Precision Balance PCE-BSK Series
User Manual
Last change: 18 December 2019
v1.0
English
Page 2
Contents
1 Safety notes ........................................................................................... 1
2 Technical specifications ...................................................................... 2
3 Delivery contents .................................................................................. 2
4 Device description ................................................................................ 3
5 First use of the balance ........................................................................ 4
6 Calibration ............................................................................................. 4
6.1 One-point calibration ........................................................................................................ 4
6.2 Three-point calibration ...................................................................................................... 5
7 Function settings .................................................................................. 5
7.1 Setting the weight unit ...................................................................................................... 5
7.2 Counting function ............................................................................................................. 6
7.3 User-specific functions ..................................................................................................... 6
8 Serial interface - communication protocol ......................................... 7
9 Troubleshooting .................................................................................... 7
10 Contact ................................................................................................. 8
11 Disposal ................................................................................................. 8
© PCE Instruments
Page 3
© PCE Instruments
1
1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.
The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
Do not expose the device to shocks or strong vibrations.
The case should only be opened by qualified PCE Instruments personnel.
Never use the instrument when your hands are wet.
You must not make any technical changes to the device.
The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
The device must only be used with accessories from PCE Instruments or equivalent.
When weighing powder and similar products, use a suitable container to make sure no
small parts enter the scales.
Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
Do not use the instrument in explosive atmospheres.
Do not expose the device to radioactive environments or corrosive substances.
The measurement range as stated in the specifications must not be exceeded under
any circumstances.
Non-observance of the safety notes can cause damage to the device and injuries to the user.
Do not use any sharp objects such as pens to press the keys of the balance. Only use your fingers to press the keys.
We do not assume liability for printing errors or any other mistakes in this manual. We expressly point to our general guarantee terms which can be found in our general terms of
business. If you have any questions please contact PCE Instruments. The contact details can be found at
the end of this manual.
Page 4
© PCE Instruments
2
2 Technical specifications
PCE-BSK 310
PCE-BSK 1100
PCE-BSK 5100
Weighing range
310g
1,100g
5,100g
Accuracy class
II
III
III
Resolution/readability
0.001g
0.01g
0.1g
Repeatability
±0.002g
±0.02g
±0.2g
Linearity
±0.003g
±0.03g
±0.3g
Settling time
<3 s
Operating temperature
17.5 ... 22.5 °C
Dimensions of weighing plate
90 mm
130 mm
160x160 mm
Calibration
external
Overall dimensions
270 x 265 x 190 mm
270 x 200 x 80 mm
Overall weight
4.0 kg
2.5 kg
Interface
RS232
Mains adaptor
Primary: 100 ... 240 V, 50 / 60 Hz
Secondary: 6 V, 500 mA
outside minus, inside plus
Batteries
3 x AA (LR6)
not included, not rechargeable within the balance
3 Delivery contents
PCE-BSK 310
PCE-BSK 1100
PCE-BSK 5100
Electronic balance
Weighing plate
90 mm
130 mm
160 x 160 mm
Mains adaptor • •
User manual • •
Test weight
200 g
500 g
---
Wind shield • ---
---
only PCE-BSK 310 includes wind shield
Page 5
© PCE Instruments
3
4 Device description
1 2
3
4
9 10 11
7
8
5 6
12 13 14 15 16
1. Weighing plate
2. Type plate
3. Adjustable feet
4. Level for alignment
5. RS232-Schnittstelle
6. Connection for mains adaptor
7. Stable weight indicator
8. Power supply lamp
9. Weight display
10. Battery level indicator
11. Current weight unit
12. ON/OFF key
13. Counting key
14. “>0<“ key, e. g. for zeroing
15. “CAL“ key, e. g. for calibration
16. Unit key for changing weight unit
Page 6
© PCE Instruments
4
5 First use of the balance
Open the packaging carefully and keep it for possible later returns. Place the weighing plate on the scales and connect the mains adaptor. If you prefer battery operation, insert the batteries. (If the weighing plate has not been placed on the balance correctly, a notification will be displayed to inform you that weighing is not possible.) The balance will start after a short beeping sound
and will show 300.00 or 1000.00 or 5000 respectively, followed by a dashed line. With a second short beeping sound, it will then display 0.00 kg. The balance is now in weighing mode.
Set-up location
1. Keep the set-up location clean and dry.
2. Place the balance on an even and stable surface.
3. Align the balance by means of the adjustable feet. The bubble of the level must be exactly within the circle marked in black.
4. Avoid draught and fluctuations in temperature, e. g. coming from open windows and doors as well as placing the scales below heaters or exposed to direct sunlight.
5. Avoid exposure to vibration.
6. Avoid interference from other devices.
7. Please note that static charges can occur when the scales are used in dry environments or for bulk materials.
8. Keep the balance away from objects with magnets and from magnetic fields.
9. Do not use the scales in explosive areas.
10. Do not use the scales in areas with high air humidity or in dusty environments.
11. When the balance is carried from a cold to a warm environment, moisture within the balance can condense. This has an effect on the accuracy and the reliability of the weighing. To avoid this, allow 2 hours for the balance to take on room temperature before switching it on.
6 Calibration
The weighing accuracy of the balance depends, among other factors, on the environmental conditions (gravitation, temperature, humidity). With the calibration function, the scales can be adapted to the current environmental conditions. The balance must have been exposed to stable environmental conditions for at least 30 minutes before the calibration can be started.
6.1 One-point calibration
In weighing mode, press
- CAL for approx. 4 seconds – “I.CAL Ig“ should be displayed.
- Press the >0< key – a weight value for the calibration should be displayed.
- Press >0< to apply or press the counting key as well as the CAL key to make a change – then confirm with >0<.
- Apply the weight that flashes in the display. A line will appear, followed by the non-flashing weight.
- Remove the weight – a line will be displayed, followed by 0.00 g“. The calibration is now completed.
Page 7
© PCE Instruments
5
6.2 Three-point calibration
In weighing mode, press
- CAL for approx. 4 seconds – “I.CAL Ig“ should be displayed.
- Press the counting key to select “2.CAL 3“.
- Press >0< - a weight value for the calibration should be displayed.
- Apply the weight that flashes in the display. A line will appear, followed by the non-flashing weight.
- Remove the first weight.
- Apply the next weight which flashes in the display. A line will appear, followed by the non-flashing weight.
- Remove the second weight.
- Apply the next weight which flashes in the display. A line will appear, followed by the non-flashing weight.
- Remove the third weight - a line will be displayed, followed by “0.00 g“. The calibration is now completed.
For the PCE-BSK 1100, 200 g, 500 g and 1000 g weights are required.
7 Function settings
7.1 Setting the weight unit
Go to F01 to define your own custom unit with decimal places and conversion factor related to grams. Go to F02 to deactivate or activate the pre-set units.
To enter the menu, first turn off the scales and then: a) press and hold the unit key and press and release the ON/OFF key or b) short-press the ON/OFF key and then the unit key.
F-01“ Decimal character and factor for custom unit
- Press >0< to open the function.
-
- - - - - - -
g“ will be displayed.
- Use the counting key to change the decimal position.
- Confirm with >0<.
- “00000.00g“ will be displayed for example, with the selected decimal places.
- Enter the conversion factor from grams to the custom unit with the counting key (for the position) and the CAL key (for the value) The customer-specific unit will be saved so that it is still available after turning off the scales.
- Confirm with >0<. F-02“ will be displayed.
- Enter weighing mode with the unit key.
F-02“ Activation/deactivation of weight units
- Press >0< to open the function.
- Select the unit with the CAL key.
- Use the counting key to switch between ON and OFF.
- These units can be selected: g / CT / OZ / FRE / dWT / OZT / gn /G.S.M. (grams, carat, ounces, free unit (user-defined), pennyweight, troy ounce, grain, g/m²).
- Confirm with >0<. F-01“ will be displayed.
- Enter weighing mode with the unit key.
Page 8
© PCE Instruments
6
7.2 Counting function
To enable the scales to count pieces, the reference quantity must be saved first:
- In weighing mode, press and hold the counting key for approx. 4 seconds.
- ---.COU --- will be displayed and after some time, 5 PCS will appear in the display.
- The number of reference pieces can be selected with >0<. You can select 5, 10, 20, 40, 50,100, 200, 300, 400 or 500.
- Place the selected number of pieces on the scales and confirm with the counting key.
- A line will be displayed, followed by the number of pieces placed on the scales.
- Remove the reference pieces from the weighing plate. The display will now show 0 and you can use the counting function.
- To switch between the counting and weighing functions, short-press the counting key.
- The determined reference value will be saved so that it is still available after turning off the balance.
7.3 User-specific functions
In weighing mode, press the counting key and the unit key at the same time until I. FILis displayed. You can go to the next menu item by pressing the counting key. You can confirm your selection with the >0< key or return to the menu with the unit key.
Display
Function
Description
I. FIL
Sensitivity
Selection with counting key:
1.1 FIL 1- fast,
1.2 FIL 2 - normal,
1.3 FIL 3 - slow,
2. bA_LI
Backlight
Selection with counting key:
2.1 ON
2.2 OFF
2.3 AUTO (energy saving mode, backlight will turn off after 30 seconds)
3. AU_Po
Automatic power off
Selection with counting key:
3.1 ON (power off after 3 minutes)
3.2 OFF
4. PEAK
Maximum value / PEAK
Selection with counting key:
4.1 ON (last maximum weight will be displayed until a higher weight is added; a new maximum value will be determined after restarting the scales)
4.2 OFF
5 SENd
Data transfer to PC
5.1 KEY CAL key for transmission
5.2 ASK.....Query from PC
5.3 Cont.....Continuous transmission
5.4 StAb Automatic transmission when value is stable
Page 9
© PCE Instruments
7
8 Serial interface - communication protocol
1: Tare Format: ST + CR + LF 2: Entering tare value Format: ST***** ***** (i. e. from 9e to maximum weight)
The signal has 16 digits 1 icon + or – 2 space 3 ...10 digits for the weight value or space 11 space
12...14 display unit (Newton, kg, g, lb or PCS (pieces)) 15 CR 16 LF
3: Query of current weight display Format: Sx + CR + LF 4: Turn off balance (not possible in standby mode – can only be turned back on via the ”ON“ key
of the balance.) Format: SO + CR + LF
9 Troubleshooting
Problem
Possible cause
Solution
No display indication
Mains plug not connected, mains adaptor defective, red light below Power is glowing but nothing is displayed after pressing ON/OFF
Connect mains cable, replace mains adaptor where required
Displayed value not stable
Environmental conditions, wind shield open, foreign object on/within the balance, weight exceeds weighing range, change in weighed product (evaporation, absorption, movement), power supply not stable
Avoid air flow and vibration, close wind shield, remove foreign objects, ensure stable power supply by mains adaptor Displayed value different from actual value
No calibration, no taring, zero point not adapted
Calibrate, tare, adapt the zero point by pressing >0< before weighing
If your problem cannot be solved with the help of the chart above, please contact our technical support.
Page 10
10 Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact
us. You will find the relevant contact information at the end of this user manual.
11 Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re­use them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations.
If you have any questions, please contact PCE Instruments.
© PCE Instruments
8
Page 11
© PCE Instruments
9
PCE Instruments contact information
Germany France Spain
PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L. Im Langel 4 23, rue de Strasbourg Calle Mayor, 53 D-59872 Meschede 67250 Soultz-Sous-Forets 02500 Tobarra (Albacete) Deutschland France España Tel.: +49 (0) 2903 976 99 0 Téléphone: +33 (0) 972 3537 17 Tel. : +34 967 543 548 Fax: +49 (0) 2903 976 99 29 Numéro de fax: +33 (0) 972 3537 18 Fax: +34 967 543 542 info@pce-instruments.com info@pce-france.fr info@pce-iberica.es www.pce-instruments.com/deutsch www.pce-instruments.com/french www.pce-instruments.com/espanol
Germany United Kingdom Italy
PCE Produktions- und PCE Instruments UK Ltd PCE Italia s.r.l. Entwicklungsgesellschaft mbH Unit 11 Southpoint Business Park Via Pesciatina 878 / B-Interno 6 Im Langel 26 Ensign Way, Southampton 55010 Loc. Gragnano D-59872 Meschede Hampshire Capannori (Lucca) Deutschland United Kingdom, SO31 4RF Italia Tel.: +49 (0) 2903 976 99 471 Tel: +44 (0) 2380 98703 0 Telefono: +39 0583 975 114 Fax: +49 (0) 2903 976 99 9971 Fax: +44 (0) 2380 98703 9 Fax: +39 0583 974 824 info@pce-instruments.com info@pce-instruments.co.uk info@pce-italia.it www.pce-instruments.com/deutsch www.pce-instruments.com/english www.pce-instruments.com/italiano
The Netherlands China Hong Kong
PCE Brookhuis B.V. PCE (Beijing) Technology Co., Limited PCE Instruments HK Ltd. Institutenweg 15 1519 Room, 6 Building Unit J, 21/F., COS Centre 7521 PH Enschede Zhong Ang Times Plaza 56 Tsun Yip Street Nederland No. 9 Mentougou Road, Tou Gou District Kwun Tong Telefoon: +31 (0)53 737 01 92 102300 Beijing, China Kowloon, Hong Kong info@pcebenelux.nl Tel: +86 (10) 8893 9660 Tel: +852-301-84912 www.pce-instruments.com/dutch info@pce-instruments.cn jyi@pce-instruments.com www.pce-instruments.cn www.pce-instruments.cn
United States of America Turkey
PCE Americas Inc. PCE Teknik Cihazları Ltd.Şti. 711 Commerce Way suite 8 Halkalı Merkez Mah. Jupiter / Palm Beach Pehlivan Sok. No.6/C 33458 FL 34303 Küçükçekmece - İstanbul USA Türkiye Tel: +1 (561) 320-9162 Tel: 0212 471 11 47 Fax: +1 (561) 320-9176 Faks: 0212 705 53 93 info@pce-americas.com info@pce-cihazlari.com.tr www.pce-instruments.com/us www.pce-instruments.com/turkish
Loading...