PCE-EM 890 Environmental Tester
User Manual
Last change: 12 January 2021
v1.0
English
Page 2
Contents
1 Safety notes ........................................................................................... 1
2 Specifications ........................................................................................ 2
3 Delivery scope ....................................................................................... 3
4 System description ............................................................................... 4
5 Measurement functions ........................................................................ 5
6 Operation ............................................................................................... 5
6.1 Preparation ...................................................................................................................... 5
6.2 Functions ......................................................................................................................... 5
6.3 Measurement ................................................................................................................... 7
7 Contact ............................................................................................... 10
8 Disposal ............................................................................................... 10
© PCE Instruments
Page 3
© PCE Instruments
1
1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.
The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
Do not expose the device to shocks or strong vibrations.
The case should only be opened by qualified PCE Instruments personnel.
Never use the instrument when your hands are wet.
You must not make any technical changes to the device.
The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
The device must only be used with accessories from PCE Instruments or equivalent.
Before each use, inspect the case for visible damage. If any damage is visible, do not
use the device.
Do not use the instrument in explosive atmospheres.
The measurement range as stated in the specifications must not be exceeded under
any circumstances.
Non-observance of the safety notes can cause damage to the device and injuries to the user.
We do not assume liability for printing errors or any other mistakes in this manual. We expressly point to our general guarantee terms which can be found in our general terms of
business. If you have any questions please contact PCE Instruments. The contact details can be found at
the end of this manual.
Page 4
© PCE Instruments
2
2 Specifications
Measuring function
Measurement range
Resolution
Accuracy
Wind speed
80 ... 3937 ft/min
1 ft/min
±3 % of range
0.4 ... 20 m/s
0.1 m/s
1.4 ... 72 km/h
0.1 km/h
0.9 ... 44.7 mph
0.1 mph
0.8 ... 38.8 knots
0.1 knots
Air temperature
0 ... 50 °C
0.1 °C
±1.2 °C
Humidity
10 ... 95 % RH
0.1 % RH <70 % RH: ±4 % RH
>70 % RH: ±4 % of rdg. + 1.2 % RH
Barometric air pressure 10 ... 999.9 hPa
0.1 hPa
±1.5 hPa
1000 ... 1100 hPa
1 hPa
±2 hPa
UV radiation
0 ....1999 µW/m²
0.1 µW/m²
±(4% of range + 2 digits) 290 ... 390 nm
2 ... 20 mW/m²
0.01 mW/m²
External temperature (Pt1000 input)
-10 ... 70 °C
0.1 °C
±1.2 °C
Volume flow
0.024 ... 3600 cmm
0.001 cmm
0.01 cmm
0.1 cmm
1 cmm
Dew point
-25.3 ... 49 °C
0.1 °C
Wet bulb temperature
-5.4 ... 49 °C
0.1 °C
Heat stress index
0 ... 100 °C
0.1 °C
±2 °C
Wind chill
-9.4 ... 44.2 °C
0.1 °C
±2 °C
Altitude above sea level
-2000 ... 9000 m
1 m
±15 m
Page 5
© PCE Instruments
3
General technical data of the climate/weather measuring device
Display
LC display 8 mm digit height
Measurement functions
Wind speed / temperature
Air humidity / temperature
UV
Air pressure
Volume flow
Dew point
Wet bulb temperature
Wind chill
Heat stress index
Altitude above sea level
External temperature via Pt1000
Power supply
CR 2032 3 V battery
Power consumption
approx. 5 mA
Operating conditions
0 ... 50 °C / max. 80 % RH
Dimensions
120 x 45 x 20 mm
Weight
approx. 160 g
3 Delivery scope
1 x environmental tester PCE-EM 890 1 x CR2032 battery 1 x protective bag 1 x user manual
Page 6
© PCE Instruments
4
4 System description
Vane
REC key (recording)
LC display
Hand strap
Hold key (freeze display)
Pt1000 connection for external sensor
UV light sensor
Brief description of key function
Temperature / humidity sensor
Pressure sensor barometer
Power key (On)
Battery compartment
Page 7
© PCE Instruments
5
5 Measurement functions
Measuring function
Display text
Unit of measurement
Wind / air speed
AN m/s, km/h, mph, Knot, FPM
Temperature
°C, °F
Volume flow
AirFL
CMM, CFM
Wind chill
CHiLL
°C, °F
Humidity
rH % r. F.
Temperature
°C, °F
Dew point
dP
°C, °F
Wet bulb temperature
_Et
°C, °F
Heat index
HEAt
°C, °F
UV radiation (UVA)
U A
mW/cm², µW/cm², automatic
Air pressure
bAr
hPa (HPA), mmHg (mH9), inHg (inH9)
Altitude above sea level
HigH
m, Ft
External temperature via Pt1000
Pt
°C, °F
The units of measurement in brackets are the display representations that differ from the unit of measurement.
6 Operation
6.1 Preparation
Open the battery compartment on the back of the device by turning it slightly to the left and insert the CR 2032 3 V battery, observing correct polarity. Close the battery compartment.
6.2 Functions
6.2.1 Switch meter on / off
Press the "Power" key to switch on the meter.
To switch off the meter, press and hold the "Power" key until it switches off.
6.2.2 Display backlight
Press and release the "Power" key when the meter is switched on.
The display backlight switches on for 5 seconds.
6.2.3 Select measuring function
Press and hold the "Hold" key when the meter is switched on.
The display shows the measuring functions one after the other.
Release the key as soon as the desired measuring function is displayed.
Page 8
© PCE Instruments
6
6.2.4 Select unit of measurement
When the meter is switched on, press and hold the "REC" key until the display shows "Unit".
Press the "Power" key repeatedly until the desired unit of measurement appears on the display.
Confirm the selection with the "REC" key.
If you have selected a measurement function with dual measurement, you can now
change the unit of measurement for the second measurement function as described above.
6.2.5 Hold function
Press and release the "Hold" key to freeze the current reading on the display.
Press and release the "Hold" key again to return to normal measuring mode.
The hold function is not available when the memory function is activated.
6.2.6 Rec. function
Briefly press the "REC" key to save the maximum and minimum readings of the current measurement.
To call up the max value, press and release the "REC" key.
To call up the min value, press and release the "REC" key again.
To exit the Rec function and delete the saved values, press and hold the "REC" key for
more than 2 seconds.
6.2.7 Automatic power off
The tester powers off after approx. 10 minutes of inactivity to save battery power. Automatic power off is not active in "REC" mode.
Page 9
© PCE Instruments
7
6.3 Measurement
6.3.1 Wind / air speed
Switch on the tester and select the wind / air speed measurement function as described above.
Select the desired units of measurement as described above.
Hold the meter in such a way that the vane is placed in wind direction.
Do not cover or enclose the openings of the temperature / humidity sensor. This will
lead to incorrect measurement results.
6.3.2 Volume flow
Switch on the meter and select the volume flow measurement function as described above.
Select the desired unit of measurement as described above.
Press the "Hold" key to set the opening cross-section.
When the hold indicator is displayed, press and hold the "REC" key until "m-2" or "F-2"
appears at the bottom of the display.
Use the "Hold" and "Power" keys to increase or decrease the cross-section value.
Press the "REC" key to confirm the entry.
Hold the meter in such a way that the vane is placed in front of the discharge opening
and read the value. To obtain a more accurate measurement result, measure at several points of the discharge opening and calculate the average of the measured values.
6.3.3 Wind chill
Switch on the tester and select the wind chill measurement function as described above.
Select the desired unit of measurement as described above.
Hold the meter in such a way that the vane is placed in wind direction.
Do not cover or enclose the openings of the temperature / humidity sensor. This will
lead to incorrect measurement results.
6.3.4 Air humidity / temperature
Switch on the tester and select the humidity / temperature measuring function as described above.
Select the desired units of measurement as described above.
Do not cover or enclose the openings of the temperature / humidity sensor. This will
lead to incorrect measurement results.
6.3.5 Dew point
Switch on the meter and select the dew point measurement function as described above.
Select the desired unit of measurement as described above.
Do not cover or enclose the openings of the temperature / humidity sensor. This will
lead to incorrect measurement results.
Page 10
© PCE Instruments
8
6.3.6 Wet bulb temperature
Power on the meter and select the wet bulb temperature measurement function as described above.
Select the desired unit of measurement as described above.
Do not cover or enclose the openings of the temperature / humidity sensor. This will
lead to incorrect measurement results.
6.3.7 Heat index
Power on the meter and select the heat index measurement function as described above.
Select the desired unit of measurement as described above.
Do not cover or enclose the openings of the temperature / humidity sensor. This will
lead to incorrect measurement results.
6.3.8 UV radiation
Power on the meter and select the UV radiation measurement function as described above.
Do not cover or enclose the openings of the UV light sensor. This will lead to incorrect measurement results.
The changeover from µW/cm² to mW/cm² is automatic.
6.3.9 Air pressure
Power on the meter and select the air pressure measuring function as described above.
Select the desired unit of measurement as described above.
Do not cover or enclose the opening of the pressure sensor on the back of the unit.
This will lead to incorrect measurement results.
Attention! The actual air pressure depends on the weather-related atmospheric air pressure and the altitude above sea level. The unit is set to sea level when delivered. To adjust the altitude above sea level to your location, proceed as follows:
With the air pressure measuring function selected, press the "Hold" key.
When the hold indicator is displayed, press and hold the "REC" key until "m" or "F"
appears at the bottom of the display.
With the "Hold" and "Power" keys, you can increase or decrease the site altitude above sea level.
Press the "REC" key to confirm the entry.
Page 11
© PCE Instruments
9
6.3.10 Altitude above sea level
Power on the meter and select the measurement function for the altitude above sea level as described above.
Select the desired unit of measurement as described above.
Do not cover or enclose the opening of the pressure sensor on the back of the unit.
This will lead to incorrect measurement results.
Attention! The altitude measurement depends on the air pressure. The higher your location is above sea level, the lower the air pressure. For an accurate altitude measurement, it is therefore necessary to set the device accordingly. For an accurate setting, it is necessary to know the current altitude above sea level or the current air pressure.
6.3.10.1 Setting via "Altitude above sea level
With the altitude above sea level measurement function selected, press the "Hold" key.
When the hold indicator is displayed, press and release the "REC" key. m" or "F"
appears at the bottom of the display.
With the "Hold" and "Power" keys, you can increase or decrease the current site altitude above sea level.
Press the "REC" key to confirm the entry.
6.3.10.2 Setting via "Air pressure
With the altitude above sea level measurement function selected, press the "Hold" key.
When the hold indicator is displayed, press and hold the "REC" key until "HPA",
"mmHg" or "inHg" appears at the bottom of the display.
Use the "Hold" and "Power" keys to increase or decrease the current air pressure value.
Press the "REC" key to confirm the entry.
6.3.11 External temperature via Pt1000
Connect the external Pt1000 sensor via the jack socket on the side of the unit.
Power on the meter and select the measuring function external temperature via Pt1000,
as described above.
Select the desired unit of measurement as described above.
The Pt1000 connection is designed as a 2-pin 3.5 mm jack socket.
Page 12
7 Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact
us. You will find the relevant contact information at the end of this user manual.
8 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations.
If you have any questions, please contact PCE Instruments.
© PCE Instruments
10
Page 13
© PCE Instruments
11
PCE Instruments contact information
Germany France Spain
PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L. Im Langel 4 23, rue de Strasbourg Calle Mayor, 53 D-59872 Meschede 67250 Soultz-Sous-Forets 02500 Tobarra (Albacete) Deutschland France España Tel.: +49 (0) 2903 976 99 0 Téléphone: +33 (0) 972 3537 17 Tel. : +34 967 543 548 Fax: +49 (0) 2903 976 99 29 Numéro de fax: +33 (0) 972 3537 18 Fax: +34 967 543 542 info@pce-instruments.com info@pce-france.fr info@pce-iberica.es www.pce-instruments.com/deutsch www.pce-instruments.com/french www.pce-instruments.com/espanol
United Kingdom Italy Turkey
PCE Instruments UK Ltd PCE Italia s.r.l. PCE Teknik Cihazları Ltd.Şti. Unit 11 Southpoint Business Park Via Pesciatina 878 / B-Interno 6 Halkalı Merkez Mah. Ensign Way, Southampton 55010 Loc. Gragnano Pehlivan Sok. No.6/C Hampshire Capannori (Lucca) 34303 Küçükçekmece - İstanbul United Kingdom, SO31 4RF Italia Türkiye Tel: +44 (0) 2380 98703 0 Telefono: +39 0583 975 114 Tel: 0212 471 11 47 Fax: +44 (0) 2380 98703 9 Fax: +39 0583 974 824 Faks: 0212 705 53 93 info@pce-instruments.co.uk info@pce-italia.it info@pce-cihazlari.com.tr www.pce-instruments.com/english www.pce-instruments.com/italiano www.pce-instruments.com/turkish
The Netherlands China Hong Kong
PCE Brookhuis B.V. PCE (Beijing) Technology Co., Limited PCE Instruments HK Ltd. Institutenweg 15 1519 Room, 6 Building Unit J, 21/F., COS Centre 7521 PH Enschede Zhong Ang Times Plaza 56 Tsun Yip Street Nederland No. 9 Mentougou Road, Tou Gou District Kwun Tong Telefoon: +31 (0)53 737 01 92 102300 Beijing, China Kowloon, Hong Kong info@pcebenelux.nl Tel: +86 (10) 8893 9660 Tel: +852-301-84912 www.pce-instruments.com/dutch info@pce-instruments.cn jyi@pce-instruments.com www.pce-instruments.cn www.pce-instruments.cn
United States of America
PCE Americas Inc. 1201 Jupiter Park Drive, Suite 8 Jupiter / Palm Beach 33458 FL USA Tel: +1 (561) 320-9162 Fax: +1 (561) 320-9176 info@pce-americas.com www.pce-instruments.com/us
Loading...