Technical Support ......................................................................... Inside Back Cover
i
Optics Table012-06557A
Copyright, Warranty and Equipment Return
Please—Feel free to duplicate this manual
subject to the copyright restrictions below.
Copyright Notice
The PASCO scientific Model OS-8536 Optics Table
manual is copyrighted and all rights reserved. However,
permission is granted to non-profit educational institutions for reproduction of any part of this manual providing the reproductions are used only for their laboratories
and are not sold for profit. Reproduction under any other
circumstances, without the written consent of PASCO
scientific, is prohibited.
Limited Warranty
PASCO scientific warrants this product to be free from
defects in materials and workmanship for a period of one
year from the date of shipment to the customer. PASCO
will repair or replace, at its option, any part of the product
which is deemed to be defective in material or workmanship. This warranty does not cover damage to the product
caused by abuse or improper use. Determination of
whether a product failure is the result of a manufacturing
defect or improper use by the customer shall be made
solely by PASCO scientific. Responsibility for the return
of equipment for warranty repair belongs to the customer.
Equipment must be properly packed to prevent damage
and shipped postage or freight prepaid. (Damage caused
by improper packing of the equipment for return shipment will not be covered by the warranty.) Shipping
costs for returning the equipment, after repair, will be
paid by PASCO scientific.
Equipment Return
Should this product have to be returned to PASCO
scientific, for whatever reason, notify PASCO scientific
by letter or phone BEFORE returning the product. Upon
notification, the return authorization and shipping instructions will be promptly issued.
➤ NOTE:
NO EQUIPMENT WILL BE ACCEPTED FOR
RETURN WITHOUT AN AUTHORIZATION.
When returning equipment for repair, the units must be
packed properly. Carriers will not accept responsibility
for damage caused by improper packing. To be certain
the unit will not be damaged in shipment, observe the
following rules:
➀ The carton must be strong enough for the item
shipped.
➁ Make certain there is at least two inches of packing
material between any point on the apparatus and the
inside walls of the carton.
➂ Make certain that the packing material can not shift in
the box, or become compressed, thus letting the instrument come in contact with the edge of the box.
Address:PASCO scientific
10101 Foothills Blvd.
Credits
This manual authored by: Dave Griffith
P.O. Box 619011
Roseville, CA 95678-9011
Phone:(916) 786-3800
FAX:(916) 786-8905
ii
012-06557AOptics Table
Introduction
The PASCO OS-8536 Optics Table includes:
– cylindrical lens
– ray table
– ray table base and mounting hardware
The lens, ray table, and ray table base are designed to be
used with the Optics Bench, Ray Optics Kit, and Light
Source which are included in the OS-8515 Basic Optics
System.
Cylindrical Lens
The Cylindrical Lens is a “D”-shaped piece of clear
acrylic plastic. The lens is one inch (2.54 cm) thick, and
the radius of curvature is one inch (2.54 cm).
Ray Table
The Ray Table is a metal disk six inches (15.24 cm) in
diameter with a degree scale printed on both sides. In addition, the center of one side has a Cartesian grid marked
in millimeters (mm). The Ray Table has a hole in its center which fits over the post on the top of the Ray Table
Base.
Ray Table Base
The top side of the Ray Table Base has a ring of magnetic
material that holds the Ray Table in position when the
Ray Table is placed on the post on the top of the table.
The mounting hardware on each leg of the Ray Table
Base consists of a square nut and a thumbscrew. The
square nuts fit into the T-slot on each side of the Optics
Bench that is a part of the OS-8515 Basic Optics System.
Tightening the thumbscrews holds the Ray Table Base in
position when it is mounted on the Optics Bench.
Mounting the Ray Table Base t o the Optics
Bench
Loosen each thumbscrew by turning counter-clockwise.
Leave the square nut on the end of each thumbscrew. Attach the Ray Table Base to the Optics Bench by inserting
the square nuts into the T-slots on each side of the track.
The base can be moved to any position along the track
while the thumbscrews are loose. Tighten the thumbscrews to secure the base in position.
Cylindrical Lens
thumbscrew
Ray Table
Ray Table
Base
Figure 1: Components
Optics
Bench
T-Slot
Ray Table Base
square nut
Figure 2: Mounting Ray Table Base
1
Optics Table012-06557A
About the Experiments
Experiments 1, 2, and 6 use the OS-8536 (Cylindrical
Lens, Ray Table, and Ray Table Base), and the Light
Source and Optics Bench from the OS-8515 Basic Optics
System.
Experiment 3 through 9 are described in more detail in the Instruction Manual and Experiment Guide for the OS-8515
Basic Optics System.
The other experiments use the Ray Table and Ray Table
Base of the OS-8536 and the Light Source and Optics
Bench of the OS-8515 and the following additional
equipment from the OS-8515 Basic Optics System:
2
012-06557AOptics Table
Experiment 1:Reversibility
EQUIPMENT NEEDED
– Cylindrical Lens– Ray Table
– Ray Table Base– Optics Bench
– Light Source
Purpose
The purpose is to determine the relationship that exists
7
0
80
60
9
between the angle of incidence and the angle of refraction for light passing from air into a more optically dense
medium (the Cylindrical Lens). The second purpose is to
determine whether the same relationship holds between
the angles of incidence and refraction for light passing
out of a more optically dense medium back into air. That
is to say, if the light is traveling in the opposite direction
through the lens, is the law of refraction the same or
different? In this experiment, you will find the answer to
this question.
Angle of
Incidence
50
40
0
3
20
0
1
0
40
10
30
20
20
30
40
1
50
10
60
40
70
80
90
0
80
70
60
40
5
30
20
10
10
20
10
20
30
60
80
7
0
0
40
R
A
Y
T
A
m
B
L
m
30
E
S
C
A
L
E
2
0
10
30
40
0
10
20
30
40
50
Angle of
Refraction
1
Procedure
Mount the Ray Table Base and the Light Source on the
Optics Bench.
1. Put the Ray Table on the base with the Cartesian grid
2. Put the Light Source on its bracket so that the mul-
3. For the first trial, put the Cylindrical Lens
See Figure 1.2
(mm SCALE) facing up. Turn the Ray Table so the 0
(zero) degree line points to the Light Source.
tiple slits are facing the Ray Table. Position the Light
Source so it is about two centimeters from the edge of
the Ray Table. Adjust the slit mask on the front of the
Light Source so the Light Source projects
one ray of light across the middle of the
top surface of the Ray Table.
on the Ray Table so the flat surface of the
lens faces the Light Source and the edge
of the lens is on the 90 (ninety) degree
line with the lens exactly centered on the
0 (zero) degree line.
SLITS
1
3
5
Angle of
Incidence
Figure 1.1
R
COLO
Y
A
R
BEFORE CHANGING BULB.
DISCONNECT POWER
WITH FINGERS.
DO NOT TOUCH BULB
BULB: 12V 10W G-4
INSTRUCTION MANUAL.
PASCO
scientific
TO REPLACE BULB SEE
80
70
90
60
50
40
30
20
10
0
40
0
30
1
20
0
2
SOURCE
LIGHT
B
O
A
S
S
-
8
I
C
5
1
O
7
P
T
I
C
S
12V @ 800mA
GENTLY SPREAD BRACKET.
TO REMOVE LIGHT SOURCE
20
10
3
40
50
0
6
70
80
90
POWER SUPPLY JACK
80
70
60
40
30
20
10
10
10
20
30
40
70
0
8
20
6
0
RAY TABLE
mm SCALE
30
50
40
40
30
5
0
4
0
30
20
1
0
0
1
0
2
0
Angle of
Refraction
2
Figure 1.2
3
Loading...
+ 15 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.