PASCO PS-2102 User Manual

®
pH Sensor
1
3
2
to PASPORT interface
pH electrode
PS-2102
Instruction Sheet
012-07308G
Included Parts
1. pH amplifier
3. Storage bottle containing storage solution
Additional Part Required
PASPORT interface or datalogger
Related Parts
Replacement electrode (PASCO part 699-195)
Flat-surface electrode (PS-2182)
pH buffer set (SC-2321)
Quick Start
1. Connect the amplifier to your PASPORT interface.
2. Connect the pH electrode to the amplifier.
3. If you are using a computer, connect the PASPORT interface
to it and start DataStudio.
Introduction
The pH Sensor measures the pH of a solution. The measurement is recorded and displayed by the connected computer or datalog­ger. The pH Sensor is well-suited for continuous recording and discrete measurements.
Set-up
Connecting the Parts
1. Plug the sensor’s PASPORT connector into any port of a PASPORT interface or datalogger.
2. Push the connector at the end of the electrode cable onto the sensor’s electrode connector. Twist the connector to lock it in place.
4. Remove the storage bottle from the electrode.
5. Rinse the electrode with distilled water.
6. Immerse the end of the electrode in the solution to be mea-
sured.
7. Press or click the start button to begin recording data.
Connecting the electrode and interface
3. If you are using a computer, connect the PASPORT interface to it and start DataStudio.
®
pH Sensor PS-2102
cap
bottle
Removing
the bottle
Removing the Storage Bottle
1. Hold the electrode vertically so that the solution will not spill out of the bottle.
2. Unscrew and remove bottle. Keep the storage solution for later use.
3. Push the bottle cap up the electrode shaft to keep it out of the way.
Collecting Data
1. Rinse the electrode with distilled water.
2. Immerse the end of the electrode in the solution to be mea-
sured. The bulb-shaped glass membrane should be entirely immersed.
3. Press or click the start button to begin recording data.
4. Wait for the reading to stabilize.
5. Rinse the electrode again before placing it in another solu-
tion.
Electrode Maintenance
Electrode Storage
For a storage period of a few weeks or less, rinse the electrode with distilled water and place it in the storage bottle filled with storage solution. To make storage solution, combine equal parts 4 M potassium chloride (KCl) and pH 4 buffer solution, with a few drops of pH buffer preservative (buffer preservative is optional).
You may keep the electrode in the storage solution indefinitely, but for long term storage, PASCO recommends storing the elec­trode dry. After dry storage, the electrode must be restored (see below) to rehydrate the glass membrane.
Restoring the Electrode
Follow this procedure to improve the response of a slow elec­trode or to rehydrate the membrane after dry storage.
1. Clean the membrane using one or more of these methods:
If the electrode is contaminated with proteins, soak the electrode in a solution of 1% pepsin in 0.1 Molar hydro­chloric acid (HCl).
Calibration
Calibration is not always necessary, especially if you are interest­ing in measuring a change in pH rather than absolute pH values. For accurate measurements, first calibrate the sensor. You will need two different buffer solutions (pH 4 and pH 7 in this exam­ple) and distilled water. The solutions should be at the same tem­perature as the solutions that you will measure later.
The calibration procedure varies depending on which software or datalogger you are using. For more specific instructions, see the documentation for your software or datalogger.
1. Open the calibration window or screen. (In DataStudio, click Setup; then click Calibrate Sensors.)
2. Select the “2 Point” calibration type.
3. Rinse the electrode with distilled water.
4. Place the electrode in pH 4 buffer solution. Wait for the read-
ing to stabilize.
5. Click or select Read for the first calibration point.
If the electrode is contaminated with inorganic deposits, rinse the electrode with 0.1 M ethylene dinitric tetra-acidic acid (EDTA) tetrasodium solution.
If the electrode is contaminated with oil or grease, wash the electrode in a mild detergent or solvent known to be effective for the particular film.
If the electrode is not responding quickly, soak the elec­trode alternately in 12 M NaOH (sodium hydroxide) and 1 M HCl (hydrogen chloride). Leave it in each solution for one minute. Rinse completely between soakings and end with HCl.
2. Soak the electrode in 0.1 M hydrochloric acid (HCl) for 15 minutes.
3. Soak the electrode in a pH 7 buffer solution for 30 minutes.
If the restoring procedure fails to improve the response of the electrode, replace the electrode.
Replacement and Alternative Electrodes
For a replacement electrode, purchase PASCO part 699-195.
6. Remove the electrode from the solution and rinse it with dis­tilled water.
7. Place the electrode in pH 7 buffer solution. Wait for the read­ing to stabilize.
8. Click or select Read for the second calibration point.
2
A flat-surface electrode, part PS-2182, is also available.
The pH Sensor is compatible with other combination pH electrodes that have BNC connectors.
Loading...
+ 1 hidden pages