Pasport SN-7927A Instruction Sheet

®
Geiger-Müller Counter
Protective
Plastic Cap
Power
Indicator
(LED)
SN-7927A
Instruction Sheet
012-07905C
Introduction
The SN-7927A Geiger-Müller Counter is a sensor with a Tip-Ring-Sleeve type plug that can be connected to a PASCO Interface. The Geiger-Müller Counter can sense alpha, beta, and gamma radiation when used with PASCO Data Collection Software. (NOTE: If your Interface has only PASPORT input ports, use a Digital Adapter (PS-2159) to connect the sensor to the interface.) Features include the following:
Versatile positioning options, either hand-held or mounted on a rod stand.
Audible feedback (the device beeps when it registers a count).
Equipment Included
Item Number
Geiger-Müller Counter SN-7927A
Required
Item*
PASCO Interface
PASCO Data Collection Software
(*See the PASCO web site at www.pasco.com for information about PASCO Interfaces and Data Collection Software.)
Recommended
Item* Number
Radioactive Sources (set of 3) SN-8110
Sensor Mounting Rod ME-8965
(*See the PASCO web site at www.pasco.com for information about radioactive sources, rods and bases, and clamps.)
Equipment
The Geiger-Müller Counter is equipped with a built-in Gei­ger-Müller tube and integral power supply. The light emitting diode (LED) Power Indicator lights when an operating voltage is supplied to the sensor. The unit has a buzzer that emits an audible signal for each pulse that is recorded. The sensor includes a plastic cap that provides protection for the delicate mica membrane. (The cap has a small venting hole to prevent the creation of a vacuum or high pressure when the cap is removed or replaced.)
®
Geiger-Müller Counter
Sensor
Plug
Sensor
DIGITAL
Port
Digital Adapter
PASPORT Interface
CAUTION
The mica window is extremely delicate. DO NOT cover the small venting hole of the plastic cap when you are removing or replacing the cap. Leave the cap in place except when alpha radiation is measured. Never touch the mica window!
Setup
1. Connect the sensor plug to a DIGITAL input port on the PASCO Interface.
2. Connect the interface to a computer and turn on the inter­face.
Eating, drinking, and the application of cosmetics in the laboratory are not permitted.
Pipetting by mouth is never permitted. (Use pipette fill­ers.)
Wear gloves and lab coats when working with all liquid radioisotopes.
Wash your hands thoroughly before leaving the lab.
Pour radioactive liquid waste into an approved liquid waste container - NEVER into a sink.
Report all spills, wounds, or other emergencies to the instructor.
Maintain good housekeeping at all times.
Store radioactive materials only in the designated storage area, and do not remove sources from the lab.
Specifications
3. Start the Data Collection Software. Setup a display.
For information on selecting and configuring the sensor, refer to the Online Help System and User’s Guide for the data col­lection software.
NOTE: Use a Digital Adapter (PS-2159) to connect the sensor to a PASCO interface that does not have a DIGI­TAL input port. See the illustration.
Connect the sensor to port 1 on the Digital Adapter.
Connect the Digital Adapter to a PASPORT input port on the interface.
NOTE: The Geiger-Müller counter displays the number of counts in a given period of time. To change the period of time, adjust the sampling rate.
Nuclear Safety
Most of the radioactive sources available to educators are very low level isotopes referred to as “license free” sources. How­ever, observe the following guidelines:
Window material: Mica
Window diameter: 9.14 millimeters (mm)
Window density: 1.52 milligrams per square centimeter
Count rate (max.): 11,111 per sec (for a minimum dead time of
90 microseconds)
Technical Support
Before calling PASCO Technical Support, please have the appa­ratus, this user's guide, and the following information available:
Product name, model number, and serial number.
Approximate age of the apparatus.
Detailed description of the problem or sequence of events required to duplicate the problem.
For assistance with any PASCO product, contact PASCO at:
Address: PASCO scientific
10101 Foothills Blvd. Roseville, CA 95747-7100
Phone: +1 916-462-8384 (worldwide)
877-373-0300 (U.S.)
Web: www.pasco.com
Email: support@pasco.com
2
012-07905C
®
Model No. SN-7927A
Limited Warranty
For a description of the product warranty, see the PASCO catalog.
Copyright
The PASCO scientific 012-07905C Geiger-Müller Counter Instruction Sheet is copyrighted with all rights reserved. Permission is granted to
non-profit educational institutions for reproduction of any part of this man­ual, providing the reproductions are used only in their laboratories and classrooms, and are not sold for profit. Reproduction under any other cir­cumstances, without the written consent of PASCO scientific, is prohibited.
Trademarks
PASCO, PASCO scientific, and PASPORT are trademarks or registered trademarks of PASCO scientific, in the United States and/or in other coun­tries. All other brands, products, or service names are or may be trade­marks or service marks of, and are used to identify, products or services of, their respective owners. For more information visit www.pasco.com/legal.
Product End of Life Disposal Instructions:
This electronic product is subject to disposal and recycling regu­lations that vary by country and region. It is your responsibility to recycle your electronic equipment per your local environmental laws and regulations to ensure that it will be recycled in a manner that protects human health and the environment. To find out where you can drop off your waste equipment for recycling, please contact your local waste recycle/disposal service, or the place where you purchased the product.
The European Union WEEE (Waste Electronic and Electrical Equipment) symbol (to the right) and on the product or its packaging indicates that this product must not be disposed of in a standard waste container.