Before weighing ............................................................................................................................................... 5
When to calibrate: ............................................................................................................................................ 6
How to calibrate: .............................................................................................................................................. 6
FEATURES .................................................................................................................................................... 6
FUNCTION SETTINGS ...................................................................................................................................... 6
1. SAFETY PRECAUTIONS
All safety message
the following:
˜W ARNING
CAUTION
When using the scale/balance, the following safety precautions should always be followed.
˜ WARNING
Use only the AC adaptor with the scale/balance. Other adaptor may cause damage. (AC adaptor is
optional)
˜ CAUTION
Avoid installing the scale/balance in direct sunlight; this may cause discoloration or malfunctions.
Do not mix the type of batteries. Replace all other batteries at the same time.
If the scale/balance is not to be used for a long period of time, remove all batteries from the battery
compartment to avoid leakage which may cause damage to the instrument.
s are identified by the following the words "WARNING" and "CAUTION". These words mean
Important information to alert you to a
situation that might cause serious injury
and damage to your property if
instructions are not followed.
Important information that tells how to prevent
damage to equipment.
Avoid overloading that could cause damage to the scale/balance.
Do not drop water in the scale/balance that is not water-resistant. It causes damage, if inside of the
scale/balance is wet.
Matter charged static electricity could influence the weighing. Discharge the static electricity. As example, the
method is to use the electrification prevention spray, to spray it to both side of weighing platform.
2
2. PARTS DESCRIPTION
3
3. PREPARATION
1. Installing batteries/Connecting the AC adaptor
Batteries
Battery Remove the battery compartment
compartment cover and insert six batteries(R6P/LR6
/AAsize) into the battery compartment,
taking extreme care that the polarities
plus and minus is observed.
R6P/LR6/AA size dry batteries.
Battery compartment cover (batteries are not included)
AC adapter (optional)
Plug the AC adapter to the AC
AC adaptor adapter socket on the side.
socket The AC input requirement could
be 110,120,220,230 or 240Volts
(50/60Hz) depending on the area
where used, so please verify that
the adaptor is correct
2. Setting up the scale/balance
&AUTION!!
Avoid installing the scale/balance in
scale/balance on a firm weighing table so that the scale/balance is level. (The scale/balance will not perform
accurately when it is not level.)
Place the scale/balance on the firm surface that is flat and level for an accurate weighing.
1.
Selection of auto backlight mode.
When the balance enter function setting menu. Press [UNIT] key, the display will show L_ON or L_OFF
ss [ZERO], selection display L_ON, this means auto backlight function active, press [ZERO] agai
pre
ay will show L_OFF, this means auto backlight function inactive, i.e. shut off backlight function.
displ
Selection of auto shut off mode.
2.
When the balance enter function setting menu. Press [UNIT] key, the display will show A_ON or A_OFF
ss [ZERO], selection display A_ON, this means auto shut off function active. Press [ZERO]
pre
display will show A_OFF, this means auto shut off function inactive.
3.Return to weighing mode.
Press [ON/OFF] to turn the power off, press [ON/OFF] again to turn on the power and the scale return to the
weig
hing mode.
Counting procedures
1.Press [ON/OFF] to turn on the sc
Wait for "0" to appear on the display.
2.Start the Count Procedure
If necessary, press [ZERO] key to set the display to “0”.
direct
ale.
sunlight that may cause discoloration or malfunctions. Place your
,
n, the
,
again, the
4
3. Place a given number of samples of an item on the pan (the Sample Size should be either 102050 or 100
pieces). The weight of these samples will show on the display.
4. Press the [PC
5. Select the sampl
necessary to put it in the correct sample size ,the sample size is the same as in step three =102050or
6. Press the [UNIT] key, the displ
sele
cted and show the starting sample size on the di
(you ca
7. Place the items that you
8. Presse
to turn the scale off to exit .
NOTE: the weight of unit sample > 1
n now remove the samples if you want to return the scale to
s the “PCS” key to exit the counting function and return to normal weighing or you can press ON/
S] key several times to put the scale in PCS mode
e size (the same as you chose above) by pressing the [PCS] key. (press it as many times as
ay will show “selected pcs”, then the scale will remember the sample
splay
want counted onto the tray and the total number of items will show on the displa
0e.
.
100.
size you
0)
y.
OFF
4. WEIGHING
Before weighing
Whenever possible, please allow the balance to warm up for 10 minutes after first turning to power on, so that
the balance will function. Properly and accurately.
Error messages
EEEE: Overload
: Low battery
Calibration may be required before weighing.
Read "CALIBRATION" first and if necessary, calibrate your scale/balance for accurate weighing.
Weighing procedures
1. Press [ON/OFF] to turn on the scale/balance.
When power is turned on, all display segments appear on LCD for a few seconds and finally "0" will appear on
the display.
2. Select the weighing unit with [UNIT].
[UNIT] to select a weighing unit, eg. "g"" ct " " oz "" dwt ""ozt""gn""lb""n". Once the unit has
Press
been selected, the selected unit will be displayed to the weight value.
3. Start weighing
If you do not use a container for weighing,
Æ
Verify the reading is "0". If not, press [ZER
When the reading becomes stable, the stable indicator is displayed.
Æ If you use a container for weighing,
Place an empty container on the platform.
Wait for the stability indicator to be displayed and press [ZERO].
Place the objects to be weighed in the container.
When the reading becomes stable, the stable indicator is displayed.
O] to display "0". Place objects on the weighing platform to weigh.
5. CALIBRATION
When to calibrate:
When first installing the balance, the factory will still be valid, of course. However, if it is exposed to severe
environmental conditions such as dust, contamination, high humidity, a re-calibration can be mades. The common calibration intervals should also be adhered to (e.g. annual re-calibration according to ISO).
How to calibrate:
1. The scale must be turned off.
2. Press and hold the [Unit] k
mode and display a random number such as 0247581811, which is tha indication of A/D converter (the signal of
the weighing sell). When loading the scale, this number should increase and when removing the load, it should
decrease and stabilise. Now press [UNIT] key to confirm the zero point. The weighing platform should be free from
any load. The display will show "00SAVE" for a short time and go to the first calibration point "2000.0"
(PCE-BS 3000) or "200.00" (PCE-BS 300). Now load the scale with the 2000 g weight. Then press the [Unit] key
again. The display will show "00SAVE" again and then go to the second calibration point "3000.0". You can either
cancel the calibration by pressing the [ON/OFF] key or load the scale with a 3000 g weight and press the [Unit]
key again. "00SAVE" will be displayed for a short time. The two-point calibration is now completed and the display
will again show the signal of the A/D converter.
. Turn off the balance and turn it back on by pressing the [ON/OFF] key. Carry out a few weighings. In case you
believe there are still some deviations, repeat the calibration procedure. If you prefere so, you can also contact PCE
Instruments for professional calibration at the calibration laboratory.
ey and at the same time press the [ON/OFF] key. The scale will now enter calibration
The calibration is complete.
FUNCTION SETTINGS
Enter function setting mode.
Press [ON/OFF] to turn the power off. Press [ZERO] first, then press [ON/OFF], while keeping [ZERO] pressed,
the display will show A_ON or A_OFF.
NOTE: This instrument doesn't have ATEX protection, so it should not be used in potentially explosive atmospheres
(powder, flammable gases).