READ THIS MANUAL BEFORE using or servicing the instrument.
FOLLOW these instructions carefully.
KEEP this manual for future use.
CAUTION
The purpose of this manual is to make the operator aware of the
texts and explanatory figures, the basic requirements and criteria
for the installation and correct use of the instrument.
Installation, maintenance and repair must carried be out only by
specialized personnel who must have read and understood this manual. “Specialized personnel” means personnel who, due to training
and professional experience, have been expressly authorized by the
System Safety Manager to perform the installation.
Supply the instrument with voltage whose value is within the limits
specified in the specifications.
It is the user’s responsibility to ensure that the installation complies
with the applicable regulations.
Any attempt to disassemble or modify the instrument not expressly
authorized will invalidate the warranty and relieve Pavone Sistemi
of any responsibility.
The installation and maintenance of this instrument should only be
permitted by qualified personnel.
Pay attention when checking, testing, and adjusting with the instrument turned on.
Make the electrical connections in the absence of the power supply
voltage.
Failure to observe these precautions may result in danger.
DO NOT ALLOW untrained personnel to work, clean, inspect, repair
or tamper with this instrument.
Page 2
INTRODUCTION
The MCT 1302 is a touch screen weight indicator and batcher to be combined with load cells to meet
the industrial applications.
The instrument is installed in a frame in a slot having a drilling template of 138 x 53 mm and is secured
by means of the 2 tensioning screws supplied.
The MCT 1302 uses the RS232 serial port with ASCII protocols, to be connected to the PC, PLC and remote units with a maximum distance of 15m beyond which it is necessary to use the RS422/RS485 serial
output that allows the connection also with MODBUS RTU protocol up to 32 addressable instruments.
The availability of the most popular field buses, as an alternative to the RS485 port, also allows interfacing the instrument with any supervision device currently offered by the market.
A USB 2.0 port is also available and facilitates interfacing with PCs using AN Utility Software available
in the kit.
6 digital inputs and 6 digital outputs configurable by Set-up are always available.
It is also possible to have the analog output in voltage or current even in the presence of the FIELDBUS.
Available versions:
•MCT 1302: weight indicator with RS232 serial output, USB, RS485. The supported protocols are
continuous Modbus RTU.
•MCT 1302/A: version with analog output.
•MCT 1302/PROFINET: weight indicator with RS232, USB and PROFINET serial output.
•MCT 1302/ETHERNET IP: weight indicator with RS232, USB and ETHERNET IP serial output.
•MCT 1302/ETHERCAT: weight indicator with RS232, USB and ETHERCAT serial output.
•MCT 1302/PROFIBUS: weight indicator with RS232, USB and PROFIBUS serial output.
The analog output can also be present with the FIELDBUS option
Page 3
IDENTIFICATION PLATE OF THE INSTRUMENT
It is important to communicate this data, in case of request for information or indications concerning
the instrument, together with the program and version number, shown on the cover of the manual and
displayed when the instrument is switched on.
PAVONE SISTEMI
WARNINGS
The procedures listed below must be performed by specialized personnel.
All connections must be made when the instrument is switched off.
Page 4
TECHNICAL FEATURES
Power supply 18÷30 Vdc
Max. absorption 5 W
Isolation Class II
Installation category Cat. II
Working temperature -10°C ÷ +50°C (Max. humidity 85% without
condensation)
Storage temperature -20°C ÷ +70°C
Display Graphic 240x128 pixel LCD
Keyboard 4 wire resistive touch screen
Overall dimensions 150 x 95 x 26 mm (l x h x w) including terminal
blocks
150 x 95 x 56 mm (l x h x w) with FIELDBUS option
Drilling template: 138 x 82 mm
Assembly Panel Mount
Case material Aluminum
Connections Removable screw terminal blocks, 3.81 pitch.
Load cell power supply 5 Vdc/120 mA (max. 8 cells of 350 in parallel)
protected against short circuit
Input sensitivity 0.02 µV min.
Linearity < 0.01% of the full scale
Temperature drift < 0.001% of the full scale / C°
Internal resolution 24 bits
Displayed weight resolution Up to 999,999 divisions on the payload
Measuring range From -3.9 mV/V to +3.9 mV/V
Weight acquisition frequency 12 Hz - 1000 Hz
Digital filter Selectable 0,1 ÷ 250 Hz
Weight decimal number from 0 to 4 decimal places
Zero and Full Scale Calibration Automatic (theoretical) or executable from the
keyboard.
Logic outputs 6 opto-insulated (dry contact) max 24 Vdc / 100 mA
each
Logic inputs 6 opto-insulated at 24 Vdc PNP (external power
supply)
Serial port (#2) RS232C and RS422/485
Maximum cable length 15 m (RS232C) and 1000 m (RS422 and RS485)
Serial protocols ASCII, Modbus RTU
Baud rate 1200, 2400, 4800, 9600, 19200, 38400, 57600,
115200 selectable
USB port: compliant with USB 2.0; speed up to 12 Mbps
Analog output (optional) 16-bit opto-insulated
Voltage: 0 ÷ 5/10 V (R min 10 KOhm),
Current: 0/4 ÷ 20 mA (R max 300 Ohm)
Analog output calibration From the keyboard
Linearity <0.02% FS
Thermal drift 0.001% FS/°C
Microcontroller: ARM Cortex M0 + 32 bit, 256 KB Flash
reprogrammable on-board from USB.
Data memory 64 Kbytes expandable up to 1024 Kbytes
Fieldbus (as an alternative to RS485) PROFINET, ETHERNET IP, ETHERCAT, PROFIBUS
Compliance with Standards EN61000-6-2, EN61000-6-3, EN61010-1,
EN45501
Page 5
INSTALLATION
GENERAL INFORMATION
The MCT 1302 is a compact panel mounting instrument made of aluminum.
The instrument is installed in a panel in a slot with a drilling template of 138 x 82 mm and is secured
by means of the 2 tensioning screws supplied.
82±1
Rmax 2
138±1
The MCT 1302 must not be immersed in water, subjected to jets of water and cleaned or washed with
solvents.
Do not expose to heat or direct sunlight.
Do not install the instrument near power equipment (motors, inverters, contactors, etc.) or near devices
that do not comply with CE regulations for electromagnetic compatibility.
The connection cable for the load cells must have a maximum length of 140 m/mm².
The RS232 serial line must have a maximum length of 15 meters (EIA RS-232-C standards).
The warnings given on the connection of the individual peripheral devices must be observed.
OVERALL DIMENSIONS
MCT 1302
94
149
Page 6
30
20
5.0
94
136
149
52
80
MCT S 1302 DIMENSIONS
237
32
20
68
62
113
M4
169
37
146
202
Page 7
ELECTRICAL INSTALLATION
The MCT 1302 instrument uses for the electrical connection of the 3.81 mm removable screw terminal
blocks. The load cell cable must be shielded and channeled away from power cables to avoid electromagnetic interference.
The field wiring of the instrument must be suitable for the environment in which it will be used
and must comply with all national regulations.
A switch or disconnector must be included in the electrical system.
The switch must be close to the instrument and easily reachable by the operator.
The switch must be marked as a device to stop the machine.
The switch or circuit breaker used as a disconnecting device must comply with the applicable requirements of IEC 60947-1 and IEC 60947-3.
Disconnect the power supply to the instrument before making electrical connections.
POWER SUPPLY OF THE INSTRUMENT
OPTIONAL ETHERNET
FIELDBUS
The instrument is powered through terminals 25 and 26. The power cable must be routed separately
from other cables.
The electric zero (terminal 26) is connected to the metal case. Connect
the terminal 26, in addition to the GND of the power supply, also
on the ground using the screw near the power terminals.
Supply voltage: 18÷30 Vdc, max 5W
+
-
Power Supply
24 Vdc
25 +Vdc
26 GND
Screw
Page 8
LOAD CELL CONNECTION
The load cell cable must not be routed with other cables, but must follow its own path.
Any cable extension connections must be carefully shielded, respecting the color code and using the
cable of the type supplied by the manufacturer. The extension connections must be made by welding,
or through support terminal blocks or through the junction box supplied separately.
Up to a maximum of 8 350 ohm load cells in parallel can be connected to the instrument. The supply
voltage of the load cells is 5 Vdc and is protected by a temporary short circuit.
The measuring range of the instrument involves the use of load cells with sensitivity up to 3.9 mV/V.
The load cell cable must be connected to terminals 3 to 9. In the case of a 4-wire cell cable, jumper
the terminals 5 to 8 and 6 to 7.
Connect the load cell cable shield to terminal 9.
JBOX
3 +Signal
4 -Signal
+Sig
+Exc-Exc
-Sig
5 -Sense
6 +Sense
7 +Exc.
8 -Exc
9 Shield
In case of use of two or more load cells, use appropriate junction boxes (CEM4/C or CSG4/C) of
which the connection is shown below.
+EXC
-EXC
+SGN
-SGN
SHD
+EXC
-EXC
+SGN
-SGN
SHD
+EXC
-EXC
+SGN
-SGN
SHD
+EXC
-EXC
+SGN
-SGN
SHD
J-BOX CGS4
+ALM
1
-ALM
2
+SGN
3
-SGN
4
SCH
5
+ALM
1
-ALM
2
+SGN
3
-SGN
4
SCH
5
+ALM
1
-ALM
2
+SGN
3
-SGN
4
SCH
5
+ALM
1
-ALM
2
+SGN
3
-SGN
4
SCH
5
-SGN
+SGN
-ALM
-SNS
+SNS
+ALM
SCH
1
2
3
4
5
6
7
25 PIN CONNECTOR
+24V
04
03
08
05
06
07
09
24
23
22
15
16
20
21
25
26
-SIG1
+SIG1
-EXC
-SENSE
+SENSE
+EXC
SHIELD
Iout+
Vout+
AGND
TX+/RX+
TX-/RX-
S I/O+
S I/O-
+
24V
-
MCT 1302
I COM
OUT1
OUT2
OUT3
OUT4
OUT5
OUT6
O COM
S.GND
IN1
IN2
IN3
IN4
IN5
IN6
RXD
TXD
1
2
3
4
5
6
7
8
9
10
11
12
13
14
FIELDBUS
19
18
17
Page 9
LOGIC INPUTS
The logic inputs are insulated from the instrument through optocouplers.
The connection cables of the logic inputs must not be channeled with power cables.
Use a shorter connection cable as possible.
To activate a logic input, it must be brought to the positive of a 24Vdc power supply while
the common item must be connected to its negative. The function of the inputs is established
in the relevant Setup menu.
1 In1
2 In2
3 In3
4 In4
5 In5.
24 Vdc
+
6 In6
7 I Comm
-
LOGIC OUTPUTS
The 6 logic outputs are of the Photorelay (dry contact) type with a common item. The capacity of each
contact is 100 mA / 30Vdc. By enabling the output, the contact closes (NA contact).
The environment where the equipment is installed can normally be subject to strong magnetic
fields and electrical noise caused by the machinery present, so it is advisable to implement
the usual precautions in order to avoid that these affect the typical signals of a precision
electronic equipment. (filters on contactor, diodes on 24 Vdc relays, etc.)
The connection of the outputs is shown below
8 Out 1
9 Out 2
10 Out 3
Page 10
11 Out 4
12 Out 5.
13 Out 6
+
14 O Comm
-
24 Vdc
CONNECTION WITH ADDITIONAL INPUTS/OUTPUTS BOARD
In applications where additional inputs/outputs are required, optional I/O boards can be connected
using the dedicated serial connection.
To make the connection, use a shielded cable, taking care to connect the screen to ground at only one
of the two ends. If the cable has more conductors than those used, connect the free wires to the screen.
The serial connection cable must have a maximum length of 100 meters. The cable must
not be channeled with other cables, but must possibly follow its own path
MCT 1302
External
I/O Board
19
20
20 S I/O+
21 S I/O-
SERIAL RS232
COMMUNICATION
The serial connection cable must have a maximum length of 15 meters (EIA RS-232-C standards), beyond
which it is necessary to adopt the RS485 interface with which the instrument is equipped.
The cable must not be channeled with other cables (i.e. outputs connected to contactors or
power cables), but must possibly follow its own path.
The PC used for the connection must comply with the EN 60950 standard.
The RS232 serial port is normally used for PC, printer and repeater connections.
To make the serial connection, use a suitable shielded cable, taking care to connect the screen to ground
at only one of the two ends.
RS232
INTERFACE
RX
TX
MCT 1302
17 TX
18 RX
19 GNDGND
Page 11
RS485
Through serial RS485 interface, it is possible to make serial connections for long distances (up to 1000 m).
The RS485 serial connection is of the 2-wire type, and allows to connect up to 32 instruments to a
single MASTER unit (personal computer, PLC, etc.) using a twisted and shielded cable, taking care to
connect the screen to the ground at only one of the two ends.
The cable must not be channeled with other cables (i.e. outputs connected to contactors or power ca-
bles), but must possibly follow its own path.
The PC used for the connection must comply with the EN 60950 standard.
NOTE: If there is a fieldbus, the RS485 is not available.
MCT 1302
15 TX+/RX+
16 TX-/RX-
USER
RS485
TX+/RX+
TX-/RX-
INTERFACE
ANALOG OUTPUT (OPTIONAL)
The instrument provides an opto-insulated analog output in current and/or voltage.
Features:
• Analog voltage output: range from 0 to 10 Volt or from 0 to 5 Volt, minimum load 10K
• Analog current output: range from 0 to 20 mA or from 4 to 20 mA. The maximum load is 300.
To make the connection, use a shielded cable, taking care to connect the screen to ground at only one
of the two ends.
The analog sending can be sensitive to electromagnetic disturbances, it is therefore recommended that
the cables are as short as possible and that they follow their own path.
Warning: do not connect the analog output to active devices.
Page 12
ANALOG
INTERFECE
CURRENT
VOLTAGE
MCT 1302
24 +mA
23 +Volt
22 GNDGND
USB DEVICE (SPECIFICATION 2.0 COMPLIANT, FULL-SPEED 12 MBPS)
Use this communication port to directly interface a PC via a USB port.
Use a standard USB cable for the connection.
To connect the instrument via the USB device port, the appropriate driver for the operating system used
must be installed on the PC. For the installation, follow the specific instructions.
FIELDBUS CONNECTIONS
As an alternative to the RS485 serial port, some of the most popular fieldbuses are available. Only
one fieldbus can be used, which must be specified when ordering.
ETHERNET CONNECTION
In the lower left part of the instrument there is an RJ45 connector for Ethernet network.
Features:
10 Mbps transmission speed
Network compatible with 10/100/1000 Base-T networks
Indicator LEDs (2) Ethernet line presence and communication/diagnostics
Buffer size 256 bytes
Connection Timeout Min. 30 seconds - Max. 90 seconds
Link Timeout (cable disconnected) 30 seconds
PINDESCRIPTION
1TX+
2TX3RX+
4
5
6RX7
8
To connect to the MASTER, use an Ethernet twisted pair cable with the relevant RJ45 connector.
• The RJ45 Ethernet connection cable has a variable maximum length depending on the type of cable.
A common shielded Cat5 cable can have a maximum length of about 180 m.
• It is possible to connect the ethernet communication port directly to the PC, without switching from
other network devices (routers, switches, hubs, lan-bridges or other), but special RJ45 cables, called
“crossovers”, must be used.
• Usually the cables are “direct”, and allow the connection to network devices such as routers or hubs,
but not to directly connect two PCs (although currently there are network cards with auto-sensing
technology, which recognize the type of cable and the type of connection, allowing direct PC-PC
connections even using non-crossover cables).
• The following page shows the diagrams of the two types of cables mentioned above and the related
connection diagram.
Page 13
6RX-
3RX+
2TX-
1TX+
1 2 3 4 5 6 7 8
CROSSOVER ETHERNET CABLE
6RX-
3RX+
2TX-
1TX+
1 2 3 4 5 6 7 8
DIRECT ETHERNET CABLE
6RX-
3 RX+
2TX-
1 TX+
TX+
RX+
RX-
TX-
RX+
TX+
RX-
TX-
1 2 3 4 5 6 7 8
6TX-
3 TX+
2RX-
1 RX+
TX+
RX+
RX-
TX-
RX+
TX+
RX-
TX-
1 2 3 4 5 6 7 8
ETHERNET/IP CONNECTION
Ethernet/IP is a real-time industrial protocol based on the Ethernet network.
There are 2 RJ45 connectors to allow the connection of several instruments under the same network.
Refer to the description above for connection notes and warnings.
Features:
10 and 100 Mbit operations, Full and Half Duplex
Modbus-TCP server
Up to 128 bytes of fieldbus I/O in every direction.
PROFINET CONNECTION
There are 2 RJ45 connectors to allow the connection of several instruments under the same network.
Refer to the previous page for connection notes and warnings.
Features:
PROFINET IO Real Time (RT) communications
Modbus-TCP server
Up to 128 bytes of fieldbus I/O in every direction.
ETHERCAT CONNECTION
EtherCAT is a real-time industrial protocol based on the Ethernet network.
The EtherCAT protocol requires that the RJ45 connectors have the function of IN and OUT.
By putting several MCT 1302 instruments in series, the MASTER will be connected to the IN connector
of the first MCT 1302 whose OUT connector will connect to the IN connector of the next, etc.
Refer to the previous page for connection notes and warnings.
Page 14
MAC ADDRESS IN THE INSTRUMENTS WITH ETHERNET INDUSTRIAL FIELDBUS.
The instruments that mount Hilscher modules with Industrial Ethernet protocol (Profinet, EthernetIP, Ethercat, etc.) have a label under the connectors, as shown below.
This label shows the MAC Address of the module (red box), an identification number of the module
(blue box) and a QR code that contains the MAC Address. The latter can be read with a smartphone
using a QR reading app i.e., on the Google Play Store, “QR Code Reader”).
PROFIBUS CONNECTION
PinSignalDescription
3B line+RxD/+TxD, level RS485
4RTSRequest to send
96
15
3
8
5GNDGround (insulated)
6+ 5V Bus Output +5V end (insulated)
8A line-RxD/-TxD, level RS485
Housing Cable shield
Internally connected to the protective
ground according to Profibus
A_LINE
B_LINE
specifications
To connect to the Profibus MASTER, use a standard Profibus cable. The typical impedance of the cable
should be between 100 and 130 Ohm (f > 100 kHz). The cable capacity (measured between conductor
and conductor) should be less than 60 pF/meter and the minimum conductor cross-section should not
be less than 0.22 mm
2
.
In a Profibus-DP network, both type A and type B cables can be used, depending on the required performance. The following table summarizes the characteristics of the cable to be used:
FeatureType A cableType B cable
Impedance135 to 165 ohm (f = 3÷20 MHz)100 to 300 ohm (f >100 kHz)
Capacity< 30 pF/m< 60 pF/m
Resistor< 110 ohm/kmConductor section> 0.34 mm
2
> 0.22 mm
2
The following table shows the maximum length of the line with type A cable and with type B cable
according to the different communication speeds required:
Length (m) of the type A cable1200 1200 1000400200100100100
Length (m) of the type B cable1200 1200600200----
For reliable operation of the fieldbus, a line terminator should be used at both ends.
In case of multiple MCT 1302 instruments, use the line end only on an instrument.
For the setup of the board, the GSD file (hms_1810.gsd) is available and must be installed in the master.
Page 15
94
FRONT PANEL OF THE INSTRUMENT
The MCT 1302 is a touch screen instrument dedicated to batching applications.
The main operating characteristics are:
• Programmable recipes, standard - 100 recipes of 20 steps. With Alibi memory option - 1000
recipes with 20 steps.
• 50 programmable activities.
• Management of 38 configurable outputs (6 internal outputs to the instrument plus 8 outputs for each
connected I/O module; maximum 4 modules that can be connected to the instrument). Depending
on the type of activity, up to 2 outputs can be selected.
• Coarse and Fine batching control, tolerance control of the dosed weight and lack of product.
• Automatic repetition of the batching cycles.
• Management of the batched totals by activity and recipe.
The setup parameters are easily accessible and modifiable through the use of the function keys that
from time to time appear on the display used to select, edit, confirm and save the new settings.
MCT 1302
DISPLAY
In the operating mode the display shows all the information necessary for a complete control of the
system. According to the various programming procedures, the display is used for programming the
parameters to be entered in the memory, i.e. messages indicating the type of operation being carried
out and therefore helping the operator in the management and programming of the instrument.
POWER UP OF THE INSTRUMENT
When switched on, the display temporarily shows an introductory mask, which shows the firmware
code and the version number. After a few seconds, the main screen is displayed, from which all the
instrument operations can be accessed.
Page 16
MCT 1302 - DOSAGGIO
PW1507 Rev. 0.9
MAIN SCREEN
From this screen it is possible to access all the operational and programming functions of the instrument,
following the indicated commands.
Date
Batching cycles
Recipe correction
factor
F1 Operation key
F2 Operation key
The display mode of the main screen can be selected using the appropriate parameter in the “Operating Mode” menu:
• “Standard” main screen, in which all the quadrants are displayed (weight, recipe selection, cycle
setting, set correction factor setting, batching information).
• Main screen “weight only”, in which only the weight dial is displayed. In this case the weight value
is displayed with a larger size than the standard display. Moreover, in case of batch in progress,
a bargraph is displayed, which represents the net weight batched in relation to the set value.
The “weight only” main screen is activated under the following conditions:
Menu key
F3 Operation key
Time
Weight quadrant
Recipe
quadrant
F4 Operation key
• After 5 seconds from switching on the instrument.
• After 5 seconds from the start of the batch.
To temporarily restore the “standard” main screen, press on the weight value. After 5 seconds the
“weight only” main screen appears again.
The functions of the operation keys (F1, F2, F3 and F4) can be programmed using the appropriate
parameters in the “Operating Mode” menu.
All the keys on the main screen can be locked using the appropriate parameters in the “Operating
Mode” menu.
Weight quadrant
F1 Operation key
F4 Operation key
F2 Operation key
Menu key
F3 Operation key
Page 17
OPERATION KEYS
Access to the parameter programming menu.
Press in the weight display dial to switch the weight display (net, gross).
Access to the recipe selection screen.
Press in the setting screen for the number of batching cycles.
Access to the recipe correction factor setting screen. The correction value
change creates the printout of the recipe with the recalculated batching
set point. The value is expressed in percentage or weight (Parameter
"CORRECTION MODE" - page 22)
Semi-automatic Zero Command.
Tare Command.
Deleting Tare Command.
Start batching command.
Weight Print Command
Page 18
INFO DISPLAY
ERROR INDICATIONS AND MESSAGES OF WEIGHT
Weight calibration not performed; flashing message, alternating
with the display of the detected weight.
Indication displayed when the instrument is switched on, while
waiting for the conditions necessary for automatic zeroing of the
weight; in case of automatic reset disabled, this indication is not
displayed.
Overload warning, displayed when the gross weight exceeds the
maximum load of the weighing system by more than 9 divisions or
when the displayed weight exceeds 999999.
Signal weight absent or out of the weighing range.
Fieldbus network disconnected.
Connection error with the Fieldbus interface.
Communication error with external input/output module (the address
of the module that generated the error is reported).
When communicating with PC software “Optimation” for remote instrument setup, the following screen
appears.
WEIGHT STATUS REPORTS
Display switched to net weight value.
Indication of stable weight.
Page 19
J
J
J
J
OPERATING FUNCTIONS
SWITCHING THE GROSS/NET WEIGHT DISPLAY
Press on the weight dial on the instrument main screen to switch the weight display. Each time the displayed weight changes between the following values:
• Gross - the gross weight is displayed in the weight dial.
• Net - the net weight is displayed in the weight dial.
SEMI-AUTOMATIC ZERO
This operation is performed to correct small scale zero displacements.
The command to reset the gross weight is not run under the following conditions:
• Unstable weight (the weight does not stabilize within 3 seconds of the reset command).
• Gross weight, compared to the original zero calibration, greater in positive or negative of the
value entered in the Zero Band parameter (instrument setup menu, metrological parameters). If this
parameter is set to 0, the semi-automatic zero function is disabled.
The operation to reset the gross weight is stored when the instrument is switched off.
AUTOMATIC TARE
Press on the button to run the automatic tare.
The automatic tare command is not executed under the following conditions:
• Unstable weight (the weight does not stabilize within 3 seconds of the automatic tare command).
• Gross weight equal to or greater than the maximum capacity of the weighing system.
The tare value is stored when the instrument is switched off.
DELETE TARE
This operation deletes the previously entered tare value.
Page 20
J
PRINT
This operation allows to print the weight. (if the printer protocol is selected, the shown recipe is prepared).
This operation is always executable; the conditions necessary for totalizing the weight are not checked.
12/11/2015 10:30
OPERATOR ID 1
GROSS WEIGHT 211.5 kg
TARE 2.5 kg
NET WEIGHT 209.0 kg
Page 21
RECIPE BATCHING SELECTION
JJJ
Press on the recipe dial, in the instrument main screen, to access the active recipe selection screen.
The available recipes are displayed, divided into
pages of 4 elements. Press on the recipe line you
want to activate. To quickly select an item in the
archive, hold down the page change key for a long
time, and the recipe index setting screen will be
displayed. After setting the index, the page where
the required element is present is automatically
displayed.
The recipes can also be selected via inputs.
PROGRAMMING THE RECIPE CYCLE NUMBER
From the general screen, it is possible to access the programming of the number of cycles that will be
automatically executed following a START batching.
• The number of cycles can be set from 1 to 99; by setting the value 99, the batching cycle is repeated
indefinitely until the manual stop by the operator.
• At the end of the programmed cycle sequence, the value of this parameter is automatically reset to 1.
RECIPE CORRECTION FACTOR PROGRAMMING
From the general screen, it is possible to access the recipe correction factor programming.
According to the “Correction Mode” parameter in the “Batching Parameters” menu, the correction
factor is managed in the following ways:
• FACTOR - The correction of the recipe is performed by entering a multiplying factor of the setpoints.
Parameter programmable from 0.01 to 10.00, which allows you to reduce up to one cent or multiply
up to ten times the total weight of the recipe.
• WEIGHT - The correction of the recipe is performed by entering the total weight value that you
wish to dose, so the recipe setpoints will be automatically recalculated to obtain the desired total.
Changing the correction value creates the printout of the recipe with the recalculated batching set point,
if the “PRINT DOS” parameter is “ON”.
Page 22
J
BATCHING
Press the START operation button to start the batching, so the following screen is displayed:
Indication of batched
weight and setpoint
Batching
Indications of the
recipe status
reports
Operation key F1
Operation key F4
Operation key F2
OPERATION KEYS
Batching stop command. Press this command to end the Batching procedure and
return to the basic screen display. The stop command must be pressed for the time
programmed in the “Timer stop dos.” Parameter, which can be set in the Batching
parameters menu. The confirmation of the Batching stop operation is always requested.
Batching pause command. The batching is temporarily suspended; to resume the
batching, press the start button.
Stop control of the sequence of batching cycles. The current cycle is ended and the
batching procedure is terminated.
Batching start command. Following a batching pause command or after a blackout,
the batching is suspended or interrupted.
Operation key F3
Alarm acknowledgment command. Following a batching alarm, this command
allows to continue the batching in progress.
Confirmation command, it is managed in the following conditions:
• During a "wait for confirmation" activity, to continue with the next step in the
recipe, press the confirmation button.
• During a "manual batching" activity, to confirm the weight loaded manually
and to continue with the next step of the recipe it is necessary to press the
confirmation key.
• At the beginning of a total unloading activity, in case of "end batching mode"
parameter = CONFIRMATION. In this condition the dosed net weight is displayed (in the area dedicated to the display of the batching set), to activate the
discharge output, it is necessary to press the confirmation key.
Page 23
TECHNICAL NOTES REGARDING BATCHING
• At the beginning of the batching, different compatibility checks are performed between the selected
recipe and the system conditions. Any errors are promptly selected.
• In the case of Coarse and Fine batching, the slow-down set point is so determined: SET - PRESET
- Inflight.
• At the end of the batching in case of active tolerance control, it is verified (at stable weight and
after any waiting time) that the net batching is within the tolerance window, calculated as: SET
+/- TOLERANCE.
• The net weight is displayed during the set unloading steps, while the gross weight is displayed
during the total unloading steps.
• After a blackout all the partial batching data are maintained and the next time the instrument is
switched on, the suspended batching can be resumed from the same point.
Area 1
Area 2
Area 3
Batching REPORTS - AREA 1
STOPNo batching operation in progress.
PAUSETemporarily suspended batching.
IN PROGRESSBatching in progress.
Alarm condition; the description of the alarm condition is displayed in
ALARM
COARSECoarse batching phase in progress.
AREA 2.
Batching SIGNALS - AREA 2
Batching REPORTS - AREA 2
BATCHING
REPORTS
FINEFine batching phase in progress
MANUALManual batching phase in progress
WAITING TIME
WAITING FOR CONFIRMATION
CHECKING INPUTCheck phase of the input status
Page 24
Waiting phase for the stabilization of the weight and waiting time at the
end of the batching of the product.
Phase of waiting for confirmation; to continue with the next step of the
recipe you need to press the confirmation button.
RESTART
Restart phase after a blackout; the instrument is waiting to receive the
batching restart command or the suspended batching stop command.
UNLOADINGUnloading phase in progress (set unloading or total unloading).
RECIPE NOT SELECTED
EMPTY RECIPE
ERR. OF INFLIGHT > SET
WEIGHT ERROR
ERR. OF MIN. WEIGHT
TOLERANCE
NO PRODUCT
Recipe error not selected; this condition is checked during the start of the
batching.
Unscheduled recipe error; this condition is checked during the start of
the batching.
Error concerning the product parameters (INFLIGHT> SET); this condition
is checked during the start of the batching.
Error concerning the batching recipe; the total weight of the recipe is
greater than the weighing system capacity. This condition is checked
during the start of the batching.
Error in the zero control of the weight; the gross weight in the scale is
greater than the set zero set point. This condition is checked during the
start of the batching.
Batching error of the product out of tolerance; the net batched product
does not fall within the tolerance window (calculated as SET +/- TOLERANCE).
Error of lack of product during the batching. During the loading phase it
is checked that the product is actually dosed; if the weight remains stable for the time set in the TIMEOUT parameter, this error is displayed.
Batching REPORTS - AREA 3
Batching status bar in progress.
BATCHING INPUTS AND OUTPUTS
The operation of the batching inputs and outputs can be programmed from the instrument setup menu.
The maximum number of inputs and outputs depends on the number of additional I/O modules connected (maximum 4 modules that can be connected to the instrument):
• 38 configurable outputs (6 internal outputs to the instrument plus 8 outputs for each additional I/O
module). Depending on the type of activity, up to 2 outputs can be selected.
• 22 configurable inputs (6 internal inputs to the instrument plus 4 inputs for each additional I/O
module). Depending on the
type of activity, an input can be
associated.
EXAMPLE OF PRINTING AT THE END
OF BATCHING CYCLE
02/02/1610:30
ID1
Cacao biscuits
CYCLE01/10
PROD.SETDOSED
Flour[*]
TOTAL
• If enabled, printing takes place at the end of each
batching step.
200.00199.85
300.00301.20
600.00601.40
• The identification code of the instrument is printed (para-
kgkg
Sugar
100.00100.35
meter “Address” of COM2); if the identification code is
0, the relevant line is not printed.
Cacao
Page 25
• In case of product dosed out of tolerance, an asterisk is printed.
• The indicated set corresponds to the actual set multiplied by the possible correction factor.
MENU STRUCTURE
The menu screens are subdivided into two types: command menu and parameter menu, depending on
the context and structure of the data programming menus.
By pressing the key containing the description, you enter the relevant submenu.
Each screen allows you to view up to 6 parameters; the display of the other parameters will take place
via page buttons that will appear automatically.
Pressing the key that contains the description of the parameter, you access the relevant setting.
OPERATION KEYS
F1 operation key; it switches to the previous menu page (this button is
only displayed if the menu has more than 6 items, it is never displayed
on the first page).
Operation key F2; it switches to the next menu page (this button is
only displayed if the menu has more than 6 items, it is never displayed
on the last page).
Page 26
Exit button from the menu (the top-level menu or the base screen is
displayed.
USER MENU
USER MENU (COMMANDS)
MESSAGEDESCRIPTION
ACT./REC. ARCHIVEProgramming menu of activities and recipes.
HARDWARE TESTMenu of the instrument's hardware test operation procedures.
INPUTProgramming menu for the operation of the inputs.
SETUP MENU
ACTIVITY/ RECIPE PARAMETER PROGRAMMING MENU
MESSAGEDESCRIPTION
ACTIVITYIt accesses the management of the activity archive.
RECIPEIt accesses the management of the recipe archive.
ENTER ACTIVITYIt enters a new activity in the archive.
EDIT ACTIVITYIt edits an activity in the archive.
Menu for viewing and managing totals, divided by activity and
recipes.
Programming menu of the selections related to the operation of the
recipe.
Summary screen of the setup parameters and the parameters identifying the firmware (program code and software version number).
Menu of the programmable parameters that determine the start-up
of the instrument.
DELETE ACTIVITYIt permanently deletes an activity in the archive.
RESET ARCHIVE
It deletes all activities in the archive (function protected by password
13205).
Page 27
ENTER AND EDIT ACTIVITY
MessageDescriptionTypeUnits DefaultRange
Max. 14
Alphanumeric name of the activity.
Name
Activity
Selection of the activity type. (*)
Type of activity
Fine is the amount of weight that
you want to batch at reduced
speed. This value must be less than
the set value set in the recipe. By
Fine
programming this value to zero, the
component is batched entirely at
Coarse output.
Parameter used only for "activity
type”” = AUTO DOS..
The Inflight is the amount of material
that by inertia of the system is
added to the batched material
after the Fine batching output has
switched off.
The settable value must be less
Inflight
than the set value set. Load output
deactivation set point = value set Inflight
Parameter used only for "activity
type”” = AUTO DOS..
This parameter, if different from
zero, applies an automatic
correction of the inflight value. At
the end of each automatic batching
activity, the difference between
the batched value and the set is
checked. This difference is used
to calculate the new inflight value,
based on the percentage set in this
parameter. The value obtained is
Inflight compensation
limited to 10% of the activity set
value and in any case cannot be
higher than the tolerance value.
The set tolerance value is checked
at the end of the batching. If the
net weight batched does not fall
within the range determined by set
+ tolerance and set - tolerance, an
error message will be displayed
until the alarm is manually silenced.
By programming the tolerance value
Tolerance
to 0, the control of the batched
net weight is disabled (during the
batching sequence the outputs will
not be re-connected).
Parameter used only for "activity
type”” = AUTO DOS..
This function, if enabled, allows the
fine output to be activated again
Fine
if the batched net weight is lower
than the value set in the recipe - the
reactivation
tolerance value.
This parameter represents the
time that passes between the
deactivation of the outputs at the
end of the batching and the stable
weight control for the acquisition
of the net weight dosed. It is used
to wait for the fall of the flying
Waiting
material. By programming the value
to zero, no timing is performed.
Parameter used only for "activity
type”” = AUTO DOS..
This parameter represents the
maximum time for the waiting of
the product during the batching
of a component, before the alarm
of lack of product is enabled. By
programming the zero value, no
Timeout
batching check is performed.
Parameter used only for "activity
type”” = AUTO DOS..
This parameter allows you to set
the fast output associated with this
activity (the same output can be
shared by several activities). By
programming this parameter to
zero, no fast output is associated
with the activity. Parameter used
Coarse output
in case of "activity type” = AUTO
DOS./ MAN DOS./ TIMER /
OUTPUT.
Numeric setup
Weight measurement unit
YES
Selection
Seconds
Numeric setup
Seconds
Numeric setup
Numeric setup
0264
(MSW)
20÷Cell capacity
[0] YES
[1] NO
00.0÷999.90266
00.0÷999.90267
00÷320268
-
0265
(LSW)
0276
Page 29
This parameter allows you to set
the Fine output associated with
this activity (the same output can
be shared by several activities).
By programming this parameter to
zero, no fast output is associated
with the activity. Parameter used
Operation logic of fast output.
Parameter used in case of "activity
type” = AUTO DOS./ MAN DOS./
Coarse
Fine output
TOT. REJ. / INT. REJ / TIMER /
output logic
OUTPUT..
Operation logic of Fine output.
Parameter used in case of "activity
type” = AUTO DOS./ MAN DOS./
logic
TOT. UNL. / INT. UNL / TIMER /
OUTPUT.
This parameter allows you to
associate an input to the activity.
Parameter used in case of "activity
type” = MAN. DOS. / TOT. UNL.
(only if the parameter END Batching
Input
MODE = CONFIRM) / INT. UNL.
(only if the parameter END Batching
MODE = CONFIRM) / EXP. CONF.
/ INPUT.
This parameter acts only if
programmed the tolerance
parameter. In case of batching
weight less than the weight value
- tolerance, the fine output is
reactivated for the time set in this
Spill ON
parameter. If the final weight value
is reached before the expiration
of this time, the fine output is still
disabled.
At the end of the reactivation of the
fine output, if the batching weight
has not yet reached the set value,
the instrument checks during the
time set in this parameter that the
Spill OFF
batching weight has been reached.
Otherwise, reactivate the fine
output.
Numeric setup
Selection
Selection
Numeric setup
Seconds
Numeric setup
Seconds
Numeric setup
00÷320269
0
0
00÷220272
00.0÷999.90273
00.0÷999.90274
[0] OPEN NO.
[1] CLOSE NO.
[0] OPEN NO.
[1] CLOSE NO.
0270
0271
(*) Each activity can be associated with one of the following functions:
• Automatic batching: Automatic batching under the load.
• Manual batching: Manual batching; to end this activity, the operator must press the confirmation
key or close the selected input.
• Total unloading: complete unloading; the scale is considered as unloaded when the gross weight
is lower than the programmable set point of unloading weight.
Page 30
• Intermediate unloading: Unloading to set; the amount of weight to be unloaded is programmed in
the recipe.
• Waiting for confirmation: Manual confirmation activity; to continue with the next recipe step, the
operator must press the confirmation key or close the selected input.
• Timer: Waiting step; the waiting time is programmed in the recipe.
• Output: Set output status. The outputs to be checked are programmed in the activity parameters;
the status that the outputs must take is programmed in the recipe (ON/OFF).
• Input: waiting input activation step.. The input to be checked is programmed in the activity para-
meters; the status to be controlled is programmed in the recipe (ON/OFF).
• Unloading batching: Automatic batching at the discharge of a single component.
DELETE ACTIVITY
The activities stored in the archive are displayed, divided into pages of 4 elements. To activate fast
selection, long-press the page change key. Press on the line of the activity that you wish to delete; the
confirmation of the deletion operation is always requested.
If the deleted activity is recalled in an archived recipe, this recipe is automatically reset and
must be re-programmed by the operator.
RESET ACTIVITY ARCHIVE
Procedure of total cancellation of the activity archive; this function is protected by password (13205).
RECIPE ARCHIVE
RECIPE ARCHIVE (commands)
Message Description
ENTER RECIPE It enters a new recipe in the archive.
EDIT RECIPE It edits a recipe in the archive.
DELETE RECIPE It permanently deletes a recipe in the archive.
RESET ARCHIVE It deletes all recipes in the archive (function protected by password 13205).
Page 31
ENTER/EDIT RECIPE (parameters, commands)
Message Description Type Units Default Range
Recipe
name
Step no.Number of recipe steps (*)
RECIPE
SETUP
(*) In case of unloading batching, the parameter “Steps no.” must be set to 1; also in the recipe the
selected activity must be of the “UNL.DOS.” type.
ENTER/EDIT RECIPE
Step number being
programmed
Alphanumeric name of the
recipe.
It accesses the recipe
parameter setting function.
Alphanumeric
setup
Numeric
setup
Activity in the archive
0 0 - 20 0209
Max. 14
alphanumeric
characters
Alarm: total recipe
greater than the
capacity of cells
Fieldbus
address
0202-
0208
Total recipe
OPERATION KEYS
It aborts the recipe programming procedure without saving the changes
made.
It selects the previous step or the next step in the recipe.
It saves the recipe and returns to the main screen.
DELETE RECIPE
The recipes stored in the archive are displayed,
divided into pages of 4 elements. Press on the
recipe line that you wish to delete; the confirmation of the deletion operation is always requested.
To activate fast selection, long-press the page
change key.
RESET RECIPE ARCHIVE
Procedure of total cancellation of the recipe archive; this function is protected by password (13205).
ACTIVITY/RECIPE TOTAL MENU
ACTIVITY/RECIPE TOTALS (commands)
MessageDescription
ACTIVITYIt accesses the management of the totals by activity.
RECIPESIt accesses the management of the totals by recipe.
ACTIVITY/RECIPE TOTALS MANAGEMENT (commands)
Message Description
DISPLAY TOTALS It accesses the display of the totals by recipe/activity.
RESET TOTALS It deletes completely the totals.
DISPLAY TOTALS
The totals of the activities/recipes in the archive are displayed; in the case of displaying the totals by
activity, only the activities of batching upon the load, manual batching and batching upon unloading
are shown. Press on the line of the total you wish to view. To activate fast selection, long-press the page
change key.
Page 33
OPERATION KEYS
It deletes the activity/recipe total displayed.
Totals print function.
EXAMPLE OF TOTAL PRINT
03/08/1711:09
ID1
ACTIVITY TOTAL
Sugar
230,587 kg
Cacao
750,879 kg
Flour 00
826,742 kg
Salt
326,820 kg
TOTAL
AMOUNT
2135.028 kg
• Only non-zero totals are printed.
• The identification code of the instrument is printed (“Address” parameter of COM2); if the identifi-
cation code is 0, the relevant line is not printed.
RESET TOTALS
Procedure for the complete cancellation of the totals; confirmation of the deletion operation is always
requested.
Page 34
BATCHING PARAMETERS MENU
BATCHING PARAMETERS
Message Description Type Units Default Range
This parameter represents the
control set point of zero.
This set point determines the
maximum weight value allowed at
the start of the BATCHING; with a
higher weight, the batching will not
start and the instrument will display
Zero Control
an error message. To disable this
control, reset the value of this set
point.
This parameter represents the empty
scale set point. This set point is
checked during unloading: when
the weight falls below this set point,
the unloading phase is considered
complete.
Unloading threshold
This parameter represents the
time of the unloading extension:
during complete unloading (not set)
when the weight becomes lower
than empty scale tset point, the
unloading phase is prolonged for
Unloading Prol.
the programmed time, to ensure the
complete unloading of the tank.
This parameter represents the
unloading end time: at the end of
the unloading phase, this time is
waited before moving on to the
next step of the recipe (in the case
of intermediate unloading), or
before deactivating the cycle output
Unloading end
in progress (in the case of final
unloading of the batching cycle).
This parameter represents the end
of cycle time that passes from
the end of the last phase of the
batching cycle to the beginning of
a new batching cycle (in case of
Cycle end
automatic repetition of the cycles)
or the enabling of a new START
batching command.
This parameter represents the
activation time of the end of cycle
output, which is activated at the end
of the last phase of the batching
output t.
Cycle end
cycle.
Numeric setup
Weight measurement unit
unit
Numeric setup
Weight measurement
Seconds
Numeric setup
Seconds
Numeric setup
Seconds
Numeric setup
setup
Numeric
Seconds
0
0
00.0÷999.90105
00.0÷999.90106
00.0÷999.90107
00.0÷999.90108
0÷Load Cell
capacity
0÷Load Cell
capacity
Fieldbus
address
0101
(MSW)
-
0102
(LSW)
0103
(MSW)
-
0104
(LSW)
Page 35
In the case of CONFIRM selection,
at the beginning of the total
unloading activity, the net weight
batched in the recipe is displayed.
To activate the unloading output,
the operator must press the
confirmation key or close the input
Batch. End mode
set in the activity.
In case of FACTOR selection,
the correction of the recipe
is performed by entering a
multiplication factor of the sets
(from 0.01 to 10.00). In the case of
WEIGHT selection, the correction of
the recipe is performed by entering
Corr. mode
the total weight value that you wish
to batch; the recipe sets will be
automatically recalculated to obtain
the desired total.
This parameter represents the
minimum level set point compared
to the gross weight. The minimum
level
Minimum
Maximum
Min. level
level output activates when the
weight is lower than the set value.
This parameter represents the
maximum level set point compared
to the gross weight. The maximum
level
level output activates when the
weight is greater than the set value.
Output associated with the minimum
level value. "NA" operating logic,
activated when the gross weight is
lower than the minimum level set
output
point. In the case of a parameter set
to 0, no output is associated.
Output associated with the
maximum level value. "NA"
operating logic, activated when
the gross weight is greater than the
maximum level set point. In the case
of a parameter set to 0, no output is
Max. level output
associated.
Output associated with the
execution of a batching cycle.
"NA" operating logic, it is activated
with the start of the batching
and deactivated at the end of
the batching. In the case of a
Cycle output
parameter set to 0, no output is
associated.
Selection
Selection
setup
Numeric
setup
Numeric
Numeric setup
Numeric setup
Numeric setup
[0] NORMAL
0
[1] CONFIRM
0
0
0
00÷320115
00÷320116
00÷320117
[0] FACTOR
[1] WEIGHT
0÷Load Cell
capacity
0÷Load Cell
capacity
0109
0110
0111
(MSW)
-
0112
(LSW)
0113
(MSW)
-
0114
(LSW)
Page 36
Output associated with the end of
a batching cycle. "NA” operating
logic, it is activated at the end of
the last phase of the batching cycle
and remains ON for the cycle end
time (settable parameter). In the
case of a parameter set to 0, no
Cycle end output
output is associated.
This output activates in an alarm
condition ("NA” operating logic). In
Alarm
the case of a parameter set to 0, no
output
output is associated.
Numeric setup
setup
Numeric
00÷320118
00÷320119
This parameter allows to select the
Timer
Batching
Batching
Print
unit of measurement of the time for
mode
the TIMER activity.
Value in seconds that indicates how
long it should be kept the STOP
command (both from touchscreen
that by logic input) to ensure that
Stop Timer
the batching is interrupted.
Autotare execution selection at the
beginning of the batching.
start tare
Automatic print selection at the end
of each batching cycle.
Batch.
Selection
setup
Numeric
Selection
Selection
0
30.0 ÷ 10.00121
OFF
OFF
[0] SECONDS
[1] MINUTES
[0] OFF
[1] ON
[0] OFF
[1] ON
0120
0151
0152
Page 37
INSTRUMENT DATA DISPLAY
The instrument data display screen is divided into 4 sections:
• FIRMWARE: Code and firmware version installed; if assistance is required, it’s important to communicate these data.
• PARAMETERS: Indication of the set full scale, the status of the internal jumper enabling metrological
calibration and the value of the power supply voltage of the instrument.
• CONNECTIONS: Specification of the setup fieldbus type and the set fieldbus address.
• OPTIONS: Indication of the type of analog output setup, of the presence of additional memory
(None, Alibi Memory).
In the case of PROFINET fieldbus, this screen displays the IP address programmed by the
PLC; this field is updated only when the instrument is switched on. Following a change of
the IP address from the PLC, the instrument must be switched off and on again to display the
correct value.
Page 38
MESSAGEDESCRIPTION
LOAD CELLS
HARDWARE TEST MENU
HARDWARE TEST (COMMANDS)
Indication of the weight with 10x resolution, of the signal acquired by
the instrument in mV/V and of the percentage with respect to the full
scale of the weighing system.
MEMORIES
INPUT / OUTPUT
COMMUNICATION PORTS
ANALOG OUTPUT
Functional test of the optional memory (not present in case of no memory).
Display of the status of the digital inputs, and manual setting of the
status of the digital outputs. In the case of setup external input/output
modules, the outputs of the external modules are also managed.
Test of setup ON communication ports. The test consists in resending
the string received from the relevant serial line (echo test); moreover,
the number of strings and the number of characters received in the last
communication are displayed.
Function test of the analog output with manual setting at preset percentage values (not present in the absence of the analog output).
(**) (**) The “SEL. REC.” selection is only available for inputs 3, 4, 5 and 6.
(*) One of the following functions can be associated with each logical input:
• Zero: It performs zero calibration.
• Tare: It performs auto tare.
• Del. Tare: It deletes the tare.
• Print: It prints the weight.
• Alarm reset: It resets the alarm condition.
• Start: Batching start command.
• Pause: Batching pause command.
• Restart: Batching restart command, following a batching pause command or following a blackout.
• Stop: Batching stop command (the input must be kept ON for the number of seconds set in the
batching parameter “Batch. Stop Timer”).
• Net/gross weight selection: It switches weight display (net/gross weight).
• Recipe selection: Selection of the active recipe; function available only for inputs 3, 4, 5 and 6.
Where the input 3 represents the least significant bit and the input 6 the most significant bit of the
selection (i.e. Input 3 ON + Input 4 ON + Input 5 OFF + Input 6 OFF = recipe 3).
Programming menu of parameters related to weight calibration. In this
menu it is also possible, through a specific selection, to execute the sample
weight calibration function or the table calibration function. (*)
Parameter programming menu for the analog output. In this menu it is also
possible to execute the adjustment function of the analog output. This menu
is only displayed in the case of an active analog option.
Parameter programming menu for serial communication ports.
Programming menu of the metrological weighing parameters. (*)
Programming menu of the weight filter value. The filter factor can be
selected from 9 preset values, or the individual parameters that determine
the behavior of the weight filter (output rate, average number, etc.) can be
setup manually.
Parameter programming menu related to the functional characteristics of
the instrument (stand-by, password to access the setup menu, etc.).
Calendar/clock adjustment function. Access to the adjustment can be
locked with a special password tath can be set in this menu.
Setup memory upload/download function.
Page 41
The following pages describe all the parameters that can be set. At the end of the description of each
parameter, if any, the fieldbus address corresponding to the parameter is shown. If the parameter is of
the selectable type, the value to be entered in the register for the desired selection is indicated between
“[ ]”.
CALIBRATION MENU
MEASUREMENT UNIT
Measurement unit related to the weight value; it appears on the main screen and in all weight parameter settings.
Selectable values:
[0] kg, [1] g [2] t [3]
lb, [4] N, [5] kN
Default: kg
DIVISION VALUE [1101÷1102]*
Value of a single division, expressed in the chosen measurement unit. The relationship between the
system’s capacity and the division value constitutes the resolution of the system (number of divisions).
Following the change in the division value, if the system capacity is not changed, the weight calibration
is automatically corrected.
Selectable values:
0.0001 - 0.0002 - 0.0005
0.001 - 0.002 - 0.005
0.01 - 0.02 - 0.05
0.1 - 0.2 - 0.5
1 -2 - 5
10 - 20 - 50
Default: 1
(*) The setting of the division values via fieldbus takes place differently from that made by the instrument.
Refer to the 1101 and 1102 addresses of the MODBUS register table.
LOAD CELLS CAPACITY [1103-1104]
It defines the value corresponding to the sum of the nominal capacity of the load cells expressed in the selected
measurement unit. In the case of systems with only one load cell and “N” fixed supports, enter the value of the
cell capacity for the total number of supports. This data constitutes the full scale value of the weighing system.
Following the change of the parameter value, the theoretical weight calibration is recalculated.
Values: from 0 to 999999
Default: 0
LOAD CELL SENSITIVITY [1105]
Set the value corresponding to the average sensitivity of the load cells, in mV/V. Values between 0.0
and 4 mV/V are accepted. If no value is programmed, 2 mV/V is assumed.
Following the change of the sensitivity value, the theoretical weight calibration is performed.
Values: from 0.0001 to 4.0000 mV/V
Default: 2.0000
Page 42
FULL SCALE [1301-1302]
Programming of the capacity (net) of the weighing system.
Values: from 0 to 999999
Default: 0
FIXED TARE OF WEIGHING SYSTEM [1106-1107]
Programming of the fixed tare value of the weighing system.
Values: from 0 to 999999
Default: 0
CALIBRATION TYPE
Selection of the type of calibration. Upon confirmation, one of the following procedures is started.
DEAD WEIGHT CALIBRATION [501 ÷ 503]
Calibration of Zero and Full Scale up to 5 points of linearization with the use of sample weights.
TABLE CALIBRATION [1151 ÷ 1172]
It allows manual programming of up to 5 calibration points. The values correspond to those determined
by the linearization procedure with sample weights. In this way, it is possible to copy the calibration
values made with sample weights.
Page 43
EXAMPLE OF SETUP/CALIBRATION
By setting the parameters listed above, the theoretical calibration of the full scale of the MCT 1302 is
carried out. This procedure must be completed with the zero calibration described below (*page 45).
The procedure guarantees, in the absence of mechanical problems, a good accuracy of the system
(max. error <1% F.S.).
When the division value selection is changed, the theoretical full scale calibration is automatically
recalculated. Selections incompatible with the calibration parameters or with the calibration present
in the memory are not accepted.
It is necessary to weigh a tank, with a weight of 750 Kg and with a capacity of 1000 liters, containing
a product with a specific weight of 1.3 Kg/dm
3
whose weight is to be read with a display resolution
of 0.2 kg.
Before proceeding with the setup, make sure that the load cells are connected correctly to the unit and
that the tank is empty; you can proceed with the setting of the parameters.
You can use:
#3 load cells capacity 1000 Kg
Sensitivity of 2.0015, 2.0008 and 1.9998 mV/V respectively (average value = 2.0007 mV/V)
Set the following values in the setup parameters:
Measurement unit = kg
Division value = 0.2
Load Cell capacity = 3000
Load Cell sensitivity = 2.0007
Full scale = 1500
Dead Weight = 0
Make sure that the value of the signal read in the HARDWARE TEST, LOAD CELLS menu corresponds
to the tare weight of the system according to the following proportion:
3000:2.0007=750:X
Where X is the value of the signal expressed in mV/V corresponding to the theoretical value of the
weight of the empty tank. The value should be about 0.5 mV/V
(*) At this point, you can proceed to the calibration described in the next paragraph or you can exit
the setup menu by saving the data entered.
The instrument should indicate the value corresponding to the weight of the empty tank (for example
756.8).
You can enter the setup menu again and enter the value of the weight read in the “Pretare” parameter
and enter the value 756.8
Exit the setup menu again by saving the data.
For greater accuracy, prepare sample weights or pre-weighed material on a certified scale and proceed
with the calibration described in the next paragraph.
Page 44
CALIBRATION WITH SAMPLE WEIGHTS
The calibration method described here must be carried out using sample weights and/or product preweighed on a sample scale.
Always perform zero calibration before proceeding with the full scale calibration.
CAUTION! If the instrument is switched off without leaving the set-up menu, the programmed settings
are not stored.
If, after calibration, the system has linearity errors, it must be checked that the weighed structure
is completely free from mechanical constraints.
OPERATION KEYSDESCRIPTION
Zero calibration: carry out the operation with the scale unloaded but
complete with the tare weight, with stabilized weight. The displayed
weight must be reset; this operation can be repeated several times.
Calibration with sample weight: before carrying out this operation,
load the sample weight on the scale and wait for it to stabilize. The
display shows the measured value to be calibrated; set the corresponding weight value via touchscreen. If the set value is higher than the
resolution offered by the instrument, the setting is not accepted and
the display shows an error message for a few seconds. This operation
can be repeated several times.
Activate linearization with sample weights. Up to 5 linearization
points on a positive scale can be set. During the linearization phase,
in the weight dial the progressive linearization points are displayed
(eg "LIN 1") alternating with the current weight.
Press the F.S button to set the value of the sample weight loaded on the
scale. To end the procedure, press the LIN button again; it is possible
to set a number of points lower than 5.
Exit button from the calibration function.
Page 45
TABLE CALIBRATION
It allows you to manually program up to 5 calibration points, in addition to the zero value. The values
correspond to those determined by the linearization procedure with sample weights. In this way, it is
possible to display the values determined automatically with this procedure, or modify and program
them according to preset values.
TABLE CALIBRATION
MESSAGEDESCRIPTION UNITS
Zero signalSignal value corresponding to the zero scale.mV/V
Value of weight P1Weight value corresponding to the 1st calibration
point.
Signal P1Signal value corresponding to the 1st calibration point. mV/V
Value of weight P2Weight value corresponding to the 2nd calibration
point.
Signal P2
Value of weight P3Weight value corresponding to the 3rd calibration
Signal P3
Value of weight P4Weight value corresponding to the 4th calibration
Signal P4Signal value corresponding to the 4th calibration point. mV/V
Value of weight P5Weight value corresponding to the 5th calibration
Signal P5Signal value corresponding to the 5th calibration point. mV/V
Signal value corresponding to the 2nd calibration
point.
point.
Signal value corresponding to the 3rd calibration
point.
point.
point.
Weight measurement
unit
Weight measurement
unit
mV/V
Weight measurement
unit
mV/V
Weight measurement
unit
Weight measurement
unit
Zero Signal
Acquisition
The values programmed to zero are not taken into account. Table calibration is automatically canceled
when a new theoretical calibration or a sample weight is performed.
Following the execution of the zero signal acquisition function, the signals in the table are recalculated.
At each signal value an offset is added, obtained from the difference between the new acquired zero
signal and the old zero signal value.
Page 46
Signal acquisition function, in mV/V, corresponding to
the zero scale
ANALOG OUTPUT PARAMETERS (OPTIONAL)
ANALOG OUTPUT RANGE [1506]
Selection of the analog output range.
Selectable value:
0÷10 VDC [0]
0÷5 VDC [1]
4÷20 mA [2]
0÷20 mA [3]
Default: 0÷10 VDC
ANALOG OUTPUT OPERATION MODE [1505]
Selection of the value to be associated to the analog output, corresponding to the net or gross weight
or to the peak value.
Selectable value:
NET [0]
GROSS [1]
Default: NET
ZERO OFFSET[1501-1502]
Analog output zero offset. If set, the zero value of the analog output corresponds to the weight value
set in this parameter.
Analog value to subtract referred to the analog output full scale.
FULL SCALE [1503-1504]
It is the weight corresponding to the full scale of the analogue output.
Value that can be set from 0 to Capacity
Default: Capacity
OUTPUT ADJUSTMENT: This parameter adjusts the zero and full scale value of the selected analog output
so that the PLC and the display of the MCT 1302 will indicate the same weight.
- ZERO ADJUSTMENT:
Measure the analog output value with a tester to perform zero calibration (0).
Use the + and - keys to adjust the analog output. Press and hold the key for a quick change. Press the
SAVE key to store the data.
- F.S. ADJUSTMENT: [1508]
Measure the analog output value with a tester to perform full scale calibration (FS).
Use the + and - keys to adjust the analog output. Press and hold the key for a quick change. Press the
SAVE key to store the data.
N.B. The analog output is calibrated at the factory for each selectable range. This procedure is
available to the user for further adjustment, for each selectable range. In case of complete reset of
the setup memory, the factory calibration is re-established.
Page 47
COMMUNICATION PORTS
This menu allows you to setup the serial ports COM1, COM2 and COM3 and the communication
parameters. The instrument has two independent serial ports:
COM1 always with RS232 interface.
COM2 can alternately mount the following interfaces: RS485, ETHERCAT, ETHERNET, ETHERNET IP,
PROFINET.
COM3 always with RS485 interface: it is normally used to connect external I/O modules.
COM1 (RS232)
VALUE SENT
Selection of the value sent on RS 232 output.
Selectable values:
NET, GROSS
Default: NET
PROTOCOL
It defines how to use the RS232 serial port:
Selectable values:
NONE Serial communication deactivated
CONTINUOUS: Continuous sending of the weight string. It can be used, for example, to drive a
weight repeater. See details in the appropriate paragraph.
PRINT: ASCII protocol. See details in the appropriate paragraph.
Default: NONE
BAUD RATE
It defines the baudrate of the RS232 serial port:
The value must be set to the same value as the PC/PLC or remote display.
PARAMETERS COM 2 WHEN PROFINET / ETHERCAT IS PRESENT
ENABLING FIELDBUS
Enabling the PROFINET / ETHERCAT; if OFF, no error message regarding the fieldbus communication
is displayed:
Selectable values:
OFF, ON
Default: OFF
INPUT AREA SIZE
Dimension of the input area for fieldbus (values expressed in bytes).
Selectable values:
32, 64, 96, 128
Default: 128
OUTPUT AREA SIZE
Dimension of the output area for fieldbus (values expressed in bytes).
Selectable values:
32, 64, 96, 128
Default: 128
In the case of an PROFINET fieldbus, the XML “GSDML-V2.3-HILSCHER-NIC 50-RE PNS 32-20160122.
xml” configuration file is provided. The size of the input and output areas set in the PLC (possible selections: 32, 64, 96 or 128 bytes) must correspond to the size of the input and output areas selected in
the instrument (“INP.ADJ.” And “OUT.ADJ.” Parameters).
The instruments are supplied with the parameter “Profinet Name” not setup and with an IP address of
0.0.0.0.
In case of ETHERCAT fieldbus: the devices must be connected with a ring type (as specified by EtherCAT),
so refer to the installation manual for the use of INPUT and OUTPUT ports.
4 different XML setting files are provided:
“Hilscher NIC 50-RE ECS V2.2 32 Byte.xml” (input area 32 bytes, output area 32 bytes).
“Hilscher NIC 50-RE ECS V2.2 64 Byte.xml” (input area 64 bytes, output area 64 bytes).
“Hilscher NIC 50-RE ECS V2.2 96 Byte.xml” (input area 96 bytes, output area 96 bytes).
“Hilscher NIC 50-RE ECS V2.2 128 Byte.xml” (input area 128 bytes, output area 128 bytes).
The file that corresponds to the size of the input and output areas selected in the device must be imported
into the PLC (for example, if the device is set to ADJ. SET. = 128 and OUT.ADJ.=128, the file “Hilscher
NIC 50-RE ECS V2.2 128 Byte.xml” must be imported into the PLC). Multiple files with different sizes
can be imported, but, in this case, it will not be possible to perform the automatic search and setup of
devices on the network.
Page 51
PARAMETERS COM 2 WHEN ETHERNET IP IS PRESENT
ENABLING FIELDBUS
Enabling the ETHERNET IP; if OFF, no error message regarding the fieldbus communication is displayed:
Selectable values:
OFF, ON
Default: OFF
IP ADDRESS
IP protocol address ETHERNET.
Values from 0.0.0.0 to 255.255.255.255
Default: 0.0.0.0
SUBNET MASK
Mask protocol subnet ETHERNET IP.
Values from 0.0.0.0 to 255.255.255.255
Default: 0.0.0.0
INPUT AREA
Dimension of the input area for fieldbus (values expressed in bytes).
Selectable values:
32, 64, 96, 128
Default: 128
OUTPUT AREA
Dimension of the output area for fieldbus (values expressed in bytes).
Selectable values:
32, 64, 96, 128
Default: 128
In the case of an ETHERNET IP fieldbus, the EDS configuration file “HILSCHER NIC 50-RE EIS V1.1.EDS” is
provided. The size of the input and output areas set in the PLC (default input area of 128 bytes, default
output area of 128 bytes) must match the size of the input and output areas selected in the instrument
(“INP.ADJ.” Parameters and “OUT.ADJ.”).
Page 52
PARAMETERS COM 2 WHEN ETHERNET IS PRESENT
VALUE SENT
Selection of the value sent on ETHERNET output.
Selectable values:
NET, GROSS
Default: NET
PROTOCOL
It defines how to use the ETHERNET serial port:
Selectable values:
NONE Serial communication OFF
CONTINUOUS: Continuous sending of the weight string. It can be used, for example, to drive a
weight repeater. See details in the appropriate paragraph.
ON DEMAND: When the operator presses the relevant front button or via Input 2, a weight string
is sent. The command is accepted if the weight is stable. Between two successive transmissions the
weight must undergo a variation equal to at least 20 divisions.
AUTO: A weight string is automatically sent when the weight stabilizes at a value above the minimum
weight (20 divisions).
SLAVE: ASCII protocol. See details in the appropriate paragraph.
MODBUS/TCP: MODBUS TCP protocol. See details in the appropriate paragraph.
Default: NONE
IP ADDRESS
IP protocol address ETHERNET.
Values from 0.0.0.0 to 255.255.255.255
Default: 192.168.0.201
SUBNET MASK
Mask protocol subnet ETHERNET.
Values from 0.0.0.0 to 255.255.255.255
Default: 255.255.255.0
GATEWAY
Protocol gateway ETHERNET.
Values from 0.0.0.0 to 255.255.255.255
Default: 192.168.0.1
PORT
Communication port for ETHERNET protocol.
Value from 1 to 65535
Default: 1800
Page 53
PARAMETERS COM 2 WHEN PROFIBUS IS PRESENT
ENABLING FIELDBUS
Enabling the PROFIBUS fieldbus; if OFF, no error message regarding the fieldbus communication is
displayed:
Selectable values:
OFF, ON
Default: OFF
PROFIBUS ADDRESS
Programming of the address used in the PROFIBUS protocol.
Value: from 0 to 126
Default: 01
INPUT AREA
Dimension of the input area for fieldbus (values expressed in bytes).
Selectable values:
32, 64, 96, 128
Default: 128
OUTPUT AREA
Dimension of the output area for fieldbus (values expressed in bytes).
Selectable values:
32, 64, 96, 128
Default: 128
Page 54
COM 3 / RS485
NUMBER OF MODULE
Number of input/output modules managed by the instrument.
Selectable values:
0÷4
Default: 0
BAUD RATE
The communication baudrate with external and fixed input/output modules at 38400 b/s.
Page 55
WEIGHING PARAMETERS
The parameters allow to adjust the acquisition and updating times of the display and the manual or
automatic resetting of the instrument.
STABILITY FACTOR [1303]
This parameter defines the number of divisions necessary to consider the weight as stable. A large
number of divisions allows the instrument to quickly detect weight stability, which is necessary when
performing tare and printing commands.
N.B. If the value selected is 0, the value 0 must be set in the TIMEOUT parameter.
Selectable value
0÷4
Default: 2
AUTOZERO SETPOINT [1304-1305]
This parameter defines the value of the maximum weight that can be reset at power up.
This operation corresponds to a zero calibration of the system and is only performed if the weight is
stable and lower than the set value.
Value from 0 to Cell capacity.
Default: 0
ZERO TRACKING [1306]
The function allows to perform a momentary zero calibration, thus compensating for any thermal drift
of the weight. When the instrument is turned off, the previous Zero calibration is automatically restored.
The maximum weight that can be zeroed from this parameter is 2% of the system capacity. To disable
the function, set the value 0.
This parameter defines the number of divisions that can be reset from the touchscreen or via the associated input.
Value from 0 to 200 divisions.
Default: 100
Page 56
FILTER PARAMETERS
FILTER VALUE [1201]
This parameter adjusts not only the refresh rate of the display, but mostly the serial and analogue output.
The maximum refresh rate of the display is limited to 10 Hz
The following parameters are visible and therefore can be set, only if the parameter selection is MANUAL.
OUTPUT RATE [1202]
Settling
Time
(mS)
Freq ADC
(Hz)
Reading no.
To be set by
the
Monotony
time (mS)
To be set by
the
Oscillation
time (mS)
To be set by
the
Oscillation
range (div)
To be set by
the
With this parameter, the frequency of weight acquisition is adjusted
Selectable values:
12.5 [0]
50 [1]
100 [2]
250 [3]
Default: it depends on the value set in the Filter Value parameter
AVERAGE NUMBER [1203]
This parameter allows you setting the number of readings that the filter will use to establish the average
value.
Value from 0 to 50.
Default: it depends on the value set in the Filter Value parameter
Page 57
MONOTONY [1204]
Parameter used to stabilize the weight when continuous variation of the last digit is detected. Normally
used in case of resolution of the weight exceeding 10,000 divisions or with low sensitivity of the input
signal. Value expressed in mS.
Value from 0 to 999.
Default: it depends on the value set in the Filter Value parameter
OSCILLATION TIME [1205]
Parameter used in conjunction with oscillation Range to reduce the slow and repetitive weight changes
typical in lifting systems. Enter the value of the oscillation time expressed in mS.
Value from 0 to 9999.
Default: it depends on the value set in the Filter Value parameter
OSCILLATION RANGE [1206]
As for the parameter Oscillation time, used to reduce the oscillations. Enter the value of the oscillation
expressed in weight divisions.
Value from 0 to 99.
Default: it depends on the value set in the Filter Value parameter
Page 58
SETTING FUNCTIONAL FEATURES
STANDBY TIME [1001]
Inactivity time after which the instrument automatically decreases the brightness of the display. 0 =
function OFF.
Value from 0 to 999.
Default: 0
PASSWORD SETUP [1002]
If programmed, enter the password to access the SETUP menu. In the case of subsequent accesses, it
is no longer required to enter the password until the standby enables or the instrument is switched off.
Value from 0000 to 9999.
Default: 0000 (no Password)
LANGUAGE [1003]
It allows you selecting the language of the operator interface.
Selectable value:
ITALIAN [0]
ENGLISH [1]
Default: ITALIAN
KEY F1 [1004]
Selection of the function associated with the F1 operation key.
Selectable value:
[0] Blocked
[1] Zero
[2] Tare
[3] Delete tare
[4] Print
[5] Start
Default: Zero
KEY F2 [1005]
Selection of the function associated with the F2 operation key.
Selectable value:
[0] Blocked
[1] ON
Default: Active
KEY WEIGHT [1009]
Weight dial lock function.
Selectable value:
[0] Blocked
[1] ON
Default: Active
KEY FACTOR [1010]
Lock function of the recipe correction factor setting key.
Selectable value:
[0] Blocked
[1] ON
Default: ON
Page 60
KEY SEL. RECIPE [1011]
Block function of the dial for selecting the recipe.
Selectable value:
[0] Blocked
[1] ON
Default: ON
Page 61
MENU - CLOCK / CALENDAR
TIME /DATE
MessageDescriptionType Units DefaultRange
Adjust
date/time
Password
date/time
Adjustment function of date and
time.
Protection password setting on the
date and time adjustment menu
Function
Imp.
Numerica
00 ÷ 9999
DISPLAY CONTRAST
Fieldbus
address
Page 62
UPLOAD/DOWNLOAD FUNCTION
The TESTER 1008 must be connected to the serial COM1 (RS232) of the instrument.
This function allows you downloading or uploading the setup configuration and calibration data stored
in the instrument.
• Download function: the setup parameters of the instrument are stored in a file.
• Upload function: the instrument is configured with the setup parameters read from a file.
To use these functions, it is necessary to activate the relative procedure (“receive files” or “send
files”) in the TESTER 1008 instrument.
Page 63
SERIAL COMMUNICATION PROTOCOLS
CONTINUOUS ASCII PROTOCOL
The continuous sending is performed at the weight update frequency, compatibly with the serial
transmission baud rate. In case of communication on the Ethernet port, the frequency of the continuous
transmission is limited to 12.5 Hz.
STX<status><weight>ETX<chksum>EOT
where:
STX (start of text) = 0x02h
ETX (end of text) = 0x03h
EOT (end of transmission) = 0x04.
<status> = character coded as in the following table (bit = 1 if condition is TRUE):
Bit 7Bit 6Bit 5Bit 4Bit 3Bit 2Bit 1Bit 0
0011
<weight> = field consisting of 8 ASCII characters with the weight value justified on the right without
non-significant zeros, with any decimal point and negative sign.
The sent weight value can be the net weight, the gross weight or the peak value, based on the selection of the sent data (MODE parameter) in the setup menu of the serial communication ports (see the
relevant paragraph).
In overweight conditions, the field assumes the value: “^^^^^^^^”.
In conditions of weight greater than 99999, the field assumes the value: “_ _ _ _ _ _ _ _”.
In conditions of wrong reading, the field assumes the value: “ O-L ”.
<cheksum> = sum of control of string data. It is calculated by executing the exclusive OR (XOR) of all the
characters from STX (or from <Addr>) to ETX excluding the latter; the result of the XOR is decomposed
into 2 characters considering separately the 4 upper bits (first character) and the lower 4 bits (second
character); the 2 characters obtained are then ASCII coded;
The addresses listed in the tables follow the standard addressing specified in the reference guide of the
Modicom PI-MBUS-300 below that is an extract that helps the user to communicate with the instrument.o.
“All data addresses in Modbus messages are referenced to zero. The first occurrence of a data item is
addressed as item number zero. For example:
The coil known as ‘coil 1’ in a programmable controller is addressed as coil 0000 in the data
address field of a Modbus message.
Coil 127 decimal is addressed as coil 007E hex (126 decimal).
Holding register 40001 is addressed as register 0000 in the data address field of the message. The
function code field already specifies a ‘holding register’ operation. Therefore the ‘4XXXX’ reference
is implicit.”
To confirm the entry of a new value in E
2
prom, execute the MAKE - BACKUP function. If this function is
not run, by turning off the MCT 1302, the value before the change will be restored.
Unless otherwise specified, numerical values (such as addresses, codes and data) are expressed as
decimal values.
The MODBUS RTU protocol is only available on COM2 RS485.
HANDLING OF COMMUNICATION ERRORS
The communication strings are controlled by CRC (Cyclical Redundancy Check). In the event of a communication error, the slave does not respond with any string. The master must consider a timeout to receive
the response. If it does not get an answer, it must deduce that a communication error has occurred.
HANDLING OF DATA ERRORS RECEIVED
In the case of a string is received correctly but is not executable, the slave responds with an EXCEPTION
RESPONSE as shown in the following table.
CODEDESCRIPTION
1ILLEGAL FUNCTION (The function is not valid or is not supported)
2ILLEGAL DATA ADDRESS (The specified data address is not available)
3ILLEGAL DATA VALUE (Data received is invalid)
SUPPORTED FUNCTIONS:
FUNCTIONDESCRIPTION
01READ COIL STATUS (Reading of the logic output status)
02READ INPUT STATUS (Reading of the logic input status)
03READ HOLDING REGISTERS (Reading of programmable registers)
04READ INPUT REGISTERS (Reading of “read only” registers)
05FORCE SINGLE COIL (Writing of the single output status)
06PRESET SINGLE REGISTER (Writing of a programmable register)
15FORCE MULTIPLE COILS (Multiple writing of outputs)
16PRESET MULTIPLE REGISTERS (Multiple writing of registers)
Funct + 80h EXCEPTION RESPONSE
Page 65
LIST OF HOLDING REGISTERS OF THE MODBUS PROTOCOL
The parameters of the instrument that can be read or programmed through the communication interfaces
available on the instrument, according to the hardware setup, are listed in the following table.
The registers of the R type are those readable while those of the W type are those writable.
In the case of the Modbus TCP protocol, the instrument address (“Unit Identifier” field) must always be FFh.
If a fieldbus (different from Modbus) is used, only the R or R/W registers will be present in the input
area and only the W or R/W registers will be present in the output area.
The registers have a size of 16 bits.
AddressHolding RegisterR/W Notes
0001Register StatusRSee the relevant table
0002Gross weight (MSW)RValue INT. - Most significant word
0003Gross weight (LSW)RValue INT. - Less significant word
0004Net weight (MSW)RValue INT. - Most significant word
0005Net weight (LSW)RValue INT. - Less significant word
0006(Reserved)R
0007(Reserved)R
0008Digital inputs on board the MCT 1302RSee the relevant table.
0009Digital outputs on board the MCT 1302RSee the relevant table.
0010Cell signalRValue INT.
0011Tare (MSW)R/W Value INT. - Most significant word
0012Tare (LSW)R/W Value INT. - Less significant word
0101Zero control (MSW)R/W Value INT. - Most significant word
0102Zero control (LSW)R/W Value INT. - Less significant word
0103Unload set point (MSW)R/W Value INT. - Most significant word
0104Unload set point (LSW)R/W Value INT. - Less significant word
0105Unload prolonged timeR/W Value INT.
0106Unload end timeR/W Value INT.
0107End cycle timeR/W Value INT.
0108Time to enable the end cycle output R/W Value INT.
0109Dose end mode R/W Value INT.
0110Correction mode R/W Value INT.
0111Minimum level (MSW) R/W Value INT. - Most significant word
0112Minimum level (LSW) R/W Value INT. - Less significant word
0113Maximum level (MSW) R/W Value INT. - Most significant word
0114Maximum level (LSW) R/W Value INT. - Less significant word
0115Minimum level output R/W Value INT.
0116Maximum level output R/W Value INT.
0117Cycle output R/W Value INT.
0118Cycle end output R/W Value INT.
0119 Alarm output R/W Value INT.
0120 Timer mode R/W Value INT.
0151 Batching start tare R/W Value INT.
0152 Batching print R/W Value INT.
0201 Number of recipes in the archive R Value INT.
0202 Recipe name (character 01 + character 02) R/W Value INT. To manage the archive from serial item.
0203 Recipe name (character 03 + character 04) R/W Value INT. To manage the archive from serial item.
Page 66
0204 Recipe name (character 05 + character 06) R/W Value INT. To manage the archive from serial item.
0205 Recipe name (character 07 + character 08) R/W Value INT. To manage the archive from serial item.
0206 Recipe name (character 09 + character 10) R/W Value INT. To manage the archive from serial item.
0207 Recipe name (character 11 + character 12) R/W Value INT. To manage the archive from serial item.
0208 Recipe name (character 13 + character 14) R/W Value INT. To manage the archive from serial item.
0209 Number of recipe steps R/W Value INT. To manage the archive from serial item.
0210 Recipe/activity step R/W Value INT. To manage the archive from serial item.
0211
0212
0251 Number of activities in the archive R Value INT.
0252 Activity name (character 01 + character
0253 Activity name (character 03 + character
0254 Activity name (character 05 + character
0255 Activity name (character 07 + character
0256 Activity name (character 09 + character
0257 Activity name (character 11 + character
0258 Activity name (character 13 + character
0259 Type of activity R/W Value INT. To manage the archive from serial item.
0260
0261
0262
0263 Inflight (LSW) R/W
0264 Tolerance (MSW) R/W
0265 Tolerance (LSW) R/W
0266 Waiting time R/W Value INT. To manage the archive from serial item.
0267 Product timeout R/W Value INT. To manage the archive from serial item.
0268 Fast output R/W Value INT. To manage the archive from serial item.
0269 Fine output R/W Value INT. To manage the archive from serial item.
0270 Fast logic output R/W Value INT. To manage the archive from serial item.
0271 Fine logic output R/W Value INT. To manage the archive from serial item.
0272 Entrance R/W Value INT. To manage the archive from serial item.
0273 Spill time ON R/W Value INT. To manage the archive from serial item.
0274 Spill time OFF R/W Value INT. To manage the archive from serial item.
0275Inflight correctionR/W Value INT. To manage the archive from serial item.
0276Fine reactivationR/W Value INT. To manage the archive from serial item
Recipe step, batching setpoint (MSW) R/W
Recipe step, batching setpoint (LSW) R/W
02)
04)
06)
08)
10)
12)
14)
Fine (MSW) R/W
Fine (LSW) R/W
Inflight (MSW) R/W
Value INT. - Most significant word
To manage the archive from serial item.
Value INT. Less significant word
To manage the archive from serial item.
R/W Value INT. To manage the archive from serial item.
R/W Value INT. To manage the archive from serial item.
R/W Value INT. To manage the archive from serial item.
R/W Value INT. To manage the archive from serial item.
R/W Value INT. To manage the archive from serial item.
R/W Value INT. To manage the archive from serial item.
R/W Value INT. To manage the archive from serial item.
Value INT. - Most significant word
To manage the archive from serial item.
Value INT. Less significant word
To manage the archive from serial item.
Value INT. - Most significant word
To manage the archive from serial item.
Value INT. Less significant word
To manage the archive from serial item.
Value INT. - Most significant word
To manage the archive from serial item.
Value INT. Less significant word
To manage the archive from serial item.
Page 67
0301 Selected recipe total (MSW) R/W
0302 Selected recipe total (LSW) R/W
0303 Selected activity total (MSW) R/W
0304 Selected activity total (LSW) R/W
0401 Active recipe R/W Value INT.
0402 Active recipe correction factor (MSW) R/W Value INT. - Most significant word
0403 Active recipe correction factor (LSW) R/W Value INT. - Less significant word
0404 Batching cycles R/W Value INT.
0451 Batching operation R See the relevant table
0452 Batching status R See the relevant table
0453 Current cycle R Value INT.
0454 Current step R Value INT.
0455 Current activity R Value INT.
0456 Batching setpoint (MSW) R Value INT. - Most significant word
0457 Batching setpoint (LSW) R Value INT. - Less significant word
0458 Dosed net (last MSW step finished)R Value INT. - Most significant word
0459 Dosed net (last LSW step finished) R Value INT. - Less significant word
0460 Dosed net alarm (last step finished) R Value INT.
0461 Digital inputs of external modules R See the relevant table
0462 Digital outputs of external modules (MSW) R See the relevant table
0463 Digital outputs of external modules (LSW) R See the relevant table
0501Data Register (MSW)WValue INT. - Most significant word (See the relevant table)
0502Data Register (LSW)WValue INT. Less significant word (See the relevant table)
0503Command RegisterWSee the relevant table
1001 Standby function R/W Value INT.
1002 Password Setup function R/W Value INT.
1003 Language R/W Value INT.
1004 Key F1 R/W Value INT.
1005 Key F2 R/W Value INT.
1006 Key F3 R/W Value INT.
1007 Key F4 R/W Value INT.
1008 Cycle key block R/W Value INT.
1009 Weight key block R/W Value INT.
1010 Factor key block R/W Value INT.
1011 Recipe key block R/W Value INT.
1101Weight division value (*)R/W See the relevant table
Value INT. - Most significant word
See command/data register
Value INT. - Less significant word
See command/data register
Value INT. - Most significant word
See command/data register
Value INT. - Less significant word
See command/data register
1102Decimals (*)R/W See the relevant table
1103Load cell capacity (MSW) (*)R/W Value INT. Most significant word
1104Load cell capacity (LSW) (*)R/W Value INT. Less significant word
1105Load cell sensitivity (*)R/W Value INT.
1106Fixed tare (MSW) (*)R/W Value INT. Most significant word
1107Fixed tare (LSW) (*)R/W Value INT. Less significant word
1108Calibration severity (MSW) (*)R/W Value INT. Most significant word
Page 68
1109Calibration gravity (LSW) (*)R/W Value INT. Less significant word
1110Use gravity (MSW) (*)R/W Value INT. Most significant word
1111Use gravity (MSW) (*)R/W Value INT. Less significant word
1151Table cal. Zero signal (MSW) (*)R/W Value INT. - Most significant word
1152Table cal. Zero signal (LSW) (*)R/W Value INT. - Less significant word
1153Table cal. P1 signal (MSW) (*)R/W Value INT. - Most significant word
1154Table cal. P1 signal (LSW) (*)R/W Value INT. - Less significant word
1155Table cal. P2 signal (MSW) (*)R/W Value INT. - Most significant word
1156Table cal. P2 signal (LSW) (*)R/W Value INT. - Less significant word
1157Table cal. P3 signal (MSW) (*)R/W Value INT. - Most significant word
1158Table cal. P3 signal (LSW) (*)R/W Value INT. - Less significant word
1159Table cal. P4 signal (MSW) (*)R/W Value INT. - Most significant word
1160Table cal. P4 signal (LSW) (*)R/W Value INT. - Less significant word
1161Table cal. P5 signal (MSW) (*)R/W Value INT. - Most significant word
1162Table cal. P5 signal (LSW) (*)R/W Value INT. - Less significant word
1163Table cal. Value P1 (MSW) (*)R/W Value INT. - Most significant word
1164Table cal. Value P1 (LSW) (*)R/W Value INT. - Less significant word
1165Table cal. Value P2 (MSW) (*)R/W Value INT. - Most significant word
1166Table cal. Value P2 (LSW) (*)R/W Value INT. - Less significant word
1167Table cal. Value P3 (MSW) (*)R/W Value INT. - Most significant word
1168Table cal. Value P3 (LSW) (*)R/W Value INT. - Less significant word
1169Table cal. Value P4 (MSW) (*)R/W Value INT. - Most significant word
1170Table cal. Value P4 (LSW) (*)R/W Value INT. - Less significant word
1171Table cal. Value P5 (MSW) (*)R/W Value INT. - Most significant word
1172Table cal. Value P5 (LSW) (*)R/W Value INT. - Less significant word
1201Filter factorR/W See match on page 46
1202Output rate ADCR/W See match on page 46
1203Average reading numberR/W Value INT.
1204Monotony timeR/W Value INT.
1205Oscillation time R/W Value INT.
1206Oscillation rangeR/W Value INT.
1301Full scale (MSW) (*)R/W Value INT. - Most significant word
1302Full scale (LSW) (*)R/W Value INT. - Less significant word
1303Weight stability (*)R/W See match on page 46
1304Autozero at power up (MSW) (*)R/W Value INT. - Most significant word
1305Autozero at power up (LSW) (*)R/W Value INT. - Less significant word
1306Tracking of zero (*)R/W See match on page 47
1307Resettable divisions (MSW) (*)R/W Value INT. - Most significant word
1308Resettable divisions (LSW) (*)R/W Value INT. - Less significant word
1401Function of input 1R/W See Input/Output menu match
1402Function of input 2R/W See Input/Output menu match
1403Function of input 3R/W See Input/Output menu match
1404Function of input 4R/W See Input/Output menu match
1405Function of input 5R/W See Input/Output menu match
1406Function of input 6R/W See Input/Output menu match
1501Analog tare (MSW)R/W Value INT. - Most significant word
Page 69
1502Analog tare (LSW)R/W Value INT. - Less significant word
1503Analog full scale (MSW)R/W Value INT. - Most significant word
1504Analog full scale (LSW)R/W Value INT. - Less significant word
1505Analog output modeR/W See match on page 35
1506Analog output rangeR/W See match on page 35
Value INT. Points of zero of the analog output; to end
1507Analog zero adjustmentR/W
1508Analog full scale adjustmentR/W
2000Monitor registerW
2100Monitor registerR
the adjustment procedure it is necessary to send the
command to save data in permanent memory in the
Command Register.
Value INT. Points of full scale of the analog output; to
end the adjustment procedure it is necessary to send
the command to save data in permanent memory in the
Command Register.
The programmed value is automatically copied to the
Monitor Register (2100).
TABLE A - CODING THE STATUS REGISTER
BITS15141312111098
DescriptionSetup (***)
Weight
delta
000
0= Net displayed
1= Gross displayed
Run backup0
BITS76543210
Description
Not
calibrated
Weight
error
Overload
Under-
load
Tare
entered
Zero
band
Stable
weight
TABLE B - CODING THE INTERNAL DIGITAL INPUTS/OUTPUTS OF MCT 1302
BITS15÷6543210
DescriptionReserved
In 6 - Out 6 ONIn 5 - Out 5 ONIn 4 - Out 4 ONIn 3 - Out 3 ONIn 2 - Out 2 ONIn 1 - Out 1
WARNING: bits 15 to 4 are not managed and always are equal to 0.
Zero
center
ON
Page 70
TABLE C - CODING THE DIGITAL INPUTS/OUTPUTS OF EXTERNAL MODULES
TABELE E - CODING THE COMMAND REGISTER / DATA REGISTER
REGISTER
VALUE
0x0001 Semi-automatic zero
0x0002 Automatic tare
0x0004 Zero calibration (**)
0x0005 Full scale calibration (**)
0x0006 Analog test
0x0007 Saving data in permanent memory
0x000B
0x000C
0x000D
0x000E Tare cancellation command.
COMMAND REGISTER FUNCTION DATA REGISTER FUNCTIONNOTE
Sample weight value in
MSW and LSW
Value between 0 and 100 in
intervals of 10 in LSW
Switching command from gross to
net weight
Switching command from net to
gross weight
Acquisition of zero signal (table
calibration)
Page 71
0x000FPrint command.
The general recipe parameters (name
0x0010 Reading the recipe parameters.
0x0011 Saving the recipe parameters.
0x0012 Reading the recipe step.Recipe step in LSB
0x0013 Saving the recipe step. Recipe step in LSB
0x0014 Reading the activity parameters. Activity index in LSB See registers from 0252 to 0274.
0x0015 Saving the activity parameters. Activity index in LSB See registers from 0252 to 0274.
0x0016 Reading the recipe total.
0x0017 Reading the activity total. Activity index in LSB
0x0018 Delete the recipe total.
0x0019 Delete the activity total. Activity index in LSB
0x001A Delete recipe
0x001B Delete activity
0x001C Reset the batching alarm - -
Recipe index in MSB and
LSB
Recipe index in MSB and
LSB
Recipe index in MSB and
LSB
Recipe index in MSB and
LSB
Recipe index in MSB and
LSB
Recipe index in MSB and
LSB
and number of steps) and the parameters
related to step 1 of the recipe are read. See
registers from 0202 to 0212.
Registers 0202 to 0209 are programmed
in the recipe specified in the Data Register.
If the change also concerns the recipe
steps, the recipe steps must first be saved by
command
The "activity” registers (0210) and the
"batching setpoint" registers (0211-0212)
are updated with the values of the step
specified in the Data Register.
The step specified in the Data Register is
programmed with the values written in the
"activity” (0210) and "batching setpoint"
registers(0211-0212). To save the changes
in permanent memory, it is necessary to send
the command 0x0011, which can be sent
only once at the end of the recipe changes.
The total of the recipe specified in the Data
Register is read. See registers from 0301 to
0302.
The total of the activity specified in the Data
Register is read. See registers from 0303 to
0304.
The total of the recipe specified in the Data
Register is deleted. Write 0xFFFF in the Data
Register to delete the totals of all the recipes.
The total of the activity specified in the Data
Register is deleted. Write 0xFFFF in the
Data Register to delete the totals of all the
activities.
The recipe specified in the Data Register is
deleted.
The recipe specified in the Data Register is
deleted.
0x001D Start batching. - -
0x001E Pause batching. - -
0x001F Restart batching. - -
0x0020 Stop batching. - -
0x3FFFEnabling the Output Data Area (*)
(*) The instrument parameters managed in the Fieldbus Output Data Area are not changed until this
command is sent. When the instrument is switched on, the Output Data Area is completely zeroed,
so the Fieldbus master must read the parameter values from the Input Data Area and copy them to
the relevant Output Data Area registers, and then it must send the enable command to the Command
Register. Otherwise, all the parameters managed in the Output Data Area would be reset at power
up.
(***) The instrument is in the setup phase (TRUE flag during access to the SETUP menu of the
instrument or during connection with PC software “Optimation”).
Page 72
TABLE F - CODING FOR BATCHING OPERATION
ValueDescriprion
0x0000Batch in stop
0x0001Batch in pause
0x0002Batch in progress
0x0003Batch Alarm
TABLE G - CODING FOR BATCHING STATUS
ValueDescriprion
0x0001Coarse
0x0002Fine
0x0003Manual
0x0004Wait
0x0005Wait for acknowldgee
0x0006Input control
0x0007Restart
0x0008Dump
0x0009Error no selected formula
0x000AError, empty formula
When the system is empty and stable, write the hexadecimal value 0004 in the Command Register
(0503).
To save the new value of Zero permanently in the memory, write the hexadecimal value 0007 in the
Command Register.
FULL SCALE CALIBRATION
Load the sample weight on the scale, example 1256 Kg.
Write the hexadecimal value of the sample weight 04E8 in the Data Register (0501 and 0502).
Write the hexadecimal value 0005 in the Command Register (0503).
The Command Register and the Data Register can be written simultaneously using the multiple register
function.
To save the new full scale value permanently in the memory, write the hexadecimal value 0007 in
the Command Register.
Page 73
TIMES OF RESPONSE OF THE INSTRUMENT
=
+ ∗
+ ,
=
+ ∗
+ ,
The instrument, to respond most queries, takes a variable time depending on the programming of the
instrument parameters and the type of request as shown in the table:
Frequency of the measure update 12.5 Hz50 Hz100 Hz250 Hz1000 Hz
Frequency of the analog output update.12.5 Hz50 Hz100 Hz250 Hz1000 Hz
Frequency of update and reading of the
fieldbus memory area.
250 Hz250 Hz250 Hz250 Hz250 Hz
Frequency of the logic output update.12.5 Hz50 Hz100 Hz250 Hz1000 Hz
Weight and status acquisition cycle from
Modbus RTU (*)
Weight and status acquisition cycle from
ASCII protocol (*)
(*) nchar = sum of the characters that make up the request string of the master (Query) and the response
string of the MCT 1302 (Response).
Example of status query, net weight and gross weight (5 registers) in baud rate = 115.2 kbit/sec:
• the writing of the registers Cell capacity, Cell sensitivity, Weighing net, System tare, Filter (Max
time = 550mSec).
USE OF OPTIMATION VIA THE USB PORT
The PC OPTIMATION software supplied with the instrument allows:
• the total setup of all the setup parameters;I
• the testing of the different hardware sections;
• the consultation of the instrument documentation;
• the firmware update of the instrument;
• the storage over time of the weight values acquired
by the instrument by means of Data-logger function;
• saving and loading of the setup parameters of the
instrument on file.
All the operating procedures are shown in the dedicated
manual.
Page 74
FIELDBUS PROTOCOL
INPUT DATA AREA
The following table lists the input area registers (prepared by the instrument and read from the master),
common to all PROFINET, ETHERCAT, ETHERNET/IP fieldbuses. The registers have a size of 16 bits.
The input area is updated to a maximum frequency of 90 Hz (60 Hz for PROFIBUS fieldbus). The size
of the Input area set in the fieldbus master must coincide with the dimension set in the instrument.
Byte
1-20Register StatusSee the relevant table
3-41Gross weight (MSW)Value INT. - Most significant word
5-62Gross weight (LSW)Value INT. - Less significant word
7-83Net weight (MSW)Value INT. - Most significant word
9-104Net weight (LSW)Value INT. - Less significant word
11-125
13-146
15-167
17-188
19-2010Cell signalValue INT.
21-2211Tare (MSW)Value INT. - Most significant word
23-2412Tare (LSW)Value INT. - Less significant word
25-2613Batching operationSee the relevant table
27-2814Batching statusSee the relevant table
29-3015Current cycleValue INT.
31-3216Current stepValue INT.
33-3417Current activityValue INT.
35-3618Current batching setpoint (MSB) Value INT. - Most significant word
37-3819Current batching setpoint (LSB)Value INT. - Less significant word
39-4020
41-4221
43-4422
45-4623Monitor Register
47-4824Active recipeValue INT.
49-5025
51-5226
53-5427Batching cyclesValue INT. - Less significant word
55-5628Selected recipe total (MSW)Value INT. - Most significant word See command/data register
57-5829Selected recipe total (LSW)Value INT. - Less significant word See command/data register
Register
address
9Digital outputs of external
INPUT AREA REGISTERNotes
Digital inputs on board the
MCT
Digital outputs on board the
MCT
Digital inputs of external
modules
Digital outputs of external
modules (MSB)
modules (LSB)
Dosed net (last MSW step
finished)
Dosed net (last LSW step
finished)
Dosed net alarm (last step
finished)
Active recipe correction factor
(MSW)
Active recipe correction factor
(LSW)
See the relevant table.
See the relevant table.
See the relevant table.
Value INT. - Most significant word
See the relevant table
. Value INT. - Most significant word
See the relevant table
Value INT. - Most significant word
Value INT. - Less significant word
Value INT.
This value corresponds to the equivalent register in the output
area.
Value INT. - Most significant word
Value INT. - Less significant word
Page 75
59-6030Selected activity total (MSW)Value INT. - Most significant word See command/data register
61-6231Selected activity total (LSW)Value INT. - Less significant word See command/data register
63-6432
65-6633
67-6834
69-7035
71-7236
73-7437
75-7638
77-7839
79-8040
81-8241Number of receipt stepsINT value for archive management from fieldbus
83-8442Recipe/activity stepsINT value for archive management from fieldbus
85-8643
87-8844
89-9045Type of activityINT value for archive management from fieldbus
91-9246Fine (MSW)
93-9447Fine (LSW)
95-9648Inflight (MSW)
97-9849Inflight (LSW)
99-10050Tolerance (MSW)
101-10251Tolerance (LSW)
103-10452Waiting timeINT value for archive management from fieldbus
105-10653Product timeoutINT value for archive management from fieldbus
107-10854Fast outputINT value for archive management from fieldbus
109-11055Fine outputINT value for archive management from fieldbus
111-11256Fast logic outputINT value for archive management from fieldbus
113-11457Fine logic outputINT value for archive management from fieldbus
115-11658EntranceINT value for archive management from fieldbus
117-11859Spill time ONINT value for archive management from fieldbus
119-12060Spill time OFFINT value for archive management from fieldbus
121-12261Inflight correctionINT value for archive management from fieldbus
123-12462Fine reactivationINT value for archive management from fieldbus
Number of recipes in the
archive
Number of activities in the
archive
Recipe/activity name
(character01 + character02)
Recipe/activity name
(character03 + character04)
Recipe/activity name
(character05 + character06)
Recipe/activity name
(character07 + character08)
Recipe/activity name
(character09 + character10)
Recipe/activity name
(character11 + character12)
Recipe/activity name
(character13 + character14)
Recipe step, batching setpoint
(MSW)
Recipe step, batching setpoint
(LSW)
Value INT.
Value INT.
INT value for archive management from fieldbus
INT value for archive management from fieldbus
INT value for archive management from fieldbus
INT value for archive management from fieldbus
INT value for archive management from fieldbus
INT value for archive management from fieldbus
INT value for archive management from fieldbus
- Most significant INT word for archive management from
fieldbus
- Most significant INT word for archive management from
fieldbus
- Most significant INT word for archive management from
fieldbus
- Most significant INT word for archive management from
fieldbus
- Most significant INT word for archive management from
fieldbus
- Most significant INT word for archive management from
fieldbus
Value INT. Most significant word for archive management from
fieldbus
Value INT. Most significant word for archive management from
fieldbus
Page 76
READING EXAMPLE
To read the gross weight from the MCT 1302, you must the address 3 to 6 of the Input Area.
To read the net weight, instead, you must read the bytes from 7 to 10 of the Input Area.
If the instrument shows on the display the gross weight value of 12351 in the relative bytes, you shall
read:
Byte 3Byte 4Byte 5Byte 6
Hex0000303F
Page 77
OUTPUT DATA AREA
The following table lists the registers of the output area (written by the master and acquired by the
instrument), common to all PROFINET, ETHERCAT, ETHERNET / IP fieldbuses.
The registers have a size of 16 bits. The registers written by the master in the output area are read by
the instrument at a maximum frequency of 90 Hz (60 Hz in the case of PROFIBUS fieldbus).
The size of the Output area set in the fieldbus master must coincide with the dimension set in the instrument.
Byte
1-20Command RegisterSee the relevant table
3-41Data Register (MSW)Value INT. - Most significant word (See Table)
5-62Data Register (LSW)Value INT. - Less significant word (See Table)
7-83Monitor registerValue corresponds to the equivalent register in the input area.
9-104Active recipeINT value for archive management from fieldbus
11-125
13-146
15-167Batching cyclesINT value for archive management from fieldbus
17-188
19-209
21-2210
23-2411
25-2612
27-2813
29-3014
31-3215Number of receipt stepsINT value for archive management from fieldbus
33-3416Recipe/activity stepsINT value for archive management from fieldbus
35-3617
37-3818
39-4019Type of activityINT value for archive management from fieldbus
41-4220Fine (MSW)
43-4421Fine (LSW)
45-4622Inflight (MSW)
47-4823Inflight (LSW)Value INT. - Less significant word
49-5024Tolerance (MSW)Value INT. - Most significant word
51-5225Tolerance (LSW)Value INT. - Less significant word
53-5426Waiting timeINT value for archive management from fieldbus
55-5627Product timeoutINT value for archive management from fieldbus
57-5828Fast outputINT value for archive management from fieldbus
Register
address
OUTPUT AREA REGISTERNotes
Active recipe correction factor
(MSW)
Active recipe correction factor
(LSW)
Recipe/activity name
(character01 + character02)
Recipe/activity name
(character03 + character04)
Recipe/activity name
(character05 + character06)
Recipe/activity name
(character07 + character08)
Recipe/activity name
(character09 + character10)
Recipe/activity name
(character11 + character12)
Recipe/activity name
(character13 + character14)
Recipe step, batching setpoint
(MSW)
Recipe step, batching setpoint
(LSW)
Value INT. - Most significant word
Value INT. - Less significant word
INT value for archive management from fieldbus
INT value for archive management from fieldbus
INT value for archive management from fieldbus
INT value for archive management from fieldbus
INT value for archive management from fieldbus
INT value for archive management from fieldbus
INT value for archive management from fieldbus
Value INT. - Most significant word for archive management from
fieldbus
Value INT. - Most significant word for archive management from
fieldbus
Value INT. - Most significant word for archive management from
fieldbus
Value INT. - Most significant word for archive management from
fieldbus
Value INT. - Most significant word for archive management from
fieldbus
Page 78
59-6029Fine outputINT value for archive management from fieldbus
61-6230Fast logic outputINT value for archive management from fieldbus
63-6431Fine logic outputINT value for archive management from fieldbus
65-6632EntranceINT value for archive management from fieldbus
67-6833Spill time ONINT value for archive management from fieldbus
69-7034Spill time OFFINT value for archive management from fieldbus
71-7235InInflight correctionINT value for archive management from fieldbus
73-7436Fine reactivationINT value for archive management from fieldbus
75-7637
77-7838
Number of recipes in the
archive
Number of activities in the
archive
INT value for archive management from fieldbus
INT value for archive management from fieldbus
WRITING EXAMPLES
To write the Set-up parameters, follow the example below:
In bytes 1-2 (Command Register) write the value HEX 3FFF that opens the internal writing area of the
MCT 1302.
Example: You want to change the values of Set-Point 1 and Set-Point 2 to 120 and 9740 respectively.
Set-Point 1Byte 21Byte 22
Hex0078
Set-Point 2Byte 23Byte 24
Hex260C
Save the data by writing the HEX value 7 in the Byte 1-2
N.B. The instrument does not accept writing a value equal to the value already present.
Page 79
PROBLEMPOSSIBLE CAUSESOLUTION
The display shows the
message O-L
The display shows
the underscore on the
upper display.
The display shows
the underscore on the
lower display.
The number of
decimals is incorrect.
Serial communication
does not work
correctly.
The semiautomatic
zero function does not
work.
The semi-automatic
tare function does not
work.
TROUBLESHOOTING
The weight acquired is not detectable
because the load cell is absent or
incorrectly connected.
The weight acquired cannot be
represented because it exceeds the
five figures available or is greater
than the capacity of the load cells.
The weight acquired cannot be
represented because negative over
-9999.
The correct division value was not
selected.
The installation was not performed
correctly.
The selection of the serial interface
operation is incorrect.
The gross weight exceeds the action
limit of semiautomatic zero.
The weight does not stabilize.
The gross weight is negative or
exceeds the maximum capacity
value.
The weight does not stabilize.
Check the load cell connections.
Select the correct division value in
the main menu.
Check the connections as described
in the installation manual.
Select the settings appropriately.
To restore zero, the weight must be
calibrated.
Wait for the weight to stabilize or
adjust the weight filter parameter.
Check the gross weight.
Wait for the weight to stabilize or
adjust the weight filter parameter.
Page 80
EU DECLARATION OF CONFORMITY (DOC)
The Company
Pavone Sistemi S.r.l.
Via Tiberio Bianchi, 11/13/15
20863 Concorezzo, MB, Italy
declares that the DoC was released under its own responsibility, and belong to the following product:
Model of tool / Product: MCT 1302
Type: Weighing Instrument
The object of the above statement used as shown in the installation and use manual complies with the
relevant European Union harmonization standards:
Directive 2014/30/UE on electromagnetic compatibility
The below harmonized standards and technical specifications have been applied:
EN 61000-6-2: 2005
EN 61000-6-3: 2007 + A1 2011
Directive 2014/35/UE Low Voltage
The below harmonized standards and technical specifications have been applied:
EN 61010-1: 2011
Directive 2014/31/UE Measurement instruments
The below harmonized standards and technical specifications have been applied:
EN 45501: 2015
Signed in the name and on behalf of:
Concorezzo: 16/01/2017
Di Reda Donato - Head
Page 81
PAVONE SISTEMI S.R.L.
Via Tiberio Bianchi, 11/13/15, 20863 Concorezzo (MB)
T 039 9162656 F 039 9162675 W en.pavonesistemi.it
Industrial Electronic Weighing Systems since 1963
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.