Instruction Sheet for the PASCO Models CI-6718 and CI-6719
012-08925A
ELVIS Adapters
Introduction
The PASCO ELVIS Adapters are designed to allow the use of PASCO sensors with National Instruments Educational Laboratory Virtual Instrument Suite (NI ELVIS). LabVIEW software can be used with NI ELVIS to collect and analyze data with the sensors. The adapters are also compatible with most electronics prototyping boards.
Two different adapters are available:
Analog Adapter (CI-6718) - enables the use of a single ScienceWorkshop analog sensor with ELVIS.
Connection of Analog ScienceWorkshop Sensors
Analog ScienceWorkshop sensors have 8-pin DIN connections. Not all sensors require all connections, but connecting all pins as specified will not damage any of the ScienceWorkshop sensors.
Please see the specific connection charts on the following pages.
Using Analog Sensors
1. Plug the analog adapter into a prototyping board.
2. Connect wires from the inputs/outputs of the adapter to the desired inputs and outputs as specified in the connection chart.
Please note that the two sets of pins nearest the DIN connector are not electrically connected. They simply provide additional mechanical support for the adapter.
Digital Adapter (CI-6719) - enables the use of one or two ScienceWorkshop digital sensors with ELVIS.
ELVIS Adapters 012-08925A
3. Steady the DIN connector with one hand and plug the desired analog sensor into the adapter.
4 . Create a Virtual Instrument (VI) using
LabVIEW software to collect and analyze data with the sensor.
Using Digital Sensors
1. Plug the digital adapter into a prototyping board.
2. Connect wires from the inputs/outputs of the adapter to the desired inputs and outputs. (Reference the appropriate connection chart.)
Please note that the two sets of pins nearest the stereo jacks are not electrically connected. They simply provide additional mechanical support for the adapter.
Chart: Connection of Analog ScienceWorkshop Sensors
3 . Steady the adapter with one hand and plug the
desired digital sensor into the adapter. The adapter will accommodate:
- One Motion Sensor or Rotary Motion Sensor
- Two Photogates, Smart Pulleys, Time of Flight Pads, or Nuclear Sensors
4 . Create a Virtual Instrument (VI) using
LabVIEW software to collect and analyze data with the sensor.
Note: If you are plotting the input from an unused channel, its output will tend to follow that of the last input in the multiplex scan order unless its inputs are shorted.
ELVIS I/O Analog Adapter I/O
ACH0+ Analog channel #0 Positive Sense AIN+ Positive Sensor Analog Input ACH0- Analog channel #0 Negative Sense AIN- Negative Sensor Analog
Input Port DC Power Supplies + 5 V Power Supply Output +5 V + 5 V Sensor Power Input +5 V
DC Power Supplies Power Ground PGND Sensor Power Ground
GND
Variable Power Supplies + 12 V Power Output.* +12V +12V Sensor Power Input
SUPPLY +
Variable Power Supplies - 12 V Power Output.* -12V -12V Sensor Power Input
SUPPLY -
N/C No Connection AOUT Analog Output **
* Set the dials for the variable power supplies to "manual" and change the values to +/- 12 V. **Usually not connected.