PASCO PS-2179 User Manual

®
Water Quality Colorimeter
Included Equipment Part Number
Water Quality Colorimeter PS-2179
Additional Required Equipment
ezSample Snap Vial Test Kit Various, see
www.pasco.com or PASCO catalog
Xplorer GLX PS-2002
or
other PASPORT interface
PASport Extension Cable 514-07281
Recommended Equipment
Water Quality Field Guide PS-2829
PASPORT Extension Cable PS-2500
Water Quality Colorimeter
PS-2179
PS-2179
Instruction Sheet
012-09873C
Introduction
The PASPORT Water Quality Colorimeter is designed for chemi­cal analysis of water samples using PASCO's ezSample Snap Vial water quality test kits.
ezSample Snap Vials contain a reagent (specific to a particular ion) in a vacuum-sealed ampoule. When the scored tip of the ampoule is snapped off in the water sample, the sample is drawn into the ampoule and mixes with the color-forming reagent. The absorbance of the resulting solution depends on the ion concen­tration.
The colorimeter accurately measures the absorbance by project­ing light through the ampoule and measuring the intensity of transmitted light. It uses Beer's Law and built-in calibration curves to calculate the ion concentration.
The colorimeter’s measurements are sent to your PASPORT interface for display, recording, and analysis.
®
Water Quality Colorimeter PS-2179
Measurement Procedure
With the Xplorer GLX (Standalone Mode)
Follow these steps to use the colorimeter connected to a GLX without a computer. The colorimeter should be horizontal when it makes a measurement. If you connect it directly to the GLX with­out a cable, keep the GLX horizontal.
1. Connect the colorimeter to the GLX (may require Extension Cable 514-07281). The GLX detects the colorimeter and opens the Digits screen.
2. Press to highlight the data source field.
3. Press again to
open the data source menu (right).
4. Use the arrow keys to highlight the desired measurement and press . (Select the measurement corre­sponding to the test kit that you will use.)
5. (Optional) Perform the calibration procedure (see “Calibra­tion” on page 3).
6. Follow the instructions provided with your ezSample Snap Vial kit to mix your water sample with the reagent in the ampoule.
7. Handling the ampoule by the tip, wipe the glass clean using a non-abrasive cleaning tissue. Avoid touching the glass after it is clean.
8. Align the mark on the ampoule as illustrated and insert the ampoule into the colorimeter.
9. Place the cap over the ampoule.
With a Computer
Follow these steps to use the colorimeter with any PASPORT interface connected to a computer.
1. Connect your PASPORT interface to the computer and start DataStudio.
2. Connect the colorimeter to the PASPORT interface. DataStudio detects the colorimeter and creates a digits dis­play showing all available measurement.
3. (Optional) Perform the calibration procedure (see “Calibra­tion” on page 3).
4. Follow the instructions provided with your ezSample Snap Vial kit to mix your water sample with the reagent in the ampoule.
5. Handling the ampoule by the tip, wipe the glass clean using a non-abrasive cleaning tissue. Avoid touching the glass after it is clean.
6. Align the mark on the ampoule as illustrated and insert the ampoule into the colorimeter.
7. Place the cap over the ampoule.
8. Click Start to start recording and displaying data. (Or open
the Experiment menu and select Monitor to display mea­surements without recording them.)
DataStudio displays all available measurements. Read the one corresponding to the test kit that you used. (The other live mea­surements are not meaningful and should be ignored.)
To remove unneeded measurements from the digits display, click each measurement and press delete.
The GLX automatically monitors and displays live measurements in the Digits screen.
10. (Optional) To start recording data, press .
To remove the measurement from the lower segment of the Digits screen: Press . Press the down arrow to high­light the data source field of the lower segment. Press again to open the menu. Use the arrow keys to highlight Clear Input and press
2
®
Water Quality Colorimeter PS-2179
For Best Results
Place the cap fully over the ampoule. A blinking red light on the calibration button indicates that the colorimeter is not ready to make a valid measurement: either stray light is entering the colorimeter or the measurement is out of range. Ensure that the cap is properly in place to make the light stop blinking.
Keep the colorimeter horizontal.
The ambient temperature and the temperature of the test solution should be about equal and between 5° C and 40° C. A cold ampoule may develop condensation, which would affect the reading.
Ensure that the outside of the ampoule is clean and dry.
Do not test turbid water samples.
Shield the colorimeter from direct sunlight.
Calibration
Specifications
Each type of ezSample test kit has its own accuracy and range specifications. For that information, refer to the instruction sheet included with the kit.
These specifications apply to the colorimeter itself, independent of the reagent used.
Range 0 to 100% transmittance
Wavelengths 660 nm (red), 610 nm (orange), 565 nm
(green), 468 nm (blue)
Accuracy ±0.5% transmittance
Resolution 0.1% transmittance
Default sample rate 1 sample/s
Maximum sample rate 5 samples/s
Temperature range 5° C to 40° C
Technical Support
Calibration is not required but is recommended for better accu­racy. Calibration information is stored in the colorimeter; when you unplug the sensor and reconnect it, the sensor retains the last calibration.
To calibrate the colorimeter you will need a calibration ampoule, which is included in each ezSample Snap Vial kit.
1. Handling the calibration ampoule by the tip, wipe the glass clean using a non-abrasive cleaning tissue. Avoid touching the glass after it is clean.
2. Align the mark on the calibration ampoule as illustrated on page 2 and insert the ampoule into the colorimeter
3. Place the cap over the ampoule.
4. Press the green button on the sensor. The button illuminates
(in green) to indicate calibration is in progress.
5. Wait for the light to turn off.
6. Remove the calibration ampoule.
If the red light on the button blinks, ensure that the cap is properly in place and repeat the calibration procedure.
For assistance with any PASCO product, contact PASCO at:
Address: PASCO scientific
10101 Foothills Blvd. Roseville, CA 95747-7100
Phone: 916-786-3800 (worldwide)
800-772-8700 (U.S.)
Fax: (916) 786-7565
Web: www.pasco.com
Email: support@pasco.com
Limited Warranty For a description of the product warranty, see the PASCO catalog.
Copyright The PASCO scientific 012-09873C Water Quality Colorimeter Instruction Sheet is copyrighted with all rights reserved. Permission is
granted to non-profit educational institutions for reproduction of any part of this manual, providing the reproductions are used only in their laboratories and classrooms, and are not sold for profit. Reproduction under any other circumstances, without the written consent of PASCO scientific, is prohib­ited. Trademarks PASCO, PASCO scientific, DataStudio, ezSample, PASPORT, and Xplorer GLX are trademarks or registered trademarks of PASCO scientific, in the United States and/or in other countries. All other brands, products, or service names are or may be trademarks or service marks of, and are used to identify, products or services of, their respective owners. For more information visit www.pasco.com/legal.
The interface cannot collect data from the colorimeter during calibration. Wait until the green light turns off before recording or monitoring data.
After calibration, the colorimeter should register the concentration of the calibration ampoule as close to zero (within the specified accuracy of the test kit).
3
Loading...