PASCO PASPort PS-2117 Instruction Sheet

®
Drop Counter
Micro Stir
Bar
Syringe
Support Rod
Stopcock (2)
Drop Tip
PS-2117 Drop Counter
PS-2117
Instruction Sheet
012-08470D
Included Parts
Drop Counter Stopcock (2)
Micro Stir Bar Drop Tip
Syringe (60 mL) Support Rod
Additional Parts Required
Three-finger Clamp (SE-9445)
Support Rod and Stand (SE-9451)
Multi Clamp (ME-9507)
Magnetic Stir Plate (SE-77 00)
Beaker , 150 mL
Additional Parts Recommended
Buret, 50 mL (SE-7703)
Chemistry Sensor (PS-2170)
pH Sensor (PS-2102)
Temperature Sensor (PS-2125)
Other sensors for measuring conductivity or oxygen-reduction potential (ORP)
Micro Stir Bars, set of 5 (PS-2565)
Quick Start
1. Connect the Drop Counter to your PASPORT compatible PASCO interface.
2. If you are using a computer, connect the interface to it and start the data acquisition software.
3. Press or click the start button to begin recording data.
4. Allow drops of liquid to fall through the rectangular open-
ing.
Introduction
The Drop Counter measures the number of drops of fluid that fall through the rectangular opening of the sensor. It is typically used in conjunction with a pH sensor and other equipment to perform
®
Drop Counter PS-2117
150 mL
beaker with
100 mL of
0.005 M HCl
Stir Plate
pH
Probe
Drop
Counter
Figure 1: Titration set-up
Micro
Stir Bar
60 mL syringe
with 50 mL of
0.1 M NaOH
Stopcock
Stopcock
Drop
Counter
Drop Tip
Three-finger
Clamp
Multi Clamp
equivalence point
final drop count
a titration. Data from the Drop Counter and other sensors are recorded and displayed in software such as PASCO Capstone, or on a datalogging interface such as an Xplorer GLX.
The Drop Counter can be mounted on a support rod and the sup­port rod can be clamped to a rod stand. The counter is equipped with three integrated probe holders. The included Micro Stir Bar fits onto the end of a pH probe or probe of the same diameter.
Acid-base Titration
These steps explain a typical use of the Drop Counter.
Procedure
1. Set up the equipment and consumables as illustrated in Fig-
ure 1. Attach both stopcocks to the syringe, and attach the drop tip to the bottom stopcock. Line up the drop tip with the rectangular opening in the Drop Counter.
2. Start with the valves of the stopcocks turned parallel to the
table top. Add the solutions to the syringe and beaker.
3. Write down the initial volumes of both solutions (in the
syringe and in the beaker), and the molarity of the titrant (in the syringe).
4. Connect the Drop Counter and pH sensor to your PASPORT
compatible PASCO interface (or interfaces).
5. If you are using a computer, connect the interface to it and
start the data acquisition software.
6. Start the stir plate.
7. Click or press the start button to begin recording data.
8. Open the top stopcock. Slowly adjust the bottom stopcock to
start delivering titrant at about 2 drops per second.
9. Observe the data appearing on the pH versus drop count
graph (Figure 2). After the equivalence point is reached, wait until the pH curve flattens.
Figure 2: Graph of pH versus drop count
10. Close the stopcocks.
11. Stop data recording.
12. Read the final volume of fluid in the syringe and calculate
the net volume of titrant that was dispensed.
13. Read the final drop count on the graph.
14. In the calculator window of the software or datalogger, enter
a calculation as shown below, but using your own values of total volume dispensed and final drop count. The calculation gives the volume of titrant dispensed over time based on the number of drops and the average volume per drop.
volume = [Drop Count (drops)] * 100/3050
In this example, “100” is the total volume dispensed (in mL) and “3050” is the final drop count.
15. In the graph, change the horizontal axis to the calculated
“volume” (Figure 3).
2
®
Drop Counter PS-2117
Select volume calculation
for horizontal axis
detector
collimating
lens
focusing lens
sheet of light
dropinfrared light
source
Figure 4: Diagram of Drop Counter
When the Drop Counter detects a falling drop, the green indicator light emitting diode (LED) flashes. If something stationary is blocking the photodetector, the indicator LED remains on.
Troubleshooting
Problem Cause Solution
Figure 3: Graph of pH versus volume
Tips
Except for opening and closing the stopcocks, do not touch the syringe during the titration; otherwise, the drop size may change significantly .
The drop rate must remain approximately constant (about 2 drops per second) in order for the drop size to remain con­stant.
Drop size depends on the exact position of the bottom stop­cock. Therefore, the average drop size will be different for every titration. For each titration, re-enter the volume used and drop count values in the calculation.
Other Titrations
Titrations with different solutions can be performed using the same equipment and procedure. Other probes (such as conductiv­ity or ORP) can be used in place of the pH probe.
Indicator light flashes twice for a single drop. Drop Counter registers more drops than actually dispensed.
Indicator LED remains on constantly. Drop Counter does not register drops.
Fluid is splashing back into the drop window.
Lenses in drop window are dirty or wet.
Adjust the position of the beaker, Drop Counter, or syringe to eliminate splash back
Clean lenses with water and dry with a cotton swab or tissue.
Specifications
Maximum drop rate 40 drops per second Default sample rate 5 samples per second Maximum sample rate 10 samples per second Minimum object detection size 0.64 mm diameter Enclosure Polypropylene, water resistant,
Detector type Infrared photodetector with
chemically resistant to mild acids and bases
collimating lens
Technical Support
For assistance with any PASCO product, contact PASCO at:
Theory of Operation
The Drop Counter uses an infrared light source and a photodetector. Lenses spread the light into a “sheet” and refocus it onto the photodetector. When a drop of fluid passes through the Drop Counter, it partially blocks this sheet of light, and the photodetector registers a momentary decrease in light intensity.
The Drop Counter uses infrared light and ignores visible light. On power up, it automatically adjusts the light level for the best sensitivity. Use the Drop Counter away from direct sunlight or other sources of infrared that may interfere with it.
Address: PASCO scientific
10101 Foothills Blvd. Roseville, CA 95747-7100
Phone: +1 916 786 3800 (worldwide)
800-772-8700 (U.S.) Web: www.pasco.com Email: support@pasco.com
For more information about the Drop Counter and the latest revi­sion of this Instruction Sheet, visit the PASCO web site and enter PS-2117 in the Search window.
Limited Warranty For a description of the product warranty, see the PASCO catalog. Copyright The PASCO scientific 012-08470D Drop Counter Instruction Sheet is copyrighted with all rights reserved. Permis­sion is granted to non-profit educational institutions for reproduction of any part of this manual, providing the reproductions are used only in their labo­ratories and classrooms, and are not sold for profit. Reproduction under
3
®
Drop Counter PS-2117
any other circumstances, without the written consent of PASCO scientific, is prohibited. Trademarks PASCO, PASCO scientific, PASCO Cap­stone, PASPORT, and Xplorer GLX are trademarks or registered trade­marks of PASCO scientific, in the United States and/or in other countries. All other brands, products, or service names are or may be trademarks or service marks of, and are used to identify, products or services of, their respective owners. For more information visit www.pasco.com/legal.
Product End of Life Disposal Instructions:
This electronic product is subject to disposal and recycling regu­lations that vary by country and region. It is your responsibility to recycle your electronic equipment per your local environmental laws and regulations to ensure that it will be recycled in a manner that protects human health and the environment. To find out where you can drop off your waste equipment for recycling, please contact your local waste recycle/disposal service, or the place where you purchased the product.
The European Union WEEE (Waste Electronic and Electrical Equipment) symbol (to the right) and on the product or its packaging indicates that this product must not be disposed of in a standard waste container.
4
Loading...