PCE Instruments PCE-ATP 1, PCE-ATP 1-KIT Users guide

Page 1
PCE Instruments UK Ltd.
Jupiter FL-33458 USA From outside US: +1 Tel: (561) 320-9162 Fax: (561) 320-9176 info@pce­americas.com
Ensign way
Hampshire / Southampton United Kingdom, SO31 4RF
From outside UK: +44
Tel: (0) 2380 98703 0
Fax: (0) 2380 98703 9
info@industrial-needs.com
PCE Americas Inc.
711
Commerce Way
Suite 8
Units 12/13
Southpoint Business
Park
www.pce-instruments.com/english www.pce-
instruments.com
Manual
Food/HygieneTester
PCE-ATP 1
Page 2
II
Disclaimer and Copyright
All contents of this manual (including but not limited to text, trademark, logo, button icons, graphics, tables, data, etc.) shall not be reproduced in any form, neither print nor electronic, nor translated into any other languages.
This manual is subject to change without prior notice. The updated version prevails.
Page 3
III
Special Declaration
Please read this manual carefully before the usage of portable ATP hygiene monitoring system!
Symbols Used in This Manual
Symbols used in this manual that directly refers to the safe and proper operations of portable ATP hygiene monitoring system.
Icon
Description
Warning
Warning operation personnel pay attention to a certain operation. Operating the portable ATP hygiene monitoring system in any manner unspecified in this manual may
result in device damage or abnormal function.
Reminding
Providing important information that critical to the success of the operation or use of the device, including the information explained in further detail elsewhere in this manual.
Prohibit
Prohibiting operation personnel from a certain dangerous operation. Otherwise may result in device damage or abnormal function.
Conventions Used in This Manual
Convention
Meaning
Page 4
IV
Ordered list
Procedure steps must be performed follow the list order.
Click/Double
click A
Click or choose A on the ATP PC software.
Press B
Press B key on the keyboard of the detector.
italic + Bold
Refers to the key/button or operation instructions of the ATP PC software.
< Bold >
Refers to the key/button or options of the detector.
[ ]
Refers to keys on the computer keyboard.
Safety and Regulatory Compliance
General Precautions:
1. The portable PCE-ATP1 hygiene monitoring system is safely and
reliably designed and manufactured. Personal injury can be avoided if the product is operated properly (as instructed in the manual) and the related precautions are strictly followed.
2. Users shall be aware of hazards that may be caused by the portable
PCE-ATP1 hygiene monitoring system and its accessories.
3. All operators shall be familiar with the safety precautions and
warnings described in this section before the usage of portable PCE-
ATP1 hygiene monitoring system.
4. Non-observance of the instruction provided or performing any
operations not stated in the manual may affect the safety protection provided by the portable PCE-ATP1 hygiene monitoring system.
Page 5
V
Operation Environmental Precautions
Icon
Meaning
Prohibit
Never run the portable PCE-ATP1 hygiene monitoring system in
places that have or may have flammable and explosive gas.
Warning
in an extreme temperature environment. Do not place the portable PCE-ATP1 hygiene monitoring system
ATP Quickswab Precautions
Icon Meaning
Warning
Please carefully read the related information and precautions of PCE-ATP1 Quickswab before the usage of detector.
Reminding
Please follow national and local environment protection regulations and laws for the usage of PCE-ATP1 Quickswab.
Warning
Do not insert any other articles into the detector except the PCE-ATP1 Quickswab, and do not squeeze the PCE-ATP1 Quickswab while inserting.
Warning
Page 6
VI
Please make sure the outside surface of PCE-ATP1 Quickswab is clean and dry before inserting it into the detector.
Battery Precautions
Icon
Meaning
Warning
Please use the supplied rechargeable lithium-ion battery and
adaptor. The recharging voltage is limited to 4.2 V.
Keyboard and Keys Precautions
Icon
Meaning
Warning
Do not continuously press the keys on the keyboard of detector.
USB Interface Precautions
Icon
Meaning
Warning
The computer that connected to the USB interface of the detector shall comply with BSEN60950/IEC950 standard.
W
arning
Please discard the used battery according to local
regulations
Page 7
VII
Device Parts Precautions
Icon
Meaning
Warning
The detector is not provided with spare parts. Do not remove any
part from the device without permission.
Use Restrictions
The detector is designed to following universality, safety and EMC requirements:
Universality
Low voltage indicator 73/23/EEC EMC indicator 89/336/EEC
Safety
IEC 61010-1:2010 IEC 61326-1:2013 IEC 61326-2-6:2013 IEC 61010-2-081:2015 IEC 61010-2-101:2015
Declaration
The design of detector conforms to and able by the requirements in section 11 of low voltage specification 73/23/EEC. The design of detector also accords with the regulation that the design of electronic products should be work under specified voltage regulation and the requirements of BS EN 61010-1: 2001.
Page 8
VIII
Content
1. Overview ................................................................................................................. 1
1.1 Applications and Features ....................................................................................... 1
1.1.1 Scope of Application ......................................................................................... 1
1.1.2 Features ........................................................................................................... 3
1.2 Technical Specifications .......................................................................................... 5
1.3 Terms and Abbreviations ........................................................................................ 7
1.4 Accessories and Consumables ................................................................................ 7
1.5 Working Principle .................................................................................................... 7
1.6 Quick Operation Instructions .................................................................................. 8
1.6.1 ATP Quickswab Structure Diagram .................................................................. 9
1.6.2 ATP Quickswab Operating Procedures ............................................................. 9
1.6.3 ATP Quickswab Storage ................................................................................. 12
1.6.4 ATP Quickswab Safety .................................................................................... 12
2. Basic Operation Instructions of Detector ............................................................... 13
2.1 Structural Diagrams of Detector ........................................................................... 13
2.2 Function of Keys .................................................................................................... 14
2.3 Battery Installation ................................................................................................ 15
2.4 Power-on Self-test ................................................................................................ 16
2.5 Internal Calibration ............................................................................................... 17
2.5.1 Power-on Calibration ..................................................................................... 17
2.5.2 Recalibration .................................................................................................. 19
2.5.3 Auto-calibration ............................................................................................. 19
2.6 Power-off .............................................................................................................. 20
2.7 Power Saving Mode On and Resuming ................................................................. 20
2.8 Low Battery Alert .................................................................................................. 21
2.9 Icons and Meanings .............................................................................................. 22
3. Setting and Operation ........................................................................................... 23
Page 9
3.1 Setting Interface .................................................................................................... 23
3.2 User ....................................................................................................................... 23
3.3 Program ................................................................................................................. 25
3.4 Plan ....................................................................................................................... 27
3.5 Records ................................................................................................................. 28
3.6 Statistics ................................................................................................................ 31
3.7 System Settings ..................................................................................................... 33
3.8 Template ............................................................................................................... 33
3.9 Help ....................................................................................................................... 34
3.10 About ................................................................................................................... 34
4. Test and Test Result............................................................................................... 35
4.1 To-Be-Tested Interface.......................................................................................... 35
4.2 Program Selection and Setting .............................................................................. 35
4.2.1 Selection and Setting of User-defined Program (Upper and Lower
Limit) ....................................................................................................................... 36
4.2.2 Plan Setting and Program Selection ............................................................... 37
4.2.3 Calling of Template Program ......................................................................... 38
4.3 Start Test ............................................................................................................... 39
4.4 Records Regarding Operations .............................................................................. 41
4.4.1 View Records .................................................................................................. 41
4.4.2 Print Records .................................................................................................. 41
4.4.3 Delete Records ............................................................................................... 42
5. Connect Detector to PC or Other Terminals ........................................................... 42
5.1 Connect Detector to PC ......................................................................................... 42
5.2 Disconnect Detector from the connected PC........................................................ 43
5.3 Connect Detector to Bluetooth Printer ................................................................. 43
6. Operation and Maintenance .................................................................................. 44
IX
Page 10
X
6.1 Daily Precautions................................................................................................... 44
6.2 Battery Charge or Replacement ............................................................................ 44
6.3 Test Chamber Clean or Replacement .................................................................... 45
7. Troubleshooting .................................................................................................... 45
8. Commitment Statements of After-sale Services .................................................... 50
8.1 Warranty Service ................................................................................................... 50
8.2 Response Time ...................................................................................................... 50
8.3 Spare and Accessory Parts .................................................................................... 51
8.4 Special Statement ................................................................................................. 51
9. Portable ATP Hygiene Monitoring System Software Specification ......................... 52
9.1 Overview ............................................................................................................... 52
9.2 Setup ..................................................................................................................... 53
9.3 Remove ................................................................................................................. 56
10. Software Interfaces and Features ......................................................................... 57
10.1 Software Interfaces ............................................................................................. 57
10.2 Overview of Features .......................................................................................... 57
10.2.1 Menu Bar...................................................................................................... 57
10.2.2 Tool Bar ........................................................................................................ 58
10.2.3 Function Options .......................................................................................... 59
11. Operating Guidance .............................................................................................. 59
11.1 Connect Device to PC .......................................................................................... 59
11.2 Connect Device to Software ................................................................................ 60
11.3 Record ................................................................................................................. 60
11.3.1 Description ................................................................................................... 60
11.3.2 Records Editing............................................................................................. 62
11.3.3 Records Deletion .......................................................................................... 63
11.3.4 Records Export ............................................................................................. 63
Page 11
XI
11.4 User ..................................................................................................................... 63
11.4.1 New User ...................................................................................................... 63
11.4.2 Edit User ....................................................................................................... 64
11.4.3 Delete User ................................................................................................... 64
11.5 Program ............................................................................................................... 64
11.5.1 New Program ............................................................................................... 65
11.5.2 Edit Program ................................................................................................ 65
11.5.3 Delete Program ............................................................................................ 65
11.6 Plan...................................................................................................................... 65
11.6.1 New Plan ...................................................................................................... 66
11.6.2 Edit Plan ....................................................................................................... 66
11.6.3 Delete Plan ................................................................................................... 67
11.7 Report ................................................................................................................. 67
11.8 Exchange ............................................................................................................. 69
11.8.1 Activation Method ....................................................................................... 69
11.8.2 Device Firmware Upgrade ............................................................................ 70
11.8.3 Synchronization ............................................................................................ 70
11.8.4 Command Send ............................................................................................ 71
Page 12
Page 13
1. Overview
The portable PCE-ATP1 hygiene monitoring system adopts the PCE­ATP1 inescent theory to realize simple hygiene monitoring in t o order help reaching HACCP and food hygienic standard.
The portable PCE-ATP1 hygiene monitoring system is consisted of two parts: ATP Quickswab and detector. This manual mainly introduces the operation, maintenance and troubleshooting of the detector in details. For details of ATP Quickswab, please refer to the user's manual of ATP Quickswab.
Warning: The detector belongs to highly sensitive measuring
device that shall be used with due care and protected from damp condition and impact.
1.1 Applications and Features
1.1.1 Scope of Application
The portable PCE-ATP1 hygiene monitoring system is intended for on-site quick cleanliness (microbial content) test in various industries such as food processing, catering, medical treatment, sanitation, daily chemicals, paper making, water treatment, environmental protection, water administration, entry-exit inspection and quarantine and other law-enforcing departments.
Examples are as follows :
1. Food processing industry
The portable PCE-ATP1 hygiene monitoring s ystem is capable of testing
bacteria, microorganism or food residue in
the production environment of food, beverage and catering industry. It is very suitable for HACCP system cleanliness test;
Cleanliness control during production; and the cleanliness test of food production line;
1
Page 14
➢ ➢ Disinfection evaluation of food packages;
Microorganism contents measurement of finished products and raw
materials;
Hygiene monitoring of the processing environment, capable of
detecting organic matter residue and thus prevent growing environment of microorganism.
2. Catering industry
Used by law-enforcing departments for hygienic security screening for catering services.
Cleanliness control of kitchens, dining tables, work bench and operating
tools;
Disinfection evaluation of tableware as well as the disinfection effect
evaluation of one-off disinfection table-wear;
Disinfection control of tableware used in airline, train and high-speed
rail;
Sanitary control of quality control department;
Quick cleanliness test of large dining places such as Olympic Games and
World Expo.
3. Healthcare Industry
Used by infection control departments to detect hospital hygiene and disinfection & sterilization conditions.
Object surface cleanliness detection of main department of the
hospital, such as disinfection center and ICU;
2
Page 15
Hand cleanliness inspection of medical staff;
Cleanliness and disinfection detection of medical apparatus and instruments, such as surgical instruments and endoscope;
Hospital environment cleanliness detection. Ensure the hospital is
entirely clean, safe and free of microbial contamination;
Disinfection effect evaluation of medial disinfection products.
Comparison of test result before and after disinfection to ensure disinfection products are high quality and reliable.
4. Environment Protection
Biological contamination assessment of water or waste water samples.
5. Other Industries
Daily chemicals manufacturing industry;
Quality supervision department;
Sanitary supervision of hotel and lodging industry; ➢ Port supervision.
1.1.2 Features
Small in size: handheld design, total weight is less than 300g, one-
hand operation;
Low power consumption: lithium battery, duration time is up to 10
hours, standby time is up to 600 hours;
3
Page 16
➢ ➢ Auto operating mode: 3.5’ color screen, simple keys, and friendly
HMI;
4
Page 17
➢ ➢
5
Quantified result: the test result is accurate to 1×10
-18
mol ATP;
Quick test: 10 seconds per sample. The detector can be connected to a thermal printer via Bluetooth. The test result can be printed in real time;
Test protection: the detector possesses built-in inclinometer which
will stop the test in case the incline angle is out of range to ensure test accuracy;
Control network: the detector can smartly check whether the
bacterial colony amount is out of limit. It can be connected to electronic terminals such as PC;
Maximum storage capacity: the detector is capable of storing 256
Users, 256 Plans, 2000 Programs and 10000 Records;
Data template: the detector is integrated with data templates to
facilitate user for viewing or calling;
Intelligent software: test data can be uploaded to the dedicated PC
software after test. Combine with this software, users could track, save the test results and analyze the trend of the tested locations.
Open reagent: suitable for ATP test reagents from multiple
manufacturers. It is strongly recommended to use the supplied consumables to ensure the test result accuracy.
1.2 Technical Specifications
Dimension: 189mm×70mm×35mm; Weight: 280g;
Display screen: 3.5’ color screen, graphic HMI;
Page 18
6
Start-up time: 15s and 60s available;
Test time: 10s;
Storage capacity: 256 Users, 256 Plans, 2000 Programs and 10000 Results;
Battery type: 3.7V rechargeable lithium battery, whose recharging
voltage is limited to 4.2 V;
Battery capacity: 2300mAh (duration time is up to 10 hours, standby
time is up to 600 hours);
Communication mode: the detector can be connected to your PC via USB
cable; and also can be connected to mobile, tablet PC or printer via Bluetooth;
Real-time print: the detector can be connected to a printer via Bluetooth,
and the test result can be printed in real time;
Incline angle detection: this feature of detector ensures the test
accuracy. The test will stop processing if the device incline angle is over 30°, and the screen will display icon as an alert;
Auto calibration: the detector possesses light source auto-calibration
system and temperature detection system that can automatically be adapted to changes of environment;
Test range: 0-999999RLUs; Test accuracy: 1×10
-18
mol ATP;
Test error: ±5% or ±5 RLUs; Test repeatability: 8%-20%; Correlation coefficient: ≥0.995;
Operating temperature range: 5-40; Operating humidity range: 20-80%;
Page 19
Storage temperature range: -10-40; Storage humidity range: ≤
60%.
1.3 Terms and Abbreviations
ATP: Adenosine tri-phosphate (energy transfer molecule)
RLU: Relative light unit (measuring unit)
USB: Communication port between the device and PC
1.4 Accessories and Consumables
Accessories include lithium battery, USB cable, USB charger, hang rope, Bluetooth printer (optional). The software can be downloaded
here: https://www.pce-instruments.com/english/download-win_4.htm
.
For details about other accessories and consumables, please ask local
dealer for assistance.
1.5 Working Principle
The portable PCE-ATP1 hygiene monitoring system adopts the inescence technique to convert invisible ATP concentration (ATP content in sample) into visible light output, the basic working principle is as shown in the Figure 1.1.
7
Page 20
Fig 1.1 Working principle
The portable PCE-ATP1 hygiene monitoring system takes the light energy as reference to output the test value and displays the test result in quantitative and qualitative form. The test result exhibits the cleanliness of the tested specimen, which should be between 0 and 999999 relative light units (RLU, 1 RLU = 1× 10
-18
mol of ATP). According to the user-deined
upper & lower limits, the detector will automatically offer the determination of the test result and displayed as < Pass > , < Fail > or < Caution > .
1.6 Quick Operation Instructions
The PCE-ATP1 Quickswab contains exclusive high-sensitivity liquid stable integrative reagent. It is capable of detecting bacteria or other microorganisms on object surfaces and total ATP activities within food residues in order to quickly provide cleanliness test results.
The PCE-ATP1 Quickswab should be used together with the portable ATP hygiene monitoring systems.
8
Page 21
1.6.1 ATP Quickswab Structure Diagram
The structure diagram of the ATP Quickswab is as shown in figure
1.2.
Fig 1.2 ATP Quickswab Structure Diagram
1. Cap 2. Spring cap (inside) 3. Joint port
4. Unplug point 5.Test tube 6. Swab tip.
1.6.2 ATP Quickswab Operating Procedures
The operating procedures of the ATP Quickswab are shown in the figure 1.3.
1. Unfreezing: Take out the ATP Quickswab from the refrigerator. Wait about 20 minutes until the internal reagent reaches the room temperature.
2. Sampling: Hold the joint port of the ATP Quickswab and unplug the cap at the unplug point, as shown in figure 1.2. Remove the test tube and pull out the pre-moistened swab tip. Keep 15 to 30 degrees angle of the swab tip and zigzag swab the sampling area, please remember to rotate the swab tip while swabbing to ensure the closely contact with the sampling area. (The sampling area should be around 10x10 cm2 and it could be marked within the sampling card).
3. Installation: After sampling, please hold the joint port of the ATP Quickswab and insert the swab tip back into the test tube. (The
9
Page 22
end face of the test tube should be aligned with the lower end face of the blue joint port).
4. Injection: Remove the cap of the ATP Quickswab, make sure it is griped in upright position, and forcibly press down the spring cap several times to the reagent fully flows into the bottom of the test tube and submerge the swab tip.
5. Mixing: Hold the upper spring cap of the ATP Quickswab and swing 30 degrees to the left and right (for five seconds), allowing the reagent to fully react with the sample.
6. Insertion: Insert the ATP Quickswab into the test camber of detector which is on the To-be-Tested interface, close the top cover and start the test.
Fig1.3-1 Unfreezing Fig1.3-2 Sampling
10
Page 23
Fig1.3-3 Installation Fig1.3-4 Injection
Fig1.3-5 Mixing Fig1.3-6 Insertion
11
Page 24
Warning:
a. The swab tip shall not contact any other object surfaces, in
order to avoid the test result from being affected;
b. Let the internal reagent fully reacts with the sample inside the
swab, and then insert the ATP Quickswab into the test chamber and complete the test within 60s.
1.6.3 ATP Quickswab Storage
1. The ATP Quickswab test tube shall be stored under 2 to 8℃, with
shelf life of 12 months.
2. Direct sunshine shall be avoided. Please maintain the ATP Quickswab
with aluminized foil bag for storage. Do not use the reagent beyond the warranty period.
1.6.4 ATP Quickswab Safety
1. The internal reagent of ATP Quickswab is diluted and can be safely
used for detection in food processing industry.
2. If standard lab operation procedures are strictly followed, the
compositions of ATP Quickswab will not be harmful for human health. The internal reagent contains 0.05% w/v of Sodium azide to exhibit its preservative effect. Please dilute the waste solution with large amount of water before disposal.
Warning: In case the internal reagent of ATP Quickswab into eyes
or onto skin, please flush eyes or skin with plenty of water. The material safety data sheet (MSDS) can be provided in demand.
12
Page 25
2. Basic Operation Instructions of Detector
2.1 Structural Diagrams of Detector
The structural diagrams of Detector are as shown in figure 2.1-1/2.
Fig 2.1-1 Detector structural diagram
1.Test chamber 2. Screen 3. Keyboard
4. Top cover 5. Hang rope hole 6. USB interface
13
Page 26
Fig 2.1-2 Detector structural diagram
1.Battery compartment 2. Label 3. Back scaffold
2.2 Function of Keys
The keyboard of detector is as shown in figure 2.2. The corresponding key functions are listed in table 1.
Fig 2.2 Detector keyboard
14
Page 27
Table1.
Key
Icon
Function
Power
Power on/off; Standby
mode.
OK
Confirm;
Start the test.
Setting
Return;
Switch between test and setting interface.
Up
Move the cursor up;
Shortcut key for records.
Down
Move the cursor down;
Shortcut key for statistics.
Left
Move the cursor left; Shortcut
key for plan
Right
Move the cursor right;
Shortcut key for program.
2.3 Battery Installation
The detector battery is installed by opening the cover of battery compartment, inserting the battery and re-installing the cover.
15
Page 28
16
2.4 Power-on Self-test
Press the < Power > key to power on the detector. In non-Silent mode, the device will give the power-on alert (one ‘beep’ sound) to enter the initial interface, as shown in figure 2.3.
In case the detector is charging, please long-press the < Power > key (about 2 seconds) on the keyboard to enter the initial interface.
Fig 2.3 Initial Interface
Reminding:
a. In case of low battery power, the detector may fail to power
on or you will be alerted with a low battery alert following with the automatically shuts down. In this condition, please charge the detector or change its battery.
Page 29
b. If users are alerted with < Calibration is halted! >, it indicates
that the power-on self-test of detector failed. Please eliminate the fault as instructed.
c. For troubleshooting, please refer to‘7. Troubleshooting’.
2.5 Internal Calibration
2.5.1 Power-on Calibration
After power on and the initial interface is displayed, the detector will enter the calibration interface, as shown in figure2.4.
Fig 2.4 Calibration Interface
Two self-test modes are available, with the calibration time of 60s and 15s (< Fast boot > mode) accordingly. The difference is that with 60s calibration time, the detector will obtain more stable and
17
Page 30
18
accurate test result. It is recommended to choose 60s self-test mode when the machine is powered on for the first time of the day.
Reminding:
a. While the calibration time the power-on self-test is counting
down, users could press < Setting > key to enter the setting interface.
b. On the setting interface users could select < Settings > to
enter the system setting page, where the < Fast boot >is shown, select the < Fast boot >, the calibration time will be 15s; Unselect the < Fast boot >, the calibration time will be 60s.
c. Press the < Setting > key again to return to the previous
interface and the calibration time counting down restarts. The detector will start the calibration, after this it will enter the test interface.
Reminding:
a. The calibration pre-conditions of the detector are: there is no
ATP Quickswab within its test chamber and its top cover is properly closed.
b. If the icon and < Calibration is halted! > are displayed on
the screen of detector, as shown in figure 2.5-1. Please take out the ATP Quickswab from the detector to continue the calibration process.
c. If the icon and < Calibration is halted! > are displayed on
the screen of detector, as shown in figure 2.5-2. Please close the top cover again to continue the calibration process.
Page 31
19
Fig 2.5-1 Fig 2.5-2
Take out the ATP Quickswab Close the top cover again
2.5.2 Recalibration
User could re-calibrate the detector if it is necessary. Please take out the ATP Quickswab and close the top cover, long press (about
3 sec) < OK > key on the test interface. After sounding a short ‘beep’, the detector will enter the calibration interface again for recalibration.
2.5.3 Auto-calibration
1. If the detector is on the test interface and all the calibration
conditions are met, it will automatically perform the calibration process;
Page 32
20
2. If the detector is not on the test interface, please press the < Setting > key to return to this interface, the detector will
automatically enter the calibration interface.
Reminding: In case the detector has been continuously working
for over 30 minutes, or the temperature change of working environment is over 5.The device will automatically perform the calibration process to ensure the accuracy of test result.
2.6 Power-off
icon
Fig 2.6 Power off
2.7 Power Saving Mode On and Resuming
Two kinds of power saving mode are provided with the detector: Sleep mode and Manual mode.
Press the
Power >
< key to display the power option menu on
the screen of detector, as shown in figure 2.6. Please select
and can press
<
OK >
key to power off the detector.
Page 33
1. Sleep mode: sleep power saving mode can be started
Fig 2.7 Screen Locker
automatically according to user-defined sleep time. When the device is on, and no activity happens during the set time, the device will automatically on standby and the screen will turn off.
Users could press < Power > key for resuming.
2. Manual mode: manual power saving mode can be started by
pressing < Power > key to display the power option menu, where users could select icon as shown in figure 2.7. Press < OK > key to let the detector on standby, press < Power > key
for resuming.
2.8 Low Battery Alert
The detector will display icon on the status bar in case the electric quantity is low. Please charge the battery in this condition; when the
status bar displays icon, it indicates that the electric quantity is very low and the device will be power off immediately. Users will be
alerted with three ‘beep’ sounds before the automatically low battery
power off.
Reminding: If battery low battery alert displays, please charge the detector as soon as possible. In case the device will be unused for a long time, please remove the battery and place it at cool and dry
place.
21
Page 34
22
2.9 Icons and Meanings
Table 2.
Icon
Meaning
The top cover is not properly closed, close it again.
Please insert the
ATP
Quickswab
.
Please take out the
ATP
Quickswab
.
The incline angle of device is too large
. The incline
(
angle of device should be less than 30°)
Ready, press the
<
OK >
key
to start the test
.
Pass
,
test result is lower than the lower limit
.
Caution
,
test result is between the
upper and lower
limit
.
Fail
,
test result is high than the upper limit
.
USB cable is not connected
.
USB cable is connected
.
Bluetooth is not connected
.
Bluetooth is connected
.
Charging
.
Electric quantity is
low .
Electric quantity is
very low, charge or power off
.
Page 35
23
3. Setting and Operation
3.1 Setting Interface
On the test interface, press < Setting > key to enter the setting interface, as shown in figure 3.1.
Fig 3.1 Setting Interface
Users could also enter the setting interface by pressing < Setting > key while the power-on self-test. Press < Setting > key
again to restart the self-test and start counting down the calibration time. After the calibration, users will be directed to the test interface.
3.2 User
Select < User > on the setting interface, and press < OK > key to access the user select option, as shown in the figure 3.2.
Page 36
24
Fig 3.2 User Select
Press < Up > key or < Down > key to view the existing users for
selection;
Press < OK > key to confirm the selection; Press < Setting > key, users could return to the previous interface.
Reminding: The device configures <User 0> by default. Users could
also execute the same related operations via PC software. For details,
please refer to ‘ Portable ATP Hygiene Monitoring System Software Specification’.
Page 37
3.3 Program
Select < Program > on the setting interface, and press < OK > key to access the < Set Program > option, as shown in the figure 3.3.
Fig 3.3 Set Program
Press < Up > key or < Down > key to select the desired
program;
Press < Left > key or < Right > key to view the < Location >
and < Surface > of the selected program.
Reminding: The detector offers two kinds of program:
a. One kind of program is written by computer and marked with
icon before program number, and its upper and lower limits cannot be changed on the device;
25
Page 38
26
b. The other kind of program is modified on the device and without
the icon before the program number, and its upper and lower limits can be user customized;
c. For a program not marked with the icon, users could press <
Up > key or < Down > key to select the procedure
(as shown in Figure 3.4) and press < OK > for confirmation;
d. Move the cursor to the upper limit setting box (as shown in Figure
3.5), using < Up > key or < Down > key to set the value (range: 1~9999) and press < OK > for confirmation;
e. Move the cursor to the lower limit setting box (as shown in Figure
3.6), using < Up > key or < Down > key to set the value (range: 1~9999) and press < OK > for
confirmation and return to the setting interface;
f. Press < Setting > key to quit setting and return to the previous
interface.
Select Setting Setting
Fig 3.4
Program
Fig 3.5 Upper Limit
Fig 3.6 Lower Limit
Page 39
27
Reminding: If the detector is stored with programs. Users could
press < Right > key on the test interface to access the set program option.
3.4 Plan
Select < Plan > on the setting interface, and press < OK > key to access the < Plan Select > option, as shown in the figure 3.7.
Press < Up > key or < Down > key to select the desired
plan;
Press < OK > key to confirm the selection; Press < Setting > key to return to the previous interface.
Fig 3.7 Plan Select
Page 40
Reminding: If the detector is stored with plans. Users could press
< Left > key on the test interface to access the set program
option.
3.5 Records
Select < Records > on the setting interface, and press < OK > key to enter the test records interface, as shown in figure 3.8.
Fig 3.8 Test Records Interface
Press < Left > key or < Right > key to page up or down;
28
Page 41
29
Press < Up > key or < Down > to move the cursor and
select test record.
Press < OK > key to confirm the selection and enter the test
record detail interface, as shown in figure 3.9.
Press < Setting > key to return to the previous interface.
Fig 3.9 Test Records Detail Interface
On the test record detail interface, and options are available.
Press < Down > key to move the cursor and select option as shown in Figure 3.10.
Page 42
30
Press < Left > key or < Right > key to select and
options for test record deletion or print;
Press < Up > key to cancel the selection, press < OK > key
to confirm and execute the selection;
Press < Setting > key to return to the previous interface.
Reminding: If the detector is stored with test records. Users
could press < Up > key on the test interface to access the test records interface.
Reminding: Before printing the test records, please check
whether the matched Bluetooth printer is connected and power on.
Fig 3.1
0 Test
Record
Dele
tion
Fig 3 . 11 Test
Record
Print
Page 43
31
3.6 Statistics
Select < Statistics > on the setting interface, and press < OK > key to access the < Statistics > option, where the statistical
Fig 3.12 Statistics
Press < Right > key to check the storage capacity of the
detector, as shown in Figure 3.13;
Press < Left > key return to the < Statistics > option; Press < Down > key to move the cursor and select for
storage deletion, as shown in Figure 3.14.
information about
< Pass >
, <
Caution >
and
< Fail >
are displayed as shown in figure 3.12.
Page 44
32
Fig 3.13 Storage Capacity Fig 3.14 Storage Deletion
Press < Up > key to cancel the selection; Press < OK > key to confirm the selection and display the
secondary confirmation for all test records deletion, as shown in figure 3.15.
Fig 3.15 All Test Records Deletion
Press < Left > key or < Right > key to select < Yes > or <
No > ( < No > is selected by default);
Press < OK > key to confirm the selection; Press < Setting > to return to the previous interface.
Reminding: If the detector is stored with test records. Users
could press < Down > key on the test interface to access the < Statistics > option, with other operations same as the above mentioned method.
Page 45
33
3.7 System Settings
Select < Settings > on the setting interface, and press < OK > key to enter the system setting interface. The options on this interface and their corresponding functions are listed in table 3 as follow.
Table 3.The options on system setting interface.
Option
Function
Date & Time
Set the date and time and its format
Sleep time
Set the auto-standby duration
Save test
Save test results or not
Fast boot
15s and 60s calibration time selection
Silent Mode
Buzzer On / Off
Language
Select the display language
Factory reset
Clear all user information
Bluetooth
Bluetooth On / Off
Brightness
Adjust the screen brightness
3.8 Template
Select < Template > on the setting interface, and press < OK > key to enter the reference template interface where the commonly­used templates are provided. Users could select a certain reference
template and press < OK > to confirm the selection and access the reference template for this industry. The template itself provides the test < Sites>, < Upper > and < Lower > limits.
Press < Left > key or < Right > key to page up or down; Press < Up > key or < Down > key for line feed;
Page 46
34
Press < OK > key to confirm the selection; Press < Setting > to return to the test interface and the selected
template will be automatically set as the test program.
3.9 Help
Select < Help > on the setting interface, and press < OK > key to enter the help interface where the < Company > name and < Website > information of the manufacture are provided.
3.10 About
Select < About > on the setting interface, and press < OK > key to enter the about interface where the information of < Battery level
>, < Temperature >, < Device name >, < Hardware version >, < Software version > and < Serial number > are provided.
Page 47
35
4. Test and Test Result
4.1 To-Be-Tested Interface
After power-on self-test and calibration, the detector will
enter the To-be-tested interface, as shown in figure 4.1.
Fig 4.1 To-be-tested Interface
1. Date & Time
2.Condition prompt
3. Plan No.
4. Status bar
5. Test result
6. Upper & lower limits
7. Program No.
8. User No.
4.2 Program Selection and Setting
There are three program selection methods, including calling of plan
program; calling of template program and user-defined program.
Page 48
36
4.2.1 Selection and Setting of User-defined
Program (Upper and Lower Limit)
1. User-defined programs are those not marked with the icon
before program number.
Press < Right > key on the test interface to access the < Set
Program > option;
< Up > or < Down > key to select a program and Press
press < OK > key to confirm the selection;
Move the cursor to the upper limit setting box, press <
Up > key or < Down > key to set its value and press < OK > key to confirm the setting;
Move the cursor to the lower limit setting box, and press <
Up > key or < Down > key to set its value;
Press < OK > key tree times to confirm the setting and return
to the test interface, whose current test program will be the one has been just modified.
2. User could also select < Program > on the setting interface,
and press < OK > key to access the < Set Program > option;
Press < Up > key or < Down > key to select a program and
press < OK > key to confirm the selection;
Move the cursor to the upper limit setting box, press <
Up > key or < Down > key to set its value and
press < OK > key to confirm the setting;
Page 49
37
Move the cursor to the lower limit setting box, press < Up >
key or < Down > key to set its value;
Press < OK > key tree times to confirm the setting and return
to the test interface, whose current test program will be the one has been just modified.
Reminding: The detector is stored with 1999 programs that not
marked with the icon by default ranging 1~1999 (0 is the default procedure and cannot be modified);
Reminding: In case a program is written by users from PC
software, such program will increase progressively from 1 and replace the original program with the same number on the detector. The information about the upper & lower limits, sites of
the program can be freely set on PC software according to users’
requirements.
Reminding: The programs written on the PC software whose
program number is marked with icon cannot be edited again
on the detector. For details, please refer to ‘Portable ATP
Hygiene Monitoring System Software Specification’.
4.2.2 Plan Setting and Program Selection
Plan is written on PC software and imported into the detector via USB
cable. One plan can contain several programs. For details, please refer
to ‘Portable ATP Hygiene Monitoring System Software Specification’.
1. Select < Plan > on the setting interface, and press < OK >
key to access the < Plan Select > option.
Press < Up > key or < Down > key to select the desired
plan;
Page 50
38
Press < OK > key to lock the selected plan; Press < Up > key or < Down > key to select the desired
program;
Press < Right > key to view the < Location > and <
Surface > of the selected program;
Press < Left > key to return to the previous step; Press < OK > key to confirm the selection and return to the
setting interface;
Press < Setting > key to return to the test interface and the
selected program will be automatically set as the current test program.
2. Users could also press < OK > key to access the < Plan Select >
option for the plans contained.
Press < Up > key or < Down > key to select the desired
program within the selected plan;
Press < Right > key to view the < Location > and <
Surface > of the selected program.
Press < Left > key to return to the previous step Press < OK > key to return to the test interface and the
selected program will be automatically set as the current test program.
4.2.3 Calling of Template Program
1. Select < Template > on the setting interface, and press <
Page 51
39
OK > key to enter the reference template interface where the
commonly-used templates are provided.
2. Users could select a certain reference template and press < OK
> to confirm the selection and access the reference template for
this industry. The template itself provides the test < Sites >, <
Upper > and < Lower >limits.
Press < Left > key or < Right > key to page up or down; Press < Up > key or < Down > key for line feed; Press < OK > key to confirm the selection; Press < Setting > key to return to the test interface and the
selected template will be automatically set as the test program.
Reminding: After selecting and confirming a certain template
program, please press < Setting > key to return to the test interface, the selected template program will be set as the current test program.
4.3 Start Test
1. After completing the program setting, users could open the top
cover of the detector, insert the test ATP Quickswab, and properly
close the top cover.
2. Keep the detector upright and ensure its incline angle is less than
30°during the test. The detector will display the test interface, as
shown in figure 4.2.
Page 52
40
Fig 4.2 Test Interface
Press < OK > key to start the test and counting down the test
time (10s). After this, the detector will display the test result.
The detector will automatically analyze the test result according
to the program upper and lower limits and offer the
determination of the test result, displayed as < Pass >, <
Caution > or < Fail >.
In non-silent mode, one beep will sound before the detector
automatically return back to test interface, where users could press < OK > key again for second or more tests.
Page 53
41
4.4 Records Regarding Operations
4.4.1 View Records
1. Select < Records > on the setting interface, and press < OK
> key to enter the test records interface.
Press < Left > key or < Right > key to page up or down; Press < Up > key or < Down > to move the cursor line
feed.
Press < OK > key to confirm the selection and enter the test
record detail interface to view the test result.
Press < Setting > key to return to the previous.
2. User could also press < Up > key on the test interface to directly
enter the test records interface.
Press < Left > key or < Right > key to page up or down; Press < Up > key or < Down > key to move the cursor line
feed.
Press < OK > key to confirm the selection and enter the test
record detail interface to view the test result.
Press < Setting > key to return to the previous.
4.4.2 Print Records
On the test record detail interface, and
options are
available.
Press < Down > key to move the cursor to
option;
Press < Right > key to select the option;
Page 54
42
Press < OK > key to confirm the selection and print the
current record.
Reminding: Before printing the test records, please check
whether the matched Bluetooth printer is connected and power on.
4.4.3 Delete Records
On the test record detail interface, and options are available.
Press < Down > key to move the cursor and select
option;
Press < OK > key to confirm the selection and delete the
current record.
5. Connect Detector to PC or Other
Terminals
5.1 Connect Detector to PC
1. Open the PC software on the computer and use the provided USB
cable to connect this computer with the detector while it is power
on;
2. The < USB > icon on the status bar of detector will be lighted
in blue and a prompt < Connected > will pop up on the screen of
Page 55
43
detector, reminding users that the device is successfully
connected to the control computer;
3. Click the Connect icon in the toolbar of the PC software. The
screen of detector will display a prompt < Communicating>,
reminding users that the device is communicating with the
control computer to realize data exchange and other operations.
5.2 Disconnect Detector from the connected PC
After the detector is successfully connected to the control computer.
1. Users could press < OK > key on the detector to disconnect
the device from the conncted computer;
2. Users could also click Disconnect icon in the toolbar of the PC
software to disconnect the device from the conncted computer.
5.3 Connect Detector to Bluetooth Printer
1. Select < Settings > on the setting interface, and press < OK
> key to enter the system setting interface.
2. Open the < Bluetooth > option on this interface and the <
Bluetooth > icon on the status bar of detector will be lighted
in blue (The device is named is and its password is 0000);
3. After the Bluetooth is enabled, the device can be matched with the
Bluetooth printer;
4. After matching, the test result can be printed on paper.
Page 56
44
6. Operation and Maintenance
6.1 Daily Precautions
Please carefully read this manual before using the detector; Please power off the detector after test; In case the device will be
unused for a long time, please remove the battery and place it at cool and dry place;
The detector belongs to high-precision optical device which shall
be stored and used against damp as well as water and rain;
The detector shall not be operated in dusty environment to
ensure the test result accuracy;
If the detector automatically powers off due to the low battery.
Please charge it as soon as possible and do not forcibly start it again and again;
The detector shall not be operated in environment with strong
magnetic field interference;
No person other than the qualified maintenance staff assigned by
our company shall disassemble the detector.
6.2 Battery Charge or Replacement
In case the low battery alert displayed in the status bar of the
detector, please charge the battery and be sure to use the supplied power adaptor.
In case the discharge time of the provided battery is short or the
battery fails to recharge or discharge, please contact our service support or local distributor for battery replacement.
Page 57
45
6.3 Test Chamber Clean or Replacement
The test chamber of detector can be removed for cleaning or replacement.
1. Users could open the top cover of detector, insert one finger into
the test chamber, and use the friction force between the finger and test chamber to pull out the test chamber.
2. As the acquisition channel for optical signal, the bottom structure
of the test chamber is designed as a special transparent cap. In order to ensure the optical performance, thus the performance of detector, the test chamber cap shall be well protected during disassembly. If any part of it is damaged, please contact our service support or local distributor for test chamber replacement.
3. Users could clean the test chamber interior with warm water or
mild detergent. Let the test chamber natural air drying, after this please insert it back into the detector till it reaches the bottom and is locked. Please pay attention to ensure the right insertion direction.
Warning: Please power off the detector and remove the battery
before cleaning or replacing the test chamber.
Prohibit: It is prohibited to clean the test chamber with strong
detergent.
7. Troubleshooting
Table 4. The main possible errors during the self-test process of
detector and their possible causes.
Page 58
46
Corrective instruction: for errors during self-test process, the detector can be restarted on the condition that the battery has sufficient voltage and the working environment meets the pre­mentioned requirements.
Reminding: If the error persists, please contact our service
support or local distributor.
Table 5. Other possible errors and their corrective instructions.
No. Description
P ossible causes
1 Storage unit error
Memory storage fails
2
File system error
Initialization of the file system
fails
3 Temperature test error
The temperature test module fails
4 Temperature error
The environment
temperature is
not within 5~40°C
5 Incline angle test error
The incline angle sensor is invalid
6 Battery error
The battery voltage is very low
7 Background error
The background value of the test
environment is out of specification
8 Test module error
The test module fails
Page 59
47
NO .
Error
Description
Possible Cause
Corrective
Instructions
1
Press the
<
Power
> key,
while the
detector
cannot
be powered on
a.
Low battery;
b.
Battery is
loosened;
c.
Device or
keyboard is
damaged.
a.
Connect to an
external power
supply. If the
device starts
normally, charge
the battery or
re - install the
battery;
b.
Otherwise please
contact us.
2
Press the
<
Power
>
key,
while the
detector cannot
be powered off
a. Improper
operation;
b. Device or
keyboard is
damaged.
a. Replace the
keyboard;
b. Restart or
power
-
down
restart.
3
Abnormal
shutdown
a. Low battery;
b. Battery is
loosened;
c. The device has
been fell
damaged or
strongly
vibrated;
d. The device is
automatically
powered off
within the
standby time;
e. The device is
a.
Re - install the
battery;
b.
Check the
integrity of the
device;
c.
Restart the
device.
Page 60
48
damaged or malfunctioned.
4
The screen has no display or partial display
a. The screen is
damaged;
b. Auto power
saving mode;
c. The device is
automatically powered off due to low battery.
a. Wake up power
saving mode;
b. Charge the
battery.
5
The test records cannot be stored
a. The system has
been set not to store the test records;
b. A certain part of
the device is loosened or damaged
a. Check if the
system is enabled to store the test records;
b. Check the
integrity of the device.
6
The test results always reads 0 or a value lower or higher than the set limits
a. The ATP
Quickswab is not properly used;
b. The ATP
Quickswab is expired;
c. The test is
performed in an unstable environment;
d. The device is
contaminated.
a. Properly use a
functional ATP Quickswab;
b. Recalibrate the
instrument;
c. Operate the
device in a compliant
environment; d.
Clean the test chamber
Page 61
49
Reminding: If the error persists, please contact our service
support or local distributor.
Warning: in case of any following condition occurs, please
immediately power off the device, and contact our service support or local distributor to ask for qualified maintenance staff for assistance:
1. Any liquid has enter into the device;
2. Any abnormal sound or smell appear inside the device;
3. Device is soaked with water or rain;
4. Any housing damage caused by accidently drop of the device;
5. Obvious functional changes of the device.
Page 62
50
8. Commitment Statements of
After-sale Services
8.1 Warranty Service
We will provide 24 months of warranty period beginning from the delivery of the detector. Within the warranty period, our company guarantees the maintenance of the device in case any malfunction is caused by the defects of materials and device manufacture. Users could contact us, and we will arrange maintenance personnel for processing (We promise to provide the list of components and the electric circuit for the maintenance personnel). Or directly send the device back to the maintenance department assigned by our company (Users shall be responsible for the transportation fees), and we will send back the well maintained device to the users for free.
Reminding: The following circumstances are not within the
scope of warranty:
1. The device damages caused by improper, negligent operation or
force majeure including: war, fire, flood, earthquake, typhoon and any other unforeseen accidents.
2. The parts or components damages caused by abnormal voltage.
3. Users did not comply with the rules of transportation and use.
4. Users did not comply with the maintenance instructions.
5. The device has been opened or maintained by person,
manufacturers or agents that are not authorized by our company
8.2 Response Time
We will respond within 24 hours upon receiving the notification, no matter whether the device is within the warranty period or not. For
Page 63
51
any problems that cannot be solved through the phone, the device will be sent back to our company for repair.
8.3 Spare and Accessory Parts
For post warranty service, our company still provides the maintenance service for the device, we will charge for the maintenance fee according to the specific situation and our company will supply the commonly used spare parts at most preferential price for long-term.
8.4 Special Statement
It is a product of PCE-Instruments and provided with dedicated ATP Quickswabs. Please purchase our genuine ATP Quickswabs to ensure the accuracy of test result and avoid the unnecessary trouble due to the unmatched ATP Quickswab. The company assumes no responsibility for the non-genuine device or ATP Quickswab.
Warning: in case the device occurring any problem, please
contact our service support or local distributor. Do not disassemble the device without permission.
Warning: our company assumes no responsibility for any
problem raised from disassembly of the device without permission which will invalid the warranty.
Page 64
52
9. Portable ATP Hygiene Monitoring
System Software Specification
9.1 Overview
ATP data management system is mainly used to realize functions such as data exchange between PC and the device, test record analyzing and processing and unidirectional command control.
ATP data management software has following main features and advantages:
Compatible with current mainstream Windows OS versions such as
Win8/Win7/Vista/XP;
Fully independent from any third-party software; Its communication employs the standard Windows driver,
requiring no third-party driver;
Chinese and English are fully supported; Capable of directly uploading test data from one or more devices; Editable templates such as User, Program and Plan, which can be
downloaded to the device for usage;
Uploaded data and edited templates can be directly exported as
Excel-compatible format;
Test data can be output as statistical graph based on specified
conditions (histogram, sector diagram, line chart and so on) and can be printed;
Uploaded test data is encrypted.
Page 65
53
9.2 Setup
Open the included CD and locate ‘setup.exe’ to run the executable file.
You will be automatically directed to the ‘Installation Confirmation’
interface, as shown in figure 9.1.
Fig 9.1 Installation Confirmation
Click the Next button to enter the ‘Installation Directory Selection' interface, as shown in figure 9.2.
Page 66
54
Fig 9.2 Installation Directory Selection
Users could click the Change button to change the Installation directory. Click the Next button to enter the ‘Software Installation’ interface, as shown in figure 9.3.
Page 67
55
Fig 9.3 Setup
Click the Install button to enter the ‘Installation Status’ interface;
After the installation, you will be directed to the ‘Installation
Complete’ interface, as shown in figure 9.4. Please click Finish to
confirm.
Page 68
56
Fig 9.4 Software Installation
The software will run automatically upon completion of installation, and generate a shortcut icon ‘ ’ on the desktop for easy operation.
9.3 Remove
Users could find the ‘ uninstall’ icon in the Start menu of Windows operation system, click it and a uninstall dialog box will pop up, according to whose prompt, the uninstall operation can be done.
Users could also uninstall the software from the ‘Control panel - Program and function’ menu of the computer.
Page 69
57
10. Software Interfaces and Features
10.1 Software Interfaces
The initial interface is as shown in figure 10.1.
Fig 10.1 Initial Interface
10.2 Overview of Features
10.2.1 Menu Bar
File Pull-down Menu
File
Icon
Feature
New...
Generate a statistical graph report according to test data;
Open...
Open a statistical graph report from the disc;
Export... Export list contents such as test records and
Page 70
58
templates to the disc;
Exit / Exit the software.
Option Pull-down Menu
Option
Icon
Feature
Login…
Switch to Chinese
Connect…
Exit
/
/
Login as administrator
Switch interface language
Connect the device Exit the software.
Help Pull-down Menu
Help Icon
Feature
Contents F1 / Display help document information
About / Software Version Information
10.2.2 Tool Bar
This area contains shortcuts of New, Open, Export and Connect.
Page 71
59
10.2.3 Function Options
Function option
Icon
Feature
Record
View, sort, search, delete and edit of uploaded test records
User
Edit, add or delete user templates
Program
Edit, add or delete procedure templates
Plan
Edit, add or delete plan templates
Report
View, print or delete statistical graphs
Exchange
Data sync with device, firmware update and issue of control command
11. Operating Guidance
11.1 Connect Device to PC
Start up the device and use the data cable to connect the device to the USB port of your PC.
Page 72
60
11.2 Connect Device to Software.
Open the software, select the connect icon from the pull­down menu of the ‘Option’, or directly click the icon on the tool bar. After the device is successfully connected to the software, the device interface is as shown in Figure 11.1, and the connect icon also changes
to
Fig 11.1 Connecting
11.3 Record
11.3.1 Description
Click the < Record> from the function options on the main interface, the software will directly enter the record interface as shown in figure 11.2.
icon.
Page 73
61
Fig 11.2 Record
After the device is successfully connected to the software, and data synchronization is completed, the test record will be displayed in the record management section to display detailed record information.
Users could edit, delete, export and generate the statistical report based on the test records, and also analyze these data by time, location, user or schedule.
Filter the displayed options
Users could select the records within a
certain period of time.
Displays the record of a certain location
Displays the record of a certain schedule
Displays the records of a certain user
Page 74
62
11.3.2 Records Editing
Administrator privilege is required for management of the test records. Under the administrator model, The < User >, < Plan >, < Program >, < Location > and < Surface > information of these test records can be edited.
11.3.2.1 Login as Administrator
From the pull-down menu of ‘Option’ on the menu bar, click the icon
Login… to display the ‘Input administrator password’ dialogue box, as shown in figure 11.3.
Fig 11.3 Input administrator password
In the ‘Input administrator password’ dialogue box, you can input the password ‘0000’ to get the administrator privilege, and click OK to
log in, as shown in figure 11.4.
Fig 11.4 Login Successfully
You can double click the cell to be edited to edit the related records.
Page 75
63
11.3.2.2 Logoff
From the pull-down menu of ‘Option’ on the menu bar, click the icon
Logoff to exit from the system.
11.3.3 Records Deletion
After login as administrator, users could select a certain record and right click to delete the selected test records.
11.3.4 Records Export
Click the icon Export from the pull-down menu of ‘File’ on the menu bar, or directly click the icon ( ) on the shortcut on the tool
bar to display the Export dialogue box. Select the save position, and enter the file name. Click OK to save the file in ‘CSV’ format that can be opened by Excel or in txt format which facilitates data entry of users.
11.4 User
11.4.1 New User
When the button Add in the work area is enabled, click it to add a new user. Input user name and press the [Enter] key to confirm it, as shown in figure 11.5.
Fig 11.5 Add New User
Page 76
64
Note: When the user name is emptyalready exists or exceeds 10 characters, the software will reminding that the user name is invalid, please re-set the user name again.
11.4.2 Edit User
Double click the user entries to be edited to enter the Edit mode, after editing the user name please press the [Enter] key to confirm the modification.
11.4.3 Delete User
After selecting the user entries to be deleted, the button Delete in the work area will be enabled. Press the button and click Yes in the displayed Confirm Delete dialogue box to delete the selected user entries.
11.5 Program
This module provides the feature to customize test program which
essentially consist of test ‘Location’, ‘Surface’, ‘Upper’ reference limit and ‘Lower’ reference limit.
Among them, the test ‘Location’ and ‘Surface’, have the maximum
character length of 30 characters and the range of ‘Upper’ and ‘Lower’ reference limit is 1 - 9999, while the upper limit is not less than the lower limit. Up to 1999 programs can be added, as shown in figure 11.6.
Page 77
65
11.5.1 New Program
When the button Add in the work area is enabled, click it to add a new program. Input all the corresponding information of the new program and press the [Enter] key to confirm it.
11.5.2 Edit Program
Double click the cell to be modified to enter Edit mode, edit the corresponding content and press the [Enter] key to confirm the modification.
11.5.3 Delete Program
After selecting the program entries to be deleted, the Delete button in the work area will be enabled. Press the button and click Yes in the displayed Confirm Delete dialogue box to delete the selected program entries.
Reminding: Delete the program in the ‘Program’ module, the
corresponding program in the ‘Plan’ module will be also deleted.
11.6 Plan
This module provides the feature to customize test plan that consist of several test programs. As shown in the above figure, the left section
Fig 11.6 Program
Page 78
66
displays the t test plan, while the right section displays test programs contained in the selected test plan.
Up to 255 test plans can be added, each of which contains up to 255 test programs, as shown in figure 11.7.
11.6.1 New Plan
When the button Add on the schedule column in the work area is enabled, please press the Add button to add a new plan. Users could use the system automatically generated plan name or manually customize the new plan name. After adding a new plan, the right program column will add the first test record of the current program template by default. Users are allowed to add or change program contents and sequence in the right program column.
Reminding: A plan can be added only when the test program is not
null.
11.6.2 Edit Plan
Plan editing is to modify the program information of the selected plan. You can select the desired program from the pull down list in the program column.
Fig 11.7 Plan
Page 79
67
11.6.3 Delete Plan
After selecting the plan entries to be deleted, the button Delete in the work area will be enabled. Press the button and click ‘Yes’ in the displayed Confirm Delete dialogue box to delete the selected plan entries.
11.7 Report
The contents displayed are statistical graphs report generated and saved by New Wizard under the current ‘Report’ directory, As shown in figure 11.8.
Double click a certain report to open it, the statistical graph is as shown in figure 11.9. User could also right click it to display the Delete option.
Fig 11.8 Report
Page 80
68
Fig 11.9 Statistical Graph
Right click on the statistical graph to display options menu, as shown in figure 11.10.
Fig 11.10 Options Menu
Page 81
69
Functional description
Options Description
Close
Close the current statistical graph
Print…
Print the current page
Print Preview
Preview the current page printing effect
Histogram
Display the histogram for statistical report
Pie chart
Display the statistical report in form of pie chart
Line chart
Display the change trend of statistical results
in form of line chart
Hide/show Label Hide/show the percent comment of charts
11.8 Exchange
This module is not enabled when the software starts.
11.8.1 Activation Method
1. Check that the device is properly connected to your PC through the
data cable;
2. Click the button Connect on the tool bar. This button will change to
after connection succeeds.
If it is not properly connected, check if the device is reliably connected to your PC.
After successfully connected the device with PC, ‘Device Firmware Upgrade’, ‘Synchronization’ and ‘Control Command’ buttons will be
Page 82
70
enabled for data exchange between your PC and device, as shown in figure 11.11.
Fig 11.11 Exchange Interface
11.8.2 Device Firmware Upgrade
This module is used to update the device firmware program. Please make sure the unit is successfully connected to computer via the provided USB cable update. Do not shutdown the computer or the device during the update process.
11.8.3 Synchronization
This module is used to synchronize the data between the device and PC. After successfully connecting the unit to the computer, please click Synchronization to upload the test results from the unit to the computer. After the completion of the upload, the software will
Page 83
71
automatically reminding the users ‘Whether to clean all the data in device?’ Click NO the data device will not be deleted; Click YES the
test data in the device and downloaded template will be cleared, and the < User >, < Program > and < Plan > templates created on the computer will be downloaded into the device in turn to complete data synchronization.
11.8.4 Command Send
This module is used to send control commands to the device, in order to realize the remote control of PC software on the device. There are two control commands: ‘Single Test’ and ‘Shutdown’.
Single Test: place the swab and close the top cover to ensure the swab is being tested. Press the Single Test key to start test.
Shutdown: Shut down the device.
Loading...