PASCO PS-2164 User Manual

Page 1
Quad Pressure Sensor
PS-2164
Quad Pressure
Sensor
1
QUAD PRESSURE SENSOR
234
PRESSURE SENSORS
Instruction Sheet
012-09012A
Pressure Ports
Polyurethane Tubing
Included Equipment Part Number
Quad Pressure Sensor
Polyurethane Tubing (length 2.4 m)
Quick-release Connectors (qty. 4)
Tubing Connectors (qty. 4)
PS-2164
640-023
640-021
640-030
Required Equipment
PASPORT Interface
1 See PASCO catalog
or www.pasco.com
PS-2164
POWER
Connect to PASPORT interface
Quick release Connectors
Tubing Connectors
Introduction
The four-port Quad Pressure Sensor is a multi-sensor capable of reading up to four absolute pressures, or two differential pressures, or two gauge pressures. Dynamic variable over-sampling automatically reduces measurement noise at low sampling rates. Sample rates up to 1000 Hz make studies of both transient and steady-state pressure possible. The sen­sor measures pressure in units of kPa, N/m square inch (psi).
2
, and pounds per
1
PASPORT interfaces include the AirLink (IW-4100 or PS-2005), Xplorer GLX (PS-2002), Xplorer (PS-2000), PowerLink (PS-2001), and USB Link (PS-2100)
®
Setup
Types of Pressure Measurements
To measure pressure in an apparatus, connect one or more of the four pressure ports using the included tubing and connec­tors. The way that you connect the Quad Pressure Sensor
Page 2
Model No. PS-2164 Setup
depends on whether you will measure absolute pressure, dif­ferential pressure, or gauge pressure:
Absolute Pressure The multi-sensor can measure up to four absolute pressures simultaneously. One connection is required for each measurement, and any of the pressure ports may be used.
Differential Pressure The multi-sensor measures differ­ential pressure by determining the pressure difference between two ports. It can measure one or two differential pressures simultaneously, with two connections required for each mea­surement. Differential pressure can be measured between ports 1 and 2 or between ports 3 and 4 (or both).
Gauge Pressure Gauge pressure is a special case of dif­ferential pressure in which the second port of the pair is left unconnected. The multi-sensor measures the pressure differ­ence between the first port (connected to the apparatus) and the second port (left open to the atmosphere). It can measure up to two gauge pressures simultaneously, with one connec­tion required for each measurement. Gauge pressure can be measured by port 1 (with port 2 left open) and by port 3 (with port 4 left open).
5. Connect the multi-sensor’s cable to your PASPORT inter-
face.
DataStudio Setup
If you are using a computer, connect the PASPORT interface to the computer and start DataStudio. (Refer to the documen­tation that came with your PASPORT interface for instructions on connecting it to your computer.) In DataStudio, click the Start button to begin data collection; click the Stop button to stop data collection.
By default, DataStudio displays the absolute pressures mea­sured by all four ports in a Graph display. To make the differ­ential measurements available for display, click the Setup button to open the Experiment Setup window; in that win­dow, select 1-2 Differential Pressure or 3-4 Differential Pressure. There you can also change the sampling rate (up to 1000 Hz) and select different units of measure. For more instructions on using DataStudio, press F1 to open DataStu­dio’s on-line help.
Xplorer GLX Setup
Sensor Setup
1
QUAD PRESSURE SENSOR
234
PRESSURE SENSORS
1. Cut a piece (or pieces) of tubing to the desired length.
2. Insert a quick-release connector into each piece of tubing.
If you are using an Xplorer GLX in standalone mode, press
to start data collection. By default, the GLX displays the
absolute pressure from port 1 (Absolute Pressure 1) in units of kPa, but you can select any of the other absolute or differ­ential measurements for display. For more instructions on using the Xplorer GLX, refer to the GLX Users’ Guide.
Multiple-measurement Alignment
When you are using multiple pressure ports, you may find that their readings differ slightly when they are measuring the same thing. In the example illustrated below, measurements from four ports open to the atmosphere vary by about 0.2 kPa. The multiple-measurement alignment procedure adjusts the calibrations of the measurements so they agree more closely.
3. Connect each piece of tubing to the multi-sensor by
inserting the quick-release connector into one of the pres­sure ports and twisting clockwise.
4. Connect the free ends of the tubing to the apparatus using
connectors supplied with the apparatus or the tubing con­nectors include with the Quad Pressure Sensor.
You can use the included tubing connectors to connect to a 5 mm ( inch) hole in a rubber stopper.
®
The procedure described here for DataStudio can also be performed on the Xplorer GLX. See the Xplorer GLX Users’ Guide for detailed instructions.
1. Click the Setup button to open the Experiment Setup window. Click Calibrate Sensors to open the calibration dialog box.
2
Page 3
Model No. PS-2164 Specifications
2. From the Sensor field, select the Quad Pressure Sensor.
(See screenshot below.)
Select Quad Pressure Sensor
Select Calibrate all...
Select 1 Point (Adjust Offset Only)
click Read From Sensor
measurement resolution. This effect is especially noticeable when very small pressure changes or differences are mea­sured. The degree of dynamic variable over-sampling that takes place within the Quad Pressure Sensor depends on the sample rate. To maximize the over-sampling, set the sample rate as low as possible for a given application. Maximum over-sampling occurs at sampling rates of 1 Hz or slower.
Specifications
Range Absolute pressure: 0 to 200 kPa
Differential pressure: −100 kPa to +100 kPa
Resolution 0.01 kPa at 10 Hz
Repeatability 1 kPa
Units of Measure kPa, N/m
Max. Sample Rate 1000 Hz
Included tubing Material: polyurethane
Inside diameter : 3.2 mm ( inch) Length: 2.4 m (8 feet)
2
, psi
click OK
3. Select the Calibrate all similar measurements simulta- neously option.
4. From the Calibration Type menu, select 1 Point (Adjust Offset Only).
5. Apply the same pressure to all four ports (for instance by
opening them to the atmosphere).
6. Click the Read From Sensor button for Point 1.
7. Click OK to accept the new calibration.
One-point and Two-point Calibrations
Though it is usually not necessary, a one-point or two-point calibration can be performed on any of the absolute or relative measurements to make them more accurate. To do so, you must have a very accurate independent means of measuring pressure, such as a barometer. For step-by-step instructions refer to the documentation for DataStudio version 1.9.5, the Xplorer, or the Xplorer GLX.
Over-sampling
Technical Support
For assistance with any PASCO product, contact PASCO at:
Address: PASCO scientific
10101 Foothills Blvd. Roseville, CA 95747-7100
Phone: 916-786-3800 (worldwide)
800-772-8700 (U.S.)
Fax: (916) 786-3292
Web: www.pasco.com
Limited Warranty
For a description of the product warranty, see the PASCO catalog.
Copyright
The PASCO scientific 012-09012A
Sheet
is copyrighted with all rights reserved. Permission is granted to non-profit educational institutions for reproduction of any part of this man­ual, providing the reproductions are used only in their laboratories and classrooms, and are not sold for profit. Reproduction under any other cir­cumstances, without the written consent of PASCO scientific, is prohibited.
Trademarks
PASCO, PASCO scientific, DataStudio, PASPORT, and ScienceWorkshop are trademarks or registered trademarks of PASCO scientific, in the United States and/or in other countries. All other brands, products, or ser­vice names are or may be trademarks or service marks of, and are used to identify, products or services of, their respective owners. For more infor­mation visit www.pasco.com/legal.
Quad Pressure Sensor Instruction
The Quad Pressure Sensor uses dynamic variable over-sam­pling to reduce noise, produce smoother data, and improve the
®
3
Loading...