PASCO PS-2117 User Manual

Drop Counter
Stopcock (2)
Micro Stir Bar
Drop Tip
Drop Counter
Syringe
Support Rod
PS-2117
Instruction Sheet
012-08470E
.
Included Items Included Items
Drop Counter Stopcock (2) Micro Stir Bar Drop Tip Syringe (60 mL) Support Rod
Required Items Required Items
PASCO Interface* Data Acquisition Software*
*See the PASCO catalog or the PASCO web site at www.pasco.com for more information
.
Recommended Recommended
pH Sensor* Temperature Senso r*
Consumables Consumables
0.1 M NaOH 0.005 M HCl
Other Needed Items*
Three-finger Clamp (SE-9445) Support Rod and Stand (SE-9451) Multi Clamp (ME-9507) Magnetic Stir Plate (SE-7770) Beaker, 150 mL Graduated Cylinder, 10 mL

Quick Start

ble PASCO interface.
2. If you are using a computer, connect the interface to it
and start the data acquisition software.
3. Touch, press, or click to begin recording data.
4. Allow drops of liquid to fall through the rectangular
opening.
Drop Counter
60 mL syringe
with 50 mL of
0.1 M NaOH
150 mL
beaker with
100 mL of
0.005 M HCl
Three-finger
Clamp
Multi
Clamp
Stopcock Stopcock
Drop Tip
pH Probe
Micro Stir
Bar
Stir Plate

Introduction

The Drop Counter measures the number of drops of fluid that fall through the rectangular opening of the sensor. It is typically used in conjunction with a pH sensor and other equipment to perform a titration. Data from the Drop Counter and other sensors are recorded and displayed in soft­ware such as SPARKvue, PASCO Capstone, or on a datalog­ging interface such as the SPARK SLS or Xplorer GLX.
The Drop Counter can be mounted on a support rod and the support rod can be clamped to a rod stand. The counter is equipped with three integrated probe holders. The included Micro Stir Bar fits onto the end of a pH probe or probe of the same diameter.

Assembling the Drop Dispenser

1. Remove the plunger from the syringe.
2. Connect the two stopcocks together and set the valves at
right angles. Connect the stopcocks to the end of the syringe.
3. Connect the drop tip to the bottom of the stopcocks.

Acid-Base Titration (Calibration Method)

Calibration
1. Attach the Drop Dispenser (syringe, stopcocks and drop
tip) to a support rod.
2. Fill the syringe with titant and adjust the valves on the
stopcocks so that the drops fall at a rate of about 1 drop per second. Close the bottom valve after the rate is achieved.
3. Connect the Drop Counter to a PASPORT-compatible
interface.
4. Align the drop tip with the opening in the Drop Counter.
Put a 10 ml graduated cylinder under the Drop Counter opening.
5. Open the bottom stopcock valve and monitor the num-
ber of drops per 10 mL of titrant.
6. Open the calibration screen in the data acquisition soft-
ware and input the number of drops and the volume.
Procedure
1. Setup the equipment and consumables as shown.
2. Connect the Drop Counter and pH Sensor to the
PASPORT-compatible interface.
3. Start the data acquisition software.
4. Use the software to create a graph of pH vs. Fluid Vol-
ume.
5. Start the stir plate.
6. Touch, press, or click to begin recording data.
7. Open the valve on the bottom stopcock so the titrant
drops fall at about 1 drop per second.
2
Model No.PS-2117
Select fluid volume for
horizontal axis.
equivalence point
final drop count
Select volume calculation
for horizontal axis
8. Observe the data on the pH versus Fluid Volume graph.
After the equivalence point is reached, continue collect­ing data until the pH curve flattens.
10. Close the stopcocks.
11. Stop data recording.
12. Read the final volume of fluid in the syringe and calcu-
9. Close the stopcock valve and stop data recording.
late the net volume of titrant that was dispensed.

Acid-Base Titration (Alternative Method)

Procedure
1. Set up the equipment and consumables as illustrated.
Attach both stopcocks to the syringe, and attach the drop tip to the bottom stopcock. Line up the drop tip with the rectangular opening in the Drop Counter.
2. Start with the valves of the stopcocks turned parallel to
the table top. Add the solutions to the syringe and bea­ker.
3. Write down the initial volumes of both solutions (in the
syringe and in the beaker), and the molarity of the titrant (in the syringe).
4. Connect the Drop Counter and pH sensor to your
PASPORT compatible PASCO interface (or interfaces).
5. If you are using a computer, connect the interface to it
and start the data acquisition software. Set up a graph display of pH versus Drop Count.
6. Start the stir plate.
13. Read the final drop count on the graph.
14. In the calculator window of the software or datalogger,
enter a calculation as shown below, but using your own values of total volume dispensed and final drop count. The calculation gives the volume of titrant dispensed over time based on the number of drops and the average volume per drop.
volume = [Drop Count (drops)] * 100/3050
In this example, “100” is the total volume dispensed (in mL) and “3050” is the final drop count.
15. In the graph, change the horizontal axis to the calculated
“volume”.
7. Touch, press, or click to begin recording data.
8. Open the top stopcock. Slowly adjust the bottom stop-
cock to start delivering titrant at about 2 drops per sec­ond.
9. Observe the data appearing on the pH versus drop count
graph. After the equivalence point is reached, continue collecting data until the pH curve flattens.
3
Drop Counter Specifications
detector
collimating
lens
focusing lens
sheet of light
dropinfrared light
source
Tips
Except for opening and closing the stopcocks, do not touch the syringe during the titration; otherwise, the drop size may change significantly.
The drop rate must remain approximately constant (about 2 drops per second) in order for the drop size to remain constant.
Drop size depends on the exact position of the bottom stopcock. Therefore, the average drop size will be dif­ferent for every titration. For each titration, re-enter the volume used and drop count values in the calculation.

Other Titrations

Titrations with different solutions can be performed using the same equipment and procedure. Other probes (such as conductivity or ORP) can be used in place of the pH probe.

Theory of Operation

The Drop Counter uses an infrared light source and a photo­detector. Lenses spread the light into a “sheet” and refocus it onto the photodetec­tor. When a drop of fluid passes through the Drop Counter, it partially blocks this sheet of light, and the photodetector registers a momentary decrease in light intensity.
The Drop Counter uses infrared light and ignores visible light. On power up, it automatically adjusts the light level for the best sensitivity. Use the Drop Counter away from direct sunlight or other sources of infrared that may interfere with it.

Troubleshooting

Problem Cause Solution
Indicator light flashes twice for a single drop. Drop Counter registers more drops than actually dispensed.
Indicator LED remains on constantly. Drop Counter does not register drops.
Fluid is splashing back into the drop window.
Lenses in drop window are dirty or wet.
Adjust the position of the beaker, Drop Counter, or syringe to eliminate splash back
Clean lenses with water and dry with a cotton swab or tissue.

Specifications

Item Value Range: -35 to 135°C Accuracy: ±0.5°C Resolution: 0.01°C or better Response TIme Wait 15 seconds for stable readings
in liquids, and wait 30 to 60 seconds for stable readings in gases.

Technical Support

For assistance with any PASCO product, contact PASCO at:
Address: PASCO scientific
10101 Foothills Blvd. Roseville, CA 95747-7100
Phone: +1 916-786-3800 (worldwide)
800-772-8700 (U.S.) E-mail: support@pasco.com Web www.pasco.com
Limited Warranty For a description of the product warranty, see the PASCO catalog. Copyright The PASCO scientific Instruction Sheet is copyrighted with all rights reserved. Permission is granted to non-profit educational institutions for reproduction of any part of this manual, providing the reproductions are used only in their laborato­ries and classrooms, and are not sold for profit. Reproduction under any other circumstances, without the written consent of PASCO sci­entific, is prohibited. Trademarks PASCO, PASCO Capstone, PASPORT, SPARK Science Learning System, SPARK SLS, and SPARKvue are trademarks or registered trademarks of PASCO sci­entific, in the United States and/or in other countries. For more infor­mation visit www.pasco.com/legal.
Product End of Life Disposal Instructions:
This electronic product is subject to disposal and recycling regulations that vary by country and region. It is your responsibility to recycle your electronic equipment per your local environmental laws and regulations to ensure that it will be recycled in a manner that protects human health and the environment. To find out where you can drop off your waste equipment for recycling, please contact your local waste recycle/disposal service, or the place where you pur­chased the product.
The European Union WEEE (Waste Elec­tronic and Electrical Equipment) symbol (to the right) and on the product or its packaging indicates that this product must not be disposed of in a standard waste con­tainer.
4
Loading...