Theory .................................................................................................................................................. 1
Technical Support ............................................................................................................................. 17
i
Copyright, Warranty , and Equipment Return
PleaseFeel free to duplicate this manual
subject to the copyright restrictions below.
Copyright Notice
The PASCO scientific 012-06616A manual is
copyrighted and all rights reserved. However,
permission is granted to non-profit educational
institutions for reproduction of any part of the Nitrate
Ion Selective Electrode manual providing the
reproductions are used only for their laboratories and
are not sold for profit. Reproduction under any other
circumstances, without the written consent of PASCO
scientific, is prohibited.
Limited Warranty
PASCO scientific warrants the product to be free from
defects in materials and workmanship for a period of
one year from the date of shipment to the customer.
PASCO will repair or replace at its option any part of
the product which is deemed to be defective in
material or workmanship. The warranty does not
cover damage to the product caused by abuse or
improper use. Determination of whether a product
failure is the result of a manufacturing defect or
improper use by the customer shall be made solely by
PASCO scientific. Responsibility for the return of
equipment for warranty repair belongs to the
customer. Equipment must be properly packed to
prevent damage and shipped postage or freight
prepaid. (Damage caused by improper packing of the
equipment for return shipment will not be covered by
the warranty.) Shipping costs for returning the
equipment after repair will be paid by PASCO
scientific.
Credits
Author: Peter Boyle
Editor: Steve Miller
Equipment Return
Should the product have to be returned to PASCO
scientific for any reason, notify PASCO scientific by
letter, phone, or fax BEFORE returning the product.
Upon notification, the return authorization and
shipping instructions will be promptly issued.
ä
NOTE:NO EQUIPMENT WILL BE
ACCEPTED FOR RETURN WITHOUT AN
AUTHORIZATION FROM PASCO.
When returning equipment for repair, the units must
be packed properly. Carriers will not accept
responsibility for damage caused by improper
packing. To be certain the unit will not be damaged in
shipment, observe the following rules:
➀ The packing carton must be strong enough for the
item shipped.
➁ Make certain there are at least two inches of packing
material between any point on the apparatus and the
inside walls of the carton.
➂ Make certain that the packing material cannot shift in
the box or become compressed, allowing the
instrument come in contact with the packing carton.
Address:PASCO scientific
10101 Foothills Blvd.
P.O. Box 619011
Roseville, CA 95678-9011
Phone:(916) 786-3800
FAX:(916) 786-3292
email:techsupp@pasco.com
web:www.pasco.com
ii
012–06616ANitrate Ion Selective Electrode
E=E
X
E
X
C
f
C
t
C
X
γ
C
f
Introduction
The PASCO scientific Nitrate Ion Selective Electrode is used to quickly, simply, accurately, and economically
measure nitrate in aqueous solutions.
Theory
The Nitrate Ion Selective Electrode consists of an electrode body containing a liquid internal filling solution in
contact with a gelled organophilic membrane containing a nitrate ion exchanger. When the membrane is in contact
with a solution containing free nitrate ions, an electrode potential develops across the membrane. This electrode
potential is measured against a constant reference potential, using an ISE Amplifier and ScienceWorkshop Interface.
The level of nitrate ions, corresponding to the measured potential, is described by the Nernst equation:
–Slog
0
where:
= measuredelectrode potential
= referencepotential(a constant)
E
0
mV
S = electrodeslope( ≈ 55
decade
)
= level ofnitrateions in solution
The activity, X, represents the effective concentration of the ions in solution. The total nitrate ion concentration,
C
is the sum of free nitrate ion, Cƒ, and complexed or bound perchlorate ion, Cb. The electrode is able to respond
t
to only the free ions, whose concentration is
=
=
b
Since nitrate ions form very few stable complexes, the free ion concentration may be equated to the total ion
concentration.
The activity is related to the free ion concentration, C
, by the activity coefficient, γ, by:
ƒ
=
Activity coefficients vary, depending on total ions strength, I, defined as:
I =
1
2
ΣCXZ
2
X
1
Nitrate Ion Selective Electrode012–06616A
C
X
where:
= concentrationofion
X
ZX= chargeof ion X
Σ = sum of all of the types of ionsin the solution
In the case of high and constant ionic strength relative to the sensed ion concentration, the activity coefficient, γ ,
is constant and the activity, X, is directly proportional to the concentration.
To adjust the background ionic strength to a high end constant value, ionic strength adjuster (ISA) is added to
samples and standards. The recommended ISA for nitrate is (NH
ionic strength adjusters as long as ions that they contain do not interfere with the electrode’s response to nitrate ions.
Strongly acidic (pH = 0 - 2) and strongly basic (pH = 12 - 14) solutions are particularly troublesome to measure.
The high mobility of hydrogen and hydroxide ions in samples make it impossible to mask their effect on the junction
potential with any concentration of equitransferent salt. One must either calibrate the electrode in the same pH range
as the sample or use a known incremental method for ion measurement.
. Solutions other than this may be used as
4) 2S04
2
012–06616ANitrate Ion Selective Electrode
Equipment
• Nitrate Ion Selective Electrode
• Nitrate Ion Selective Electrode fill solution
• Pipette for fill solution
to ISE
Amplifier
filling pipette
Figure 1.
Included Equipment
Sodium Ion Selective
Electrode
R
O
E
O
L
1
P
0
1
M
3
A
S
R
n
e
f
e
tio
r
e
u
n
l
c
o
e
S
Fi
l
l
4
I
M
C
K
filling
solution
Additional Required:
Required Equipment
PASCO CI-6738 ISE (Ion Selective Electrode)
Amplifier
• ScienceWorkshop 2.2.5 or higher
• PASCO ScienceWorkshop Computer
Interface
• Semi-logarithmic 4-cycle graph paper
for preparing calibration curves.
• Magnetic stir plate
• Lab-ware made of plastic, not glass, for
all low level measurements
Required Solutions
The stock solutions listed in this section may be
created as described in the text or ordered directly
from PASCO. The solutions available for order,
and their respective prices are listed on the ‘ISE
Working Solution Price List’.
• Deionized or distilled water for solution and standard preparation
• Ionic Strength Adjuster (ISA) , 2 M (NH
4) 2SO4
To prepare this solution, half fill a 1,000 ml volumetric flask with distilled water and add 264 grams of
reagent-grade ammonium sulfate, (NH
. Swirl the flask gently to dissolve the solid. Fill the flask to
4) 2S04
the mark with distilled water, cap, and upend several times to mix the contents. ISA is added at the rate of
2 ml of ISA to each 100 ml of standard or sample to adjust the ionic strength to about 0.12 M.
• Reference Filling Solution, 0.l M (NH
• Nitrate Standard, 0.1 M NaNO
3
4) 2SO4
/0.0l M KCl
To prepare this solution, add 8.50 grams of reagent-grade sodium nitrate to a one liter volumetric flask about
half full of distilled water. Swirl the flask gently to dissolve the solid. Fill to the mark with distilled water,
cap, and upend several times to mix the solution.
• Nitrate Standard, 1,000 ppm NO
-1
3
To prepare this solution, add 1.37 grams of reagent-grade sodium nitrate to a one liter volumetric flask about
half full with distilled water. Swirl the flask gently to dissolve the solid. Fill to the mark with distilled water,
cap, and upend several times to mix the solution.
• Nitrate Standard, 100 ppm NO
-1
3
To prepare this solution, add 0.61 grams of reagent-grade sodium nitrate to a one liter volumetric flask about
half full with distilled water. Swirl the flask gently to dissolve the solid. Fill to the mark with distilled water,
cap, and upend several times to mix the solution.
3
Nitrate Ion Selective Electrode012–06616A
General Preparation
fill hole
Electrode Preparation
1. Remove the rubber cap covering the electrode tip. Slide
the rubber sleeve down away from the filling hole of the
Nitrate Ion Selective Electrode. Fill the electrode with
the included filling solution to a level just below the fill
hole. Slide the rubber sleeve back over the filling hole
(Figure 2a). Gently shake the electrode downward in the
same manner as a clinical thermometer to remove any air
bubbles that may be trapped behind the nitrate membrane.
Prior to first usage, or after long term storage, immerse
the nitrate membrane in nitrate standard for thirty minutes.
The electrode is now ready for use.
2. Connect the Nitrate Ion Selective Electrode to the ISE
Amplifier and insert the DIN connector of the ISE
Amplifier into analog port A or B on a PASCO Computer
Interface (Figures 2b and 2c).
a
b
rubber
sleeve
rubber cap
Insert DIN
connector into
analog channel
A or B.
Electrode Slope Check Using Science Workshop
(check electrodes each day)
1. To a 150 ml glass beaker, add 100 ml of distilled water
and 2 ml of ISA. Place the beaker on a magnetic stirrer
and begin stirring at a constant rate. Start the
ScienceWorkshop software, select the Ion Selective
Electrode sensor, open a Digital display, change the
number of digits to the right of the decimal from 1 to 3,
and begin monitoring data. Lower the electrode tip into
the solution.
2. Using a pipette, add 1 ml of 0.l M or 1,000 ppm nitrate
standard to the beaker. When the reading has stabilized,
record the voltage reading indicated in the Digits display.
3. Using a pipette, add 10 ml of the same nitrate standard used above to the beaker. When the reading has
stabilized, record the voltage reading indicated in the Digits display.
4. Determine the difference between the two readings. A difference of 55 ±2 mV indicates correct electrode
operation assuming the solution temperature is between 20 °C and 25 °C. See the Troubleshooting
sections if the potential change is not within this range.
Rotate ring
c
one-quarter
turn to secure
electrode
connector
Figure 2
Equipment Setup. a: filling the electrode
with filling solution; b & c: connecting the
electrode to the ISE Amplifier and to the
computer interface
ISE Amplifier
ä
Note: Slope is defined as the change in potential observed when the concentration
changes by a factor of 10.
4
012–06616ANitrate Ion Selective Electrode
Measurement
Measuring Hints
• All samples and standards should be at the same temperature for precise measurement. A difference of 1 °C
in temperature will result in a 2% measurement error.
• The sensing membrane is normally subject to water uptake and might appear milky. This has no effect on
performance.
• Constant, but not violent, stirring is necessary for accurate measurement. Magnetic stirrers can generate
sufficient heat to change the solution temperature. To counteract this effect, place a piece of insulating
material, such as Styrofoam sheet, between the stirrer and beaker.
• Always rinse the electrode with distilled water and blot dry between measurements. Use a clean, dry tissue
to prevent cross-contamination.
• For samples with high ionic strength, prepare standards whose composition is similar to the sample.
• Always check to see that the membrane is free from air bubbles after immersion into standard or sample.
• A slow responding electrode may be caused by interferences to the electrode. To restore proper
performance, soak the electrode in distilled water for about 5 minutes to clean the membrane, rinse, and soak
in standard solution for about 5 minutes.
Sample Requirements
• All samples must be aqueous and not contain organics which can dissolve in the membrane or extract out
the liquid ion exchanger.
• The temperature of the standard solutions and of the sample solutions should be the same and below 40 °C.
About a 2% error will be introduced for a l °C difference in temperature.
• Interference’s should be absent. If they are present, use the procedures found in the Interference’s section
to remove them.
• The pH range for the nitrate ion electrode is 2.5 - 11. Neutralize samples outside this range with acid or base
to bring them in range.
Units of Measurement
Nitrate concentrations are measured in units of ppm as sodium nitrate, ppm as nitrate, moles per liter, or any other
convenient concentration unit. Table 1 indicates some of the concentration units.
T ABLE 1. Concentration Unit Conversion Factors
ppm NaNO
3
ppm NO
850.00620.001.0 X 10
85.062.01.0 X 10
8.56.21.0 X 10
-1
3
moles/liter
-2
-3
-4
5
Nitrate Ion Selective Electrode012–06616A
Measurement Procedure
Direct Measurement
Direct measurement is a simple procedure for measuring a large number of samples. A single reading is all that is
required for each sample. The ionic strength of samples and standards should be made the same by adjustment with
ISA for all nitrate solutions. The temperature of both sample solution and standard solutions should be the same.
Direct Measurement
ä
Note: A calibration curve is constructed on semilogarithmic paper. The measured
electrode potential (linear axis) is plotted against the standard concentration (log axis). In
the
linear region of the curve, only two standards are necessary to determine a calibration
curve. Calibration solutions close to the anticipated value of the “unknown” should be
chosen. In the
measurement procedures given are for the linear portion of the curve. The nonlinear
portion of the curve requires the use of low level procedures.
1. By serial dilution, prepare three standard solutions from the 0.l M or 1,000 ppm stock standard. The
resultant concentrations should be l0
2 ml of ISA to each 100 ml of standard. Prepare standards with a composition similar to the samples if
the samples have an ionic strength above 0.l M
2. Place the most dilute solution (l0
stirring at a constant rate. After assuring that ScienceWorkshop is operating, lower the electrode tip into
the solution. After the reading has stabilized, record the Voltage reading in the Digits Display.
3. Place the mid-range solution (l0
stirring. After rinsing the electrode with distilled water, blot dry and immerse the electrode tip in the
solution. When the reading has stabilized, record the Voltage reading in the Digits Display.
4. Place the most concentrated solution (l0
begin stirring. After rinsing the electrode in distilled water, blot dry, and immerse the electrode tip in the
solution. When the reading has stabilized, record the Voltage reading in the Digits Display.
nonlinear region, additional points must be measured. The direct
-2
, l0-3, and 10-4 M or 1,000, 100 and 10 ppm nitrate standards. Add
-4
M or 10 ppm) in a 150 ml beaker on the magnetic stirrer and begin
-3
N or 100 ppm) in a 150 ml beaker on the magnetic stirrer and begin
-2
M or 1,000 ppm) in a 150ml beaker on the magnetic stirrer and
5. Using the semi-logarithmic graph paper, plot the Voltage reading in the Digits Display (linear axis)
against the concentration (log axis). Extrapolate the calibration curve down to about l.0X10
-1
NO
). A typical calibration curve can be found in Figure 3.
3
-5
6
M (1.0 ppm
012–06616ANitrate Ion Selective Electrode
+30
+70
+110
electrode
potential
+150
(mV)
+190
+230
Figure 3
Typical nitrate electrode calibration curve
ä
Note: A calibration curve is constructed on semi-logarithmic paper The measured
electrode potential in (linear axis) is plotted against the standard concentration (log axis).
In the linear region of the curve, only three standards are necessary to determine a
calibration curve. In the non-linear region, additional points must be measured. The direct
measurement procedures given are for the linear portion of the curve. The non-linear
portion of the curve requires the use of low level procedures.
10
10-fold change
~56mV
(ppm as N)
0.11101001,000
(ppm as NO3)
0.1
-6
1101001,000
-5
10
-1
NO
concentration (M)
3
10
- 4
10
-3
10
-2
10
-1
6. To a clean, dry 150 ml beaker, add 100 ml of the sample and 2 ml of ISA. Place the beaker on the magnetic
stirrer and begin stirring at a constant rate. Rinse the electrode tip with distilled water, blot dry, and lower
the electrode tip into the solution. When the reading has stabilized, record the Voltage reading in the Digits
Display. Using the calibration curve, determine the sample concentration.
7. The calibration should be checked every two hours. Assuming no change in ambient temperature, place
the electrode tip in the mid-range standard. After the reading has stabilized, compare it to the original
reading recorded in step 3 above. A reading differing by more than 0.5mV or a change in the ambient
temperature will necessitate the repetition of steps 2 - 6 above. A new calibration curve should be
prepared daily.
Low Level Nitrate Measurements
This procedure is recommended for solutions with ionic strengths less than l.0Xl0-4M. If the solution is high in ionic
strength, but low in nitrate, use the same procedure, but prepare a calibration solution with a composition similar
to the sample.
1. Using 20 ml of standard ISA, dilute to 100 ml with distilled water. This low level ISA [0.4 M (NH
is added at the rate of 1 ml low level ISA to each 100 ml of sample. The background ionic strength will
be 4.0X l0
-3
M.
4) 2SO4
]
7
Nitrate Ion Selective Electrode012–06616A
2. Dilute 1 ml of 0.l M standard to 100 ml to prepare a l.0Xl0-3 M NO
per liter. Use the 1,000 ppm standard for preparing a 100 ppm NO
-1
solution for measurements in moles
3
-1
standard by diluting 10 ml of the
3
1,000 ppm standard to 100 ml. Standards should be prepared fresh daily.
3. Add 1 ml of the low level ISA to a 100 ml volumetric flask and fill to the mark with distilled water. Pour
this solution into a 150 ml beaker and place the beaker on the magnetic stirrer. Begin stirring at a constant
rate.
4. Place the electrode tip in the solution. Assure that ScienceWorkshop is operating.
5. Add increments of the l.0Xl0
-3
M or 100 ppm standards as given in Table 2 below.
6. After the reading has stabilized, record the Voltage reading in the Digits Display after each addition.
T ABLE 2: Step-wise Calibration for Low Level Nitrate Measurements
AddedConcentration
StepPipetteVolume (mL)M NO
1A0.11.0 X 10
2A0.12.0 X 10
3A0.24.0 X 10
4A0.26.0 X 10
5A0.49.9 X 10
6B2.02.9 X 10
7B2.04.8 X 10
-1
3
-6
-6
-6
-6
-6
-5
-5
ppm NO
-1
3
0.1
0.2
0.4
0.6
1.0
2.9
4.8
Pipette A = 1 ml graduated pipette
Pipette 8 = 2 ml pipette
Solutions: additions of l.0X10
-3
M or 100 ppm standard to 100 ml of ISA as prepared in step 3 above.
7. On semi-logarithmic graph paper, plot the voltage reading on the Digits display (linear axis) against the
concentration (log axis) as in Figure 3.
8. Rinse the electrode and blot dry.
9. Measure out 100 ml of the sample into a 150 ml beaker, add 1 ml of low level ISA. Place the beaker on
the magnetic stirrer and begin stirring. Lower the electrode tip into the solution. After the reading has
stabilized, record the Voltage reading indicated in the Digits Display and determine the concentration
from the low level calibration curve.
10.Prepare a new low level calibration curve daily. Check the calibration curve every two hours by repeating
steps 2 - 7.
8
012–06616ANitrate Ion Selective Electrode
Electrode Characteristics
Reproducibility
Electrode measurements reproducible to +2% can be obtained if the electrode is calibrated every hour. Factors such
as temperature fluctuations, drift, and noise limit reproducibility. Reproducibility is independent of concentration
within the electrode’s operating range.
Interferences
Certain anions are electrode interferences and will cause electrode malfunction, drift or measurement errors if
present in high enough levels. The level of interfering common anions that will cause a 10% error at three levels
of nitrate is given in Table 3.
T ABLE 3:Concentration of Possible Interference’s Causing a 10% Error at Various
Levels of Nitrate; Background Ionic Strength of 0.12 M (NH
4) 2SO4
.
Interference
moles/liter10
Cl
NO
Br
CN
ClO
I
ClO
-1
-1
2
-1
-1
-1
3
-1
-1
4
3.0X10
7.0X10
7.0X10
1.0X10
5.0X10
5.0X10
1.0X10
-2
M10
-1
-3
-3
-3
-4
-5
-6
-3
M10
3.0X10
7.0X10
7.0X10
1.0X10
5.0X10
5.0X10
1.0X10
-2
-4
-4
-4
-5
-6
-7
3.0X10
7.0X10
7.0X10
1.0X10
5.0X10
5.0X10
1.0X10
-4
M
Interference
moles/liter100 ppm N10 ppm N1 ppm N
-1
Cl
NO
Br
CN
ClO
I
ClO
-1
2
-1
-1
-1
3
-1
-1
4
7,60076076.0
23023.02.30
40040.04.00
20.02.000.20
30.03.000.30
4.000.400.04
0.070.0070.0007
-3
-5
-5
-5
-6
-7
-8
Interferences such as chloride, bromide, iodide and cyanide can be removed by precipitation with 0.5 grams of silver
sulfate added to 100 ml of sample. Nitrite interference can be removed by adding 0.3 grams of sulfamic acid to 100
ml of sample. Carbonate and bicarbonate, which are weak interferences, can be removed by acidifying the sample
to pH 4.5 with sulfuric acid. Organic (carboxylic) anions hinder the nitrate electrode response and can be removed
by adding 10 grams of aluminum sulfate to 100 ml of sample.
9
Nitrate Ion Selective Electrode012–06616A
The above interference removal procedures require similar treatment of standards as well as samples.
If the electrode is exposed to high levels of interfering ions which cannot be removed, the electrode reading may
drift and the response may become sluggish. Restore performance by soaking in distilled water for 30 minutes
followed by soaking in nitrate standard for 30 minutes.
Temperature Influences
Samples and standards should be at the same temperature, since electrode potentials are influenced by changes in
temperature. A l °C difference in temperature results in a 2% error at the l0
equilibria on which the electrode depends, the absolute potential of the reference electrode changes slowly with
temperature. The slope of the nitrate electrode, as indicated by the factor “S’ in the Nernst equation, also varies with
temperature. Table 4 gives values for the “S” factor in the Nernst equation for the nitrate ion.
The operating range of the nitrate ion electrode is 0 °C - 40 °C, provided that temperature equilibrium has occurred.
If the temperature varies substantially from room temperature, equilibrium times up to one hour are recommended.
-3
M level. Because of the solubility
T ABLE 4: T emperature vs. Theoretical V alues for the Electrode Slope
Plotting the electrode potential against the nitrate concentration on semi-logarithmic paper results in a straight line
with a slope of about 55 mV/decade. (Refer to Figure 3.)
The time needed to reach 99% of the stable electrode potential reading, the electrode response time, varies from
one minute or less in highly concentrated solutions to several minutes near the detection limit. (Refer to Figure 4.)
10
012–06616ANitrate Ion Selective Electrode
+50
+75
10-3M to 10-2M NaNO
+100
3
electrode
potential
(mV)
+125
+150
+175
+200
+225
10-3M to 10-6M NaNO
1
10-3M to 10-4M NaNO
10-3M to 10-5M NaNO
3
234
3
3
time (minutes)
Figure 4
Typical electrode time response to step changes in NaNO3
Limits of Detection
The upper limit of detection in pure sodium nitrate solutions is 1 M. In the presence of other ions, the upper limit
of detection is above l0
potential developing at the reference electrode and the malt extraction effect influence this upper limit. Some salts
may infuse into the electrode membrane at high salt concentrations, causing deviation from the theoretical response.
Either dilute samples between 1 M and l0
-1
M nitrate, but two factors influence this upper limit. Both the possibility of a liquid junction
-1
M or calibrate the electrode at 4 or 5 intermediate points.
The lower limit of detection is influenced by the slight water solubility of the ion exchanger used in the sensing
portion of the electrode. Refer to Figure 3 for a comparison of the theoretical response to the actual response at low
levels of nitrate. Nitrate measurements below l0
-5
M NO
-1
(0.6 ppm as N0
3
-1
) should employ low level procedures.
3
pH Effects
The operating range of the nitrate electrode is from pH 2.5 to pH 11.
Electrode Life
The nitrate electrode will last six months in normal laboratory use. On-line measurement might shorten operational
lifetime to several months. In time, the response time will increase and the calibration slope will decrease to the point
calibration is difficult and electrode replacement is required.
11
Nitrate Ion Selective Electrode012–06616A
Maintenance
Electrode Storage
The nitrate electrode may be stored for short periods of time in l0-2 M nitrate solution. For longer storage (longer
than two weeks), rinse and dry the nitrate membrane and cover the tip with any protective cap shipped with the
electrode. The reference portion of the combination electrode (or the outer chamber of the reference electrode)
should be drained of filling solution, if refillable and the rubber insert placed over the filling hole.
Specifications
Concentration Range:1 M to 7X10-6 M (6.2X104 to 0.5 ppm as NO
pH Range:2.5 to 11
-1
)
3
Temperature Range:0
Resistance:100M ohms
Reproducibility:±2%
Samples:aqueous solutions only, no organic solvents
Storage:store in dilute nitrate solution
Size:110 mm length
°C
– 40 °C
12 mm diameter
cable length = 1 m
Troubleshooting Guide
The goal of troubleshooting is the isolation of a problem through checking each of the system components in turn:
the meter, the plastic-ware, the electrode, the standards & reagents, the sample, and the technique.
Glassware/Plastic-ware
Clean glassware is essential for good measurement. Be sure to wash the glassware/plastic-ware well with a mild
detergent and rinse very well with distilled or deionized water.
Electrode
The electrode may be checked by using the procedure found in the sections entitled Electrode Slope Check.
1. Be sure to use distilled or deionized water when following the procedures given in Electrode SlopeCheck.
2. If the electrode fails to respond as expected, see the section Measuring Hints. Repeat the slope check.
3. If the electrode still fails to respond as expected, substitute another Nitrate Ion Selective Electrode (if
available) that is known to be in good working order for the questionable electrode. If the problem persists
and you are using an electrode pair, try the same routine with a working reference electrode.
12
012–06616ANitrate Ion Selective Electrode
4. If the problem persists, the standards or reagent may be of poor quality, interferences in the sample may
be present or the technique may be faulty. (See StandardsReagents, Sample, and Technique sections
below.)
5. If another electrode is not available for test purposes, or if the electrode in use is suspect, review the
instruction manual and be sure to:
- Clean and rinse the electrode thoroughly.
- Prepare the electrode properly.
- Use the proper filling solution.
- Adjust the pH and the ionic strength of the solution by the use of the proper ISA.
- Measure correctly and accurately.
- Review Troubleshooting Hints.
Standards & Reagents
Whenever problems arise with the measuring procedure that has been used successfully in the past, be sure to check
the standard and reagent solutions. If in doubt about the credibility of any of the solutions, prepare them again. Errors
may result from contamination of the ISA, incorrect dilution of standards, poor quality distilled/deionized water,
or a simple mathematical miscalculation.
Sample
Look for possible interferences, complexing agents, or substances which could affect the response or physically
damage the sensing electrode if the electrode works perfectly in the standard, but not in the sample.
Try to determine the composition of the samples prior to testing to eliminate a problem before it starts. (See
Measuring Hints, Sample Requirements, and Interferences.)
Technique
Be sure that the electrode’s limit of detection has not been exceeded. Be sure that the analysis method is clearly
understood and is compatible with the sample. Refer to the instruction manual again. Reread General Preparation
and Electrode Characteristics.
If trouble still persists, call PASCO scientific Technical Support.
13
Nitrate Ion Selective Electrode012–06616A
Troubleshooting Hints
SymptomPossible CausesNext Step
Out of Range Readingdefective electrodecheck electrode operation
electrodes not plugged inunplug electrodes and reseat electrodes
properly
reference electrode not filledbe sure reference electrode is filled
air bubble on membraneremove bubble by re-dipping electrode
electrode not in solutionput electrodes in solution
Noisy or Unstableelectrode exposed to interferencessoak electrode in nitrate standard
Readings (readings
continuously or rapidlydefective electrodereplace electrode
changing)
ISA not useduse recommended ISA
stirrer not groundedground stirrer
air bubble on membraneremove bubble by re-dipping electrode
outer filling solution level too lowfill electrode to level just below the fill hole
Drift (reading slowlysamples and standards at differentallow solutions to come to room temperature
changing in one direction)temperaturesbefore measurement
electrode exposed to interferencessoak electrode in nitrate standard
incorrect reference filling solutionuse recommended filling solution
14
012–06616ANitrate Ion Selective Electrode
SymptomPossible CausesNext Step
Low Slope or No Slopestandards contaminated orprepare fresh standards
incorrectly made
ISA not useduse recommended ISA
standard used as ISAuse ISA
electrode exposed to interferencessoak electrode in nitrate standard
defective electrodecheck electrode operation
air bubble on membraneremove bubble by re-dipping probe
“Incorrect Answer”incorrect scaling of semilog paperplot voltage potential on the linear axis.
(but calibration curveOn the log axis, be sure concentration
is good)numbers within each decade are increasing
with increasing concentration.
incorrect signbe sure to note sign of millivolt reading
sample carryoverrinse electrode thoroughly between samples
15
Nitrate Ion Selective Electrode012–06616A
16
T echnical Support
Feedback
If you have any comments about the product or
manual, please let us know. If you have any
suggestions on alternate experiments or find a problem
in the manual, please tell us. PASCO appreciates any
customer feedback. Your input helps us evaluate and
improve our product.
To Reach PASCO
For technical support, call us at 1-800-772-8700
(toll-free within the U.S.) or (916) 786-3800.
fax:(916) 786-3292
e-mail:techsupp@pasco.com
web:www.pasco.com
Contacting Technical Support
Before you call the PASCO Technical Support staff,
it would be helpful to prepare the following
information:
➤ If your problem is with the PASCO apparatus,
note:
- Title and model number (usually listed on the
label);
- Approximate age of apparatus;
- A detailed description of the problem/
sequence of events (in case you can’t call
PASCO right away, you won’t lose valuable
data);
- If possible, have the apparatus within reach
when calling to facilitate description of
individual parts.
➤ If your problem relates to the instruction manual,
note:
- Part number and revision (listed by month and
year on the front cover);
- Have the manual at hand to discuss your
questions.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.