PASCO OS-8494 User Manual

Instruction Sheet
1
2
3
*012-12360*
PASCO Optics
Adjustable Focal Length Lens
OS-8494
012-12360A
Included Equipment Quantity
1. Adjustable Focal Length Lens 2
2. Tubing, 5 mm O.D. 30 cm
3. 10 mL Syringe 1
Related Equipment
Human Eye Model (OS-8477)
Eye Model Bracket (OS-8469)
Adjustable Lens Holder* (OS-8474
Basic Optics Light Source* (OS-8470)
Basic Optics Bench*, 60 cm (OS-8541)
Basic Optics Viewing Screen* (OS-8460)
*Part of Basic Optics System, OS-8515A
See www.pasco.com for other PASCO Optics System equipment.
800-772-8700 www.pasco.com
Introduction
The Adjustable Focal Length Lens includes two lenses, a length of plastic tubing, and a 10 milliliter (mL) syringe. Each lens consists of a plastic housing with two flexible membranes held onto the housing with removable rings. Use the syringe to fill the lens with a fluid such as water and to increase or decrease the amount of fluid between the two flexible mem­branes. By adjusting the curvature of the membranes, you can adjust the focal length of the lens.
The Adjustable Focal Length Lens is designed to be used with the PASCO Human Eye Model (OS-8477) to demonstrate accomodation, the ability of the eye’s crystalline lens to change its focal length. It can also be used with components of the PASCO Basic Optics System (OS-8515A) such as the Light Source, Optics Bench, Viewing Screen, and Adjustable Lens Holder.
®
Assembly and Maintenance
Syringe
Tubing
Lens Housing
Connector
Figure 1
Membrane
Figure 2
Retainer ring
Membrane
Lens Housing
Cut a length of tubing about 15 cm (6 inch) long. Adjust the syringe so that the piston is beyond the midpoint of the cylinder. Attach the piece of plastic tubing to the 10 mL syringe and the connector on the edge of the lens housing (see Figure 1).
NOTE: Avoid touching the surface of the flexible membranes. If you need to clean the membrane’s surface, use a soft, lint-free cloth moistened with water.
To remove a membrane for cleaning, carefully remove the retainer ring that holds the membrane in place and lift the mem­brane off the lens housing holding just the edges of the mem­brane (see Figure 2).
Put the clean membrane on the edge of the lens housing so that some of the membrane material extends over the edge. Be sure to center the membrane on the lens housing. Carefully press the retainer ring over the membrane and onto the lens housing (see Figure 2).
Filling the Lens
Fill the lens housing with a fluid such as water (or corn syrup) before using it with the Human Eye Model or the Basic Optics components. To fill the lens housing with a fluid, use differences in pressure to allow the fluid to fill the lens. Start with the lens housing and syringe connected to the plastic turing as in Figure 1.
Disconnect the syringe from the plastic tubing.
To fill the syringe, push the piston all the way into the cylin­der. Put the end of the syringe into the fluid and slowly pull the piston outward so that the fluid is forced up into the cylinder. Stop when the fluid level is at the midpoint of the cylinder.
Re-attach the plastic tubing to the syringe. Hold the syringe vertically so the lens holder hangs down from the end of the tub­ing.
DO NOT force the liquid from the syringe into the lens holder. Instead, slowly pull the piston out so that air from inside the lens holder bubbles up through the liquid in the cylinder. The fluid should begin moving drop-by-drop into the lens.
When the piston is almost to the end of the cylinder, start pushing it back into the cylinder so that fluid moves from the cylin­der into the lens housing.
Repeat the process until the lens holder is filled with fluid and there are no air bubbles in the lens.
After the lens holder is filled, make sure that the tubing is also filled with fluid. Refill the syringe until it is about one-quarter full and reconnect it to the syringe.
2
Loading...
+ 2 hidden pages