PASCO OS-9287B User Manual

Instruction Sheet for the PASCO Model OS-9287B
mounting filters
To grounded
115 VAC, 60 Hz or
220V, 50 Hz outlet,
depending upon model.
Sodium Light Source
Rails for
CAUTION
RISK OF ELECTRIC SHOCK
DO NOT OPEN
TO AVOID ELECTRIC SHOCK THE POWER CORD PROTECTIVE GROUNDING CONDUCTOR MUST BE CONNECTED TO GROUND. REFER SERVICING T0 QUALIFIED PERSONNEL.
DISCONNECT INPUT POWER BEFORE REPLACING BULB.
LAMP CONTAINS ELEMENTAL SODIUM. CONSULT YOUR MANUAL for replacement and disposal of bulb.
IMPORTANT
Turn on lamp 20 minutes before use. Avoid turning lamp on and off for short periods.
Na
LIGHT
SOURCE
OS-9287
ON
OFF
ON/OFF Switch
012-05364C
DISCONNECT INPUT POWER BEFORE REPLACING FUSE. FOR CONTINUED FIRE PROTECTION REPLACE ONLY WITH 4.0A, 250 V, TYPE T FUSE, (PER IEC 127-2/III).
LINE VOLTAGE RANGE: 108 - 132 VAC POWER: 125 W MAX FREQ.: 47 - 63 HZ
115 VOLTS
FOR CONTINUED FIRE PROTECTION REPLACE ONLY WITH 2.0A, 250 V, TYPE T FUSE, (PER IEC 127-2/III).
LINE VOLTAGE RANGE: 198 - 242 VAC POWER: 125 W MAX FREQ.: 47 - 63 HZ
220 VOLTS
FOR CONTINUED FIRE PROTECTION REPLACE ONLY WITH 2.0A, 250 V, TYPE T FUSE, (PER IEC 127-2/III).
LINE VOLTAGE RANGE: 216 - 264 VAC POWER: 125 W MAX FREQ.: 47 - 63 HZ
240 VOLTS
Introduction
The PASCO scientific Model OS-9287B Low Pressure Sodium Light Source provides 6 candela/cm2, with better than 99.5% of the visible output in the 5889 and 5895 angstrom spectral lines. The light source comes with a built­in power supply so the unit can be powered from a standard 115 VAC, 60 Hz or 220 VAC, 50 Hz outlet, depending upon the model. Cooling fins and air vents on the sturdy aluminum case ensure cool, safe operation. The rails on the front and rear of the case allow easy mounting of standard 2­inch by 2-inch filters.
To operate the light source, simply plug it into a grounded 115 VAC outlet; then flip the switch on the front of the light source to On. Allow approximately 20 minutes warm-up time. (The first time you use your light source, you will need to install the sodium lamp. See the following page for initial setup instructions.)
Fuse (3.15 amp, slo-blo for OS-9287B)
~
~
~
(1.6 amp, slo-blo for OS-9287B-220)
The spectrum for the lamp is shown below. Better than
99.5% of the visible output is in the sodium doublet at 5889* and 5895* angstroms. Slight impurities (1% neon and argon) are added to the sodium gas to improve operating efficiency. Because of this, the lamp is not recommended for use with instruments having a resolving power in excess of 1 part in 104. The sodium pressure in the lamp is 5 microns Hg.
*
589.2 (the average of the doublet)
Note: Spectral lines (of very low intensity) also occur at 497.8,
515.4, and 616.0 nm.
Important Note—For maximum lamp life:
1. Always operate the light source in its upright position. Operation in other orientations will severely reduce the life of the lamp.
2. If you are going to use the light source more than once during the day, leave it on. Lamp wear results more from turning the light source on and off than from steady operation.
10101 Foothills Blvd. • P.O. Box 619011 • Roseville, CA 95678-9011 USA
Phone (916) 786-3800 • FAX (916) 786-8905 • TWX 910-383-2040
819
300 400 500 600 700 800
wavelength (nm)
© 1994 PASCO scientific This instruction sheet edited by: Dave Griffith
better
ways to
Sodium Light Source 012-05364C
Initial Setup
Before using the Light Source for the first time, you will need to install the lamp. To do this, just follow the numbered instructions shown below.
(1) Remove 4 screws
(shown shaded) on top and bottom panel, then remove the rear panel.
CAUTION
RISK OF ELECTRIC SHOCK
DO NOT OPEN
CAUTION: TO PREVENT THE RISK OF ELECTRIC SHOCK, DO NOT REMOVE BACK COVER. NO USER SERVICEABLE PARTS INSIDE. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL.
The lightning flash with arrowhead, within an equilateral triangle, is intended to alert the user of the presence of uninsulated “dangerous voltage” within the product’s enclosure that may be of sufficient magni­tude to constitute a risk of electric shock to persons.
CAUTION--HIGH VOLTAGE: Do not open the light
source with the unit plugged in.
(2) Align pins to fit
lamp in socket
(4) Reverse procedure
to remove lamp.
Maintenance
The light source lamp, when used correctly (see the note on the front panel of the unit), can be expected to provide 14,000+ hours of trouble free operation. However, if the light at any time fails to come on, first check the fuse on the front of the light source case. If the fuse is not blown, you may need to replace the lamp. In rare cases, the problem may be with the ballast. However, ballast replacement should be performed only by experienced personnel. If the fuse is intact and replacing the bulb does not restore opera­tion, we recommend you return the unit to PASCO scientific for repair.
Only the following replacement parts should be used:
Fuse (3.15 amp) 530-033
Fuse (1.6 amp) 530-034
(3) Push up and twist
to mate lamp with socket.
Item PASCO Part #
Lamp 526-034
The exclamation point within an equi­lateral triangle is intended to alert the user of the presence of important oper­ating and maintenance (servicing) in­structions in the literature accompany­ing the appliance.
110 volt Ballast 322-042 220 volt Ballast 322-05440 220 volt Ignitor 550-05441
2
012-05364C Sodium Light Source
Replacing the Sodium Lamp:
Caution—High Voltage: Do not open the light
source with the unit plugged in.
1. Unplug the light source, and allow several minutes for the sodium lamp to cool.
2. Remove the rear panel and replace the lamp as shown in the Initial Setup instructions above.
3. Replace the rear panel. Caution—Disposing of the Sodium Lamp: The
sodium lamp contains a small quantity of metallic sodium which produces heat when dampened. To dispose of a used lamp, wrap the lamp in several rolls of dry paper, then place it in a dry container. Then, in a dry atmosphere, break the lamp carefully into smaller pieces. Finally, stand back and fill the container with water. Within a few minutes the metallic sodium will be deactivated. Dispose of the lamp pieces as you would any broken glass.
Note: If additional maintenance is required, it
should be performed only by qualified personnel (see the schematic on the following page. )
Limited Warranty
PASCO scientific warrants this product to be free from defects in materials and workmanship for a period of one year from the date of shipment to the customer. PASCO will repair or replace, at its option, any part of the product which is deemed to be defective in material or workmanship. This warranty does not cover damage to the product caused by abuse or improper use. Determination of whether a product failure is the result of a manufacturing defect or improper use by the customer shall be made solely by PASC0 scien­tific. Responsibility for the return of equipment for warranty repair belongs to the customer. Equipment must be properly packed to prevent damage and shipped postage or freight prepaid. (Damage caused by improper packing of the equipment for return shipment will not be covered by the warranty.) Shipping costs for returning the equipment, after repair, will be paid by PASCO scientific.
Equipment Return
Should this product have to be returned to PASCO scien­tific, for whatever reason, notify PASCO scientific by letter or phone BEFORE returning the product. Upon notification, the return authorization and shipping instruc­tions will be promptly issued.
Other PASCO Spectral Equipment
PASCO scientific offers a variety of spectral light sources, and a precision student spectrometer for accurate spectral measurements in the student lab. For more information, check our catalog, or call toll-free at 1-800-772-8700.
Schematic
115V
115 VAC
SW 1
On/Off
3.15 amp
F 1
slo-blo
Ballast
transformer
L 1
NOTE: NO EQUIPMENT WILL BE
ACCEPTED FOR RETURN WITHOUT AN AUTHORIZATION.
220V
220 VAC
SW 1
On/Off
F 1
1.6 amp slo-blo
Ballast
transformer
Ignitor
L 1
COMMON
GROUND
COMMON
GROUND
3
Sodium Light Source 012-05364C
Notes:
4