AB series balances are destined for high accuracy weighing in laboratory
practice.Balances are equipped with internal calibration system for proper accuracy
control during operation. The user should also own weight standard of OIML E2 class
for periodical control of the balance (weight value stated in Technical Data sheet) available separatelly. All balances are metrologically tested - calibration or legal
verification certificate on demand.Legal verification is required for balances used in
some applications: direct sale, pharmaceutical prescriptions, medical and
pharmaceutical analysis, goods packing and others.Balances with legal verification are
mark with the following:
- protective seal placed on the casing mounting screw at the back of the
balance,
- calibration switch protective seal,
- notified body stamps and green metrological marking placed on the balance name
Type
Capacity (Max) 100 g 200 g
Min load (Min) 100 mg 100 mg
Reading unit (d) 0,1mg 0,1mg
Verification unit (e) 1mg 1mg
Tare range -100 g -200 g
Accuracy class I
Working
temperature
Weighing time <5 s
Pan dimension
Balance base dim. 210 x 510 x 350 mm
Balance weight external feeder.ZN 12V/ 500 mA: AC: 230 V 50 Hz 6VA / DC:
To avoid electrical shock or damage of the
balance or connected peripheral devices, it is
necessary to follow the security rules below.
• All repairs and necessary regulations can be made by authorised
personnel only.
• To avoid fire risk use a feeder of an appropriate type (supplied with the
balance). Pay attention that supply voltage is compatible with specified
technical data.
• Do not use the balance when its cover is opened.
• Do not use the balance in explosive conditions.
• Do not use the balance in high humidity.
• If the balance seems not to operate properly, unplug it from the mains
and do not use until checked by authorised service.