manual.
15 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European
Parliament applies. Due to the contained pollutants, batteries must not be disposed of as
household waste. They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either reuse them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance
with your local waste regulations.
If you have any questions, please contact PCE Instruments.
© PCE Instruments
7

© PCE Instruments
8
PCE Instruments contact information
Germany France Spain
PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L.
Im Langel 4 23, rue de Strasbourg Calle Mayor, 53
D-59872 Meschede 67250 Soultz-Sous-Forets 02500 Tobarra (Albacete)
Deutschland France España
Tel.: +49 (0) 2903 976 99 0 Téléphone: +33 (0) 972 3537 17 Tel. : +34 967 543 548
Fax: +49 (0) 2903 976 99 29 Numéro de fax: +33 (0) 972 3537 18 Fax: +34 967 543 542
info@pce-instruments.com info@pce-france.fr info@pce-iberica.es
www.pce-instruments.com/deutsch www.pce-instruments.com/french www.pce-instruments.com/espanol
Germany United Kingdom Italy
PCE Produktions- und PCE Instruments UK Ltd PCE Italia s.r.l.
Entwicklungsgesellschaft mbH Unit 11 Southpoint Business Park Via Pesciatina 878 / B-Interno 6
Im Langel 26 Ensign Way, Southampton 55010 Loc. Gragnano
D-59872 Meschede Hampshire Capannori (Lucca)
Deutschland United Kingdom, SO31 4RF Italia
Tel.: +49 (0) 2903 976 99 471 Tel: +44 (0) 2380 98703 0 Telefono: +39 0583 975 114
Fax: +49 (0) 2903 976 99 9971 Fax: +44 (0) 2380 98703 9 Fax: +39 0583 974 824
info@pce-instruments.com info@pce-instruments.co.uk info@pce-italia.it
www.pce-instruments.com/deutsch www.pce-instruments.com/english www.pce-instruments.com/italiano
The Netherlands China Hong Kong
PCE Brookhuis B.V. PCE (Beijing) Technology Co., Limited PCE Instruments HK Ltd.
Institutenweg 15 1519 Room, 6 Building Unit J, 21/F., COS Centre
7521 PH Enschede Zhong Ang Times Plaza 56 Tsun Yip Street
Nederland No. 9 Mentougou Road, Tou Gou District Kwun Tong
Telefoon: +31 (0)53 737 01 92 102300 Beijing, China Kowloon, Hong Kong
info@pcebenelux.nl Tel: +86 (10) 8893 9660 Tel: +852-301-84912
www.pce-instruments.com/dutch info@pce-instruments.cn jyi@pce-instruments.com
www.pce-instruments.cn www.pce-instruments.cn
United States of America Turkey
PCE Americas Inc. PCE Teknik Cihazları Ltd.Şti.
711 Commerce Way suite 8 Halkalı Merkez Mah.
Jupiter / Palm Beach Pehlivan Sok. No.6/C
33458 FL 34303 Küçükçekmece - İstanbul
USA Türkiye
Tel: +1 (561) 320-9162 Tel: 0212 471 11 47
Fax: +1 (561) 320-9176 Faks: 0212 705 53 93
info@pce-americas.com info@pce-cihazlari.com.tr
www.pce-instruments.com/us www.pce-instruments.com/turkish