PCE Instruments CAL-PCE-PA 8000, PCE-PA 8000, PCE-PA 8000-ICA Users guide

Page 1
SD card real time data logger
Your purchase of
this 3 PHASE
marks a step
forward for you into
t h e f i e l d o f
p r e c i s i o n
measurement.
A l t h o u g h t h i s
is a complex and
delicate instrument,
its durable structure
will allow many
years of use if
proper operating
tech niq ues are
developed. Please
read the following
i n s t r u c t i o n s
carefully and always
keep this manual
within easy reach.
3 PHASE POWER ANALYZER
Model : PCE-PA 8000
OPERATION MANUAL
Page 2
Caution Symbol
Caution :
* Risk of electric shock ! * During the measurement,
do not open the cabinet.
Caution :
* Do not apply the overload
voltage, current to the input terminal !
* Remove test leads before open
the battery cover !
* Cleaning - Only use the dry
cloth to clean the plastic case !
Environment Conditions
* Installation Categories III 600V.
* Pollution Degree 2.
* Altitude up to 2000 meters.
* Indoor use.
* Relative humidity 80% max.
Page 3
1.
FEATURES...............................................................................................
2. SPECIFICATIONS.................................................................................................
General Specifications......................................................................
Electrical Specifications..................................................
3.
FRONT PANEL DESCRIPTION.....................................................
MEASURING PREPARATION ......................................................
4-2 Entry the measurement Screen...............................
4-3 The summary description of keyboard...................
4-4 SETUP KEY description.........................................
4-5 Setting function description before measuring...........
MEASURING PROCEDURES ......................................................
measurement......................................................
measurement....................................................................
measurement....................................................................
measurement....................................................................
5-5 The CT and PT measurement.................................
5-6 ZERO adjustment for Watt Hour.................................
5-7 Data Logger function............................................
5-9 Backlight key........................................................
5-10 A ( Current ) Range key............................................................
46
5-11 LOWBAT screen............................................................
5-12 Appendix 1............................................................
MAINTENANCE.....................................................................................
6-1 Cleaning................................................................................
6-2 Replacement of batteries .................................................................................
RS232 PC SERIAL INTERFACE.............................................
the computer ( EXCEL software ) ...................................
PATENT.....................................................................
10. THE ADDRESS OF AFTER SERVICE CENTER ..................
2-1 2-2
TABLE OF CONTENTS
1 2 2 4 8
4.
4-1 The original screen.............................................. 10
5. 5-1 1Φ 2W ( one phase by two wires )
5-2 1Φ 3W ( one phase by three wires )
5-3 3Φ 3W ( three phase by three wires )
5-4 3Φ 4W ( three phase by four wires )
5-8 Data Hold function............................................ 45
10
10 12 13 14 35
35
36
38
39 41 42 43
46
48 49
6.
7.
8. Download the saving data from the SD card to
9.
50 50 50 51
53 57 58
Page 4
1. FEATURES
* Analysis for 3 phase multi-power system, 1P/2W,
1P/3W, 3P/3W, 3P/4W
* Voltage & Current are the True RMS value. * True Power ( KWMWGW ) measurement. * Apparent Power ( KVAMVAGVA ) measurement. * Reactive Power ( KVAR MVARGVAR) measurement. * Watt-Hour ( WHSHQHPFH ). * Power Factor( PF )Phase Angle( Φ ). * Voltage measurement range : 10 to 600 ACV * Current measurement range: 0.2A to 1200 ACA. * Programmable CT ratio (1 to 600) and PT ratio (1 to 1000). * ACV input impedance is 10 Mega ohms. * Safety Standard : IEC 1010, CAT III 600V * Built-in clock and Calendar, real time data record with
SD memory card , sampling time set from 2 to 7200 seconds. Just slot in the SD card into the computer, it can down load the all the measured value with the time information ( year/month/data/ hour/minute/second ) to the Excel directly, then user can make the further data analysis by themselves.
* Complete set with 4 PCs Test Leads, 4 PCs Alligator
clips, 3 PCs Clamp Probe, AC to DC 9V adapter, 2 G SD memory card and Carrying bag.
* Computer data output, can cooperate with USB Cable
/USB-01 RS232 cable/UPCB-02 and Data Acquisition software, SOFT-LUT-USB.
1
Page 5
2. SPECIFICATIONS
2-1 General Specifications:
Circuit Custom one-chip of microprocessor LSI
circuit
Display * LCD Size :
81.4 X 61 mm ( 3.2 X 2.4 inch )
* Dot Matrix LCD (320 X 240 pixels )
with back light.
Measurement * ACV
* ACA * AC WATT ( True Power )
AC WATT( Apparent Power )
AC WATT( Reactive Power ) * Power factor * Phase angle * Frequency
Wire 1P/2W, 1P/3W, 3P/3W, 3P/4W. connections Voltage ranges 10 ACV to 600 ACV, auto range. Current ranges 0.2 ACA to 1200 ACA,
auto range/manual range.
Safety IEC1010 CAT III 600 V. standard ACV input 10 Mega ohms. impedance Range select ACV Auto range.
ACA Auto range & manual range.
Clamp 40 Hz to 1 KHz. frequency response Spec. tested 45 to 65 Hz. frequency Over load ACV 720 ACV rms protection ACA 1300 ACA with clamp probe
CP-1200
2
Page 6
Over Indicator Show " OL ". Under Indicator Show " UR ". Data Hold Freeze the display reading. Data Record SD Card Record. Sampling Time Approx. 1 second. Power ON/OFF Manual OFF by push button. Real time * Real time data logger, saved the data data logger into SD memory card and down load
the all the measured value with the
time information ( year/month/data/
hour/minute/second ) down load
to the Excel * Integration time for data logger :
2 seconds to 7200 seconds, the during
of setting step are 2 seconds.
Data Output RS232 computer serial interface : USB/RS232 * Connect the optional USB cable
* Computer
interface
USB-01 will get the USB plug. * Connect the optional RS232 cable
UPCB-02 will get the RS232
plug.
Operating 0 to 50 ( 32 to 122 ). Temperature Operating Less than 80% R.H.. Humidity Power Supply * DC 1.5V, AA ( UM-3 ) Battery X 8 PCs
(Alkaline or heavy-duty battery). * AC to DC 9V power adapter.
Power * Meter : 300 DCmA. Consumption * Clamp : 34 DCmA. Clamp max. 50 mm ( 2.0 inch ) Dia. conductor Size
3
Page 7
Weight * Meter : 948g ( includes batteries )
Instruction manual...............
Test Leads (TL88-4AT)...........
Clamp Probe ( CP-1200 ).........
AC to DC 9V adapter.................
Carrying bag.........................
* Clamp ( includded cable ) : 467g
Dimension
Meter :
225 X 125 X 64 mm ( 8.86 X 4.92 X 2.52 inch )
Clamp :
210 X 64 X 33mm ( 8.3 X 2.5 X 1.3 inch ) Clamp Jaw : 86 mm (3.4 inch)- outside
Accessories * Included *
* Alligator clips (TL88-4AC) 1 Set (4 PCs) * *
* SD card ( 2 G ).................... 1 PC
*
1 PC 1 Set (4 PCs)
3 PCs 1 PC
1 PC
Optional * USB Cable , USB-01 Accessories * RS232 cable, UPCB-02
* Data Acquisition software,
SW-U811-WIN
2-2 Electrical Specifications:
ACV
Range Resolution Accuracy
10.0V to 600.0V 0.1V ± (0.5%+0.5V)
* Phase to neutral line
10.0V to 600.0V
* Phase to phase
4
Page 8
ACA
Range Resolution Accuracy
20A 0.001A,
0.01A,
200A 0.01A,
0.1A,
1200A 0.1A,
1A,
< 10 A
10 A
< 100 A
100 A
< 1000 A
1000 A
± (0.5%+0.1A)
± (0.5%+0.5A)
± (0.5%+5A)
Power factor
Range Resolution Accuracy
0.00 to 1.00 0.01 ± 0.04
Remark :
* PFH : Long term power factor * PFΣ :
For 3Φ 4W, 3Φ 3W
PFΣ = ( PF1 + PF2 + PF3 )/3
For 1Φ 3W
PFΣ = ( PF1 + PF2 )/2
Φ ( Phase angle )
Range Resolution Accuracy
-180° to 180° 0.1° ± 1°
* ACOS ( PF )
Frequency
Range Resolution Accuracy
45 to 65 Hz 0.1 Hz 0.1 Hz
5
Page 9
Active (Real) Power
Range Resolution Accuracy
0.000 to 9.999 KW *0.001/0.01/0.1 KW ± (1%+0.008KW)
10.00 to 99.99 KW *0.01/0.1 KW ± (1%+0.08KW)
100.0 to 999.9 KW 0.1 KW ± (1%+0.8KW)
1.000 to 9.999 MW 0.001 MW ± (1%+0.008MW)
* : The resolution is changed according the different ACA range.
Apparent Power
Range Resolution Accuracy
0.000 to 9.999 KVA *0.001/0.01/0.1KVA ± (1%+0.008KVA)
10.00 to 99.99 KVA *0.01/0.1 KVA ± (1%+0.08KVA)
100.0 to 999.9 KVA 0.1 KVA ± (1%+0.8KVA)
1.000 to 9.999 MVA 0.001 MVA ± (1%+0.008MVA)
* : The resolution is changed according the different ACA range.
Reactive Power
Range Resolution Accuracy
0.000 to 9.999 KVAR *0.001/0.01/0.1KVAR ± (1%+0.008 KVAR)
10.00 to 99.99 KVAR *0.01/0.1 KVAR ± (1%+0.08 KVAR)
100.0 to 999.9 KVAR 0.1 KVAR ± (1%+0.8 KVAR)
1.000 to 9.999 MVAR 0.001 MVAR ± (1%+0.008 MVAR)
* : The resolution is changed according the different ACA range.
6
Page 10
Watt Hour ( Active Power Hour) : WH
Range Resolution Accuracy
0.000 to 9.999 KWH 0.001 KWH ± (2%+0.008 KWH)
10.00 to 99.99 KWH 0.01 KWH ± (2%+0.08 KWH)
100.0 to 999.9 KWH 0.1 KWH ± (2%+0.8 KWH)
1.000 to 9.999 MWH 0.001 MWH ± (2%+0.008 MWH)
VA Hour ( Apparent Power Hour ) : SH
Range Resolution Accuracy
0.000 to 9.999 KVAH 0.001 KVAH ± (2%+0.008 KVAH)
10.00 to 99.99 KVAH 0.01 KVAH ± (2%+0.08 KVAH)
100.0 to 999.9 KVAH 0.1 KVAH ± (2%+0.8 KVAH)
1.000 to 9.999 MVAH 0.001 MVAH ± (2%+0.008 MVAH)
VAR Hour ( Reactive Power Hour ) : QH
Range Resolution Accuracy
0.000 to 9.999 KVARH 0.001 KVARH ± (2%+0.008 KVARH)
10.00 to 99.99 KVARH 0.01 KVARH ± (2%+0.08 KVARH)
100.0 to 999.9 KVARH 0.1 KVARH ± (2%+0.8 KVARH)
1.000 to 9.999 MVARH 0.001 MVARH ± (2%+0.008 MVARH)
7
Page 11
3. FRONT PANEL DESCRIPTION
8
Page 12
3-1 Display 3-2 1Φ 3Φ ( Phase/wire ) key button 3-3 key button 3-4 key button 3-5 Hold key button 3-6 Backlight key button 3-7 Power key button 3-8 Exit key button 3-9 REC key button 3-10 A ( current ) range key button 3-11 Shift key button 3-12 Setup key button 3-13 Voltage input terminals 3-14 Current probe input sockets 3-15 SD card socket 3-16 RS232 socket 3-17 Reset button 3-18 DC 9V power adapter socket 3-19 Battery Cover/Battery compartment 3-20 Stand 3-21 Current Sense Jaw 3-22 Trigger 3-23 Plug of current probe
9
Page 13
4. MEASURING PREPARATION
1)
2)
4-1 The original screen
4-2 Entry the measurement Screen
The bottom right display of screen 1 will show as " SD Check " along with blinking while inserting SD CARD then disappears after several seconds that indicates the data from SD CARD has been read completed. The bottom right display of screen 2 will show as " NO DISK " along with blinking when SD CARD is not inserted.
10
Page 14
screen 1 ( 4-2 )
QΣ :
QΣ :
V12: 0.0 V V1: 0.0 V A1: 0.00 A V23: 0.0 V V2: 0.0 V A2: 0.00 A V31: 0.0 V V3: 0.0 V A3: 0.00 A
P1: -0.000 KW S1: 0.000KVA Q1: -0.000 KVAR P2: -0.000 KW S2: 0.000KVA Q2: -0.000 KVAR P3: -0.000 KW S3: 0.000KVA Q3: -0.000 KVAR
PΣ : -0.000 KW SΣ : 0.000KVA PF1: -0.00 PF 2: -0.00 PF 3: -0.00 PFΣ : 0.00 PF H: 0.00 Φ 1: - 0.0° Φ 2: - 0.0° Φ 3: - 0.0°
WH: 0.000 KWH SH: 0.000KVAH QH: 0.000 KVARH FREQ: 0.0 Hz
AUTO SD 20A 3Φ 4W SEC: 2 CT: 1 PT: 1 Check
screen 1 ( 4-2 )
V12: 0.0 V V1: 0.0 V A1: 0.00 A V23: 0.0 V V2: 0.0 V A2: 0.00 A V31: 0.0 V V3: 0.0 V A3: 0.00 A
P1: -0.000 KW S1: 0.000KVA Q1: -0.000 KVAR P2: -0.000 KW S2: 0.000KVA Q2: -0.000 KVAR P3: -0.000 KW S3: 0.000KVA Q3: -0.000 KVAR
-0.000 KVAR
PΣ : -0.000 KW SΣ : 0.000KVA PF1: -0.00 PF 2: -0.00 PF 3: -0.00 PFΣ : 0.00 PF H: 0.00 Φ 1: - 0.0° Φ 2: - 0.0° Φ 3: - 0.0°
WH: 0.000 KWH SH: 0.000KVAH QH: 0.000 KVARH FREQ: 0.0 Hz
AUTO NO 20A 3Φ 4W SEC: 2 CT: 1 PT: 1 DISK
11
-0.000 KVAR
Page 15
4-3 The summary description of keyboard
1)
2)
3)
Press the key to change AUTO RANGE to MANUAL
4)
5)
6)
7)
8)
9)
POWER KEY ( 3-7, Fig. 1 ) : Press the key to turn the instrument ON/OFF. 1Φ 3Φ ( phase/wire ) KEY ( 3-2, Fig. 1 ) : Press the key to select (1P/2W1P/3W3P/3W3P/4W) measurement function mode. A ( current ) RANGE KEY ( 3-10, Fig. 1 ) :
RANGE mode for current. REC KEY ( 3-9, Fig. 1 ) : The data record key for SD CARD. HOLD KEY ( 3-5, Fig. 1 ) : Press the key to freeze the display reading. BACKLIGHT KEY ( 3-6, Fig. 1 ) : Press the key to switch LCD backlight to ON/OFF. SETUP KEY ( 3-12, Fig. 1 ) : Press the key to setup the function before measuring. EXIT KEY ( 3-8, Fig. 1 ) : Press the key to exit setting screen. SHIFT KEY ( 3-11, Fig. 1 ) Press the key to set the different functions in setting screen.
10) UP ( ) KEY ( 3-3, Fig. 1 ) : Press the key to move the cursor up in setting
screen.
11) DOWN ( ) KEY ( 3-4, Fig. 1 ) : Press the key to move the cursor down in setting screen.
12
Page 16
4-4 SETUP KEY description:
File Name:
3P401001.XLS
SHIFT 1
REC Date: 2008-11-28 00:03:17
Sampling Time: 2
Delet File:
0%SD Format:
0%Use Size:
388 KB
Decimal: Basic
screen 1 ( 4-4 )
Free Size:
1946 MB
Clamp Type: 1200A
Total Size:
1946 MB
RS232 Out Sel:
PT:
1 : 1
V1I1P1
CT:
1 : 1
S1Q1PF1
Beep:
ON
Φ1
WH
FREQ
Year
Month
Date
Hour
Minute
Second
2008
1205111518
File Name:
3P401001.XLS
SHIFT 2
REC Date: 2008-11-28 00:03:17
Sampling Time: 2
Delet File:
0%SD Format:
0
%
screen 2 ( 4-4 )
Use Size:
388 KB
Decimal: Basic
Free Size:
1946 MB
Clamp Type: 1200A
Total Size:
1946 MB
RS232 Out Sel:
PT:
1 : 1
V1I1P1
CT:
1 : 1
S1Q1PF1
Beep:
ON
Φ1
WH
FREQ
Year
Month
Date
Hour
Minute
Second
2008
1205111518
4-4-1 SHIFT KEY
* SHIFT 1 : When the symbols " SETUP " and " SHIFT 1 "
are appeared on up right display of screen 1 in the meantime, and then use the or to select the expect item.
* SHIFT 2 : When the symbols " SETUP " and " SHIFT 2 "
are appeared on up right display of screen 2 in the mean time, and then use the or to select ( 1P/2W 1P/3W3P/3W3P/4W ) in File Name function.
Folder Name: WTA01 SETUP
Folder Name: WTA01 SETUP
13
Page 17
4-4-2 The Setting Function menu
* Folder Name : Set the expect folder name for SD CARD,
the range is between WTA01 and WTA10.
* File Name: Set the file name for SD CARD, It allows setting
50 filenames in this function.
* REC Date: Show the recorded time of existing files
( Year/Month/Date, Hour/Min./Sec. ) * Sampling Time : Set the sampling time from 2 to 7200 seconds. * Delete File : To delete the existing data from SD CARD. * SD Format : to Format SD CARD fast. * PT : Set the potential transformer from 1 to 1000. * CT : Set the current transformer from 1 to 600. * Beep : Set to ON/OFF for buzzer. * Clamp Type : Select the Clamp Type to 200A or 1200A. * RS232 out Sel. : Set RS232 output function, maximum up
to nine items can be selected to output. screen 1 screen 2. * Year : Set the year. * Month : Set the month. * Date : Set the date. * Hour : Set the hour. * Minute : Set the minute. * Second : Set the second.
4-5 Setting function description before measuring
Press SETUP KEY to enter setting function screen, the selected item will be displayed in highlight.
14
Page 18
4-5-1 Folder Name: Set the folder name for SD
Delet File:
SD Format:
Delet File:
SD Format:
screen 1 ( 4-5-1 )
Folder Name: WTA01 SETUP
File Name: 3P401001.XLS REC Date: 2008-11-28 00:03:17 Sampling Time: 2
0 %
0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 14 49
screen 2 ( 4-5-1 )
Folder Name: WTA01 SETUP
File Name: 3P401001.XLS SHIFT 1 REC Date: 2008-11-28 00:03:17 Sampling Time: 2
0 %
0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 14 34
15
Page 19
A : Folder Name range: WTA01 to WTA10.
Delet File:
SD Format:
B : Press or to select the expect folder number, the number consists of " 01 to 10 " (as screen 1). C : Press or continuously at least two seconds can skip the numbers faster. D : Press SHIFT KEY once, the symbol " SHIFT1" will
appear on up right display, and then press to entry next setting function as screen 2 (Folder Name File Name).
4-5-2 File Name: Set the file name for SD
A : The screen will show " NO File " indicator in REC Date option when the selected file is new ( as screen 1 ). B : The screen will show recording date and time in REC
Date option when the selected file has been recorded as screen 2.
screen 1 ( 4-5-2 ) Folder Name: WTA03 SETUP
File Name: 3P401001.XLS
REC Date: NO File Sampling Time: 2
0 %
0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 15 10 55
16
Page 20
screen 2 ( 4-5-2 )
Delet File:
SD Format:
Folder Name: WTA01 SETUP
File Name: 3P401001.XLS
REC Date: 2008-11-28 00:03:17 Sampling Time: 2
0 %
0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 15 31
C : File Name description : press or in screen 2 to
select expect file number from 001 to 050.
Remark : When press or > 2 sec, the setting
no. will change fast.
* 1P201001 : 1P2 means one phase by two wires, 01
means folder number, 001 means file number.
* 1P301001: 1P3 means one phase by three wires,
01 means folder number, 001 means file number.
* 3P301001 : 3P3 means three phases by three wires,
01 means folder number, 001 means file number.
* 3P401001 : 3P4 means three phases by four
wires, 01 means folder number, 001 means file number.
D : The up right display will show " SHIFT1 " symbol
while pressing SHIFT KEY once in screen 2, and then press to enter next setting function as screen 3 ( File Name Sampling Time ).
17
Page 21
E : The up right display will show " SHIFT2 " symbol
Delet File:
SD Format:
Delet File:
SD Format:
while pressing SHIFT KEY again in screen 4, at this time press or to select 1P/2W(1P2)1P/3W(1P3) 3P/3W(3P3) and 3P/4W(3P4) as screen 4.
F: One by one to press SHIFT KEY to select different
functions circularly.
Folder Name: WTA01 SETUP
File Name: 3P401001.XLS SHIFT 1
REC Date: 2008-11-28 00:03:17 Sampling Time: 2
0 % screen 3
0 % (4-5-2) Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 15 06
Folder Name: WTA01 SETUP
File Name: 3P401001.XLS SHIFT 2
REC Date: 2008-11-28 00:03:17 Sampling Time: 2 screen 4
0 % ( 4-5-2 )
0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 15 18
18
Page 22
4-5-3 Sampling time: Set the data logger sampling time
screen 1 ( 4-5-3 )
Folder Name:
WTA01
SETUP
File Name:
3P401001.XLS
SHIFT 1
REC Date: 2008-11-28 00:03:17
Sampling Time:
2
Delet File:
0%SD Format:
0%Use Size:
388 KB
Decimal: Basic
Free Size:
1946 MB
Clamp Type: 1200A
Total Size:
1946 MB
RS232 Out Sel:
PT:
1 : 1
V1I1P1
CT:
1 : 1
S1Q1PF1
Beep:
ON
Φ1
WH
FREQ
Year
Month
Date
Hour
Minute
Second
2008
1205111551
screen 2 ( 4-5-3 )
Folder Name:
WTA01
SETUP
File Name:
3P401001.XLS
REC Date: 2008-11-28 00:03:17
Sampling Time:
2
Delet File:
0%SD Format:
0%Use Size:
388 KB
Decimal: Basic
Free Size:
1946 MB
Clamp Type: 1200A
Total Size:
1946 MB
RS232 Out Sel:
PT:
1 : 1
V1I1P1
CT:
1 : 1
S1Q1PF1
Beep:
ON
Φ1
WH
FREQ
Year
Month
Date
Hour
Minute
Second
2008
1205111601
for SD
A : When press SHIFT KEY once, the symbol " SHIFT1 " will disappear on up right display, at this time press
or to adjust expect sampling time as screen 2, adjusting numbers are from 2 to 7200 seconds.
Remark : When press or > 2 sec, the setting no. will change fast.
B : The up right display will show " SHIFT1 " symbol
while pressing SHIFT KEY again, and then press to enter next setting function (Sampling Time Delete File)
19
Page 23
4-5-4 Delete File: Delete the files for SD
SD Format:
A : The indicator " Y or N " will appear on right side
display of the option while pressing SHIFT KEY continuously at least two seconds, and now press the display will show " Y " in highlight as screen 2, press SETUP KEY again to confirm, the selected file (ex: 3P401001.XLS) will be erased then return to screen 1, or else press SETUP KEY in " Y or N " option to return to screen 1.
B : Press in screen 1 to enter next setting function
(Delete File SD Format)
screen 1 ( 4-5-4 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS SHIFT 1 REC Date: 2008-11-28 00:03:17 Sampling Time: 2
Delete File: 0 %
0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 16 20
20
Page 24
screen 2 ( 4-5-4 )
SD Format:
Folder Name: WTA01 SETUP File Name: 3P401001.XLS SHIFT 1 REC Date: 2008-11-28 00:03:17 Sampling Time: 2
Delete File: Y OR N
0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 16 45
4-5-5 SD Format : Formatting function for SD CARD
A : The indicator " Y or N " will appear on right side
display of the option while pressing SHIFT KEY continuously at least two seconds, and press the display will show " Y " in highlight as screen 2, press SETUP KEY again to confirm to format SD CARD then return to screen 1, or else press SETUP KEY in " Y OR N " option return to screen 1.
B : Press in screen 1 to enter next setting function ( SD
Format PT ).
21
Page 25
screen 1 ( 4-5-5 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS SHIFT 1 REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 %
SD Format: 0 %
Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 17 05
screen 2 ( 4-5-5 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS SHIFT 1 REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 %
SD Format: Y OR N
Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 17 20
22
Page 26
4-5-6 PT: Set the Potential Transformer
A : When press SHIFT KEY once, the symbol " SHIFT1 "
will disappear as screen 2 at this time press or can adjust to expect PT values, the adjusting numbers are from 1 to 1000.
Remark : When press or > 2 sec, the setting
no. will change fast.
B : Press SHIFT KEY once again will return to screen 1
then press to enter next setting function ( PT CT ).
screen 1 ( 4-5-6 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS SHIFT 1 REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 % SD Format: 0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1
CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 17 53
23
Page 27
screen 2 ( 4-5-6 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 % SD Format: 0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1
CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 19 07
4-5-7 CT: Set the Current Transformer
A : When press SHIFT KEY once, the symbol " SHIFT1 "
will disappear as screen 2 at this time press or can adjust to expect CT values, the adjusting numbers are from 1 to 600.
Remark : When press or > 2 sec, the setting
no. will change fast.
B : Press SHIFT KEY once again will return to screen 1
then press to enter next setting function ( CT BEEP ). again will return to screen 1 then press to enter next setting function ( CT BEEP ).
24
Page 28
screen 1 ( 4-5-7 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS SHIFT 1 REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 % SD Format: 0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1
CT: 1 : 1 S1 Q1 PF1
Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 19 20
screen 2 ( 4-5-7 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 % SD Format: 0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1
CT: 1 : 1 S1 Q1 PF1
Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 19 30
25
Page 29
4-5-8 Beep: Control the buzzer to ON/OFF
A : When press SHIFT KEY once the symbol " SHIFT1 " will
disappear as screen 2, at this time press or to control the buzzer to ON/OFF.
B : Press SHIFT KEY once again will return to screen 1 then
press to enter next setting function ( BEEP Decimal type )
screen 1 ( 4-5-8 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS SHIFT 1 REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 % SD Format: 0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1
Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 19 44
26
Page 30
screen 2 ( 4-5-8 )
disappear as screen 2, at this time press or to select
Folder Name: WTA01 SETUP File Name: 3P401001.XLS REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 % SD Format: 0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1
Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 19 58
4-5-9 Decimal Type: set the Decimal type to Basic ( .) or Euro ( , )
The numerical data structure of SD card is default used the " . " as the decimal, for example "20.6" "1000.53" . But in certain countries ( Europe ...) is used the " , " as the decimal point, for example " 20,6 " "1000,53". Under such situation, it should change the Decimal character at first.
A : When press SHIFT KEY once the symbol " SHIFT1 " will
the Decimal type to " Basic " or " Euro ".
27
Page 31
* Basic type :
The numerical data structure of SD card
is default used the " . " as the decimal,
for example "20.6" "1000.53" .
* Euro type :
The numerical data structure of SD card
is default used the " , " as the decimal,
for example "20,6" "1000,53" .
B : Press SHIFT KEY once again will return to screen 1 then
press to enter next setting function ( Decimal type
Clamp type ).
screen 1 ( 4-5-9 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS SHIFT 1 REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 % SD Format: 0 % Use Size: 388 KB Decimal : Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 20 18
28
Page 32
screen 2 ( 4-5-9 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 % SD Format: 0 % Use Size: 388 KB Decimal : Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 20 18
4-5-10 Clamp Type: set the clamp type to 200A or 1200A
A : When press SHIFT KEY once the symbol " SHIFT1 " will
disappear as screen 2, at this time press or to select the CT type to 200A or 1200A. B : Press SHIFT KEY once again will return to screen 1 then
press to enter next setting function ( CLAMP TYPE
RS232 OUT SEL ).
29
Page 33
screen 1 ( 4-5-10 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS SHIFT 1 REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 % SD Format: 0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 20 18
screen 2 ( 4-5-10 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 % SD Format: 0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 19 44
30
Page 34
4-5-11 RS232 Out Sel setting
A : When press SHIFT KEY continuously at least two
seconds as screen 2 and now press or to select the item that intend to output, maximum up to nine items, when the cursor stops on the selected item and then press SHIFT KEY again, the selected item will be
displayed in highlight. B : If the selected items are over nine, the low right display will show indicator " full " as screen 3. C : After the selecting is completed, press SHIFT KEY
continuously at least two seconds again will return to screen 1 and show all the selected items at the same time.
D : Press in screen 1 to enter next setting function
( RS232 Out Sel Year )
screen 1 ( 4-5-11 ) Folder Name: WTA01 SETUP File Name: 3P401001.XLS SHIFT 1 REC Date: 2008-11-28 00:03:17 Sampling Time: 2 Delete File: 0 % SD Format: 0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ1 WH FREQ
Year Month Date Hour Minute Second 2008 12 05 11 20 43
31
Page 35
screen 2 ( 4-5-11 )
Φ 1
Φ 1
Φ 1
Φ 1
RS232 OUTPUT SELECT
1. V12 12. P3 23. PF2
2. V23 13. 24. PF3
3. V31 14. S1 25. PFΣ
4. V1 15. S2 26. PFH
5. V2 16. S3 27. Φ 1
6. V3 17. 28.
7. I1 18. Q1 29.
8. I2 19. Q2 30. WH
9. I3 20. Q3 31. SH
10. P1 21. 32. QH
11. P2 22. PF1 33. FREQ
2 3
screen 3 ( 4-5-11 )
RS232 OUTPUT SELECT
1. V12 12. P3 23. PF2
2. V23 13. 24. PF3
3. V31 14. S1 25. PFΣ
4. V1 15. S2 26. PFH
5. V2 16. S3 27. Φ 1
6. V3 17. 28.
7. I1 18. Q1 29.
8. I2 19. Q2 30. WH
9. I3 20. Q3 31. SH
10. P1 21. 32. QH
11. P2 22. PF1 33. FREQ
2 3
FULL
32
Page 36
4-5-12 Year/Month/Date/Hour/Minute/Second setting
A : When press SHIFT KEY once the symbol " SHIFT1"
will disappear as screen 2, at this time press or
to adjust expect numbers, and press or
continuously at least two seconds can skip the
numbers faster. B : Press in screen 1 to enter next setting function ( Year Month ). C : The settings about ( Month Date ), (Date
Hour ), (Hour Minute ), ( Minute Second ) are same as above step A and step B. D : In this setting function ( Year Minute ), press or
in addition to adjust the numbers, and the setting
value will also be saved during the adjusting.
E : In the function of setting " second ", press or
to adjust numbers. at this point the number of second is at a standstill condition and then press setup key that will save setting value and also start counting function of " second ".
Folder Name: WTA01 SETUP File Name: 3P401001.XLS SHIFT 1 REC Date: 2008-11-28 00:03:17 Sampling Time: 2 screen 1 Delete File: 0 % ( 4-5-12 ) SD Format: 0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ 1 WH FREQ
Year Month Date Hour Minute Second
2008 12 05 12 02 13
33
Page 37
Folder Name: WTA01 SETUP File Name: 3P401001.XLS REC Date: 2008-11-28 00:03:17 Sampling Time: 2 screen 2 Delete File: 0 % ( 4-5-12 ) SD Format: 0 % Use Size: 388 KB Decimal: Basic Free Size: 1946 MB Clamp Type: 1200A Total Size: 1946 MB RS232 Out Sel:
PT: 1 : 1 V1 I1 P1 CT: 1 : 1 S1 Q1 PF1 Beep: ON Φ 1 WH FREQ
Year Month Date Hour Minute Second
2008 12 05 12 02 28
4-5-13 When all settings are completed, press EXIT KEY to return measuring screen.
4-5-14 The descriptions about SD CARD memory space
A : Use Size - To show the space data numbers that have been used. B : Free Size - To show the data numbers of balance space. C : Total Size - To show the data numbers of total space. D : Typical SD CARD and SDHC both can be used with the
instrument, except the SD CARD memory size is less
than 32MB.
4-5-15 RESET KEY : Press this key to reboot the
instrument
34
Page 38
5. MEASURING PROCEDURES
SCREEN 1 ( 5-1 )
5-1 1Φ 2W ( one phase by two wires ) measurement
A : Diagram
B : Operation Instructions:
B-1 : Power on the instrument by pressing POWER KEY, and then press 1Φ 3Φ KEY to select the 1Φ 2W system, the selected name of system will be appeared on bottom left display of screen 2. B-2 : Connect the line voltage L1, Vn ( Neutral ) to V1 and N terminals of the instrument. B-3: Place the conductor of CP-1200 (A1)to A1 as screen 1. B-4: Connect the output of clamp meter " CP-1200(A1) " to A1 terminal of the instrument. B-5: The related measuring factors will be appeared on display, about the instruction of factor please refer appendix 1 ( 5-11, page 49 ).
35
Page 39
Φ 1 :
**
CT: 1
PT: 1
screen 2 ( 5-1 ) V 1 : 0.0 V A 1 : 0.00 A
P 1 : - 0.000KW P F 1 : - 0.00 S 1 : 0.000KVA P F H : 0.00 Q 1 : - 0.000KVAR
W H : 0.000KWH S H : 0.000KVAH Q H : 0.000KVARH F R E Q : 50.1 Hz
AUTO
20A 1Φ 2W SEC: 2
: - 0.0°
1
1
5-2 1Φ 3W (one phase by three wires) measurement
A : Diagram
screen1 ( 5-2 )
36
Page 40
B : Operation Instructions:
50.0
CT: 1
PT: 1
B-1 : Power on the instrument by pressing POWER KEY, and then press 1Φ 3Φ KEY to select the 1Φ 3W system, the selected name of system will be appeared on bottom left display of screen 2. B-2 : Connect the line voltage L1, L2 and Vn (Neutral) to V1, V2 and N terminals of the instrument. B-3 : Place the conductor of CP-1200(A1), CP-1200(A2) hook to A1 and A2 as screen 1. B-4 : Connect the outputs of clamp meter CP-1200(A1)CP-1200(A2) to A1 and A2 terminals of the instrument. B-5 : The related measuring factors will be appeared on display, about the instruction of factor please refer appendix 1 ( 5-11, page 49 ).
screen 2 ( 5-2 ) V 1 : 0.0 V P 1 : - 0.000KW V 2 : 0.0 V P 2 : - 0.000KW A 1 : 0.00 A S 1 : 0.000KVA A 2 : 0.00 A S 2 : 0.000KVA Q 1 : - 0.000KVAR Q 2 : - 0.000KVAR
PΣ : 0.000 KW SΣ : 0.000 KVA QΣ : 0.000 KVAR PF1: - 0.00 PF2: - 0.00 PFΣ : 0.00 PFH: 0.00 Φ1: - 0.0° Φ2: - 0.0°
WH: 0.000 KWH SH: 0.000 KVAH QH: 0.000 KVARH FREQ:
AUTO
20A 1Φ 3W SEC: 2
Hz
1
1
37
Page 41
5-3 3Φ 3W (three phases by three wires)
, CP-1200(A2), CP-1200(A3) to A1, A2, A3 terminals
measurement
A : Diagram
screen 1 ( 5-3 )
B : Operation Instructions:
B-1 : Power on the instrument by pressing POWER KEY, and then press 1Φ 3Φ KEY to select the 3Φ 3W system, the selected name of system will be appeared on bottom left display of screen 2. B-2: Connect the line voltage L1, L2 and L3 to V1, V2 and V3 terminals of the instrument. B-3: Place the conductor of CP-1200(A1), CP-1200(A2), CP-1200(A3) hook to A1, A2 ,A3 as screen 1. B-4 : Connect the outputs of clamp meter CP-1200(A1)
of the instrument. B-5: The related measuring factors will be appeared on display, about the instruction of factor please refer appendix 1 ( 5-11, page 49 ).
38
Page 42
screen 2 ( 5-3 )
50.0
CT: 1
PT: 1
screen 1 ( 5-4 )
V 1 2 : 0.0 V A 1 : 0.00 A V 2 3 : 0.0 V A 2 : 0.00 A V 3 1 : 0.0 V A 3 : 0.00 A
P Σ : - 0.000 KW S Σ : 0.000 KVA Q Σ : 0.000 KVAR PFΣ : 0.00 P F H : 0.00
WH: 0.000 KWH SH: 0.000 KVAH QH: 0.000 KVARH FREQ:
AUTO
20A 3Φ 3W SEC: 2
Hz
1
5-4 3Φ 4W (three phases by four wires)
measurement
A : Diagram
1
39
Page 43
B: Operation Instructions:
:
CT: 1
PT: 1
B-1 : Power on the instrument by pressing POWER KEY, and then press 1Φ 3Φ KEY to select the 3Φ 4W system, the selected name of system will be appeared on bottom left display of screen 2. B-2 : Connect the line voltage L1, L2, L3 and Vn to V1, V2, V3 and N terminals of the instrument. B-3 : Place the conductor of CP-1200(A1), CP-1200(A2) CP-1200(A3) hook to A1, A2, A3 as screen 1. B-4 : Connect the outputs of clamp meter CP-1200(A1), CP-1200(A2), CP-1200(A3) to A1A2A3 terminals of the instrument. B-5 : The related measuring factors will be appeared on display, about the instruction of factor please refer appendix 1 ( 5-11, page 49 ).
screen 2 ( 5-4 )
V12: 0.0 V V1: 0.0 V A1: 0.00 A V23: 0.0 V V2: 0.0 V A2: 0.00 A V31: 0.0 V V3: 0.0 V A3: 0.00 A
P1: - 0.000 KW S1: 0.000 KVA Q1: - 0.000 KVAR P2: - 0.000 KW S2: 0.000 KVA Q2: - 0.000 KVAR P3: - 0.000 KW S3: 0.000 KVA Q3: - 0.000 KVAR
- 0.000 KW SΣ : 0.000 KVA QΣ : - 0.000 KVAR
PF1: - 0.00 PF2: - 0.00 PF3: - 0.00 PFΣ : 0.00 PFH: 0.00 Φ 1: - 0.0° Φ2: - 0.0° Φ 3: - 0.0°
WH: 0.000 KWH SH: 0.000 KVAH QH: 0.000 KVARH FREQ: 0.0 Hz
AUTO
20A 3Φ 4W SEC: 2
1
1
40
Page 44
5-5 The CT and PT measurement
A : Diagram
screen 1 ( 5-5 )
B : Operation Instructions
B-1 : Power on the instrument by pressing POWER KEY, and then press 1Φ 3Φ KEY to select the 3Φ 4W system, the selected name of system will be appeared on bottom left display of screen 2. B-2 : Connect the line voltage L1, L2, L3 and Vn to V1, , V3 and N terminals of the instrument. B-3 : Place the conductor of CP-1200(A1), CP-1200(A2), CP-1200(A3) hook to A1, A2 , A3 as screen 1. B-4: Connect the outputs of clamp meter CP-1200(A1),
CP-1200(A2), CP-1200(A3) to A1, A2, A3 terminals of the instrument. B-5: The related measuring factors will be appeared on display, about the instruction of factor please refer appendix 1 ( 5-11, page 49 ).
41
Page 45
screen 2 ( 5-5 )
:
CT: 1
PT: 1
V12: 0.0 V V1: 0.0 V A1: 0.00 A V23: 0.0 V V2: 0.0 V A2: 0.00 A V31: 0.0 V V3: 0.0 V A3: 0.00 A
P1: - 0.000 KW S1: 0.000 KVA Q1: - 0.000 KVAR P2: - 0.000 KW S2: 0.000 KVA Q2: - 0.000 KVAR P3: - 0.000 KW S3: 0.000 KVA Q3: - 0.000 KVAR
- 0.000 KW SΣ : 0.000 KVA QΣ : - 0.000 KVAR
PF1: - 0.00 PF2: - 0.00 PF3: - 0.00 PFΣ : 0.00 PFH: 0.00 Φ 1: - 0.0° Φ2: - 0.0° Φ 3: - 0.0°
WH: 0.000 KWH SH: 0.000 KVAH QH: 0.000 KVARH FREQ: 0.0 Hz
AUTO
20A 3Φ 4W SEC: 2
1
1
5-6 ZERO adjustment for Watt Hour
If reset the " Exit key button " ( 3-8, Fig. 1 ) continuously and > 6 seconds, the measurement value of " WH ", " SH ", " QH " will reset to Zero value.
42
Page 46
5-7 Data Logger Function
:
CT: 1
PT: 1
A : Press REC KEY once to start the data record function.
A-1 : If the bottom right shows as " Change Card ", it indicates the memory space is already full either or the SD CARD exist some wrong. A-2 : If the SD CARD is normal, the data logger function will start to be executed.
V12: 0.0 V V1: 0.0 V A1: 0.00 A V23: 0.0 V V2: 0.0 V A2: 0.00 A V31: 0.0 V V3: 0.0 V A3: 0.00 A
P1: - 0.000 KW S1: 0.000 KVA Q1: - 0.000 KVAR P2: - 0.000 KW S2: 0.000 KVA Q2: - 0.000 KVAR P3: - 0.000 KW S3: 0.000 KVA Q3: - 0.000 KVAR
- 0.000 KW SΣ : 0.000 KVA QΣ : - 0.000 KVAR
PF1: - 0.00 PF2: - 0.00 PF3: - 0.00 PFΣ : 0.00 PFH: 0.00 Φ 1: - 0.0° Φ2: - 0.0° Φ 3: - 0.0°
WH: 0.000 KWH SH: 0.000 KVAH QH: 0.000 KVARH FREQ: 0.0 Hz
AUTO Change
20A 3Φ 4W SEC: 2
1
1 Card
B : The bottom right display will show the recorded data
points. B-1 : Each file can record up to 30,000 data points as screen 1 when the record points exceed 30,000 points, system will create a new file automatically. (For example, WTA01001.XLS will be replaced by WTA01002.XLS) B-2 : While pressing REC KEY twice, the data logger function will stop to execute, the record points will disappear on bottom right display as screen 2.
43
Page 47
screen 1 ( 5-6 B )
:
CT: 1
PT: 1
:
CT: 1
PT: 1
V12: 0.0 V V1: 0.0 V A1: 0.00 A V23: 0.0 V V2: 0.0 V A2: 0.00 A V31: 0.0 V V3: 0.0 V A3: 0.00 A
P1: - 0.000 KW S1: 0.000 KVA Q1: - 0.000 KVAR P2: - 0.000 KW S2: 0.000 KVA Q2: - 0.000 KVAR P3: - 0.000 KW S3: 0.000 KVA Q3: - 0.000 KVAR
- 0.000 KW SΣ : 0.000 KVA QΣ : - 0.000 KVAR
PF1: - 0.00 PF2: - 0.00 PF3: - 0.00 PFΣ : 0.00 PFH: 0.00 Φ 1: - 0.0° Φ2: - 0.0° Φ 3: - 0.0°
WH: 0.000 KWH SH: 0.000 KVAH QH: 0.000 KVARH FREQ: 0.0 Hz
AUTO REC
20A 3Φ 4W SEC: 2
1
1 9
screen 2 ( 5-6 B )
V12: 0.0 V V1: 0.0 V A1: 0.00 A V23: 0.0 V V2: 0.0 V A2: 0.00 A V31: 0.0 V V3: 0.0 V A3: 0.00 A
P1: - 0.000 KW S1: 0.000 KVA Q1: - 0.000 KVAR P2: - 0.000 KW S2: 0.000 KVA Q2: - 0.000 KVAR P3: - 0.000 KW S3: 0.000 KVA Q3: - 0.000 KVAR
- 0.000 KW SΣ : 0.000 KVA QΣ : - 0.000 KVAR
PF1: - 0.00 PF2: - 0.00 PF3: - 0.00 PFΣ : 0.00 PFH: 0.00 Φ 1: - 0.0° Φ2: - 0.0° Φ 3: - 0.0°
WH: 0.000 KWH SH: 0.000 KVAH QH: 0.000 KVARH FREQ: 0.0 Hz
AUTO
20A 3Φ 4W SEC: 2
1
1
44
Page 48
5-8 Data HOLD Function
screen 1 ( 5-7 )
V12:
0.0VV1:
0.0VA1:
0.00
A
V23:
0.0VV2:
0.0VA2:
0.00
A
V31:
0.0VV3:
0.0VA3:
0.00
A
P1:
-
0.000 KW
S1:
0.000 KVA
Q1:
-
0.000 KVAR
P2:
-
0.000 KW
S2:
0.000 KVA
Q2:
-
0.000 KVAR
P3:
-
0.000 KW
S3:
0.000 KVA
Q3:
-
0.000 KVAR
:
-
0.000 KW
SΣ :
0.000 KVA
QΣ :
-
0.000 KVAR
PF1:
-
0.00
PF2:
-
0.00
PF3:
-
0.00
PFΣ :
0.00
PFH:
0.00
Φ 1:
-
0.0°
Φ 2:
-
0.0°
Φ 3:
-
0.0°
WH:
0.000 KWH
SH:
0.000 KVAH
QH:
0.000 KVARH
FREQ:
0.0HzAUTO
20A
3Φ 4W
SEC: 2
CT: 1
1
PT: 1
1
HOLD
screen 2 ( 5-7 )
V12:
0.0VV1:
0.0VA1:
0.00
A
V23:
0.0VV2:
0.0VA2:
0.00
A
V31:
0.0VV3:
0.0VA3:
0.00
A
P1:
-
0.000 KW
S1:
0.000 KVA
Q1:
-
0.000 KVAR
P2:
-
0.000 KW
S2:
0.000 KVA
Q2:
-
0.000 KVAR
P3:
-
0.000 KW
S3:
0.000 KVA
Q3:
-
0.000 KVAR
:
-
0.000 KW
SΣ :
0.000 KVA
QΣ :
-
0.000 KVAR
PF1:
-
0.00
PF2:
-
0.00
PF3:
-
0.00
PFΣ :
0.00
PFH:
0.00
Φ 1:
-
0.0°
Φ 2:
-
0.0°
Φ 3:
-
0.0°
WH:
0.000 KWH
SH:
0.000 KVAH
QH:
0.000 KVARH
FREQ:
0.0
Hz
AUTO
20A
3Φ 4W
SEC: 2
CT: 1
1
PT: 1
1
A: During the measurement, press HOLD KEY once, the bottom right display will show “ HOLD symbol as screen 1. B: Press the HOLD KEY twice will disable the Data HOLD function and the “ HOLD” symbol will disappear in the meantime
45
Page 49
5-9 BACKLIGHT KEY
:
CT: 1
PT: 1
screen 1 ( 5-9 )
Control the backlight function of LCD to ON/OFF
5-10 A ( Current ) RANGE KEY function
Change AUTO RANGE to MANUAL RANGE
A : One by one to press A RANGE KEY will entry to different ranges as following screens in sequence. Range 200A (as show on bottom left display of screen 1) to 1200A (as screen 2) to 20A (as screen 3) B : When hold A RANGE KEY down at least for 2 seconds will change MANUAL to AUTO as show on bottom left display of screen 4.
V12: 0.0 V V1: 0.0 V A1: 0.00 A V23: 0.0 V V2: 0.0 V A2: 0.00 A V31: 0.0 V V3: 0.0 V A3: 0.00 A
P1: - 0.000 KW S1: 0.000 KVA Q1: - 0.000 KVAR P2: - 0.000 KW S2: 0.000 KVA Q2: - 0.000 KVAR P3: - 0.000 KW S3: 0.000 KVA Q3: - 0.000 KVAR
- 0.000 KW SΣ : 0.000 KVA QΣ : - 0.000 KVAR
PF1: - 0.00 PF2: - 0.00 PF3: - 0.00 PFΣ : 0.00 PFH: 0.00 Φ 1: - 0.0° Φ2: - 0.0° Φ 3: - 0.0°
WH: 0.000 KWH SH: 0.000 KVAH QH: 0.000 KVARH FREQ: 0.0 Hz
MANU
200A 3Φ 4W SEC: 2
1
1
46
Page 50
V12: 0.0 V V1: 0.0 V A1: 0.00 A
:
CT: 1
PT: 1
screen 2 ( 5-9 )
:
CT: 1
PT: 1
screen 3 ( 5-9 )
V23: 0.0 V V2: 0.0 V A2: 0.00 A V31: 0.0 V V3: 0.0 V A3: 0.00 A
P1: - 0.000 KW S1: 0.000 KVA Q1: - 0.000 KVAR P2: - 0.000 KW S2: 0.000 KVA Q2: - 0.000 KVAR P3: - 0.000 KW S3: 0.000 KVA Q3: - 0.000 KVAR
- 0.000 KW SΣ : 0.000 KVA QΣ : - 0.000 KVAR
PF1: - 0.00 PF2: - 0.00 PF3: - 0.00 PFΣ : 0.00 PFH: 0.00 Φ 1: - 0.0° Φ2: - 0.0° Φ 3: - 0.0°
WH: 0.000 KWH SH: 0.000 KVAH QH: 0.000 KVARH FREQ: 0.0 Hz
MANU
1200A 3Φ4W SEC: 2
1
1
V12: 0.0 V V1: 0.0 V A1: 0.00 A V23: 0.0 V V2: 0.0 V A2: 0.00 A V31: 0.0 V V3: 0.0 V A3: 0.00 A
P1: - 0.000 KW S1: 0.000 KVA Q1: - 0.000 KVAR P2: - 0.000 KW S2: 0.000 KVA Q2: - 0.000 KVAR P3: - 0.000 KW S3: 0.000 KVA Q3: - 0.000 KVAR
- 0.000 KW SΣ : 0.000 KVA QΣ : - 0.000 KVAR
PF1: - 0.00 PF2: - 0.00 PF3: - 0.00 PFΣ : 0.00 PFH: 0.00 Φ 1: - 0.0° Φ2: - 0.0° Φ 3: - 0.0°
WH: 0.000 KWH SH: 0.000 KVAH QH: 0.000 KVARH FREQ: 0.0 Hz
MANU
20A 3Φ 4W SEC: 2
1
1
47
Page 51
V12:
0.0VV1:
0.0VA1:
0.00
A
V23:
0.0VV2:
0.0VA2:
0.00
A
V31:
0.0VV3:
0.0VA3:
0.00
A
P1:
-
0.000 KW
S1:
0.000 KVA
Q1:
-
0.000 KVAR
P2:
-
0.000 KW
S2:
0.000 KVA
Q2:
-
0.000 KVAR
P3:
-
0.000 KW
S3:
0.000 KVA
Q3:
-
0.000 KVAR
:
-
0.000 KW
SΣ :
0.000 KVA
QΣ :
-
0.000 KVAR
PF1:
-
0.00
PF2:
-
0.00
PF3:
-
0.00
PFΣ :
0.00
PFH:
0.00
Φ 1:
-
0.0°
Φ 2:
-
0.0°
Φ 3:
-
0.0°
WH:
0.000 KWH
SH:
0.000 KVAH
QH:
0.000 KVARH
FREQ:
0.0HzAUTO
20A
3Φ 4W
SEC: 2
CT: 1
1
PT: 1
1
screen 4 ( 5-9 )
V12:
0.0VV1:
0.0VA1:
0.00
A
V23:
0.0VV2:
0.0VA2:
0.00
A
V31:
0.0VV3:
0.0VA3:
0.00
A
P1:
-
0.000 KW
S1:
0.000 KVA
Q1:
-
0.000 KVAR
P2:
-
0.000 KW
S2:
0.000 KVA
Q2:
-
0.000 KVAR
P3:
-
0.000 KW
S3:
0.000 KVA
Q3:
-
0.000 KVAR
:
-
0.000 KW
SΣ :
0.000 KVA
QΣ :
-
0.000 KVAR
PF1:
-
0.00
PF2:
-
0.00
PF3:
-
0.00
PFΣ :
0.00
PFH:
0.00
Φ 1:
-
0.0°
Φ 2:
-
0.0°
Φ 3:
-
0.0°
WH:
0.000 KWH
SH:
0.000 KVAH
QH:
0.000 KVARH
FREQ:
0.0HzAUTO
LOWBAT
20A
3Φ 4W
SEC: 2
CT: 1
1
PT: 1
1
5-11 The LOWBAT screen: as show on lower right
display of the following screen.
48
Page 52
5-12 Appendix 1
* V12, V23, V31 : Line Voltage
* V1, V2, V3 : Phase Voltage
* A1, A2, A3 : Line Current
* P1, P2, P3 : True Power of each phase. (W)
* S1, S2, S3 : Apparent Power of each phase. (VA)
* Q1, Q2, Q3 : Reactive Power of each phase (VAR)
* PΣ : Total True Power (W)
* SΣ : Total Apparent Power (VA)
* QΣ : Total Reactive Power (VAR)
* PF1, PF2, PF3 : Power Factor of each phase
* PFΣ : Total Power Factor
* PFH : Long Term Average Power Factor ( WH/SH )
* Φ 1, Φ 2, Φ 3 : Phase Angle of each phase
* WH : Watt Hour
* SH : Apparent Power Hour
* QH : Reactive Power Hour
* 1Φ 2W : One phase by two wires
* 1Φ 3W : One phase by three wires
* 3Φ 3W : Three phases by three wires
* 3Φ 4W : Three phases by four wires
* SEC : The sampling time of data logger
* CT : Current transformer
* PT : Potential transformer
49
Page 53
6. MAINTENANCE
1)
2)
open the " Battery Cover " ( 3-19, Fig. 1 ) away from the
3)
4)
Caution :
Remove test leads before
opening the battery cover
or housing case !
6-1 Cleaning
Caution :
Cleaning - Only use
the dry cloth to clean
the plastic case !
6-2 Replacement of batteries
When Display show the " LOWBAT " indicator ( ref. 5-10 page 48 ), it should change the batteries.
instrument and remove the battery. Replace with batteries ( DC 1.5V, AA/UM-3 battery X 8 PCs ) and reinstate the cover.
* When install the batteries, should make
attention the battery polarity.
Make sure the battery cover is secured after changing the batteries.
50
Page 54
7. RS232 PC SERIAL OUTPUT
Center Pin............................................
Ground/shield...........................................
The instrument is provided an 3.5 mm dia. phone socket ( 3-16, Fig. 1 ) for RS232 computer interface socket.
The connector output is a 16 digits data stream which can be utilized to the user's specific application.
A RS232 lead with the following connection will be required to link the instrument with the PC serial input.
Meter PC
(3.5 mm jack plug) (9W 'D" Connector)
Pin 4
Pin 2
2.2 K resister
Pin 5
The 16 digits data stream will be displayed in the following format :
D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0
51
Page 55
Each digit indicate the following status :
D15 Start Word D14 4 D13 1 D12 & D11 Annunciator for Display
31=HZ C0 = MW D1 = GW/Hr 32=DEGREE C1 = GW D2 = TW/Hr 48=K WATT C2 = TW D3 = KVA/Hr 50=ACV C3 = MVA D4 = MVA/Hr 52=ACA C4 = GVA D5 = GVA/Hr 64=KVA C5 = TVA D6 = TVA/Hr 65=KW/HR C6 = KVAR D7 = KVAR/Hr B6 = KACV C7 = MVAR D8 = MVAR/Hr B7 = MACV C8 = GVAR D9 = GVAR/Hr B8 = KACA C9 = TVAR E0 = TVAR/Hr B9 = MACA D0 = MW/Hr
D10 Polarity
0 = Positive 1 = Negative
D9 Decimal Point(DP), position from right to the left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D8 to D1 Display reading, D1 = LSD, D8 = MSD
For example : If the display reading is 1234, then D8 to D1 is :
00001234
D0 End Word
RS232 setting
Baud rate 9600 Parity No parity Data bit no. 8 Data bits Stop bit 1 Stop bit
52
Page 56
8. Download the saving data from the
1)
2)
3)
software screen ( for example as following EXCEL data
SD card to the computer ( EXCEL software )
After execute the Data Logger function, take away the SD card out from the " SD card socket " ( 3-15, Fig. 1 ). Plug in the SD card into the Computer's SD card slot ( if your computer build in this installation ) or insert the SD card into the " SD card adapter ". then connect the " SD card adapter " into the computer. Power ON the computer and run the " EXCEL software ". Down load the saving data file ( for example the file name : 3P401001.XLS, 1P201001.XLS, 1P301001.XLS,
3P301001.XLS........) from the SD card to the
computer. The saving data will present into the EXCEL
screens ) , then user can use those EXCEL data to make the further Data or Graphic analysis usefully.
EXCEL data screen 1 ( for example )
53
Page 57
EXCEL data screen 2 ( for example )
EXCEL data screen 3 ( for example )
EXCEL data screen 4 ( for example )
54
Page 58
EXCEL data screen 5 ( for example )
EXCEL data screen 6 ( for example )
EXCEL graphic screen 1 ( for example )
55
Page 59
EXCEL graphic screen 2 ( for example )
EXCEL graphic screen 3 ( for example )
56
Page 60
EXCEL graphic screen 4 ( for example )
EXCEL graphic screen 5 ( for example )
Loading...