The DO2 Probe comes with a bottle of filling solution, a syringe, a replacement membrane cartridge
and O-ring, and a plunger. See illustration on page 7.
2
PASPORT interfaces include the AirLink (IW-4100 or PS-2005), Xplorer GLX (PS-2002), Xplorer
(PS-2000), PowerLink (PS-2001), and USB Link (PS-2100)
®
Water Quality MultiMeasure SensorIntroduction
Introduction
The PS-2169 Water Quality MultiMeasure Sensor combines four sensors in a single
unit:
•Temperature
•Conductivity
•pH, ORP and ion-selective electrode voltage
•Dissolved Oxygen (DO
)
2
When connected to a PASPORT interface, the multi-sensor collects data at up to 20
samples per second from each component sensor. You can use just one component
sensor at a time or any combination simultaneously. If you have a PASPORT interface
that supports multiple sensors, or if you have more than one interface connected to
your computer, you can use the Water Quality MultiMeasure Sensor in combination
with other PASPORT sensors.
Note: No electrical interference will occur between two or more probes connected to a single
Water Quality MultiMeasure Sensor.
Sensor, Interface, and Software Setup
Connect the multi-sensor to your PASPORT interface as pictured (right).
Connect any or all of the included probes to the multi-sensor. (You can leave
any probe disconnected if you do not plan to use it.) If you will be using an
ion-selective electrode or oxygen reduction potential electrode, connect it to
the pH/ISE/ORP port.
Follow the instructions below to set up the multi-sensor with DataStudio
software (if you are using a computer) or the Xplorer GLX (if you are using
it standalone without a computer).
For detailed information about each of the multi-sensor’s component sensors
see “Temperature” on page 3, “Conductivity” on page 3, “pH/ISE/ORP” on
page 4, and “Dissolved Oxygen” on page 6.
DO
ConductivitypH/ISE/ORP
CONDUCTIVITYpH/ISE/ORP
2
QUALITY
SENSOR
WATER
DISSOLVED OXYGEN
Temperature
TEMPERATURE
ä
PS-2169
Note: Only the tips and handles of the probes should be immersed in liquid, keep the
cables, connectors and body of the multi-sensor dry.
DataStudio Setup
If you are using a computer, connect the PASPORT interface to the computer and start
DataStudio. (Refer to the documentation that came with your PASPORT interface for
instructions on connecting it to your computer.) In DataStudio, click the Start button
to begin data collection; click the Stop button to stop data collection.
Note that the sensor may collected data for probes that are not connected. Hide or
ignore these measurements. To hide measurements that you do not need, open the
Experiment Setup Window by clicking the Setup button. In the Setup window you
can also change the sampling rate and open the calibration dialog box (see Appendix
A). For more instructions on using DataStudio, press F1 to open DataStudio’s on-line
help.
2
PASPORT
Interface
Click to Start
®
Model No. PS-2169 Temperature
Xplorer GLX Setup
If you are using an Xplorer GLX in standalone mode, press to start data collection.
Note that the sensor may collected data for probes that are not connected. Hide or
ignore these measurements. To hide measurements that you do not need, open the
Sensors screen (from the Home screen, press ). In the Sensors screen you can
F4
also change the sampling rate and open the calibration dialog box. For more instructions on using the Xplorer GLX, refer to the GLX Users’ Guide.
Temperature
Use this component of the multi-sensor to measure the temperature of a fluid or
object. The measurement can be displayed in units of °C, °F, or K. (In DataStudio,
click the Setup button to change units.) You can use the included stainless steel probe
or a different probe, such as the Fast-response Probe (PS-2135) or Skin/Surface Probe
(PS-2131). PASCO physics apparatus containing an embedded 10 kΩ thermistor can
also be connected to the multi-sensor. The multi-sensor automatically detects the
presence of a temperature probe, and it will only collect temperature data if a probe is
connected.
Measuring Temperature
To measure temperature, connect the probe and start data collection. Immerse the tip
of the probe in a fluid or place it in contact with an object. The included stainless steel
probe can be used in both dry conditions and in liquids, such as water and other mild
chemicals and solutions.
Tip: For better chemical resistance, use a Teflon® cover (CI-6549).
Temperature Calibration
The temperature measurement can be calibrated; however, for most applications calibration is not necessary. Use a two-point calibration with two standards of known
temperature. For instructions on calibrating in DataStudio, see Appendix A. For
instructions on calibrating on the Xplorer GLX, refer to the Xplorer GLX Users’
Guide.
Conductivity
The conductivity component sensor measures the electrolytic conductivity of aqueous
solutions. It determines the conductivity by applying an alternating voltage to the
electrode cell at the end of the probe and measuring the resulting current. You can use
the included 10x electrode or a 1x electrode (PASCO part 699-06620). In the software
or on the interface select the Conductivity (10x) or Conductivity (1x) measurement,
depending on which electrode you are using.
glass rod
Note: The 10x probe can be used to measure conductivity up to 100000 µS/cm. The 1x probe
has one tenth of the range (up to 10 000
µ
S/cm) but ten times better resolution.
Measuring Conductivity
Before using the conductivity electrode, soak the end in distilled water for 5 to 10
minutes. Connect the electrode to the multi-sensor and start data collection. Immerse
®
platinized
platinum
conductors
Conductivity Electrode
3
Water Quality MultiMeasure SensorpH/ISE/ORP
the end of the electrode in the solution to be measured and wait for the reading to stabilize. Rinse the electrode with deionized water before measuring a different solution.
Range (mS/cm) 10x probe
The conductivity sensor has three ranges, which you select using the three buttons on
the multi-sensor. To select a range, push one of the buttons: for 0
µS/cm, for or 0 µS/cm to 10000 µS/cm, or for 0 µS/cm to 100000 µS/cm.
µS/cm to 1000
Lights on the buttons indicate which range is selected. To determine which range is
appropriate, look at the data (on a Graph display, for instance) while it is being collected; if the measurement appears to be “railed” at the top of the selected range
(1000
µS/cm or 10000 µS/cm), select the next higher range. You can push a button to
change the range without stopping data collection.
Note: The ranges specified above are for a 10x probe. Divide each range by 10 if you are using a
1x probe.
The conductivity electrode is sensitive to very low concentrations of dissolved solids,
so you will probably never measure a sample with a conductivity of zero. A reading
of 25 µS/cm for “pure” water is typical. True pure water is difficult to obtain and
store. The table (right) lists typical conductivity values for common aqueous solutions
at 25 °C.
As a rule of thumb, to estimate total dissolved solids (TDS) in parts per million
(ppm), divide the conductivity in µS/cm by 2:
The conductivity measurement can be calibrated; however, for most applications calibration is not necessary. Use a one-point calibration with a standard solution of
known conductivity. For instructions on preparing a standard solution, see Appendix
C. For instructions on calibrating in DataStudio, see Appendix A. For instructions on
calibrating on the Xplorer GLX, refer to the Xplorer GLX Users’ Guide.
Conductivity Electrode Maintenance And Storage
To ensure accurate and reproducible results, the electrode must be clean. Substances
on the electrode may contaminate the solution being tested and change the conductivity. To clean the electrode, dip the end into a detergent solution or dilute nitric acid
(1%) and stir for three minutes, then rinse with deionized water.
For storage, dry the electrode and return it to its box. Before use, soak the electrode in
distilled water for at least 5 minutes.
pH/ISE/ORP
The pH/ISE/ORP input of the multi-sensor is a specialized voltage sensor. Its BNC
connector accepts the included pH electrode, as well as ISE and ORP electrodes. The
sensor measures the voltage produced by any of these electrodes. When used with a
pH probe, the sensor also computes the pH based on the measured voltage.
Connect the pH electrode (or other electrode) to the pH/ISE/ORP port and start data
collection. On your computer or interface, display the ISE Voltage measurement (this
measurement is valid for pH and ORP electrodes as well as ISE) or the pH measurement for the calculated pH (assuming that the pH electrode is connected).
4
®
Model No. PS-2169 pH/ISE/ORP
Measuring pH
The pH electrode produces a voltage proportional to the pH of the solution that it is
immersed in. This voltage is measured by the multi-sensor, which computes pH.
Unscrew and remove the storage bottle from the electrode (be careful not to spill the
storage solution). Push the O-ring and bottle cap up the electrode handle. Rinse the
electrode tip with distilled water. If you see bubbles in the electrode bulb, gently
shake the electrode downward (similar to shaking down a thermometer). Start data
collection. Place the tip of the electrode in the solution to be measured and wait for
the reading on your computer or interface to stabilize. Rinse the electrode with distilled water before measuring another solution.
pH Calibration
The pH measurement can be calibrated; however, for most applications calibration is
not necessary. Perform a two-point calibration with two buffer solutions of known
pH. For instructions on calibrating in DataStudio, see Appendix A. For instructions
on calibrating on the Xplorer GLX, refer to the Xplorer GLX Users’ Guide.
pH Electrode Maintenance and Storage
Cleaning
If the pH electrode becomes contaminated, use one of these methods to clean and
restore it. After any of these procedures, soak the electrode in a pH 7 buffer solution
for 30 minutes.
•General Contamination: Soak the electrode in 0.1 molar hydrochloric acid
(HCl) for 15 minutes.
•Protein Deposits: Soak the electrode in a solution of 1% pepsin in 0.1 molar
HCl.
•Inorganic Deposits: Rinse the electrode with 0.1 molar ethylene dinitric
tetra-acidic acid (EDTA) tetrasodium solution.
•Oil and Grease Film: Wash the electrode carefully in a mild detergent or a sol-
vent known to be effective for the particular film.
•Unknown Contamination: Soak the electrode alternately in 12 molar sodium
hydroxide (NaOH) and 1 molar HCl. Leave it in each solution for one minute.
Rinse completely between soakings. End with HCl. (The NaOH etches the glass
and the HCl reestablishes hydrogen ions on the surface.)
If these steps fail to improve the response of the electrode, replaced it.
Storage
Store the pH electrode in the included electrode storage bottle with one of the following solutions. (Never store the electrode in distilled water.)
•Short-term (up to one week): pH 4 buffer solution or tap water.
•Long-term (over one week): pH 4 buffer solution with 1 g per 100 mL of potas-
sium chloride (KCl) added.
®
5
Water Quality MultiMeasure SensorDissolved Oxygen
Dissolved Oxygen
The DO2 component of the multi-sensor is designed for use in aqueous media at temperatures ranging from 0 °C to 50 °C. For greatest accuracy, the following requirements should be met:
•Although the unit is temperature compensated, it should be calibrated at approximately the same temperature as the test solution.
silver
anode
filling
solution
•The sensor must equilibrate for a short period when the temperature of the test
solution changes.
•The test solution should constantly flow past the membrane of the probe. This
can be accomplished by gently swirling the electrode during measurement.
The DO
rounded by potassium chloride (KCl
probe consists of a platinum (Pt) cathode and a silver (Ag) anode sur-
2
) electrolyte filling solution with a silicon
(aq)
membrane separating the filling solution from the test solution. When the probe is
placed in an aqueous medium that contains dissolved oxygen, O
across the membrane and react with water molecules (H
trons from the cathode to form hydroxide ions (OH
O) in the presence of elec-
2
-
). Hydroxide ions diffuse to the
anode and react with silver atoms, forming silver oxide (Ag
molecules diffuse
2
O), H2O, and free elec-
2
trons. The net movement of negative charge (electrons) from the cathode to the anode
is measured by the sensor as an electric current. Since the rate of O
the membrane is proportional DO
resulting electric current is also proportional to DO
concentration, the rate of these reactions and the
2
concentration. The rate at which
2
diffusion across
2
the probe processes oxygen also depends on temperature. To correct for temperature
variation, the probe contains a temperature-sensing thermistor, which adjusts the gain
of an amplifier in the multi-sensor.
Adding Filling Solution
Prior to its initial use, and after long-term storage or cleaning, prepare the probe by
filling it with electrolyte solution as follows:
cartridge
housing
cathode
insulator
membrane
Detail of DO2 Probe Tip
platinum
cathode
Unscrew
cartridge
housing
1.Remove the storage bottle from the end of the probe.
2.Unscrew and remove the cartridge housing at the end of the probe, as illustrated
(right).
3.Pull about 10 mL of new filling solution into the syringe (included with the
probe). Be careful not to introduce air bubbles.
4.Place the tip of the syringe in the cartridge housing, very close to—but not touching—the membrane. Slowly fill the housing to about 5 mm from the top. Agitate
the housing while filling to avoid introducing air bubbles.
5.Replace the housing over the electrode and screw it into place. Excess filling
solution will spill from the housing as you do this.
6.Rinse and dry the outside of the probe.
6
Fill cartridge
housing
fill to here
Replace
cartridge
housing
®
Model No. PS-2169 Dissolved Oxygen
DO2 Calibration
Perform a one-point calibration using 100% humidified air as the known standard,
which is equivalent to 100% saturated (pure) water. Prepare 100% humidified air as
follows:
1.Place about 5 mL (to a height of about 1 cm) of deionized water into a clean electrode storage bottle.
2.Slip the cap and O-ring of the soaker bottle over the end of the probe. Insert the
probe into the soaker bottle and screw on the cap. Adjust the height of the end of
the probe to about 2 cm above the water.
3.Shake the probe and bottle vigorously for about 10 seconds. Shake off any large
water drops from the membrane.
4.Hold the probe vertically so that it is in the air above the water, not in the water
itself.
5.Perform a one-point calibration on either the Dissolved Oxygen (%) or Dis-
solved Oxygen (mg/L) measurement. If you are calibrating the percentage mea-
surement, enter “100” as the known value. If you are calibrating the mg/L
measurement, look up the known concentration in Appendix D. For instructions
on calibrating in DataStudio, see Appendix A. For instructions on calibrating on
the Xplorer GLX, refer to the Xplorer GLX Users’ Guide.
Measuring DO2
Follow the instructions above to fill the probe with electrolyte solution and calibrate
the measurement. Start data collection. Remove the storage bottle from the probe.
Immerse the end of the probe in the solution to be measured. Gently swirl the probe or
stir the solution to prevent localized oxygen depletion at the probe end. Wait for the
reading to stabilize. Rinse the electrode with deionized water before measuring a different solution.
Shake
vigorously
Electrode
storage bottle
Probe tip
Water level
Preparation of 100%
humidified air for
calibration
DO2 Probe Maintenance and Storage
The probe comes with extra filling solution, a replacement membrane cartridge and
O-ring, and a plunger for installing the cartridge. Order the Electrode Maintenance
Kit (CI-6541) to replace these items.
Syringe
Filling
Solution
Plunger
Electrode Cleaning
To maintain optimal performance, periodically clean the silver electrode and replace
the electrolyte filling solution as follows:
1.Unscrew and remove the cartridge housing at the end of the probe. Discard the
old filling solution from the cartridge housing.
®
Replacement
membrane cartridge
Replacement
O-ring
Remove loose silver oxide
from cathode
7
Loading...
+ 14 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.