High Current Sensor
PS-2193
Introduction
The PS-2193 High Current Sensor measures current between
10 A and +10 A. The sensor connects to a PASPORT interface and collects data at up to 1000 samples/s.
Connecting to a Circuit
Use cables with 4 mm banana plugs to place the sensor in a cir­cuit as illustrated. Positive current flows into the positive termi­nal and out of the negative terminal.
Instruction Sheet
012-10102A
Included Equipment Part Number
High Current Sensor PS-2193
Additional Equipment Required
PASPORT Interface See www.pasco.com
or PASCO catalog
Quick Start
1. Connect the High Current Sensor to you PASPORT inter-
face.
to it and start DataStudio.
3. Connect the sensor to a circuit.
4. Press or click the Star t button to begin recording data.
OVERCURRENT
MAXIMUM
CIRCUIT VOLTAGE
10V
CURRENT
CONNECT
+-
IN SERIES
Positive current
10 AMP MAX
CURRENT
HIGH
SENSOR
PS-2193
ä
The voltage anywhere within the connected circuit should not exceed 10 V above or below earth ground.
Overcurrent
If the current through the sensor exceeds ±10 A, the overcurrent indicator on the sensor will light up. Reduce the current or dis­connect the sensor to prevent damage.
Over-sampling
The High Current Sensor uses dynamic variable over-sampling to reduce noise and produce smoother data. The degree of over-sampling that takes place within the sensor depends on the sample rate. To maximize over-sampling, set the sample rate as low as possible for a given application. Maximum over-sampling occurs at sample rates of 10 Hz or slower.
®
High Current Sensor PS-2193
Specifications Technical Support
Measurable current range 10 A to +10 A
Measurement resolution 16 bits (approximately 300 µA)
Maximum circuit voltage 10 V above or below earth ground
Measurement accuracy 0.5% or 2.5 mA, whichever is greater
Sensing element series resistance
Maximum sample rate 1000 samples/s
Maximum continuous current without damage
Maximum continuous circuit overvoltage without damage
Term inal s 4 mm banana jacks
0.01
12 A
40 V above or below earth ground
For assistance with any PASCO product, contact PASCO at:
Address: PASCO scientific
10101 Foothills Blvd. Roseville, CA 95747-7100
Phone: 916-786-3800 (worldwide)
800-772-8700 (U.S.)
Fax: (916) 786-7565
Web: www.pasco.com
Email: support@pasco.com
Limited Warranty For a description of the product warranty, see the PASCO catalog. Copyright The PASCO scientific 012-10102A
Instruction Sheet
granted to non-profit educational institutions for reproduction of any part of this manual, providing the reproductions are used only in their laboratories and classrooms, and are not sold for profit. Reproduction under any other circumstances, without the written consent of PASCO scientific, is prohib­ited. Trademarks PASCO, PASCO scientific, PASPORT, and DataStudio are trademarks or registered trademarks of PASCO scientific, in the United States and/or in other countries. All other brands, products, or service names are or may be trademarks or service marks of, and are used to identify, products or services of, their respective owners. For more infor­mation visit www.pasco.com/legal.