Instruction Sheet
012-09741A
Galvanometer
PS-2160
R
E
T
E
M
O
R
O
N
S
A
N
E
V
S
L
A
G
E
R
A
T
PS-2160
Included Equipment Part Number
Galvanometer PS-2160
BNC-to-banana plug cable 514-015
BNC-to-binding post adapter 515-012
Resistor, 10 Ω, 5 W, 5% 111-100
Resistor, 0.1 Ω, 3 W, 1% 130-019
Required Equipment
PASPORT Interface
1
See PASCO catalog or
www.pasco.com
1
PASPORT interfaces include the AirLink (PS-2005), Xplorer GLX
(PS-2002), Xplorer (PS-2000), PowerLink (PS-2001), and USB Link
(PS-2100)
Introduction
The PASPORT Galvanometer is a high-resolution,
high-impedance voltage sensor for use with PASPORT interfaces. It measures differential voltage in the range of −2 V to
+2 V. It can also be used in combination with a resistor as a
current sensor.
Sensor Set-up
Connect the Galvanometer Sensor to a PASPORT interface. If
you are using a computer, start DataStudio.
For measuring voltage, connect the included cable or binding
post adapter to the sensor's BNC connector, or connect a
device to the Galvanometer with a BNC cable.
Before making a measurement, short the + and − inputs
together and press the Tare button on the sensor. This adjusts
the sensor's measurement to 0 V.
Connect BNC
cable or adapter
R
E
T
E
M
O
N
SENSOR
A
V
L
A
G
Tare Button
PS-2160
TARE
PASPORT Interface
®
Model No. PS-2160 Over-sampling
To start data collection, click the Start button in DataStudio or
press on the Xplorer GLX.
By default, the sensor collects data at 10 Hz and displays it in
units of volts. In DataStudio, click the Setup button to open
the Experiment Setup window, where you can change the
sample rate and units. On the Xplorer GLX, press ,
to open the Sensors screen, where you can change the sample
rate and units.
Note: For complete instructions on collecting and displaying and data,
press F1 for DataStudio's on-line help, or refer to the Xplorer GLX
Users' Guide.
Over-sampling
The Galvanometer uses dynamic variable over-sampling to
reduce noise, produce smoother data, and improve measurement resolution. This effect is especially noticeable when very
small voltage changes are measured. The degree of dynamic
variable over-sampling that take place within the Galvanometer depends on the sample rate. To maximize the over-sampling, set the sample rate as low as possible for a given
application. Maximum over-sampling occurs at sampling rates
of 1 Hz or slower.
Before making a measurement, ensure that no current is flowing through the resistor and press the Tare button.
Current Calculation
F4
To calculate the current (I ) from the measured voltage (V ), use
Ohm's Law:
V
I
---=
R
where R is the combined resistance of the resistor and the
leads connecting it to the binding posts.
A reliable way to measure R is to apply an accurately known
current (I' ) and note the measure voltage (V'). The resistance
is
V′
R
-----=
I′
Once you have determined R, you can configure DataStudio
or the Xplorer GLX to automatically calculate the current. In
DataStudio, click the Calculate button and define a calcula-
tion such as
Measuring Current
Current Measurement Set-up
To measure current, connect the included binding post adapter
to the sensor's BNC connector. Loosen the binding post
screws and connect one of the included resistors (or another
resistor) as pictured.
where, in this example, “9.89” is the value of R in Ω.
On the GLX (in standalone mode), press , to open
F3
the Calculator screen and define a calculation such as
In this configuration, the sensor will measure the voltage drop
across the resistor as current flows through it. Connect a circuit or device so that positive current will flow through the
resistor from the red post to the black post.
®
Note: For complete instructions on defining calculations, press F1 for
DataStudio's on-line help, or refer to the Xplorer GLX Users' Guide.
2