Table 1: Conductivity of Various Water Samples at 25°C ...........................................23
Table 2: Table of Conductivity vs. Concentration for Common Solutions ................... 24
Table 3: Table of Conductivity vs. Concentration for Common Solutions ................... 25
Table 4: Conversion Chart to Estimate TDS of Aqueous Solutions .............................26
Table 5: Table of High Accuracy Reference Solution for Calibration ..........................26
Table 6: Sample Illustrating the Application of Conductivity to Agriculture ................27
Technical Support ..................................................................................... Back Cover
i
Conductivity Sensor012–06485B
Copyright, Warranty, and Equipment Return
Please—Feel free to duplicate this manual
subject to the copyright restrictions below.
Copyright Notice
The PASCO scientific Conductivity Sensor manual
(012-06485A) is copyrighted and all rights reserved.
However, permission is granted to nonprofit
educational institutions for reproduction of any part
of the Conductivity Sensor manual providing the
reproductions are used only for their laboratories
and are not sold for profit. Reproduction under any
other circumstances, without the written consent of
PASCO scientific, is prohibited.
Limited Warranty
PASCO scientific warrants the product to be free from
defects in materials and workmanship for a period of
one year from the date of shipment to the customer.
PASCO will repair or replace at its option any part
of the product which is deemed to be defective in
material or workmanship. The warranty does not
cover damage to the product caused by abuse or
improper use. Determination of whether a product
failure is the result of a manufacturing defect or
improper use by the customer shall be made solely by
PASCO scientific. Responsibility for the return of
equipment for warranty repair belongs to the
customer. Equipment must be properly packed to
prevent damage and shipped postage or freight
prepaid. (Damage caused by improper packing of the
equipment for return shipment will not be covered by
the warranty.) Shipping costs for returning the equipment after repair will be paid by PASCO scientific.
Equipment Return
Should the product have to be returned to PASCO
scientific for any reason, notify PASCO scientific by
letter, phone, or fax BEFORE returning the product.
Upon notification, the return authorization and
shipping instructions will be promptly issued.
➤➤
➤ NOTE: NO EQUIPMENT WILL BE
➤➤
ACCEPTED FOR RETURN WITHOUT AN
AUTHORIZATION FROM PASCO.
When returning equipment for repair, the units must
be packed properly. Carriers will not accept responsibility for damage caused by improper packing. To
be certain the unit will not be damaged in shipment,
observe the following rules:
➀ The packing carton must be strong enough for the
item shipped.
➁ Make certain there are at least two inches of packing
material between any point on the apparatus and the
inside walls of the carton.
➂ Make certain that the packing material cannot shift in
the box or become compressed, allowing the
instrument come in contact with the packing carton.
Address:PASCO scientific
10101 Foothills Blvd.
P.O. Box 619011
Roseville, CA 95678-9011
Credits
Authors: Dominic Calabrese and Steve Miller
Editor:Sunny Bishop
ii
Phone:(916) 786-3800
FAX:(916) 786-3292
email:techsupp@pasco.com
web:www.pasco.com
012–06485BConductivity Sensor
CONDUCTIVITY
SENSOR
CI-6729
CI-6739A
2K
20K
2
00
RANGE
SELEC
T
699-06620
C
O
N
D
U
C
TIV
ITY
699-06621
C
O
N
D
U
C
TIV
ITY
10X
Introduction
The PASCO CI-6729 and CI-6739A Conductivity Sensors are
used with a PASCO Science Workshop
®
computer interface to
investigate factors that influence the electrical conductivity of
liquids. The Conductivity Sensor consists of a signal conditioning
amplifier and either a 1X or 10X conductivity electrode. The
signal conditioning amplifier box is capable of working with
either electrode. Students can use this sensor to explore the
effects of temperature and concentration on the electrical
transport properties of aqueous solutions, especially in
applications relating to ecological systems.
The Conductivity Sensor is designed for use in aqueous solutions
at temperatures ranging from 0 °C to 80 °C. The 1X conductivity
electrode provided with the CI-6729 Conductivity Sensor has a
range of up to 20,000 microsiemens per centimeter. The 10X
conductivity electrode, provided with the CI-6739A, has a range
up to 200,000 microsiemens per centimeter. In order to achieve
maximum performance of the Conductivity Sensor, soak the
electrode in distilled water for 5 to 10 minutes before use to assure
complete wetting of the electrodes. The electrode may also be
calibrated at approximately the same temperature as the test solution
for concentration or total dissolved solids (TDS) measurements.
➤ Conductivity Sensors are used
in many practical applications
including:
• Environmental Monitoring
• Oceanography-Salinity
• Agriculture
• Waste Systems
• Beverage Industries
• Electroplating
interface cable with 8-pin
DIN connectors
connector for
amplifier box
Equipment
connector for
computer
interface
Provided With CI-6729 (1X)
• 1X conductivity electrode with 3-foot BNC cable
• Conductivity Sensor Amplifier box with 8-pin DIN
connector and BNC connector
• interface cable with DIN connector
amplifier box
Provided With CI-6739A (10X)
• 10X conductivity electrode with 3-foot BNC cable
• Conductivity Sensor Amplifier box with 8-pin DIN
connector and BNC connector
• interface cable with DIN connector
Additional equipment required
• Any PASCO Science Workshop® computer interface
• computer
Replacement Electrodes
• 1X conductivity electrode: 699-06620
• 10X conductivity electrode: 699-06621
conductivity
electrodes
1
Conductivity Sensor012–06485B
Maintenance
Cleaning
To ensure accurate and reproducible results, the electrode cell must be clean.
A dirty cell will contaminate the solution and change the conductivity of the
liquid.
Cells can be cleaned with detergent and/or dilute nitric acid (1%) by dipping
or filling the cell with cleaning solution and stirring for three minutes.
Storage
Although the best method to store the electrode is by immersing it in deionized
water, the electrode can be stored dry in its container. If the cell is stored dry,
it should be soaked in distilled water for 5 to 10 minutes before use.
Theory—Principles of Operation
of the Conductivity Sensor
What is Conductivity?
Conductivity (or specifically, electrolytic conductivity) is defined as the ability
of a substance to conduct electric current. It is the reciprocal of the more
commonly encountered term, resistivity. All substances possess conductivity
to some degree, but the amount varies widely, ranging from extremely low
(insulators such as benzene and glass) to very high (silver, copper, and metals
in general). Most industrial interest is in the conductivity measurement of
liquids that generally consist of ionic compounds dissolved in water. These
solutions have conductivities approximately midway between insulators and
metallic conductors. This conductivity can be measured quite easily by
electronic means, allowing a simple test that can tell much about the quality
of the water or the makeup of the solution. A broad line of conductivity
equipment is available to measure liquids ranging from ultrapure water (low
conductivity) to concentrated chemical streams (high conductivity).
Units of Conductivity
The units of measurement used to describe conductivity and resistively are
quite fundamental and are frequently misused. Once the units are known,
various waters can be quantitatively described.
The basic unit of resistance is the familiar ohm. Conductance is the
reciprocal of resistance, and its basic unit is the siemens (S), formerly called
mho. In discussions of bulk material, it is convenient to talk of its specific
conduct-ance, now commonly called its conductivity. This is the conductance
as measured between the opposite faces of a 1 cm cube of the material
(Figure 1).
This measurement has units of siemens/centimeter (S/cm). The units
microsiemens/centimeter (µS/cm) and millisiemens/centimeter (mS/cm)
are most commonly used to describe the conductivity of aqueous solutions.
The corresponding terms for specific resistance (or resistivity) are
Users of ultra pure water prefer to use resistivity units of M
measurement in these units tends to spread the scale out in the range of
interest. These same users frequently use k
Ω
· cm when dealing with less
pure water such as tap water.
Ω
· cm, because
conducting surface
A
Pt
Figure 1
Conductivity Cube
B
Pt
L
L
H
A
A
W
3
Conductivity Sensor012–06485B
However when dealing with a chemical solution ranging from extremely
dilute to very concentrated chemical, use of conductivity units of µS/cm and
mS/cm are common. In these applications, conductivity has the advantage
of an almost direct relationship with impurities, especially at low
concentration. Hence, a rising conductivity reading shows increasing
impurities, or a generally increasing concentration in the case of a chemical
stream (with some exceptions in concentrated solutions). See Table 1 for a
comparison of resistance and conductivity.
Table 1. Conductivity Of Various Aqueous Solutions At 25 0C
SolutionConductivityResistivity
Pure water0.05µS/cm18MΩ · cm
Power plant boiler water0.05–1µS/cm1–18MΩ · cm
Distilled water0.5µS/cm2MΩ · cm
Deionized water0.1–10µS/cm0.1–10 MΩ · cm
Demineralized water1–80µS/cm0.01–1 MΩ · cm
Mountain water10µS/cm0.1MΩ · cm
Drinking water0.5 –1mS/cm1–2kΩ · cm
Waste water0.9–9mS/cm0.1–1kΩ · cm
KCl solution (0.0l M)1.4mS/cm0.7Ω · cm
Potable water maximum1.5mS/cm0.7Ω · cm
Brackish water1–80mS/cm0.01–1 Ω · cm
Industrial process water7–140mS/cmrarely stated
Ocean water53mS/cmrarely stated
10% NaOH355mS/cmrarely stated
100/0 H
2504
31% HNO
3
432mS/cmrarely stated
865mS/cmrarely stated
Conductivity Electrodes (Cells)
Simple conductivity sensors are constructed of an insulating material imbedded
with platinum, graphite, stainless steel or other metallic pieces. These metal
contacts serve as sensing elements and are placed at a fixed distance apart to
make contact with a solution whose conductivity is to be determined. The length
between the sensing elements, as well as the surface area of the metallic piece
determine the electrode cell constant, defined as length/area (Figure 1). The cell
constant is a critical parameter affecting the conductance value produced by the
cell and handled by the electronic circuitry.
A cell constant of 1.0 will produce a conductance reading approximately equal
to the solution conductivity. For solutions of low conductivity, the sensing
4
012–06485BConductivity Sensor
electrodes can be placed closer together, reducing the length between them
and producing cell constants of 0.1 or 0.01. This will raise the conductance
reading by a factor of 10 to 100 to offset the low solution conductivity and
give a better signal to the conductivity meter. On the other hand, the sensing
electrodes can be placed farther apart to create cell constants of 10 or 100
for use in highly conductive solutions. Table 2 lists the optimum conductivity
range for cells with different cell constants.
Table 2. Optimum Conductivity Range For Cells
With Different Cell Constants
Cell ConstantConductivity
0.010–20µS/cm
0.10–200µS/cm
1.00 –2000µS/cm
10.00–200,000 µS/cm
to amplifier box
glass rod
The conductivity electrode provided with the CI- 6729 has a cell constant of
1.0 and is designed to achieve optimum performance over a range of 0 to
20,000 µS (Figure 2). This performance is achieved by using a cylindrical
cell geometry and platinized platinum conductors embedded on a glass rod.
For measurements of conductivity greater than 20,000 µS, the 10x electrode
should be used.
Conductivity Sensor Amplifier
The Conductivity Sensor amplifier has two distinct functions. First, it provides
a signal or voltage used to drive the conductivity electrode, and second, it
senses the electrical current the electrode passes when placed in the solution
to be tested.
If the voltage and current are known, then the resistance of the cell can be
determined. If the resistance of the cell is known, then the conductivity can
be determined by taking the inverse of the resistance and multiplying by the
conductivity cell constant. While it is instructional to know how the sensor
determines conductivity, Science Workshop handles the calculations and
reports conductivity for the 1x cell directly. If the 10x conductivity electrode
is used, the value reported by Science Workshop should be multiplied by ten.
platinized
platinum
conductors
Figure 2
Schematic view of the cell for the
CI-6729 Conductivity Electrode.
E
E
+
-
When a potential is a applied to the conductivity cell, the ions in solution are
influenced by the charge on the cell’s electrodes and begin to migrate toward
the electrodes (Figure 3).
Figure 3
Conductivity Cell in operation
5
Conductivity Sensor012–06485B
+
Figure 4
Polarized Electrode
➤ Approximate conductivities
°°
(at 25
°
C) and impurity
°°
concentrations (in ppm) of
various water samples are given
below.
-
SampleConductivity
Pure H200.055 µS/cm (0.027 ppm)
Distilled H20 0.5 µS/cm (0.206 ppm)
City H2050 µS/cm (25 ppm)
Ocean H2053 mS/cm (35,000 ppm)
After a fairly short period of time most of the ions in solution will move
to the electrodes and the current flow through probe will begin to
decrease. This process is called polarization (Figure 4). Polarization
causes the conductivity probe’s output to be low if not corrected.
One method used to reduce the effects of polarization is to alternate the
polarity of the voltage applied to the electrode. Polarization of the
electrode and the associated build up of oxidation/reduction byproducts is prevented if the voltage is alternated quickly enough.
Conductivity Measurements
The conductivity of an electrolytic solution is dependent on several
factors, including the concentration of the solute, the degree of dissociation of the solute, the degree of dissociation of the molecules
present in the solution, the valiancy of the ion(s) present in the
solution, the mobilities of ions that are formed upon dissociation, and
the temperature of the solution. In general, ion mobilities remain
constant over a specific concentration range, depending on the solute
in solution. Within the specific concentration range, the conductivity
of a solution will increase in proportion to an increase in solute
concentration.
Since temperature is a measure of the average kinetic energy of an atom,
ion, or molecule, any change in temperature of a solution will strongly
affect the mobility of the species present in the electrolyte. Therefore,
the conductivity of the solution will also change. As a result, the
conductivity of a solution whose concentration is known is always
quoted for a specific temperature. Several tables in the Appendix
illustrate the temperature dependence of conductivity for specific
electrolytic solutions.
Concentration or total dissolved solid (TDS) measurements can also be
accurately determined by correlating the conductivity of a solution with
reference tables or graphs. TDS measurements are generally used to
estimate the salt content of water samples. A rule of thumb used to
estimate TDS in mg/L (ppm) is to divide the measured conductivity in
microsiemens per centimeter (µS/cm) by two. Table 4 in the Appendix
is a conversion chart for TDS measurements involving NaCl and
CaCO
Appendix.
. Additional references of conductivity data are also listed in the
3
6
012–06485BConductivity Sensor
Setup and Calibration
Set up
Science Workshop
1. Attach electrode to the amplifier box (Figure 5a).
2. Plug the 8-pin DIN connector of the amplifier box into any analog
channel on a PASCO computer interface (Figure 5b).
3. Launch Science Workshop, drag the analog plug icon to the analog
channel icon, and select Conductivity Sensor.
4. In the Experiment Setup window, drag a Graph display to the Conductivity
Sensor icon (
), and select the appropriate conductivity range to be
displayed. (The range setting selected on the amplifier box must match
the display range for the graph.)
➤➤
➤
A Digits display may also be used to display conductivity.
➤➤
Calibration
Prepare one of the weight percent NaCl solutions given in the table below
by weighing out in air the desired mass (mg) of NaCl in a 1-liter flask. Add
500 ml of deionized water and stir the solution to dissolve the salt. Add the
remaining 500 ml of deionized water and stir the solution.
RAN
SEL
20K
GE
ECT
2K
20
a
0
CONDUCTIVITY
CONDU
SENSOR
SENSOR
CTIVITY
I-6729
I-6729
C
C
I-6739A
C
b
R
S
A
E
N
2
L
G
0
E
E
K
C
2
K
T
2
0
0
CONDUCTIVITY
SENSOR
9
2
7
6
-
I
C
Figure 5
Plug the sensor into the computer
interface.
%Weight
(approx.)
0.001
0.01
0.1
1.0
10.0
➤➤
➤
For direct measurements of conductivity greater than 20,000
➤➤
Mass of
NaCl (mg)
10
100
1,000
10,000
100,000
TDS ppm
or mg/liter
10
100
1,000
10,000
100,000
Conductivity (µS/cm)
at 25 °C
21.4
210
1,990
17,600
140,000
µS the 10x electrode should be used. The 1x electrode may be
used if a 10:1 dilution of the solution to be measured is made. If
the 10:1 dilution method is used remember to multiply the value
indicated by 10.
➤➤
➤
For measurements that require very high accuracy, a KCl
➤➤
standard should be used (see Table 5 in the Appendix).
Other calibration solutions can be prepared using the tables and graphs in the
Appendix.
7
Loading...
+ 23 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.