PCE Health and Fitness PCE-BSK Series, PCE-BSK 310, PCE-BSK 1100, PCE-BSK 5100 User Manual

© PCE Instruments
Precision Balance PCE-BSK Series
User Manual
Last change: 18 December 2019
v1.0
English
© PCE Instruments
Contents
1 Safety notes ........................................................................................... 1
2 Technical specifications ...................................................................... 2
3 Delivery contents .................................................................................. 2
4 Device description ................................................................................ 3
5 First use of the balance ........................................................................ 4
6 Calibration ............................................................................................. 4
6.1 One-point calibration ........................................................................................................ 4
6.2 Three-point calibration ...................................................................................................... 5
7 Function settings .................................................................................. 5
7.1 Setting the weight unit ...................................................................................................... 5
7.2 Counting function ............................................................................................................. 6
7.3 User-specific functions ..................................................................................................... 6
8 Serial interface - communication protocol ......................................... 7
9 Troubleshooting .................................................................................... 7
10 Warranty ................................................................................................. 8
11 Disposal ................................................................................................. 8
© PCE Instruments
1
1 Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.
The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
The instrument may only be used if the environmental conditions (temperature, relative
humidity, …) are within the ranges stated in the technical specifications. Do not expose
the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
Do not expose the device to shocks or strong vibrations.
The case should only be opened by qualified PCE Instruments personnel.
Never use the instrument when your hands are wet.
You must not make any technical changes to the device.
The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner,
no abrasives or solvents.
The device must only be used with accessories from PCE Instruments or equivalent.
When weighing powder and similar products, use a suitable container to make sure no
small parts enter the scales.
Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
Do not use the instrument in explosive atmospheres.
Do not expose the device to radioactive environments or corrosive substances.
The measurement range as stated in the specifications must not be exceeded under
any circumstances.
Non-observance of the safety notes can cause damage to the device and injuries to the user.
Do not use any sharp objects such as pens to press the keys of the balance. Only use your fingers to press the keys.
We do not assume liability for printing errors or any other mistakes in this manual. We expressly point to our general guarantee terms which can be found in our general terms of
business. If you have any questions please contact PCE Instruments. The contact details can be found at
the end of this manual.
© PCE Instruments
2
2 Technical specifications
PCE-BSK 310
PCE-BSK 1100
PCE-BSK 5100
Weighing range
310g
1,100g
5,100g
Accuracy class
II
III
III
Resolution/readability
0.001g
0.01g
0.1g
Repeatability
±0.002g
±0.02g
±0.2g
Linearity
±0.003g
±0.03g
±0.3g
Settling time
<3 s
Operating temperature
17.5 ... 22.5 °C
Dimensions of weighing plate
90 mm
130 mm
160x160 mm
Calibration
external
Overall dimensions
270 x 265 x 190 mm
270 x 200 x 80 mm
Overall weight
4.0 kg
2.5 kg
Interface
RS232
Mains adaptor
Primary: 100 ... 240 V, 50 / 60 Hz
Secondary: 6 V, 500 mA
outside minus, inside plus
Batteries
3 x AA (LR6)
not included, not rechargeable within the balance
3 Delivery contents
PCE-BSK 310
PCE-BSK 1100
PCE-BSK 5100
Electronic balance
Weighing plate
90 mm
130 mm
160 x 160 mm
Mains adaptor • •
User manual • •
Test weight
200 g
500 g
---
Wind shield • ---
---
only PCE-BSK 310 includes wind shield
Loading...