* Monitor with real time data logger, save the measuring
data along the time information ( year, month, date,
hou, minute, second ) into the SD memory card and can be
down load to the Excel, extra software is no need. user
can make the further data or graphic analysis by
themselves.
* Show Humidity, Temperature and Barometer value in the
same LCD.
* 0.1 %RH resolution for the humidity reading,
0.1 degree resolution for the Temp. reading.
* Used the precision capacitance type humidity sensor,
professional and high accuracy for humidity measurement.
* Professional barometer, unit : hPa, mmHg, inHg.
* SD card capacity : 1 GB to 16 GB.
* Large LCD display, easy readout.
* Microcomputer circuit, high reliability.
* The circuit design are deigned to consider low power
consumption and long battery life when use battery
power.
* DC 1.5V ( UM-4, AAA ) battery x 6 PCs or DC 9V
adapter in.
* RS232/USB computer interface.
* Patented
2. SPECIFICATIONS
2-1 General Specifications
CircuitCustom one-chip of microprocessor LSI
circuit.
DisplayLCD size : 60 mm x 50 mm
MeasurementHumidity, Temperature and Barometer
Memory CardSD memory card, 1 GB to 16 GB.
1
Page 4
Datalogger5/10/30/60/120/300/600 seconds
* AC/DCpoweradapterisoptional.
* AC/DCpoweradapterisoptional.
Sampling Time or Auto.
*
Default sampling time is 60 seconds.
*
The " Auto " sampling . means when
the measuring value is changed ( >
± 1 %RH or > ± 1 or > ± 1℃
hPa ) will save the data one time only.
Data error no. 0.1 % no. of total saved data typically.≦
Advanced * SD memory card Format
setting* Set clock time ( Year/Month/Date,
Hour/Minute/ Second )
* Set sampling time value
* Set beep sound ON/OFF.
* Decimal point of SD card setting
* Set temperature unit to or ℃℉
* Set RS232 out ON/OFF.
* Set the unit of Barometer
Update TimeApprox. 1 second if measuring data
of Displayis changed.
Data OutputRS 232/USB PC computer interface.
* Connect the optional RS232 cable
UPCB-02 will get the RS232 plug.
* Connect the optional USB cable
USB-01 will get the USB plug.
Operating 0 to 50 .℃
Temperature
Operating Less than 90% R.H.
Humidity
Power Supply
Alkaline or heavy duty DC 1.5 V battery
( UM4, AAA ) x 6 PCs, or equivalent.
DC 9V adapter input. ( AC/DC power
adapter is optional ).
Battery lifeIf use the new battery ( alkaline type )
and sampling time set to 60 seconds,
the battery life will be > one month
typically.
2
Page 5
Weight285 g/0.63 LB.
Instruction manual..............................
Dimension132 x 80 x 32 mm
( 5.2 x 3.1 x 1.3 inch )
Accessories
1 PC
IncludedHanging unit ( with sticker )..................
OptionalSD Card ( 2 GB )
AccessoriesAC to DC 9V adapter.
USB cable, USB-01.
RS232 cable, UPCB-02.
Data Acquisition software, SW-U801-WIN.
Range0 to 50 ,32 to 122 .℃℃℉℉
Resolution0.1 degree
Accuracy - 0.8 .℃℃
- 1.5 .℉℉
Barometer
Measurement RangeResolutionAccuracy @ reading
hPa10.0 to 1000.0 0.1
1001 to 1100 1
mmHg
inHg
225.0 to 825.0
8.85 to 32.48
3
0.1
0.01
±2 hPa (at 750.1...1100.0 hPa)
±3 hPa (at 300.0...750.0 hPa)
± 2.0 mmHg
± 0.2 inHg
Page 6
3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display
3-2 Logger button, Enter button
3-3 button, Time button▲
3-4 button▼
3-5 SET button
3-6 Humidity/Temp./Barometer sensor
3-7 Hanging holes
3-8 Stand
3-9 Battery cover/Battery compartment
3-10 Screw for battery cover
3-11 Reset button
3-12 RS-232 output terminal
3-13 SD card socket
3-14 DC 9V power sdapter input socket
3-15 Hanging unit ( with sticker )
4
Page 7
4. MEASURING PROCEDURE
*
How to change the Temp. unit ( , ), refer to chapter 7-6,℃ ℉
page 14.
*
How to change the Barometer unit ( hPa, mmHg, inHg ), refer to
chapter 7-8, page 15.
5
1)Install the batteries into the battery compartment :
* Loose the " Screw of the battery cover " ( 3-10, Fig. 1 )
and take away the " Battery Cover " ( 3-9, Fig. 1 ) from
the meter.
* Replace with DC 1.5 V battery ( UM4/AAA, Alkaline/Heavy
duty type ) x 6 PCs, and reinstate the cover.
* Make sure the battery cover is secured after changing
the battery.
2)The " Display " ( 3-1, Fig. 1 ) will show both the
Humidity, Temperature and Barometer value that sensing from
Humidity/Temp./Barometer sensor " ( 3-6, Fig. 1 ).
Remark :
5. DATALOGGER
5-1 Preparation before execute datalogger function
a. Insert the SD card
Prepare a " SD memory card " ( 1 GB to 16 GB, optional ),
insert the SD card into the " SD card socket " ( 3-13, Fig. 1)
with the correct direction exactly.
* It recommend use memory card's capacity is 4 GB.≦
b. SD card Format
If SD card just the first time use into the meter, it
recommend to make the " SD card Format " at first,
please refer chapter 7-1 ( page 11 ).
* It recommend strongly, do not use memory cards that have been
formatted by other meter or by other installation ( such as camera...)
Reformat the memory card with your meter.
* If the SD memory card exist the trouble during format by the meter,
use the Computer to reformat again can fix the problem.
Page 8
c. Time setting
If the meter is used at first time, it should to adjust the
clock time exactly, please refer chapter 7-2 ( page 11 ).
d. Decimal format setting
The numerical data structure of SD card is
default used the " . " as the decimal, for
example "20.6" "1000.53" . But in certain
countries ( Europe ...) is used the " , " as the
decimal point, for example " 20, 6 "
"1000,53". Under such situation, it should
change the Decimal character at first, details
of setting the Decimal point, refer to Chapter
7-5, page 13.
e. 3 Information of LCD display
* If the Display show :
It means that the SD card exist the problem
CH-or the SD card memory is full, it should
CArdchange SD memory card.
* If the Display show :
It means that the battery is low voltage.
LobAtUnder such condition, the Datalogger
function is disable.
* If the Display show :
It means that the SD card is not plugged
nointo the meter.
CArd
6
Page 9
5-2 Datalogger
*
*
a. Start the datalogger
Press the " Logger button ( 3-2, Fig. 1 ) > 2 seconds
continuously, until the Display show the indicator "
DATALOGGER ", release the " Logger Button " ( 3-2, Fig.
1 ), then the measuring data along the time information
will be saved into the memory circuit.
Remark :
How to set the sampling time, refer to Chapter 7-3,
page 12.
How to set the beeper sound is enable, refer to
Chapter 7-4, page 12.
c.. Finish the Datalogger
During execute the Datalogger function ( Display show the
" Datalogger " indicator ), press the " Logger button " ( 3-2,
Fig. 1 ) > 2 seconds continuously, until the Display
indicator " DATALOGGER " is disappeared, release the
" Logger Button " will finish the Datalogger function.
Before take away the SD card from the
meter, it should execute the
procedures of " Finish the Datalogger ",
otherwise some existing recent save
data may loss.
5-3 Check time information
Press " Time button " ( 3-3, Fig. 1 ) > 2 seconds
continuously, the LCD display will present the time
information of Year/Month/Date, Hour/Minute/Second and
the sampling value.
7
Page 10
5-4 SD Card Data structure
1)WhentheSDcardisusedintothemeter,theSDcard
2)
3)
4)
When the first time, the SD card is used into the meter,
the SD card will generate a folder :
HBA01
If the first time to execute the Datalogger, under the
route HBA01\, will generate a new file name
HBA01001.XLS.
After exist the Datalogger, then execute again, the data
will save to the HBA01001.XLS until Data column reach
to 30,000 columns, then will generate a new file, for
example HBA01002.XLS
Under the folder HBA01\, if the total files more
than 99 files, will generate anew route, such as
HBA02\ ........
The file's route structure :
HBA01\
HBA01001.XLS
HBA01002.XLS
.....................
HBA01099.XLS
HBA02\
HBA02001.XLS
HBA02002.XLS
.....................
HBA02099.XLS
HBAXX\
.....................
.....................
Remark :
XX : Max. value is 10.
8
Page 11
6. Saving data from the SD card
EXCEL data screen ( for example )
to the computer ( EXCEL software )
1)After execute the Data Logger function, take away the
SD card out from the " SD card socket " ( 3-13, Fig. 1 ).
2)Plug in the SD card into the Computer's SD card slot
( if your computer build in this installation ) or
insert the SD card into the " SD card adapter ". then
connect the " SD card adapter " into the computer.
3)Power ON the computer and run the " EXCEL software ".
Down load the saving data file ( for example the file
name : HBA01001.XLS, HBA01002.XLS ) from the SD
card to the computer. The saving data will present into
the EXCEL software screen ( for example as following
EXCEL data screens ) , then user can use those EXCEL
data to make the further Data or Graphic analysis
usefully.
9
Page 12
EXCEL graphic screen ( for example )
dAtE......
SP-t.......
bEEP........
dEC........
t-CF.......
rS232......
bAro..........
7. ADVANCED SETTING
Under do not execute the Datalogger function, press the
" SET button " ( 3-5, Fig. 1 ) > 2 seconds continuously will
enter the " Setting " mode., then release the " SET button ".
Following press the " SET button " (3-5, Fig. 1 ) once a
while in sequence to select the eight main function, the
display will show :
Sd F..... SD memory card Format
Set clock time ( Year/Month/Date, Hour/Minute/
Second )
Set sampling time
Set beeper sound ON/OFF
Set SD card Decimal character
Select the Temp. unit to or ℃℉
Set RS232 data output ON/OFF
Set the unit of Barometer
10
Page 13
Remark :
During execute the " Setting " function, if within
5 seconds , do not press any buttons further, the
LCD Display will return to normal screen.
7-1 SD memory card Format
When the Display show " Sd F "
1)Use the " Button " ( 3-3, Fig. 1 ) or " Button " (▲▼
3-4, Fig. 1 ) to select the upper value to " yES " or
" no ".
yES - Intend to format the SD memory card
no - Not execute the SD memory card format
2)If select the upper to " yES ", press the " Enter Button
" ( 3-2, Fig. 1 ) once again, the Display will show text
" yES Enter " to confirm again, if make sure to do the
SD memory card format, then press " Enter Button "
once will format the SD memory clear all the existing
data that already saving into the SD card.
7-2 Set clock time ( Year/Month/Date,
Hour/Minute/ Second )
When the Display show " dAtE "
1)Use the " Button " ( 3-3, Fig. 1 ) or " Button " ▲▼
( 3-4, Fig. 1 ) to adjust the value ( Setting start from
Year value ). After the desired value is set, press the
" Enter button " ( 3-2, Fig. 1 ) once will going to
next value adjustment ( for example, first setting
value is Year then next to adjust Month, Date, Hour,
Minute, Second value ).
11
Page 14
Remark :
The adjusted unit will be flashed.
2)After set all the time value ( Year, Month, Date, Hour,
Minute, Second ), press the " SET button " ( 3-5, Fig.
1 ) once will save the time value, then the screen will
jump to Sampling time " setting screen ( Chapter 7-3 ).
Remark :
After the time value is setting, the internal clock will
run precisely even Power off if the battery is under
normal condition ( No low battery power ).
7-3 Set sampling time
When the Display show " SP-t "
1)Use the " Button " ( 3-3, Fig. 1 ) or " Button " ▲▼
( 3-4, Fig. 1 ) to adjust the sampling value :
5 seconds, 10 seconds, 30 seconds, 60 seconds,
120 seconds, 300 seconds, 600 seconds, Auto
After the desired value is set, press the " Enter Button "
( 3-2, Fig. 1 ) to save the adjusting value with default.
Remark :
The " Auto " sampling time means when the measuring
value is changed ( > ± 1 %RH or > ± 1 ) will save℃
the data to the memory circuit one time.
7-4 Set beeper sound ON/OFF
When the Display show " bEEP "
1)Use the " Button " ( 3-3, Fig. 1 ) or " Button "▲▼
( 3-4, Fig. 1 ) to select the data to " yES " or " no ".
12
Page 15
yES - Meter's beep sound will be ON with default
when save each data.
no - Meter's beep sound will be OFF with default.
when save each data.
2)After select the upper text to " yES " or " no ", press the
" Enter Button " ( 3-2, Fig. 1 ) will save the setting
function with default.
7-5 Decimal point of SD card setting
When the Display show " dEC "
The numerical data structure of SD card is used the " . "
as the decimal with default, for example "20.6"
"1000.53" . But in certain countries ( Europe ...) is
used the " , " as the decimal point, for example " 20,6 "
"1000,53". Under such situation, it should change the
Decimal character at first.
1)Use the " Button " ( 3-3, Fig. 1 ) or " Button " ▲▼
( 3-4, Fig. 1 ) to select the upper text to " USA " or
" Euro ".
USA - Use " . " as the Decimal point with default.
Euro - Use " , " as the Decimal point with default.
2)After select the text to " USA " or " Euro ",
press the " Enter Button " ( 3-2, Fig. 1 ) will save the
setting function with default.
13
Page 16
7-6 Select the Temp. unit to or ℃℉
When the Display show " t-CF "
1)Use the " Button " ( 3-3, Fig. 1 ) or " Button "▲▼
( 3-4, Fig. 1 ) to select the upper Display text to " C " or
" F ".
C - Temperature unit is ℃
F - Temperature unit is ℉
2)After Display unit is selected to " C " or " F ", press the
" Enter Button " ( 3-2, Fig. 1 ) will save the setting
function with default.
7-7 Set RS232 data output ON/OFF
When the Display show " rS232 "
1)Use the " Button " ( 3-3, Fig. 1 ) or " Button "▲▼
( 3-4, Fig. 1 ) to select the upper Display text to "yES " or
" no ".
yES - RS-232 output terminal ( 3-12, Fig. 1 )
will send the RS232 signal output.
no - RS-232 output terminal ( 3-12, Fig. 1 )
will not send the RS232 signal output.
2)After Display text is selected to " yES " or " no ",
press the " Enter Button " ( 3-2, Fig. 1 ) will save the
setting function with default.
14
Page 17
7-8 Set the unit of Barometer
When the Display show " bAro "
1)Use the " Button " ( 3-3, Fig. 1 ) or " Button " ▲▼
( 3-4, Fig. 1 ) to select the upper Display text to
hPA ", " - Hg " or " InHg ".
Text Unit
hPA hPa
- HGmmHg
InHginHg
2)After the unit is selected, press the " Enter Button " (
3-2, Fig. 1 ) will save the setting function with
default.
8. POWER SUPPLY from DC
ADAPTER
The meter also can supply the power supply from the
DC 9V Power Adapter ( optional ). Insert the plug of
Power Adapter into " DC 9V Power Adapter Input Socket
" ( 3-14, Fig. 1 ).
15
Page 18
9. BATTERY REPLACEMENT
When the left corner of LCD display show " ", it
1)
is necessary to replace the battery. However, in-spec.
measurement may still be made for several hours after
low battery indicator appears before the instrument
become inaccurate.
2)Loose the " Screw of the battery cover " ( 3-10, Fig. 1 )
and take away the " Battery Cover " ( 3-9, Fig. 1 ) from
the instrument and remove the battery.
3)Replace with DC 1.5 V battery ( UM4/AAA,
Alkaline/heavy duty ) x 6 PCs, and reinstate the cover.
4)Make sure the battery cover is secured after changing
the battery.
10. SYSTEM RESET
If the meter happen the troubles such as :
CPU system is hold ( for example, the key button can
not be operated... ).
Then make the system RESET will fix the problem.
The system RESET procedures will be either following
method :
During the power on, use a pin to press the " Reset
Button " ( 3-11, Fig. 1 ) once a while will reset the
circuit system.
16
Page 19
11. RS232 PC SERIAL INTERFACE
Center Pin............................................
The instrument has RS232 PC serial interface via a 3.5 mm
terminal ( 3-12, Fig. 1 ) if the RS232 function already select
to " ON ", refer to chapter 7-7, page 14.
The data output is a 16 digit stream which can be
utilized for user's specific application.
A RS232 lead with the following connection will be
required to link the instrument with the PC serial port.
MeterPC
(9W 'D" Connector)
Pin 4
(3.5 mm jack plug)
Pin 2
2.2 K
resistor
Pin 5
The 16 digits data stream will be displayed in the
following format :