PCE Instruments PCE-3000ULS Users guide

Page 1
PCE-3000ULS
HARDNESS TESTER
OPERATING MANUAL
Page 2
This operating manual contains information on the purpose, functions, technical characteristics, the principle of operation, design, and operation of the portable hardness tester PCE-3000ULS (hereinafter - the hardness tester) and the rules of its operation, transportation and storage.
1
Page 3
Page 4
2.6.1.1 "Create scale" item .......................................................................................... 35
2.6.2 "Change scale" item ........................................................................................... 38
2.6.2.1 Calibration adjustment .................................................................................... 38
2.6.2.2 Change the calibration name ........................................................................... 40
2.6.2.3 Restore settings ............................................................................................... 41
2.6.3 Delete scale item ................................................................................................ 42
2.7 Custom calibration ................................................................................................ 43
2.7.2 Edit name ............................................................................................................ 47
2.7.3 Delete scale item ................................................................................................ 48
2.8 One point calibration ............................................................................................. 49
2.9 Settings tab ............................................................................................................ 52
2.9.1 Description of the "Settings" tab items .............................................................. 52
2.9.2 Date and time ..................................................................................................... 53
2.9.3 Saving a series of measurements ....................................................................... 53
2.9.4 Probe type ........................................................................................................... 54
2.9.5 Language ............................................................................................................ 55
2.9.6 Backlight brightness ........................................................................................... 55
2.9.7 Change password ................................................................................................ 56
2.9.8 Sound settings .................................................................................................... 57
2.9.9 Auto-shutdown time ........................................................................................... 57
2.9.10 USB settings ..................................................................................................... 58
2.9.11 Service menu .................................................................................................... 59
2.9.12 Statistics ........................................................................................................... 59
2.9.13 About the device............................................................................................... 60
2.10 Marking of the hardness tester ............................................................................ 61
2.11 Hardness tester packaging ................................................................................... 61
3. MAINTENANCE .................................................................................................. 62
4. TRANSPORTATION AND STORAGE ............................................................. 63
3
Page 5
5. PRECAUTIONS .................................................................................................... 64
6. RECYCLING ........................................................................................................ 65
ATTACHMENT 1 ..................................................................................................... 66
4
Page 6
1. DESCRIPTION OF THE HARDNESS TESTER
1.1 Purpose of the hardness tester
Universal hardness tester PCE-3000ULS is designed to control the hardness of metal products.
The PCE-3000ULS hardness tester combines two methods of hardness measurement: ultrasonic and dynamic. This makes the instrument the most versatile and efficient solution for incoming, in-operational and outgoing material quality control.
The universal hardness tester is used for various tasks. For example, it is well suited for measuring carbon and structural steels, surface-hardened products, heat-resistant, corrosion-resistant, stainless steels, galvanic coatings, surfacing, products of complex configuration, thin-walled and compact products.
The device is pre-calibrated Leeb scale for steel, cast iron, stainless steel, aluminum, bronze and brass, copper. Standard Rockwell, Vickers, Brinell, and Shore calibrations are also available. The user can create and correct (if required) additional calibrations for different materials. The UCI method is pre-calibrated for steel
The hardness tester can be used to control hardness:
- high-alloy, heat-resistant, corrosion-resistant, stainless, and other steels
- non-ferrous metals and alloys
- cast iron
- surface-hardened and other layers on steel products (HFC hardening, carburizing, nitriding, etc.)
- products from fine-grained materials
In the case when the properties of the controlled material differ from the material set
as sample material, measurements are carried out after programming an additional adjustment (or additional scale) on hardness samples of the corresponding material by
the user of the device or at the manufacturer, by the user's order.
The PCE-3000ULS is available with pre-installed optional scales for testing a range of
5
Page 7
different materials.
The hardness tester is designed for use in laboratory, workshop and field conditions.
1.2 The principle of operation of the hardness tester
The hardness tester consists of a probe and an electronic unit for converting signals
from the probe and processing the measurement results.
The electronic unit receives a frequency signal from the probe of the device, converts it into units of hardness, and displays the measurement results on the display, statistical processing, and other functions.
The principle of operation of the hardness tester is based on two methods of measuring hardness - dynamic and ultrasonic.
Dynamic method
The technique is to determine the rebound rate of the carbide indenter from the surface of the controlled product.
The components of the hardness tester are a probe and an electronic unit for converting signals from the probe and processing the measurement results.
The main parts of the probe are the indenter and the electromagnetic coil. When the rebound of the indenter from the test product in the coil is the EMF, which is proportional to the rate of rebound of the surface of the product. The rebound speed is determined by the hardness of the productу.
Since the bounce rate of the indenter is an indicator of hardness, there is a relationship between the bounce rate V and the hardness of the material H:
Н = f(V)
The electronic unit of the hardness tester receives a signal from the probe of the device, its conversion into units of hardness, the output of results of measurements on the display, statistical processing, and other functions of this hardness tester.
6
Page 8
Ultrasonic method
The principle of operation of ultrasonic probes of the hardness tester is based on the method of measuring ultrasonic contact impedance (UCI - ultrasonic contact impedance).
At the end of the metal rod, which is part of the hardness tester probe, a diamond tip is fixed. The rod oscillates at its own resonant frequency. When a load is created by the user's hand, the diamond tip is embedded in the material and changes the resonant frequency of the rod. The change in the natural resonant frequency of the rod is proportional to the depth of penetration of the tip into the material. Since the penetration depth of the tip into the material is an indicator of hardness, there is a relationship between the change in the resonant frequency of rod F and the hardness of the material H: H = f ( F ).
1.3 Hardness tester functions
The control is carried out:
according to the main scales: Rockwell (HRC), Brinell (HB), Vickers (HV), and Shore for measuring the hardness of carbon structural steels;
scale conversion: Rockwell (HRA), Rockwell (HRB), to control the hardness of carbon structural steels;
on additional scales set in the hardness tester by the manufacturer for hardness control or scales created by the user:
alloy, tool steels,
aluminum alloys in Brinell units (НВ),
brass (copper-zinc alloys) in Brinell units (НВ),
bronzes (copper-tin, copper-aluminum) in Brinell units (НВ);
using custom scales of hardness tester (UCI) - for cases when the
physical and mechanical properties of the controlled material differ from carbon structural steels (high-alloy steels, specialized cast irons, non-ferrous metals, alloys, etc.).
It is possible to calibrate the main scales in the event of an additional error after
prolonged use.
The hardness tester is equipped with a bright color display and a strong case for
7
Page 9
protection against dust and moisture.
Hardness tester during measurements allows calculating the average value of a series
of results of measurements and filtering incorrect measurements that are out of range.
Hardness tester during measurements allows to carry out additional statistical processing of a series of results of measurements - search of the minimum, maximum values, calculation of average value, construction of graphs and display of data in the form of the table.
Hardness tester during measurements allows you to display additional information – previous measurement results.
The device allows you to organize the data archive in the form of nominal blocks of measurement results, save it when the power is off and transfer data to a computer.
Allows to carry out various types of the analysis of the saved results of measurements, to carry out the construction of various types of graphs directly on the device display.
The hardness tester allows selecting the information which is in addition displayed on the device display in the course of measurements.
The hardness tester constantly displays the battery charge status and signals its discharge during operation.
The device allows you to set the time of automatic shutdown of the device during pauses in its operation to further save battery power.
The hardness tester allows you to choose the type of probe.
Allows you to adjust the display backlight mode to further save battery power.
The hardness tester allows you to set the backlight brightness of the display.
Allows you to select the language of the interface.
8
Page 10
1.4 Technical specifications
Measuring scale
Rockwell
Brinell
Vickers
Measuring range
20 – 70
HRC
30 – 650 HB
230 – 940 HV
Accuracy
2 HRC
10 HB –
in range 90-180 HB
15 HB –
in range 180-250 HB
20 HB –
in range 250-460 HB
15 HV – in range
240-500 HV
20 HV – in range
500-800 HV
25 HV – in range
800-940 HV
Standards
ASTM A1038 and ASTM A956
Indenter
Probe UCI – Diamond indenter
Leeb Probe - Hardened Ball
Diameter of the platform for installing the probe
For ultrasonic probe:
– from 5 mm/0.197" in
a blind hole (groove)
For dynamic probe:
from 10 mm/0.394" on the
plane
Materials
Ultrasonic Sensor (UCI) - pre-calibrated for steel
Dynamic (Leeb) with pre-calibration for steel, cast iron,
stainless steel, aluminium, bronze, brass, copper
Additional Custom Calibration Materials
Algorithm for discarding incorrect measurements
Yes
Calculation
Average value from 1 to 20 measurements
Minimum, maximum, average value
search for results of incorrect measurements
Scale conversion
Converting the measured hardness into different scales
Instrument housing
Drop-proof plastic housing with rubber bumper
Operating temperature
- 20 ... +45 °С / -4 …113 °F / non-condensing
Graphs
All points in the series included in the calculation of the
average
Display
LCD TFT 3.5" 320x480 px
Language
PC connection
Product surface roughness
for probe type "UCI":
1,6 Ra
for probe type "D": 3.2 Ra for probe type "G": 7.2 Ra
for probe type "E": 1.6 Ra
Power supply
built-in battery
Working hours
approx. 9h
Dimensions
183×93×42 mm / 7.2 x 3.7 x 1.7 "
Weight
300 g / < 1 lb. (without probe)
Table 1.1
English, German, Turkish, Spanish, Ukrainian, Russian
USB-C
9
Page 11
1.5 Hardness tester probes
The standard delivery set includes 2 probes – ultrasonic contact impedance and dynamic probe type "D" (Leeb).
The Rebound Leeb Probe – purposed for hardness measurement by dynamic method.
It is intended for measuring hardness of large-sized objects and also coarse-grained materials.
To ensure optimal measurement conditions on products, hardness testers can use additional dynamic probes with different dimensions and different spring stiffness, which provide different impact energy of the indenter against the tested product.
In a dynamic probe, the support washer can be removed (unscrewed) from the probe to reduce the area required for its placement on the surface of the product during measurement.
The Ultrasonic Contact Impedance Probe – measurement of hardness by Ultrasonic Contact Impedance method. The use of a diamond indenter allows the probe to be mounted precisely at any tiny point, and leaves a small-size imprint, making the measurement of the UCI probe the least destructive.
The probe is perfectly suited for the following tasks: measuring the hardness of complex shapes, fine-grained materials, heat-treated materials, thin layers and coatings, surface hardened parts, thinwalled pipes, small parts, etc.
In the ultrasonic probe, the protective nozzle (nose) serves to protect the rod from overload and contact with a foreign object or the user's hand during the measurement. When measuring hardness in hard-to-reach places, it is permissible to remove the protective cap, while the permissible parameters of the holes/grooves will be changed (table 1.5).
Probes are manufactured with built-in connecting cables for connection to the electronic unit. The list of typical probes is given in Table 1.2. Only the D and UCI probe
are available at PCE Instruments.
10
Page 12
Table 1.2
Model
Features
Main application
D
The universal probe comes standard with the hardness tester.
Used for the main number of
measurement tasks
G
The probe with increased impact energy (relative to probe "D")
For coarse grained materials up to 450
HB with high surface roughness.
Е
Probe with reduced (to probe "D") impact energy and diamond indenter
For testing very hard materials
up to 72 HRC
UCI
The main type of ultrasonic probe for solving most hardness measurement tasks. 5kg/11lb. load is to be applied for measurement (automatically controled by the probe). Average surface cleanliness requirements.
- Heat-treated and cemented parts, for example, shafts, turbines, gears, teeth, welds, heat affected zones.
- Measuring in grooves, on teeth, in grooves, on radius surfaces.
- Measurement on the blades, on the internal surface of the pipes, openings.
UCI-S
The reduced load probe is designed to measure the hardness of material with increased requirements to the print size (polished surfaces), to the hardness measure of surface hardened layers. 1 kg / 2.2 lb. load is to be applied for measurement (automatically controlled by the probe). More sensible to the surface cleanliness
- Control of the hardness of galvanic coatings: chrome, copper, tin, nickel, zinc
- Nitrided and cemented surface layers of molds, stamps, stamps, thin-walled parts.
- Bearings, lateral surfaces of saws teeth.
- Measurement of hardness of hardening coatings.
- Measurement on the blades, on the inner surface of the pipes, inside the holes.
UCI-R
Control of the products hardness with a rough surface up to Ra5. 10 kg / 22 lb. load is to be applied for measurement (set automatically by the probe). Low requirements for surface cleanliness.
- Heat-treated and cemented details.
- Measuring in grooves, on teeth, on radius surfaces.
- Measurement on the blades, on the internal surface of the pipes, openings.
UCI-L
Measurements in hard-to-reach places
- on products with teeth, grooves, notches
UCI-P
Measurements in other hard-to-reach places
in the middle of containers and pipes
from 80 mm / 3.2"
the product when measuring different hardness (for carbon structural steels) are shown in
The average conditional diameters of the impressions (mm) created on the surface of
11
Page 13
table 1.3.
Marking of
the probe
type
103 НВ (103 НV)
209 НВ (212
НV)
406 НВ (420
НV) (42,5 НRС)
763 НV (63,0
НRС)
D
G
-
Е
UCI
Marking of
the probe
type
103 HB (103 НV)
209 НВ (212
НV)
406 НВ (420
НV) (42,5 НRС)
763 НV (63,0
НRС)
D
G
-
Е
UCI
Hole/groove diameter from, μm
Hole/groove depth up to, μm
with spout (normal)
without spout
with spout (normal)
without spout
10 5 18
23
Table 1.3
0.80 0.72 0.67 0.57
1.29 1.22 0.93
0.80 0.72 0.67 0.57
0.23 0.16 0.13 0.09
The average conditional depths of impressions (mm) created on the surface of the product when measuring different hardness (for carbon structural steels) are shown in table
1.4.
Table 1.4
0.054 0.043 0.038 0.027
0.084 0.075 0.044
0.054 0.043 0.038 0.027
0.066 0.045 0.05 0.037
Table 1.5 - Parameters of holes/grooves for the Ultrasonic probe
12
Page 14
1.6 Requirements for the controlled product
Marking of the probe type
Minimum weight, kg
Minimum weight, lb.
D 5 11 G 15
33 Е 5
11
UCI
0.67
Marking of the probe type
Minimum thickness, mm
Minimum thickness, inch
D
25 1 G
70
2.8 Е 25
1
UCI
5
0.2
Marking of the probe type
Surface roughness, no more
D
G
Е
UCI
The minimum recommended weight of the controlled product is in table 1.6
Table 1.6
0.3
If the weight of the controlled product is less than the specified - at measurements it is necessary to be guided in addition by paragraph 1.10 "Measurements on light and thin-walled products".
The minimum recommended thickness of the controlled area of the product - is according to table 1.7.
Table 1.7
If the thickness of the controlled product is less than specified - when measuring must be further guided by paragraph 1.10 "Measurements on light and thin-walled products".
The maximum recommended surface roughness of the product (surface area) on which measurements are made by table 1.8.
Table 1.8
Rа 3.2
Rа 7.2
Rа 3.2
Rа 1.6
13
Page 15
If the surface roughness of the product is higher than specified, the measurements must be additionally guided by paragraph 1.11 "Work on products with high roughness".
The minimum radius of curvature of the convex controlled surface is 18 mm /0.7" (for probes D, and E). For the UCI probe - 4 mm / 0.15".
The minimum radius of curvature of the concave controlled surface is 200 mm/ 7.9"
(with the support washer removed, for probes D, and E). For UCI probe - 6 mm/0.23".
Minimum diameter of the platform for measuring the hardness on the surface of the controlled product (for installation of the probe).
Probes D, E - 21 mm/ 0.82". (with removed support washer - 6 mm/0.23".).
G probes - 29 mm/1.14". (with removed support washer - 7 mm/0.27".).
The surface of the inspected product must be dry, free of scale, rust, dust, dirt and degreased.
In case of deviation of material properties of the controlled product from carbon structural steels (heat-resistant, corrosion-resistant, stainless, etc. steels, non-ferrous metals, and alloys, etc.), the measurements must be additionally guided by the requirements of 1.7 "Work with products that differ from carbon structural steel".
1.7 Measurement of specimens that differ in properties from carbon steel
The hardness tester's operation principle is based on the rebound method according to Leeb. Unlike static hardness testers (benchtop hardness testers), the measurement results
are influenced not only by the properties of the metal during plastic deformation but also by other physical and mechanical properties. Mostly - the modulus of elasticity of the controlled metal (Young's modulus). In hardness tester according to Leeb, material
characteristic curves are saved for the different metals.
The specific material characteristic curves of the various metal materials to be tested
may deviate from the pre-installed material characteristic curves under certain
circumstances. Deviations may occur in particular due to different heat treatments or alloys.
14
Page 16
1.8 Requirements for samples for additional adjustment of the hardness tester
At least 3 measurements are required for additional adjustment.From these 3 tests, the average is
calculated and set accordingly (adjustment).
The surface of the sample must fulfil the following requirements:
metallic bright without colour or scale coating
the surface roughness of the sample - is not worse than defined in the table 1.8.
the test surface must be even.
15
Page 17
When control specific samples it is necessary to be guided in addition by paragraph
1.10 "Measurements on light and thin-walled products".
1.9 Measurements on light and thin-walled products
Elimination of error on light and thin-walled products
In case of the insufficient thickness of a sample for the elimination of an additional error, it is necessary to couple a sample to a massive basic flat plate. Recommended hob parameters:
Weight and thickness - greater than the minimum mass and thickness of the controlled product specified in paragraph 1.6 "Requirements for the controlled product ".
The surface roughness should be approx. Ra 0.4 μm.
The modulus of elasticity of the plate metal (Young's modulus) is close to the modulus
of elasticity of the controlled product.
"Coupling" the product to the surface of the plate through a layer of coupling gel so that there are no, even small, stains of air layers between the surfaces of the product and the plate.
16
Page 18
plate.
Marking of the probe type
Minimum thickness, mm
( Coupled )
Minimum thickness, inch
( Coupled )
D
3
0.12
G
10
0.4 Е 3
0.12
UCI
1
0.04
Marking of the probe
type
Minimum weight, kg
(rigid support)
Minimum weight, lb.
(rigid support)
D 3 6.6 G 6
13.2
Е 3 6.6
UCI
0,1
0.2
It may be required to affix a workpiece which is not heavy enough to a hevier base
The thickness of the coupled sample - is according to table 1.9.
Table 1.9
The "Coupling" method should be used for thin products (in the form of sheets), products having a flat shape and a mass comparable to the minimum weight of the controlled product specified in paragraph 1.6 "Requirements for the controlled product".
It is necessary to affix small products of insignificant weight having "volume" form to a heavier base plate. The weight of the sample with using rigid support is in table 1.10
Table 1.10
In cases where the products have a small weight and/or thickness, these measures may not give the desired results. For control of such products, it is necessary to use hardness
17
Page 19
gauges of the static principle of action with the created small and ultralow loading.
1.10 Measurements on materials with high surface roughness
When measuring a product with a high surface roughness (above specified in paragraph 1.6 "Requirements for the controlled product"), there may be high additional variance in the readings of the device.
There are two ways to eliminate the additional deviation:
Finish the measuring surface (for example, using a grinder) to obtain the required roughness in the area defined in item 1.6 "Requirements for the controlled product".
If required, make some aditional measurements on initial adjustment (odd number) to
get a more accurate average value.
1.11 Influence of surface properties of coatings and hardened layers
Unlike devices of the static principle of action when carrying out measurements, the indenter of the probe of the hardness tester creates prints of small depth (see item 1.5 "H ardness tester probes", table 1.4). Therefore, the hardness of the surface layer of the product is measured directly.
On the controlled product in the manufacturing process may be thin surface layers that differ in hardness from the bulk of the metal. Example:
Carbon-free layer with low hardness formed due to high-temperature heat treatment (hardening, normalization, hot rolling, forging, etc.).
18
Page 20
Work hardening - can be formed in the surface layer after turning and milling, as well as rough grinding.
Hardening - the presence of such layers (the presence, in some cases, can be determined using a hardness tester) has a much greater impact on the readings of the hardness tester than the readings of static instruments.
The thickness of such layers usually does not exceed 0.2 mm / 0.4". To ensure the accuracy of the measurements, if any, they must be removed from the measuring area by grinding.
1.12 Hardness measurement of the reinforced surface layers
Depending on the type of probe used and the hardness of the product, the indenter creates prints of different depths (see section 1.5 "Probes of the hardness tester", table 1.4). Plastic deformation of the metal is formed at much greater depths.
As a rule, the hardness of the surface reinforced layers differs sharply from the hardness of the base metal of the product (for example, HDTV hardening). It is recommended to measure hardness in such cases when the thickness of the layer is not less than 1.5 mm.
To measure the hardness of reinforced surface layers, it is recommended to use only probes of types D and E, or specialized probes made for this purpose.
When measuring hardness, it is necessary to additionally take into account paragraph "1.7 Measurement of specimens that differ in properties from carbon steel."
The device corresponds to: ASTM A956 “Standard Test Method for Leeb Hardness
Testing of Steel Products”; ASTM A 1038 “Standard Test Method for Portable Hardness
Testing by the Ultrasonic Contact Impedance Method”.
19
Page 21
1.13 Delivery set of the hardness tester
1 x Hardness tester PCE-3000ULS
1 x UCI measuring probe
1 x Rebound probe type D
1 x charger
1 x USB cable
1 x Carrying case
1 x instruction manual
20
Page 22
1.14 Hardness tester design
Functionally, the hardness tester consists of an electronic unit and a probe.
The electronic unit of the hardness tester receives a signal from the electromagnetic coil of the probe, converts it into units of hardness, and displays the measurement results, statistical processing, and other functions of this hardness tester. The front panel of the electronic unit has a color graphic display and keyboard shown in Figure 1.1.
Figure 1.1 - The front panel of the electronic unit
21
Page 23
. A schematic representation of the electronic unit body is shown in Figure 1.2.
Figure 1.2 - Schematic representation of the body of the electronic unit
On the upper-end wall of the hardness tester, there is a button for turning on the hardness tester.
On the lower end wall of the hardness tester is a connector for connecting a probe and a cable for connecting to a computer.
On the rear panel of the hardness tester is a plate containing the serial number of the hardness tester.
Control of work of hardness tester - switching of scales, installations, creation of additional calibrations and scales, a record of results in memory, the analysis of results, etc., occurs on the display using the keyboard of the device.
To facilitate the use of the hardness tester, the electronic unit implements an interactive user interface that meets generally accepted standards used in modern computer technology.
22
Page 24
1.15 UCI Probe handling
The distance between the imprint center and the edge of the specimen or adjacent indentation should be at least 2.5 diagonal length imprint
Figure 1.3 – Setting the probe
Figure 1.4 – Charging the spring of the probe
The protective sleeve of UCI probe serves two purposes: It protects the UCI rod against damages (as distortion). Furthermore, during the measurements, it functions as a mechanical stop for the deflection of the rod. To carry out a measurement, the probe must be held perpendicular to the specimen surface. The Vickers diamond may touch the surface slightly, but not for too long (otherwise an error occurs). Then the probe is pressed to the specimen until the protective sleeve hits the surface. An acoustic signal indicates the completion of the measurement. To achieve an accurate measurement, the probe must be pressed steadily and vertically onto the specimen. The load is set by the built in the probe controlled load spring.
1.16 Rebound Leeb Probe handling
Set the probe onto the surface of the material at the zone of control (Fig. 1.3).
Hold the bottom part of the probe housing with one hand, and charge the spring of the probe by mowing the upper part of the housing down with the other hand (Fig. 1.4).
23
Page 25
Smoothly push the trigger button on the top of the probe (Fig. 1.5). Make sure that the
The minimal distance between the points of measuring should be not less than 3 mm/0.118". Do not repeat measurements on the same point, because repeated measurements give overestimated indications of the hardness of the product due to metal cold work in the imprint zone
probe does not move and secured to the surface of the controlled zone.
Figure 1.5 – Pushing the trigger button of the probe
After pushing the trigger button and the Impact Body hits the surface the measured hardness value will appear on the display.
24
Page 26
2. OPERATION OF HARDNESS TESTER
Before starting work, make sure that the device has no external damage and is ready for use.
2.1 Preparation for use
To get started - connect the probe and turn on the device.
The appliance is switched on by pressing the power button at front panel of the appliance for 1-2 seconds. After switching on the device, the following image will appear on the display (Fig. 2.1):
Figure 2.1 - Image on the display of the hardness tester when turning on the device
If the appliance does not switch on, check the battery charge. If the battery is low, charge the device.
To charge the battery, connect the charger cable connector to the USB type-C connector located on the bottom end of the device, and connect the charger to the AC mains with an operating voltage of ~ 220V and a frequency of 50 Hz.
25
Page 27
2.2 Hardness tester buttons
In the "Measurements" tab, the , , , buttons perform the following functions:
The button allows you to select the calibration scale for a given material.
The button allows you to select the measurement scale to which you want to convert values.
The button is used to select the direction of the probe.
The button is used to enter the number of measurements.
In other tabs, the buttons perform the following functions:
The button are used to move up / down the tabs of the hardness tester and enter data.
The buttons are used to move up / down the tabs of the hardness tester and enter data.
The button allows you to enter data.
The buttons are used to move left / right on the tabs of the hardness tester and data entry.
The button allows you to switch between the main tabs of the device and enter data.
The button allows you to enter data.
The button allows you to enter data.
The button allows you to make a space, put a period, comma, dash or enter the desired data.
The button is used to cancel the action, to delete one or more graphic characters, to return to the previous tab.
The button is used to select an item in the tab, submit a command, confirm an action or user readiness.
26
Page 28
2.3 Switching the tabs of the hardness tester
2.3.1 General description of hardness tester tabs items
There are 4 tabs in this hardness tester:
1) Measurements (Figure 2.1)
2) Archive (Fig. 2.2)
3) Calibration (Fig. 2.3)
Figure 2.2 – “Archive” tab
Figure 2.3 – “Calibration” tab
27
Page 29
4) Settings (Fig. 2.4)
Figure 2.4 - "Settings" tab
Select the desired tab of the hardness tester by briefly pressing the button . When pressed, the icons are highlighted in blue (Fig. 2.5).
Figure 2.5 - Select the desired item in the hardness tester tab
Once the desired tab is selected, you need to press the button to enter it.
Use the up / down or left / right arrows to select the desired items
in the tab, then press the login button or the cancel button .
28
Page 30
2.4 Measurement tab
In the measurements tab you can choose:
calibration on a given scale,
scales for measurement conversion,
probe directions,
a number of measurements in series.
2.4.1 Selection of calibration scale
To select the calibration according to a given scale, press the button , then the calibration selection window will open (Fig. 2.6)
Figure 2.6 - Calibration selection window
You can use the buttons to select standard or custom calibration. Use the
buttons to select the desired calibration value and press the button , then return to the "Measurements" tab. If calibration has not been selected, press the button to
return to the "Measurement" tab.
29
Page 31
2.4.2 Selecting the scale for measurement conversion
To select the scale to convert the measurement, press the button . The required
window will open (Fig. 2.7). Use the buttons to select the desired value of the
scale, then press the button . The values obtained will now be automatically converted
from HRC to the scale we selected. To return to the "Measurements" tab, click .
Figure 2.7 - Scale selection window for measurement conversion
2.4.3 Select the direction of the probe
To select the direction of the probe, press the button . In the window that opens (Fig.
2.8) use the buttons to select the desired direction, then press the button .
To return to the "Measurements" tab, click the button .
Figure 2.8 - Probe direction selection window
30
Page 32
2.4.4 Choice of number of measurements
To select the number of measurements, press the button . In the window, use the
buttons to enter the required number
of measurements, but not more than 20. Then press the button . To return to the
"Measurements" tab, click the button .
2.5 Archive tab
To enter the "Archive" tab, press the button , select the appropriate icon on the
screen of the hardness tester (Fig. 2.9) and press the button .
Figure 2.9 - Archive tab icon
In the window (Fig. 2.10) you can select previously saved measurements.
Figure 2.10 - "Archive" tab
31
Page 33
Measurements are sorted by date and time. Use the buttons to select the desired
date, and use the buttons to select the desired value of the time when the
measurement was made. After the desired value is selected, press the button . There will be an entrance to the tab with the saved values of measurements (fig. 2.11).
Then you can use the buttons to select the following values: "Graph" ­display of saved values in the form of a graph, "Table" - display of saved values in the form of a table, "Delete" - delete saved values from the device memory (Fig. 2.12).
Figure 2.11 - Entrance to the tab with saved measurement values
Figure 2.12 - Options for displaying saved measurement values
To enter the "Graph", "Table" or "Delete" tabs, select the desired value and press the
button . To go back, press the button .
32
Page 34
2.6 Calibration tab
To enter the "Calibration" tab, press the button to select the appropriate icon on
the screen of the hardness tester (Fig. 2.13) and press the button . After that, the "Calibration" tab will be entered (Fig. 2.14)
Figure 2.13 - Calibration tab icon
Figure 2.14 - "Calibration" tab
There are three ways to calibrate in this tab:
Standard (only for service technician of PCE Instruments!)
Custom (only for service technician of PCE Instruments!)
One point calibration
33
Page 35
2.6.1 Standard calibration
This function is only to be used by trained PCE Instruments service staff! This applies to chapters 2.6.1 to 2.6.3
Figure 2.15 - Service password entry window
34
Page 36
2.6.1.1 "Create scale" item
Use the buttons to select "Create scale" and click in the window that opens, the user will be prompted to enter the calibration name (Fig. 2.17).
Figure 2.17 - Enter the calibration name
Use the buttons to enter the
calibration name and press the button .
The next step is to select the hardness measurement scale (Fig. 2.18), use the buttons
to select the desired hardness measurement scale and press
Figure 2.18 - Choice of hardness measurement scale
35
Page 37
Then a window will open where you need to enter the number of hardness measures
for calibration (Fig. 2.19), use the buttons
to enter the desired number of measures and press .
Figure 2.19 - Enter the number of hardness measures for calibration
A window will appear (Fig. 2.20), where measurements should be made on calibration measures one after another. After carrying out several impacts on one of the measures of
hardness it is necessary to press the button and in a window (Fig. 2.21) by means buttons
to enter the specific value of a
measure and to press .Then you need to repeat the same steps on the remaining measures.
36
Page 38
Figure 2.20 - Calibration of the device on calibration measures
Figure 2.21 - Enter the specific value of the calibration measure
After performing these steps, a calibration completion message will be displayed.
In order to exit the process of creating a new calibration, you need to press the button
, a dialog box will appear (Fig. 2.22) where you need to use the buttons to
select "Yes" or "No" and press the button .
37
Page 39
Figure 2.22 - Exit the calibration mode
2.6.2 "Change scale" item
To change an already saved scale, use the buttons to select "Change scale"
and press . The following items will appear in the window (Fig. 2.23): "Adjustment", "Change name", "Restore settings".
Figure 2.23 - Scale change window
2.6.2.1 Calibration adjustment
To adjust the calibration, use the buttons to select the "Adjustment" item
and press the button (Fig. 2.24.
38
Page 40
In the window that opens (Fig. 2.25) - use the buttons to select the desired
calibration and press .
Figure 2.24 - Selecting the item "Adjustment"
Figure 2.25 - Select the desired calibration scale
Then, using the buttons in the
window (Fig. 2.26), enter the number of adjustment points and press . Then make
measurements as hardness, then press and adjust the value (Fig. 2.27).
39
Page 41
Figure 2.26 - Enter the number of points (measures of hardness) adjustment
Figure 2.27 - Value adjustment
Press , repeat the measurement on another sample, adjust the hardness value, and
press the button . The calibration will then be adjusted.
2.6.2.2 Change the calibration name
To change the name of the calibration, use the buttons , to select "Change
name" (Fig. 2.24). In the window (Fig. 2.28), use the buttons to select the name of
the calibration you want to change and press the button .
40
Page 42
Figure 2.28 - Calibration name change window
Use the buttons to enter the desired calibration name and press . Then, by pressing the button you can return to
the "Calibration" tab.
2.6.2.3 Restore settings
Select "Restore settings" (Fig. 2.24) and press . Then the tab with calibration will open (Fig. 2.29). Use the buttons , to select the desired scale and press , and the
message "Factory settings loaded" should appear.
To return to the Calibration tab, double-click the button .
Figure 2.29 - View the "Restore Settings" tab
41
Page 43
2.6.3 Delete scale item
To delete the calibration scale, use the buttons , to select "Delete scale" and
press the button (Fig. 2.16). In the tab that opens, select the scale you want to delete and
click (Fig. 2.30). After that the window of confirmation of removal of calibration will
open (fig. 2.31), buttons select "Yes" or "No" and press .
Figure 2.30 - View tab "Delete scale"
Figure 2.31 - Delete calibration
42
Page 44
2.7 Custom calibration
Figure 2.32 -
Calibration" tab
Figure 2.33 -
Password entry window
Figure 2.34 - Items in
the custom calibration
Custom calibration is intended for the the service technician of PCE Instruments
only to create their calibrations on their materials. This applies to chapters 2.7 to
2.73. The user can add new calibrations for different materials and adjust the
values of custom calibrations already created by him. To do this, use the buttons to select
"Custom" (Fig. 2.32), then press the button . A window for entering the password will
appear (Fig. 2.33), enter the password using the buttons
. After entering the password, a window will open with three options:
"Create scale", "Edit name" and "Delete scale" (Fig. 2.34).
2.7.1 "Create scale" item
Use the buttons to select "Create scale" and click on the window that opens, the user will be prompted to enter the calibration name (Fig. 2.35).
43
Page 45
Figure 2.35 - Enter the calibration name
Use the buttons o enter the
calibration name and press the button .
The next step is to select the hardness measurement scale (Fig. 2.36), use the buttons
to select the desired hardness measurement scale, and press .
Figure 2.36 - Choice of hardness measurement scale
44
Page 46
Then a window will open where you need to enter the number of hardness measures
for calibration (Fig. 2.37), use the buttons
to enter the desired number of measures, and press .
Figure 2.37 - Enter the number of measures to calibrate
A window will appear (Fig. 2.38), where measurements should be made on calibration measures one after another. After making several impacts on one of the measures, press the
button and in the window (Fig. 2.39) use the buttons
to enter the specific value of the measure and press . Then you need
to repeat the same steps on the remaining hardness measures.
Figure 2.38 - Calibration of the device on calibration measures
45
Page 47
Figure 2.39 - Enter the specific value of the calibration measure
After performing these steps, a calibration completion message will be displayed.
To exit the process of creating a new calibration, you need to press a button , on an
empty line of name or number, and a dialog box will appear (Fig. 2.40), where you
need to select "Yes" or "No" and press the button .
Figure 2.40 - Exit the calibration mode
46
Page 48
2.7.2 Edit name
Figure 2.41 - "Edit name" tab with user
calibration names
Figure 2.42 - Editing the calibration
username
To change the calibration user name, use the buttons to select "Edit name"
and press the button (Fig. 2.34). In the tab that opens (Fig. 2.41), use the buttons
to select the calibration whose name you want to change and press . A window will
open where you will be prompted to enter the calibration name (Fig. 2.42).
Use the buttons to enter a new
calibration name and press . To exit the calibration name, change mode without making
name changes, press the button , only the blinking cursor should remain, then press You will go to the "Edit name" tab without renaming the calibration.
47
Page 49
2.7.3 Delete scale item
Figure 2.43 - Items in the custom calibration
Figure 2.44 - View tab "Delete scale"
To delete a custom calibration scale, use the buttons , to select "Delete scale"
and press the button (Fig. 2.43). In the tab that opens, select the scale you want to delete
and click (Fig. 2.44).
After that the confirmation window of calibration removal will open (fig. 2.45), buttons
select "Yes" or "No" and press .
Figure 2.45 - Deleting the calibration
48
Page 50
2.8 One point calibration
If the device is calibrated to standard measures, for example, carbon steel 35, the readings on products made of similar steels, with the same heat treatment - will not differ much. But the values of the hardness of alloy steels and other alloys, which have a completely different alloy compared to standard measures - will differ significantly.
One point calibration is designed to adjust the standard calibration scale of a specific material for a similar material (alloy; mechanical properties).
The accuracy of the new scale from the specific measure to which the calibration was performed is 15-20%.
To perform one point calibration in the tab "Calibration" (Fig. 2.46) use the buttons
to select the item "One point" and press the button .
Figure 2.46 - "Calibration" tab, item "One point"
To access the calibration, you will be prompted to enter the user password (Fig. 2.47).
Use the buttons to enter the user
password and press . The password is 1111.
In the window that opens (Fig. 2.48), use the buttons to select an
existing calibration, Standard or Custom, and press .
49
Page 51
Figure 2.47 - Password entry window
Figure 2.48 - Calibration selection
You will then be prompted to enter a new calibration name. Use the buttons
button .
A window will appear (Fig. 2.49), where you should take a series of measurements.
to enter the new calibration name and press the
After a series of strikes press the button
the buttons
the event and press .
, and in the window that opens (Fig. 2.50) using
to enter the avg value of
50
Page 52
Figure 2.49 - One point calibration of the
device
Figure 2.50 - Enter the specific value of the
calibration measure
After performing these actions, a message will be displayed stating that the calibration has been adjusted, and it will be saved in the "Custom" section under a new name.
To exit the process of creating a calibration on one reading, you need to press the button
, a dialog box will appear (Fig. 2.51) where you need to use the buttons to
select "Yes" or "No" and press the button .
Figure 2.51 - Exit the One-point calibration mode
51
Page 53
2.9 Settings tab
Date and time
Save series
Probe type
Language
Brightness
To enter the "Settings" tab, press the button to select the appropriate icon on the
screen of the hardness tester (Fig. 2.52) and press the button . After that, enter to the "Settings" tab (Fig. 2.53).
Figure 2.52 - Settings tab icon
2.9.1 Description of the "Settings" tab items
Figure 2.53 - Settings tab
Change password - only for use by
PCE Instruments!
Sound settings
Shutdown
USB settings
Service menu - only for use by PCE
Instruments!
52
Page 54
2.9.2 Date and time
To enter the tab, use the buttons to select "Date and time" and press . A window will open (Fig. 2.54), where you need to set the date using the keyboard, then press
the button , then you need to set the time and press . The date and time will change signaling it with a sound signal.
Figure 2.54 - Tab window "Date and time»
2.9.3 Saving a series of measurements
This device allows you to enable or disable the automatic recording of a complete series of measurements.
To enter the tab, use the buttons to select "Saving the series" and press
. A window will open (Fig. 2.55), where you need to use the buttons , to select
"Yes" or "No", then press .
53
Page 55
Figure 2.55 - "Save the series" tab window
2.9.4 Probe type
On this tab, you can select from four different probe types: Type D probe, G probe, E probe, and UCI probe.
To enter the tab, use the buttons to select "Sensor type" and press . A
window will open (Fig. 2.56), where you need to use the buttons , to select the type of probe. Once the probe type is selected, press the button and the desired probe will be selected.
Figure 2.56 - "Sensor type" tab window
54
Page 56
2.9.5 Language
In this tab, you can choose English, German, Turkish, Spanish, Russian or Ukrainian
language.
To change the interface language, use the buttons to select "Language" and
press the button , in the window (Fig. 2.57), use the buttons to select the desired
language and press the button .
Figure 2.57 - "Language" tab window
2.9.6 Backlight brightness
In this tab, you can change the backlight brightness of the display.
Select "Brightness" with the buttons then press the button In the window that
opens (Fig. 2.58) use the buttons to increase or decrease the brightness.
Figure 2.58 - "Brightness" tab window
55
Page 57
2.9.7 Change password
This tab is only for use by PCE Instruments service personnel!
Figure 2.59 - "Password change" tab window
56
Page 58
2.9.8 Sound settings
In this tab, the user can enable or disable audible confirmation of actions.
To turn the audible confirmation on or off, use the buttons to select "Sound
Settings" and press the button . In the window (Fig. 2.62), use the buttons to
select "Yes" or "No" and press .
Figure 2.61 - "Sound settings" tab window
2.9.9 Auto-shutdown time
In this tab, the user can set the time to turn off the device. You can select items such as:
Off
1 minute
2 minutes
5 minutes
10 minutes
15 minutes
To set the auto-shutdown time, use the buttons to select "Auto-off time" and
press the button . In the window that opens (Fig. 2.62), select the desired item with the
buttons and press .
57
Page 59
Figure 2.62 - "Auto-shutdown time" tab window
2.9.10 USB settings
In this tab, the user can enable or disable the USB-C
Figure 2.63 - "USB Settings" tab window
port to connect to a PC.
To turn the USB-C port on or off, use the buttons to select "USB Settings" and
press the button . In the window (Fig. 2.63), use the buttons to select "Yes" or "No" and
press .
58
Page 60
2.9.11 Service menu
The service menu is only to be used by authorised PCE Instruments service staff! (Fig. 2.64).
Figure 2.64 - "Service menu" tab window
2.9.12 Statistics
This tab is only to be used by authorised PCE Instruments service personnel (Fig. 2.65).
Figure 2.65 - "Statistics" tab window
59
Page 61
2.9.13 About the device
In this tab, the user can view information about the device (Fig. 2.66). The user can see: the gain of the probe, the coefficient of standard deviation, the error after calibration at one point, and the firmware version.
Figure 2.66 - "About the device" tab window
2.11 Hardness tester packaging
To avoid mechanical damage to the cable and connectors of the device, it is necessary to disconnect the sensor from the device before packing it into the package.
60
Page 62
3. MAINTENANCE
Checking the technical condition of the hardness tester is carried out at least once
a year in the following sequence:
check the completeness of the hardness tester according to clause 1.13 "Delivery set of the hardness tester";
Carry out visual external inspection of the hardness tester, make sure that there are no mechanical damages to the electronic unit, the connector and the connecting cable;
No parts of the probes should have any signs of corrosion and mechanical damage.
Check performance;
compliance with the operating conditions;
battery charge level;
To eliminate the identified deficiencies you should contact PCE Instruments.
61
Page 63
4. TRANSPORTATION AND STORAGE
Packed instruments can be transported by any rail, road, sea, or air transport provided
the following conditions are met:
- transportation is carried out in factory packaging;
- there is no direct exposure to moisture;
- the temperature does not exceed -50 ° C to +50 ° C / -58 ° F to 122 ° F
- humidity does not exceed 95% at temperatures up to 35 ° С / 95 ° F;
- vibration in the range from 10 to 500 Hz with amplitude up to 0.35 mm and
acceleration up to 49 m / s2;
- impacts with a peak acceleration value of up to 98 m / s2;
- the devices placed in the transport are fixed to avoid falling and collision. In the case of air transport, the transport must be carried out in sealed heated
compartments.
To prevent moisture condensation inside the hardness tester when transporting it from
frost to a warm room, it is necessary to hold the device for 6 hours at room temperature.
The hardness tester is stored in a case indoors with an air temperature (25 ± 10) ° C,
relative humidity from 45 to 80%, and atmospheric pressure from 630 to 800 mm Hg/0,84 to 1,07 bar. The room should not have mold, acid fumes, reagents, paints, and other chemicals. Abrupt changes in temperature and humidity because of dew should not be allowed in the room.
For long-term storage, the device must be preserved, for which the electronic unit,
probe, charger and hardness measures, cleaned of dirt and oil, are placed in separate plastic bags and placed in separate pockets of the transportation bag of the device.
62
Page 64
5. PRECAUTIONS
A hardness tester is a technically complex measuring device that requires careful
handling. It must be protected from:
shocks, loads that can lead to mechanical damage to the hardness tester;
exposure to chemically aggressive environments;
ingress of liquids;
prolonged exposure to direct sunlight;
other influences which can cause damage to the functionality of the device.
The use of a hardness tester in the conditions of the sharp difference in temperatures is
not allowed. After transporting the device to the place of operation at a negative ambient temperature and placing it into a room with a positive temperature, it is necessary to keep the product in its packaging for at least 6 hours in order to avoid failure due to condensation of moisture.
Do not use batteries or chargers that are not approved by the manufacturer in the
hardness tester.
Dissection of the electronic unit and probes, and self-repair of the hardness tester is not
allowed.
63
Page 65
6. RECYCLING
The hardness tester does not contain in its design any dangerous or poisonous
substances that can harm human health or the environment and do not pose a threat to life, health of people and the environment at the end of their service life. In this regard, the recycling of the product can be made according to the rules for the disposal of general industrial waste. Recycling is carried out separately by groups of materials: plastic elements, metal fasteners.
The content of precious metals in the components of the product (electronic cards,
connectors, etc.) is extremely small, so it is not appropriate to produce their secondary processing.
Hardness tester batteries are disposed of in accordance with the current regulations for
the disposal of products.