Analog Adapter
PS-2158
Instruction Sheet
012-08921A
Introduction
The PS-2158 Analog Adapter allows an analog
ScienceWorkshop sensor to be used with a PASPORT interface. Multiple Analog Adapters can be used simultaneously to
support multiple ScienceWorkshop sensors.
Compatible PASPORT Interfaces Part Number
Xplorer GLX
Xplorer1,
Power Link
USB Link
1
When the Analog Adapter is used with an Xplorer or Xplorer GLX
in logging mode (not connected to a computer), the logger records
output voltage from the sensor. A future free upgrade to the GLX
will support calibrated data logging. Visit www.pasco.com for the
latest GLX upgrade.
2
When the Analog Adapter is used with an Xplorer (PS-2000) in
logging mode, the Xplorer Power Adapter (PS-2530) is recommended (and in some cases required) due to the high current draw
of some ScienceWorkshop sensors.
1 PS-2002
2
PS-2000
PS-2001
PS-2100
Equipment and Software Setup
DataStudio version 1.9.5 (or later) is required. Visit
www.pasco.com to download the latest version of DataStudio.
The first three steps can be performed in any order.
1. Connect an analog ScienceWorkshop sensor to the cable
of the Analog Adapter.
2. Connect the Analog Adapter to a PASPORT interface.
3. Connected the PASPORT interface to a computer.
4. If DataStudio is not running, it will launch automatically.
5. The ScienceWorkshop Analog Sensors list will open
automatically. Select the connected sensor from the list
and click OK.
6. To view and change the sampling rate, gain, and other set-
tings of the sensor and adapter, press the Setup button in
DataStudio’s main tool bar; the Experiment Setup window will open. Refer to DataStudio documentation for
further instructions.
®
Analog Adapter
Specifications
Maximum input voltage range
1
−10 V to +10 V
Absolute maximum input
voltage range without damage
Input impedance 1 MΩ
Gain 1, 10, and 100
Maximum Sampling Rate 50 kHz with Xplorer GLX
Analog-to-Digital Conversion 12 bit
Voltage Resolution 5 mV at gain = 1
Offset voltage accuracy < ±5 mV
Full-scale voltage accuracy < ±15 mV
1
When used with a ScienceWorkshop Voltage Sensor (CI-6503) or
Current Sensor (CI-6556), the sensor measures differential voltage,
but the maximum input voltage range to either terminal is ±10 V.
−40 V to +40 V
(selectable in DataStudio)
1000 Hz with other interfaces
0.5 mV at gain = 10
0.05 mV at gain = 100
Technical Support
For assistance with any PASCO product, contact PASCO at:
Address: PASCO scientific
10101 Foothills Blvd.
Roseville, CA 95747-7100
Phone: 916-786-3800 (worldwide)
800-772-8700 (U.S.)
Fax: (916) 786-3292
Web: www.pasco.com
Email: support@pasco.com
Limited Warranty
For a description of the product warranty, see the PASCO catalog.
Copyright
The PASCO scientific 012-08921A
copyrighted with all rights reserved. Permission is granted to non-profit
educational institutions for reproduction of any part of this manual, providing the reproductions are used only in their laboratories and classrooms,
and are not sold for profit. Reproduction under any other circumstances,
without the written consent of PASCO scientific, is prohibited.
Trademarks
PASCO, PASCO scientific, DataStudio, PASPORT, and ScienceWorkshop
are trademarks or registered trademarks of PASCO scientific, in the
United States and/or in other countries. All other brands, products, or service names are or may be trademarks or service marks of, and are used to
identify, products or services of, their respective owners. For more information visit www.pasco.com/legal.
Analog Adapter Instruction Sheet